DICOM PS3.3 2016a - Information Object Definitions |
---|
This Section describes the MR Image Module. This Module contains all Attributes that are specific to MR images.
Table C.8-4 contains the Attributes that describe MR images.
Table C.8-4. MR Image Module Attributes
Image identification characteristics. See Section C.8.3.1.1.1 for specialization. |
|||
Number of samples (planes) in this image. See Section C.8.3.1.1.2 for specialization. |
|||
Specifies the intended interpretation of the pixel data. See Section C.8.3.1.1.3 for specialization. |
|||
Number of bits allocated for each pixel sample. Each sample shall have the same number of bits allocated. See Section C.8.3.1.1.4 for specialization. |
|||
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. |
|||
Most significant bit for pixel sample data. Each sample shall have the same high bit. High Bit (0028,0102) shall be one less than Bits Stored (0028,0101). |
|||
Description of the type of data taken. Enumerated Values: |
|||
The period of time in msec between the beginning of a pulse sequence and the beginning of the succeeding (essentially identical) pulse sequence. Required except when Scanning Sequence (0018,0020) is EP and Sequence Variant (0018,0021) is not SK. |
|||
Time in ms between the middle of the excitation pulse and the peak of the echo produced (kx=0). In the case of segmented k-space, the TE(eff) is the time between the middle of the excitation pulse to the peak of the echo that is used to cover the center of k-space (i.e.,-kx=0, ky=0). |
|||
Number of lines in k-space acquired per excitation per image. |
|||
Time in msec after the middle of inverting RF pulse to middle of excitation pulse to detect the amount of longitudinal magnetization. Required if Scanning Sequence (0018,0020) has values of IR. |
|||
Time, in msec, between peak of the R wave and the peak of the echo produced. In the case of segmented k-space, the TE(eff) is the time between the peak of the echo that is used to cover the center of k-space. Required for Scan Options (0018,0022) that include heart gating (e.g., CG, PPG, etc.) |
|||
User defined name for the combination of Scanning Sequence (0018,0020) and Sequence Variant (0018,0021). |
|||
Number of times a given pulse sequence is repeated before any parameter is changed |
|||
Nucleus that is resonant at the imaging frequency. Examples: 31P, 1H |
|||
The echo number used in generating this image. In the case of segmented k-space, it is the effective Echo Number. |
|||
Spacing between slices, in mm. The spacing is measured from the center-to-center of each slice. |
|||
Total number of lines in k-space in the 'y' direction collected during acquisition. |
|||
Fraction of acquisition matrix lines acquired, expressed as a percent. |
|||
Ratio of field of view dimension in phase direction to field of view dimension in frequency direction, expressed as a percent. |
|||
Reciprocal of the total sampling period, in hertz per pixel. |
|||
Percent of R-R interval, based on Heart Rate (0018,1088), prescribed as a window for a valid/usable trigger. |
|||
Diameter in mm. of the region from within which data were used in creating the reconstruction of the image. Data may exist outside this region and portions of the patient may exist outside this region. |
|||
Dimensions of the acquired frequency /phase data before reconstruction. Multi-valued: frequency rows\frequency columns\phase rows\phase columns. |
|||
Steady state angle in degrees to which the magnetic vector is flipped from the magnetic vector of the primary field. |
|||
Calculated whole body Specific Absorption Rate in watts/kilogram. |
|||
The rate of change of the gradient coil magnetic flux density with time (T/s). |
|||
Time delta between Images in a dynamic or functional set of Images. |
|||
Include Table 10-7 “General Anatomy Optional Macro Attributes” |
|||
Include Table 10-25 “Optional View and Slice Progression Direction Macro Attributes” |
|||
Isocenter coordinates (x,y,z), in mm. Specifies the location of the machine isocenter in the patient-based coordinate system associated with the Frame of Reference. It allows transformation from the equipment-based coordinate system to the patient-based coordinate system. |
|||
B1+ rms value in units of micro_tesla (uT) for the acquisition producing the image. See [IEC 60601-2-33]. |
See Section C.7.6.3.1.2 for definition of these terms.
DICOM PS3.3 2016a - Information Object Definitions |
---|