<?xml version="1.0" encoding="utf-8" standalone="no"?>
<book xmlns="http://docbook.org/ns/docbook" xmlns:xhtml="http://www.w3.org/1999/xhtml" xmlns:xl="http://www.w3.org/1999/xlink" label="PS3.20" version="5.0" xml:id="PS3.20">
  <title>PS3.20</title>
  <subtitle>DICOM PS3.20 2026c - Imaging Reports using HL7 Clinical Document Architecture</subtitle>
  <info>
    <author>
      <orgname>DICOM Standards Committee</orgname>
    </author>
    <copyright>
      <year>2026</year>
      <holder>NEMA</holder>
    </copyright>
      <legalnotice>
          <para xml:id="para_c5271702-dcbc-4554-ac97-800adca6ff13">A DICOM® publication</para>
      </legalnotice>
  </info>
  <chapter label="" status="1" xml:id="chapter_Notice">
    <title>Notice and Disclaimer</title>
    <para xml:id="para_adf733c8-e316-4816-8cd9-71de831f0b0a">The information in this publication was considered technically sound by the consensus of
      persons engaged in the development and approval of the document at the time it was developed.
      Consensus does not necessarily mean that there is unanimous agreement among every person
      participating in the development of this document.</para>
    <para xml:id="para_167ad6a4-509e-45e5-8f53-145e76fd37ff">NEMA standards and guideline publications, of which the document contained herein is one,
      are developed through a voluntary consensus standards development process. This process brings
      together volunteers and/or seeks out the views of persons who have an interest in the topic
      covered by this publication. While NEMA administers the process and establishes rules to
      promote fairness in the development of consensus, it does not write the document and it does
      not independently test, evaluate, or verify the accuracy or completeness of any information or
      the soundness of any judgments contained in its standards and guideline publications.</para>
    <para xml:id="para_c5f0e641-c627-450a-bbf8-89da6fb69fbd">NEMA disclaims liability for any personal injury, property, or other damages of any nature
      whatsoever, whether special, indirect, consequential, or compensatory, directly or indirectly
      resulting from the publication, use of, application, or reliance on this document. NEMA
      disclaims and makes no guaranty or warranty, expressed or implied, as to the accuracy or
      completeness of any information published herein, and disclaims and makes no warranty that the
      information in this document will fulfill any of your particular purposes or needs. NEMA does
      not undertake to guarantee the performance of any individual manufacturer or seller's products
      or services by virtue of this standard or guide.</para>
    <para xml:id="para_dbac6629-149e-4846-ab12-597c97e1809c">In publishing and making this document available, NEMA is not undertaking to render
      professional or other services for or on behalf of any person or entity, nor is NEMA
      undertaking to perform any duty owed by any person or entity to someone else. Anyone using
      this document should rely on his or her own independent judgment or, as appropriate, seek the
      advice of a competent professional in determining the exercise of reasonable care in any given
      circumstances. Information and other standards on the topic covered by this publication may be
      available from other sources, which the user may wish to consult for additional views or
      information not covered by this publication.</para>
    <para xml:id="para_974be0ee-883a-4cd0-b6ac-af871ec741fa">NEMA has no power, nor does it undertake to police or enforce compliance with the contents
      of this document. NEMA does not certify, test, or inspect products, designs, or installations
      for safety or health purposes. Any certification or other statement of compliance with any
      health or safety-related information in this document shall not be attributable to NEMA and is
      solely the responsibility of the certifier or maker of the statement.</para>
  </chapter>
  <chapter label="" status="1" xml:id="chapter_Foreword">
    <title>Foreword</title>
    <para xml:id="para_5cd3f15c-db82-4f61-b7da-e0974c5c5c5a">This DICOM Standard was developed according to the procedures of the DICOM Standards Committee.</para>
    <para xml:id="para_febe0a70-3963-4a39-b265-67834f6390a7">The DICOM Standard is structured as a multi-part document using the guidelines established in <xref linkend="biblio_ISODirectives2"/>.</para>
    <para xml:id="para_9ebbc019-87c6-4282-8978-80f20c2ed332">DICOM® is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications of medical information, all rights reserved.</para>
    <para xml:id="para_08b2d2c1-d3b1-4c75-b1dd-e233881eab68">HL7® and CDA® are the registered trademarks of Health Level Seven International, all rights reserved.</para>
    <para xml:id="para_0bce879c-1cc9-4d62-9423-271830bbd29b">SNOMED®, SNOMED Clinical Terms®, SNOMED CT® are the registered trademarks of the International Health Terminology Standards Development <foreignphrase xml:lang="en-GB">Organisation</foreignphrase> (IHTSDO), all rights reserved.</para>
    <para xml:id="para_d0ec6978-dc6d-4ae8-83ce-9820e5505da6">LOINC® is the registered trademark of Regenstrief Institute, Inc, all rights reserved.</para>
  </chapter>
  <chapter label="1" status="1" xml:id="chapter_1">
    <title>Scope and Field of Application</title>
    <para xml:id="para_442ff5b0-7139-4f11-8ed7-a4506b7bb8a2">This Part of the DICOM Standard specifies templates for the encoding of imaging reports using the HL7 Clinical Document Architecture Release 2 (CDA R2, or simply CDA) Standard. Within this scope are clinical procedure reports for specialties that use imaging for screening, diagnostic, or therapeutic purposes.</para>
    <para xml:id="para_5d8dc6a8-16b3-4804-b190-7417bc2d3e2b">This Part constitutes an implementation guide for CDA, and is harmonized with the approach to standardized templates for CDA implementation guides developed by HL7. It also provides Business Names for data elements that link data in user terminology, e.g., collected by a report authoring application, to specific CDA encoded elements.</para>
    <para xml:id="para_265c46dd-9539-4308-8d9f-d136ae780a3b">As an implementation guide for imaging reports, particular attention is given to the use and reference of data collected in imaging procedures as explicit evidence within reports. This data includes images, waveforms, measurements, annotations, and other analytic results managed as DICOM SOP Instances. Specifically, this Part includes a specification for transformation into CDA documents of DICOM Structured Report instances that represent imaging reports.</para>
  </chapter>
  <chapter label="2" status="1" xml:id="chapter_2">
    <title>Normative and Informative References</title>
    <para xml:id="para_5bf51e91-b581-4a63-9f32-4037b718f674">The following standards contain provisions that, through reference in this text, constitute provisions of this Standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this Standard are encouraged to investigate the possibilities of applying the most recent editions of the standards indicated below.</para>
    <bibliography>
      <biblioentry xml:id="biblio_ISODirectives2">
        <abbrev>ISO/IEC Directives, Part 2</abbrev>
        <author>
          <orgname>ISO/IEC</orgname>
        </author>
        <date>2021</date>
        <edition>9.0</edition>
        <title>Rules for the structure and drafting of International Standards</title>
        <bibliosource>
          <link xl:href="http://www.iso.org/sites/directives/current/part2/index.xhtml"/>
        </bibliosource>
      </biblioentry>
    </bibliography>
    <para xml:id="para_700045e7-1290-4dc7-adfb-77a032059e6f">ANSI/HL7 CDA®, R2-2005 HL7 Version 3 Standard: Clinical Document Architecture (CDA) Release 2, 2005 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=7"/>)</para>
    <para xml:id="para_23c02427-c34a-4ff3-b12a-558c648cf9be">CDA® is a registered trademark of HL7 International.</para>
