    <title>Digital Imaging and Communications in Medicine (DICOM)</title>
    <subtitle>Sup 243 - Label Map Segmentation</subtitle>
    <info>
        <author>
            <orgname>DICOM Standards Committee - Working Group 6 - Base Standard</orgname>
            <address>
            <street>1300 N. 17th Street Suite 1752</street>
            <city>Rosslyn</city>
            <state>VA</state>
            <postcode>22209</postcode>
            <country>USA</country>
         </address>
        </author>
        <copyright>
            <year>2017</year>
            <holder>NEMA</holder>
        </copyright>
        <releaseinfo>Status: Final Text</releaseinfo>
        <pubdate>2024/09/14</pubdate>
        <legalnotice>
            <para>This supplement is prepared pursuant to work item: 2023-10-B</para>
        </legalnotice>
    </info>

    <chapter>
        <title>Scope and Field of Application</title>
        <para>This Supplement describes addition of a Label Map Segmentation IOD to DICOM to encode classification of entities.</para>
        <para>Currently, the DICOM standard supports an IOD and SOP Class for pixel- or voxel-based segmentation encoding (as distinct from the representation of segmented objects as surfaces in the surface segmentation and encapsulated 3D object IODs and SOP Classes), in which each segmented property is represented as a binary bit plane (or an 8 bit probabilistic or occupancy value). While this allows for overlapping of segments, it is inefficient and difficult to encode large numbers of non-overlapping segmentations, as they require non-trivial processing both to extract from the bit plane encoded data, to assure there is no overlap, and to convert to the label map form that is very commonly used internally and persistently for clinical applications. </para>
        <para>The current DICOM bit-plane-based segmentation methods have proven to be awkward both for 3D cross-sectional imaging applications when there are very large numbers of slices and/or structures, and for whole slide microscopy imaging, when there are very large numbers of tiles and/or property classes. They are also typically large and sparse and should compress well but there are very few single bit compression schemes supported by the standard and they do not do well with these types of images.</para>
        <para>This Supplement defines a label map segmentation enhanced multi-frame IOD that specifies a data structure that provides, for each pixel or voxel in 2D, 3D or tiled pyramidal space, an index value conveying the non-overlapping segment for each pixel. Existing data elements for describing segmentations are reused where appropriate. Bit depth is sufficient (8, 16) to encode large numbers of segments but allow for more compact encoding. The existing palette color photometric interpretation may be used (instead of monochrome) if colors are to be suggested, to leverage the widespread implementations in toolkits, and to allow for the use of existing lossless compression schemes. Segment properties are conveyed in the existing segment description structure so as to be compatible with the existing bit plane segment descriptions. Re-using the segment description does not prevent the use of separately encoded or well-known DICOM color palette objects.</para>
        <para>The scope is confined to label maps for "classes" (what "class" a segment represents) but not "instances' (which "instance" of a "class" is represented), where classes and instances are separately communicated by the pixel value (e.g., if one wants to individually identify nuclei rather than treat them all as being of one class). This might be the subject of a future extension.</para>
        <para>The scope is confined to a single label map, which does not allow for overlap of different segments. If overlapping of multiple label maps is required, separate SOP Instances may be created.</para>
        <para>Issues related to the efficient representation (or avoidance) of the Per-Frame Functional Group Sequence (in which, for every frame, the Referenced Segment Number is specified) are out of scope, and may be addressed in a separate Supplement or CP if necessary.</para>
    </chapter>
    <chapter label="PS3.3" status="1" xml:id="chapter_PS3.3g">
        <title>DICOM PS3.3 - Information Object Definitions</title>
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">PS3.3 for reference:</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
        <glossary>
            <glossdiv label="3.8" status="2" xml:id="sect_3.8">
                <title>DICOM Information Object</title>
                <glossentry xml:id="glossentry_Segment" xreflabel="Segment">
                    <glossterm>Segment</glossterm>
                    <glossdef>
                        <para xml:id="para_af217440-77ca-4551-804f-37880dd15d02">A part of a whole, such as the classification of pixels in an image.</para>
                    </glossdef>
                </glossentry>
            </glossdiv>
        </glossary>
    </chapter>
    
    <chapter label="PS3.3" status="1" xml:id="chapter_PS3.3">
        <title>DICOM PS3.3 - Information Object Definitions</title>
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.3 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="A.1.4" status="3" xml:id="sect_A.1.4">
            <table frame="box" label="A.1-1e" rules="all" xml:id="table_A.1-1e">
                <caption>Composite Information Object Modules Overview - More Images</caption>
                <thead>
                    <tr valign="top">
                        <th align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5d897c24-32b7-4d83-a3d6-943571bfdf5b">
                                <emphasis role="bold">IODs</emphasis>
                            </para>
                            <para xml:id="para_99f9723a-d5a7-4145-a64f-da73ed389b7a">
                                <emphasis role="bold">Modules</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_7cc21e1d-a2d5-4a6e-889f-5d0179c3f15d">Seg</para>
                        </th>
                    </tr>
                </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_340eb2ae-e247-44fb-9cf4-f5ae956dbb69">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_385337f6-ac04-40cb-8e78-7e150695d6a0">Clinical Trial Subject</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_982e8381-d924-4fef-9714-3b64e3ab6869">General Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_78fbea4b-b163-49cf-806e-3bd25f391113">Patient Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b56c3b88-41b0-4706-8da0-6a1b8cae95bf">Clinical Trial Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0f115cd5-f9c8-4ee0-af2b-f2862c17f866">General Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bc33cc47-4643-47a3-8443-26e551884fc6">Clinical Trial Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_125c2dd5-a8d7-490a-8041-79217705db5e">Segmentation Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_76c611e1-07c1-4604-8c74-f17d22a60b14">Frame of Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>C</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d8dc6eb8-b533-4971-b033-6d51b4a468a5">General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6dbff0a3-34a1-476d-9d2d-28d0a86f47dc">Enhanced General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_be089288-6f04-43d2-aba6-762f9529e9bb">Acquisition</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1">
                                <para xml:id="para_0fbdf726-7180-4144-8dd2-cea2c9ec763b">Multi-Resolution Pyramid</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7fee515d-53e0-4b47-a140-b3daaf1934fa">General Image</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7efbfdfb-cb76-48f3-bda2-a699f86633c3">General Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1">
                                <para xml:id="para_719b53f3-1e65-40ec-8527-0396967b6b23">Microscope Slide Layer Tile Organization</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>C</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_986bf306-18e5-424d-a858-225efef97add">Image Pixel</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a47075b7-f211-4e23-9bef-b923c535bcdf"><emphasis role="bold"><emphasis role="underline">Palette Color Lookup Table</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d0619215-9561-41e1-82c5-c6e79e216c93">Multi-frame Functional Groups</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f454de2a-41af-44a2-8659-f819f8e80bf4">Multi-frame Dimension</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d13acad0-e908-4f3e-8a62-9925dccfa98b">Specimen</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c7f4e64e-0dd2-4e23-ab55-407a54d880cc">Segmentation Image</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2b4802ea-8e5b-4697-b549-82e141d604b7">Common Instance Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>C</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f9382869-2cc5-45db-82dd-35e350758af6"><emphasis role="bold"><emphasis role="underline">ICC Profile</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">C</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6d30e5c8-ff2e-4c56-9773-d9b4160d0555">SOP Common</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0669d41f-f9b1-4d8d-a23c-c14fcc5683c3">Frame Extraction</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>C</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
        </section>
        
