G.6 Example of Elements For An Encoded YCbCr RLE Three-frame Image with Basic Offset Table
Table G.6-1 is an example of encoding of RLE Compressed Frames (described in Section G.4) with the basic offset table. Table G.6-2 is an example of Item Value data for one frame.
Table G.6-1. Example of Elements for an Encoded YCbCr RLE Three-Frame Image with Basic Offset Table
Pixel Data Element Tag
|
Value Representation
|
Data Element Length
|
Data Element
|
|
|
|
|
Basic Offset Table with Item Value
|
First Fragment (Frame 1) of Pixel Data
|
|
|
|
|
Item Tag
|
Item Length
|
Item Value
|
Item Tag
|
Item Length
|
Item Value
|
(7FE0,0010) with VR of OB
|
OB
|
0000H
|
FFFF FFFFH undefined length
|
(FFFE,E000)
|
0000 000CH
|
0000 0000H 0000 02D0H 0000 0642H
|
(FFFE,E000)
|
0000 02C8H
|
RLE Compressed Frame
|
4 bytes
|
2 bytes
|
2 bytes
|
4 bytes
|
4 bytes
|
4 bytes
|
000CH bytes
|
4 bytes
|
4 bytes
|
02C8H bytes
|
Table G.6-1b. Example of Elements for an Encoded YCbCr RLE Three-Frame Image with Basic Offset Table (continued)
Data Element Continued
|
Second Fragment (Frame 2) of Pixel Data
|
Third Fragment (Frame 3) of Pixel Data
|
Sequence Delimiter Item
|
Item Tag
|
Item Length
|
Item Value
|
Item Tag
|
Item Length
|
Item Value
|
Sequence Delimiter Tag
|
Item Length
|
(FFFE,E000)
|
0000 036AH
|
RLE Compressed Frame
|
(FFFE,E000)
|
0000 0BC8H
|
RLE Compressed Frame
|
(FFFE,E0DD)
|
0000 0000H
|
4 bytes
|
2 bytes
|
036AH bytes
|
4 bytes
|
4 bytes
|
0BC8H bytes
|
4 bytes
|
4 bytes
|
Table G.6-2. Example of Encoded YCbCr RLE Compressed Frame Item Value
Offset
|
Data
|
Description of Data
|
|
0000 0000H
|
0000 0003H
|
number of RLE Segments
|
(Header)
|
|
0000 0040H
|
location of RLE Segment 1 (Y component)
|
|
|
0000 0140H
|
location of RLE Segment 2 (CB component)
|
|
|
0000 01C0H
|
location of RLE Segment 3 (CR component)
|
|
|
0000 0000H
|
|
|
|
…..
|
|
|
|
…..
|
|
|
|
0000 0000H
|
|
|
0000 0040H
|
Y - RLE Segment Data
|
|
(DATA)
|
0000 0140H
|
CB - RLE Segment Data
|
|
(DATA)
|
0000 01C0H
|
CR - RLE Segment Data
|
|
(DATA)
|