    <para xml:id="para_c2829e9f-a6d3-46ed-8501-074829ef4f38">ANSI/HL7 V3 CPPV3MODELS, R1-2012 HL7 Version 3 Standard: Core Principles and Properties of Version 3 Models, Release 1 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=58"/>)</para>
    <para xml:id="para_156ba210-8415-4573-83a1-65cd16c231c5">ANSI/HL7 V3 CMET, R2-2009 Health Level Seven Version 3 Standard: Common Message Element Types, Release 2, 2009.</para>
    <para xml:id="para_02f1befb-1d74-4e1f-a37b-249a264eaa23">ANSI/HL7 V3 DT, R1-2004 HL7 Version 3 Data Types Abstract Specification, Release 1 - November 2004. [Note: this specific release version is required by CDA R2]</para>
    <para xml:id="para_a2efb7e2-2874-4837-8f56-e4a7c9acb77f">ANSI/HL7 V3 XMLITSDT, R1-2004 HL7 Version 3 XML Implementation Technology Specification - Data Types, Release 1 - April 2004. [Note: this specific release version is required by CDA R2]</para>
    <para xml:id="para_4b85d1c9-a1c7-4cd3-957b-09724f49fd98">HL7 CDA R2 DIR IG, R1-2009 Health Level Seven Implementation Guide for CDA Release 2: Imaging Integration, Basic Imaging Reports in CDA and DICOM, Diagnostic Imaging Reports (DIR) Release 1.0 - Informative, 2009 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=13"/>)</para>
    <para xml:id="para_b065995b-1369-4d28-aee3-3c87cd590583">HL7 CDAR2_IG_IHE_CONSOL HL7 Implementation Guide for CDA® Release 2: IHE Health Story Consolidation, Release 1.1 - US Realm, Draft Standard for Trial Use, July 2012 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=258"/>)</para>
    <para xml:id="para_c24e435a-b4bc-432b-89d8-9618223ecdc9">HL7 CDAR2_IG_CCDA_CLINNOTES_R2 HL7 Implementation Guide for CDA® Release 2: Consolidated CDA Templates for Clinical Notes, Release 2 - US Realm, Draft Standard for Trial Use, November 2014 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=379"/>)</para>
    <para xml:id="para_bfd110e5-f78e-42f4-95bb-eaa455709122">HL7 CDAR2_IG_GREENMOD4CCD HL7 Implementation Guides for CDA® R2: greenCDA Modules for CCD®, Release 1 - Informative, April 2011(<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=136"/>)</para>
    <para xml:id="para_41b5304d-f52a-4bc1-a490-f214f8b17681">HL7 Templates HL7 Templates Standard: Specification and Use of Reusable Information Constraint Templates, Release 1 - DSTU, October 2014 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=377"/>)</para>
    <para xml:id="para_2cea67c6-5e42-4520-af66-2799b9fe18c8">HL7 CDA Digital Signatures HL7 Implementation Guide for CDA® Release 2: Digital Signatures and Delegation of Rights, Release 1 - DSTU, October 2014 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=375"/>)</para>
    <para xml:id="para_3eba9fcc-5145-45e7-8e4c-bc8c05c5fd6b">HL7 v3-2014 HL7 Version 3 Interoperability Standards, Normative Edition 2014 (<link xl:href="http://www.hl7.org/implement/standards/product_brief.cfm?product_id=362]"/>
</para>
    <para xml:id="para_0f9d88be-7d15-435f-8a28-259cf48e2946">IHE Card Sup CIRC IHE Cardiology Technical Framework Supplement, Cardiac Imaging Report Content, Trial Implementation, July 2011 (<link xl:href="http://www.ihe.net/Technical_Frameworks/#cardiology"/>)</para>
    <para xml:id="para_6edc6d9e-55e8-415e-9223-6ff0122c2181">IHE ITI TF IHE IT Infrastructure Technical Framework, Revision 11.0, September 2014 (<link xl:href="http://www.ihe.net/Technical_Frameworks/#IT"/>)</para>
    <para xml:id="para_cec98f18-daf4-4819-8b9f-d5ba89179c72">IHE PCC TF IHE Patient Care Coordination Technical Framework, Revision 10.0, November 2014 (<link xl:href="http://www.ihe.net/Technical_Frameworks/#pcc"/>)</para>
    <para xml:id="para_e6686b5d-6e7c-4a9f-8c6b-8a23f3691ad2">IHE RAD TF IHE Radiology Technical Framework, Revision 13.0, July 2014 (<link xl:href="http://www.ihe.net/Technical_Frameworks/#radiology"/>)</para>
    <para xml:id="para_3c548fc5-6a77-40ac-af16-e7461d9c73e6">LOINC Logical Observation Identifier Names and Codes, Regenstrief Institute, Indianapolis 2013.</para>
    <para xml:id="para_8f1725b7-bade-4169-93d7-a8134969bd29">This product includes all or a portion of the LOINC® table, LOINC panels and forms file, LOINC document ontology file, and/or LOINC hierarchies file, or is derived from one or more of the foregoing, subject to a license from Regenstrief Institute, Inc. Your use of the LOINC table, LOINC codes, LOINC panels and forms file, LOINC document ontology file, and LOINC hierarchies file also is subject to this license, a copy of which is available at http://loinc.org/terms-of-use. The current complete LOINC table, LOINC Users' Guide, LOINC panels and forms file, LOINC document ontology file, and LOINC hierarchies file are available for download at <link xl:href="http://loinc.org"/>. The LOINC table and LOINC codes are copyright © 1995-2013, Regenstrief Institute, Inc. and the Logical Observation Identifiers Names and Codes (LOINC) Committee. The LOINC panels and forms file, LOINC document ontology file, and LOINC hierarchies file are copyright © 1995-2013, Regenstrief Institute, Inc. All rights
      reserved.</para>
    <para xml:id="para_317624b3-8ef1-4fdf-acea-8a03e26acbbf">THE LOINC TABLE (IN ALL FORMATS), LOINC PANELS AND FORMS FILE, LOINC DOCUMENT ONTOLOGY FILE, AND LOINC HIERARCHIES ARE PROVIDED "AS IS." ANY EXPRESS OR IMPLIED WARRANTIES ARE DISCLAIMED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.</para>
    <para xml:id="para_aa335c5b-c28e-4c6f-bf47-6d38526540a3">LOINC® is a registered United States trademark of Regenstrief Institute, Inc. A small portion of the LOINC table may include content (e.g., survey instruments) that is subject to copyrights owned by third parties. Such content has been mapped to LOINC terms under applicable copyright and terms of use. Notice of such third party copyright and license terms would need to be included if such content is included.</para>
    <para xml:id="para_3c76379a-ecb2-48b4-b8cf-19fd992d07de">RFC4646 Tags for Identifying Languages, The Internet Society, 2005</para>
    <para xml:id="para_d6e320d9-19a1-4e32-a1d5-a634fcc0b644">SNOMED CT® Systematized Nomenclature of Medicine - Clinical Terms, International Release, International Health Terminology Standards Development Organisation (IHTSDO), January 2015</para>
    <para xml:id="para_a9b7da55-9109-461d-b6d2-0288fe3660bc">SNOMED CT is a registered trademark of the International Health Terminology Standard Development Organisation (IHTSDO).</para>
    <para xml:id="para_ed6f6c29-b71f-462d-b761-8c2f51f37b08">UCUM Unified Code for Units of Measure, Regenstrief Institute, Indianapolis 2013.</para>
    <para xml:id="para_570677f9-599e-4b7a-9b1c-0dca5acebba7">This product includes all or a portion of the UCUM table, UCUM codes, and UCUM definitions or is derived from it, subject to a license from Regenstrief Institute, Inc. and The UCUM Organization. Your use of the UCUM table, UCUM codes, UCUM definitions also is subject to this license, a copy of which is available at http://unitsofmeasure.org. The current complete UCUM table, UCUM Specification are available for download at <link xl:href="http://unitsofmeasure.org"/>. The UCUM table and UCUM codes are copyright © 1995-2013, Regenstrief Institute, Inc. and the Unified Codes for Units of Measures (UCUM) Organization. All rights reserved.</para>
    <para xml:id="para_1137a7ee-dced-4ed9-9376-baedb127f0b8">THE UCUM TABLE (IN ALL FORMATS), UCUM DEFINITIONS, AND SPECIFICATION ARE PROVIDED "AS IS." ANY EXPRESS OR IMPLIED WARRANTIES ARE DISCLAIMED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.</para>
    <para xml:id="para_34abbbbd-35a6-4446-b797-c651a2764155">XML Extensible Markup Language (XML) 1.0 (Fifth Edition), World Wide Web Consortium, 2008 (<link xl:href="http://www.w3.org/TR/REC-xml/"/>)</para>
    <para xml:id="para_6da86eb2-a9f6-4e2f-b7f9-eb4bdd1360a5">XML Schema Datatypes XML Schema Part 2: Datatypes Second Edition, World Wide Web Consortium, 2004 (<link xl:href="http://www.w3.org/TR/xmlschema-2/"/>)</para>