        <section label="A.51" status="2" xml:id="sect_A.51">
            <title>Segmentation IOD</title>
            <section label="A.51.1" status="3" xml:id="sect_A.51.1">
                <title>Segmentation IOD Description</title>
                <para xml:id="para_baa57ae2-b0de-4fe3-901d-9ac72408deac">The Segmentation IOD specifies a multi-frame image representing a classification of pixels in one or more referenced images. Segmentations are <emphasis role="bold"><emphasis role="strikethrough">either </emphasis></emphasis>binary<emphasis role="bold"><emphasis role="underline">,</emphasis></emphasis><emphasis role="bold"><emphasis role="strikethrough"> or</emphasis></emphasis> fractional<emphasis role="bold"><emphasis role="underline">, or label map</emphasis></emphasis>. If the referenced images have a defined Frame of Reference, the Segmentation Instance shall have the same Frame of Reference and is not required to have the same spatial sampling or extent as the referenced images. If the referenced image does not have a defined Frame of Reference, the Segmentation Instance shall have the same spatial sampling and extent as the referenced image.</para>
                <para xml:id="para_a94599e8-4986-4c13-8ef5-65236ee116da">The Segmentation IOD does not include the full set of acquisition parameters of the referenced images, e.g., cardiac phase. An application rendering or processing the <emphasis role="bold"><emphasis role="strikethrough">s</emphasis><emphasis role="underline">S</emphasis></emphasis>egmentation <emphasis role="bold"><emphasis role="underline">Instance </emphasis></emphasis>may need to access the referenced images for such information.</para>
                <para><emphasis role="bold"><emphasis role="underline">The Segmentation IOD is used in two SOP Classes as defined in PS3.4 <xref linkend="sect_B.5.1.25" xrefstyle="select: label"/><!--<olink targetdoc="PS3.4" targetptr="sect_B.5.1.25" xrefstyle="template:Section %n in PS3.4"/>--> Storage Service Class, a SOP Class for storage of BINARY and FRACTIONAL segmentations, and a SOP Class for storage of LABELMAP segmentations. These are distinguished by their SOP Class UID and by the Enumerated Value of the Segmentation Type (0062,0001) of the <xref linkend="sect_C.8.20.2" xrefstyle="select: title"/></emphasis></emphasis>.</para>
            </section>
            <section label="A.51.2" status="3" xml:id="sect_A.51.2">
                <title>Segmentation IOD Entity-Relationship Model</title>
                <para xml:id="para_096f590c-fe5b-43c8-9e5c-8fa8bc249e68">This IOD uses the E-R Model in Section A.1.2 <!--<xref linkend="sect_A.1.2" xrefstyle="select: label"/>-->, with only the Image IE below the Series IE.</para>
            </section>
            <section label="A.51.3" status="3" xml:id="sect_A.51.3">
                <title>Segmentation IOD Module Table</title>
                <para xml:id="para_25af160f-377a-4db7-900a-ed8f86620d3b">
                    <xref linkend="table_A.51-1" xrefstyle="select: label"/> specifies the Modules of the Segmentation IOD.</para>
                <table frame="box" label="A.51-1" rules="all" xml:id="table_A.51-1">
                    <caption>Segmentation IOD Modules</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_74031506-f227-413e-b895-0754a78e790d">IE</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_80d21f8b-b8b3-4c7b-8b99-a35665a89f5e">Module</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_81535ed2-b3ff-4ade-b371-76238a4dbe86">Reference</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_0d537144-b05a-4d32-a2f3-6f756bfa8969">Usage</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="2">
                                <para xml:id="para_61fa9c7c-ed12-40c4-bb49-d4382808a051">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6687bc5c-0317-47b8-9332-cf47c1140ea5">Patient</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4e58ebac-6854-4658-bfb9-26e1a83ad823">
                                    C.7.1.1<!--<xref linkend="sect_C.7.1.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b01560b0-a517-450f-a1dd-b4d4d945e22a">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9ce09931-986b-4e88-b43a-ed01b383ca12">Clinical Trial Subject</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_293a0110-7d47-4e13-aa26-a0cad0bed820">
                                    C.7.1.3<!--<xref linkend="sect_C.7.1.3" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_31c8f53f-8f6c-461f-b42d-3f78fd27bf8d">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="3">
                                <para xml:id="para_bb0b42ab-4293-4c9f-895a-b92133a00571">Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d6e34c69-5f34-4074-b416-191bfddacb1d">General Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_eff0ede9-b9d3-4925-bc88-3570477fac27">
                                    C.7.2.1<!--<xref linkend="sect_C.7.2.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0962da60-5b2a-45cc-83bd-2327f01f34a5">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_24344a5c-32da-4b65-8d02-b31e57d43171">Patient Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f9c5c340-2fe2-4210-8450-0e5d03ac84d7">
                                    C.7.2.2<!--<xref linkend="sect_C.7.2.2" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c78b2ae9-81d8-4c02-aa31-e48b51fe1df6">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_167d1a11-1c4e-4a1b-835c-1dafd71c9d1e">Clinical Trial Study</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_954d042d-b169-459f-bb94-a47a8d279fdd">
                                    C.7.2.3<!--<xref linkend="sect_C.7.2.3" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b65dab7a-3d80-4df2-a1fb-c71e6ddbfa15">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="3">
                                <para xml:id="para_4087ad98-d3e6-4b93-b0f7-29325e2250a6">Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_69bd25af-7176-400d-a6c7-8e61eb9ed670">General Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a655f2d0-5439-457e-ac9a-43f19d21347a">
                                    C.7.3.1<!--<xref linkend="sect_C.7.3.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5ad6141e-0d30-4798-babb-cbf55f8aa3c4">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c5eb9609-49a6-4886-b499-95b1b99109dc">Segmentation Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_17d51764-46aa-4217-bee8-d926cfd76ea0">
                                    C.8.20.1<!--<xref linkend="sect_C.8.20.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fbe2b5d5-6fdd-4045-9988-2db57afe8e52">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_545537bb-a343-47e2-b2b5-a8ba8d242298">Clinical Trial Series</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6d3588cb-ff03-4185-8e47-0be986516583">
                                    C.7.3.2<!--<xref linkend="sect_C.7.3.2" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9971ff53-bb40-4e5a-b989-973b8119e36f">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_20b67227-7f0e-4523-aaa2-0ff552ce278f">Frame of Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bd835efa-9c35-4a14-b15f-e24a234c457b">Frame of Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_22328553-8270-44ac-8e8b-5343168d9169">
                                    C.7.4.1<!--<xref linkend="sect_C.7.4.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_737ce382-7dd6-4b66-bdb3-6863521ec17b">C - Required if Derivation Image Functional Group (C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->) is not present. May be present otherwise.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="2">
                                <para xml:id="para_7dd7e324-3cc9-4623-bc45-28cf704e46d6">Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_aa5fcdf2-928f-4b27-b9ba-363316db9211">General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1546ff89-369f-454e-a0e7-44120a02aca0">
                                    C.7.5.1<!--<xref linkend="sect_C.7.5.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f198a467-18d1-4858-9234-4e603522c3fb">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_75fb6aef-fe86-424b-8b81-5365f2712cd6">Enhanced General Equipment</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_760e3be8-8c10-42ff-914e-aae0e1643e2a">
                                    C.7.5.2<!--<xref linkend="sect_C.7.5.2" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fac4c402-c33c-4f79-9319-9cee29528671">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_1e959fcd-a484-4848-ac39-0d7c3e98bedb">Acquisition</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ebaa6a77-3955-40de-af16-e05c4731c187">General Acquisition</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8dd75090-c6f3-4cb5-af94-37591fd45674">
                                    C.7.10.1<!--<xref linkend="sect_C.7.10.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_18545258-4975-49ac-bf7a-4abd43520140">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9bd5b5b0-172f-486e-8798-8bdcb4834eb4">Multi-Resolution Pyramid</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_59d3bafc-565f-458e-a886-037f118d9582">Multi-Resolution Pyramid</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_222ac1ee-6c93-4c51-bf0b-45c0bf8ff63f">
                                    C.7.11.1<!--<xref linkend="sect_C.7.11.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5ab5828d-0d92-4de4-855b-2e7a316ecd61">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="13">
                                <para xml:id="para_a29120de-8c16-4db3-a974-7f51a2ad6955">Image</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cd8f2ea0-7bb8-4826-aada-a3212bbeee9d">General Image</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f10bcec8-5fe9-48dc-bbd2-10d1878969d2">
                                    C.7.6.1<!--<xref linkend="sect_C.7.6.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fa06a7ac-3fff-49ce-82b4-5d686af3a36e">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_955bdd4b-6459-483e-85c2-7793b57513e7">General Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_11252dc9-2a81-434e-830b-91618b693a4a">
                                    C.12.4<!--<xref linkend="sect_C.12.4" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0affef98-f05d-4596-beaf-6a077698236e">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c1ae1d16-6e1a-4434-a88a-61b02d8d12e6">Microscope Slide Layer Tile Organization</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_da7b1814-cbb1-44f5-8db1-4120a82b0e46">
                                    C.8.12.14<!--<xref linkend="sect_C.8.12.14" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c90bda27-b562-4fce-a4f3-5c04ea21ef77">C - Required if Dimension Organization Type (0020,9311) is present with a value of TILED_FULL. May be present otherwise.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e50f4b4f-58cd-4cca-b6d7-a36e358dd470">Image Pixel</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b5cda1bc-825e-4183-89e2-c47b5010627c">
                                    C.7.6.3<!--<xref linkend="sect_C.7.6.3" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_72e8f780-d9e4-424e-b40c-a339d505dc37">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c7a4cb1c-7ccd-4230-ac88-81edbe76bdb3">Segmentation Image</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_89ae50d5-4575-4b7c-8796-fb7c89dcad16">
                                    C.8.20.2<!--<xref linkend="sect_C.8.20.2" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7bb90311-4656-4b15-bd60-4208a3d069d1">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_056ca2d0-36e3-4ed5-b11e-7228e925bdf1">Multi-frame Functional Groups</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_80853fb0-7713-4efe-a26f-600fbac0dc18">
                                    C.7.6.16<!--<xref linkend="sect_C.7.6.16" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9ce25fa3-dd1f-44f7-9286-c6f7d23e33d0">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5f15d1fe-052e-4c69-b3f4-efe7714fe5ee">Multi-frame Dimension</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_bdb7811d-a57e-42bd-ad4c-6bec5e3a0eca">
                                    C.7.6.17<!--<xref linkend="sect_C.7.6.17" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e669a4ea-95b3-4515-9a9a-895843154d23">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">Palette Color Lookup Table</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>
                                    <emphasis role="bold"><emphasis role="underline">C.7.9<!--<xref linkend="sect_C.7.9" xrefstyle="select: labelnumber"/>--></emphasis></emphasis>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR.</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ddbfe10b-a6e3-4f87-9e7f-08cb30a3c004">Specimen</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ebb6b782-1aac-4494-aa37-a38e85f1c05e">
                                    C.7.6.22<!--<xref linkend="sect_C.7.6.22" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_15497950-b16e-4af6-a69c-3ace84d988ef">U</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a1c362f4-a607-437f-bfd0-761cc9eae3f3">Common Instance Reference</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d299f5a3-9292-4821-b288-b449db4d314b">
                                    C.12.2<!--<xref linkend="sect_C.12.2" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9294d1cc-c81b-4a81-9524-36a22cdf238a">C - Required if Derivation Image Functional Group (C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->) is present.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">ICC Profile</emphasis></emphasis></para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para>
                                    <emphasis role="bold"><emphasis role="underline">C.11.15<!--<xref linkend="sect_C.11.15" xrefstyle="select: labelnumber"/>--></emphasis></emphasis>
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para><emphasis role="bold"><emphasis role="underline">C - Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR.</emphasis></emphasis></para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9238de34-6acd-4a48-b258-9a90f3847315">SOP Common</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_886867bf-6abb-41c4-8480-4358af65241d">
                                    C.12.1<!--<xref linkend="sect_C.12.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f0a5b66f-4e3b-4891-93a6-9cb316b92c58">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_99fe4fe9-a2c5-421a-b621-7b861e4eb5b3">Frame Extraction</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_88aac54f-7f29-4492-beb7-e971cd65089b">
                                    C.12.3<!--<xref linkend="sect_C.12.3" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_13ed0766-6818-4342-bf64-698e3c976023">C - Required if the SOP Instance was created in response to a Frame-Level retrieve request</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
                <note>
                    <para xml:id="para_c5773c29-8d91-4c83-9cc8-23727fdbcbf0">The Specimen Identification Module (Retired)<!--<xref linkend="sect_C.7.1.2" xrefstyle="select: title"/>--> was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module<!--<xref linkend="sect_C.7.6.22" xrefstyle="select: title"/>-->. See PS3.3-2008.</para>
                </note>
            </section>
            <section label="A.51.4" status="3" xml:id="sect_A.51.4">
                <title>Segmentation IOD Content Constraints</title>
                <para xml:id="para_be960443-3d98-41a4-ae00-879d26774b19">The VOI LUT Module<!--<xref linkend="sect_C.11.2" xrefstyle="select: title"/>--> shall not be present.</para>
                <para xml:id="para_0566f148-4956-4422-b5fb-e918149106ac">The Modality LUT Module<!--<xref linkend="sect_C.11.1" xrefstyle="select: title"/>--> shall not be present.</para>
                <para xml:id="para_939be885-e0eb-4d74-92bd-afaf9e462503">The Overlay Plane Module<!--<xref linkend="sect_C.9.2" xrefstyle="select: title"/>--> shall not be present.</para>
                <para xml:id="para_e912144a-88b2-43b9-8cc7-bc9fa4eea774">Pixel Padding Value (0028,0120) shall not be present <emphasis role="bold"><emphasis role="underline">if Segmentation Type (0062,0001) is not LABELMAP</emphasis></emphasis>.<emphasis role="bold"><emphasis role="underline"> Pixel Padding Value (0028,0010) may be used to indicate that a particular segment is background if Segmentation Type (0062,0001) is LABELMAP; see Pixel Padding Value (0028,0120) in <xref linkend="sect_C.7.5.1.1.2" xrefstyle="select: label"/>.</emphasis></emphasis></para>
                <para><emphasis role="bold"><emphasis role="underline">Pixel Padding Range Limit (0028,0121) shall not be present.</emphasis></emphasis></para>
                <para xml:id="para_61a37ab1-94da-46d1-8cf7-7735bd0dc51b">For Purpose of Reference Code Sequence (0040,A170) within Source Instance Sequence (0042,0013) in the General Reference Module<!--<xref linkend="sect_C.12.4" xrefstyle="select: title"/>--> D<olink targetdoc="PS3.16" targetptr="sect_CID_7019" xrefstyle="select: labelnumber quotedtitle"/> shall be used.</para>
            </section>
            <section label="A.51.5" status="3" xml:id="sect_A.51.5">
                <title>Segmentation Functional Groups</title>
                <para xml:id="para_5ead2168-a679-4842-a1fd-7f6f8a18114a">
                    <xref linkend="table_A.51-2" xrefstyle="select: label"/> specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module<!--<xref linkend="sect_C.7.6.16" xrefstyle="select: title"/>--> for the Segmentation IOD.</para>
                <table frame="box" label="A.51-2" rules="all" xml:id="table_A.51-2">
                    <caption>Segmentation Functional Group Macros</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_f4dac09b-ec34-48a8-bc04-61d6eaeccf62">Functional Group Macro</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3ebeea64-2c0d-4aec-8ac4-34ca38a578f5">Section</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_07abb63b-8340-4a86-a957-b6129678667b">Usage</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d225a1f4-1580-4be6-8b8a-b0a0d94fc407">Pixel Measures</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_d718e2c3-00cc-447d-8313-f46175fe2cb6">
                                    C.7.6.16.2.1<!--<xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_edf2f027-832b-4bbb-b1d9-3a0b61c50a47">C - Required if Derivation Image Functional Group (C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->) is not present and the Frame of Reference is defined in the patient-relative Reference Coordinate System. May be present otherwise if the Frame of Reference is defined in the patient-relative Reference Coordinate System. See <xref linkend="sect_A.51.5.1" xrefstyle="select: label"/>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_666fccea-1bd6-4436-af0f-b03a6b64d2d5">Plane Position (Patient)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2ac099dc-e14c-4c20-8b5f-05a98a7a0626">
                                    C.7.6.16.2.3<!--<xref linkend="sect_C.7.6.16.2.3" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ec30dd1b-a5c6-41dd-a445-4423a8865a87">C - Required if Derivation Image Functional Group (C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->) is not present and the Frame of Reference is defined in the patient-relative Reference Coordinate System. May be present otherwise if the Frame of Reference is defined in the patient-relative Reference Coordinate System. See <xref linkend="sect_A.51.5.1" xrefstyle="select: label"/>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8eb3193e-9456-45f6-9124-a167bce5f9b9">Plane Orientation (Patient)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_f15036a1-4360-4ff4-a3be-6632cbb4c87a">
                                    C.7.6.16.2.4<!--<xref linkend="sect_C.7.6.16.2.4" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_46755a9c-db45-40ba-9cf0-b13114a7bbc2">C - Required if Derivation Image Functional Group (C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->) is not present and the Frame of Reference is defined in the patient-relative Reference Coordinate System. May be present otherwise if the Frame of Reference is defined in the patient-relative Reference Coordinate System. See <xref linkend="sect_A.51.5.1" xrefstyle="select: label"/>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3e0ef8b9-c64d-417e-b072-4372f1322139">Plane Position (Slide)</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6b261bbd-473c-409b-9b08-573396606725">
                                    C.8.12.6.1<!--<xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_20e543a1-9c89-42ad-b619-c161f3322d35">C - Required if Derivation Image Functional Group (C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->) is not present and the Frame of Reference is defined in the Slide Coordinate System and Dimension Organization Type (0020,9311) is not TILED_FULL. May be present otherwise if the Frame of Reference is defined in the Slide Coordinate System. See <xref linkend="sect_A.51.5.1" xrefstyle="select: label"/>.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_91c06f6e-a310-4239-b87b-152f277e8bb2">Derivation Image</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a526bf4f-3813-48b6-b5fe-82258a8f6126">
                                    C.7.6.16.2.6<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_59530188-7dc9-4e95-a726-ec244a4ffb19">C - Required if Pixel Measures (C.7.6.16.2.1<!--<xref linkend="sect_C.7.6.16.2.1" xrefstyle="select: labelnumber"/>-->) or either Plane Position (Patient) (C.7.6.16.2.3<!--<xref linkend="sect_C.7.6.16.2.3" xrefstyle="select: labelnumber"/>-->) or Plane Orientation (Patient) (C.7.6.16.2.4<!--<xref linkend="sect_C.7.6.16.2.4" xrefstyle="select: labelnumber"/>-->) (if the Frame of Reference is defined in the patient-relative Reference Coordinate System), or Plane Position (Slide) (C.8.12.6.1<!--<xref linkend="sect_C.8.12.6.1" xrefstyle="select: labelnumber"/>-->) (if the Frame of Reference is defined in the Slide Coordinate System) Functional Groups are not present. May be present otherwise. See <xref linkend="sect_A.51.5.1" xrefstyle="select: label"/>
                                </para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_12f90a9f-cba9-4b47-a388-469cca7cdd07">Frame Content</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_0c59c690-b0f2-4164-872c-5c273eb4521e">
                                    C.7.6.16.2.2<!--<xref linkend="sect_C.7.6.16.2.2" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_901e8b81-50fe-4fee-b851-028ef2058a47">M</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_cb0c73da-b198-4fa8-b66c-f49d967a5244">Segmentation</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3be6feb6-0c95-4371-817b-313342a455bc">
                                    C.8.20.3.1<!--<xref linkend="sect_C.8.20.3.1" xrefstyle="select: labelnumber"/>-->
                                </para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_ec95e942-28e9-461b-8564-0a619598f9fb">C - Required if Dimension Organization Type (0020,9311) is not TILED_FULL<emphasis role="bold"><emphasis role="underline"> and Segmentation Type (0062,0001) is not LABELMAP.</emphasis></emphasis></para>
                            </td>
                        </tr>
                    </tbody>
                </table>
                <note>
                    <para xml:id="para_37c644ed-7f9b-433a-b4b0-3360fe635c72">The Plane Position (Slide), Frame Content and Segmentation Macros are Type C, which allows the Per-Frame Functional Group Sequence (5200,9230)
                        to be entirely omitted in those cases in which there are no other Per-Frame Functional Group Macros with content
                        (i.e., the Frame Content Macro is empty or absent).</para>
                </note>
                <section label="A.51.5.1" status="4" xml:id="sect_A.51.5.1">
                    <title>Segmentation Functional Groups Description</title>
                    <para xml:id="para_31f0915e-245f-409a-8307-e332181a2e3c">When a Frame of Reference UID is present the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment shall be specified within that coordinate system, using the Pixel Measures and either the Plane Position (Patient) and Plane Orientation (Patient), or the Plane Position (Slide) Functional Groups. Since this defines the spatial relationship of the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment, the size of the segmentation frames need not be the same size, or resolution, as the image data used to generate the segment data. The Derivation Image Functional Group may also be present, to specify on which images the segmentation was actually performed (since there may be others in the same Frame of Reference that are spatially co-located, but were not used to perform the segmentation).</para>
                    <para xml:id="para_d40a5ab6-c9c4-4a90-9da0-36f91fa8f599">If the Frame of Reference UID is not present, each pixel of the segmentation shall correspond to a pixel in a referenced image, using the Derivation Image Functional Group. Hence, the rows and columns of each referenced image will match the segmentation image. If both the Frame of Reference UID and the Derivation Image Functional Group are present, the segmentation and referenced image pixels need not correspond.</para>
                    <para xml:id="para_4a034de9-2d78-47e1-b66b-663baca9730d">The value of Purpose of Reference Code Sequence (0040,A170) in the Derivation Image Macro<!--<xref linkend="sect_C.7.6.16.2.6" xrefstyle="select: title"/>--> shall be <olink targetdoc="PS3.16" targetptr="DCM_121322">(121322, DCM, "Source Image for Image Processing Operation")</olink>. The value of Derivation Code Sequence (0008,9215) shall be <olink targetdoc="PS3.16" targetptr="DCM_113076">(113076, DCM, "Segmentation")</olink>.</para>
                    <note>
                        <para xml:id="para_a53a9c5c-79b1-416e-9c6c-ea96717b9f7e">Non-image source Instances used during segmentation, such as Real World Value maps, can be described in the top level Data Set in the Source Instance Sequence (0042,0013) of the General Reference Module<!--<xref linkend="sect_C.12.4" xrefstyle="select: title"/>--> and are implied to have been used for the derivation of all frames. I.e., there is no mechanism for selectively specifying on a per-frame basis which non-Image Instances were used. Real World Value Map Instances already contain a means of selectively applying different scale factors to different frames.</para>
                    </note>
                </section>
            </section>
        </section>
        
