DICOM PS3.3 2016a - Information Object Definitions
PS3.3
Next
PS3.3
PS3.3
DICOM PS3.3 2016a - Information Object Definitions
DICOM Standards Committee
Copyright © 2016 NEMA
Table of Contents
Notice and Disclaimer
Foreword
1. Scope and Field of Application
2. Normative References
Bibliography
3. Definitions
Glossary
4. Symbols and Abbreviations
5. Conventions
5.1. Entity-Relationship Model
5.1.1. Entity
5.1.2. Relationship
5.2. Sequences
5.3. Triplet Encoding of Structured Data (Retired)
5.4. Attribute Macros
5.5. Types and Conditions in Normalized IODs
6. DICOM Information Model
6.1. Information Object Definition
6.1.1. Composite IOD
6.1.2. Normalized IOD
6.2. Attributes
6.3. On-line Communication and Media Storage Services
6.3.1. DIMSE-C Services
6.3.2. DIMSE-N Services
6.4. DIMSE Service Group
6.5. Service-Object Pair (SOP) Class
6.5.1. Normalized and Composite SOP Classes
6.6. Association Negotiation
6.7. Service Class Specification
7. DICOM Model of the Real World
7.1. DICOM Information Model
7.2. Organization of Annexes A, B and C
7.3. Extension of the DICOM Model of the Real World
7.3.1. Definition of the Extensions of the DICOM Real World Model
7.3.1.1. Patient
7.3.1.2. Service Episode and Visit
7.3.1.3. Imaging Service Request
7.3.1.4. Procedure Type
7.3.1.5. Requested Procedure
7.3.1.6. Scheduled Procedure Step
7.3.1.7. Procedure Plan
7.3.1.8. Protocol
7.3.1.9. Modality Performed Procedure Step
7.3.1.10. General Purpose Scheduled Procedure Step (Retired)
7.3.1.11. General Purpose Performed Procedure Step (Retired)
7.3.1.12. Workitem (Retired)
7.3.1.13. Clinical Document
7.4. Extension of the DICOM Model of the Real World for the General Purpose Worklist (Retired)
7.5. Organizing Large Sets of Information
7.5.1. Concatenation
7.5.2. Dimension Organization
7.6. Extension of the DICOM Model of the Real World for Clinical Trials and Research
7.6.1. Clinical Trial and Research Information Entities
7.6.1.1. Clinical Trial Sponsor
7.6.1.2. Clinical Trial Protocol
7.6.1.3. Clinical Trial Subject
7.6.1.4. Clinical Trial Site
7.6.1.5. Clinical Trial Time Point
7.6.1.6. Clinical Trial Coordinating Center
7.7. Extension of the DICOM Model of the Real World for Hanging Protocols
7.7.1. Hanging Protocol Information Entity
7.8. Extension of the DICOM Model of the Real World for Color Palettes
7.8.1. Color Palette Information Entity
7.9. Extension of the DICOM Model of the Real World for Specimens
7.9.1. Specimen
7.9.2. Container
7.9.3. Container Component
7.9.4. Preparation Step
7.10. Extension of DICOM Model of the Real World for Implant Templates
7.11. Extension of the DICOM Model of the Real World for the Unified Procedure Step (UPS)
7.11.1. Unified Procedure Step
7.11.2. Worklist
7.12. Extension of The DICOM Model of The Real World For Display System
8. Encoding of Coded Entry Data
8.1. Code Value
8.2. Coding Scheme Designator and Coding Scheme Version
8.3. Code Meaning
8.4. Mapping Resource
8.5. Context Group Version
8.6. Context Identifier and Context UID
8.7. Context Group Extensions
8.8. Standard Attribute Sets for Code Sequence Attributes
8.9. Equivalent Code Sequence
8.10. Coded Entry Data Examples
9. Template Identification Macro (Retired)
10. Miscellaneous Macros
10.1. Person Identification Macro
10.2. Content Item Macro
10.3. Image SOP Instance Reference Macro
10.4. Series and Instance Reference Macro
10.5. General Anatomy Macros
10.6. Request Attributes Macro
10.6.1. SOP Class UID in Referenced Study Sequence
10.7. Basic Pixel Spacing Calibration Macro
10.7.1. Basic Pixel Spacing Calibration Macro Attribute Descriptions
10.7.1.1. Pixel Spacing
10.7.1.2. Pixel Spacing Calibration Type
10.7.1.3. Pixel Spacing Value Order and Valid Values
10.8. SOP Instance Reference Macro
10.9. Content Identification Macro
10.10. General Contributing Sources Macro
10.11. Contributing Image Sources Macro
10.12. Patient Orientation Macro
10.12.1. Relation With Other Positioning Attributes
10.13. Performed Procedure Step Summary Macro
10.14. HL7v2 Hierarchic Designator Macro
10.15. Issuer of Patient ID Macro
10.16. Algorithm Identification Macro
10.17. Selector Attribute Macro
10.17.1. Selector Attribute Macro Attribute Descriptions
10.17.1.1. Referencing Nested Elements
10.17.1.2. Private Attribute References
10.18. Data Set Identification Macro
10.19. Exposure Index Macro
10.20. Mandatory View and Slice Progression Direction Macro
10.20.1. Mandatory View and Slice Progression Direction Macro Attributes
10.20.1.1. Slice Progression Direction
10.21. Optional View and Slice Progression Direction Macro
10.22. Numeric Value Macro
10.23. RT Equipment Correlation Macro
10.24. Device Motion Control Macro
10.24.1. Device Motion Control Macro Attributes
10.24.1.1. Device Motion Execution Mode
10.24.1.2. Device Motion Observation Mode
A. Composite Information Object Definitions (Normative)
A.1. Elements of An Information Object Definition
A.1.1. IOD Description
A.1.2. IOD Entity-Relationship Model
A.1.2.1. Patient IE
A.1.2.2. Study IE
A.1.2.3. Series IE
A.1.2.4. Equipment IE
A.1.2.5. Frame of Reference IE
A.1.2.6. Image IE
A.1.2.7. Overlay IE
A.1.2.8. Curve IE
A.1.2.9. Modality LUT IE
A.1.2.10. VOI LUT IE
A.1.2.11. Presentation State IE
A.1.2.12. Waveform IE
A.1.2.13. SR Document IE
A.1.2.14. MR Spectroscopy IE
A.1.2.15. Raw Data IE
A.1.2.16. Encapsulated Document IE
A.1.2.17. Real World Value Mapping IE
A.1.2.18. Surface IE
A.1.2.19. Measurements IE
A.1.2.20. Tractography Results IE
A.1.2.21. Plan IE
A.1.3. IOD Module Table and Functional Group Macro Table
A.1.3.1. Mandatory Modules
A.1.3.2. Conditional Modules
A.1.3.3. User Option Modules
A.1.4. Overview of the Composite IOD Module Content
A.2. Computed Radiography Image IOD
A.2.1. CR Image IOD Description
A.2.2. CR Image IOD Entity-Relationship Model
A.2.3. CR Image IOD Module Table
A.3. Computed Tomography Image IOD
A.3.1. CT Image IOD Description
A.3.2. CT Image IOD Entity-Relationship Model
A.3.3. CT Image IOD Module Table
A.4. Magnetic Resonance Image IOD
A.4.1. MR Image IOD Description
A.4.2. MR Image IOD Entity-Relationship Model
A.4.3. MR Image IOD Module Table
A.5. Nuclear Medicine Image IOD
A.5.1. NM Image IOD Description
A.5.2. NM Image IOD Entity-Relationship Model
A.5.3. NM Image IOD Module Table (Retired)
A.5.4. NM Image IOD Module Table
A.5.4.1. Acquisition Context Module
A.6. Ultrasound Image IOD
A.6.1. US Image IOD Description
A.6.2. US Image IOD Entity-Relationship Model
A.6.3. US Image IOD Module Table (Retired)
A.6.4. US Image IOD Module Table
A.6.4.1. Mutually Exclusive IEs
A.7. Ultrasound Multi-frame Image IOD
A.7.1. US Image IOD Description
A.7.2. US Multi-frame Image IOD Entity-Relationship Model
A.7.3. US Image IOD Module Table (Retired)
A.7.4. US Multi-frame Image IOD Module Table
A.7.4.1. Mutually Exclusive IEs
A.8. Secondary Capture Image IOD
A.8.1. SC Image Information Objection Definition
A.8.1.1. SC Image IOD Description
A.8.1.2. SC Image IOD Entity-Relationship Model
A.8.1.3. SC Image IOD Module Table
A.8.2. Multi-frame Single Bit Secondary Capture Image IOD
A.8.2.1. Multi-frame Single Bit SC Image IOD Description
A.8.2.2. Multi-frame Single Bit SC Image IOD Entity-Relationship Model
A.8.2.3. Multi-frame Single Bit SC Image IOD Module Table
A.8.2.4. Multi-frame Single Bit SC Image IOD Content Constraints
A.8.3. Multi-frame Grayscale Byte Secondary Capture Image IOD
A.8.3.1. Multi-frame Grayscale Byte Image IOD Description
A.8.3.2. Multi-frame Grayscale Byte SC Image IOD Entity-Relationship Model
A.8.3.3. Multi-frame Grayscale Byte SC Image IOD Module Table
A.8.3.4. Multi-frame Grayscale Byte SC Image IOD Content Constraints
A.8.3.5. Multi-frame Grayscale Byte SC Image Functional Group Macros
A.8.4. Multi-frame Grayscale Word Secondary Capture Image IOD
A.8.4.1. Multi-frame Grayscale Word SC Image IOD Description
A.8.4.2. Multi-frame Grayscale Word SC Image IOD Entity-Relationship Model
A.8.4.3. Multi-frame Grayscale Word SC Image IOD Module Table
A.8.4.4. Multi-frame Grayscale Word SC Image IOD Content Constraints
A.8.4.5. Multi-frame Grayscale Word SC Image Functional Group Macros
A.8.5. Multi-frame True Color Secondary Capture Image IOD
A.8.5.1. Multi-frame True Color Image IOD Description
A.8.5.2. Multi-frame True Color SC Image IOD Entity-Relationship Model
A.8.5.3. Multi-frame True Color SC Image IOD Module Table
A.8.5.4. Multi-frame True Color SC Image IOD Content Constraints
A.8.5.5. Multi-frame True Color SC Image Functional Group Macros
A.9. Standalone Overlay IOD
A.10. Standalone Curve IOD
A.11. Basic Study Descriptor IOD
A.12. Standalone Modality LUT IOD
A.13. Standalone VOI LUT IOD
A.14. X-Ray Angiographic Image IOD
A.14.1. XA Image IOD Description
A.14.2. XA Image IOD Entity-Relationship Model
A.14.3. XA Image IOD Module Table
A.15. X-Ray Angiographic Bi-plane Image Information Object Definition (Retired)
A.16. X-Ray RF Image IOD
A.16.1. XRF Image IOD Description
A.16.2. XRF Image IOD Entity-Relationship Model
A.16.3. XRF Image IOD Module Table
A.17. RT Image IOD
A.17.1. RT Image IOD Description
A.17.2. RT Image IOD Entity-Relationship Model
A.17.3. RT Image IOD Module Table
A.18. RT Dose IOD
A.18.1. RT Dose IOD Description
A.18.2. RT Dose IOD Entity-Relationship Model
A.18.3. RT Dose IOD Module Table
A.19. RT Structure Set IOD
A.19.1. RT Structure Set IOD Description
A.19.2. RT Structure Set IOD Entity-Relationship Model
A.19.3. RT Structure Set IOD Module Table
A.20. RT Plan IOD
A.20.1. RT Plan IOD Description
A.20.2. RT Plan IOD Entity-Relationship Model
A.20.3. RT Plan IOD Module Table
A.20.3.1. RT Fraction Scheme Module
A.20.3.2. RT Prescription Module
A.20.3.3. RT Tolerance Tables Module
A.20.3.4. RT Patient Setup Module
A.21. Positron Emission Tomography Image IOD
A.21.1. PET Image IOD Description
A.21.2. PET Image IOD Entity-Relationship Model
A.21.3. PET Image IOD Module Table
A.21.3.1. Acquisition Context Module
A.22. Standalone PET Curve IOD
A.23. Stored Print IOD
A.24. Hardcopy Grayscale Image IOD
A.25. Hardcopy Color Image IOD
A.26. Digital X-Ray Image IOD
A.26.1. DX Image IOD Description
A.26.2. DX Image IOD Entity-Relationship Model
A.26.3. DX Image IOD Module Table
A.26.4. Overlay Plane Module
A.26.5. Acquisition Context Module
A.27. Digital Mammography X-Ray Image IOD
A.27.1. Digital Mammography X-Ray Image IOD Description
A.27.2. Digital Mammography X-Ray Image IOD Module Table
A.27.3. Overlay Plane Module
A.28. Digital Intra-Oral X-Ray Image IOD
A.28.1. Digital Intra-Oral X-Ray Image IOD Description
A.28.2. Digital Intra-Oral X-Ray Image IOD Module Table
A.28.3. Overlay Plane Module
A.29. RT Beams Treatment Record IOD
A.29.1. RT Beams Treatment Record IOD Description
A.29.2. RT Beams Treatment Record IOD Entity-Relationship Model
A.29.3. RT Beams Treatment Record IOD Module Table
A.30. RT Brachy Treatment Record IOD
A.30.1. RT Brachy Treatment Record IOD Description
A.30.2. RT Brachy Treatment Record IOD Entity-Relationship Model
A.30.3. RT Brachy Treatment Record IOD Module Table
A.31. RT Treatment Summary Record IOD
A.31.1. RT Treatment Summary Record IOD Description
A.31.2. RT Treatment Summary Record IOD Entity-Relationship Model
A.31.3. RT Treatment Summary Record IOD Module Table
A.32. Visible Light Image Information Object Definitions
A.32.1. VL Endoscopic Image IOD
A.32.1.1. VL Endoscopic Image IOD Description
A.32.1.2. VL Endoscopic Image IOD Entity-Relationship Model
A.32.1.3. VL Endoscopic Image IOD Content Constraints
A.32.1.3.1. Modality
A.32.2. VL Microscopic Image IOD
A.32.2.1. VL Microscopic Image IOD Description
A.32.2.2. VL Microscopic Image IOD Entity-Relationship Model
A.32.2.3. VL Microscopic Image IOD Content Constraints
A.32.2.3.1. Modality
A.32.3. VL Slide-coordinates Microscopic Image IOD
A.32.3.1. VL Slide-coordinates Microscopic Image IOD Description
A.32.3.2. VL Slide-coordinates Microscopic Image IOD Entity-Relationship Model
A.32.3.3. VL Slide-coordinates Microscopic Image IOD Content Constraints
A.32.3.3.1. Modality
A.32.4. VL Photographic Image IOD
A.32.4.1. VL Photographic Image IOD Description
A.32.4.2. VL Photographic Image IOD Entity-Relationship Model
A.32.4.3. VL Photographic Image IOD Content Constraints
A.32.4.3.1. Modality
A.32.5. Video Endoscopic Image IOD
A.32.5.1. Video Endoscopic Image IOD Description
A.32.5.2. Video Endoscopic Image IOD Entity-Relationship Model
A.32.5.3. Video Endoscopic Image IOD Content Constraints
A.32.5.3.1. Modality
A.32.5.3.2. Image Related Data Encoding
A.32.5.3.3. Anatomic Region Sequence
A.32.6. Video Microscopic Image IOD
A.32.6.1. Video Microscopic Image IOD Description
A.32.6.2. Video Microscopic Image IOD Entity-Relationship Model
A.32.6.3. Video Microscopic Image IOD Content Constraints
A.32.6.3.1. Modality
A.32.6.3.2. Image Related Data Encoding
A.32.7. Video Photographic Image IOD
A.32.7.1. Video Photographic Image IOD Description
A.32.7.2. Video Photographic Image IOD Entity-Relationship Model
A.32.7.3. Video Photographic Image IOD Content Constraints
A.32.7.3.1. Modality
A.32.7.3.2. Image Related Data Encoding
A.32.8. VL Whole Slide Microscopy Image IOD
A.32.8.1. VL Whole Slide Microscopy Image IOD Description
A.32.8.2. VL Whole Slide Microscopy Image IOD Entity-Relationship Model
A.32.8.3. VL Whole Slide Microscopy Image IOD Module Table
A.32.8.3.1. VL Whole Slide Microscopy Image IOD Content Constraints
A.32.8.3.1.1. Optical Path Module
A.32.8.4. VL Whole Slide Microscopy Image Functional Group Macros
A.32.8.4.1. VL Whole Slide Microscopy Image Functional Group Macros Content Constraints
A.32.8.4.1.1. Referenced Image
A.32.8.4.1.2. Plane Position (Slide)
A.33. Softcopy Presentation State Information Object Definitions
A.33.1. Grayscale Softcopy Presentation State IOD
A.33.1.1. Grayscale Softcopy Presentation State IOD Description
A.33.1.2. Grayscale Softcopy Presentation State IOD Module Table
A.33.2. Color Softcopy Presentation State IOD
A.33.2.1. Color Softcopy Presentation State IOD Description
A.33.2.2. Color Softcopy Presentation State IOD Module Table
A.33.3. Pseudo-color Softcopy Presentation State IOD
A.33.3.1. Pseudo-color Softcopy Presentation State IOD Description
A.33.3.2. Pseudo-color Softcopy Presentation State IOD Module Table
A.33.4. Blending Softcopy Presentation State IOD
A.33.4.1. Blending Softcopy Presentation State IOD Description
A.33.4.2. Blending Softcopy Presentation State IOD Module Table
A.33.5. Basic Structured Display IOD
A.33.5.1. Basic Structured Display IOD Description
A.33.6. XA/XRF Grayscale Softcopy Presentation State IOD
A.33.6.1. XA/XRF Grayscale Softcopy Presentation State IOD Description
A.33.6.2. XA/XRF Grayscale Softcopy Presentation State IOD Module Table
A.34. Waveform Information Object Definitions
A.34.1. Waveform IOD Entity-Relationship Model
A.34.2. Basic Voice Audio IOD
A.34.2.1. Basic Voice Audio IOD Description
A.34.2.2. Basic Voice Audio IOD Entity-Relationship Model
A.34.2.3. Basic Voice Audio IOD Module Table
A.34.2.4. Basic Voice Audio IOD Content Constraints
A.34.2.4.1. Modality
A.34.2.4.2. Waveform Sequence
A.34.2.4.3. Number of Waveform Channels
A.34.2.4.4. Sampling Frequency
A.34.2.4.5. Waveform Sample Interpretation
A.34.3. 12-Lead Electrocardiogram IOD
A.34.3.1. 12-Lead ECG IOD Description
A.34.3.2. 12-Lead ECG IOD Entity-Relationship Model
A.34.3.3. 12-Lead ECG IOD Module Table
A.34.3.4. 12-Lead ECG IOD Content Constraints
A.34.3.4.1. Modality
A.34.3.4.2. Acquisition Context Module
A.34.3.4.3. Waveform Sequence
A.34.3.4.4. Number of Waveform Channels
A.34.3.4.5. Number of Waveform Samples
A.34.3.4.6. Sampling Frequency
A.34.3.4.7. Channel Source
A.34.3.4.8. Waveform Sample Interpretation
A.34.3.4.9. Waveform Annotation Module
A.34.4. General Electrocardiogram IOD
A.34.4.1. General ECG IOD Description
A.34.4.2. General ECG IOD Entity-Relationship Model
A.34.4.3. General ECG IOD Module Table
A.34.4.4. General ECG IOD Content Constraints
A.34.4.4.1. Modality
A.34.4.4.2. Waveform Sequence
A.34.4.4.3. Number of Waveform Channels
A.34.4.4.4. Sampling Frequency
A.34.4.4.5. Channel Source
A.34.4.4.6. Waveform Sample Interpretation
A.34.4.4.7. Waveform Annotation Module
A.34.5. Ambulatory Electrocardiogram IOD
A.34.5.1. Ambulatory ECG IOD Description
A.34.5.2. Ambulatory ECG IOD Entity-Relationship Model
A.34.5.3. Ambulatory ECG IOD Module Table
A.34.5.4. Ambulatory ECG IOD Content Constraints
A.34.5.4.1. Modality
A.34.5.4.2. Waveform Sequence
A.34.5.4.3. Number of Waveform Channels
A.34.5.4.5. Sampling Frequency
A.34.5.4.6. Channel Source
A.34.5.4.7. Waveform Sample Interpretation
A.34.6. Hemodynamic IOD
A.34.6.1. Hemodynamic IOD Description
A.34.6.2. Hemodynamic IOD Entity-Relationship Model
A.34.6.3. Hemodynamic IOD Module Table
A.34.6.4. Hemodynamic IOD Content Constraints
A.34.6.4.1. Modality
A.34.6.4.2. Acquisition Context Module
A.34.6.4.3. Waveform Sequence
A.34.6.4.4. Number of Waveform Channels
A.34.6.4.5. Sampling Frequency
A.34.6.4.7. Channel Source
A.34.6.4.8. Waveform Sample Interpretation
A.34.6.4.9. Waveform Annotation Module
A.34.7. Basic Cardiac Electrophysiology IOD
A.34.7.1. Basic Cardiac EP IOD Description
A.34.7.2. Basic Cardiac EP IOD Entity-Relationship Model
A.34.7.3. Basic Cardiac EP IOD Module Table
A.34.7.4. Basic Cardiac EP IOD Content Constraints
A.34.7.4.1. Modality
A.34.7.4.2. Acquisition Context Module
A.34.7.4.3. Waveform Sequence
A.34.7.4.4. Sampling Frequency
A.34.7.4.5. Channel Source
A.34.7.4.6. Waveform Sample Interpretation
A.34.7.4.7. Waveform Annotation Module
A.34.8. Arterial Pulse Waveform IOD
A.34.8.1. Arterial Pulse Waveform IOD Description
A.34.8.2. Arterial Pulse Waveform IOD Entity-Relationship Model
A.34.8.3. Arterial Pulse Waveform IOD Module Table
A.34.8.4. Arterial Pulse Waveform IOD Content Constraints
A.34.8.4.1. Modality
A.34.8.4.2. Waveform Sequence
A.34.8.4.3. Number of Waveform Channels
A.34.8.4.4. Sampling Frequency
A.34.8.4.5. Channel Source
A.34.8.4.6. Waveform Sample Interpretation
A.34.9. Respiratory Waveform IOD
A.34.9.1. Respiratory Waveform IOD Description
A.34.9.2. Respiratory Waveform IOD Entity-Relationship Model
A.34.9.3. Respiratory Waveform IOD Module Table
A.34.9.4. Respiratory Waveform IOD Content Constraints
A.34.9.4.1. Modality
A.34.9.4.2. Waveform Sequence
A.34.9.4.3. Number of Waveform Channels
A.34.9.4.4. Sampling Frequency
A.34.9.4.5. Channel Source
A.34.9.4.6. Waveform Sample Interpretation
A.34.10. General Audio Waveform IOD
A.34.10.1. General Audio Waveform IOD Description
A.34.10.2. General Audio Waveform IOD Entity-Relationship Model
A.34.10.3. General Audio Waveform IOD Module Table
A.34.10.4. General Audio Waveform IOD Content Constraints
A.34.10.4.1. Modality
A.34.10.4.2. Waveform Sequence
A.34.10.4.3. Number of Waveform Channels