    <para xml:id="para_81922d63-79d7-4aff-92bb-4097cbff861b">xml:id xml:id Version 1.0, World Wide Web Consortium, 2005 (<link xl:href="http://www.w3.org/TR/xml-id"/>)</para>
    <para xml:id="para_ea577820-35e2-48f0-8072-5cdbd7fb3cf9">XPath XML Path Language (XPath), Version 1.0, World Wide Web Consortium, 1999 (<link xl:href="http://www.w3.org/TR/xpath/"/>)</para>
  </chapter>
    <chapter label="3" status="1" xml:id="chapter_3">
    <title>Definitions</title>
    <para xml:id="para_80c4cfbf-8fa3-4a99-af91-6ba52690f148">For the purposes of this Standard the following definitions apply.</para>
        <glossary>
            <glossdiv label="3.1" status="2" xml:id="sect_3.1">
                <title>Codes and Controlled Terminology Definitions</title>
                <para xml:id="para_7b61465b-852f-47ec-b0b0-1b5f8d193b2a">The following definitions are commonly used in this Part of the DICOM Standard:</para>
                <glossentry>
                    <glossterm>Context Group</glossterm>
                    <glossdef>
                        <para xml:id="para_0509cb75-2f71-47c0-8f82-443c5d89be47">A set of coded concepts defined by a Mapping Resource forming a set appropriate to use in a particular context.</para>
                    </glossdef>
                </glossentry>
                <glossentry>
                    <glossterm>Context ID</glossterm>
                    <acronym>CID</acronym>
                    <glossdef>
                        <para xml:id="para_64d131bc-bfd6-4658-baf6-a582fb123dcb">Identifier of a Context Group.</para>
                    </glossdef>
                </glossentry>
                <glossentry>
                    <glossterm>Template</glossterm>
                    <glossdef>
                        <para xml:id="para_41de6367-201c-416b-8e5b-907ec84eccf8">A pattern that describes the Content Items, Value Types, Relationship Types and Value Sets that may be used in part of a Structured Report content tree, or in other Content Item constructs, such as Acquisition Context or Protocol Context. Analogous to a Module of an Information Object Definition.</para>
                    </glossdef>
                </glossentry>
                <glossentry>
                    <glossterm>Template ID</glossterm>
                    <acronym>TID</acronym>
                    <glossdef>
                        <para xml:id="para_745793a9-ab79-4b9c-8a5a-2e5c60758877">Identifier of a Template.</para>
                    </glossdef>
                </glossentry>
                <glossentry>
                    <glossterm>Coding Schemes</glossterm>
                    <glossdef>
                        <para xml:id="para_a5c7948c-2b09-40d4-8bfc-8b6fa41634b3">Dictionaries (lexicons) of concepts (terms) with assigned codes and well defined meanings.</para>
                    </glossdef>
                </glossentry>
            </glossdiv>
            <glossdiv label="3.2" status="2" xml:id="sect_3.2">
                <title>Vocabulary Model Definitions</title>
                <para xml:id="para_bf0a4f1a-a595-45d8-a1f2-35477a692393">The following terms used in this Part of the DICOM Standard are defined in HL7 Core Principles and Properties of Version 3 Models:</para>
                <glossentry>
                    <glossterm>Concept Domain</glossterm>
                    <glossdef>
                        <para xml:id="para_51445782-b2e1-4287-a85f-3702f501ecca">A named category of like concepts (a semantic type) that is specified in the vocabulary declaration of an attribute in an information model. It constrains the intent of the coded element while deferring the binding of the element to a specific set of codes until later in the specification process.</para>
                    </glossdef>
                </glossentry>
                <glossentry>
                    <glossterm>Value Set</glossterm>
                    <glossdef>
                        <para xml:id="para_22c1650d-ed8a-4a18-8e6d-952c79b6cc0e">A uniquely identifiable set of valid concept identifiers. Value sets constrain the permissible content for a coded element in an information model or data type specification.</para>
                    </glossdef>
                </glossentry>
                <glossentry>
                    <glossterm>Vocabulary Binding</glossterm>
                    <glossdef>
                        <para xml:id="para_a53c399a-a876-43b5-a4f1-477670a249a3">The mechanism of identifying specific codes to be used to express the semantics of coded model elements in information models or coded data type properties. Vocabulary Binding may bind the coded element or data type property to a single fixed value code, or may bind it to a Value Set Assertion.</para>
                    </glossdef>
                </glossentry>
            </glossdiv>
            <glossdiv label="3.3" status="2" xml:id="sect_3.3">
                <title>Template Definitions</title>
                <para xml:id="para_a7ec167b-35d1-48f5-bff2-73f3dc690218">The following term used in this Part of the DICOM Standard is defined in the HL7 Templates Standard, and applies to CDA template specifications:</para>
                <glossentry>
                    <glossterm>Template</glossterm>
                    <glossdef>
                        <para xml:id="para_9f246ba5-b966-445e-83c3-82e4a380527f">A set of conformance statements which further constrain an existing information model.</para>
                    </glossdef>
                </glossentry>
            </glossdiv>
            <glossdiv label="3.4" status="2" xml:id="sect_3.4">
                <title>Imaging Report Definitions</title>
                <para xml:id="para_8e1c3a99-3487-4e18-92de-39ddcd8662c8">The following definitions apply to terms used in this Part of the Standard:</para>
                <glossentry>
                    <glossterm>Business Name</glossterm>
                    <glossdef>
                        <para xml:id="para_63c6a187-7de1-4585-9de4-2fd5e650cd29">Identifier for a CDA Data Element, Attribute, or structure of Data Elements that corresponds to a business requirement for information exchange.</para>
                    </glossdef>
                </glossentry>
            </glossdiv>
        </glossary>
  </chapter>
  <chapter label="4" status="1" xml:id="chapter_4">
    <title>Symbols and Abbreviations</title>
    <para xml:id="para_30063e8b-b2bb-4b9a-851e-2a6203cf49ae">The following symbols and abbreviations are used in this Part of the Standard.</para>
    <variablelist>
      <varlistentry>
        <term>ANSI</term>
        <listitem>
          <para xml:id="para_b4448a14-207e-4e12-b44a-fb87c720a265">American National Standards Institute</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>CDA</term>
        <listitem>
          <para xml:id="para_ccad847f-57f0-4a59-bbd7-fe2b7f4eec19">Clinical Document Architecture (HL7)</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>DICOM</term>
        <listitem>
          <para xml:id="para_ae49e673-7128-4d31-a36d-8388b0f70295">Digital Imaging and Communications in Medicine</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>HL7</term>
        <listitem>
          <para xml:id="para_cbdc98f4-b997-4027-a5fb-cf47e2a95803">Health Level 7</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>HMD</term>
        <listitem>
          <para xml:id="para_e35cdf9f-d1ef-488c-a555-fa91f810e523">Hierarchical Message Description (HL7)</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>IE</term>
        <listitem>
          <para xml:id="para_a48f4b6c-10fa-4bdd-b117-c2477fc66d90">Information Entity</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>IHE</term>
        <listitem>
          <para xml:id="para_01db106d-89e7-4bcf-b2a9-eeb494383a67">Integrating the Healthcare Enterprise</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>IOD</term>
        <listitem>
          <para xml:id="para_040dea16-93ee-411a-88bd-111a27f86011">Information Object Definition</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>ISO</term>