        <section label="C.8.20" status="3" xml:id="sect_C.8.20">
                <title>Segmentation</title>
                <para xml:id="para_9f68ac61-5fe1-4c52-a493-b58faf91e6c7">This section describes the specific Modules for the Segmentation IOD.</para>
                <section label="C.8.20.1" status="4" xml:id="sect_C.8.20.1">
                    <title>Segmentation Series Module</title>
                    <para xml:id="para_fcfb1b1a-7130-40ad-b650-ec6a22f77c91">
                        <xref linkend="table_C.8.20-1" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.8.20.1" xrefstyle="select: title"/>.</para>
                    <table frame="box" label="C.8.20-1" rules="all" xml:id="table_C.8.20-1">
                        <caption>Segmentation Series Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f6574bb2-3e21-4697-a341-b4fb36778b78">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_384f1b12-7b3f-42ef-b2fb-e047b5da1ccf">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_79921e92-443b-4db4-b7ae-7b0f14901303">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a60bf9e5-a53e-4910-be2c-a66f6ac40de1">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_539fbf3e-7809-489b-890e-35bbee3cef00">Modality</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_18093d63-c429-4fbd-8f2c-1a85aa056aa9">(0008,0060)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_887fb7e9-dca0-4462-9fc9-3c225fce899e">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_5339460d-d3c8-4e1c-8345-656b63f27652">Type of device, process or method that created the Instances in this Series<emphasis role="bold"><emphasis role="underline">.</emphasis></emphasis></para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>SEG</term>
                                            <listitem>
                                                <para xml:id="para_82483ecf-fb9b-469c-8de0-de4815766af7"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8c2c4600-97fe-44c0-8048-175e369ea4a6">Series Number</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1d6ce601-8b35-4a33-aa5f-0809ea0476ab">(0020,0011)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f6a120cf-e7e8-4e25-b48c-82644947648b">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bf545cb9-d336-4903-ad37-1e03dfc655f4">A number that identifies this Series.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cae8995b-e627-4a56-a42b-d0c8a73fae54">Referenced Performed Procedure Step Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_d79ca6c9-fedc-4ff3-b8ce-58e5234598ca">(0008,1111)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2470a1b7-1b8f-4322-8442-36028210aa26">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9ad16d8b-a398-4a11-84d8-9dcc518fb0ed">Uniquely identifies the Performed Procedure Step SOP Instance to which the Series is related<emphasis role="bold"><emphasis role="underline">.</emphasis></emphasis></para>
                                    <para xml:id="para_6169e6e1-b0b7-45de-9cb9-e355513cbad4">Only a single Item shall be included in this Sequence.</para>
                                    <para xml:id="para_9d19cbf1-4b5f-4013-9fd4-ae83c3c1bb4a">Required if a Performed Procedure Step SOP Class was involved in the creation of this Series.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="4" rowspan="1">
                                    <para xml:id="para_89802988-3c9e-482a-9ebe-bc095320c267">
                                        <emphasis role="italic">&gt;Include Table 10-11 "SOP Instance Reference Macro Attributes"<!--<xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis>
                                    </para>
                                </td>
                            </tr>
                        </tbody>
                    </table>
                </section>
                <section label="C.8.20.2" status="4" xml:id="sect_C.8.20.2">
                    <title>Segmentation Image Module</title>
                    <para xml:id="para_5888c90a-6c54-4953-949c-c175c1cde813">
                        <xref linkend="table_C.8.20-2" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.8.20.2" xrefstyle="select: title"/>.</para>
                    <table frame="box" label="C.8.20-2" rules="all" xml:id="table_C.8.20-2">
                        <caption>Segmentation Image Module Attributes</caption>
                        <thead>
                            <tr valign="top">
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_5ac516a3-1159-4ff7-b037-2e9d8f9240fd">Attribute Name</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f09fb47c-78cd-4775-9d95-171f90cd203a">Tag</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a1bccd2a-d522-40b7-9120-db796ab8fc01">Type</para>
                                </th>
                                <th align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b9de38ed-1b90-4723-afe4-5080da14ee7e">Attribute Description</para>
                                </th>
                            </tr>
                        </thead>
                        <tbody>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_a8da15f4-7601-47ac-9c43-de55402d4337">Image Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_271bfb8b-b010-4ffa-9cce-fb0bbeac9551">(0008,0008)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ad86d6e1-7636-4e59-a32c-ff37ac0c3ce6">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Image identification characteristics.</emphasis></emphasis></para>
                                    <para xml:id="para_191f0ff0-a887-43e2-b703-3bdac8373a02">Value 1 shall be DERIVED. Value 2 shall be PRIMARY. No other values shall be present.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="4" rowspan="1">
                                    <para xml:id="para_d70090a9-0afc-4819-b19d-e8fe5b19805c">
                                        <emphasis role="italic">Include Table 10-12 "Content Identification Macro Attributes"<!--<xref linkend="table_10-12" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7e8be640-10b6-441f-a367-e3c402ff596c">Samples Per Pixel</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_6b9640c4-2823-4086-a0b0-96634e5d206c">(0028,0002)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4f4a9610-dacc-43d5-aa79-27199179ca98">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Number of samples (planes) in this image.</emphasis></emphasis></para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>1</term>
                                            <listitem>
                                                <para xml:id="para_c478ef23-25d1-4945-a777-2c1d7233a865"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_cc83b18b-874b-4399-be39-30c3f5c64eae">Photometric Interpretation</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8516540d-6b48-44e6-baf1-e3190c9bb69f">(0028,0004)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_ce7e340f-cc1b-48fb-93e3-479746014e84">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Specifies the intended interpretation of the pixel data.</emphasis></emphasis></para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values<emphasis role="bold"><emphasis role="underline"> if Segmentation Type (0062,0001) is BINARY or FRACTIONAL</emphasis></emphasis>:</title>
                                        <varlistentry>
                                            <term>MONOCHROME2</term>
                                            <listitem>
                                                <para xml:id="para_8e13fa16-ef94-4273-9bf1-4476d60c86f2"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact" termlength="3in">
                                        <title><emphasis role="bold"><emphasis role="underline">Enumerated Values if Segmentation Type (0062,0001) is LABELMAP</emphasis></emphasis>:</title>
                                        <varlistentry>
                                            <term><emphasis role="underline">MONOCHROME2</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="underline">PALETTE COLOR</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_57418235-bd41-44c7-93d8-c2d04312ed95">Pixel Representation</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_395d07b6-513c-4d93-93c8-348cffe6ad7d">(0028,0103)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_12b54ebe-3073-4445-a509-4d6a6745f6e3">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Data representation of pixel samples.</emphasis></emphasis></para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>0</term>
                                            <listitem>
                                                <para xml:id="para_751cd3e0-bcf5-4bb9-bcaa-0518af37cbf5"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9acabd73-b59b-435e-9f23-985bd2eb285a">Bits Allocated</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b4e6df18-6c87-4d41-b852-942c23a89173">(0028,0100)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f28ef735-fb70-417a-b159-bceb26d1bf7e">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Number of bits allocated for each pixel sample.</emphasis></emphasis></para>
                                    <para xml:id="para_4e664ec9-26a8-49b2-a725-163407bbfeb0">See <xref linkend="sect_C.8.20.2.1" xrefstyle="select: label"/>.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Segmentation Type (0062,0001) is BINARY:</title>
                                        <varlistentry>
                                            <term>1</term>
                                            <listitem>
                                                <para xml:id="para_b60940ab-61b1-4256-83a7-735e03f76107"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Segmentation Type (0062,0001) is <emphasis role="strikethrough">not BINARY</emphasis><emphasis role="underline">FRACTIONAL</emphasis>:</title>
                                        <varlistentry>
                                            <term>8</term>
                                            <listitem>
                                                <para xml:id="para_2f16b24f-2311-4b82-b01e-e6035dbc359f"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title><emphasis role="underline">Enumerated Values if Segmentation Type (0062,0001) is LABELMAP</emphasis>:</title>
                                        <varlistentry>
                                            <term><emphasis role="underline">8</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="underline">16</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7e8fb1c6-2c53-440c-bd95-45bc14e3ff3c">Bits Stored</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c3dcfd44-a6e9-4194-b127-eff220bd0652">(0028,0101)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7529cab1-cea9-4e62-b131-b401fd543b31">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Number of bits stored for each pixel sample.</emphasis></emphasis></para>
                                    <para xml:id="para_c1555166-96ec-4e51-afc7-2903cc32f363">See <xref linkend="sect_C.8.20.2.1" xrefstyle="select: label"/>.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Segmentation Type (0062,0001) is BINARY:</title>
                                        <varlistentry>
                                            <term>1</term>
                                            <listitem>
                                                <para xml:id="para_479cf21e-11cc-40bb-a11f-5a2c8ff3d15f"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Segmentation Type (0062,0001) is <emphasis role="strikethrough">not BINARY</emphasis><emphasis role="underline">FRACTIONAL</emphasis>:</title>
                                        <varlistentry>
                                            <term>8</term>
                                            <listitem>
                                                <para xml:id="para_f3ae8cef-3801-4bbf-891c-f7d60e6a5e70"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title><emphasis role="underline">Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 8</emphasis>:</title>
                                        <varlistentry>
                                            <term><emphasis role="underline">8</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title><emphasis role="underline">Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 16</emphasis>:</title>
                                        <varlistentry>
                                            <term><emphasis role="underline">16</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_f52df353-8ce1-42a2-a0ea-252c9d08a2e5">High Bit</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_c64fb64f-3890-437d-8b75-0dfc789afad1">(0028,0102)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_93b22e49-d575-45b4-aed6-e43d59a374e9">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para><emphasis role="bold"><emphasis role="underline">Most significant bit for pixel sample data.</emphasis></emphasis></para>
                                    <para xml:id="para_bf037e32-4365-4758-8c93-d1ed649c4b82">See <xref linkend="sect_C.8.20.2.1" xrefstyle="select: label"/>.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Segmentation Type (0062,0001) is BINARY:</title>
                                        <varlistentry>
                                            <term>0</term>
                                            <listitem>
                                                <para xml:id="para_9dafecc5-a258-44ed-a568-1f3d0bb694ef"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values if Segmentation Type (0062,0001) is <emphasis role="strikethrough">not BINARY</emphasis><emphasis role="underline">FRACTIONAL</emphasis>:</title>
                                        <varlistentry>
                                            <term>7</term>
                                            <listitem>
                                                <para xml:id="para_f5f5e12a-e7c5-4dc2-b9d6-c63e523f8525"/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title><emphasis role="underline">Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 8</emphasis>:</title>
                                        <varlistentry>
                                            <term><emphasis role="underline">7</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <variablelist spacing="compact">
                                        <title><emphasis role="underline">Enumerated Values if Segmentation Type (0062,0001) is LABELMAP and Bits Allocated (0028,0100) is 16</emphasis>:</title>
                                        <varlistentry>
                                            <term><emphasis role="underline">15</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0b5cd0fa-a61a-4f26-b0f3-41a44997787b">Lossy Image Compression</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a98b0b45-79ff-483a-99d4-786447c3c781">(0028,2110)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f56f8f6a-0725-4fe9-b8fb-8e712b63aa72">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_8692e564-6833-4231-b575-7a7a5144fea5">Specifies whether an Image has undergone lossy compression (at a point in its lifetime), or is derived from lossy compressed images.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>00</term>
                                            <listitem>
                                                <para xml:id="para_f1aa33cd-3de9-4ab0-84f2-1d77d2f7a478">Image has NOT been subjected to lossy compression.</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>01</term>
                                            <listitem>
                                                <para xml:id="para_d2059e43-ff7e-45ea-a11f-ba0095eee65b">Image has been subjected to lossy compression.</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para xml:id="para_10384443-9205-4dac-8378-b66f4d12fb4b">Once this value has been set to "01" it shall not be reset.</para>
                                    <para xml:id="para_33ff8435-c837-4797-b09e-7f2f23936b34">See <xref linkend="sect_C.8.20.2.2" xrefstyle="select: label"/> and Section C.7.6.1.1.5<!--<xref linkend="sect_C.7.6.1.1.5"/>-->.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_65c81f8e-c2ca-4244-b8d9-5bc297a47f01">Lossy Image Compression Ratio</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_500c1085-8132-4329-a5f7-e6ba37772a61">(0028,2112)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b9025667-4aed-43e4-8db1-ca6e0712179b">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_53e16485-d6e3-4a36-9767-9138e826cd16">Describes the approximate lossy compression ratio(s) that have been applied to this image.</para>
                                    <para xml:id="para_ebf96a34-b7b1-4fb4-9f9c-0f44f10ac00d">See Section C.7.6.1.1.5.2 <!--<xref linkend="sect_C.7.6.1.1.5.2" xrefstyle="select: label"/>-->.</para>
                                    <para xml:id="para_0df4d4f7-642b-4b2c-9f59-c311e47f154a">Required if present in the source images or this IOD Instance has been compressed.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_04e490ae-b525-4f70-8958-4a3ae81d9791">Lossy Image Compression Method</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_f8a4fba4-1847-4a5f-97fa-be6106de176a">(0028,2114)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_7848f244-2ebc-4d13-bbc7-459a58ce5eed">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_79ebdebe-626c-4f54-8544-23f28c4a91c9">A label for the lossy compression method(s) that have been applied to this image.</para>
                                    <para xml:id="para_8a003f0e-c2e6-4085-ae18-af26a2629cdf">See Section C.7.6.1.1.5.1 <!--<xref linkend="sect_C.7.6.1.1.5.1" xrefstyle="select: label"/>-->.</para>
                                    <para xml:id="para_a9392969-5c1c-470b-8225-f116593d1286">Required if present in the source images or this IOD Instance has been compressed. See <xref linkend="sect_C.8.20.2.2" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_678b7133-d3bc-4ca7-8886-fbbaffdd5fbb">Segmentation Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b3bcb2e1-ecdb-47c2-adab-44c8021c7743">(0062,0001)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_db339dac-4982-4579-ae46-26c38672f845">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_30da4277-2b17-4813-ac33-06bb5ac40caa">The type of encoding used to indicate the presence of the segmented property at a pixel/voxel location.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>BINARY</term>
                                            <listitem>
                                                <para xml:id="para_a83fdb8b-a3b8-4b66-8a42-7984af131335"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>FRACTIONAL</term>
                                            <listitem>
                                                <para xml:id="para_4505e7c6-9d6f-4909-b22d-d4e2e01d30d6"/>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term><emphasis role="underline">LABELMAP</emphasis></term>
                                            <listitem>
                                                <para/>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para xml:id="para_780d5a72-ad66-40b1-bdf8-8438ad5160b2">See <xref linkend="sect_C.8.20.2.3" xrefstyle="select: label"/>.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_adc4b8c2-6e00-47b1-85df-5958eb2dfde2">Segmentation Fractional Type</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_46f09e41-8540-4972-a301-4d077886bb5a">(0062,0010)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_2005e708-2dca-4ee8-8713-e99a803e2a14">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_ac475378-81d7-4cfb-b390-be868e620a3a">For fractional segmentation encoding, the meaning of the fractional value.</para>
                                    <para xml:id="para_eb1e0912-08d5-4e58-8506-85b01e514ef3">Required if Segmentation Type (0062,0001) is FRACTIONAL.</para>
                                    <para xml:id="para_ef9659e5-e169-4d95-9ce6-64b05def310c">See <xref linkend="sect_C.8.20.2.3" xrefstyle="select: label"/> for Enumerated Values.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fa78e6a8-c8b7-47ac-816b-0a921a026651">Maximum Fractional Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9c8e6af4-6d02-46f1-9b27-01ac0e83d092">(0062,000E)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_8195956a-7b31-46a3-a11e-8649ae637811">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_bb2c60c5-2ecd-4518-9686-1108470c458c">Specifies the value that represents a probability of 1 or complete occupancy. There shall be no values in Pixel Data (7FE0,0010) greater than this value. Required if Segmentation Type (0062,0001) is FRACTIONAL.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_7e7e9f0d-bb13-4c4f-afb2-d21208a2cf15">Segments Overlap</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_9ed5178a-6052-460c-b3ef-6393bed965af">(0062,0013)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_e4c5baad-18c7-46f7-8c02-b36a28758652">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0ae928c4-6e03-4d4f-a17a-fc057c44ac1a">Whether or not any <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egments in this Instance overlap. I.e., whether or not any pixel is or might be in more than one <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment.</para>
                                    <variablelist spacing="compact">
                                        <title>Enumerated Values:</title>
                                        <varlistentry>
                                            <term>YES</term>
                                            <listitem>
                                                <para xml:id="para_f2b467c7-f1b7-4289-bd1b-1687d494faba">Some <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egments overlap</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>UNDEFINED</term>
                                            <listitem>
                                                <para xml:id="para_1363e172-001a-47e0-a57b-fbb2072cb748">Some <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egments might overlap</para>
                                            </listitem>
                                        </varlistentry>
                                        <varlistentry>
                                            <term>NO</term>
                                            <listitem>
                                                <para xml:id="para_4990160b-1c14-4dc9-8aa9-3d4854302a1d">No <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egments overlap</para>
                                            </listitem>
                                        </varlistentry>
                                    </variablelist>
                                    <para xml:id="para_b4466fbd-8f47-4d47-abd1-df1a77747429">See <xref linkend="sect_C.8.20.2.3" xrefstyle="select: label"/>.</para>
                                <para><emphasis role="bold"><emphasis role="underline">If present, shall be NO if Segmentation Type (0062,0001) is LABELMAP.</emphasis></emphasis></para>
                                    <note>
                                        <para xml:id="para_c5d47887-c2c5-4c3f-a2c7-ef870c7e09f9">If the value is NO, then a receiving application to which this matters can be assured that no <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egments overlap and does not need to check every pixel. If the value is UNDEFINED or YES, or the Attribute is absent, then a receiving application might need to check every pixel in every <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_d4c0699f-5c9c-4417-8fe5-346dedcbc9f4">Segment Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_1c35dc93-2e58-40b6-b2de-5afd23db4c70">(0062,0002)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_afad81ee-3943-47f3-be0a-2d311501711b">1</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_07748b92-199f-45a8-9cb8-84e406f20c1c">Describes the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egments that are contained within the data.</para>
                                    <para xml:id="para_8e7e6240-4092-42f2-982d-1bd2da9fdd36">One or more Items shall be included in this Sequence.</para>
                                    <note>
                                        <para><emphasis role="bold"><emphasis role="underline">The Items of this Sequence are not required to be in any particular order, i.e., are not required to be ordered by Segment Number (0062,0004).</emphasis></emphasis></para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="4" rowspan="1">
                                    <para xml:id="para_47ce1e6b-9c8e-4af9-b39f-a6cf1a38d939">
                                        <emphasis role="italic">&gt;Include <xref linkend="table_C.8.20-4" xrefstyle="select: label quotedtitle"/>
                                        </emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_9023426b-9d82-48f8-b30e-25c65ae7e699">&gt;Segment Algorithm Name</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_95d61e8a-980c-4823-82fe-0edb77396cd1">(0062,0009)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cf002e94-d7ce-45d5-9161-b2ff14e0adac">1C</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_0b98cef6-7890-409a-8ea2-0ad1c76871e1">The name(s) of algorithm(s) used to generate the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment. Required if Segment Algorithm Type (0062,0008) is not MANUAL.</para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_4bb57699-8baa-4632-8ea3-601c1cc5288b">&gt;Segmentation Algorithm Identification Sequence</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_4303e8ad-97fe-43d8-a17a-176e2a1eed69">(0062,0007)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_cb0ae3b2-f659-4e9b-ab33-caffc4e16080">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_94cafa9b-9d0c-4210-b020-62b6684b919c">A description of how this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment was derived.</para>
                                    <para xml:id="para_93218838-ce90-49cb-85af-b61f53675fe4">Algorithm Name (0066,0036) within this Sequence may be identical to Segment Algorithm Name (0062,0009).</para>
                                    <para xml:id="para_c2676a92-5bcf-40fa-ad7c-113e301e9c60">One or more Items are permitted in this Sequence.</para>
                                    <note>
                                        <para xml:id="para_030b06d5-3ff2-43a9-8c8e-b110a5525d23">Previously, the Segment Surface Generation Algorithm Identification Code Sequence (0066,002D) was used, but it has been replaced in this Module, since not all segmentation algorithms involve surface generation. See <link xl:href="http://dicom.nema.org/MEDICAL/Dicom/2016d/output/pdf/part03.pdf">PS3.3-2016d</link>.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="3" rowspan="1">
                                    <para xml:id="para_c2516bcb-2661-4309-9898-4c5f28b1688a">
                                        <emphasis role="italic">&gt;&gt;Include Table 10-19 "Algorithm Identification Macro Attributes"<!--<xref linkend="table_10-19" xrefstyle="select: label quotedtitle"/>-->
                                        </emphasis>
                                    </para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_3849d786-8f85-4a31-a9cb-fd4a281c1494">
                                        <emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_7162" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                                    </para>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_563351ad-a948-4882-97d3-d7f0b89a6db9">&gt;Recommended Display Grayscale Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_24f9453c-dbad-4d90-be9d-b8d60737874d">(0062,000C)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b4fec89c-f7e2-476c-bd47-05d9b2bc2200">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_66007bb9-159e-4cf3-b349-2b878c8b3a61">A default single gray unsigned value in which it is recommended that the maximum pixel value in this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment be rendered on a monochrome display. The units are specified in P-Values from a minimum of 0000H (black) up to a maximum of FFFFH (white).</para>
                                    <note>
                                        <para xml:id="para_1cfe33cf-043c-4a55-b393-0eeef96cda83">The maximum P-Value for this Attribute may be different from the maximum P-Value from the output of the Presentation LUT, which may be less than 16 bits in depth.</para>
                                    </note>
                                </td>
                            </tr>
                            <tr valign="top">
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_10a6e23d-6fee-4fc5-ba40-b1f0073f6ffc">&gt;Recommended Display CIELab Value</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_b697d7b7-6945-444b-a814-9e74c75a912a">(0062,000D)</para>
                                </td>
                                <td align="center" colspan="1" rowspan="1">
                                    <para xml:id="para_a376baac-a18b-44bd-b0f4-0fe4b3f9bc2f">3</para>
                                </td>
                                <td align="left" colspan="1" rowspan="1">
                                    <para xml:id="para_fea81a93-95f6-4da5-988a-84358398da66">A default triplet value in which it is recommended that <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment be rendered on a color display. The units are specified in PCS-Values, and the value is encoded as CIELab. See Section C.10.7.1.1<!--<xref linkend="sect_C.10.7.1.1" xrefstyle="select: label"/>-->.</para>
                                