A.34.10.4.4. Sampling Frequency
A.34.10.4.5. Channel Source
A.34.10.4.6. Waveform Sample Interpretation
A.35. Structured Report Document Information Object Definitions
A.35.1. Basic Text SR IOD
A.35.1.1. Basic Text SR IOD Description
A.35.1.2. Basic Text SR IOD Entity-Relationship Model
A.35.1.3. Basic Text SR IOD Module Table
A.35.1.3.1. Basic Text SR IOD Content Constraints
A.35.1.3.1.1. Value Type
A.35.1.3.1.2. Relationship Constraints
A.35.2. Enhanced SR IOD
A.35.2.1. Enhanced SR IOD Description
A.35.2.2. Enhanced SR IOD Entity-Relationship Model
A.35.2.3. Enhanced SR IOD Module Table
A.35.2.3.1. Enhanced SR IOD Content Constraints
A.35.2.3.1.1. Value Type
A.35.2.3.1.2. Relationship Constraints
A.35.3. Comprehensive SR IOD
A.35.3.1. Comprehensive SR IOD Description
A.35.3.2. Comprehensive SR IOD Entity-Relationship Model
A.35.3.3. Comprehensive SR IOD Module Table
A.35.3.3.1. Comprehensive SR IOD Content Constraints
A.35.3.3.1.1. Value Type
A.35.3.3.1.2. Relationship Constraints
A.35.4. Key Object Selection Document IOD
A.35.4.1. Key Object Selection Document IOD Description
A.35.4.2. Key Object Selection Document IOD Entity-Relationship Model
A.35.4.3. Key Object Selection Document IOD Module Table
A.35.4.3.1. Key Object Selection Document IOD Content Constraints
A.35.4.3.1.1. Value Type
A.35.4.3.1.2. Relationship Constraints
A.35.4.3.1.3. Template Constraints
A.35.5. Mammography CAD SR IOD
A.35.5.1. Mammography CAD SR IOD Description
A.35.5.2. Mammography CAD SR IOD Entity-Relationship Model
A.35.5.3. Mammography CAD SR IOD Module Table
A.35.5.3.1. Mammography CAD SR IOD Content Constraints
A.35.5.3.1.1. Template Constraints
A.35.5.3.1.2. Value Type
A.35.5.3.1.3. Relationship Constraints
A.35.6. Chest CAD SR IOD
A.35.6.1. Chest CAD SR IOD Description
A.35.6.2. Chest CAD SR IOD Entity-Relationship Model
A.35.6.3. Chest CAD SR IOD Module Table
A.35.6.3.1. Chest CAD SR IOD Content Constraints
A.35.6.3.1.1. Template Constraints
A.35.6.3.1.2. Value Type
A.35.6.3.1.3. Relationship Constraints
A.35.7. Procedure Log IOD
A.35.7.1. Procedure Log IOD Description
A.35.7.2. Procedure Log IOD Entity-Relationship Model
A.35.7.3. Procedure Log IOD Module Table
A.35.7.3.1. Procedure Log IOD Content Constraints
A.35.7.3.1.1. Template
A.35.7.3.1.2. Observation DateTime
A.35.7.3.1.3. Value Type
A.35.7.3.1.4. Relationship Constraints
A.35.8. X-Ray Radiation Dose SR IOD
A.35.8.1. X-Ray Radiation Dose SR IOD Description
A.35.8.2. X-Ray Radiation Dose SR IOD Entity-Relationship Model
A.35.8.3. X-Ray Radiation Dose SR IOD Module Table
A.35.8.3.1. X-Ray Radiation Dose SR IOD Content Constraints
A.35.8.3.1.1. Template
A.35.8.3.1.2. Value Type
A.35.8.3.1.3. Relationship Constraints
A.35.8.3.1.4. Completion Flag
A.35.9. Spectacle Prescription Report IOD
A.35.9.1. Spectacle Prescription Report IOD Description
A.35.9.2. Spectacle Prescription Report IOD Entity-Relationship Model
A.35.9.3. Spectacle Prescription Report IOD Module Table
A.35.9.3.1. Spectacle Prescription Report IOD Content Constraints
A.35.9.3.1.1. Value Type
A.35.9.3.1.2. Relationship Constraints
A.35.9.3.1.3. Template Constraints
A.35.10. Colon CAD SR IOD
A.35.10.1. Colon CAD SR IOD Description
A.35.10.2. Colon CAD SR IOD Entity-Relationship Model
A.35.10.3. Colon CAD SR IOD Module Table
A.35.10.3.1. Colon CAD SR IOD Content Constraints
A.35.10.3.1.1. Template Constraints
A.35.10.3.1.2. Value Type
A.35.10.3.1.3. Relationship Constraints
A.35.11. Macular Grid Thickness and Volume Report IOD
A.35.11.1. Macular Grid Thickness and Volume Report IOD Description
A.35.11.2. Macular Grid Thickness and Volume Report IOD Entity-Relationship Model
A.35.11.3. Macular Grid Thickness and Volume Report IOD Module Table
A.35.11.3.1. Macular Grid Thickness and Volume Report IOD Content Constraints
A.35.11.3.1.1. Value Type
A.35.11.3.1.2. Relationship Constraints
A.35.11.3.1.3. Template Constraints
A.35.12. Implantation Plan SR Document IOD
A.35.12.1. Implantation Plan SR Document IOD Description
A.35.12.2. Implantation Plan SR Document IOD Entity-Relationship Model
A.35.12.3. Implantation Plan SR Document IOD Module Table
A.35.12.3.1. Implantation Plan SR Document IOD Content Constraints
A.35.12.3.1.1. Template Constraints
A.35.12.3.1.2. Value Type
A.35.12.3.1.3. Relationship Constraints
A.35.13. Comprehensive 3D SR IOD
A.35.13.1. Comprehensive 3D SR IOD Description
A.35.13.2. Comprehensive 3D SR IOD Entity-Relationship Model
A.35.13.3. Comprehensive 3D SR IOD Module Table
A.35.13.3.1. Comprehensive 3D SR IOD Content Constraints
A.35.13.3.1.1. Value Type
A.35.13.3.1.2. Relationship Constraints
A.35.14. Radiopharmaceutical Radiation Dose SR IOD
A.35.14.1. Radiopharmaceutical Radiation Dose SR IOD Description
A.35.14.2. Radiopharmaceutical Radiation Dose SR IOD Entity-relationship Model
A.35.14.3. Radiopharmaceutical Radiation Dose SR IOD Module Table
A.35.14.3.1. Radiopharmaceutical Radiation Dose SR IOD Content Constraints
A.35.14.3.1.1. Template
A.35.14.3.1.2. Value Type
A.35.14.3.1.3. Relationship Constraints
A.35.15. Extensible SR IOD
A.35.15.1. Extensible SR IOD Description
A.35.15.2. Extensible SR IOD Entity-Relationship Model
A.35.15.3. Extensible SR IOD Module Table
A.35.15.3.1. Extensible SR IOD Content Constraints
A.35.15.3.1.1. Value Type
A.35.15.3.1.2. Relationship Constraints
A.35.16. Acquisition Context SR IOD
A.35.16.1. Acquisition Context SR IOD Description
A.35.16.2. Acquisition Context SR IOD Entity-Relationship Model
A.35.16.3. Acquisition Context SR IOD Module Table
A.35.16.3.1. Acquisition Context SR IOD Content Constraints
A.35.16.3.1.1. Value Type
A.35.16.3.1.2. Relationship Constraints
A.35.16.3.1.3. Template Constraints
A.36. Enhanced MR Information Object Definitions
A.36.1. Relationship Between Enhanced MR IODs
A.36.2. Enhanced MR Image IOD
A.36.2.1. Enhanced MR Image IOD Description
A.36.2.2. Enhanced MR Image Entity-Relationship Model
A.36.2.3. Enhanced MR Image IOD Module Table
A.36.2.3.1. Enhanced MR Image IOD Content Constraints
A.36.2.4. Enhanced MR Image Functional Group Macros
A.36.3. MR Spectroscopy IOD
A.36.3.1. MR Spectroscopy IOD Description
A.36.3.2. MR Spectroscopy Entity-Relationship Model
A.36.3.3. MR Spectroscopy IOD Module Table
A.36.3.4. MR Spectroscopy Functional Group Macros
A.36.4. Enhanced MR Color Image IOD
A.36.4.1. Enhanced MR Color Image IOD Description
A.36.4.2. Enhanced MR Color Image Entity-Relationship Model
A.36.4.3. Enhanced MR Color Image IOD Module Table
A.36.4.3.1. Enhanced MR Color Image IOD Content Constraints
A.36.4.4. Enhanced MR Color Image Functional Group Macros
A.37. Raw Data IOD
A.37.1. Raw Data IOD Description
A.37.2. Raw Data Entity-Relationship Model
A.37.3. Raw Data IOD Module Table
A.38. Enhanced Computed Tomography Image IOD
A.38.1. Enhanced CT Image IOD
A.38.1.1. Enhanced CT Image IOD Description
A.38.1.2. Enhanced CT Image IOD Entity-Relationship Model
A.38.1.3. Enhanced CT Image IOD Module Table
A.38.1.3.1. Enhanced CT Image IOD Content Constraints
A.38.1.4. Enhanced CT Image Functional Group Macros
A.39. Spatial Registration Information Object Definitions
A.39.1. Spatial Registration IOD
A.39.1.1. Spatial Registration IOD Description
A.39.1.2. Spatial Registration IOD Entity-Relationship Model
A.39.1.3. Spatial Registration IOD Module Table
A.39.2. Deformable Spatial Registration IOD
A.39.2.1. Deformable Spatial Registration IOD Description
A.39.2.2. Deformable Spatial Registration IOD Entity-Relationship Model
A.39.2.3. Deformable Spatial Registration IOD Module Table
A.40. Spatial Fiducials IOD
A.40.1. Spatial Fiducials IOD Description
A.40.2. Spatial Fiducials IOD Entity-Relationship Model
A.40.3. Spatial Fiducials IOD Module Table
A.41. Ophthalmic Photography 8 Bit Image IOD
A.41.1. Ophthalmic Photography 8 Bit Image IOD Description
A.41.2. Ophthalmic Photography 8 Bit Image IOD Entity-Relationship Model
A.41.3. Ophthalmic Photography 8 Bit Image IOD Modules
A.41.4. Ophthalmic Photography 8 Bit Image IOD Content Constraints
A.41.4.1. Bits Allocated, Bits Stored, and High Bit
A.41.4.2. Contrast/Bolus Agent Sequence
A.42. Ophthalmic Photography 16 Bit Image IOD
A.42.1. Ophthalmic Photography 16 Bit Image IOD Description
A.42.2. Ophthalmic Photography 16 Bit Image IOD Entity-Relationship Model
A.42.3. Ophthalmic Photography 16 Bit Image IOD Modules
A.42.4. Ophthalmic Photography 16 Bit Image IOD Content Constraints
A.42.4.1. Bits Allocated, Bits Stored, and High Bit
A.42.4.2. Contrast/Bolus Agent Sequence
A.43. Stereometric Relationship IOD
A.43.1. Stereometric Relationship IOD Entity-Relationship Model
A.43.2. Stereometric Relationship IOD Modules
A.44. Hanging Protocol IOD
A.44.1. Hanging Protocol IOD Description
A.44.2. Hanging Protocol IOD Entity-Relationship Model
A.44.3. Hanging Protocol IOD Module Table
A.45. Encapsulated Document IOD
A.45.1. Encapsulated PDF IOD
A.45.1.1. Encapsulated PDF IOD Description
A.45.1.2. Encapsulated PDF Entity-Relationship Model
A.45.1.3. Encapsulated PDF IOD Module Table
A.45.1.4. Encapsulated PDF IOD Content Constraints
A.45.1.4.1. MIME Type of Encapsulated Document
A.45.2. Encapsulated CDA IOD
A.45.2.1. Encapsulated CDA IOD Description
A.45.2.2. Encapsulated CDA Entity-Relationship Model
A.45.2.3. Encapsulated CDA IOD Module Table
A.45.2.4. Encapsulated CDA IOD Content Constraints
A.46. Real World Value Mapping IOD
A.46.1. Real World Value Mapping IOD Entity-Relationship Model
A.46.2. Real World Value Mapping IOD Modules
A.47. Enhanced X-Ray Angiographic Image IOD
A.47.1. Enhanced XA Image IOD Description
A.47.2. Enhanced XA Image IOD Entity-Relationship Model
A.47.3. Enhanced XA Image IOD Module Table
A.47.3.1. Enhanced XA Image IOD Content Constraints
A.47.3.1.1. Modality Type Attribute
A.47.3.1.2. Overlay Plane, Curve, VOI LUT and Specimen Identification Modules
A.47.3.1.3. Positioner Type
A.47.4. Enhanced XA Image Functional Group Macros
A.47.4.1. Enhanced XA Image Functional Group Macros Content Constraints
A.47.4.1.1. Frame Anatomy Functional Group Macro
A.48. Enhanced X-Ray RF Image IOD
A.48.1. Enhanced XRF Image IOD Description
A.48.2. Enhanced XRF Image IOD Entity-Relationship Model
A.48.3. Enhanced XRF Image IOD Module Table
A.48.3.1. Enhanced XRF Image IOD Content Constraints
A.48.3.1.1. Modality Type Attribute
A.48.3.1.2. Overlay Plane, Curve, VOI LUT and Specimen Identification Modules
A.48.3.1.3. Positioner Type
A.48.4. Enhanced XRF Image Functional Group Macros
A.48.4.1. Enhanced XRF Image Functional Group Macros Content Constraints
A.48.4.1.1. Frame Anatomy Functional Group Macro
A.49. RT Ion Plan IOD
A.49.1. IOD Description
A.49.2. IOD Modules
A.50. RT Ion Beams Treatment Record IOD
A.50.1. IOD Description
A.50.2. IOD Modules
A.51. Segmentation IOD
A.51.1. Segmentation IOD Description
A.51.2. Segmentation IOD Entity-Relationship Model
A.51.3. Segmentation IOD Module Table
A.51.4. Segmentation IOD Content Constraints
A.51.5. Segmentation Functional Groups
A.51.5.1. Segmentation Functional Groups Description
A.52. Ophthalmic Tomography Image IOD
A.52.1. Ophthalmic Tomography Image IOD Description
A.52.2. Ophthalmic Tomography Image IOD Entity-Relationship Model
A.52.3. Ophthalmic Tomography Image IOD Modules
A.52.4. Ophthalmic Tomography Image IOD Content Constraints
A.52.4.1. Contrast/Bolus Agent Sequence
A.52.4.2. Overlay Plane Module and VOI LUT Module
A.52.4.3. Ophthalmic Tomography Image Functional Group Macros
A.53. X-Ray 3D Angiographic Image IOD
A.53.1. X-Ray 3D Angiographic Image IOD Description
A.53.2. X-Ray 3D Angiographic Image IOD Entity-Relationship Model
A.53.3. X-Ray 3D Angiographic Image IOD Image Module Table
A.53.3.1. X-Ray 3D Angiographic Image IOD Content Constraints
A.53.3.1.1. Modality Type Attribute
A.53.3.1.2. Restrictions for Standard Extended SOP Classes
A.53.3.1.3. Image - Equipment Coordinate Relationship Module
A.53.4. X-Ray 3D Angiographic Image Functional Group Macros
A.53.4.1. X-Ray 3D Angiographic Image Functional Group Macros Content Constraints
A.53.4.1.1. Frame Anatomy Functional Group Macro
A.54. X-Ray 3D Craniofacial Image IOD
A.54.1. X-Ray 3D Craniofacial Image IOD Description
A.54.2. X-Ray 3D Craniofacial Image IOD Entity-Relationship Model
A.54.3. X-Ray 3D Craniofacial Image IOD Module Table
A.54.3.1. X-Ray 3D Craniofacial Image IOD Content Constraints
A.54.3.1.1. Modality Type Attribute
A.54.3.1.2. Restrictions for Standard Extended SOP Classes
A.54.4. X-Ray 3D Craniofacial Image Functional Group Macros
A.54.4.1. X-Ray 3D Craniofacial Image Functional Group Macros Content Constraints
A.54.4.1.1. Frame Anatomy Functional Group Macro
A.55. Breast Tomosynthesis Image IOD
A.55.1. Breast Tomosynthesis Image IOD Description
A.55.2. Breast Tomosynthesis Image IOD Entity-Relationship Model
A.55.3. Breast Tomosynthesis Image IOD Module Table
A.55.3.1. Breast Tomosynthesis Image IOD Content Constraints
A.55.3.1.1. Restrictions for Standard Extended SOP Classes
A.55.3.1.2. Image - Equipment Coordinate Relationship Module
A.55.4. Breast Tomosynthesis Image Functional Group Macros
A.55.4.1. Breast Tomosynthesis Image Functional Group Macros Content Constraints
A.55.4.1.1. Frame Anatomy Functional Group Macro
A.56. Enhanced PET Image IOD
A.56.1. Enhanced PET Image IOD Description
A.56.2. Enhanced PET Image IOD Entity-Relationship Model
A.56.3. Enhanced PET Image IOD Module Table
A.56.3.1. Enhanced PET Image IOD Content Constraints
A.56.4. Enhanced PET Image Functional Group Macros
A.56.5. Acquisition Context Module
A.57. Surface Segmentation IOD
A.57.1. Surface Segmentation IOD Description
A.57.2. Surface Segmentation IOD Entity-Relationship Model
A.57.3. Surface Segmentation IOD Module Table
A.58. Color Palette IOD
A.58.1. Color Palette IOD Description
A.58.2. Color Palette IOD Entity-Relationship Model
A.58.3. Color Palette IOD Module Table
A.59. Enhanced US Volume IOD
A.59.1. Enhanced US Volume IOD Description
A.59.2. Enhanced US Volume IOD Entity-Relationship Model
A.59.3. Enhanced US Volume IOD Module Table
A.59.3.1. Enhanced US Volume IOD Content Constraints
A.59.3.1.1. Associated Physiological Waveforms
A.59.3.1.2. Contrast
A.59.4. Enhanced US Volume Functional Group Macros
A.59.4.1. Enhanced US Volume Functional Group Macros Content Constraints
A.59.4.1.1. US Image Description Macro
A.59.4.1.2. Plane Position (Volume) and Plane Orientation (Volume) Macros
A.60. Ophthalmic Refractive Measurements Information Object Definitions
A.60.1. Lensometry Measurements IOD
A.60.1.1. Lensometry Measurements IOD Description
A.60.1.2. Lensometry Measurements IOD Entity-Relationship Model
A.60.1.3. Lensometry Measurements IOD Module Table
A.60.2. Autorefraction Measurements IOD
A.60.2.1. Autorefraction Measurements IOD Description
A.60.2.2. Autorefraction Measurements IOD Entity-Relationship Model
A.60.2.3. Autorefraction Measurements IOD Module Table
A.60.3. Keratometry Measurements IOD
A.60.3.1. Keratometry Measurements IOD Description
A.60.3.2. Keratometry Measurements IOD Entity-Relationship Model
A.60.3.3. Keratometry Measurements IOD Module Table
A.60.4. Subjective Refraction Measurements IOD
A.60.4.1. Subjective Refraction Measurements IOD Description
A.60.4.2. Subjective Refraction Measurements IOD Entity-Relationship Model
A.60.4.3. Subjective Refraction Measurements IOD Module Table
A.60.5. Visual Acuity Measurements IOD
A.60.5.1. Visual Acuity Measurements IOD Description
A.60.5.2. Visual Acuity Measurements IOD Entity-Relationship Model
A.60.5.3. Visual Acuity Measurements IOD Module Table
A.60.6. Ophthalmic Axial Measurements IOD
A.60.6.1. Ophthalmic Axial Measurements IOD Description
A.60.6.2. Ophthalmic Axial Measurements IOD Entity-Relationship Model
A.60.6.3. Ophthalmic Axial Measurements IOD Module Table
A.60.7. Intraocular Lens Calculations IOD
A.60.7.1. Intraocular Lens Calculations IOD Description
A.60.7.2. Intraocular Lens Calculations IOD Entity-Relationship Model
A.60.7.3. Intraocular Lens Calculations IOD Module Table
A.61. Generic Implant Template IOD
A.61.1. Generic Implant Template IOD Description
A.61.2. Generic Implant Template IOD Entity-relationship
A.61.3. Generic Implant Module IOD Module Table
A.62. Implant Assembly Template IOD
A.62.1. Implant Assembly Template IOD Description
A.62.2. Implant Assembly Template IOD Entity Relationship
A.62.3. Implant Assembly Template IOD Module Table
A.63. Implant Template Group IOD
A.63.1. Implant Template Group IOD Description
A.63.2. Implant Template Group IOD Entity Relationship
A.63.3. Implant Template Group IOD Module Table
A.64. RT Beams Delivery Instruction IOD
A.64.1. RT Beams Delivery Instruction IOD Description
A.64.2. RT Beams Delivery Instruction IOD Entity-Relationship Model
A.64.3. RT Beams Delivery Instruction IOD Module Table
A.64.4. RT Beams Delivery Instruction IOD Content Constraints
A.64.4.1. Modality
A.65. Ophthalmic Visual Field Static Perimetry Measurements IOD
A.65.1. Ophthalmic Visual Field Static Perimetry Measurements IOD Description
A.65.2. Ophthalmic Visual Field Static Perimetry Measurements IOD Entity-Relationship Model
A.65.3. Ophthalmic Visual Field Static Perimetry Measurements IOD Modules
A.66. Intravascular OCT IOD
A.66.1. Intravascular OCT Image IOD Description
A.66.2. Intravascular OCT Image IOD Entity-Relationship Model
A.66.3. Intravascular OCT Image IOD Modules
A.66.3.1. Intravascular OCT Image IOD Content Constraints
A.66.3.1.1. Contrast/Bolus Agent Sequence
A.66.3.1.2. Prohibited Modules
A.66.4. Intravascular OCT Image Functional Group Macros
A.66.4.1. Intravascular OCT Image Functional Group Macros Content Constraints
A.66.4.1.1. Frame Anatomy Functional Group Macro
A.67. Ophthalmic Thickness Map IOD
A.67.1. Ophthalmic Thickness Map IOD Description
A.67.2. Ophthalmic Thickness Map IOD Entity-Relationship Model
A.67.3. Ophthalmic Thickness Map IOD Modules
A.67.4. Ophthalmic Thickness Map IOD Content Constraints
A.67.4.1. Prohibited Modules
A.68. Surface Scan Mesh IOD
A.68.1. Surface Scan Mesh IOD Description
A.68.2. Surface Scan Mesh IOD Entity-Relationship Model
A.68.3. Surface Scan Mesh IOD Module Table
A.69. Surface Scan Point Cloud IOD
A.69.1. Surface Scan Point Cloud IOD Description
A.69.2. Surface Scan Point Cloud IOD Entity Relationship Model
A.69.3. Surface Scan Point Cloud IOD Module Table
A.70. Legacy Converted Enhanced CT Image IOD
A.70.1. Legacy Converted Enhanced CT Image IOD Description
A.70.2. Legacy Converted Enhanced CT Image IOD Entity-Relationship Model
A.70.3. Legacy Converted Enhanced CT Image IOD Module Table
A.70.3.1. Legacy Converted Enhanced CT Image IOD Content Constraints
A.70.4. Legacy Converted Enhanced CT Image Functional Group Macros
A.71. Legacy Converted Enhanced MR Image IOD
A.71.1. Legacy Converted Enhanced MR Image IOD Description