        <listitem>
          <para xml:id="para_a4ed2202-4abf-4dcd-b33f-22e242cd39d9">International Standards Organization</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>LOINC</term>
        <listitem>
          <para xml:id="para_02fc39bf-7ace-4f95-9de5-2da0c3fc8751">Logical Observation Identifiers Names and Codes</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>MRRT</term>
        <listitem>
          <para xml:id="para_006ee372-1a98-4749-a784-246bd507460c">IHE Management of Radiology Report Templates Profile</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>NEMA</term>
        <listitem>
          <para xml:id="para_bef081ce-6222-4e2c-bb58-cec4b0203423">National Electrical Manufacturers Association</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>OID</term>
        <listitem>
          <para xml:id="para_1b9d12f5-1773-4b2e-90b3-65a3f423af01">Object Identifier (ISO 8824)</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>RSNA</term>
        <listitem>
          <para xml:id="para_1f1dda72-957d-4746-bb9d-54344e7a7bbd">Radiological Society of North America</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>SNOMED</term>
        <listitem>
          <para xml:id="para_6894db8f-18b0-448d-9372-3b458db07285">Systematized Nomenclature of Medicine</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>SR</term>
        <listitem>
          <para xml:id="para_6f8493da-d051-475b-85e0-14649dfbeca7">Structured Reporting</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>UCUM</term>
        <listitem>
          <para xml:id="para_ed5392cc-a4ae-40f1-b77a-3235620e0be4">Unified Code for Units of Measure</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>UID</term>
        <listitem>
          <para xml:id="para_4d4ae8e7-c07b-4bb0-8d78-cd87953b76dd">Unique Identifier</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>XML</term>
        <listitem>
          <para xml:id="para_caba6196-887c-4baf-b698-efd7165ec465">Extensible Markup Language</para>
        </listitem>
      </varlistentry>
    </variablelist>
    <para xml:id="para_a39107a4-864c-4189-9f39-1b219d3b8285">The following symbols and abbreviations for HL7 v3 Data Types are used in this Part of the Standard.</para>
    <variablelist>
      <varlistentry>
        <term>AD</term>
        <listitem>
          <para xml:id="para_908c5b23-56f2-49d1-a426-afaafad5fc9d">Postal Address</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>CE</term>
        <listitem>
          <para xml:id="para_fc626b3c-27ca-4999-a605-daebc98c58c6">Coded With Equivalents</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>CD</term>
        <listitem>
          <para xml:id="para_477590e1-8554-4bd2-883b-24e41da7fd76">Concept Descriptor</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>CS</term>
        <listitem>
          <para xml:id="para_91b9a9c4-c62b-4d16-81cf-991efcaa0991">Coded Simple Value</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>ED</term>
        <listitem>
          <para xml:id="para_7afd5dc0-a842-4dd3-8e1f-52ad028c6c59">Encapsulated Data</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>EN</term>
        <listitem>
          <para xml:id="para_f438f505-c466-4561-a0fb-d20e59287fb1">Entity Name</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>II</term>
        <listitem>
          <para xml:id="para_b453fbe5-ed29-4487-8c3d-3fcc1efeae6b">Instance Identifier</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>INT</term>
        <listitem>
          <para xml:id="para_642c2d0c-a2f5-4938-868e-e6da0ae69ddd">Integer Number</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>IVL&lt;&gt;</term>
        <listitem>
          <para xml:id="para_ad0d77e1-aebb-4f2b-bfde-4e36ae8eb186">Interval</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>LIST&lt;&gt;</term>
        <listitem>
          <para xml:id="para_07165c60-6243-4000-9fbc-71201e58d7bd">List</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>OID</term>
        <listitem>
          <para xml:id="para_d884b641-7b1c-4a19-8d51-fb5f5a9bf391">ISO Object Identifier</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>ON</term>
        <listitem>
          <para xml:id="para_d1c05e4c-5816-421d-a34f-7af88f3f4bcd">Organization Name</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>PN</term>
        <listitem>
          <para xml:id="para_9b3e1284-1253-471b-bca0-4f73c184d799">Person Name</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>PQ</term>
        <listitem>
          <para xml:id="para_f4c9b408-7329-481c-a064-1df30a21b7ca">Physical Quantity</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>REAL</term>
        <listitem>
          <para xml:id="para_74f3536c-003f-4a2b-bcf6-2935187847eb">Real Number</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>ST</term>
        <listitem>
          <para xml:id="para_2b6c2d42-c34e-4022-9c0d-1dc8a17829b4">Character String</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>TEL</term>
        <listitem>
          <para xml:id="para_a273a542-7f04-4f89-b5fd-8d44a7754abb">Telecommunication Address</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>TS</term>
        <listitem>
          <para xml:id="para_04fab938-df4a-420a-aa16-de9f90c5bcb4">Point in Time</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>UID</term>
        <listitem>
          <para xml:id="para_e0148d2d-1727-4448-9c33-1cc40a64fb3e">Unique Identifier String</para>
        </listitem>
      </varlistentry>
      <varlistentry>
        <term>URL</term>
        <listitem>
          <para xml:id="para_9d1bd267-74cc-4b72-9e7b-b3f0e3b44a03">Universal Resource Identifier</para>
        </listitem>
      </varlistentry>
    </variablelist>
  </chapter>
  <chapter label="5" status="1" xml:id="chapter_5">
    <title>Conventions</title>
    <section label="5.1" status="2" xml:id="sect_5.1">
      <title>Template Metadata</title>
      <para xml:id="para_22df0f42-9948-4e85-b4fa-d7984abbd656">Each template has a set of metadata, as specified in the HL7 Templates Specification. The metadata is presented as a table, as shown in <xref linkend="table_5.1-1" xrefstyle="select: label"/>.</para>
      <table frame="box" label="5.1-1" rules="all" xml:id="table_5.1-1">
        <caption>Template metadata table format</caption>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1b25cd9b-f851-49a6-9488-8303432b9fd3">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd6bc2cf-404a-4786-845a-2c6ee4c35be1">OID (see <xref linkend="sect_5.1.1" xrefstyle="select: label"/>)</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e63c5c6-c080-4911-ae98-837f55091cbe">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf8bb71a-6b62-4d19-ac2f-58ed61fdf0e4">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc472a6c-68d2-4b81-a8c4-1dde9a025c60">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c54f250-b80f-49cd-9510-3d5a9f54548e">(see <xref linkend="sect_5.1.1" xrefstyle="select: label"/>)</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df70ba72-09b7-4c66-a60b-9c9c9f39c807">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64cae269-4684-466d-a1b0-52385eb80358">"draft", "active", "review" or "retired"</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40cea5fc-5436-47f0-a740-2f0323a4cfb3">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c16759d-a9b2-4549-bb9d-14945d5dae5f">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9970eb06-90aa-4ab6-bc45-453cfb2f4a7c">type of the template, e.g., CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_edafa855-3e9a-4804-bbb0-c48f25445ee1">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_271dbf45-0c58-486c-9cf3-1094b55c9a51">relationships to other templates or model artifacts</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_276ff672-bb17-4695-9b35-f1c86183b350">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8a32d01-4e4b-4b0a-9252-b2463d29c563">"parent node", "sibling node" (see <xref linkend="sect_5.1.2" xrefstyle="select: label"/>)</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_52fabe17-168d-4eee-9636-4e6bc8b3a2a3">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4b69a9c-3b1b-48e3-ac23-e31811751578">"open", "closed"(see <xref linkend="sect_5.1.3" xrefstyle="select: label"/>)</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19fa5674-a0e3-4cba-b6e4-1601ffff7295">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </table>