                                    <para><emphasis role="bold"><emphasis role="underline">Shall not be present if Segmentation Type (0062,0001) is LABELMAP and Photometric Interpretation (0028,0004) is PALETTE COLOR.</emphasis></emphasis></para>
                                
                                </td>
                            </tr>
                        </tbody>
                    </table>
                    <section label="C.8.20.2.1" status="5" xml:id="sect_C.8.20.2.1">
                        <title>Bits Allocated<emphasis role="bold"><emphasis role="strikethrough"> and</emphasis><emphasis role="underline">,</emphasis></emphasis> Bits Stored<emphasis role="bold"><emphasis role="underline">, and High Bit</emphasis></emphasis></title>
                        <para xml:id="para_368ed49a-cf87-4df4-a191-403141130645">As a consequence of the <emphasis role="bold"><emphasis role="strikethrough">e</emphasis><emphasis role="underline">E</emphasis></emphasis>numerated <emphasis role="bold"><emphasis role="underline">Values for </emphasis></emphasis>Bits Allocated<emphasis role="bold"><emphasis role="strikethrough"> and</emphasis><emphasis role="underline">,</emphasis></emphasis> Bits Stored<emphasis role="bold"><emphasis role="underline">, and High Bit</emphasis></emphasis> Attribute<emphasis role="bold"><emphasis role="strikethrough"> value</emphasis></emphasis>s, single bit pixels shall be packed 8 to a byte as defined by the encoding rules in <olink targetdoc="PS3.5" targetptr="PS3.5" xrefstyle="select: labelnumber"/>.</para>
                    </section>
                    <section label="C.8.20.2.2" status="5" xml:id="sect_C.8.20.2.2">
                        <title>Lossy Image Compression and Lossy Image Compression Method</title>
                        <para xml:id="para_4904ac59-5981-4778-847a-926095bf8938">If Lossy Image Compression (0028,2110) in any of the source images is "01", the value shall be "01" for the Segmentation Instance.</para>
                        <para xml:id="para_da361aea-ca87-4f7b-af69-0ff8ed864c33">The process of segmentation itself is defined not to be lossy compression, even though it involves loss. If the Segmentation Instance is encoded using a lossy compression Transfer Syntax, then the value shall be set to "01".</para>
                        <note>
                            <para xml:id="para_2325bba0-5f6b-4fb8-9133-dd37a2897152">It is not advisable to lossy compress a Segmentation <emphasis role="bold"><emphasis role="strikethrough">SOP </emphasis></emphasis>Instance. In particular, <emphasis role="bold"><emphasis role="strikethrough">a binary</emphasis></emphasis>
                            <emphasis role="bold"><emphasis role="underline">BINARY or LABELMAP S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egmentation <emphasis role="bold"><emphasis role="underline">Instances </emphasis></emphasis>should not be lossy compressed.</para>
                        </note>
                    </section>
                    <section label="C.8.20.2.3" status="5" xml:id="sect_C.8.20.2.3">
                    <title>Segmentation Type, Segmentation Fractional Type and Segments Overlap</title>
                        <section label="C.8.20.2.3.1" status="6" xml:id="sect_C.8.20.2.3.1">
                            <title><emphasis role="underline">BINARY Segmentation Type</emphasis></title>
                    <para xml:id="para_e82002ca-f15f-462a-9f3d-b275d3cb23ca">A Segmentation Type (0062,0001) of BINARY indicates <emphasis role="bold"><emphasis role="underline">that </emphasis></emphasis>the segmented property is present with a value of 1 and absent with a value of 0.</para>
                    </section>
                        <section label="C.8.20.2.3.2" status="6" xml:id="sect_C.8.20.2.3.2">
                            <title><emphasis role="underline">FRACTIONAL Segmentation Type</emphasis></title>
                            