A.71.2. Legacy Converted Enhanced MR Image IOD Entity-Relationship Model
A.71.3. Legacy Converted Enhanced MR Image IOD Module Table
A.71.3.1. Legacy Converted Enhanced MR Image IOD Content Constraints
A.71.4. Legacy Converted Enhanced MR Image Functional Group Macros
A.72. Legacy Converted Enhanced PET Image IOD
A.72.1. Legacy Converted Enhanced PET Image IOD Description
A.72.2. Legacy Converted Enhanced PET Image IOD Entity-Relationship Model
A.72.3. Legacy Converted Enhanced PET Image IOD Module Table
A.72.3.1. Legacy Converted Enhanced PET Image IOD Content Constraints
A.72.4. Legacy Converted Enhanced PET Image Functional Group Macros
A.73. Corneal Topography Map IOD
A.73.1. Corneal Topography Map IOD Description
A.73.2. Corneal IOD Entity-Relationship Model
A.73.3. Corneal Topography Map IOD Modules
A.73.4. Corneal Topography Map IOD Content Constraints
A.73.4.1. Prohibited Modules
A.74. Breast Projection X-Ray Image IOD
A.74.1. Breast Projection X-Ray Image IOD Description
A.74.2. Breast Projection X-Ray Image IOD Entity-relationship Model
A.74.3. Breast Projection X-Ray Image IOD Module Table
A.74.3.1. Breast Projection X-Ray Image IOD Content Constraints
A.74.3.1.1. Modality Type Attribute
A.74.3.1.2. Overlay Plane Module, Curve Module and VOI LUT Module
A.74.4. Breast Projection X-Ray Image Functional Group Macros
A.74.4.1. Breast Projection X-Ray Image Functional Group Macros Content Constraints
A.74.4.1.1. Frame Anatomy Functional Group Macro
A.75. Parametric Map IOD
A.75.1. Parametric Map IOD Description
A.75.2. Parametric Map IOD Entity-Relationship Model
A.75.3. Parametric Map IOD Module Table
A.75.4. Parametric Map IOD Content Constraints
A.75.5. Parametric Map Functional Groups
A.75.5.1. Parametric Map Functional Groups Description
A.76. Wide Field Ophthalmic Photography Stereographic Projection Image IOD
A.76.1. Wide Field Ophthalmic Photography Stereographic Projection Image IOD Description
A.76.2. Wide Field Ophthalmic Photography Stereographic Projection Image IOD Entity-Relationship Model
A.76.3. Wide Field Ophthalmic Photography Stereographic Projection Image IOD Modules
A.76.4. Wide Field Ophthalmic Photography Stereographic Projection Image IOD Content Constraints
A.76.4.1. Bits Allocated, Bits Stored, and High Bit
A.76.4.2. Contrast/Bolus Agent Sequence
A.76.4.3. ICC Profile Module
A.77. Wide Field Ophthalmic Photography 3D Coordinates Image IOD
A.77.1. Wide Field Ophthalmic Photography 3D Coordinates Image IOD Description
A.77.2. Wide Field Ophthalmic Photography 3D Coordinates Image IOD Entity-Relationship Model
A.77.3. Wide Field Ophthalmic Photography 3D Coordinates Image IOD Modules
A.77.4. Wide Field Ophthalmic Photography 3D Coordinates Image IOD Content Constraints
A.77.4.1. Bits Allocated, Bits Stored, and High Bit
A.77.4.2. Contrast/bolus Agent Sequence
A.77.4.3. ICC Profile Module
A.78. Tractography Results IOD
A.78.1. Tractography Results IOD Description
A.78.2. Tractography Results IOD Entity-Relationship Model
A.78.3. Tractography Results IOD Module Table
A.79. RT Brachy Application Setup Delivery Instruction IOD
A.79.1. RT Brachy Application Setup Delivery Instruction IOD Description
A.79.2. RT Brachy Application Setup Delivery Instruction IOD Entity-Relationship Model
A.79.3. RT Brachy Application Setup Delivery Instruction IOD Module Table
A.79.4. RT Brachy Application Setup Delivery Instruction IOD Content Constraints
A.79.4.1. Modality
A.80. Volumetric Presentation State Information Object Definitions
A.80.1. Planar MPR Volumetric Presentation State IOD
A.80.1.1. Planar MPR Volumetric Presentation State IOD Description
A.80.1.2. Planar MPR Volumetric Presentation State IOD Module Table
A.80.1.3. Planar MPR Volumetric Presentation State IOD Content Constraints
A.80.1.3.1. Presentation Input Restrictions
A.80.1.3.2. Multi-Planar Reconstruction Style
A.80.1.3.3. Graphic Annotation Module
A.80.1.3.4. Volume Cropping Module
A.80.1.3.5. Volumetric Presentation State Reference Coordinate System
B. Normalized Information Object Definitions (Normative)
B.1. Patient Information Object Definition
B.2. Visit Information Object Definition
B.3. Study Information Object Definition
B.4. Study Component Information Object Definition
B.5. Results Information Object Definition
B.6. Interpretation Information Object Definition
B.7. Basic Film Session Information Object Definition
B.7.1. IOD Description
B.7.2. IOD Modules
B.8. Basic Film Box Information Object Definition
B.8.1. IOD Description
B.8.2. IOD Modules
B.9. Basic Image Box Information Object Definition
B.9.1. IOD Description
B.9.2. IOD Modules
B.10. Basic Annotation Box Information Object Definition
B.10.1. IOD Description
B.10.2. IOD Modules
B.11. Print Job Information Object Definition
B.11.1. IOD Description
B.11.2. IOD Modules
B.12. Printer Information Object Definition
B.12.1. IOD Description
B.12.2. IOD Modules
B.13. VOI LUT Box Information Object Definition (Retired)
B.14. Image Overlay Box Information Object Definition (Retired)
B.15. Storage Commitment Information Object Definition
B.15.1. Storage Commitment IOD Description
B.15.2. Storage Commitment IOD Modules
B.16. Print Queue Information Object Definition
B.17. Modality Performed Procedure Step Information Object Definition
B.17.1. IOD Description
B.17.2. IOD Modules
B.18. Presentation LUT Information Object Definition
B.18.1. IOD Description
B.18.2. IOD Modules
B.19. Pull Print Request Information Object Definition
B.20. Printer Configuration Information Object Definition
B.20.1. IOD Description
B.20.2. IOD Modules
B.21. Basic Print Image Overlay Box Information Object Definition
B.22. General Purpose Scheduled Procedure Step Information Object Definition (Retired)
B.23. General Purpose Performed Procedure Step Information Object Definition (Retired)
B.24. Instance Availability Notification Information Object Definition
B.24.1. IOD Description
B.24.2. IOD Modules
B.25. Media Creation Management Information Object Definition
B.25.1. IOD Description
B.25.2. IOD Modules
B.26. Unified Procedure Step Information Object Definition
B.26.1. IOD Description
B.26.2. IOD Modules
B.27. RT Conventional Machine Verification Information Object Definition
B.27.1. IOD Description
B.27.2. IOD Modules
B.28. RT Ion Machine Verification Information Object Definition
B.28.1. IOD Description
B.28.2. IOD Modules
B.29. Display System Information Object Definition
B.29.1. IOD Description
C. Information Module Definitions (Normative)
C.1. Elements of a Module Definition
C.1.1. Module Description
C.1.2. Module Definition
C.1.2.1. Attribute Name
C.1.2.2. Attribute Tag
C.1.2.3. Type Designation
C.1.2.4. Attribute Definition
C.1.3. Attribute Descriptions
C.2. Patient Modules
C.2.1. Patient Relationship Module
C.2.2. Patient Identification Module
C.2.2.1. Patient Identification Module Attributes
C.2.2.1.1. Referenced Patient Photo Sequence
C.2.3. Patient Demographic Module
C.2.4. Patient Medical Module
C.3. Visit Modules
C.3.1. Visit Relationship Module
C.3.2. Visit Identification Module
C.3.3. Visit Status Module
C.3.4. Visit Admission Module
C.3.5. Visit Discharge Module
C.3.6. Visit Scheduling Module
C.4. Study Modules
C.4.1. Study Relationship Module
C.4.2. Study Identification Module
C.4.3. Study Classification Module
C.4.4. Study Scheduling Module
C.4.5. Study Acquisition Module
C.4.6. Study Read Module
C.4.7. Study Component Module
C.4.8. Study Component Relationship Module
C.4.9. Study Component Acquisition Module
C.4.10. Scheduled Procedure Step Module
C.4.10.1. Protocol Context Sequence
C.4.11. Requested Procedure Module
C.4.12. Imaging Service Request Module
C.4.13. Performed Procedure Step Relationship
C.4.14. Performed Procedure Step Information
C.4.15. Image Acquisition Results Module
C.4.16. Radiation Dose Module
C.4.17. Billing and Material Management Code Module
C.4.18. General Purpose Scheduled Procedure Step Relationship Module (Retired)
C.4.19. General Purpose Scheduled Procedure Step Information Module (Retired)
C.4.20. General Purpose Performed Procedure Step Relationship Module (Retired)
C.4.21. General Purpose Performed Procedure Step Information Module (Retired)
C.4.22. General Purpose Results (Retired)
C.4.23. Instance Availability Notification Module
C.4.23.1. Instance Availability Notification Module Attribute Definitions
C.4.23.1.1. Instance Availability
C.5. Results Modules
C.6. Interpretation Modules
C.7. Common Composite Image IOD Modules
C.7.1. Common Patient IE Modules
C.7.1.1. Patient Module
C.7.1.1.1. Patient Module Attributes
C.7.1.1.1.1. Patient Breed Description and Code Sequence
C.7.1.1.1.2. Responsible Person Role
C.7.1.1.1.3. Patient Species (Taxonomic Rank Value)
C.7.1.1.1.4. Patient Strain
C.7.1.2. Specimen Identification Module
C.7.1.3. Clinical Trial Subject Module
C.7.1.3.1. Clinical Trial Subject Attribute Descriptions
C.7.1.3.1.1. Clinical Trial Sponsor Name
C.7.1.3.1.2. Clinical Trial Protocol ID
C.7.1.3.1.3. Clinical Trial Protocol Name
C.7.1.3.1.4. Clinical Trial Site ID
C.7.1.3.1.5. Clinical Trial Site Name
C.7.1.3.1.6. Clinical Trial Subject ID
C.7.1.3.1.7. Clinical Trial Subject Reading ID
C.7.1.4. Patient Group Macro
C.7.1.4.1. Patient Group Macro Attributes
C.7.1.4.1.1. Groups of Subjects
C.7.1.4.1.1.1. Subject Relative Position in Image and Patient Position
C.7.2. Common Study IE Modules
C.7.2.1. General Study Module
C.7.2.1.1. General Study Attribute Descriptions
C.7.2.1.1.1. Referring Physician, Physician of Record, Physician Reading Study, Consulting Physician
C.7.2.2. Patient Study Module
C.7.2.3. Clinical Trial Study Module
C.7.2.3.1. Clinical Trial Study Attribute Descriptions
C.7.2.3.1.1. Clinical Trial Time Point
C.7.2.3.1.2. Consent For Clinical Trial Use Sequence
C.7.3. Common Series IE Modules
C.7.3.1. General Series Module
C.7.3.1.1. General Series Attribute Descriptions
C.7.3.1.1.1. Modality
C.7.3.1.1.2. Patient Position
C.7.3.2. Clinical Trial Series Module
C.7.3.2.1. Clinical Trial Series Attribute Descriptions
C.7.3.2.1.1. Clinical Trial Coordinating Center Name
C.7.3.2.1.2. Clinical Trial Series Identifier and Description
C.7.3.3. Enhanced Series Module
C.7.4. Common Frame of Reference Information Entity Modules
C.7.4.1. Frame of Reference Module
C.7.4.1.1. Frame of Reference Attribute Descriptions
C.7.4.1.1.1. Frame of Reference UID
C.7.4.1.1.2. Position Reference Indicator
C.7.4.2. Synchronization Module
C.7.4.2.1. Synchronization Attribute Descriptions
C.7.4.2.1.1. Synchronization Frame of Reference UID
C.7.4.2.1.2. Time Source and Time Distribution Protocol
C.7.4.2.1.3. Synchronization Channel
C.7.4.2.1.4. Acquisition Time Synchronized
C.7.5. Common Equipment IE Modules
C.7.5.1. General Equipment Module
C.7.5.1.1. General Equipment Attribute Descriptions
C.7.5.1.1.1. Date of Last Calibration, Time of Last Calibration
C.7.5.1.1.2. Pixel Padding Value and Pixel Padding Range Limit
C.7.5.1.1.3. Software Versions
C.7.5.2. Enhanced General Equipment Module
C.7.6. Common Image IE Modules
C.7.6.1. General Image Module
C.7.6.1.1. General Image Attribute Descriptions
C.7.6.1.1.1. Patient Orientation
C.7.6.1.1.2. Image Type
C.7.6.1.1.3. Derivation Description
C.7.6.1.1.4. Source Image Sequence
C.7.6.1.1.5. Lossy Image Compression
C.7.6.1.1.5.1. Lossy Image Compression Method
C.7.6.1.1.5.2. Lossy Image Compression Ratio
C.7.6.1.1.6. Icon Image Sequence
C.7.6.1.1.7. Irradiation Event UID
C.7.6.2. Image Plane Module
C.7.6.2.1. Image Plane Attribute Descriptions
C.7.6.2.1.1. Image Position and Image Orientation
C.7.6.2.1.2. Slice Location
C.7.6.3. Image Pixel Module
C.7.6.3.1. Image Pixel Attribute Descriptions
C.7.6.3.1.1. Samples Per Pixel
C.7.6.3.1.2. Photometric Interpretation
C.7.6.3.1.3. Planar Configuration
C.7.6.3.1.4. Pixel Data
C.7.6.3.1.5. Palette Color Lookup Table Descriptor
C.7.6.3.1.6. Palette Color Lookup Table Data
C.7.6.3.1.7. Pixel Aspect Ratio
C.7.6.4. Contrast/Bolus Module
C.7.6.4b. Enhanced Contrast/Bolus Module
C.7.6.4b.1. Enhanced Contrast/Bolus Module Attributes
C.7.6.4b.1.1. Contrast/Bolus Ingredient Opaque for X-Ray Equipment
C.7.6.5. Cine Module
C.7.6.5.1. Cine Attribute Descriptions
C.7.6.5.1.1. Frame Time
C.7.6.5.1.2. Frame Time Vector
C.7.6.5.1.3. Multiplexed Audio
C.7.6.6. Multi-frame Module
C.7.6.6.1. Multi-frame Attribute Descriptions
C.7.6.6.1.1. Number of Frames and Frame Increment Pointer
C.7.6.6.1.2. Frame Increment Pointer
C.7.6.6.1.3. Stereoscopic Pairs Present
C.7.6.7. Bi-plane Sequence Module (Retired)
C.7.6.8. Bi-plane Image Module (Retired)
C.7.6.9. Frame Pointers Module
C.7.6.10. Mask Module
C.7.6.10.1. Mask Subtraction Attribute Descriptions
C.7.6.10.1.1. Mask Operation
C.7.6.10.1.2. Mask Sub-pixel Shift
C.7.6.11. Display Shutter Module
C.7.6.12. Device Module
C.7.6.12.1. Device Attribute Descriptions
C.7.6.12.1.1. Device Type and Size
C.7.6.13. Intervention Module
C.7.6.14. Acquisition Context Module
C.7.6.15. Bitmap Display Shutter Module
C.7.6.16. Multi-frame Functional Groups Module
C.7.6.16.1. Multi-frame Functional Groups Module Attribute Description
C.7.6.16.1.1. Functional Group
C.7.6.16.1.2. Per-frame Functional Groups Sequence
C.7.6.16.1.3. SOP Instance UID of Concatenation Source
C.7.6.16.2. Common Functional Group Macros
C.7.6.16.2.1. Pixel Measures Macro
C.7.6.16.2.2. Frame Content Macro
C.7.6.16.2.2.1. Timing Parameter Relationships
C.7.6.16.2.2.2. Frame Reference DateTime
C.7.6.16.2.2.3. Frame Acquisition Duration
C.7.6.16.2.2.4. Concatenations and Stacks
C.7.6.16.2.2.5. Frame Label
C.7.6.16.2.2.6. Temporal Position Index and Stack ID in PET images
C.7.6.16.2.2.7. Stack ID usage in PET static, whole body and gated images
C.7.6.16.2.2.8. Usage of Temporal Position Index, Stack ID and In-Stack Position Number in fMRI Volumes
C.7.6.16.2.3. Plane Position (Patient) Macro
C.7.6.16.2.3.1. Position and Orientation for SAMPLED Frames
C.7.6.16.2.4. Plane Orientation (Patient) Macro
C.7.6.16.2.5. Referenced Image Macro
C.7.6.16.2.5.1. Use of Referenced Image Macro
C.7.6.16.2.6. Derivation Image Macro
C.7.6.16.2.7. Cardiac Synchronization Macro
C.7.6.16.2.7.1. Relationship of Cardiac Timing Attributes
C.7.6.16.2.8. Frame Anatomy Macro
C.7.6.16.2.9. Pixel Value Transformation Macro
C.7.6.16.2.9b. Identity Pixel Value Transformation Macro
C.7.6.16.2.10. Frame VOI LUT Macro
C.7.6.16.2.10b. Frame VOI LUT With LUT Macro
C.7.6.16.2.11. Real World Value Mapping Macro
C.7.6.16.2.11.1. Real World Value Representation
C.7.6.16.2.11.1.1. Real World Value Mapping Sequence
C.7.6.16.2.11.1.2. Real World Values Mapping Sequence Attributes
C.7.6.16.2.12. Contrast/Bolus Usage Macro
C.7.6.16.2.13. Pixel Intensity Relationship LUT Macro
C.7.6.16.2.13.1. Pixel Intensity Relationship LUT
C.7.6.16.2.13.2. Pixel Intensity Relationship LUT Data Attribute
C.7.6.16.2.14. Frame Pixel Shift Macro
C.7.6.16.2.14.1. Subtraction Item ID Description
C.7.6.16.2.15. Patient Orientation in Frame Macro
C.7.6.16.2.16. Frame Display Shutter
C.7.6.16.2.17. Respiratory Synchronization Macro
C.7.6.16.2.17.1. Relationship of Respiratory Timing Attributes
C.7.6.16.2.18. Irradiation Event Identification Macro
C.7.6.16.2.19. Radiopharmaceutical Usage Macro
C.7.6.16.2.20. Patient Physiological State Macro
C.7.6.16.2.21. Plane Position (Volume) Macro
C.7.6.16.2.22. Plane Orientation (Volume) Macro
C.7.6.16.2.23. Temporal Position Macro
C.7.6.16.2.24. Image Data Type Macro
C.7.6.16.2.24.1. Data Type
C.7.6.16.2.24.2. Aliased Data Type
C.7.6.16.2.24.3. Zero Velocity Pixel Value
C.7.6.16.2.25. Unassigned Shared and Per-frame Converted Attributes Macros
C.7.6.16.2.25.1. Unassigned Shared Converted Attributes Macro
C.7.6.16.2.25.2. Unassigned Per-Frame Converted Attributes Macro
C.7.6.16.2.25.3. Image Frame Conversion Source Macro
C.7.6.17. Multi-frame Dimension Module
C.7.6.17.1. Dimension Indices
C.7.6.17.2. Dimension Organization UID
C.7.6.18. Physiological Synchronization
C.7.6.18.1. Cardiac Synchronization Module
C.7.6.18.1.1. Attribute Descriptions
C.7.6.18.1.1.1. Cardiac Framing Type
C.7.6.18.2. Respiratory Synchronization Module
C.7.6.18.3. Bulk Motion Synchronization Module
C.7.6.19. Supplemental Palette Color Lookup Table Module
C.7.6.20. Patient Orientation Module
C.7.6.21. Image - Equipment Coordinate Relationship Module
C.7.6.21.1. Image to Equipment Mapping Matrix
C.7.6.21.2. Equipment Coordinate System Identification
C.7.6.22. Specimen Module
C.7.6.22.1. Specimen Module Attributes
C.7.6.22.1.1. Container Identifier and Specimen Identifier
C.7.6.22.1.2. Specimen Identifier and Specimen UID
C.7.6.22.1.3. Specimen Preparation Sequence and Specimen Preparation Step Content Item Sequence
C.7.6.22.1.4. Specimen Localization Content Item Sequence
C.7.6.23. Enhanced Palette Color Lookup Table Module
C.7.6.23.1. Description of the Enhanced Blending and Display Pipeline
C.7.6.23.2. Data Path Assignment
C.7.6.23.3. Bits Mapped to Color Lookup Table
C.7.6.23.4. Blending LUT Transfer Function
C.7.6.23.5. Blending LUT Descriptor
C.7.6.23.6. Lossy Compression and Palette Color Lookup Tables (Informative)
C.7.6.24. Floating Point Image Pixel Module
C.7.6.25. Double Floating Point Image Pixel Module
C.7.7. Patient Summary Module
C.7.8. Study Content Module
C.7.9. Palette Color Lookup Table Module
C.7.9.1. Palette Color Lookup Table UID
C.7.9.2. Segmented Palette Color Lookup Table Data
C.7.9.2.1. Discrete Segment Type
C.7.9.2.2. Linear Segment Type
C.7.9.2.3. Indirect Segment Type
C.8. Modality Specific Modules
C.8.1. Computed Radiography Modules
C.8.1.1. CR Series Module
C.8.1.2. CR Image Module
C.8.2. CT Modules
C.8.2.1. CT Image Module
C.8.2.1.1. CT Image Attribute Descriptions
C.8.2.1.1.1. Image Type
C.8.2.1.1.2. Samples Per Pixel
C.8.2.1.1.3. Photometric Interpretation
C.8.2.1.1.4. Bits Allocated
C.8.2.1.1.5. Bits Stored
C.8.2.1.1.6. High Bit
C.8.2.1.1.7. Calcium Scoring Mass Factor Patient and Device
C.8.3. MR Modules
C.8.3.1. MR Image Module
C.8.3.1.1. MR Image Attribute Descriptions
C.8.3.1.1.1. Image Type
C.8.3.1.1.2. Samples Per Pixel
C.8.3.1.1.3. Photometric Interpretation
C.8.3.1.1.4. Bits Allocated
C.8.4. Nuclear Medicine Modules
C.8.4.1. NM Series Module (Retired)
C.8.4.2. NM Equipment Module (Retired)
C.8.4.3. NM Image Module (Retired)
C.8.4.4. NM Spect Acquisition Image Module (Retired)
C.8.4.5. NM Multi-gated Acquisition Image Module (Retired)
C.8.4.6. NM/PET Patient Orientation Module
C.8.4.6.1. NM/PET Patient Orientation Attribute Descriptions
C.8.4.6.1.1. Patient Orientation Code Sequence
C.8.4.6.1.2. Patient Orientation Modifier Code Sequence
C.8.4.6.1.3. Patient Gantry Relationship Code Sequence
C.8.4.7. NM Image Pixel Module
C.8.4.7.1. NM Image Pixel Attribute Descriptions
C.8.4.7.1.1. Photometric Interpretation
C.8.4.8. NM Multi-frame Module