      <section label="5.1.1" status="3" xml:id="sect_5.1.1">
        <title>Template IDs and Version</title>
        <para xml:id="para_197da84c-3c5f-4992-b63e-ea3e2ca41828">Template identifiers (templateId) are assigned for each document, section, and entry level template. When valued in an instance, the template identifier signals the imposition of a set of template-defined constraints. The value of this attribute (e.g., @root="2.16.840.1.113883.10.20.22.4.8") provides a unique identifier for the template in question.</para>
        <para xml:id="para_f2b98a15-c4eb-4fcf-a75a-15ade1269174">A template may be further qualified by a version label. This label may be used as the extension attribute of the templateID (e.g., @extension="20150309"). All versions of a template, regardless of the version label, must be compatible; i.e., they may vary only by optional content conformance requirements. Thus the version label is typically not used as a conformance constraint.</para>
        <para xml:id="para_11071905-6f6d-414b-aba6-8c9f1f21ce5d">Within this Standard, template versions are identified by the string "DICOM" and the date of adoption (represented as YYYYMMDD), separated by a hyphen (e.g., DICOM-20150523).</para>
      </section>
      <section label="5.1.2" status="3" xml:id="sect_5.1.2">
        <title>Context</title>
        <para xml:id="para_724d94c9-a9a6-4835-ae9b-503d71f4c387">As described in the HL7 Template specification section 2.9.9.4, the context identifies whether the template applies to the parent node in which the templateID is an element, or applies to its sibling nodes in the template table. These typically are applied respectively to templates with a single parent element with child element structure, and to templates with flat list of sibling elements (see <xref linkend="sect_5.2.8" xrefstyle="select: label"/>).</para>
      </section>
      <section label="5.1.3" status="3" xml:id="sect_5.1.3">
        <title>Open and Closed Templates</title>
        <para xml:id="para_36cfafab-2bae-4010-9124-fa43acf53509">Each template is defined as being either "open" or "closed". In "open" templates, all of the features of the CDA Specification are allowed except as constrained by the templates. By contrast, a "closed" template specifies everything that is allowed and nothing further may be included.</para>
      </section>
    </section>
    <section label="5.2" status="2" xml:id="sect_5.2">
      <title>Template Table Structure</title>
      <para xml:id="para_81948ecc-78b1-4387-938d-1bbae75f3362">Each template is specified in tabular form, as shown in <xref linkend="table_5.2-1" xrefstyle="select: label"/>.</para>
      <table frame="box" label="5.2-1" rules="all" xml:id="table_5.2-1">
        <caption>Template table format</caption>
        <thead>
          <tr valign="top">
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb035efd-3db0-41fe-948a-47f36f8c87ac">Business Name</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_dec13d50-5ff3-42e3-af95-27ebf5f94a11">Nest Level</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0aa0535-430f-4ef1-ac4b-d5f3db4e7e24">Element/​Attribute</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_acdacb1b-ceb9-4801-8106-93004c5b8eed">Card</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_8bb607a1-ebfe-4d01-840d-da4180f73a52">Elem/Attr Conf</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_934e6425-d267-4787-9ae9-11dcf22ac00f">Data Type</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9796642-344b-44fc-80df-f6caf27e1402">Value Conf</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_e317bed5-e27c-45f8-b165-41891f7fdcab">Value</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d7dd2b2-25b5-40e0-a25d-614cdcddf338">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d348f6e7-4e9d-4e48-8989-7abbaac1ae1a">
                <emphasis role="bold">Scoping​Business​Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c21667d5-c5fa-48bd-98aa-0db4999c5d03">Element​Business​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_90e90cdb-b590-48a1-a98e-4be3f4a9bd89">
                <emphasis role="italic">Referenced​Business​Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </table>
      <section label="5.2.1" status="3" xml:id="sect_5.2.1">
        <title>Business Name</title>
        <para xml:id="para_c564fd04-de25-42c5-9804-72e3bab041a9">This Part uses Business Names to identify and map common data elements from clinical imaging reports into the proper context-specific CDA/XML structure.</para>
        <para xml:id="para_6a5adc3d-fe06-4e32-9b9f-b207f3ed1267">A Business Name is assigned to each element or attribute that may have a user-specified value assigned in the production of the clinical document instance. Business Names are specified to facilitate the implementation of production logic for clinical report authoring applications. The benefit is that developers of clinical report authoring applications are not required to have an in depth knowledge of CDA, the HL7 v3 R-MIM data model, or the XML structures. The use of readable and intuitive Business Names provides a method of direct access to insert data that is specific to each clinical report instance.</para>
        <note>
          <para xml:id="para_acfd4acb-2ffd-44fc-a8d8-7b7ab49dec2a">Business Names are also described in the HL7 greenCDAModules for CCD specification, but that specification implies the use of a specific XML structure for production logic that is not required by this Part. The specification of production logic using Business Names is outside the scope of this Part.</para>
        </note>
        <para xml:id="para_e8d3463f-f154-422f-b3d9-5be58bda0d24">Business Names are not specified for elements that are expected to receive an automatically generated value, e.g., the id element for each document, section, and entry.</para>
        <para xml:id="para_1edf3f0b-d2ce-47c5-9732-9e3bc7118f59">As a convention, Business Names are represented in CamelCase (alternating upper and lower case, no spaces, initial letter in upper case) in the Business Name column of the template tables.</para>
        <para xml:id="para_903aab1c-be1c-4d9f-8051-7c072439f682">Business Names are hierarchically organized, and contextually scoped by higher level Business Names.</para>
        <itemizedlist>
          <listitem>
<para xml:id="para_9548184c-2d12-4594-9ae3-48eb45bba658">Data element/attribute level Business Names are shown in normal font</para>
</listitem>
          <listitem>
<para xml:id="para_5239784b-ca8f-433b-900c-30b31956ae8b">Business Names that provide scoping for subsidiary Business Names are shown in bold font.</para>
</listitem>
          <listitem>
<para xml:id="para_fdd509ea-419e-4619-a505-65ea0a5a6df2">Referenced Business Names from included templates are shown in italic (see <xref linkend="sect_5.2.9" xrefstyle="select: label"/>)</para>
</listitem>
          <listitem>
<para xml:id="para_d5bf2bb9-e10b-4c98-94c5-7c4622307fd1">As a convention, hierarchical relationship between Business Names is shown with the : character.</para>
</listitem>
        </itemizedlist>
        <para xml:id="para_718415bb-1965-410f-a8fb-ca8d10179198">Scoping Business Names scope all attributes and elements subsidiary to the element to which the name is assigned.</para>
        <para xml:id="para_58c8dcfd-8dbf-4100-97c8-1ed74d4cc6e3">Examples:</para>
        <itemizedlist>
          <listitem>
<para xml:id="para_8b4cd881-8112-4e57-ac7c-f75b2728abdb">The top level scoping Business Name for an Imaging Report is "ImagingReport", and it scopes all attributes and elements in the document, i.e., including and subsidiary to the &lt;ClinicalDocument&gt; XML element</para>