                        <para xml:id="para_771823c9-ccd3-42e5-ba90-55967e3cb95d">For a Segmentation Type (0062,0001) of FRACTIONAL the segmented property is defined as a value from zero to the Maximum Fractional Value (0062,000E). A FRACTIONAL segmentation shall be further specified via Segmentation Fractional Type (0062,0010).</para>
                    <variablelist spacing="compact">
                        <title>Enumerated Values of Segmentation Fractional Type (0062,0010):</title>
                        <varlistentry>
                            <term>PROBABILITY</term>
                            <listitem>
                                <para xml:id="para_304a5ea6-fa5d-421d-83d0-912952b62342">Defines the probability, as a ratio of the pixel value to the Maximum Fractional Value, that the segmented property occupies the spatial area defined by the voxel.</para>
                            </listitem>
                        </varlistentry>
                        <varlistentry>
                            <term>OCCUPANCY</term>
                            <listitem>
                                <para xml:id="para_c35fbd60-4255-424b-b357-4e963e2dee7e">Defines the proportion of the pixel volume occupied by the segmented property as the ratio of the pixel value to the Maximum Fractional Value.</para>
                            </listitem>
                        </varlistentry>
                    </variablelist>
                        </section>
                        <section label="C.8.20.2.3.3" status="6" xml:id="sect_C.8.20.2.3.3">
                            <title><emphasis role="underline">LABELMAP Segmentation Type</emphasis></title>
                        <para><emphasis role="bold"><emphasis role="underline">A Segmentation Type (0062,0001) of LABELMAP indicates that the presence of the segmented property is conveyed by the integer value of the Pixel Data (7FE0,0010), which is one of the values of Segment Number (0062,0004).</emphasis></emphasis></para>
                    <note>
                        <para xml:id="para_4913e2f0-493b-4d64-8c5e-2cc91535ef30"><emphasis role="bold"><emphasis role="strikethrough">Binary</emphasis><emphasis role="underline">BINARY</emphasis></emphasis> Segmentation Instances may be transformed into or from <emphasis role="bold"><emphasis role="strikethrough">"</emphasis></emphasis>label maps<emphasis role="bold"><emphasis role="strikethrough">"</emphasis></emphasis>,<emphasis role="bold"><emphasis role="strikethrough"> in which each pixel contains a coded value that indicates the segment, rather than the Segmentation IOD representation that encodes separate bit planes for each segment</emphasis></emphasis>. This conversion is facilitated by knowing whether or not any <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment bi<emphasis role="bold"><emphasis role="underline">t</emphasis></emphasis>planes overlap. <emphasis role="bold"><emphasis role="underline">By
                            definition, each pixel value in a single LABELMAP Segmentation Instance can only represent one Segment. and cannot overlap.</emphasis></emphasis> A Segments Overlap (0062,0013) value of NO indicates <emphasis role="bold"><emphasis role="underline">that </emphasis></emphasis>they can be converted into (or may have been converted from) a label map representation without the need to check every pixel.</para>
                    </note>
                        <para><emphasis role="bold"><emphasis role="underline">For a Segmentation Type (0062,0001) of LABELMAP, every pixel value actually encoded in Pixel Data (7FE0,0010) is required to be described in an Item of Segment Sequence (0062,0002).</emphasis></emphasis></para>
                        <note>
                            <orderedlist>
                                <listitem>
                                    <para><emphasis role="bold"><emphasis role="underline">The converse is not true, in that the Segment Sequence (0062,0002) can describe Segments that are not actually present in the pixel data, e.g., to allow for re-use of a common <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment description across multiple instances, despite the inefficiency of encoding unused information.</emphasis></emphasis></para>
                                </listitem>
                                <listitem>
                                    <para><emphasis role="bold"><emphasis role="underline">The need to describe every pixel value implies that the value 0, if used, will also be described, which might serve as the background.</emphasis></emphasis></para>
                                </listitem>
                            </orderedlist>
                        </note>
                        </section>
                </section>
                    <section label="C.8.20.2.4" status="5" xml:id="sect_C.8.20.2.4">
                        <title>Segment Number</title>
                        <para xml:id="para_54bb8e6b-7101-4da9-ba2a-d9ef0d85d039">Segment Number (0062,0004) shall be unique within each Instance<emphasis role="bold"><emphasis role="underline">.</emphasis></emphasis></para>
                    <para><emphasis role="bold"><emphasis role="underline">If Segmentation Type (0062,0001) is BINARY or FRACTIONAL</emphasis></emphasis>, <emphasis role="bold"><emphasis role="underline">Segment Number (0062,0004)  shall </emphasis></emphasis>start at a value of 1, and increase monotonically by 1.</para>
                    <note>
                        <orderedlist>
                            <listitem>
                                <para><emphasis role="bold"><emphasis role="underline">When converting from LABELMAP Segmentation Instances to BINARY Segmentation Instances, care should be taken to reassign the Segment Number (0062,0004) values if necessary, such that the requirements in this section are complied with.</emphasis></emphasis></para>
                            </listitem>
                            <listitem>
                                <para><emphasis role="bold"><emphasis role="underline">Requirements on the ascending order of the Segment Number (0062,0004) values do not imply an ordering of the Sequence Items of any enclosing Sequence.</emphasis></emphasis></para>
                            </listitem>
                            <listitem>
                                <para><emphasis role="bold"><emphasis role="underline">For a Segmentation Type (0062,0001) of LABELMAP, if a value is specified in Pixel Padding Value (0028,0120), that indicates that the corresponding Segment Number is treated as background; see <xref linkend="sect_A.51.4" xrefstyle="select: label"/> and <xref linkend="sect_C.7.5.1.1.2" xrefstyle="select: label"/>. Otherwise, no particular value of Segment Number is deemed to be treated specially as background.</emphasis></emphasis></para>
                            </listitem>
                        </orderedlist>
                    </note>
                    </section>
                </section>
                <section label="C.8.20.3" status="4" xml:id="sect_C.8.20.3">
                    <title>Segmentation Functional Group Macros</title>
                    <para xml:id="para_376c9129-7639-41d0-948b-db2cbc078149">The following sections contain Functional Group Macros specific to the Segmentation IOD.</para>
                    <note>
                        <para xml:id="para_7d603a8c-dfe7-4ebf-8721-f8730ada4a76">The Attribute descriptions in the Functional Group Macros are written as if they were applicable to a single frame (i.e., the Macro is part of the Per-Frame Functional Groups Sequence). If an Attribute is applicable to all frames (i.e., the Macro is part of the Shared Functional Groups Sequence) the phrase "this frame" in the Attribute description shall be interpreted to mean "for all frames".</para>
                    </note>
                    <section label="C.8.20.3.1" status="5" xml:id="sect_C.8.20.3.1">
                        <title>Segmentation Macro</title>
                        <para xml:id="para_ab568e07-baf8-4574-ba4e-9e4e5cee6bae">
                            <xref linkend="table_C.8.20-3" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.8.20.3.1" xrefstyle="select: title"/>.</para>
                        <table frame="box" label="C.8.20-3" rules="all" xml:id="table_C.8.20-3">
                            <caption>Segmentation Macro Attributes</caption>
                            <thead>
                                <tr valign="top">
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_b2d3604e-0289-42db-9b74-5efaf44bcd6b">Attribute Name</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_e59b020d-edbb-43d6-a511-6d53cd7a188b">Tag</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_059d6df1-014a-4e9a-82a2-6970f361ffb0">Type</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_86935f4c-2c50-4a25-9c51-0079c32cabb5">Attribute Description</para>
                                    </th>
                                </tr>
                            </thead>
                            <tbody>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_514f8fb4-dfe1-4ba3-8178-88cd785d732c">Segment Identification Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_fcc5fd89-4817-4c0d-837d-6b03fe93a4fa">(0062,000A)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_968bc470-2fc6-4c74-be8b-d4693b70a7de">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_e8521205-c87d-40de-8259-221d401f450f">Identifies the characteristics of this frame.</para>
                                        <para xml:id="para_206d5957-d6e6-4a8a-808a-04df34afb3b2">Only a single Item shall be included in this Sequence.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_4c82df38-56ab-4f53-9d01-715fb364febf">&gt;Referenced Segment Number</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_581ba7bd-d847-4e79-987d-60a7241c3227">(0062,000B)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_f226b401-4e25-4d9e-a767-d0e0f5751c49">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_ddb9b36c-b07a-4714-b603-c2d88ff3dde9">Uniquely identifies the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment described in Segment Sequence (0062,0002) by reference to Segment Number (0062,0004). Referenced Segment Number (0062,000B) shall not be multi-valued.</para>
                                    </td>
                                </tr>
                            </tbody>
                        </table>
                    </section>
                </section>
                <section label="C.8.20.4" status="4" xml:id="sect_C.8.20.4">
                    <title>Segmentation Macros</title>
                    <para xml:id="para_d39448f7-fe6b-427a-ac75-7fb75b810827">The following sections contain Macros specific to the Segmentation IOD.</para>
                    <section label="C.8.20.4.1" status="5" xml:id="sect_C.8.20.4.1">
                        <title>Segment Description Macro</title>
                        <para xml:id="para_7b517577-d1cc-43a6-bade-d98a4f5c85d7">
                            <xref linkend="table_C.8.20-4" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.8.20.4.1" xrefstyle="select: title"/>.</para>
                        <table frame="box" label="C.8.20-4" rules="all" xml:id="table_C.8.20-4">
                            <caption>Segment Description Macro Attributes</caption>
                            <thead>
                                <tr valign="top">
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_e5873f75-48d1-402e-a8cc-0fec00eb11ed">Attribute Name</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_9ce3b607-5dfa-467c-ae0c-f713cb13ccf7">Tag</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_047952eb-8c41-4ce9-a835-4b42125903a0">Type</para>
                                    </th>
                                    <th align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_10477ae3-4bc4-416f-a8ad-8a433354940c">Attribute Description</para>
                                    </th>
                                </tr>
                            </thead>
                            <tbody>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_89b74565-96db-49e7-80ed-947f98554f8e">Segment Number</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_679d6547-0844-44c1-b2e2-29e9e53b79d3">(0062,0004)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_2f3320ca-b331-4e3c-9ca1-aff2d0a4c320">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_a2dc2e33-66cb-4732-8ca0-640023a2941c">Identification number of the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment. The value of Segment Number (0062,0004) shall be unique within the Segmentation Instance in which it is created. See <xref linkend="sect_C.8.20.2.4" xrefstyle="select: label"/>.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_7f4a9b09-0817-4b35-a8be-08a3e398a911">Segment Label</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_e832b0f2-7a7b-4ee0-9c02-d08ea312e98a">(0062,0005)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_dbbbd023-9265-4b62-93b1-500fbc88bd52">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_31c5221b-1ec8-407e-8543-c3a728cce054">User-defined label identifying this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment. This may be the same as Code Meaning (0008,0104) of Segmented Property Type Code Sequence (0062,000F).</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_8df80b9f-e0b5-4748-af9f-81f86f0a6df0">Segment Description</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_7eb5a6ed-a466-4462-874c-3ca981ce7f59">(0062,0006)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_e296be92-eb14-40f3-9eab-676d2594d849">3</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_49345a3f-6586-4727-baed-0b6743c895ce">User-defined description for this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_59326d43-8df5-4bad-8353-2e56652fe6b3">Segment Algorithm Type</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_3260b7d0-fa20-4cdd-846f-7a0e26634e15">(0062,0008)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_d3722e9d-3368-4bbf-b100-e44055834b8a">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_ced9b6e7-728a-426b-bc51-f0a2fe2f5a90">Type of algorithm used to generate the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment.</para>
                                        <variablelist spacing="compact">
                                            <title>Enumerated Values:</title>
                                            <varlistentry>
                                                <term>AUTOMATIC</term>
                                                <listitem>
                                                    <para xml:id="para_95e3bf8a-e3e4-48cf-a6dc-93e5cf0b2854">calculated <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment</para>
                                                </listitem>
                                            </varlistentry>
                                            <varlistentry>
                                                <term>SEMIAUTOMATIC</term>
                                                <listitem>
                                                    <para xml:id="para_c676991e-3cd3-478c-b75a-f7b25bc73285">calculated <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment with user assistance</para>
                                                </listitem>
                                            </varlistentry>
                                            <varlistentry>
                                                <term>MANUAL</term>
                                                <listitem>
                                                    <para xml:id="para_0529be8a-6c63-4684-a745-dbc8e5cb5383">user-entered <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment</para>
                                                </listitem>
                                            </varlistentry>
                                        </variablelist>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="3" rowspan="1">
                                        <para xml:id="para_5405c6ad-ec1a-49a7-b29f-4b5f46b44a95">
                                            <emphasis role="italic">Include Table 10-7b "Multiple Site General Anatomy Optional Macro Attributes"<!--<xref linkend="table_10-7b" xrefstyle="select: label quotedtitle"/>-->
                                            </emphasis>
                                        </para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_01915cc9-a15b-41e5-a999-1c7648a155ba">
                                            <emphasis role="italic">May not be necessary if the anatomy is implicit in the Segmented Property Type Code Sequence.</emphasis>
                                        </para>
                                        <para xml:id="para_94bb03d6-d3bf-4659-9282-a60e3f949904">
                                            <emphasis role="italic">More than one Item in Anatomic Region Sequence (0008,2218) may be used when a region of interest spans multiple anatomical locations and there is not a single pre-coordinated code describing the combination of locations. There is no requirement that the multiple locations be contiguous.</emphasis>
                                        </para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_f5f0349b-ab2a-436d-8ded-b0bc82a19e11">Segmented Property Category Code Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_5e3c280a-0a6b-4635-9143-679d88f966bf">(0062,0003)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_4b0500cc-fac5-42a6-b1d1-2e51c2ac94bb">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_ae19ae43-e724-4506-89f2-4ea677c3026b">Sequence defining the general category of the property the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment represents.</para>
                                        <para xml:id="para_d11f4704-6558-4bf3-8bee-2d2aa60a7247">Only a single Item shall be included in this Sequence.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="3" rowspan="1">
                                        <para xml:id="para_2a90e84e-d251-4457-8664-eb73c4e2712f">
                                            <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                            </emphasis>
                                        </para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_6b2c86d9-b05f-485b-8f22-ce22d8cde766">
                                            <emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_7150" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                                        </para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_2c17461c-ef24-4b80-9680-cdcf54031c11">Segmented Property Type Code Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_0d467a88-b426-4e98-89d7-db0109927e06">(0062,000F)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_39736199-ccab-48c1-a7c4-0af7d87420ab">1</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_5e98e7fa-102f-4676-b423-6e5a88aadc59">Sequence defining the specific property the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment represents.</para>
                                        <note>
                                            <para xml:id="para_ec8e6f7e-0860-4d73-bbf3-1fda63967705">"Property" is used in the sense of meaning "what the segmented voxels represent", whether it be a physical or biological object, be real or conceptual, having spatial, temporal or functional extent or not. I.e., it is what the <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment "is" (as opposed to some feature, attribute, quality, or characteristic of it, like color or shape or size).</para>
                                        </note>
                                        <para xml:id="para_0709ad40-b9d1-4432-b54b-eb584c52e2b3">Only a single Item shall be included in this Sequence.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="3" rowspan="1">
                                        <para xml:id="para_af07999d-7c2b-4b0c-9013-f4c2eb71bc4c">
                                            <emphasis role="italic">&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                            </emphasis>
                                        </para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_4c9a4caa-f726-4fc7-a4f3-ee4d30cce998">
                                            <emphasis role="italic">B<olink targetdoc="PS3.16" targetptr="sect_CID_7151" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                                        </para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_45dbd699-cfb1-402d-95c7-8ace55dbddef">&gt;Segmented Property Type Modifier Code Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_dc0b1bfc-2562-4615-a294-f388862cdfea">(0062,0011)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_8303257d-425a-4786-b950-714d3a4226ce">3</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_46c1197b-66e2-4c3f-ab60-2183b739e963">Sequence defining the modifier of the property type of this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment.</para>
                                        <para xml:id="para_6ff8ff0c-b63d-4cd8-87cf-009043959541">One or more Items are permitted in this Sequence.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="3" rowspan="1">
                                        <para xml:id="para_e44a4c92-6204-43fa-99b2-70cd8bb7f1f3">
                                            <emphasis role="italic">&gt;&gt;Include Table 8.8-1 "Code Sequence Macro Attributes"<!--<xref linkend="table_8.8-1" xrefstyle="select: label quotedtitle"/>-->
                                            </emphasis>
                                        </para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_8e288acf-cbfb-48ab-8644-e04b283b25f4">
                                            <emphasis role="italic">D<olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>.</emphasis>
                                        </para>
                                        <note>
                                            <para xml:id="para_8fa0b2fd-bdd0-4c8b-aaf2-b3de2f6844d4">
<emphasis role="italic">For Retinal Segmentation Surfaces, laterality is not typically specified.</emphasis>
</para>
                                        </note>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_332c16f0-7d34-4477-a4a4-e7813f47a700">Tracking ID</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_d9be05fc-e8b3-4ac2-b6d9-7b932b230089">(0062,0020)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_090dabe3-4629-4c17-8997-454a05fddd64">1C</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_34bba996-30f3-4288-afe8-cd2e5b679fd7">A text label used for tracking a finding or feature, potentially across multiple reporting objects, over time. This label shall be unique within the domain in which it is used.</para>
                                        <para xml:id="para_95cc4898-30f5-4eb7-a982-3c201a07f0bb">Required if Tracking UID (0062,0021) is present.</para>
                                        <note>
                                            <orderedlist>
                                                <listitem>
                                                    <para xml:id="para_bc9377e8-79b7-4985-ab7b-d5675e1cf2e0">May or may not have the same value as Segment Label (0062,0005).</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_6446e408-6f4d-494d-968e-a243d94feb6a">Related SR Instances may exist, for example, to record measurements related to this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment, but need not exist for this Attribute to be used.</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_6eae24e2-bfba-4094-beec-e75fd44ac5a1">This Attribute will have the same value as the value of the <olink targetdoc="PS3.16" targetptr="DCM_112039" xrefstyle="select: nopage">(112039, DCM, "Tracking Identifier")</olink> Content Item in SR Instances that reference this Segment in this Segmentation Instance.</para>
                                                </listitem>
                                            </orderedlist>
                                        </note>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_b2e24341-d2d7-4aae-8916-aae70653fe4c">Tracking UID</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_812ac62a-0bd8-4328-a13d-9a31cad6ed92">(0062,0021)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_2bd967b5-7525-42ae-ab2f-7c49741c3ff2">1C</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_5022eae5-11f9-4ec4-a3dc-876da69d0426">A unique identifier used for tracking a finding or feature, potentially across multiple reporting objects, over time.</para>
                                        <para xml:id="para_0c8ffb12-539c-458a-84d3-83141c247650">Required if Tracking ID (0062,0020) is present.</para>
                                        <note>
                                            <orderedlist>
                                                <listitem>
                                                    <para xml:id="para_8d194842-7201-4c47-a558-0ff46d6a0475">Related SR Instances may exist, for example, to record measurements related to this <emphasis role="bold"><emphasis role="underline">S</emphasis><emphasis role="strikethrough">s</emphasis></emphasis>egment, but need not exist for this Attribute to be used.</para>
                                                </listitem>
                                                <listitem>
                                                    <para xml:id="para_1674b0e1-fc50-47ad-83f5-adb781633aa4">This Attribute will have the same value as the value of the <olink targetdoc="PS3.16" targetptr="DCM_112040" xrefstyle="select: nopage">(112040, DCM, "Tracking Unique Identifier")</olink> Content Item in SR Instances that reference this Segment in this Segmentation Instance.</para>
                                                </listitem>
                                            </orderedlist>
                                        </note>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_e26ba899-0c88-4a8c-8654-3ca5186901e2">Definition Source Sequence</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_ec80049a-b160-4017-b906-26be9942791e">(0008,1156)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_1815b14f-5b1a-47df-ad45-159840fabe92">3</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_1a82db4f-9410-4a5b-b917-48b3c63cc21e">Instances containing the source of the Segment information.</para>
                                        <para xml:id="para_cc8a52bc-6a5c-4fbc-a46d-74ba82d0d261">Only a single Item is permitted in this Sequence.</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="4" rowspan="1">
                                        <para xml:id="para_f9a7344d-ec95-409b-9d4e-68e4a88786e8">
                                            <emphasis role="italic">&gt;Include Table 10-11 "SOP Instance Reference Macro Attributes"<!--<xref linkend="table_10-11" xrefstyle="select: label quotedtitle"/>-->.</emphasis>
                                        </para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_b9c57314-3d72-49eb-9dde-ec066756f565">&gt;Referenced ROI Number</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_71617b51-bb89-4fe1-9cf3-5f668c5a9ce7">(3006,0084)</para>
                                    </td>
                                    <td align="center" colspan="1" rowspan="1">
                                        <para xml:id="para_ba1faeac-068a-47b9-9dee-602daa7e9a0f">1C</para>
                                    </td>
                                    <td align="left" colspan="1" rowspan="1">
                                        <para xml:id="para_e0950fff-7a99-4b92-9c69-38977867a294">The value of ROI Number (3006,0022) in the referenced SOP Instance that identifies the ROI that is the origin of the Segment information.</para>
                                        <para xml:id="para_fee19d43-d264-4ac3-865a-aca9578f31bd">Required if Referenced SOP Class UID (0008,1150) is RT Structure Set Storage ("1.2.840.10008.5.1.4.1.1.481.3").</para>
                                    </td>
                                </tr>
                                <tr valign="top">
                                    <td align="left" colspan="4" rowspan="1">
                                        <para xml:id="para_ae67cc6a-ea80-40da-bd0e-d6b808fbe9fb">
                                            <emphasis role="italic">Include Table 10.9.3-1 "Content Creator Macro"<!--<xref linkend="table_10.9.3-1" xrefstyle="select: label quotedtitle"/>-->
                                            </emphasis>
                                        </para>
                                    </td>
                                </tr>
                            </tbody>
                        </table>
                    </section>
                </section>
            </section>
        
