DICOM PS3.3 2024c - Information Object Definitions |
---|
Table C.8.17.16-1 specifies the Attributes of the Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Image Module.
Table C.8.17.16-1. Ophthalmic Optical Coherence Tomography B-scan Volume Analysis Image Module Attributes
Most significant bit for pixel sample data. Shall be one less than the value in Bits Stored (0028,0101). |
|||
Specifies an identity transformation for the Presentation LUT such that the output of all grayscale transformations are defined to be in P-Values. |
|||
Specifies whether an Image has undergone lossy compression (at a point in its lifetime), or is derived from lossy compressed images. Enumerated Values: Once this value has been set to "01" it shall not be reset. See Section C.8.32.2.1 and Section C.7.6.1.1.5. |
|||
Describes the approximate lossy compression ratio(s) that have been applied to this image. Required if present in the source images or this IOD Instance has been compressed. |
|||
A label for the lossy compression method(s) that have been applied to this image. Required if present in the source images or this IOD Instance has been compressed. See Section C.8.32.2.1. |
|||
Indicates whether or not image contains sufficient burned in annotation to identify the patient and date the image was acquired. |
|||
Indicates whether or not the image contains sufficiently recognizable visual features to allow the image or a reconstruction from a set of images to identify the patient. |
|||
>Include Table 10-19 “Algorithm Identification Macro Attributes” |
For Algorithm Family Code Sequence (0066,002F) BCID 4270 “OCT-A Processing Algorithm Family”. |
||
Conveys raw data parameters captured during the B-scan acquisition process. One or more Items are permitted in this Sequence. More than one Item indicates that multiple scan patterns have been used to acquire the raw data. See Section C.8.17.16.2 for further explanation. |
|||
The number of B-scans performed at the same spatial location (B-scan slab). See Section C.8.17.16.2 for further explanation. |
|||
Nominal thickness of each B-scan slab, in mm. See Section C.8.17.16.2 for further explanation. |
|||
Nominal distance between adjacent B-scan slabs, in mm. See Section C.8.17.16.2 for further explanation. |
|||
Nominal time (in msec) between individual B-scans. See Section C.8.17.16.1.1 for further explanation. |
|||
An array that contains the real time increments (in msec) between B-scans. See Section C.8.17.16.1.1 for further explanation. |
|||
Frequency, in kHz, of the A-scan used to acquire the raw image data. |
|||
Frequency, in Hz, of the B-scan used to acquire the raw image data. |
|||
Offset of the first frame in a multi-frame image of a concatenation. |
|||
Identifier for one SOP Instance belonging to a concatenation. |
|||
The number of SOP Instances sharing the same Concatenation UID (0020,9161). See Section C.8.17.16.3 for further explanation. |
|||
The B-scan Cycle Time (0022,1645) is the nominal scan time (in milliseconds) plus the latency between the individual repeats of the B-scan at the same location. The B-scan Cycle Time Vector can be used if the time for the repeats are not uniform. Either the B-scan Cycle Time (0022,1645) or the B-scan Cycle Time Vector (0022,1646) are required.
B-scan Cycle Time shall be used in the following manner to calculate 'the relative time' for each B-scan:
When there is only one B-scan present, B-scan Cycle Time (0022,1645) may have either a value of 0, or a nominal value that would apply if there were multiple B-scans.
B-scan Cycle Time Vector (0022,1646) is an array that contains the time increments (in milliseconds) between the nth B-scan cycle and the previous B-scan cycle for a frame. The first B-scan cycle always has a time increment of 0. The B-scan Cycle Time Vector shall be used in the following manner to calculate 'relative time' T(n)for B-scan Cycle n:
DICOM PS3.3 2024c - Information Object Definitions |
---|