</listitem>
          <listitem>
<para xml:id="para_8fee17be-d557-4b7d-abd4-a78d411f58bc">The Business Name for the <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> report section is "ImagingReport:ClinicalInformation", and it scopes all attributes and elements including and subsidiary to the &lt;section&gt; XML element in template 1.2.840.10008.9.2</para>
</listitem>
          <listitem>
<para xml:id="para_52afc16f-7c83-4e19-a4bb-c51268706004">The Business Name for the text element of the <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> report section is "ImagingReport:ClinicalInformation:Text"</para>
</listitem>
          <listitem>
<para xml:id="para_09eeb719-ec4c-4baf-a5d0-6943be07711d">The Business Name for the text element of the <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> section is "ImagingReport:Impression:Text"</para>
</listitem>
        </itemizedlist>
        <para xml:id="para_b8612e21-4221-4aea-9a7a-8721583da5ad">Note that both <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> and <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> define a Business Name "Text", but these are distinguished by their hierarchical location under the scoping Business Names of their respective sections.</para>
        <section label="5.2.1.1" status="4" xml:id="sect_5.2.1.1">
          <title>Multiple Instantiations</title>
          <para xml:id="para_a29cc6ab-e408-4dce-b18c-f1d426427ce7">Some templates or elements may be invoked multiple times in a document instance; for example, the <xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> template is instantiated for each numeric measurement in a report.
              Each instantiation may have an identifying string, unique within the document, used as a discriminator between those multiple instantiations.
              The Business Name for each element that may have multiple instantiations has a suffix [*], indicating the use of a discriminator string.
              This allows Business Name based production logic for authoring applications to identify specific instances of an element.</para>
            <note>
                <para xml:id="para_b1aba374-8839-4edf-930c-ab15779bebc7">Production logic based on Business Names should allow single instantiations of elements without a discriminator; for instance,
                    <xref linkend="sect_8.1" xrefstyle="select: label title"/>
                    specifies Patient[*], allowing multiple patients to be recorded in special cases,
                    although the vast majority of reports will be for a single patient for which a discriminator is unnecessary.</para>
            </note>
          <example label="5.2.1.1-1" xml:id="example_5.2.1.1-1">
            <title>Example Business Name based production logic with discriminators for two measurements</title>
            <programlisting><![CDATA[-- "Q21a" is the discriminator for the first measurement
-- "Q21b" is the discriminator for the second measurement
ImagingReport:Findings:QuantityMeasurement[Q21a]:MeasurementName = ("112058", "DCM", "Calcium score")
ImagingReport:Findings:QuantityMeasurement[Q21a]:MeasurementValue = "8"
ImagingReport:Findings:QuantityMeasurement[Q21a]:MeasurementUnits = "[arb'U]"
ImagingReport:Findings:QuantityMeasurement[Q21b]:MeasurementName = ("408716009", "SNOMED", "Stenotic lesion length")
ImagingReport:Findings:QuantityMeasurement[Q21b]:MeasurementValue = "14"
ImagingReport:Findings:QuantityMeasurement[Q21b]:MeasurementUnits = "mm"]]></programlisting>
      </example>
          <para xml:id="para_96c0bc9c-0b8b-4216-bbd5-48b72a646ee4">The discriminator string shall be conformant to XML Name production requirements, as used for the XML ID attribute (see <xref linkend="sect_5.3.4" xrefstyle="select: label"/> on the use of XML ID).</para>
          <para xml:id="para_cf941fb5-d625-40ff-ba87-e67cf13892ed">Some CDA elements may include an XML ID attribute. This attribute is identified by '*' (asterisk) as its Business Name, and its value shall be the discriminator string.</para>
        </section>
        <section label="5.2.1.2" status="4" xml:id="sect_5.2.1.2">
          <title>Implicit Element Structure For Business Name</title>
          <para xml:id="para_52c256de-c3a0-465d-a6e1-580dab1f8755">A Business Name may be associated with an element subsidiary to another element that does not have an associated Business Name. In such a case, when the element with the Business Name is instantiated in a document, its entire parent element hierarchy must be instantiated, even if those elements are identified as optional.</para>
          <note>
            <para xml:id="para_dd4cdb9b-27a8-4ab6-b97f-14257ed27e42">For example, in the <xref linkend="sect_8.1" xrefstyle="select: labelnumber title"/> template, if Recipient:Name is instantiated, the entire hierarchical structure of informationRecipient/intendedRecipient/informationRecipient/name must be instantiated to hold the name element content.</para>
          </note>
        </section>
      </section>
      <section label="5.2.2" status="3" xml:id="sect_5.2.2">
        <title>Nesting Level</title>
        <para xml:id="para_eae255da-66a9-419e-b2f4-d97471295a01">CDA documents are encoded using the Extensible Markup Language (XML), and are marked up through hierarchically nested XML elements (tags). The Nesting Level column of the template tables identifies the hierarchical level of each element relative to the other elements in the table using the character right angle bracket '&gt;'. Multiple levels of nesting are identified by multiple &gt; characters.</para>
        <para xml:id="para_7e850889-220e-4784-ba43-74855446f24f">XML elements may have attributes, encoded as "&lt;name&gt;=&lt;value&gt;" pairs within the element tag. Such attributes are identified using the character at sign '@'.</para>
      </section>
      <section label="5.2.3" status="3" xml:id="sect_5.2.3">
        <title>Element/Attribute Names and XPath Notation</title>
        <para xml:id="para_4efc0c46-c855-4713-bdea-8183060992e3">The name of each XML element and attribute used in a CDA document for which specific constraints are applied is shown in the Element/Attribute column of the template tables. Optional elements whose use is not specified nor constrained are not shown.</para>
        <para xml:id="para_1b849f81-e1d1-4d8c-8e56-5b99d39eb10e">Elements and attributes that use the default value specified in CDA Specification are not shown. For example, the Section element has default attributes classCode='DOCSECT' and moodCode='EVN'; these are not listed in the templates. In accordance with the HL7 v3 specification, attributes with default values need not be included in instances, and their absence implies the default value.</para>
        <para xml:id="para_828a3b79-8a62-4689-99b8-01cd777716d5">XML Path Language (XPath) notation is used to identify the XML elements and attributes within the CDA document instance to which various constraints are applied. The implicit context of these expressions is the root node of the document, and traversing the path to the root node of the relevant template. This notation provides a mechanism that will be familiar to developers for identifying parts of an XML document.</para>
        <para xml:id="para_77299804-26bc-49e8-95c2-22cb5398afb6">XPath syntax selects nodes from an XML document using a path containing the context of the node(s). The path is constructed from node names and attribute names (prefixed by a '@') and concatenated with a '/' symbol.</para>
        <para xml:id="para_6d0822b4-4560-476f-8d93-a714c9789f9e">
<xref linkend="example_5.2.3-1" xrefstyle="select: label"/> is an example of a fragment of a CDA document.</para>
        <example label="5.2.3-1" xml:id="example_5.2.3-1">
          <title>XML document example</title>
          <programlisting><![CDATA[<author>
    <assignedAuthor>
        ...