C.8.4.8.1. NM Multi-frame Attribute Descriptions
C.8.4.8.1.1. Frame Increment Pointer
C.8.4.8.1.2. Number of Energy Windows and Energy Window Vector
C.8.4.8.1.3. Number of Detectors and Detector Vector
C.8.4.8.1.4. Number of Phases and Phase Vector
C.8.4.8.1.5. Number of Rotations and Rotation Vector
C.8.4.8.1.6. Number of R-R Intervals and R-R Interval Vector
C.8.4.8.1.7. Number of Time Slots and Time Slot Vector
C.8.4.8.1.8. Number of Slices and Slice Vector
C.8.4.8.1.9. Angular View Vector
C.8.4.8.1.10. Time Slice Vector
C.8.4.9. NM Image Module
C.8.4.9.1. NM Image Module Attribute Descriptions
C.8.4.9.1.1. Image Type
C.8.4.9.1.2. Counts Accumulated
C.8.4.9.1.3. Acquisition Termination Condition
C.8.4.9.1.4. Actual Frame Duration
C.8.4.10. NM Isotope Module
C.8.4.10.1. NM Isotope Module Attribute Descriptions
C.8.4.10.1.1. Energy Window Lower Limit
C.8.4.10.1.2. Energy Window Upper Limit
C.8.4.10.1.3. (Retired)
C.8.4.10.1.4. (Retired)
C.8.4.10.1.5. Radiopharmaceutical Start Time
C.8.4.10.1.6. Radiopharmaceutical Stop Time
C.8.4.10.1.7. Radionuclide Total Dose
C.8.4.10.1.8. Syringe Counts
C.8.4.10.1.9. Residual Syringe Counts
C.8.4.10.1.10. (Retired)
C.8.4.10.1.11. (Retired)
C.8.4.11. NM Detector Module
C.8.4.11.1. NM Detector Attribute Descriptions
C.8.4.11.1.1. Focal Distance
C.8.4.11.1.2. Focus Center
C.8.4.11.1.3. Zoom Center
C.8.4.11.1.4. Zoom Factor
C.8.4.11.1.5. Center of Rotation Offset
C.8.4.11.1.6. Gantry/Detector Tilt
C.8.4.12. NM Tomo Acquisition Module
C.8.4.12.1. NM Tomo Acquisition Attribute Descriptions
C.8.4.12.1.1. Angular Step
C.8.4.13. NM Multi-gated Acquisition Module
C.8.4.13.1. NM Multi-gated Acquisition Attribute Descriptions
C.8.4.13.1.1. Data Information Sequence
C.8.4.13.1.2. Time Slot Time
C.8.4.14. NM Phase Module
C.8.4.14.1. NM Phase Module Attributes Description
C.8.4.14.1.1. Trigger Vector
C.8.4.15. NM Reconstruction Module
C.8.5. Ultrasound Modules
C.8.5.1. US Frame of Reference Module (Retired)
C.8.5.2. US Region Calibration (Retired)
C.8.5.3. US Image Module (Retired)
C.8.5.4. US Frame of Reference Module
C.8.5.5. US Region Calibration Module
C.8.5.5.1. US Region Calibration Attribute Descriptions
C.8.5.5.1.1. Region Spatial Format
C.8.5.5.1.2. Region Data Type
C.8.5.5.1.3. Region Flags
C.8.5.5.1.4. Pixel Component Organization
C.8.5.5.1.5. Pixel Component Mask
C.8.5.5.1.6. Pixel Component Physical Units
C.8.5.5.1.7. Pixel Component Data Type
C.8.5.5.1.8. Number of Table Break Points
C.8.5.5.1.9. Table of X Break Points and Table of Y Break Points
C.8.5.5.1.10. TM-line Position X0, Y0, X1 and Y1
C.8.5.5.1.11. Number of Table Entries
C.8.5.5.1.12. Table of Pixel Values
C.8.5.5.1.13. Table of Parameter Values
C.8.5.5.1.14. Region Location Min X0, Min Y0, Max X1 and Max Y1
C.8.5.5.1.15. Physical Units X Direction and Physical Units Y Direction
C.8.5.5.1.16. Reference Pixel X0 and Reference Pixel Y0
C.8.5.5.1.16.1. 2D - Tissue or Color Flow
C.8.5.5.1.16.2. Spectral - CW or PW Doppler or Doppler Trace
C.8.5.5.1.16.3. M-Mode - Tissue or Color Flow
C.8.5.5.1.16.4. Waveform - ECG, Phonocardiogram and Pulse Traces
C.8.5.5.1.16.5. Waveform - Doppler Mode, Mean and Max Trace
C.8.5.5.1.16.6. Graphics Spatial Formats
C.8.5.5.1.16.7. Treatment of Sweeping Regions
C.8.5.5.1.17. Physical Delta X and Physical Delta Y
C.8.5.5.1.18. Pixel Value Mapping Code Sequence
C.8.5.6. US Image Module
C.8.5.6.1. US Image Attribute Descriptions
C.8.5.6.1.1. Image Type
C.8.5.6.1.2. Photometric Interpretation
C.8.5.6.1.3. Pixel Representation
C.8.5.6.1.4. Frame Increment Pointer
C.8.5.6.1.5. (Retired)
C.8.5.6.1.6. (Retired)
C.8.5.6.1.7. (Retired)
C.8.5.6.1.8. Mechanical Index, Bone Thermal Index, Cranial Thermal Index, Soft Tissue Thermal Index
C.8.5.6.1.9. Image Transformation Matrix and Image Translation Vector
C.8.5.6.1.10. Ultrasound Color Data Present
C.8.5.6.1.11. Overlay Subtype
C.8.5.6.1.12. Samples Per Pixel
C.8.5.6.1.13. Bits Allocated
C.8.5.6.1.14. Bits Stored
C.8.5.6.1.15. High Bit
C.8.5.6.1.16. Planar Configuration
C.8.5.6.1.19. View Code Sequence
C.8.5.6.1.20. (Retired)
C.8.5.6.1.21. IVUS Acquisition
C.8.5.6.1.22. IVUS Pullback Rate
C.8.5.6.1.23. IVUS Gated Rate
C.8.5.6.1.24. IVUS Pullback Start Frame Number
C.8.5.6.1.25. IVUS Pullback Stop Frame Number
C.8.5.6.1.26. Lesion Number
C.8.6. Secondary Capture Modules
C.8.6.1. SC Equipment Module
C.8.6.2. SC Image Module
C.8.6.3. SC Multi-frame Image Module
C.8.6.3.1. Scanned Film, Optical Density and P-Values
C.8.6.4. SC Multi-frame Vector Module
C.8.7. X-Ray Modules
C.8.7.1. X-Ray Image Module
C.8.7.1.1. X-Ray Image Attribute Descriptions
C.8.7.1.1.1. Image Type
C.8.7.1.1.2. Pixel Intensity Relationship
C.8.7.1.1.3. Acquisition Device Processing Description
C.8.7.1.1.4. Scan Options
C.8.7.1.1.5. Derivation Description
C.8.7.1.1.6. Bits Allocated
C.8.7.1.1.7. Bits Stored
C.8.7.1.1.8. High Bit
C.8.7.1.1.9. Synchronization of Frame and Waveform Times
C.8.7.1.1.12. Frame Dimension Pointer
C.8.7.1.1.13. Referenced Image Sequence
C.8.7.2. X-Ray Acquisition Module
C.8.7.2.1. X-Ray Acquisition Attribute Descriptions
C.8.7.2.1.1. Exposure Time
C.8.7.2.1.2. Field of View
C.8.7.3. X-Ray Collimator Module
C.8.7.3.1. X-Ray Collimator Attribute Descriptions
C.8.7.3.1.1. Collimator Vertical and Horizontal Edges
C.8.7.4. X-Ray Table Module
C.8.7.4.1. X-Ray Table Attribute Descriptions
C.8.7.4.1.1. Table Motion Increments
C.8.7.4.1.2. Table Longitudinal Increment
C.8.7.4.1.3. Table Lateral Increment
C.8.7.4.1.4. Table Motion With Patient in Relation to Imaging Chain
C.8.7.5. XA Positioner Module
C.8.7.5.1. XA Positioner Attribute Descriptions
C.8.7.5.1.1. Positioner Motion
C.8.7.5.1.2. Positioner Primary and Secondary Angles
C.8.7.5.1.3. Positioner Angle Increments
C.8.7.5.1.4. Detector Primary and Secondary Angles
C.8.7.6. XRF Positioner Module
C.8.7.7. X-Ray Tomography Acquisition Module
C.8.7.8. X-Ray Acquisition Dose Module
C.8.7.9. X-Ray Generation Module
C.8.7.10. X-Ray Filtration Module
C.8.7.11. X-Ray Grid Module
C.8.8. Radiotherapy Modules
C.8.8.1. RT Series Module
C.8.8.1.1. Modality
C.8.8.2. RT Image Module
C.8.8.2.1. Multi-frame Image Data
C.8.8.2.2. X-Ray Image Receptor Angle
C.8.8.2.3. Image Plane Pixel Spacing and RT Image SID
C.8.8.2.4. Exposure Sequence
C.8.8.2.5. Single-Frame and Multi-frame Images
C.8.8.2.6. Image Pixel Module Attributes
C.8.8.2.6.1. Samples Per Pixel
C.8.8.2.6.2. Photometric Interpretation
C.8.8.2.6.3. Bits Allocated
C.8.8.2.6.4. Bits Stored
C.8.8.2.6.5. High Bit
C.8.8.2.6.6. Pixel Representation
C.8.8.2.7. RT Image Plane, Position and Orientation
C.8.8.2.8. Exposure Time and Meterset Exposure
C.8.8.3. RT Dose Module
C.8.8.3.1. Normalization Point
C.8.8.3.2. Grid Frame Offset Vector
C.8.8.3.3. Dose Units
C.8.8.3.4. Image Pixel Module Attributes
C.8.8.3.4.1. Samples Per Pixel
C.8.8.3.4.2. Photometric Interpretation
C.8.8.3.4.3. Bits Allocated
C.8.8.3.4.4. Bits Stored
C.8.8.3.4.5. High Bit
C.8.8.3.4.6. Pixel Representation
C.8.8.3.5. Referenced Spatial Registration Sequence
C.8.8.4. RT DVH Module
C.8.8.4.1. Referenced Structure Set Sequence
C.8.8.4.2. DVH ROI Contribution Type
C.8.8.4.3. DVH Volume Units
C.8.8.5. Structure Set Module
C.8.8.5.1. Frames of Reference
C.8.8.5.2. Frame of Reference Relationship Sequence and Transformation Matrix
C.8.8.5.3. ROI Derivation Sequence
C.8.8.5.4. SOP Class UID in RT Referenced Study Sequence
C.8.8.6. ROI Contour Module
C.8.8.6.1. Contour Geometric Type
C.8.8.6.2. Contour Slab Thickness
C.8.8.6.3. Representing Inner and Outer Contours on an Image
C.8.8.7. RT Dose ROI Module
C.8.8.7.1. Contour Geometric Type of Referenced ROI
C.8.8.7.2. Referenced ROI Number
C.8.8.7.3. Dose Value
C.8.8.8. RT ROI Observations Module
C.8.8.8.1. RT ROI Interpreted Type
C.8.8.8.2. Additional RT ROI Identification Code Sequence
C.8.8.9. RT General Plan Module
C.8.8.9.1. Referenced Structure Set Sequence
C.8.8.10. RT Prescription Module
C.8.8.10.1. Target Underdose Volume Fraction
C.8.8.11. RT Tolerance Tables Module
C.8.8.12. RT Patient Setup Module
C.8.8.12.1. RT Patient Setup Module Attributes
C.8.8.12.1.1. Referenced Setup Image Sequence
C.8.8.12.1.2. Patient Position
C.8.8.13. RT Fraction Scheme Module
C.8.8.13.1. Beam Dose Verification Parameters
C.8.8.13.1.1. Referenced Control Point
C.8.8.13.1.2. Distance Parameters
C.8.8.14. RT Beams Module
C.8.8.14.1. Meterset Calculations
C.8.8.14.2. Planned Verification Image Sequence
C.8.8.14.3. X-Ray Image Receptor Angle
C.8.8.14.4. Multiple Aperture Blocks
C.8.8.14.5. Control Point Sequence
C.8.8.14.6. Absolute and Relative Machine Coordinates
C.8.8.14.7. Cumulative Dose Reference Coefficient
C.8.8.14.8. Machine Rotations
C.8.8.14.9. Compensator Thickness Data and Source to Compensator Distance
C.8.8.14.10. Compensator Transmission and Thickness Data Direction
C.8.8.14.11. Block and Compensator Precedence for Dosimetric Calculations
C.8.8.14.12. Table Top Pitch and Table Top Roll
C.8.8.14.13. Angular Values in RT Beams Module
C.8.8.14.14. Effective Wedge Angle
C.8.8.14.15. Source to External Contour Distance and External Contour Entry Point
C.8.8.15. RT Brachy Application Setups Module
C.8.8.15.1. Permanent Implants
C.8.8.15.2. Referenced ROI Number
C.8.8.15.3. Channel Length
C.8.8.15.4. Oscillating Source Movement
C.8.8.15.5. Channel Shields
C.8.8.15.6. Time Calculations
C.8.8.15.7. Brachy Control Point Sequence
C.8.8.15.8. Source Transit Time
C.8.8.15.9. Control Point Relative Position
C.8.8.15.10. Control Point 3D Position
C.8.8.15.11. Cumulative Dose Reference Coefficient
C.8.8.15.12. Nominal Thickness and Nominal Transmission
C.8.8.15.13. Reference Point for Calibration of Beta Emitting Isotopes
C.8.8.15.14. Orientation of Brachy Sources
C.8.8.15.15. Source Model ID
C.8.8.16. Approval Module
C.8.8.17. RT General Treatment Record Module
C.8.8.18. RT Treatment Machine Record Module
C.8.8.19. Measured Dose Reference Record Module
C.8.8.20. Calculated Dose Reference Record Module
C.8.8.21. RT Beams Session Record Module
C.8.8.21.1. Control Point Machine Delivery Parameters
C.8.8.21.2. Specified and Delivered Meterset Values
C.8.8.21.2.1. Beam Level
C.8.8.21.2.2. Control Point Level
C.8.8.22. RT Brachy Session Record Module
C.8.8.22.1. PDR (Pulsed Dose Rate) Treatment
C.8.8.22.2. Specified Channel Total Time
C.8.8.23. RT Treatment Summary Record Module
C.8.8.23.1. Current Treatment Status
C.8.8.24. RT Ion Tolerance Tables Module
C.8.8.25. RT Ion Beams Module
C.8.8.25.1. Beam Identifying Information
C.8.8.25.2. Treatment Machine Name
C.8.8.25.3. Leaf Position Boundaries
C.8.8.25.4. Virtual Source-Axis Distances and the Use of Trays in Ion Therapy
C.8.8.25.5. Range Shifter and Lateral Spreading Device Settings
C.8.8.25.6. Coordinate Systems
C.8.8.25.6.1. Fixed Beam Line
C.8.8.25.6.2. Table Top Pitch and Table Top Roll
C.8.8.25.6.3. Seated Treatments
C.8.8.25.6.4. Ocular Treatments
C.8.8.25.6.4.1. Gantry Beam Line
C.8.8.25.6.4.2. Fixed Beam Line
C.8.8.25.6.5. Gantry Pitch Angle
C.8.8.25.7. Ion Control Point Sequence
C.8.8.25.8. Scan Spot Maps
C.8.8.25.9. Depth Dose Parameters Sequence Attributes
C.8.8.26. RT Ion Beams Session Record Module
C.8.8.26.1. Specified and Delivered Meterset Values
C.8.8.27. Beam Limiting Device Position Macro
C.8.8.28. Patient Support Identification Macro
C.8.8.29. RT Beams Delivery Instruction Module
C.8.8.29.1. Current Fraction Number
C.8.8.29.2. Adjusted Table Positions and Angles
C.8.8.29.3. Meterset Exposure
C.8.8.29.4. Double Exposure Field Delta
C.8.8.29.5. Beam Order Index
C.8.8.29.6. Autosequence Flag
C.8.8.30. RT Brachy Application Setup Delivery Instruction Module
C.8.8.30.1. Brachy Task Sequence
C.8.8.30.1.1. Usage Scenario 1: HDR Treatment, Interruption Skipping Remainder of the Interrupted Delivery
C.8.8.30.1.2. Usage Scenario 2: PDR Treatment, Interruption and Continuation
C.8.8.30.2. Treatment Delivery Type
C.8.8.30.3. Current Fraction Number
C.8.8.30.4. Current Pulse Number
C.8.9. PET Information Module Definitions
C.8.9.1. PET Series Module
C.8.9.1.1. PET Series Attribute Descriptions
C.8.9.1.1.1. Specialization of Image Plane Module and Image Pixel Module Attributes
C.8.9.1.1.2. Series Date, Series Time
C.8.9.1.1.3. Units
C.8.9.1.1.4. Series Type
C.8.9.1.1.5. Decay Correction
C.8.9.1.1.6. Acquisition Start Condition
C.8.9.1.1.7. Gantry/Detector Tilt
C.8.9.1.1.8. Axial Mash
C.8.9.1.1.9. Transverse Mash
C.8.9.1.1.10. Energy Window Range Sequence
C.8.9.1.1.11. Temporal Relationships of Images in PET Series
C.8.9.2. PET Isotope Module
C.8.9.3. PET Multi-gated Acquisition Module
C.8.9.4. PET Image Module
C.8.9.4.1. PET Image Module Attribute Descriptions
C.8.9.4.1.1. Image Type
C.8.9.4.1.2. Photometric Interpretation
C.8.9.4.1.3. Frame Time
C.8.9.4.1.4. Acquisition Date, Acquisition Time
C.8.9.4.1.5. Frame Reference Time
C.8.9.4.1.6. Actual Frame Duration
C.8.9.4.1.7. Secondary Counts Accumulated
C.8.9.4.1.8. Dose Calibration Factor
C.8.9.4.1.9. Image Index
C.8.9.5. PET Curve Module
C.8.10. Hardcopy Modules
C.8.11. DX Modules
C.8.11.1. DX Series Module
C.8.11.1.1. DX Series Attribute Descriptions
C.8.11.1.1.1. Presentation Intent Type
C.8.11.2. DX Anatomy Imaged Module
C.8.11.2.1. DX Anatomy Imaged Attribute Descriptions
C.8.11.3. DX Image Module
C.8.11.3.1. DX Image Attribute Descriptions
C.8.11.3.1.1. Image Type
C.8.11.3.1.2. Pixel Intensity Relationship and Grayscale Transformations
C.8.11.3.1.3. Acquisition Device Processing Description
C.8.11.3.1.4. Derivation Description
C.8.11.3.1.5. VOI Attributes
C.8.11.4. DX Detector Module
C.8.11.4.1. DX Detector Attribute Descriptions
C.8.11.4.1.1. Physical, Active, Field of View, Exposed and Displayed Areas
C.8.11.5. DX Positioning Module
C.8.11.5.1. DX Positioning Attribute Descriptions
C.8.11.5.1.1. View Code Sequence
C.8.11.5.1.2. Patient Orientation Code Sequence
C.8.11.6. Mammography Series Module
C.8.11.7. Mammography Image Module
C.8.11.7.1. Mammography Image Attribute Descriptions
C.8.11.7.1.1. Mammography X-Ray Beam and X-Ray Beam Vector Definition
C.8.11.7.1.2. Detector Primary and Secondary Angles
C.8.11.7.1.3. Partial View Code Sequence
C.8.11.7.1.4. Image Type
C.8.11.8. Intra-Oral Series Module
C.8.11.9. Intra-Oral Image Module
C.8.11.9.1. Intra-Oral Image Attribute Descriptions
C.8.11.9.1.1. Primary Anatomic Structure Sequence
C.8.11.10. Enhanced Mammography Series Module
C.8.12. VL Modules and Functional Group Macros
C.8.12.1. VL Image Module
C.8.12.1.1. VL Image Module Attribute Descriptions
C.8.12.1.1.1. Photometric Interpretation
C.8.12.1.1.2. Bits Allocated, Bits Stored, and High Bit
C.8.12.1.1.3. Pixel Representation
C.8.12.1.1.4. Samples Per Pixel
C.8.12.1.1.5. Planar Configuration
C.8.12.1.1.6. Image Type
C.8.12.1.1.7. Referenced Image Sequence
C.8.12.2. Slide Coordinates Module
C.8.12.2.1. Slide Coordinates Attribute Descriptions
C.8.12.2.1.1. Image Center Point Coordinates Sequence
C.8.12.3. Whole Slide Microscopy Series Module
C.8.12.4. Whole Slide Microscopy Image Module
C.8.12.4.1. Whole Slide Microscopy Image Attribute Descriptions
C.8.12.4.1.1. Image Type
C.8.12.4.1.2. Imaged Volume Width, Height, Depth
C.8.12.4.1.3. Total Pixel Matrix Columns, Rows
C.8.12.4.1.4. Total Pixel Matrix Origin Sequence and Image Orientation (slide)
C.8.12.4.1.5. Photometric Interpretation and Samples Per Pixel
C.8.12.5. Optical Path Module
C.8.12.5.1. Optical Path Attribute Descriptions
C.8.12.5.1.1. Optical Path Sequence and Optical Path Identifier
C.8.12.5.1.2. Illumination Type Code Sequence
C.8.12.5.1.3. Light Path and Image Path Filter Type Stack Code Sequences
C.8.12.5.1.4. ICC Profile
C.8.12.6. Whole Slide Microscopy Image Functional Group Macros
C.8.12.6.1. Plane Position (Slide) Macro
C.8.12.6.2. Optical Path Identification Macro
C.8.12.6.3. Specimen Reference Macro
C.8.12.7. Multi-Resolution Navigation Module
C.8.12.8. Slide Label Module
C.8.13. Enhanced MR Image
C.8.13.1. Enhanced MR Image Module
C.8.13.1.1. Enhanced MR Image Module Attribute Description
C.8.13.1.1.1. Image Type and Frame Type
C.8.13.1.1.1.1. Pixel Data Characteristics
C.8.13.1.1.1.2. Patient Examination Characteristics
C.8.13.1.1.1.3. Image Flavor
C.8.13.1.1.1.4. Derived Pixel Contrast
C.8.13.1.1.2. Photometric Interpretation, Pixel Representation, Samples Per Pixel, Planar Configuration, Bits Allocated and Bits Stored
C.8.13.1.1.3. Pixel Presentation
C.8.13.1.1.3.1. Supplemental Palette Color LUTs
C.8.13.1.1.4. Volumetric Properties
C.8.13.1.1.5. Volume Based Calculation Technique
C.8.13.2. MR Image and Spectroscopy Instance Macro
C.8.13.2.1. MR Image and Spectroscopy Instance Macro Attribute Description
C.8.13.2.1.1. Content Qualification
C.8.13.2.1.2. Evidence Sequence Attributes
C.8.13.3. MR Image Description Macro
C.8.13.3.1. MR Image Description Attribute Description
C.8.13.3.1.1. Complex Image Component
C.8.13.3.1.2. Acquisition Contrast
C.8.13.4. MR Pulse Sequence Module
C.8.13.5. Enhanced MR Image Functional Group Macros
C.8.13.5.1. MR Image Frame Type Macro
C.8.13.5.2. MR Timing and Related Parameters Macro
C.8.13.5.2.1. RF Echo Train Length and Gradient Echo Train Length Attributes Usage
C.8.13.5.3. MR FOV/Geometry Macro
C.8.13.5.4. MR Echo Macro
C.8.13.5.5. MR Modifier Macro
C.8.13.5.6. MR Imaging Modifier Macro
C.8.13.5.7. MR Receive Coil Macro
C.8.13.5.8. MR Transmit Coil Macro
C.8.13.5.9. MR Diffusion Macro
C.8.13.5.10. MR Averages Macro
C.8.13.5.11. MR Spatial Saturation Macro
C.8.13.5.12. MR Metabolite Map Macro
C.8.13.5.13. MR Velocity Encoding Macro
C.8.13.5.13.1. Velocity Encoding Direction
C.8.13.5.14. MR Arterial Spin Labeling Macro
C.8.13.5.14.1. ASL Context
C.8.13.5.14.2. ASL Crusher Flag
C.8.13.5.14.3. Relationship of ASL Timing Attributes
C.8.13.5.15. Functional MR Macro
C.8.13.5.15.1. Functional Sync Pulse
C.8.13.6. MR Series Module
C.8.14. MR Spectroscopy Modules
C.8.14.1. MR Spectroscopy Module
C.8.14.1.1. MR Spectroscopy Attribute Multiplicity Ordering
C.8.14.1.2. MR Spectroscopy Zero Fill Explanation
C.8.14.1.3. MR Spectroscopy Water Reference Data Clarification
C.8.14.1.4. Water Reference Acquisition
C.8.14.2. MR Spectroscopy Pulse Sequence Module
C.8.14.3. MR Spectroscopy Functional Group Macros
C.8.14.3.1. MR Spectroscopy Frame Type Macro
C.8.14.3.2. MR Spectroscopy FOV/Geometry Macro
C.8.14.4. MR Spectroscopy Data Module
C.8.14.4.1. Spectroscopy Data
C.8.14.5. MR Spectroscopy Description Macro
C.8.14.5.1. MR Spectroscopy Description Attribute Description
C.8.14.5.1.1. Image Type and Frame Type
C.8.14.5.1.1.1. Pixel Data Characteristics
C.8.14.5.1.1.2. Patient Examination Characteristics
C.8.14.5.1.1.3. Image Flavor
C.8.14.5.1.1.4. Derived Pixel Contrast
C.8.14.5.1.2. Volumetric Properties
C.8.14.5.1.3. Volume Based Calculation Technique Attribute
C.8.14.5.1.4. Complex Image Component
C.8.14.5.1.5. Acquisition Contrast