        <section label="C.7.5.1.1" status="5" xml:id="sect_C.7.5.1.1">
            <title>General Equipment Module Attribute Descriptions</title>
            <para>...</para>
            <section label="C.7.5.1.1.2" status="6" xml:id="sect_C.7.5.1.1.2">
                <title>Pixel Padding Value and Pixel Padding Range Limit</title>
                <para xml:id="para_1a462888-08ea-4690-921f-eeb6fd330c0b">Pixel Padding Value (0028,0120) is typically used to pad grayscale images (those with a Photometric Interpretation (0028,0004) of MONOCHROME1 or MONOCHROME2), or color images with a Photometric interpretation (0028,0004) of PALETTE COLOR, to rectangular format. The native format of some images is not rectangular. It is common for devices with this format to pad the images, to the rectangular format required by the DICOM Standard, with a specific pixel value that is not contained in the native image. Further, when resampling, such as after spatial registration, padding may need to be used to fill previously non-existent pixels.</para>
                <para xml:id="para_a2fb18d0-b54e-4762-89b4-b13b98fc8a06">Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) are also used to identify pixels to be excluded from the rendering pipeline for other reasons, such as suppression of background air. Pixel Padding Range Limit (0028,0121) is defined in the <xref linkend="sect_C.7.6.3" xrefstyle="select: title"/>.</para>
                <note>
                    <orderedlist numeration="arabic">
                        <listitem>
                            <para xml:id="para_2af6195a-81b9-4690-8763-9fc7e8d241f1">The "native image" is that which is being padded to the required rectangular format, e.g., the area within the circular reconstruction perimeter of a CT image, or the subset of the rectangular area that contains useful image information, i.e., which is not to be suppressed, or e.g., is that part of a pseudo-colored image that might be superimposed on top of another image. For other mechanisms, see <olink targetdoc="PS3.4" targetptr="sect_N.2.6" xrefstyle="template:Section %n “%t” in PS3.4"/>.</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_3c092e2e-9153-4645-8942-b86e1f2aa2eb">The Pixel Padding Value is explicitly described in order to prevent display applications from taking it into account when determining the dynamic range of an image, since the Pixel Padding Value will be outside the range between the minimum and maximum values of the pixels in the native image</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_1965bec6-3291-41fb-904e-0c894e5e85c7">No pixels in the native image will have a value equal to Pixel Padding Value (0028,0120).</para>
                        </listitem>
                    </orderedlist>
                </note>
                <para xml:id="para_5baf3d02-dfdc-4efd-8bd7-93e63cf03799">Pixel Padding Value (0028,0120) specifies either a single value of this padding, or when combined with Pixel Padding Range Limit (0028,0121), a range of values (inclusive) that are padding.</para>
                <para xml:id="para_8ae6007e-1de5-4eb1-b921-49c98482c270">The values of Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) shall be valid values within the constraints defined by Bits Allocated (0028,0100), Bits Stored (0028,0101), and High Bit (0028,0102).</para>
                <para xml:id="para_9bf05f8b-2760-47ed-b71d-20ba14dbf72d">The Pixel Padding Value shall correspond to a value in the original stored pixel data, before the Modality LUT Transformation or any other transformations are applied.</para>
                <para xml:id="para_41537ed0-4e98-459a-89b6-d4d01672a6b3">Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) shall not be present when padding is performed but the pixel value used for padding does occur in the native image.</para>
                <para xml:id="para_84c202b8-f76d-407a-abb1-72f29fb068b6">If Photometric Interpretation (0028,0004) is MONOCHROME2 or PALETTE COLOR, Pixel Padding Value (0028,0120) shall be less than (closer to or equal to the minimum possible pixel value) or equal to Pixel Padding Range Limit (0028,0121). If Photometric Interpretation (0028,0004) is MONOCHROME1, Pixel Padding Value (0028,0120) shall be greater than (closer to or equal to the maximum possible pixel value) or equal to Pixel Padding Range Limit (0028,0121).</para>
                <note>
                    <orderedlist numeration="arabic">
                        <listitem>
                            <para xml:id="para_b505df6d-712b-4b92-bd2e-803884cb8145">When the relationship between pixel value and X-Ray Intensity is unknown, it is recommended that the following values be used to pad with black when the image is unsigned:</para>
                            <itemizedlist>
                                <listitem>
                                    <para xml:id="para_d6462a1c-34b6-4220-85a7-ffeed2a13e38">0 if Photometric Interpretation (0028,0004) is MONOCHROME2.</para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_d4ae7e06-8f1e-4ffc-b38e-eba1a6ce48af">2<superscript>Bits Stored</superscript> - 1 if Photometric Interpretation (0028,0004) is MONOCHROME1.</para>
                                </listitem>
                            </itemizedlist>
                            <para xml:id="para_6696b5d6-0ef9-4c95-90d3-02b949087074">and when the image is signed:</para>
                            <itemizedlist>
                                <listitem>
                                    <para xml:id="para_8e6351ad-e604-4e7d-a900-fccd60bdfdaa">-2<superscript>Bits Stored-1</superscript> if Photometric Interpretation (0028,0004) is MONOCHROME2.</para>
                                </listitem>
                                <listitem>
                                    <para xml:id="para_1134b5df-86ac-4ca8-b241-ff9f6b068ab5">2<superscript>Bits Stored-1</superscript> - 1 if Photometric Interpretation (0028,0004) is MONOCHROME1.</para>
                                </listitem>
                            </itemizedlist>
                        </listitem>
                        <listitem>
                            <para xml:id="para_f9a9b734-171b-4a07-ad6b-e4ef12186cca">For projection radiography, when the relationship between pixel value and X-Ray Intensity is known (for example as defined by Pixel Intensity Relationship (0028,1040) and Pixel Intensity Relationship Sign (0028,1041)), it is recommended that a pixel value equivalent to, or rendered similarly to, air (least X-Ray absorbance) be used for padding. However, if such a value may occur in the native image, the Pixel Padding Value (0028,0120) Attribute itself should not be present.</para>
                            <para xml:id="para_26da1bb6-c833-4428-8b5c-047222cdbf84">E.g., for an XRF image obtained with an image intensifier, if air is black then a padded perimeter, if any, should also appear black. Typically though, if unpadded, this area would be collimated with a circular collimator, in which case the pixels would appear natively as white (greatest X-Ray absorbance) and a circular shutter would be necessary to neutralize them as black. Whether collimated areas are detected and treated as padded, or neutralized with shutters is at the discretion of the application. See also the <xref linkend="sect_C.7.6.11" xrefstyle="select: title"/>
                                <xref linkend="sect_C.7.6.11" xrefstyle="select: label"/>.</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_a665eeee-c2cc-4187-aa63-89378bbe30bf">The conditional requirement for the Pixel Padding Range Limit (0028,0121) in the <xref linkend="sect_C.7.6.3" xrefstyle="select: title"/> means that it shall not be present unless Pixel Padding Value (0028,0120) is also present.</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_7e066e9b-7b0d-474f-87a4-6944b1b0348a">The range of values to be suppressed between Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) is specified as being inclusive, that is the values themselves as well as all values between are suppressed.</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_ec3371ea-e38d-416c-b5f8-b4f972f046c9">When Pixel Padding Range Limit (0028,0121) is present, but not supported by a rendering application, the constraint that Pixel Padding Value (0028,0120) is closest to the "blackest" value, which is typically the most frequently occurring background pixel, will most often result in an acceptable display, permitting "backward compatibility" in the majority of cases.</para>
                        </listitem>
                        <listitem>
                            <para><emphasis role="bold"><emphasis role="underline">For Segmentations with a Segmentation Type (0062,0001) of LABELMAP, Pixel Padding Value (0028,0120) may be used to indicate that a particular pixel value is to be treated as background. Only a single pixel value is so indicated, and Pixel Padding Range Limit (0028,0121) is not used; see <xref linkend="sect_A.51.4" xrefstyle="select: label"/>. Typically the Pixel Padding Value (0028,0120) will be 0, but it can be another value, or may not be specified at all. </emphasis></emphasis></para>
                        </listitem>
                    </orderedlist>
                </note>
                <para xml:id="para_9d69206c-b0b9-4726-b21d-46f13b185f97">When modifying equipment changes the Pixel Padding Value in the image, it shall change the values of Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121), if present. If modifying equipment changes the Pixel Padding Values in the image to values present in the native image, the Attribute Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) shall be removed.</para>
                <note>
                    <orderedlist numeration="arabic">
                        <listitem>
                            <para xml:id="para_c6f434b3-eada-4bfd-9498-69d296d72c35">For example, if a CT image containing signed values from -1024 to 3191 and a Pixel Padding Value (0028,0120) of -2000 and a Rescale Intercept (0028,1052) of 0 is converted to an unsigned image with a Rescale Intercept (0028,1052) of -1024 by adding 1024 to all pixels and clipping all more negative pixels to 0, then the padding pixels will be indistinguishable from some of the modified native image pixels, and hence Pixel Padding Value (0028,0120) needs to be removed.</para>
                        </listitem>
                        <listitem>
                            <para xml:id="para_72bc0aca-a39d-4f15-9925-7caf4048709f">If the modification involves lossy compression, which may result in changes to the pixel values, then the application of Pixel Padding Value (0028,0120) and Pixel Padding Range Limit (0028,0121) may result in a different appearance, and hence these Attributes may need different values also.</para>
                        </listitem>
                    </orderedlist>
                </note>
            </section>
        </section>
        <section label="C.7.9" status="3" xml:id="sect_C.7.9">
                <title>Palette Color Lookup Table Module</title>
                <para xml:id="para_55388b37-bd58-4a13-b7cc-2ec5eb5a019f">
                    <xref linkend="table_C.7-22" xrefstyle="select: label"/> specifies the Attributes of the <xref linkend="sect_C.7.9" xrefstyle="select: title"/>, which describe the Lookup table data for images with Palette Color photometric interpretation.</para>
            <para xml:id="para_621cff74-c1f5-4b4d-8743-ce5fa9e4f9d9">When the <xref linkend="sect_C.7.9" xrefstyle="select: title"/> is present in an Image IOD, the conditional requirements for the use of Palette Color Lookup Table Data (0028,1201-1203) and Segmented Palette Color Lookup Table Data (0028,1221-1223), described in <xref linkend="table_C.7-22" xrefstyle="select: label"/>, shall take precedence over the conditional requirements described in the Image Pixel Module<!--<xref linkend="sect_C.7.6.3" xrefstyle="select: title"/>--> (see Section C.7.6.3<!--<xref linkend="sect_C.7.6.3" xrefstyle="select: label"/>-->). When the <xref linkend="sect_C.7.9" xrefstyle="select: title"/> is present in a Presentation State IOD, the Palette Color Lookup Table Data (0028,1201-1203) Attributes are mandatory and the Segmented Palette Color Lookup Table Data (0028,1221-1223) shall not be present. When the <xref linkend="sect_C.7.9" xrefstyle="select: title"/> is present in a Color Palette IOD, either the Palette Color Lookup Table Data
                    (0028,1201-1203) or Segmented Palette Color Lookup Table Data (0028,1221-1223) Attributes may be used.</para>
                <para xml:id="para_78a23ebd-abd9-4eb4-9a76-9ef782670edc">When the <xref linkend="sect_C.7.9" xrefstyle="select: title"/> is present in a Color Palette IOD, the 3rd value of Palette Color Lookup Table Descriptor (0028,1101-1103) (i.e, the number of bits for each entry in the Lookup Table Data) shall be 8.</para>
                <table frame="box" label="C.7-22" rules="all" xml:id="table_C.7-22">
                    <caption>Palette Color Lookup Table Module Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c93c70de-c53a-45b3-b02d-e147834811ba">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_85748cc7-dce5-4ba5-89f7-5c0bf7c1f0d9">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_84321406-3087-4bb8-a7f6-a2ed62114f92">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_47829838-f1b9-46ed-82b7-a3214296b929">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="4" rowspan="1">
                                <para xml:id="para_3ca51964-b287-4984-87d1-da0455989e49">
                                    <emphasis role="italic">Include <xref linkend="table_C.7-22a" xrefstyle="select: label quotedtitle"/>
                                    </emphasis>
                                </para>
                            </td>
                        </tr>
                    </tbody>
                </table>
                <table frame="box" label="C.7-22a" rules="all" xml:id="table_C.7-22a">
                    <caption>Palette Color Lookup Table Macro Attributes</caption>
                    <thead>
                        <tr valign="top">
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5baedec1-ab08-4504-a897-912bf7a2197b">Attribute Name</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_29dc4762-aca5-4551-bfe2-5481a7470201">Tag</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cc35e274-60c1-43fa-a192-1972a9de183b">Type</para>
                            </th>
                            <th align="center" colspan="1" rowspan="1">
                                <para xml:id="para_75558297-30f6-4002-aed2-05c691ec74d3">Attribute Description</para>
                            </th>
                        </tr>
                    </thead>
                    <tbody>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_e4ab131b-dee8-4b76-bad7-4fd30df8ea79">Red Palette Color Lookup Table Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9a8dde96-a358-4a76-b629-3bc58648571e">(0028,1101)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_c7338a19-26d8-4af9-9528-395955f952c8">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b2799463-c18b-441a-a242-54d450a3e2c5">Specifies the format of the Red Palette Color Lookup Table Data (0028,1201). See Section C.7.6.3.1.5<!--<xref linkend="sect_C.7.6.3.1.5" xrefstyle="select: label"/>--> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_8e4b2a05-89a1-4fb1-a2a8-1ada7231be22">Green Palette Color Lookup Table Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_6589d7bd-52f5-4f6f-9518-376c3bcf66b4">(0028,1102)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_1e84806d-f480-4029-bb56-aa46b7b3f886">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_9513e3dc-1e89-4b67-b5ec-e1cfa6b5b777">Specifies the format of the Green Palette Color Lookup Table Data (0028,1202). See Section C.7.6.3.1.5<!--<xref linkend="sect_C.7.6.3.1.5" xrefstyle="select: label"/>--> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_3a3336f3-3a0d-4183-8425-599e7077e879">Blue Palette Color Lookup Table Descriptor</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_d1e1278f-767e-4e02-bb8d-2192c67d884d">(0028,1103)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_37ccb353-4b27-457b-b5a5-2bfe318e192e">1</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_c39c2f9e-5b9d-4429-a52f-a9a93ca94e40">Specifies the format of the Blue Palette Color Lookup table Data (0028,1203). See Section C.7.6.3.1.5<!--<xref linkend="sect_C.7.6.3.1.5" xrefstyle="select: label"/>--> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_495a143e-0e9d-48f7-806e-ee1e1d0eaf37">Palette Color Lookup Table UID</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_04e273c0-66a7-47e7-adcc-24f27c3b0230">(0028,1199)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_82f389a7-a4d0-437c-b4a3-0ef34ebf3393">3</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_58c7e439-c35f-4821-829b-516f02c4f491">Palette Color Lookup Table UID. See <xref linkend="sect_C.7.9.1" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_30efbc10-2ba4-4f2b-b184-6a5ee92c72d4">Red Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_181d145e-3bea-48ec-ab36-3e68424fed53">(0028,1201)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9064b619-dba1-4bcf-adca-c388f857c48c">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5b4ff6e4-f797-4161-ad59-6b30e8c2ccd4">Red Palette Color Lookup Table Data. Required if segmented data is NOT used in an Image IOD or Color Palette IOD, or if the IOD is a Presentation State IOD<emphasis role="bold"><emphasis role="underline"> or Segmentation IOD</emphasis></emphasis>. See Section C.7.6.3.1.6<!--<xref linkend="sect_C.7.6.3.1.6" xrefstyle="select: label"/>--> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_6ad29a05-6f88-451d-8614-e7d1d577281b">Green Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_73d637ce-8566-4806-8696-99d6e150f533">(0028,1202)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_3936d05f-19cf-4d13-9006-22f302aa5325">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_103d12c2-b21c-4f1f-a0da-fa4aa39f47eb">Green Palette Color Lookup Table Data. Required if segmented data is NOT used in an Image IOD or Color Palette IOD, or if the IOD is a Presentation State IOD<emphasis role="bold"><emphasis role="underline"> or Segmentation IOD</emphasis></emphasis>. See Section C.7.6.3.1.6<!--<xref linkend="sect_C.7.6.3.1.6" xrefstyle="select: label"/>--> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_2ee679bb-62c5-4ee7-b42b-2444c7f978ab">Blue Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_26082704-aac2-4621-a284-24dbbdefac0f">(0028,1203)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_b06032da-e11f-4ce0-a12b-93de37ab733a">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7a7b4b13-0be6-45e9-8ab2-76adcd717232">Blue Palette Color Lookup Table Data. Required if segmented data is NOT used in an Image IOD or Color Palette IOD, or if the IOD is a Presentation State IOD<emphasis role="bold"><emphasis role="underline"> or Segmentation IOD</emphasis></emphasis>. See Section C.7.6.3.1.6<!--<xref linkend="sect_C.7.6.3.1.6" xrefstyle="select: label"/>--> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_fb67d1ac-7c13-49ed-a2db-e271f4d1820f">Segmented Red Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_28a23325-e24a-4b58-8b30-a1dcae204797">(0028,1221)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_5a268220-5c17-4eee-92ec-c6301c2fe257">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_4bb50d53-a9b5-40e2-9b5b-b26d3cb5e333">Segmented Red Palette Color Lookup Table Data. Required if segmented data is used in an Image IOD or Color Palette IOD; shall not be present in a Presentation State IOD<emphasis role="bold"><emphasis role="underline"> or Segmentation IOD</emphasis></emphasis>. See <xref linkend="sect_C.7.9.2" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_b128854b-5337-4561-b8bb-21fb4eac429b">Segmented Green Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_9b354717-b47e-4b7d-8726-c7772521a4f7">(0028,1222)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_cd11bc94-f14b-4098-ac75-8de8d0fadcd1">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_7dde8e4c-ef72-4fb9-acda-6f6cdc1d67ca">Segmented Green Palette Color Lookup Table Data. Required if segmented data is used in an Image IOD or Color Palette IOD; shall not be present in a Presentation State IOD<emphasis role="bold"><emphasis role="underline"> or Segmentation IOD</emphasis></emphasis> See <xref linkend="sect_C.7.9.2" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                        <tr valign="top">
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_5177db41-e98c-4f8d-a528-ce05e67ee723">Segmented Blue Palette Color Lookup Table Data</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_435fca5a-41e5-4dde-86c7-3e100c485613">(0028,1223)</para>
                            </td>
                            <td align="center" colspan="1" rowspan="1">
                                <para xml:id="para_669d621e-2fb9-4c2c-95ea-4a69070c13a6">1C</para>
                            </td>
                            <td align="left" colspan="1" rowspan="1">
                                <para xml:id="para_a6dd4dad-719b-482d-a237-093741ad8200">Segmented Blue Palette Color Lookup Table Data. Required if segmented data is used in an Image IOD or Color Palette IOD; shall not be present in a Presentation State IOD<emphasis role="bold"><emphasis role="underline"> or Segmentation IOD</emphasis></emphasis>. See <xref linkend="sect_C.7.9.2" xrefstyle="select: label"/> for further explanation.</para>
                            </td>
                        </tr>
                    </tbody>
                </table>
                <section label="C.7.9.1" status="4" xml:id="sect_C.7.9.1">
                    <title>Palette Color Lookup Table UID</title>
                    <para xml:id="para_a6b90a58-8944-41cd-908e-86a48042e15c">...</para>
                </section>
                <section label="C.7.9.2" status="4" xml:id="sect_C.7.9.2">
                    <title>Segmented Palette Color Lookup Table Data</title>
                    <para xml:id="para_3e10e657-ec8c-4fba-950a-babcd6ccdf50">...</para>
                </section>
            </section>
        