        <code codeSystem='2.16.840.1.113883.6.96' codeSystemName='SNOMED CT'
                code='17561000' displayName='Cardiologist' />
            ...
        </assignedAuthor>
    </author>]]></programlisting>
      </example>
        <para xml:id="para_c50a2da6-ff0e-49eb-a9bb-49a9ecdcc1d8">
<xref linkend="table_5.2.3-1" xrefstyle="select: label"/> is an example of a fragment of a template specification table.</para>
        <table frame="box" label="5.2.3-1" rules="all" xml:id="table_5.2.3-1">
          <caption>Template element and attribute example</caption>
          <thead>
            <tr valign="top">
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_8e167a47-4fa4-4e99-888b-b7c42396d939">…</para>
              </th>
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5356674-5ac2-4875-92ac-40205b6867a4">Nest Level</para>
              </th>
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_d809076d-8b61-4853-a7b5-c1e6dcd5f772">Element/​Attribute</para>
              </th>
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a84685e-21cb-49a6-bedf-ef68019b26af">…</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_acf0a479-cc9c-435b-9eda-9124228017e5">author</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1e8123e4-cd6e-4968-b784-0d14360b16e8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7fde3840-85e9-46c6-9b88-7254e81ce6e4">assigned​Author</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc266ffa-3ead-4dfc-85bd-38278b8d2bca">…</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f6e82187-b714-4d7f-8fb5-f03548edb976">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e71e86ed-14f5-4efa-bd6d-5b8f20c9a570">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d404a15-77e7-4c66-bf71-15a768c5c11b">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_24ca564f-ce50-4ff4-b090-3e33c4d15841">@codeSystem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bb5f2c59-4572-43ef-a624-3e2091f8183c">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db0354a3-ef2a-4c28-97b3-bf3d5ddab76e">@codeSystemName</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc8d12b3-461f-4e21-be5d-0d159edf2822">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f679d158-cde4-4aba-a686-35ed3f120a72">@code</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09b9be92-f671-4349-b8fc-02ac2b15acfb">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e55e049e-2d80-474f-a14b-69d6f262bb95">@displayName</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7fa4b0aa-dd40-4112-82fd-4fecb48d98af">…</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </table>
        <para xml:id="para_3ca6763b-be63-4e75-858d-e7744e0d4757">In <xref linkend="table_5.2.3-1" xrefstyle="select: label"/>, the code attribute of the code element could be selected with the XPath expressionauthor/assignedAuthor/code/@code.</para>
      </section>
      <section label="5.2.4" status="3" xml:id="sect_5.2.4">
        <title>Cardinality</title>
        <para xml:id="para_84516985-89e5-4711-9752-28c7b90abf5a">Each element/attribute has a cardinality indicator that specifies the number of allowable occurrences within a template instance. The cardinality indicators are interpreted with the following format "m…n" where m represents the least and n the most:</para>
        <itemizedlist>
          <listitem>
<para xml:id="para_3c254729-51a2-42ca-a587-50be843d2069">0..1 zero or one</para>
</listitem>
          <listitem>
<para xml:id="para_9a4fe460-92ab-45ad-9b00-b03ec05c410a">1..1 exactly one</para>
</listitem>
          <listitem>
<para xml:id="para_992648be-9566-42db-9928-a46fbf022f7a">1..* at least one</para>
</listitem>
          <listitem>
<para xml:id="para_c44c8bdd-d9b1-42cd-b049-167676b7d793">0..* zero or more</para>
</listitem>
          <listitem>
<para xml:id="para_917d255e-2d6c-48a5-849c-e2fee856d7d4">1..n at least one and not more than n</para>
</listitem>
          <listitem>
<para xml:id="para_c93b0f57-d582-46ee-aba1-17cdeef40b1f">0..0 none [SHALL NOT]</para>
</listitem>
        </itemizedlist>
      </section>
      <section label="5.2.5" status="3" xml:id="sect_5.2.5">
        <title>Element/Attribute Conformance</title>
        <para xml:id="para_d0041cb0-f8f9-44fa-bb12-6ce7acd67a33">Each element/attribute has a conformance verb (keyword) in addition to the cardinality constraint.</para>
        <para xml:id="para_2a95f125-47cc-4b55-af2b-41c3d79db208">The keywords SHALL, SHOULD, MAY, SHOULD NOT, SHALL NOT, and NEED NOT in this document are to be interpreted as described in ISO/IEC Directives, Part 2, Annex H "Verbal forms for the expression of provisions":</para>
        <itemizedlist>
          <listitem>
<para xml:id="para_37f1df90-26e8-4da0-84a2-41173d520c1a">SHALL: an absolute requirement</para>
</listitem>
          <listitem>
<para xml:id="para_408d6c78-fa72-438f-ac5e-fc5c4b72651f">SHALL NOT: an absolute prohibition against inclusion</para>
</listitem>
          <listitem>
<para xml:id="para_5790196a-5c4f-40d7-b0f4-61e0746e982a">SHOULD/SHOULD NOT: a best practice or recommendation. There may be valid reasons to ignore a recommendation, but the full implications must be understood and carefully weighed before choosing to not adhere to the best practice.</para>
</listitem>
          <listitem>
<para xml:id="para_c5df8747-c476-4c62-8808-4004378ec918">MAY/NEED NOT: truly optional; can be included or omitted at the discretion of the content creator with no conformance implications</para>
</listitem>
        </itemizedlist>
        <para xml:id="para_00f62774-e55c-45dc-bd01-992ff31d651b">The keyword SHALL is associated with a minimum cardinality of at least 1; other keywords have a minimum cardinality of 0. If an element is required by SHALL, but is not known (and would otherwise be omitted without the SHALL requirement), it must be represented by a nullFlavor. SHALL allows the use of nullFlavor unless the requirement is on an attribute (nullFlavor does not apply to attributes), or the use of nullFlavor is explicitly precluded (see <xref linkend="sect_5.2.7" xrefstyle="select:label title"/> and <xref linkend="sect_5.3.2" xrefstyle="select:label title"/>).</para>
        <para xml:id="para_9f1f8d6d-e7db-4dbb-8a36-0d6320fbc996">Within the template, the keyword COND (conditional) may be present. In this case, the specification of the condition, and the conformance verbs associated with the condition being true or false, are described below the table in a paragraph flagged with the COND keyword.</para>
        <para xml:id="para_c5bbd6ae-4c9c-4cb9-85d3-7a3459f7a267">In an open template, additional elements and attributes allowed by the CDA Specification are not precluded by template constraints, unless there are applicable SHALL NOT template specifications.</para>
      </section>
      <section label="5.2.6" status="3" xml:id="sect_5.2.6">
        <title>Data Type</title>
        <para xml:id="para_4cc6f351-3ce2-4964-80a4-1413129861dd">The data type associated with each element/attribute is specified, as described in the CDA Specification and its referenced HL7v3 Data Types Release 1. Elements that are simply tags with subsidiary content only of nested elements, e.g., RIM class clone names, have the Data Type column empty.</para>
        <para xml:id="para_07599aed-fc8c-4371-b5a5-3964224654ba">Many data types are non-primitive, and may include constituent component elements and/or attributes. Such subsidiary components are not specified in the templates unless specific constraints are to be applied to them.</para>
      </section>
      <section label="5.2.7" status="3" xml:id="sect_5.2.7">
        <title>Value Conformance</title>
        <para xml:id="para_84e3ff3f-0d5e-46e1-9419-65deb17ee4f4">The template table may constrain values or vocabularies to be used with an element or attribute. Value constraints include a conformance verb (SHALL, SHOULD, MAY, etc.) as defined in <xref linkend="sect_5.2.5" xrefstyle="select: label"/>, and specified in the Value Conformance column of the template tables.</para>
        <para xml:id="para_cb011d3f-04f4-4510-912f-42a4c2c878a3">Elements for which nullFlavor is forbidden are indicated with an additional constraint keyword noNull.