C.8.15. Enhanced CT Image
C.8.15.1. CT Series Module
C.8.15.2. Enhanced CT Image Module
C.8.15.2.1. CT Image Description Attribute Description
C.8.15.2.1.1. Image Type and Frame Type
C.8.15.2.1.1.1. Pixel Data Characteristics
C.8.15.2.1.1.2. Patient Examination Characteristics
C.8.15.2.1.1.3. Image Flavor
C.8.15.2.1.1.4. Derived Pixel Contrast
C.8.15.3. Enhanced CT Image Functional Group Macros
C.8.15.3.1. CT Image Frame Type Macro
C.8.15.3.2. CT Acquisition Type Macro
C.8.15.3.2.1. Acquisition Type
C.8.15.3.3. CT Acquisition Details Macro
C.8.15.3.4. CT Table Dynamics Macro
C.8.15.3.4.1. Spiral Pitch Factor
C.8.15.3.5. CT Position Macro
C.8.15.3.6. CT Geometry Macro
C.8.15.3.6.1. Relationships Between CT Geometric Attributes (Informative)
C.8.15.3.7. CT Reconstruction Macro
C.8.15.3.8. CT Exposure Macro
C.8.15.3.8.1. Exposure Time in ms Attribute (Informative)
C.8.15.3.9. CT X-Ray Details Macro
C.8.15.3.10. CT Pixel Value Transformation Macro
C.8.15.3.11. CT Additional X-Ray Source Macro
C.8.16. Common CT, MR and US Descriptions
C.8.16.1. Image Type and Frame Type
C.8.16.1.1. Pixel Data Characteristics
C.8.16.1.2. Patient Examination Characteristics
C.8.16.1.3. Image Flavor
C.8.16.1.4. Derived Pixel Contrast
C.8.16.2. Common CT/MR Image Description Macro
C.8.16.2.1. Common CT/MR Image Description Attribute Description
C.8.16.2.1.1. Pixel Presentation
C.8.16.2.1.1.1. Supplemental Palette Color LUTs
C.8.16.2.1.2. Volumetric Properties
C.8.16.2.1.3. Volume Based Calculation Technique Attribute
C.8.17. Ophthalmic Photography and Tomography Modules
C.8.17.1. Ophthalmic Photography Series Module
C.8.17.2. Ophthalmic Photography Image Module
C.8.17.2.1. Ophthalmic Photography Image Module Attribute Descriptions
C.8.17.2.1.1. Referenced Image Sequence
C.8.17.2.1.2. Samples Per Pixel and Samples Per Pixel Used
C.8.17.2.1.3. Photometric Interpretation
C.8.17.2.1.4. Image Type
C.8.17.3. Ophthalmic Photographic Parameters Module
C.8.17.4. Ophthalmic Photography Acquisition Parameters Module
C.8.17.5. Ocular Region Imaged Module
C.8.17.6. Ophthalmic Tomography Series Module
C.8.17.7. Ophthalmic Tomography Image Module
C.8.17.8. Ophthalmic Tomography Acquisition Parameters Module
C.8.17.9. Ophthalmic Tomography Parameters Module
C.8.17.10. Ophthalmic Tomography Functional Group Macros
C.8.17.10.1. Ophthalmic Frame Location Macro
C.8.17.10.1.1. Reference Coordinates
C.8.17.11. Wide Field Ophthalmic Photography Stereographic Projection Module
C.8.17.11.1. Wide Field Ophthalmic Photography Stereographic Projection Attribute Descriptions
C.8.17.11.1.1. Center Pixel View Angle
C.8.17.12. Wide Field Ophthalmic Photography 3D Coordinates Module
C.8.17.12.1. Wide Field Ophthalmic Photography 3D Coordinates Attribute Descriptions
C.8.17.12.1.1. Transformation Method Code Sequence
C.8.17.12.1.2. Two Dimensional to Three Dimensional Map Data
C.8.17.13. Wide Field Ophthalmic Photography Quality Rating Module
C.8.18. Stereometric Modules
C.8.18.1. Stereometric Series Module
C.8.18.2. Stereometric Relationship Module
C.8.18.2.1. Stereometric Relationship Module Attribute Descriptions
C.8.18.2.1.1. Left and Right Image Sequences
C.8.19. Enhanced XA/XRF Image
C.8.19.1. XA/XRF Series Module
C.8.19.2. Enhanced XA/XRF Image Module
C.8.19.2.1. Enhanced XA/XRF Image Module Attribute Description
C.8.19.2.1.1. Image Type and Frame Type
C.8.19.2.1.1.1. Pixel Data Characteristics
C.8.19.2.1.1.2. Patient Examination Characteristics
C.8.19.2.1.1.3. Image Flavor
C.8.19.2.1.1.4. Derived Pixel Contrast
C.8.19.2.1.2. Bits Allocated and Bits Stored
C.8.19.2.1.3. Planes in Acquisition
C.8.19.3. XA/XRF Acquisition Module
C.8.19.4. X-Ray Image Intensifier Module
C.8.19.5. X-Ray Detector Module
C.8.19.6. Enhanced XA/XRF Image Functional Group Macros
C.8.19.6.1. X-Ray Frame Characteristics Macro
C.8.19.6.2. X-Ray Field of View Macro
C.8.19.6.3. X-Ray Exposure Control Sensing Regions Macro
C.8.19.6.3.1. X-Ray Exposure Control Sensing Regions Attributes
C.8.19.6.4. X-Ray Frame Pixel Data Properties Macro
C.8.19.6.4.1. X-Ray Frame Pixel Data Properties Attributes
C.8.19.6.4.1.1. Pixel Intensity Relationship
C.8.19.6.4.1.2. Imager Pixel Spacing
C.8.19.6.4.1.3. Pixel Data Area Relative to FOV
C.8.19.6.5. X-Ray Frame Detector Parameters Macro
C.8.19.6.6. X-Ray Calibration Device Usage Macro
C.8.19.6.7. X-Ray Object Thickness Macro
C.8.19.6.8. X-Ray Frame Acquisition Macro
C.8.19.6.8.1. X-Ray Frame Acquisition Sequence Attributes
C.8.19.6.9. X-Ray Projection Pixel Calibration Macro
C.8.19.6.9.1. Projection Calibration Method
C.8.19.6.9.2. Object Pixel Spacing in Center of Beam
C.8.19.6.10. X-Ray Positioner Macro
C.8.19.6.11. X-Ray Table Position Macro
C.8.19.6.11.1. X-Ray Table Position Macro Attribute Description
C.8.19.6.12. X-Ray Collimator Macro
C.8.19.6.12.1. X-Ray Collimator Attributes
C.8.19.6.13. X-Ray Isocenter Reference System Macro
C.8.19.6.13.1. Isocenter Reference System Attribute Description
C.8.19.6.13.1.1. Isocenter Coordinate System
C.8.19.6.13.1.2. Positioner Coordinate System
C.8.19.6.13.1.3. Table Coordinate System
C.8.19.6.13.2. Relationship Patient Coordinate System
C.8.19.6.14. X-Ray Geometry Macro
C.8.19.7. XA/XRF Multi-frame Presentation Module
C.8.19.7.1. Multi-frame Presentation Attributes
C.8.19.7.1.1. Mask Visibility Percentage (Informative)
C.8.20. Segmentation
C.8.20.1. Segmentation Series Module
C.8.20.2. Segmentation Image Module
C.8.20.2.1. Bits Allocated and Bits Stored
C.8.20.2.2. Lossy Image Compression and Lossy Image Compression Method
C.8.20.2.3. Segmentation Type and Segmentation Fractional Type
C.8.20.2.4. Segment Number
C.8.20.3. Segmentation Functional Group Macros
C.8.20.3.1. Segmentation Macro
C.8.20.4. Segmentation Macros
C.8.20.4.1. Segment Description Macro
C.8.21. X-Ray 3D
C.8.21.1. X-Ray 3D Image Module
C.8.21.1.1. X-Ray 3D Image Module Attribute Description
C.8.21.1.1.1. Image Type and Frame Type
C.8.21.1.1.1.1. Pixel Data Characteristics
C.8.21.1.1.1.2. Patient Examination Characteristics
C.8.21.1.1.1.3. Image Flavor
C.8.21.1.1.1.4. Derived Pixel Contrast
C.8.21.1.1.2. Volumetric Properties Attribute
C.8.21.1.1.3. Volume Based Calculation Technique Attribute
C.8.21.1.1.4. Source Irradiation Event Sequence
C.8.21.2. X-Ray 3D Contributing Image Sources Modules
C.8.21.2.1. X-Ray 3D Angiographic Image Contributing Sources Module
C.8.21.2.2. X-Ray 3D Craniofacial Image Contributing Sources Module
C.8.21.2.3. Breast Tomosynthesis Contributing Sources Module
C.8.21.3. X-Ray 3D Acquisition Modules
C.8.21.3.1. X-Ray 3D General Acquisition Macros
C.8.21.3.1.1. X-Ray 3D General Shared Acquisition Macro
C.8.21.3.1.2. X-Ray 3D General Per Projection Acquisition Macro
C.8.21.3.1.3. X-Ray 3D General Positioner Movement Macro
C.8.21.3.1.3.1. Positioner Angle Increments
C.8.21.3.2. X-Ray 3D Angiographic Acquisition Module
C.8.21.3.3. X-Ray 3D Craniofacial Acquisition Module
C.8.21.3.4. Breast Tomosynthesis Acquisition Module
C.8.21.4. X-Ray 3D Reconstruction Module
C.8.21.5. X-Ray 3D Image Functional Group Macros
C.8.21.5.1. X-Ray 3D Frame Type Macro
C.8.21.6. Breast View Module
C.8.21.6.1. Breast View Module Attribute Description
C.8.21.6.1.1. Image Type and Frame Type
C.8.22. Enhanced PET Modules
C.8.22.1. Enhanced PET Series Module
C.8.22.2. Enhanced PET Acquisition Module
C.8.22.2.1. Enhanced PET Acquisition Module Attribute Descriptions
C.8.22.2.1.1. Acquisition Start Condition
C.8.22.2.1.2. Acquisition Termination Condition
C.8.22.2.1.3. Energy Window Range Sequence
C.8.22.3. Enhanced PET Image Module
C.8.22.3.1. Enhanced PET Image Description Attribute Description
C.8.22.3.1.1. Image Type and Frame Type
C.8.22.3.1.1.1. Pixel Data Characteristics
C.8.22.3.1.1.2. Patient Examination Characteristics
C.8.22.3.1.1.3. Image Flavor
C.8.22.3.1.1.4. Derived Pixel Contrast
C.8.22.3.1.2. Attenuation Correction Source
C.8.22.3.1.3. Attenuation Correction Temporal Relationship
C.8.22.4. Enhanced PET Isotope Module
C.8.22.5. Enhanced PET Image Functional Group Macros
C.8.22.5.1. PET Frame Type Macro
C.8.22.5.2. PET Frame Acquisition Macro
C.8.22.5.3. PET Detector Motion Details Macro
C.8.22.5.4. PET Position Macro
C.8.22.5.4.1. Relationships Between PET Geometric Attributes (Informative)
C.8.22.5.5. PET Frame Correction Factors Macro
C.8.22.5.6. PET Reconstruction Macro
C.8.22.5.7. PET Table Dynamics Macro
C.8.22.6. Enhanced PET Corrections Module
C.8.23. Surface Segmentation
C.8.23.1. Surface Segmentation Module
C.8.24. Enhanced US Volume Modules and Functional Group Macros
C.8.24.1. Enhanced US Series Module
C.8.24.2. Ultrasound Frame of Reference Module
C.8.24.2.1. Ultrasound Frame of Reference Module Overview
C.8.24.2.1.1. Volume Frame of Reference
C.8.24.2.1.2. Transducer Frame of Reference
C.8.24.2.1.3. Table Frame of Reference
C.8.24.2.2. Ultrasound Frame of Reference Module Attributes
C.8.24.2.2.1. Volume to Transducer Mapping Matrix
C.8.24.2.2.2. Volume to Table Mapping Matrix
C.8.24.3. Enhanced US Image Module
C.8.24.3.1. Pixel Representation
C.8.24.3.2. Image Type
C.8.24.3.2.1. Pixel Data Characteristics
C.8.24.3.2.2. Patient Examination Characteristics
C.8.24.3.2.3. Image Flavor
C.8.24.3.2.4. Derived Pixel Contrast
C.8.24.3.3. Dimension Organization Type
C.8.24.3.4. Transducer Description
C.8.24.4. IVUS Image Module
C.8.24.5. Excluded Intervals Module
C.8.24.6. Enhanced US Volume Functional Group Macros
C.8.24.6.1. US Image Description Macro
C.8.25. Ophthalmic Refractive Measurements Modules
C.8.25.1. Lensometry Measurements Series Module
C.8.25.2. Autorefraction Measurements Series Module
C.8.25.3. Keratometry Measurements Series Module
C.8.25.4. Subjective Refraction Measurements Series Module
C.8.25.5. Visual Acuity Measurements Series Module
C.8.25.6. General Ophthalmic Refractive Macros
C.8.25.6.1. Cylinder Sequence Macro
C.8.25.6.2. Prism Sequence Macro
C.8.25.7. General Ophthalmic Refractive Measurements Module
C.8.25.7.1. General Ophthalmic Refractive Measurements Module Attributes
C.8.25.7.1.1. Referenced Refractive Measurements Sequence
C.8.25.8. Lensometry Measurements Module
C.8.25.9. Autorefraction Measurements Module
C.8.25.10. Keratometry Measurements Module
C.8.25.11. Subjective Refraction Measurements Module
C.8.25.12. Visual Acuity Measurements Module
C.8.25.13. Ophthalmic Axial Measurements Series Module
C.8.25.14. Ophthalmic Axial Measurements Module
C.8.25.14.1. Ophthalmic Axial Measurements Macro
C.8.25.14.1.1. Ophthalmic Axial Measurements Macro Attribute Descriptions
C.8.25.14.1.1.1. Lens Status Code Sequence
C.8.25.14.1.1.2. Vitreous Status Code Sequence
C.8.25.14.1.1.3. Ophthalmic Axial Length
C.8.25.14.1.1.4. Ophthalmic Axial Length Measurements Segment Name Code Sequence
C.8.25.14.1.1.5. Observer Type
C.8.25.14.1.1.6. Referenced Ophthalmic Axial Length Measurement QC Image Sequence
C.8.25.14.2. Ophthalmic Axial Length Segmental Measurements Macro
C.8.25.14.3. Ophthalmic Axial Measurements Related Information Macro
C.8.25.14.4. Ophthalmic Axial Measurements Selected Macro
C.8.25.14.5. Ophthalmic Axial Measurements Quality Metric Macro
C.8.25.15. Intraocular Lens Calculations Series Module
C.8.25.16. Intraocular Lens Calculations Module
C.8.25.16.1. Intraocular Lens Calculations Module Attributes
C.8.25.16.1.1. Referenced SOP Sequence
C.8.25.16.2. Intraocular Lens Calculations Macro
C.8.25.16.3. Keratometry Macro
C.8.25.16.4. IOL Ophthalmic Axial Length Macro
C.8.25.16.5. Calculated IOL Macro
C.8.25.16.6. Ophthalmic Axial Measurements Quality Image SOP Instance Reference Macro
C.8.26. Visual Field Static Perimetry Measurements Modules
C.8.26.1. Visual Field Static Perimetry Measurements Series Module
C.8.26.2. Visual Field Static Perimetry Test Parameters Module
C.8.26.3. Visual Field Static Perimetry Test Reliability Module
C.8.26.3.1. Ophthalmic Visual Field Global Index Macro
C.8.26.4. Visual Field Static Perimetry Test Measurements Module
C.8.26.4.1. Visual Field Static Perimetry Test Measurement Attribute Descriptions
C.8.26.4.1.1. Measurement Laterality
C.8.26.4.1.2. Foveal Sensitivity
C.8.26.4.1.3. Blind Spot X and Y Coordinates
C.8.26.5. Visual Field Static Perimetry Test Results Module
C.8.26.6. Ophthalmic Patient Clinical Information and Test Lens Parameters Module
C.8.26.6.1. Ophthalmic Patient Clinical Information and Test Lens Parameters Macro
C.8.27. Intravascular Optical Coherence Tomography Modules
C.8.27.1. Intravascular OCT Series Module
C.8.27.1.1. Intravascular OCT Series Attribute Descriptions
C.8.27.1.1.1. Presentation Intent Type
C.8.27.2. Intravascular OCT Image Module
C.8.27.2.1. Image Type and Frame Type
C.8.27.2.2. Volumetric Properties
C.8.27.2.3. Pixel Presentation
C.8.27.2.4. Bits Allocated, Bits Stored and Samples Per Pixel
C.8.27.3. Intravascular OCT Acquisition Parameters Module
C.8.27.4. Intravascular OCT Processing Parameters Module
C.8.27.4.1. Refractive Index Applied
C.8.27.4.2. Pixel Intensity Relationship
C.8.27.5. Intravascular Image Acquisition Parameters Module
C.8.27.5.1. IVUS Acquisition
C.8.27.5.2. IVUS Pullback Rate
C.8.27.6. Intravascular OCT Functional Group Macros
C.8.27.6.1. Intravascular OCT Frame Type Macro
C.8.27.6.2. Intravascular Frame Content Macro
C.8.27.6.3. Intravascular OCT Frame Content Macro
C.8.27.6.3.1. OCT Z Offset Correction
C.8.27.6.3.2. A-line Padding
C.8.28. Ophthalmic Thickness Map Modules
C.8.28.1. Ophthalmic Thickness Map Series Module
C.8.28.2. Ophthalmic Thickness Map Module
C.8.28.2.1. Ophthalmic Thickness Map Module Attribute Descriptions
C.8.28.2.1.1. Image Type
C.8.28.2.1.2. Referenced Instance Sequence
C.8.28.2.1.3. Ophthalmic Mapping Device Type
C.8.28.3. Ophthalmic Thickness Map Quality Rating Module
C.8.29. Optical Surface Scanner Modules
C.8.29.1. Optical Surface Scanner Series Module
C.8.29.2. Scan Procedure Module
C.8.29.2.1. Scan Procedure Module Attribute Descriptions
C.8.29.2.1.1. Shot Duration Time and Shot Offset Time
C.8.30. Corneal Topography Map Modules
C.8.30.1. Corneal Topography Map Series Module
C.8.30.2. Corneal Topography Map Image Module
C.8.30.2.1. Corneal Topography Map Image Module Attribute Descriptions
C.8.30.2.1.1. Image Type
C.8.30.2.1.2. Corneal Topography Map Real World Value and Image Transformations
C.8.30.3. Corneal Topography Map Analysis Module
C.8.30.3.1. Corneal Topography Map Analysis Module Attribute Descriptions
C.8.30.3.1.1. Ophthalmic Mapping Device Type
C.8.30.3.1.2. Source Image Sequence
C.8.30.3.1.3. Source Image Corneal Processed Data Sequence
C.8.30.3.1.4. Corneal Vertex Location
C.8.31. Breast Projection Image Modules
C.8.31.1. Enhanced Mammography Image Module
C.8.31.1.1. Positioner and Detector Motion Attribute Descriptions
C.8.31.2. Breast X-Ray Positioner Macro
C.8.31.2.1. Breast X-Ray Positioner Attribute Descriptions
C.8.31.3. Breast X-Ray Detector Macro
C.8.31.4. Breast X-Ray Geometry Macro
C.8.31.5. Breast X-Ray Acquisition Dose Macro
C.8.31.6. Breast X-Ray Isocenter Reference System Macro
C.8.31.6.1. Isocenter Reference System Attribute Descriptions
C.8.31.6.1.1. Isocenter Coordinate System
C.8.31.6.1.2. X-Ray Source Coordinate System
C.8.31.6.1.3. Breast Support Coordinate System
C.8.31.6.1.4. Detector Coordinate System
C.8.31.6.1.5. Detector Active Area Relationship
C.8.31.7. X-Ray Grid Macro
C.8.31.8. X-Ray Filter Macro
C.8.32. Parametric Map
C.8.32.1. Parametric Map Series Module
C.8.32.2. Parametric Map Image Module
C.8.32.2.1. Lossy Image Compression and Lossy Image Compression Method
C.8.32.3. Parametric Map Functional Group Macros
C.8.32.3.1. Parametric Map Frame Type Macro
C.8.33. Tractography Results Modules
C.8.33.1. Tractography Results Series Module
C.8.33.2. Tractography Results Module
C.8.33.2.1. Tractography Results Module Attributes
C.8.33.2.2. Acquisition, Model and Algorithm Attributes
C.8.33.3. Summary Statistics Macro
C.9. Overlays
C.9.1. Overlay Identification Module
C.9.2. Overlay Plane Module
C.9.2.1. Overlay Attribute Descriptions
C.9.2.1.1. Overlay Type
C.9.2.1.2. ROI Area, ROI Mean and ROI Standard Deviation
C.9.2.1.3. Overlay Subtype
C.9.3. Multi-frame Overlay Module
C.9.3.1. Multi-frame Overlay Attribute Descriptions
C.9.3.1.1. Number of Frames in Overlay
C.9.4. Bi-plane Overlay Module (Retired)
C.9.5. Basic Print Image Overlay Box Module (Retired)
C.10. Curve, Graphic and Waveform
C.10.1. Curve Identification Module (Retired)
C.10.2. Curve Module (Retired)
C.10.3. Audio Module (Retired)
C.10.4. Displayed Area Module
C.10.5. Graphic Annotation Module
C.10.5.1. Graphic Annotation Attribute Descriptions
C.10.5.1.1. Unformatted Text Value
C.10.5.1.2. Graphic Data and Graphic Type
C.10.5.1.3. Compound Graphic Sequence
C.10.5.1.3.1. Compound Graphic Instance ID
C.10.5.1.3.2. Rotation
C.10.5.1.3.3. Ellipses
C.10.5.1.3.4. Rectangles
C.10.5.1.3.5. Multi-lines
C.10.5.1.3.6. Cut and Infinite-lines
C.10.5.1.3.7. Range Lines
C.10.5.1.3.8. Ruler
C.10.5.1.3.9. Axis
C.10.5.1.3.10. Crosshairs
C.10.5.1.3.11. Arrows
C.10.5.1.3.12. Text Style Sequence
C.10.5.1.3.12.1. Text Alignment
C.10.5.1.3.13. Line Style Sequence
C.10.5.1.3.13.1. Line Dashing Style
C.10.5.1.3.13.2. Shadows
C.10.5.1.3.14. Fill Style Sequence
C.10.5.1.3.14.1. Fill Mode
C.10.6. Spatial Transformation Module
C.10.7. Graphic Layer Module
C.10.7.1. Graphic Layer Module Attributes
C.10.7.1.1. Encoding of CIELab Values
C.10.8. Waveform Identification Module
C.10.9. Waveform Module
C.10.9.1. Waveform Attribute Descriptions
C.10.9.1.1. Multiplex Group Time Offset
C.10.9.1.2. Trigger Sample Position
C.10.9.1.3. Waveform Originality
C.10.9.1.4. Channel Definition Sequence
C.10.9.1.4.1. Channel Source and Modifiers
C.10.9.1.4.2. Channel Sensitivity and Channel Sensitivity Units
C.10.9.1.4.3. Channel Skew and Channel Offset
C.10.9.1.4.4. Waveform Bits Stored
C.10.9.1.4.5. Channel Minimum and Maximum Value
C.10.9.1.5. Waveform Bits Allocated and Waveform Sample Interpretation
C.10.9.1.6. Waveform Padding Value
C.10.9.1.7. Waveform Data
C.10.9.1.8. Waveform Data Display Scale
C.10.9.1.9. Channel Position
C.10.9.1.10. Channel Display Scale
C.10.10. Waveform Annotation Module
C.10.10.1. Waveform Annotation Attribute Descriptions
C.10.10.1.1. Referenced Channels
C.10.10.1.2. Temporal Range Type
C.10.10.1.3. Referenced Sample Positions
C.10.10.1.4. Annotation Group Number
C.10.11. Graphic Group Module
C.11. Look Up Tables and Presentation States
C.11.1. Modality LUT Module
C.11.1.1. LUT Attribute Descriptions
C.11.1.1.1. LUT Descriptor
C.11.1.1.2. Modality LUT and Rescale Type
C.11.2. VOI LUT Module
C.11.2.1. LUT Attribute Descriptions
C.11.2.1.1. LUT Descriptor
C.11.2.1.2. Window Center and Window Width
C.11.2.1.2.1. Default LINEAR Function
C.11.2.1.2.2. General Requirements for Window Center and Window Width
C.11.2.1.3. VOI LUT Function