        <section label="C.7.6.3.1.5" status="6" xml:id="sect_C.7.6.3.1.5">
            <title>Palette Color Lookup Table Descriptor</title>
            <para xml:id="para_a614c81f-e874-4f62-a0e2-916d3c153311">...</para>
            <para><emphasis role="bold"><emphasis role="underline">When the Red, Green, or Blue Palette Color Lookup Table Descriptor (0028,1101-1103) are used as part of the <xref linkend="sect_C.7.9" xrefstyle="select: title"/> in the Segmentation IOD, the 3rd value of Palette Color Lookup Table Descriptor (0028,1101-1103) (i.e, the number of bits for each entry in the Lookup Table Data) shall be 8 or 16.</emphasis></emphasis></para>
            <note>
                <orderedlist numeration="arabic">
                    <listitem>
                        <para xml:id="para_a4e67be1-02b6-48b4-89e3-70bb105d7ab1">A value of 16 indicates the Lookup Table Data will range from (0,0,0) minimum intensity to (65535,65535,65535) maximum intensity.</para>
                    </listitem>
                    <listitem>
                        <para xml:id="para_8f3a1178-e335-4395-bed6-64a6f5811da2">Since the Palette Color Lookup Table Descriptor (0028,1101-1104) Attributes are multi-valued, in an Explicit VR Transfer Syntax, only one value representation (US or SS) may be specified, even though the first and third values are always by definition interpreted as unsigned. The explicit VR actually used is dictated by the VR needed to represent the second value, which will be consistent with Pixel Representation (0028,0103).</para>
                    </listitem>
                </orderedlist>
            </note>
        </section>
        <section label="C.7.6.3.1.6" status="6" xml:id="sect_C.7.6.3.1.6">
            <title>Palette Color Lookup Table Data</title>
            <para xml:id="para_ec954e64-50ac-4f38-9fc9-0c4eac04286e">...</para>
        </section>
        