C.11.2.1.3.1. SIGMOID Function
C.11.2.1.3.2. LINEAR_EXACT Function
C.11.3. LUT Identification Module
C.11.4. Presentation LUT Module
C.11.4.1. LUT Descriptor
C.11.5. Image Histogram Module
C.11.5.1. Image Histogram Attribute Descriptions
C.11.6. Softcopy Presentation LUT Module
C.11.6.1. Softcopy Presentation LUT Attributes
C.11.6.1.1. LUT Descriptor
C.11.6.1.2. Presentation LUT Shape
C.11.7. Overlay Activation Module
C.11.8. Softcopy VOI LUT Module
C.11.9. Presentation Series Module
C.11.10. Presentation State Identification Module
C.11.11. Presentation State Relationship Module
C.11.12. Presentation State Shutter Module
C.11.13. Presentation State Mask Module
C.11.14. Presentation State Blending Module
C.11.14.1. Presentation State Blending Module Attributes
C.11.14.1.1. Blending Sequence
C.11.15. ICC Profile Module
C.11.15.1. Attribute Descriptions
C.11.15.1.1. ICC Profile
C.11.15.1.2. Color Space
C.11.16. Structured Display Module
C.11.17. Structured Display Image Box Module
C.11.17.1. Structured Display Image Box Attribute Descriptions
C.11.17.1.1. Display Environment Spatial Position
C.11.17.1.2. Referenced Image Sequence and Referenced Presentation State Sequence
C.11.17.1.3. Referenced Instance Sequence
C.11.17.1.4. Referenced Stereometric Instance Sequence
C.11.17.1.5. Type of Synchronization
C.11.18. Structured Display Annotation Module
C.11.18.1. Structured Display Annotation Attribute Descriptions
C.11.18.1.1. Unformatted Text Value and Display Environment Spatial Position
C.11.19. XA/XRF Presentation State Mask Module
C.11.19.1. XA/XRF Presentation State Mask Attributes
C.11.19.1.1. Mask Sub-pixel Shift
C.11.19.1.2. Vertices of the Region
C.11.20. XA/XRF Presentation State Shutter Module
C.11.21. XA/XRF Presentation State Presentation Module
C.11.22. Volumetric Presentation State Identification Module
C.11.22.1. Presentation Display Collection UID
C.11.22.2. Presentation Sequence Collection UID
C.11.23. Volumetric Presentation State Relationship Module
C.11.23.1. Presentation Input Type Volume Input Requirements
C.11.23.2. Input Sequence Position Index
C.11.23.3. Volumetric Presentation State Reference Coordinate System
C.11.24. Volume Cropping Module
C.11.24.1. Volume Cropping
C.11.25. Presentation View Description Module
C.11.26. Multi-Planar Reconstruction Geometry Module
C.11.26.1. Multi-Planar Reconstruction Styles
C.11.26.1.1. Planar Style
C.11.27. Volumetric Presentation State Display Module
C.11.27.1. RGB LUT Transfer Function
C.11.27.2. Classification Component Usage
C.11.27.3. Compositor Component Usage
C.11.27.4. Weighting Transfer Function
C.11.27.5. Use of Segmented Palette Color Lookup Tables
C.11.28. Volumetric Graphic Annotation Module
C.11.28.1. Annotation Clipping
C.11.28.2. Referenced Structured Content Sequence
C.11.29. Presentation Animation Module
C.11.29.1. Presentation Animation Style
C.12. General Modules
C.12.1. SOP Common Module
C.12.1.1. SOP Common Attribute Descriptions
C.12.1.1.1. SOP Class UID, SOP Instance UID
C.12.1.1.2. Specific Character Set
C.12.1.1.3. Digital Signatures Macro
C.12.1.1.3.1. Digital Signature Attribute Descriptions
C.12.1.1.3.1.1. Data Elements Signed
C.12.1.1.3.1.2. Signature
C.12.1.1.3.1.3. Certified Timestamp
C.12.1.1.4. Encrypted Attribute Descriptions
C.12.1.1.4.1. Encrypted Attributes Sequence
C.12.1.1.4.2. Encrypted Content
C.12.1.1.5. Contributing Equipment Sequence
C.12.1.1.6. HL7 Structured Document Reference Sequence
C.12.2. Common Instance Reference Module
C.12.3. Frame Extraction Module
C.13. Print Management Specific Modules
C.13.1. Basic Film Session Presentation Module
C.13.2. Basic Film Session Relationship Module
C.13.3. Basic Film Box Presentation Module
C.13.3.1. Image Display Format
C.13.3.1.1. Standard Image Display Format
C.13.3.1.2. Row Symmetric Image Display Format
C.13.3.1.3. Column Symmetric Image Display Format
C.13.4. Basic Film Box Relationship Module
C.13.5. Image Box Pixel Presentation Module
C.13.5.1. Image Box Position
C.13.6. Image Box Relationship Module (Retired)
C.13.7. Basic Annotation Presentation Module
C.13.8. Print Job Module
C.13.9. Printer Module
C.13.9.1. Printer Status Info and Execution Status Info
C.13.10. Image Overlay Box Presentation Module (Retired)
C.13.11. Image Overlay Box Relationship Module (Retired)
C.13.12. Print Request Module
C.13.13. Printer Configuration Module
C.14. Storage Commitment Module
C.14.1. Storage Commitment Attribute Description
C.14.1.1. Failure Reason
C.15. Queue Management Specific Modules
C.16. Stored Print Specific Modules
C.17. SR Document Modules
C.17.1. SR Document Series Module
C.17.2. SR Document General Module
C.17.2.1. Hierarchical SOP Instance Reference Macro
C.17.2.2. Identical Documents Sequence
C.17.2.3. Current Requested Procedure Evidence Sequence and Pertinent Other Evidence Sequence
C.17.2.4. Identified Person or Device Macro
C.17.2.5. Verifying Observer, Author Observer and Participant Sequences
C.17.2.6. Equivalent CDA Document
C.17.2.7. Completion Flag
C.17.3. SR Document Content Module
C.17.3.1. SR Document Content Tree
C.17.3.2. Content Item Attributes
C.17.3.2.1. Content Item Value Type
C.17.3.2.2. Concept Name Code Sequence
C.17.3.2.3. Continuity of Content
C.17.3.2.4. Content Sequence and Relationship Type
C.17.3.2.5. Referenced Content Item Identifier
C.17.4. SR Content Tree Example (Informative)
C.17.5. Observation Context Encoding
C.17.6. Key Object Selection Modules
C.17.6.1. Key Object Document Series Module
C.17.6.2. Key Object Document Module
C.17.6.2.1. Identical Documents
C.18. Content Macros
C.18.1. Numeric Measurement Macro
C.18.2. Code Macro
C.18.3. Composite Object Reference Macro
C.18.4. Image Reference Macro
C.18.5. Waveform Reference Macro
C.18.5.1. Waveform Reference Macro Attribute Descriptions
C.18.5.1.1. Referenced Waveform Channels
C.18.6. Spatial Coordinates Macro
C.18.6.1. Spatial Coordinates Macro Attribute Descriptions
C.18.6.1.1. Graphic Data
C.18.6.1.2. Graphic Type
C.18.7. Temporal Coordinates Macro
C.18.7.1. Temporal Coordinates Macro Attribute Descriptions
C.18.7.1.1. Temporal Range Type
C.18.8. Container Macro
C.18.8.1. Container Macro Attribute Descriptions
C.18.8.1.1. Continuity of Content
C.18.8.1.2. Content Template Sequence
C.18.9. 3D Spatial Coordinates Macro
C.18.9.1. 3D Spatial Coordinates Macro Attribute Descriptions
C.18.9.1.1. Graphic Data
C.18.9.1.2. Graphic Type
C.19. Raw Data Specific Modules
C.19.1. Raw Data Module
C.19.1.1. Raw Data
C.20. Spatial Registration
C.20.1. Spatial Registration Series Module
C.20.2. Spatial Registration Module
C.20.2.1. Registration Module Attribute Descriptions
C.20.2.1.1. Frame of Reference Transformation Matrix
C.20.2.1.2. Frame of Reference Transformation Matrix Type
C.20.3. Deformable Spatial Registration Module
C.20.3.1. Deformable Spatial Registration Module Attribute Descriptions
C.20.3.1.1. Deformable Registration Sequence Application
C.20.3.1.2. Deformable Registration Grid Sequence
C.20.3.1.3. Vector Grid Data
C.21. Spatial Fiducials
C.21.1. Spatial Fiducials Series Module
C.21.2. Spatial Fiducials Module
C.21.2.1. Spatial Fiducials Module Attribute Descriptions
C.21.2.1.1. Shape Type
C.21.2.1.2. Contour Data
C.21.2.1.3. Contour Uncertainty Radius
C.22. Media Creation Management Specific Modules
C.22.1. Media Creation Management Module
C.22.1.1. Barcode Symbology
C.22.1.2. Execution Status
C.22.1.3. Execution Status Info
C.22.1.4. Failure Reason
C.23. Hanging Protocol Specific Modules
C.23.1. Hanging Protocol Definition Module
C.23.1.1. Attribute Descriptions
C.23.1.1.1. Hanging Protocol Definition Sequence Attributes
C.23.1.1.2. Image Sets Sequence
C.23.1.1.3. Image Set Selector Sequence Attributes
C.23.2. Hanging Protocol Environment Module
C.23.2.1. Hanging Protocol Environment Attribute Descriptions
C.23.2.1.1. Display Environment Spatial Position
C.23.3. Hanging Protocol Display Module
C.23.3.1. Attribute Descriptions
C.23.3.1.1. Filter Operations Sequence
C.23.3.1.2. Sorting Operations Sequence
C.23.3.1.3. Blending Operation Type
C.23.3.1.4. Presentation Intent Attributes
C.23.4. Hanging Protocol Selector Attribute Macros
C.23.4.1. Hanging Protocol Selector Attribute Context Macro
C.23.4.1.1. Hanging Protocol Selector Attribute Context Macro Attribute Descriptions
C.23.4.1.1.1. Selector Sequence Pointer
C.23.4.1.1.2. Functional Group Pointer
C.23.4.1.1.3. Private Attribute References
C.23.4.2. Hanging Protocol Selector Attribute Value Macro
C.23.4.2.1. Hanging Protocol Selector Attribute Value Macro Attribute Descriptions
C.23.4.2.1.1. Selector Attribute Value Matching
C.23.4.2.1.2. Selector Code Sequence Value
C.24. Encapsulated Document Modules
C.24.1. Encapsulated Document Series Module
C.24.2. Encapsulated Document Module
C.24.2.1. Attribute Requirements for Encapsulated CDA Document
C.25. Real World Value Mapping Modules
C.25.1. Real World Value Mapping Series Module
C.25.2. Real World Value Mapping Module
C.26. Substance Administration Modules
C.26.1. Product Characteristics Module
C.26.2. Substance Approval Module
C.26.3. Substance Administration Module
C.26.4. Substance Administration Log Module
C.27. Common Surface IE Modules
C.27.1. Surface Mesh Module
C.27.1.1. Surface Mesh Module Attribute Descriptions
C.27.1.1.1. Surface Sequence
C.27.1.1.2. Surface Processing
C.27.1.1.3. Recommended Presentation
C.27.1.1.4. Finite Volume
C.27.1.1.5. Manifold
C.27.1.1.6. Surface Points Normals Sequence
C.27.2. Points Macro
C.27.2.1. Points Macro Attribute Descriptions
C.27.2.1.1. Point Coordinates Data
C.27.3. Vectors Macro
C.27.3.1. Vectors Macro Attribute Descriptions
C.27.4. Surface Mesh Primitives Macro
C.27.4.1. Surface Mesh Primitives Macro Attribute Descriptions
C.27.5. Point Cloud Module
C.27.6. UV Mapping Module
C.27.6.1. UV Mapping Attribute Description
C.27.6.1.1. U Value Data and V Value Data
C.28. Color Palette Modules
C.28.1. Color Palette Definition Module
C.28.1.1. Attribute Descriptions
C.28.1.1.1. Content Identification
C.29. Implant Modules
C.29.1. Generic Implant Template Modules
C.29.1.1. Generic Implant Template Description Module
C.29.1.1.1. Generic Implant Template Description Attribute Descriptions
C.29.1.1.1.1. Implant Size
C.29.1.1.1.2. Frame of Reference
C.29.1.2. Generic Implant Template 2D Drawings Module
C.29.1.2.1. Generic Implant Template 2D Drawings Attribute Descriptions
C.29.1.2.1.1. HPGL Document Scaling and Coordinate System
C.29.1.2.1.2. HPGL Document
C.29.1.2.1.2.1. Scope
C.29.1.2.1.2.2. Syntax
C.29.1.3. Generic Implant Template 3D Models Module
C.29.1.3.1. Generic Implant Template 3D Models Attribute Descriptions
C.29.1.3.1.1. Surface Model Scaling Factor
C.29.1.4. Generic Implant Template Mating Features Module
C.29.1.4.1. Generic Implant Template 2D Drawings Attribute Descriptions
C.29.1.4.1.1. Mating Features
C.29.1.4.1.2. Degrees of Freedom
C.29.1.5. Generic Implant Template Planning Landmarks Module
C.29.1.5.1. Planning Landmark Macros
C.29.1.5.1.1. Planning Landmark Macros Attribute Descriptions
C.29.1.5.1.1.1. 2D Coordinates Sequences
C.29.2. Implant Assembly Template
C.29.2.1. Implant Assembly Template Module
C.29.2.1.1. Implant Assembly Template Module Attribute Description
C.29.2.1.1.1. Component Assembly Sequence
C.29.3. Implant Template Group Module
C.29.3.1. Implant Template Group Module
C.29.3.1.1. Implant Template Group Module Attribute Description
C.29.3.1.1.1. Implant Template Group Member Matching Coordinates
C.29.3.1.1.2. Implant Template Groups
C.30. Unified Procedure Step Specific Modules
C.30.1. Unified Procedure Step Progress Information Module
C.30.2. Unified Procedure Step Scheduled Procedure Information Module
C.30.3. Unified Procedure Step Performed Procedure Information Module
C.30.4. Unified Procedure Step Relationship Module
C.30.4.1. Patient Identification
C.31. Radiotherapy Workflow Modules
C.31.1. RT General Machine Verification Module
C.31.1.1. Failed Parameters and Overridden Parameters
C.31.2. RT Conventional Machine Verification Module
C.31.3. RT Ion Machine Verification Module
C.31.3.1. Range Modulator Type
C.32. Display System Modules
C.32.1. Display System Module
C.32.1.1. System Status
C.32.1.2. Measurement Equipment Type
C.32.2. Target Luminance Characteristics Module
C.32.2.1. Display Function Type
C.32.3. QA Results Module
C.32.3.1. Display QA Result Context Macro
C.32.3.2. Visual Evaluation Result Macro
C.32.3.3. Luminance Uniformity Result Macro
C.32.3.4. Luminance Result Macro
D. Codes and Controlled Terminology (Informative)
E. Explanation of Patient Orientation (Normative)
F. Basic Directory Information Object Definition (Normative)
F.1. Scope of the Basic Directory Information IOD
F.2. Basic Directory IOD Overview
F.2.1. Basic Directory IOD Organization
F.2.2. Example of A Directory
F.2.2.1. Illustration of the Overall Directory Organization
F.2.2.2. Example of a DICOMDIR File Structure
F.3. Basic Directory IOD
F.3.1. Module Table
F.3.2. Modules of the Basic Directory Information Object
F.3.2.1. File-set Identification Module
F.3.2.2. Directory Information Module
F.4. Basic Directory IOD Information Model
F.5. Definition of Specific Directory Records
F.5.1. Patient Directory Record Definition
F.5.2. Study Directory Record Definition
F.5.3. Series Directory Record Definition
F.5.4. Image Directory Record Definition
F.5.5. Standalone Overlay Directory Record Definition
F.5.6. Standalone Modality LUT Directory Record Definition
F.5.7. Standalone VOI LUT Directory Record Definition
F.5.8. Standalone Curve Directory Record Definition
F.5.9. Topic Directory Record Definition
F.5.10. Visit Directory Record Definition
F.5.11. Results Directory Record Definition
F.5.12. Interpretation Directory Record Definition
F.5.13. Study Component Directory Record Definition
F.5.14. Print Queue Directory Record Definition
F.5.15. Film Session Directory Record Definition
F.5.16. Film Box Directory Record Definition
F.5.17. Basic Image Box Directory Record Definition
F.5.18. Stored Print Directory Record Definition
F.5.19. RT Dose Directory Record Definition
F.5.20. RT Structure Set Directory Record Definition
F.5.21. RT Plan Directory Record Definition
F.5.22. RT Treatment Record Directory Record Definition
F.5.23. Presentation State Directory Record Definition
F.5.24. Waveform Directory Record Definition
F.5.25. SR Document Directory Record Definition
F.5.26. Key Object Document Directory Record Definition
F.5.27. Spectroscopy Directory Record Definition
F.5.28. Raw Data Directory Record Definition
F.5.29. Registration Directory Record Definition
F.5.30. Fiducial Directory Record Definition
F.5.31. Hanging Protocol Directory Record Definition
F.5.32. Encapsulated Document Directory Record Definition
F.5.33. HL7 Structured Document Directory Record Definition
F.5.34. Real World Value Mapping Directory Record Definition
F.5.35. Stereometric Relationship Directory Record Definition
F.5.36. Palette Directory Record Definition
F.5.37. Implant Directory Record Definition
F.5.38. Implant Assembly Directory Record Definition
F.5.39. Implant Group Directory Record Definition
F.5.40. Plan Directory Record Definition
F.5.41. Measurement Directory Record Definition
F.5.42. Surface Directory Record Definition
F.5.43. Surface Scan Mesh Directory Record Definition
F.5.44. Tractography Results Directory Record Definition
F.6. Special Directory Records
F.6.1. Private Directory Record Definition
F.6.2. Multi-referenced File Directory Record Definition
F.7. Icon Image Key Definition
G. Integration of Modality Worklist and Modality Performed Procedure Step in the Original DICOM Standard (Informative)
H. Retired Composite Information Object Definitions (Normative) (Retired)
I. Retired Modules (Normative) (Retired)
J. Waveforms (Informative)
K. SR Encoding Example (Informative)
L. Mammography CAD (Informative)
M. Chest CAD (Informative)
N. Explanation of Grouping Criteria for Multi-frame Functional Group IODs (Informative)
O. Clinical Trial Identification Workflow Examples (Informative)
List of Figures
5.1-1.
Entity Convention
5.1-2.
Relationship Convention
6-1.
Major Structures of DICOM Information Model
7-1a.
DICOM Model of the Real World
7-1b.
DICOM Model of the Real World - PRINT
7-2b.
DICOM Information Model - PRINT
7-2c.
DICOM Information Model - RADIOTHERAPY
7-2d.
DICOM Information Model - IMPLANT TEMPLATES
7-3.
Model of the Real World for the Purpose of Modality-IS Interface
7.5-1.
Extension of the Real World Model with Concatenations and Dimensions
7.6-1.
DICOM Model of the Real World - Clinical Trials and Research
7.7-1.
DICOM Model of the Real World - Hanging Protocol
7.8-1.
DICOM Model of the Real World - Color Palette
7.9-1.
DICOM Model of the Real World - Specimens
7.10-1.
Implant Template Model of the Real World
7.10-2.
Implant Assembly Template Model of the Real World
7.10-3.
Implant Template Group Model of the Real World
7.11-1.
DICOM Model of the Real World - Unified Procedure Step
7.12-1.
DICOM Model of the Real World - Display System
7.12-2.
Display Subsystem Composition in the Display System IOD
10.7.1.3-1.
Example of Pixel Spacing Value Order
A.1-1.
DICOM Composite Instance IOD Information Model
A.17-1.
DICOM RT Image IOD Information Model
A.18-1.
DICOM RT Dose IOD Information Model
A.19-1.
DICOM RT Structure Set IOD Information Model
A.20-1.
DICOM RT Plan IOD Information Model
A.29-1.
DICOM RT Beams Treatment Record IOD Information Model
A.30-1.
DICOM RT Brachy Treatment Record IOD Information Model
A.31-1.
DICOM RT Treatment Summary Record IOD Information Model
A.34-1.
DICOM Waveform IOD Information Model
A.34.3-1.
12-Lead ECG Example (Informative)
A.36-1.
Relationships between Enhanced MR IODs
A.39-1.
Spatial Registration Information Object Definition E-R Model
A.40-1.
Spatial Fiducials Information Object Definition E-R Model
A.43-1.
Stereometric Relationship Information Object Definition E-R Model
A.44.2-1.
Hanging Protocol IOD E-R Model
A.46-1.
Real World Value Mapping Information Object Definition E-R Model
A.58.2-1.
Color Palette IOD E-R Model
A.59-1.
Types of 3D Ultrasound Source and Derived Images (Informative)
A.61-1.
Generic Implant Template IOD Information Model
A.62-1.
Implant Assembly Template IOD Information Model
A.63-1.
Implant Template Group IOD Information Model
A.64-1.
DICOM RT Beams Delivery Instruction IOD Information Model
C.7.1.4-1.
Example of Subject Relative Position in Image for group of 6 mice in a 3 column, 2 row single longitudinal plane array, all feet first prone
C.7.1.4-2.
Example of Subject Relative Position in Image and differing Patient Position for group of 2 mice head-to-head
C.7.1.4-3.
Example of Subject Relative Position in Image and overlapping narrow field of view group of 2 mice head-to-head
C.7.2-1.
Functions of Physicians
C.7.3.1.1.2-1.
Representation of the Eight Different Patient Positions on the X-Ray Table
C.7.3.1.1.2-2.
Example of Right First-Prone (RFP) Patient Position Relative to the Gantry and Table for a Small Animal
C.7.6.3.1.7-1.
Example of Pixel Size and Aspect Ratio
C.7.6.4b-1.
X-Ray beam intensity vs. Contrast/Bolus Ingredient Opaque
C.7.6.10-1.
Example of TID Offset
C.7-1.
Rectangular Display Shutter (1:1 aspect ratio image)
C.7-2.
Rectangular Display Shutter (2:1 aspect ratio images as they would appear before interpolation for display)
C.7-3.
Circular Display Shutter (1:1 aspect ratio image)
C.7-4.
Circular Display Shutter (2:1 aspect ratio images as they would appear before interpolation for display)
C.7-4b.
Combined Rectangular and Circular Display Shutter
C.7.6.16-1.
A Graphical Presentation of the Multi-frame Functional Groups Structure
C.7.6.16-1a.
SOP Instance UID of Concatenation Source
C.7.6.16-2.
Relationship of Timing Related Attributes
C.7.6.16-3.
Identifying Attributes for Concatenation, SOP Instances, Frames and Stacks
C.7.6.16-4.
Example of multiple stacks
C.7.6.16-4b.
PET dynamic frame organization
C.7.6.16-4c.
Example of Functional MRI volumes
C.7.6.16-5a.
Cardiac Timing Tags
C.7.6.16-5b.
Cardiac Timing Relationships
C.7.6.16-6.
The Real World Value LUT and the Image Viewing Pipeline
C.7.6.16-7.
Example of mapping stored values to real world values
C.7.6.16-7a.
Purpose of Pixel Intensity Relationship LUT
C.7.6.16-8.
Example of usage of Subtraction Item ID in a per-frame Macro
C.7.6.16-9a.
Respiratory Timing Tags
C.7.6.16-9b.
Relationship of Respiratory Amplitude Attributes
C.7.6.17-1.
Example of Dimension Index Sequence and Dimension Index Values Attributes
C.7.6.17-2.
Example of Use of Dimension Organization Module
C.7.6.23-1.
Enhanced Blending and Display Pipeline
C.8-6.
Intersecting Spatial Format Regions and Overlapping Measurement
C.8-7.
Pixel Component Calibration
C.8-8.
Pixel Component Calibration Example
C.8-1.
2D Regions with Reference Pixel
C.8-2.
2D & Doppler Regions with Reference Pixel
C.8-3.
2D & M-Mode Regions with Reference Pixel Example
C.8-4.
2D, M-Mode, & Waveform Regions with Reference Pixel
C.8-5.
Sweep Example Using Two Regions
C.8-9.
Relationships of X-Ray Collimator
C.8-9a.
Table Longitudinal Movement
C.8-9b.
Table Lateral Movement
C.8-10.
Table Motion Vector Coordinates
C.8-11.
Positioner Primary Angle
C.8-12.
Positioner Secondary Angle
C.8.8.2.7-1.
Non-normal Image Plane
C.8.8.3-1.
Dose Grid Frame Example
C.8.8.6-1.
Example of ROI with excluded inner volume
C.8.8.14-1.
Table Top Pitch Angle
C.8.8.14-2.
Table Top Roll Angle
C.8.8.21-1.
Control Point Example 1: 2 Control Points, 2 Complete Partial Treatments
C.8.8.21-2.
Control Point Example 1: 2 Control Points, 2 Complete Partial Treatments
C.8.8.21-3.
Control Point Example 3: 7 Control Points, 2 Partial Treatments with Small Gap
C.8.8.25-1.
Virtual Source-Axis Distances
C.8.8.25-2a.
Fixed Beam Line - View Along IEC FIXED Y-axis
C.8.8.25-2b.
Fixed Beam Line - View From Top (Along IEC FIXED Z-axis)
C.8.8.25-5.
Patient's eye view
C.8.8.25-6.
Lateral view along the positive axis Xb
C.8.8.25-7.
Gantry Pitch Angle
C.8.8.25.9-1.
Attributes specifying a depth dose distribution parameters in case of Reference Dose Definition (300A,0512) = CENTER
C.8.8.25.9-2.
Attributes specifying a generic depth dose distribution. Abscissa expanded near distal edge of dose distribution.
C.8.8.25.9-3.
Representation of the different Reference Dose Definition (300A,0512) using the range modulated depth dose distribution or highest energy component depth dose distribution.
C.8.8.29-1.
Double Exposure Field Delta
C.8.8.30-1.
Relationship of brachytherapy treatment plan and delivery instruction
C.8.8.30-2.
Relation of the Fraction, Pulse and Channels within the RT Brachy Task Sequence
C.8.9.1.1.11-1a.
Example DYNAMIC PET Series
C.8.9.1.1.11-1b.
Example GATED PET Series
C.8-13.
Explanation of Presentation Intent Type
C.8-14.
Explanation of Digital Detector Attributes
C.8-15.
Explanation of Digital Detector Attributes
C.8.11.7-1.
X-Ray Beam for a Digital Mammography System
C.8.11.7-2.
CC, FB, XCCL and XCCM partial views
C.8.11.7-3.
ML, MLO, LM, LMO, SIO and ISO partial views
C.8-16.
Reference Slide Orientation
C.8-17.
Z-Axis Center Point Location, View From Right Edge of Slide
C.8.13.5.14-1.
Attributes for Quantitative ASL
C.8.14.1-1.
Example Result without and with Zero Filling
C.8.14.1-2.
Echo Acquisition with Applying Zero Fill
C.8.14.1-3.
Free Induction Decay Acquisition with Applying Zero Fill
C.8.14.1-4.
Relation Between Spectroscopy Instances
C.8-18.
Dimensions of Spectroscopy Data
C.8-19.
Geometry of CT Acquisition System
C.8-20.
MONOCHROME2 Photometric Interpretation with Supplemental Palette Color Mapping
C.8.17.10-1.
Relationship of Longitudinal (Nominally Perpendicular) OCT Frames to Reference Frame
C.8.17.10-2.
Relationship of Transverse (Nominally Parallel) OCT Frame to Reference Frame
C.8.19.6.4-1.
Explanation of Pixel Data Area Attributes
C.8.19.6-1.
Project Calibration Without Angulation of the X-Ray Beam (Beam Angle = 0)
C.8.19.6-2.
Project Calibration With Angulation of the X-Ray Beam (Beam Angle Not Equal to 0)
C.8.19.6-3.
Table Position Vectors
C.8.19.6-4.
Isocenter Coordinate System
C.8.19.6-5.
Positioner Coordinate System
C.8.19.6-6.
Positioner Isocenter Angles
C.8.19.6-7.
Positioner Isocenter Detector Rotation Angle when Ap
1
= 0 and Ap
2
= 0
C.8.19.6-8.
Table Translation with respect to the Isocenter Reference System
C.8.19.6-9.
Table Angulations with respect to the Isocenter Reference System
C.8.21.3-1.
Example of Usage Acquisition
C.8.21.3-2.
Example of Contents of Contributing Sources Sequence
C.8.21.3-3.
Example of Contents of X-Ray 3D Acquisition Sequence
C.8.21.3-4.
Example of Contents of X-Ray 3D Reconstruction Sequence
C.8.21.3-5.
Example of Contents of X-Ray 3D Frame Type Sequences
C.8.22-1.
Geometry of PET Acquisition System
C.8.24.2-1.
Volume Frame of Reference
C.8.24.2-2.
Transducer Frame of Reference
C.8.24.2-3.
Table Frame of Reference
C.8.24.3.3-1.
Dimension Organization Type Frame Relationships
C.8.24.5-1.
Relationship of Timing Related Attributes
C.8.26.4.1-1.
Convention of coordinate system used in visual field representation.
C.8.27-1.
Explanation of Presentation Intent Type
C.8.27-2.
OCT Z Offset Correction Example
C.8.28.2.1-1.
Reference/Localizer Image Related to OPM Image
C.8.28.2.1-2.
Optic Nerve Head Thickness Example
C.8.28.2.1-3.
Polarimetry Optic Nerve Head Thickness Example
C.8.29.2-1.
Timing Attributes Illustration
C.8.30.2.1-1.
Corneal Topography Map Real World Value and Image Transformation Pipeline
C.8.30.3.1-1.
Device types: A. Placido; B. Grid; C. OCT; D. Slit Based
C.8.30.3.1-2.
Source Image (left), Corneal Topography Map Image (Center), Corneal Topography Map Presented as an Overlay on the Source Image (right)
C.8.30.3.1-3.
Representation of Corneal Topography
C.8.30.3.1-4.
Sample Coordinate Data Points
C.8.30.3.1-5.
Schematic of the 3-Dimensional Representation of Corneal Elevation
C.8.30.3.1-6.
Schematic of the Ophthalmic Coordinate System of the 3-Dimensional Representation used in Wide Field Measurements
C.8.31.2-1.
Illustration of Positioner Primary Angle Direction
C.8.31.6-1.
Patient/Detector Relationship (Patient Standing)
C.8.31.6-2.
Gantry Sides
C.8.31.6-3.
Isocenter Coordinate System
C.8.31.6-4.
X-Ray Source Coordinate System
C.8.31.6-5.
X-Ray Source Isocenter Angles
C.8.31.6-6.
Breast Support Orientation
C.8.31.6-7.
Breast Support Angles with respect to the Isocenter Reference System
C.8.31.6-8.
Breast Support Translation with respect to the Isocenter Reference System
C.8.31.6-9.
Detector Coordinate System
C.8.31.6-10.
Detector Isocenter Angles
C.8.31.6-11.
Translation of the Detector Coordinate System
C.8.31.6-12.
Examples of Detector Active Area TLHC Position and Orientation
C.10.4-1.
Example of Displayed Area Selection Addressing of Pixels Before and After Spatial Transformation
C.10.5-1.
Sub-pixel Addressing Units in PIXEL Space
C.10.5-2.
ELLIPSE on the left and RECTANGLE object on the right
C.10.5-3.
Example for MULTILINE Object
C.10.5-4.
INFINITELINE
C.10.5-5.
CUTLINE on the Left and CUTLINE with gap on the Right
C.10.5-6.
RANGELINE Example
C.10.5-7.
RULER / AXIS Example Showing TOP Tick Alignment and TOP Tick Label Alignment
C.10.5-8.
CROSSHAIR Example Showing BOTTOM Tick Labels Alignment
C.10.5-9.
ARROW Example
C.10.5-10.
Example of Horizontal and Vertical CENTER Alignment
C.10.5-11.
Example for Shadow Style (0070,0244) NORMAL
C.10.5-12.
Example for Shadow Style (0070,0244) OUTLINED
C.11.5-1.
Image Histogram Example
C.11.19-1.
Applicable pixel shift in case of multiple pixel shift regions
C.11.19-2.
Example of Contents of Mask Subtraction Sequence
C.11.19-3.
Example of three different pixel shift regions
C.11.26-1.
Volume Rendering with visualization of MPR plane orientation and resulting PLANAR MPR rendering
C.11.26-2.
Planar MPR THIN Geometry
C.11.26-3.
Planar MPR SLAB Geometry
C.12-1.
Example of Nesting Digital Signatures (Informative)
C.12-2.
Example of Encoding of Encrypted Attributes Data Set (Informative)
C.12-3.
HL7 Structured Document References
C.17.3-1.
SR Information Model
C.17.3-2.
Example of an SR Content Tree (Informative)
C.17.3-3.
Use of Position as SR Content Item Identifier (Informative)
C.17.4-1.
SR Content Tree for an Example Diagnostic Interpretation (Informative)
C.17.5-1.
Definition and Inheritance of Observation Context (Informative)
C.23.2-1.
Example Hanging Protocol Screen Environment
C.27.1.1-1.
Finite Volume Illustration
C.27.1.1-2.
Manifold Illustration
C.27.1.1-3.
Triangle Normal Computation
C.27.6.1.1-1.
Example of mapping textures to point coordinates.
C.29.1.2-1.
DICOM-HPGL Document
C.29.1.2-2.
DICOM-HPGL Document - Example
C.29.3.1.1.1-1.
Implant Template Groups Example.
C.29.3.1.1.2-1.
Implant Template Group for 2 Dimensions Example
F.1-1.
The DICOMDIR File. A Central Role in a DICOM File-set
F.2-1.
Basic Directory Information Object E-R Model
F.2-2.
Example of a Directory Organization and Content
F.2-3.
Example of Data Structure for the DICOM Directory Information
F.4-1.
Basic Directory IOD Information Model
List of Tables
5.4-1.
Example Module Table
5.4-2.
Example Macro Attributes
5.4-3.
Example Module Table Without The Use of An Attribute Macro
8.8-1a.
Basic Code Sequence Macro Attributes
8.8-1.
Code Sequence Macro Attributes
10-1.
Person Identification Macro Attributes Description
10-2.
Content Item Macro Attributes Description
10-3.
Image SOP Instance Reference Macro Attributes
10-3b.
Referenced Instances and Access Macro Attributes
10-3c.
Storage Macro Attributes
10-4.
Series and Instance Reference Macro Attributes
10-5.
General Anatomy Mandatory Macro Attributes
10-6.
General Anatomy Required Macro Attributes
10-7.
General Anatomy Optional Macro Attributes
10-8.
Primary Anatomic Structure Macro Attributes
10-9.
Request Attributes Macro Attributes
10-10.
Basic Pixel Spacing Calibration Macro Attributes
10-11.
SOP Instance Reference Macro Attributes
10-12.
Content Identification Macro Attributes
10-13.
General Contributing Sources Macro Attributes
10-14.
Contributing Image Sources Macro Attributes
10-15.
Patient Orientation Macro Attributes
10-16.
Performed Procedure Step Summary Macro Attributes
10-17.
HL7v2 Hierarchic Designator Macro Attributes
10-18.
Issuer of Patient ID Macro Attributes
10-19.
Algorithm Identification Macro Attributes
10-20.
Selector Attribute Macro Attributes
10-21.
Selector Attribute Macro Example
10-22.
Data Set Identification Macro Attributes
10-23.
Exposure Index Macro Attributes
10-24.
Mandatory View and Slice Progression Direction Macro Attributes
10-25.
Optional View and Slice Progression Direction Macro Attributes
10-26.
Numeric Value Macro Attributes
10-27.
RT Equipment Correlation Macro Attributes Description
10-28.
Device Motion Control Macro Attributes Description
A.1-1a.
Composite Information Object Modules Overview - Images
A.1-1b.
Composite Information Object Modules Overview - Images
A.1-1c.
Composite Information Object Modules Overview - More Images
A.1-2.
Composite Information Object Modules Overview - Non-Images
A.1-3.
Composite Information Object Modules Overview - More Non-Images
A.1-4.
Composite Information Object Modules Overview - Radiotherapy
A.1-5.
Composite Information Object Modules Overview - Implants
A.1-6.
Composite Information Object Modules Overview - Presentation States
A.1-7.
Composite Information Object Modules Overview - Structured Reports
A.2-1.
CR Image IOD Modules
A.3-1.
CT Image IOD Modules
A.4-1.
MR Image IOD Modules
A.5-1.
NM Image IOD Modules
A.6-1.
US Image IOD Modules
A.7-1.
US Multi-frame Image IOD Modules
A.8-1.
SC Image IOD Modules
A.8-2.
Multi-frame Single Bit SC Image IOD Modules
A.8-3.
Multi-frame Grayscale Byte SC Image IOD Modules
A.8-3b.
Multi-frame Grayscale Byte SC Image Functional Group Macros
A.8-4.
Multi-frame Grayscale Word SC Image IOD Modules
A.8-4b.
Multi-frame Grayscale Word SC Image Functional Group Macros
A.8-5.
Multi-frame True Color SC Image IOD Modules
A.8-5b.
Multi-frame True Color SC Image Functional Group Macros
A.14-1.
X-Ray Angiographic Image IOD Modules
A.16-1.
XRF Image IOD Modules
A.17.3-1.
RT Image IOD Modules
A.18.3-1.
RT Dose IOD Modules
A.19.3-1.
RT Structure Set IOD Modules
A.20.3-1.
RT Plan IOD Modules
A.21.3-1.
PET Image IOD Modules
A.26-1.
Digital X-Ray Image IOD Modules
A.27-1.
Digital Mammography X-Ray Image IOD Modules
A.28-1.
Digital Intra-Oral X-Ray Image IOD Modules
A.29.3-1.
RT Beams Treatment Record IOD Modules
A.30.3-1.
RT Brachy Treatment Record IOD Modules
A.31.3-1.
RT Treatment Summary Record IOD Modules
A.32.1-1.
VL Endoscopic Image IOD Modules
A.32.1-2.
VL Microscopic Image IOD Modules
A.32.1-3.
VL Slide-Coordinates Microscopic Image IOD Modules
A.32.4-1.
VL Photographic Image IOD Modules
A.32.5-1.
Video Endoscopic Image IOD Modules
A.32.6-1.
Video Microscopic Image IOD Modules
A.32.7-1.
Video Photographic Image IOD Modules
A.32.8-1.
VL Whole Slide Microscopy Image IOD Modules
A.32.8-2.
VL Whole Slide Microscopy Image Functional Group Macros
A.33.1-1.
Grayscale Softcopy Presentation State IOD Modules
A.33.2-1.
Color Softcopy Presentation State IOD Modules
A.33.3-1.
Pseudo-Color Softcopy Presentation State IOD Modules
A.33.4-1.
Blending Softcopy Presentation State IOD Modules
A.33.5.1-1.
Basic Structured Display IOD Modules
A.33.6-1.
XA/XRF Grayscale Softcopy Presentation State IOD Modules
A.34.2-1.
Basic Voice Audio IOD Modules
A.34.3-1.
12-Lead ECG IOD Modules
A.34.4-1.
General ECG IOD Modules
A.34.5-1.
Ambulatory ECG IOD Modules
A.34.6-1.
Hemodynamic IOD Modules
A.34.7-1.
Basic Cardiac EP IOD Modules
A.34.8-1.
Arterial Pulse Waveform IOD Modules
A.34.9-1.
Respiratory Waveform IOD Modules
A.34.10-1.
General Audio Waveform IOD Modules
A.35.1-1.
Basic Text SR IOD Modules
A.35.1-2.
Relationship Content Constraints for Basic Text SR IOD
A.35.2-1.
Enhanced SR IOD Modules
A.35.2-2.
Relationship Content Constraints for Enhanced SR IOD
A.35.3-1.
Comprehensive SR IOD Modules
A.35.3-2.
Relationship Content Constraints for Comprehensive SR IOD
A.35.4-1.
Key Object Selection Document IOD Modules
A.35.4-2.
Relationship Content Constraints for Key Object Selection Document IOD
A.35.5-1.
Mammography CAD SR IOD Modules
A.35.5-2.
Relationship Content Constraints for Mammography CAD SR IOD
A.35.6-1.
Chest CAD SR IOD Modules
A.35.6-2.
Relationship Content Constraints for Chest CAD SR IOD
A.35.7-1.
Procedure Log IOD Modules
A.35.7-2.
Relationship Content Constraints for Procedure Log IOD
A.35.8-1.
X-Ray Radiation Dose SR IOD Modules
A.35.8-2.
Relationship Content Constraints for X-Ray Radiation Dose SR IOD
A.35.9-1.
Spectacle Prescription Report IOD Modules
A.35.9-2.
Relationship Content Constraints for Spectacle Prescription Report IOD
A.35.10-1.
Colon CAD SR IOD Modules
A.35.10-2.
Relationship Content Constraints for Colon CAD SR IOD
A.35.11-1.
Macular Grid Thickness and Volume Report IOD Modules
A.35.11-2.
Relationship Content Constraints for Macular Grid Thickness and Volume Report IOD
A.35.12-1.
Implantation Plan SR Document IOD Modules
A.35.12-2.
Relationship Content Constraints for Implantation Plan SR Document IOD
A.35.13-1.
Comprehensive 3D SR IOD Modules
A.35.13-2.
Relationship Content Constraints for Comprehensive 3D SR IOD
A.35.14-1.
Radiopharmaceutical Radiation Dose SR IOD Modules
A.35.14-2.
Relationship Content Constraints for Radiopharmaceutical Radiation Dose SR IOD
A.35.15-1.
Extensible SR IOD Modules
A.35.15-2.
Relationship Content Constraints for Extensible SR IOD
A.35.16-1.
Acquisition Context SR IOD Modules
A.35.16-2.
Relationship Content Constraints for Acquisition Context SR IOD
A.36-1.
Enhanced MR Image IOD Modules
A.36-2.
Enhanced MR Image Functional Group Macros
A.36-3.
MR Spectroscopy IOD Modules
A.36-4.
MR Spectroscopy Functional Group Macros
A.36-5.
Enhanced MR Color Image IOD Modules
A.37-1.
Raw Data IOD Modules
A.38-1.
Enhanced CT Image IOD Modules
A.38-2.
Enhanced CT Image Functional Group Macros
A.39.1-1.
Spatial Registration IOD Modules
A.39.2-1.
Deformable Spatial Registration IOD Modules
A.40-1.
Spatial Fiducials IOD Modules
A.41-1.
Ophthalmic Photography 8 Bit Image IOD Modules
A.42-1.
Ophthalmic Photography 16 Bit Image IOD Modules
A.43-2.
Stereometric Relationship IOD Modules
A.44.3-1.
Hanging Protocol IOD Modules
A.45.1-1.
Encapsulated PDF IOD Modules
A.45.2-1.
Encapsulated CDA IOD Modules
A.46-1.
Real World Value Mapping IOD Modules
A.47-1.
Enhanced X-Ray Angiographic Image IOD Modules
A.47-2.
Enhanced XA Image Functional Group Macros
A.48.-1.
Enhanced X-Ray RF Image IOD Modules
A.48-2.
Enhanced XRF Image Functional Group Macros
A.49-1.
RT Ion Plan IOD Modules
A.50-1.
RT Ion Beams Treatment Record IOD Modules
A.51-1.
Segmentation IOD Modules
A.51-2.
Segmentation Functional Group Macros
A.52.3-1.
Ophthalmic Tomography Image IOD Modules
A.52.4.3-1.
Ophthalmic Tomography Functional Group Macros
A.53-1.
X-Ray 3D Angiographic Image IOD Modules
A.53-2.
X-Ray 3D Angiographic Image Functional Group Macros
A.54-1.
X-Ray 3D Craniofacial Image IOD Modules
A.54-2.
X-Ray 3D Craniofacial Image Functional Group Macros
A.55-1.
Breast Tomosynthesis Image IOD Modules
A.55-2.
Breast Tomosynthesis Image Functional Group Macros
A.56-1.
Enhanced PET Image IOD Modules
A.56-2.
Enhanced PET Image Functional Group Macros
A.57-1.
Surface Segmentation IOD Modules
A.58.3-1.
Color Palette IOD Modules
A.59-1.
Enhanced US Volume IOD Modules
A.59-2.
Enhanced US Volume Functional Group Macros
A.60.1-1.
Lensometry Measurements IOD Modules
A.60.2-1.
Autorefraction Measurements IOD Modules
A.60.3-1.
Keratometry Measurements IOD Modules
A.60.4-1.
Subjective Refraction Measurements IOD Modules
A.60.5-1.
Visual Acuity Measurements IOD Modules
A.60.6-1.
Ophthalmic Axial Measurements IOD Modules
A.60.7-1.
Intraocular Lens Calculations IOD Modules
A.61-1.
Generic Implant Template IOD Modules
A.62-1.
Implant Assembly Template IOD Modules
A.63-1.
Implant Template Group IOD Modules
A.64.3-1.