        
    </chapter>
    
    <chapter label="PS3.4" status="1" xml:id="chapter_PS3.4">
        <title>DICOM PS3.4 - Service Class Specifications</title>
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.4 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="B.5" status="2" xml:id="sect_B.5">
            <title>Standard SOP Classes</title>
            <para xml:id="para_31132281-abf4-449c-bba6-9e5508758d41">The SOP Classes in the Storage Service Class identify the Composite IODs to be stored. <xref linkend="table_B.5-1" xrefstyle="select: label"/> identifies Standard SOP Classes.</para>
            <table frame="box" label="B.5-1" rules="all" xml:id="table_B.5-1">
                <caption>Standard SOP Classes</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_d5ac09f7-fad0-454d-ae01-98de6af5e14f">SOP Class Name</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_873bad43-46ef-4569-afca-d49dafa57b07">SOP Class UID</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_faee907c-fdfc-4e5a-ba41-631a1b2e811f">
                                <emphasis role="bold">IOD Specification (defined in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>)</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_4c27cc42-a225-466b-a8c7-250fc4fd9f21">Specialization</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_47951e3b-0aa8-4fa9-991f-f1031bc26d59">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_d6f07d3b-cc3c-4f11-8488-bf63f13e1775"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_e1983b90-21c1-44a2-8641-ee10d5ee98a1"> </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_3d7e1150-da2d-4926-896d-c638fbcf674e">Digital X-Ray Image Storage - For Presentation</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f84eb40a-04a2-4029-a9c3-2532b0dee008">1.2.840.10008.5.1.4.1.1.1.1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_7bf5bd6f-fd05-44fd-8d13-9900de2647e7">
                                <olink targetdoc="PS3.3" targetptr="sect_A.26" xrefstyle="select: title"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_06c885b6-b78c-4ed6-a899-31cf0a305ee0">
<xref linkend="sect_B.5.1.1" xrefstyle="select: labelnumber"/>
</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_4463cd58-7910-4947-b3bd-b0908f8dd0b1">Digital X-Ray Image Storage - For Processing</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_0d23e107-ef88-447a-8097-fb021e71a922">1.2.840.10008.5.1.4.1.1.1.1.1</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_50e7d3a3-346a-42e0-a6ef-5cd4d7bfd34e">
                                <olink targetdoc="PS3.3" targetptr="sect_A.26" xrefstyle="select: title"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_23cfcb8e-1013-443b-9afd-1ba5f2776a9c">
<xref linkend="sect_B.5.1.1" xrefstyle="select: labelnumber"/>
</para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_bbdd93f9-0b55-4211-b4bd-ce9164cef3bc">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_efd49374-8cb4-4fad-9cc4-f163e9098800"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_f3ebdb80-cd30-4488-bf2f-243e2ac2f5f9"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_5acdb83f-ec28-445c-ba5a-aac041c97e51"/>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8b7857c0-9094-49cb-955b-160475fc9bb5">Segmentation Storage</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_18a65967-60d4-48cf-8a00-3836ee4654da">1.2.840.10008.5.1.4.1.1.66.4</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_394aa49c-c0e3-4fbb-a8c4-01be5c38b195">
                                <olink targetdoc="PS3.3" targetptr="sect_A.51" xrefstyle="select: title"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_B.5.1.25" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Label Map Segmentation Storage</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.5.1.4.1.1.66.7</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para>
                                <olink targetdoc="PS3.3" targetptr="sect_A.51" xrefstyle="select: title"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline"><xref linkend="sect_B.5.1.25" xrefstyle="select: labelnumber"/></emphasis></emphasis></para>
                        </td>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_2d55540f-e2b3-46be-a1b1-c4ca3bf753a4">Surface Segmentation Storage</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_6e17d640-f0a1-4e6d-858d-c1357dc1799a">1.2.840.10008.5.1.4.1.1.66.5</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_20721dc2-9a42-401a-80a6-b877599f0d0d">
                                <olink targetdoc="PS3.3" targetptr="sect_A.57" xrefstyle="select: title"/>
                            </para>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_b6716a26-4671-407a-9728-874f38baf19b">...</para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_51931e76-63c8-4888-9ada-1adcd1348b43"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para xml:id="para_8d05dce3-6e60-44d0-9f18-a07be241eaa9"/>
                        </td>
                        <td align="left" colspan="1" rowspan="1"/>
                    </tr>
                </tbody>
            </table>
            <section label="B.5.1" status="3" xml:id="sect_B.5.1">
                <title>Specialization for Standard SOP Classes</title>
                <section label="B.5.1.1" status="4" xml:id="sect_B.5.1.1">
                    <title>Digital X-Ray Image Storage SOP Classes</title>
                    <para xml:id="para_6594ef38-3786-4fb9-9b3d-6ecbf1c40b92">The Digital X-Ray Image Storage - For Presentation SOP Class shall use the DX IOD with an Enumerated Value of FOR PRESENTATION for Presentation Intent Type (0008,0068).</para>
                    <para xml:id="para_fcc590ba-9c3a-4ee9-a63e-a87b5d08b0db">The Digital X-Ray Image Storage - For Processing SOP Class shall use the DX IOD with an Enumerated Value of FOR PROCESSING for Presentation Intent Type (0008,0068).</para>
                    <para xml:id="para_187707e5-d237-4d50-9e9a-437c0b77495c">...</para>
                </section>
                <section label="B.5.1.25" status="4" xml:id="sect_B.5.1.25">
                    <title><emphasis role="underline">Segmentation Storage SOP Classes</emphasis></title>
                    <para><emphasis role="bold"><emphasis role="underline">The Segmentation Storage SOP Class shall use the <olink targetdoc="PS3.3" targetptr="sect_A.51" xrefstyle="select: title"/> with an Enumerated Value of BINARY or FRACTIONAL for Segmentation Type (0062,0001).</emphasis></emphasis></para>
                    <para><emphasis role="bold"><emphasis role="underline">The Label Map Segmentation Storage SOP Class shall use the <olink targetdoc="PS3.3" targetptr="sect_A.51" xrefstyle="select: title"/> with an Enumerated Value of LABELMAP for Segmentation Type (0062,0001).</emphasis></emphasis></para>
                </section>
            </section>
        </section>
        
    </chapter>

    <chapter label="PS3.6" status="1" xml:id="chapter_PS3.6">
        <title>DICOM PS3.6 - Data Dictionary</title>
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.6 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        <section label="A" status="1" xml:id="chapter_A">
            <title>Registry of DICOM Unique Identifiers (UIDs) (Normative)</title>
            <table frame="box" label="A-1" rules="all" xml:id="table_A-1">
                <caption>UID Values</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_89103697-fd54-4c3f-abbe-5655ae35308b">
                                <emphasis role="bold">UID Value</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_cb5b6b8a-9aec-44f2-8cdc-800a7cd327b7">
                                <emphasis role="bold">UID Name</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_013510c8-4f44-4131-9319-6e72c777418c">
                                <emphasis role="bold">UID ​Keyword</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_08eb5599-7f50-4d17-bd10-6617063c6e31">
                                <emphasis role="bold">UID Type</emphasis>
                            </para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_42e221e9-efb2-49d7-8712-69266c9bc2d4">
                                <emphasis role="bold">Part</emphasis>
                            </para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">1.2.840.10008.​5.​1.​4.​1.​1.​66.​7</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Label Map Segmentation Storage</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">Label​Map​Segmentation​Storage</emphasis></emphasis></para>
                        </td>
                        <td align="left" colspan="1" rowspan="1">
                            <para><emphasis role="bold"><emphasis role="underline">SOP Class</emphasis></emphasis></para>
                        </td>
                        <td align="center" colspan="1" rowspan="1">
                            <para>
                                <emphasis role="bold"><emphasis role="underline"><olink targetdoc="PS3.4" targetptr="PS3.4" xrefstyle="select: labelnumber"/></emphasis></emphasis>
                            </para>
                        </td>
                    </tr>
                </tbody>
            </table>
        </section>
    </chapter>

    <chapter label="PS3.16" status="1" xml:id="chapter_PS3.16">
        <title>DICOM PS3.16 - Content Mapping Resource</title>
        <informaltable rules="all" frame="box">
            <tr>
                <td>
                    <para><emphasis role="italic">Amend PS3.16 as follows (changes to existing text
                        are bold and <emphasis role="bold"><emphasis role="underline"
                            >underlined</emphasis></emphasis> for additions and <emphasis role="bold"
                                ><emphasis role="strikethrough">struckthrough</emphasis></emphasis> for
                        removals):</emphasis></para>
                </td>
            </tr>
        </informaltable>
        
        
    <section label="CID 7019" status="3" xml:id="sect_CID_7019">
        <title>Segmentation Non-Image Source Purpose of Reference</title>
        <variablelist spacing="compact">
            <varlistentry>
                <term>
                    <emphasis role="bold">Version:</emphasis>
                </term>
                <listitem>
                    <para xml:id="para_d091c886-ada9-415c-b47d-a490a8f2ed34">
                            <emphasis role="bold"><emphasis role="strikethrough">20170914</emphasis><emphasis role="underline">20240914</emphasis></emphasis>
                        </para>
                </listitem>
            </varlistentry>
        </variablelist>
        <table frame="box" label="CID 7019" rules="all" xml:id="table_CID_7019">
            <caption>Segmentation Non-Image Source Purpose of Reference</caption>
            <thead>
                <tr valign="top">
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_1d9290e1-d48e-419d-9695-e5c5d23c335c">Coding Scheme Designator</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_27148331-5e2a-4610-abea-5bc7972d9150">Code Value</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_2e23d81a-9038-43f4-949b-05bd25445951">Code Meaning</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_018859d0-8ed2-411e-b4ba-66b9288be435">SNOMED-RT ID</para>
                    </th>
                    <th align="center" colspan="1" rowspan="1">
                        <para xml:id="para_31e75eca-82f4-4071-a303-6d421827ad1a">UMLS Concept Unique ID</para>
                    </th>
                </tr>
            </thead>
            <tbody>
                <tr valign="top">
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_662c7c7c-966e-4ed1-8474-f224347168b3">DCM</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_c61085df-b094-41a4-a020-df85e9f1c0ae">
                            <link linkend="DCM_128227" xrefstyle="select: nopage">128227</link>
                        </para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para xml:id="para_1beffa8a-3aea-448f-b02c-eb51c27a8a45">Source real world value map</para>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_e91ebd71-b5a6-446e-bee0-18fcbec75083"/>
                    </td>
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="para_d797c6a4-3204-40eb-8fae-f97143cd9c8f"/>
                    </td>
                </tr>
                <tr valign="top">
                        <td align="center" colspan="1"><para><emphasis role="bold"><emphasis role="underline">DCM</emphasis></emphasis></para></td>
                        <td align="center" colspan="1"><para><link linkend="DCM_128228" xrefstyle="select: nopage"><emphasis role="bold"><emphasis role="underline">128228</emphasis></emphasis></link></para></td>
                        <td align="left" colspan="1"><para><emphasis role="bold"><emphasis role="underline">Source segmentation</emphasis></emphasis></para></td>
                        <td align="center" colspan="1"><para/></td>
                        <td align="center" colspan="1"><para/></td>
                    </tr>
            </tbody>
        </table>
    </section>
        
        <section label="D" status="1" xml:id="chapter_D">
            <title>DICOM Controlled Terminology Definitions (Normative)</title>
            <table frame="box" label="D-1" rules="all" xml:id="table_D-1">
                <caption>DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding Scheme Version "01")</caption>
                <thead>
                    <tr valign="top">
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_f05607a1-90fe-4408-ad56-d194c3203ae5">Code Value</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_022556ee-7c69-4138-a152-b3fa16bb86b2">Code Meaning</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_e0270183-1962-48d5-877f-b63624671b37">Definition</para>
                        </th>
                        <th align="center" colspan="1" rowspan="1">
                            <para xml:id="para_5aa10e80-ebb2-45b9-a2a9-fe4b3eb4497c">Notes</para>
                        </th>
                    </tr>
                </thead>
                <tbody>
                    <tr valign="top">
            <td align="center" colspan="1" rowspan="1">
                <para xml:id="DCM_128227">128227</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58d31393-50d6-4bc9-b8e3-ba9e4a98211b">Source real world value map</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d56d4a1b-196f-4eb5-8fa7-e61c653a1588">Real world value map used as the source for derivation. E.g., the map applied to source images before processing them, such as for a threshold based segmentation operation.</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59362367-2175-4de5-ae8e-10b1b15c4b32"/>
            </td>
        </tr>
                    <tr valign="top">
                    <td align="center" colspan="1" rowspan="1">
                        <para xml:id="DCM_128228"><emphasis role="bold"><emphasis role="underline">128228</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Source segmentation</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para><emphasis role="bold"><emphasis role="underline">Segmentation used as the source for derivation. E.g., the binary segmentation that was converted to a labelmap segmentation, or vice versa.</emphasis></emphasis></para>
                    </td>
                    <td align="left" colspan="1" rowspan="1">
                        <para/>
                    </td>
                    </tr>
                    
                </tbody>
            </table>
        </section>
        
    