RT Beams Delivery Instruction IOD Modules
A.65.3-1.
Ophthalmic Visual Field Static Perimetry Measurements IOD Modules
A.66.3-1.
Intravascular OCT Image IOD Modules
A.66.4.3-1.
Intravascular OCT Functional Group Macros
A.67.3-1.
Ophthalmic Thickness Map IOD Modules
A.68-1.
Surface Scan Mesh IOD Modules
A.69-1.
Surface Scan Point Cloud IOD Modules
A.70-1.
Legacy Converted Enhanced CT Image IOD Modules
A.70-2.
Legacy Converted Enhanced CT Image Functional Group Macros
A.71-1.
Legacy Converted Enhanced MR Image IOD Modules
A.71-2.
Legacy Converted Enhanced MR Image Functional Group Macros
A.72-1.
Legacy Converted Enhanced PET Image IOD Modules
A.72-2.
Legacy Converted Enhanced PET Image Functional Group Macros
A.73.3-1.
Corneal Topography Map IOD Modules
A.74-1.
Breast Projection X-Ray Image IOD Modules
A.74-2.
Breast Projection X-Ray Image Functional Group Macros
A.75-1.
Parametric Map IOD Modules
A.75-2.
Parametric Map Functional Group Macros
A.76-1.
Wide Field Ophthalmic Photography Stereographic Projection Image IOD Modules
A.77-1.
Wide Field Ophthalmic Photography 3D Coordinates Image IOD Modules
A.78-1.
Tractography Results IOD Modules
A.79.3-1.
RT Brachy Application Setup Delivery Instruction IOD Modules
A.80.1-1.
Planar MPR Volumetric Presentation State IOD Modules
B.7-1.
Film Session IOD Modules
B.8-1.
Basic Film Box IOD Modules
B.9-1.
Basic Image Box IOD Modules
B.10-1.
Basic Annotation Box IOD Modules
B.11-1.
Print Job IOD Modules
B.12-1.
Printer IOD Modules
B.15-1.
Storage Commitment IOD Modules
B.17.2-1.
Modality Performed Procedure Step IOD Modules
B.18.2-1.
Presentation LUT IOD Modules
B.20-1.
Printer Configuration IOD Modules
B.24.2-1.
Instance Availability Notification IOD Modules
B.25.2-1.
Media Creation Management IOD Modules
B.26.2-1.
Unified Procedure Step IOD Modules
B.27.2-1.
RT Conventional Machine Verification IOD Modules
B.28.2-1.
RT Ion Machine Verification IOD Modules
B.29-1.
Display System IOD Modules
C.2-1.
Patient Relationship Module Attributes
C.2-2.
Patient Identification Module Attributes
C.2-3.
Patient Demographic Module Attributes
C.2-4.
Patient Medical Module Attributes
C.3-1.
Visit Relationship Module Attributes
C.3-2.
Visit Identification Module Attributes
C.3-3.
Visit Status Module Attributes
C.3-4.
Visit Admission Module Attributes
C.4-10.
Scheduled Procedure Step Module Attributes
C.4-11.
Requested Procedure Module Attributes
C.4-12.
Imaging Service Request Module Attributes
C.4-13.
Performed Procedure Step Relationship Module Attributes
C.4-14.
Performed Procedure Step Information Module Attributes
C.4-15.
Image Acquisition Results Module Attributes
C.4-16.
Radiation Dose Module Attributes
C.4-17.
Billing and Material Management Code Module Attributes
C.4.23-1.
Instance Availability Notification Module Attributes
C.7-1.
Patient Module Attributes
C.7-2b.
Clinical Trial Subject Module Attributes
C.7.1.4-1.
Patient Group Macro Attributes
C.7-3.
General Study Module Attributes
C.7-4a.
Patient Study Module Attributes
C.7-4b.
Clinical Trial Study Module Attributes
C.7-5a.
General Series Module Attributes
C.7-5b.
Clinical Trial Series Module Attributes
C.7-5c.
Enhanced Series Module Attributes
C.7-6.
Frame of Reference Module Attributes
C.7-7.
Synchronization Module Attributes
C.7-8.
General Equipment Module Attributes
C.7-8b.
Enhanced General Equipment Module Attributes
C.7-9.
General Image Module Attributes
C.7-10.
Image Plane Module Attributes
C.7-11a.
Image Pixel Module Attributes
C.7-11b.
Image Pixel Macro Attributes
C.7-12.
Contrast/Bolus Module Attributes
C.7-12b.
Enhanced Contrast/Bolus Module Attributes
C.7-13.
Cine Module Attributes
C.7-14.
Multi-frame Module Attributes
C.7-15.
Frame Pointers Module Attributes
C.7-16.
Mask Module Attributes
C.7.6.10-1.
Example Mask Frame Numbers for Mask Operation REV_TID
C.7-17.
Display Shutter Module Attributes
C.7-17a.
Display Shutter Macro Attributes
C.7-18.
Device Module Attributes
C.7-19.
Intervention Module Attributes
C.7.6.14-1.
Acquisition Context Module Attributes
C.7.6.15-1.
Bitmap Display Shutter Module Attributes
C.7.6.16-1.
Multi-frame Functional Groups Module Attributes
C.7.6.16-2.
Pixel Measures Macro Attributes
C.7.6.16-3.
Frame Content Macro Attributes
C.7.6.16-4.
Plane Position (Patient) Macro Attributes
C.7.6.16-5.
Plane Orientation (Patient) Macro Attributes
C.7.6.16-6.
Referenced Image Macro Attributes
C.7.6.16-7.
Derivation Image Macro Attributes
C.7.6.16-8.
Cardiac Synchronization Macro Attributes
C.7.6.16-9.
Frame Anatomy Macro Attributes
C.7.6.16-10.
Pixel Value Transformation Macro Attributes
C.7.6.16-10b.
Identity Pixel Value Transformation Macro Attributes
C.7.6.16-11.
Frame VOI LUT Macro Attributes
C.7.6.16-11b.
Frame VOI LUT with LUT Macro Attributes
C.7.6.16-12.
Real World Value Mapping Macro Attributes
C.7.6.16-12b.
Real World Value Mapping Item Macro Attributes
C.7.16-13.
Contrast/Bolus Usage Macro Attributes
C.7.6.16-14.
Pixel Intensity Relationship LUT Macro Attributes
C.7.6.16-15.
Frame Pixel Shift Macro Attributes
C.7.6.16-16.
Patient Orientation in Frame Macro Attributes
C.7.6.16-17.
Frame Display Shutter Macro Attributes
C.7.6.16-18.
Respiratory Synchronization Macro Attributes
C.7.6.16-19.
Irradiation Event Identification Macro Attributes
C.7.6.16.2-20.
Radiopharmaceutical Usage Macro Attributes
C.7.6.16.2-21.
Patient Physiological State Macro Attributes
C.7.6.16.2.21-1.
Plane Position (Volume) Macro Attributes
C.7.6.16.2.22-1.
Plane Orientation (Volume) Macro Attributes
C.7.6.16.2.23-1.
Temporal Position Macro Attributes
C.7.6.16.2.24-1.
Image Data Type Macro Attributes
C.7.6.16.2.24.1-1.
Data Type Defined Terms
C.7.6.16.2.25.1-1.
Unassigned Shared Converted Attributes Macro Attributes
C.7.6.16.2.25.2-1.
Unassigned Per-Frame Converted Attributes Macro Attributes
C.7.6.16.2.25.3-1.
Image Frame Conversion Source Macro Attributes
C.7.6.17-1.
Multi-frame Dimension Module Attributes
C.7.6.18-1.
Cardiac Synchronization Module Attributes
C.7.6.18-2.
Respiratory Synchronization Module Attributes
C.7.6.18-3.
Bulk Motion Synchronization Module Attributes
C.7.6.19-1.
Supplemental Palette Color Lookup Table Module Attributes
C.7.6.20-1.
Patient Orientation Module Attributes
C.7.6.21-1.
Image - Equipment Coordinate Relationship Module Attributes
C.7.6.22-1.
Specimen Module Attributes
C.7.6.22-2.
Specimen Macro Attributes
C.7.6.23-1.
Enhanced Palette Color Lookup Table Module Attributes
C.7.6.24-1.
Floating Point Image Pixel Module Attributes
C.7.6.25-1.
Double Floating Point Image Pixel Module Attributes
C.7-22.
Palette Color Lookup Table Module Attributes
C.7-22a.
Palette Color Lookup Table Macro Attributes
C.7-23.
Compressed Palette Color Lookup Table Data
C.7-24.
Segment Types
C.7-25.
Discrete Segment Type
C.7-26.
Linear Segment Type
C.7-27.
Indirect Segment Type
C.8-1.
CR Series Module Attributes
C.8-2.
CR Image Module Attributes
C.8-3.
CT Image Module Attributes
C.8-4.
MR Image Module Attributes
C.8-5.
NM/PET Patient Orientation Module Attributes
C.8-6.
NM Image Pixel Module Attributes
C.8-7.
NM Multi-frame Module Attributes
C.8-8.
Enumerated Values for Frame Increment Pointer
C.8-9.
NM Image Module Attributes
C.8-10.
NM Isotope Module Attributes
C.8-11.
NM Detector Module Attributes
C.8-12.
NM Tomo Acquisition Module Attributes
C.8-13.
NM Multi-Gated Acquisition Module Attributes
C.8-14.
NM Phase Module Attributes
C.8-15.
NM Reconstruction Module Attributes
C.8-17.
US Region Calibration Module Attributes
C.8-18.
US Image Module Attributes
C.8-19.
US Samples Per Pixel
C.8-20.
US Bits Allocated
C.8-23.
US Planar Configuration
C.8-24.
SC Equipment Module Attributes
C.8-25.
SC Image Module Attributes
C.8-25b.
SC Multi-frame Image Module Attributes
C.8-25c.
SC Multi-frame Vector Module Attributes
C.8-26.
X-Ray Image Module Attributes
C.8-27.
X-Ray Acquisition Module Attributes
C.8-28.
X-Ray Collimator Module Attributes
C.8-28b.
X-Ray Collimator Dimensions Macro Attributes
C.8-29.
X-Ray Table Module Attributes
C.8-30.
XA Positioner Module Attributes
C.8-31.
XRF Positioner Module Attributes
C.8-32.
X-Ray Tomography Acquisition Module Attributes
C.8-33.
X-Ray Acquisition Dose Module Attributes
C.8-34.
X-Ray Generation Module Attributes
C.8-35.
X-Ray Filtration Module Attributes
C.8-35a.
X-Ray Filtration Macro Attributes
C.8-36.
X-Ray Grid Module Attributes
C.8-36b.
X-Ray Grid Description Macro Attributes
C.8-37.
RT Series Module Attributes
C.8-38.
RT Image Module Attributes
C.8-39.
RT Dose Module Attributes
C.8-39b.
Values of Dose Grid Frame Offset Vector Under Relative (A) and Absolute (B) Interpretations
C.8-40.
RT DVH Module Attributes
C.8-41.
Structure Set Module Attributes
C.8-42.
ROI Contour Module Attributes
C.8-43.
RT Dose ROI Module Attributes
C.8-44.
RT ROI Observations Module Attributes
C.8-45.
RT General Plan Module Attributes
C.8-46.
RT Prescription Module Attributes
C.8-47.
RT Tolerance Tables Module Attributes
C.8-48.
RT Patient Setup Module Attributes
C.8-49.
RT Fraction Scheme Module Attributes
C.8-50.
RT Beams Module Attributes
C.8-51.
RT Brachy Application Setups Module Attributes
C.8-52.
Approval Module Attributes
C.8-53.
RT General Treatment Record Module Attributes
C.8-54.
RT Treatment Machine Record Module Attributes
C.8-55.
Measured Dose Reference Record Module Attributes
C.8-56.
Calculated Dose Reference Record Module Attributes
C.8-57.
RT Beams Session Record Module Attributes
C.8-58.
RT Brachy Session Record Module Attributes
C.8-59.
RT Treatment Summary Record Module Attributes
C.8.8.24-1.
RT Ion Tolerance Tables Module Attributes
C.8.8.25-1.
RT Ion Beams Module Attributes
C.8.8.26-1.
RT Ion Beams Session Record Module Attributes
C.8.8.27-1.
Beam Limiting Device Position Macro Attributes
C.8.8.28-1.
Patient Support Identification Macro Attributes
C.8.8.29-1.
RT Beams Delivery Instruction Module Attributes
C.8.8.30-1.
RT Brachy Application Setup Delivery Instruction Module Attributes
C.8-60.
PET Series Module Attributes
C.8-61.
PET Isotope Module Attributes
C.8-62.
PET Multi-Gated Acquisition Module Attributes
C.8-63.
PET Image Module Attributes
C.8-68.
DX Series Module Attributes
C.8-69.
DX Anatomy Imaged Module Attributes
C.8-70.
DX Image Module Attributes
C.8-71.
DX Detector Module Attributes
C.8-71b.
Digital X-Ray Detector Macro Attributes
C.8-72.
DX Positioning Module Attributes
C.8-73.
Mammography Series Module Attributes
C.8-74.
Mammography Image Module Attributes
C.8-74a.
Enumerated Values of Image Type (0008,0008) Value 3 for Stereotactic Mammography
C.8-74b.
Enumerated Values of Image Type (0008,0008) Value 3 for Digital Breast Tomosynthesis Projection and Generated 2D Images
C.8-74c.
Enumerated Values of Image Type (0008,0008) Value 3 for Contrast Enhanced Images
C.8-74d.
Defined Terms of Image Type (0008,0008) Value 4 for Contrast Enhanced and Generated 2D Images
C.8-74e.
Defined Terms of Image Type (0008,0008) Value 5 for Contrast Enhanced 2D Images
C.8-75.
Intra-Oral Series Module Attributes
C.8-76.
Intra-Oral Image Module Attributes
C.8-76b.
Enhanced Mammography Series Module Attributes
C.8-77.
VL Image Module Attributes
C.8-78.
Slide Coordinates Module Attributes
C.8.12.3-1.
Whole Slide Microscopy Series Module Attributes
C.8.12.4-1.
Whole Slide Microscopy Image Module Attributes
C.8.12.4-2.
Whole Slide Microscopy Image Flavors
C.8.12.4-3.
Whole Slide Microscopy Image Derived Pixels
C.8.12.5-1.
Optical Path Module Attributes
C.8.12.6.1-1.
Plane Position (Slide) Macro Attributes
C.8.12.6.2-1.
Optical Path Identification Macro Attributes
C.8.12.6.3-1.
Specimen Reference Macro Attributes
C.8.12.7-1.
Multi-Resolution Navigation Module Attributes
C.8.12.8-1.
Slide Label Module Attributes
C.8-79.
Enhanced MR Image Module Attributes
C.8-80.
MR-Specific Image Type and Frame Type Value 3
C.8-81.
MR-Specific Image Type and Frame Type Value 4
C.8-82.
Allowed Combinations of Attribute Values for Photometric Interpretation, Samples Per Pixel, Planar Configuration, Pixel Representation, Bits Allocated and Bits Stored
C.8-83.
MR Image and Spectroscopy Instance Macro Attributes
C.8-84.
MR Image Description Macro Attributes
C.8-85.
Complex Image Component Attribute Values
C.8-86.
Acquisition Contrast Values
C.8-87.
MR Pulse Sequence Module Attributes
C.8-88.
MR Image Frame Type Macro Attributes
C.8-89.
MR Timing and Related Parameters Macro Attributes
C.8-90.
MR FOV/Geometry Macro Attributes
C.8-91.
MR Echo Macro Attributes
C.8-92.
MR Modifier Macro Attributes
C.8-93.
MR Imaging Modifier Macro Attributes
C.8-94.
MR Receive Coil Macro Attributes
C.8-95.
MR Transmit Coil Macro Attributes
C.8-96.
MR Diffusion Macro Attributes
C.8-97.
MR Averages Macro Attributes
C.8-98.
MR Spatial Saturation Macro Attributes
C.8-99.
MR Metabolite Map Macro Attributes
C.8-100.
MR Velocity Encoding Macro Attributes
C.8-100b.
MR Arterial Spin Labeling Macro Attributes
C.8-100c.
Functional MR Macro Attributes
C.8-101.
MR Series Module Attributes
C.8-102.
MR Spectroscopy Module Attributes
C.8-103.
MR Spectroscopy Pulse Sequence Module Attributes
C.8-104.
MR Spectroscopy Frame Type Macro Attributes
C.8-105.
MR Spectroscopy FOV/Geometry Macro Attributes
C.8-106.
MR Spectroscopy Data Module Attributes
C.8-107.
MR Spectroscopy Description Macro Attributes
C.8-108.
MR Spectroscopy Image Type and Frame Type Value 3
C.8-109.
MR Spectroscopy Image Type and Frame Type Value 4
C.8-110.
Volume Based Calculation Technique Attribute Values
C.8-111.
Complex Image Component Attribute Values
C.8-112.
Acquisition Contrast Values
C.8-113.
CT Series Module Attributes
C.8-114.
Enhanced CT Image Module Attributes
C.8-115.
Image Type and Frame Type Value 3 for CT
C.8-116.
Image Type and Frame Type Value 4 for Ct
C.8-117.
CT Image Frame Type Macro Attributes
C.8-118.
CT Acquisition Type Macro Attributes
C.8-119.
CT Acquisition Details Macro Attributes
C.8-120.
CT Table Dynamics Macro Attributes
C.8-121.
CT Position Macro Attributes
C.8-122.
CT Geometry Macro Attributes
C.8-123.
CT Reconstruction Macro Attributes
C.8-124.
CT Exposure Macro Attributes
C.8-125.
CT X-Ray Details Sequence Macro Attributes
C.8-126.
CT Pixel Value Transformation Macro Attributes
C.8-126b.
CT Additional X-Ray Source Macro Attributes
C.8-127.
Image Type and Frame Type Value 1
C.8-128.
Image Type and Frame Type Value 2
C.8-129.
Image Type and Frame Type Value 3 Common
C.8-130.
Image Type and Frame Type Value 4 Common
C.8-131.
Common CT/MR Image Description Macro Attributes
C.8-132.
Pixel Presentation Attribute Values
C.8-133.
Volumetric Properties Attribute Values
C.8-134.
Volume Based Calculation Technique Attribute Values
C.8.17.1-1.
Ophthalmic Photography Series Module Attributes
C.8.17.2-1.
Ophthalmic Photography Image Module Attributes
C.8.17.3-1.
Ophthalmic Photographic Parameters Module Attributes
C.8.17.4-1.
Ophthalmic Photography Acquisition Parameters Module Attributes
C.8.17.5-1.
Ocular Region Imaged Module Attributes
C.8.17.6-1.
Ophthalmic Tomography Series Module Attributes
C.8.17.7-1.
Ophthalmic Tomography Image Module Attributes
C.8.17.8-1.
Ophthalmic Tomography Acquisition Parameters Module Attributes
C.8.17.8-2.
Ophthalmic Acquisition Parameters Macro Attributes
C.8.17.9-1.
Ophthalmic Tomography Parameters Module Attributes
C.8.17.10-1.
Ophthalmic Frame Location Macro Attributes
C.8.17.11-1.
Wide Field Ophthalmic Photography Stereographic Projection Module Attributes
C.8.17.12-1.
Wide Field Ophthalmic Photography 3D Coordinates Module Attributes
C.8.17.13-1.
Wide Field Ophthalmic Photography Quality Rating Module Attributes
C.8.18.1-1.
Stereometric Series Module Attributes
C.8.18.2-1.
Stereometric Relationship Module Attributes
C.8.19.1-1.
XA/XRF Series Module Attributes
C.8.19.2-1.
Enhanced XA/XRF Image Module Table
C.8.19.2-2.
Allowed Combinations of Attribute Values for Bits Allocated and Bits Stored
C.8.19.3-1.
XA/XRF Acquisition Module Attributes
C.8.19.4-1.
X-Ray Image Intensifier Module Attributes
C.8.19.5-1.
X-Ray Detector Module Attributes
C.8.19.6-1.
X-Ray Frame Characteristics Macro Attributes
C.8.19.6-2.
X-Ray Field of View Macro Attributes
C.8.19.6-3.
X-Ray Exposure Control Sensing Regions Macro Attributes
C.8.19.6-4.
X-Ray Frame Pixel Data Properties Macro Attributes
C.8.19.6-5.
X-Ray Frame Detector Parameters Macro Attributes
C.8.19.6-6.
X-Ray Calibration Device Usage Macro Attributes
C.8.19.6-7.
X-Ray Object Thickness Macro Attributes
C.8.19.6-8.
X-Ray Frame Acquisition Macro Attributes
C.8.19.6-9.
X-Ray Projection Pixel Calibration Macro Attributes
C.8.19.6-10.
X-Ray Positioner Macro Attributes
C.8.19.6-11.
X-Ray Table Position Macro Attributes
C.8.19.6-12.
X-Ray Collimator Macro Attributes
C.8.19.6-13.
X-Ray Isocenter Reference System Macro Attributes
C.8.19.6-14.
X-Ray Geometry Macro Attributes
C.8.19.7-1.
XA/XRF Multi-frame Presentation Module Attributes
C.8.20-1.
Segmentation Series Module Attributes
C.8.20-2.
Segmentation Image Module Attributes
C.8.20-3.
Segmentation Macro Attributes
C.8.20-4.
Segment Description Macro Attributes
C.8.21.1-1.
X-Ray 3D Image Module Attributes
C.8.21.1.1.3-1.
Additional Volume Based Calculation Technique Attribute Values for X-Ray 3D
C.8.21.2.1-1.
X-Ray 3D Angiographic Image Contributing Sources Module Attributes
C.8.21.2.2-1.
X-Ray 3D Craniofacial Image Contributing Sources Module Attributes
C.8.21.2.3-1.
Breast Tomosynthesis Contributing Sources Module Attributes
C.8.21.3.1.1-1.
X-Ray 3D General Shared Acquisition Macro Attributes
C.8.21.3.1.2-1.
X-Ray 3D General Per Projection Acquisition Macro Attributes
C.8.21.3.1.3-1.
X-Ray 3D General Positioner Movement Macro Attributes
C.8.21.3.2-1.
X-Ray 3D Angiographic Acquisition Module Attributes
C.8.21.3.3-1.
X-Ray 3D Craniofacial Acquisition Module Attributes
C.8.21.3.4-1.
Breast Tomosynthesis Acquisition Module Attributes
C.8.21.4-1.
X-Ray 3D Reconstruction Module Attributes
C.8.21.5.1-1.
X-Ray 3D Frame Type Macro Attributes
C.8.21.6-1.
Breast View Module Attributes
C.8.21.6-1a.
Defined Terms of Image Type (0008,0008) Value 3 for Contrast Enhanced, Tomosynthesis and Biopsy
C.8.21.6-1b.
Defined Terms of Image Type (0008,0008) Value 4 for Contrast Enhanced and Tomosynthesis
C.8.21.6-1c.
Defined Terms of Image Type (0008,0008) Value 5 for Contrast Enhanced Images
C.8.22-1.
Enhanced PET Series Module Attributes
C.8.22-2.
Enhanced PET Acquisition Module Attributes
C.8.22-3.
Enhanced PET Image Module Attributes
C.8.22-9.
Enhanced PET Isotope Module Attributes
C.8.22-10.
PET Frame Type Macro Attributes
C.8.22-11.
PET Frame Acquisition Macro Attributes
C.8.22-12.
PET Detector Motion Details Macro Attributes
C.8.22-14.
PET Position Macro Attributes
C.8.22-15.
PET Frame Correction Factors Macro Attributes
C.8.22-17.
PET Reconstruction Macro Attributes
C.8.22-18.
PET Table Dynamics Macro Attributes
<