        </para>
        <para xml:id="para_026bb572-0177-428e-89c4-d7e93e8c4cc6">Additionally, constraints specifying Value Sets include a coding strength conformance CWE (Coded With Extensibility) or CNE (Coded with No Extensibility), as defined in Core Principles and Properties of HL7 Version 3 Models, Release 1.</para>
        <para xml:id="para_a0334092-245a-4215-ae5b-3fda765ced45">Further, Value Set constraints can be static, meaning that they are bound to a specified version of a Value Set, or dynamic, meaning that they are bound to the most current version of the Value Set. By default, Value Sets have dynamic binding, unless explicitly specified with an additional constraint keyword static.</para>
      </section>
      <section label="5.2.8" status="3" xml:id="sect_5.2.8">
        <title>Value Specification</title>
        <para xml:id="para_8d8777c3-b4f8-488b-a81e-64b14b7b384e">The template table may constrain values to a single fixed value, to a Value Set from which the value is to be drawn, or to a named Concept Domain. It may non-normatively reference a mapping from a DICOM SOP Instance or an HL7 message.</para>
        <para xml:id="para_5af74239-44ea-4218-a149-e1a9c0b28ff0">The template table may contain elements without a value specification, and without a Business Name. These are typically id elements. The application creating the document instance shall fill these elements with appropriate values.</para>
        <section label="5.2.8.1" status="4" xml:id="sect_5.2.8.1">
          <title>Coded Simple Value</title>
          <para xml:id="para_3b0bbeb6-5b66-4a50-99f3-1b5b495df7c9">Values of Data Type CS (Coded Simple Value) have a fixed code system defined in the CDA Specification, and are simple strings. The template tables identify only the constraint on the code value, and do not specify the fixed code system nor the code meaning (display name), which are not encoded in the CDA instance.</para>
        </section>
        <section label="5.2.8.2" status="4" xml:id="sect_5.2.8.2">
          <title>Concept Descriptor and Coded With Equivalents</title>
          <para xml:id="para_c00b4141-00df-4163-bb1d-84482e4cc3d7">Single values of Data Type CD (Concept Descriptor) or CE (Coded With Equivalents) are specified in the template tables with the triplet notation specified in <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>:</para>
          <itemizedlist mark="none">
            <listitem>
              <para xml:id="para_0b2fe95c-7cc2-406f-ad7c-f62a111fbb6a">(CodeValue, Coding Scheme Designator, "Code Meaning")</para>
            </listitem>
          </itemizedlist>
          <para xml:id="para_8519da87-6837-442f-a48f-fc105af0d95b">The Coding Scheme Designator is a simple human readable identifier of the code system, and corresponds to the optional codeSystemName attribute of the CD or CE element. The CDA Specification requires the Code System OID to be encoded in the codeSystem attribute of the CD or CE element. The corresponding OID for each Coding Scheme Designator is provided in <olink targetdoc="PS3.16" targetptr="chapter_8" xrefstyle="template:Annex %n “%t” in PS3.16"/>. The Code Meaning is encoded in the displayName attribute of the CD or CE element.</para>
        </section>
        <section label="5.2.8.3" status="4" xml:id="sect_5.2.8.3">
          <title>Value Set</title>
          <para xml:id="para_d24af0a6-7e1c-4d16-bb33-441612baf2be">Elements whose value may be drawn from a Value Set will have that Value Set identified in the Value column introduced by the keyword ValueSet.</para>
        </section>
        <section label="5.2.8.4" status="4" xml:id="sect_5.2.8.4">
          <title>Concept Domains</title>
          <para xml:id="para_c454358d-1389-4daa-b50e-b0f4c915aa82">Concept Domains (see definition in <xref linkend="sect_3.2" xrefstyle="template:Section %n"/>) are used to provide a named category in a structural template that can be bound to a specific value or value set by an invoking template, thus specializing the structural template for a particular use case. Concept Domain names are introduced by the keyword
            ConceptDomain in the Value column. For example, the <xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> template Concept Domain "observationType" might be bound to a value set of fetal ultrasound measurements in one invoking template, or to a value set of cardiac CT measurements in another invoking template.</para>
          <para xml:id="para_03f039d4-719d-4e00-8996-6cf36a651353">Concept Domain names are similar to element Business Names in that they provide a public interface that is bound to specific values later in the document specification and production process. Concept Domains do not have a Value Conformance verb; the conformance verb is specified when the Concept Domain is bound to a specific value or value set (see <xref linkend="sect_5.2.9.1" xrefstyle="select: label"/>).</para>
        </section>
        <section label="5.2.8.5" status="4" xml:id="sect_5.2.8.5">
          <title>Mapping From DICOM SOP Instances and HL7v2 Messages</title>
          <para xml:id="para_dfffbf87-1493-411a-843f-8785a350c478">Elements whose value may be mapped from a DICOM SOP Instance or from an HL7v2 message have the source attribute name and tag identified in the Value column in italic font. Note that many of these values have their origin in IT systems outside the imaging department, and there may be alternate routes for these values to be accessed by the reporting application, e.g., from an HL7 FHIR web service.</para>
          <note>
            <para xml:id="para_3e2c3585-b253-4b8c-820e-092db3a47bf2">Due to differences in use of HL7v2 data elements, mappings should not be considered normative.</para>
          </note>
          <para xml:id="para_00657163-df3d-4de1-89bb-62416d0bb994">Data mapped from a specific Attribute in the interpreted DICOM image(s) is identified by the Attribute Name and Tag, represented in the mapping as:</para>
          <itemizedlist mark="none">
            <listitem>
              <para xml:id="para_be2adae9-357b-4ff2-8bb6-ffb1b0181404">
                <emphasis role="italic">Attribute Name (gggg,eeee)</emphasis>
              </para>
            </listitem>
          </itemizedlist>
          <para xml:id="para_bb265a65-97ed-42d2-9c65-b68f72db5243">Data mapped from Attributes within sequences is identified with the &gt; character:</para>
          <itemizedlist mark="none">
            <listitem>
              <para xml:id="para_a1d27c2a-fd24-4543-b04a-e738a0e6bda6">
                <emphasis role="italic">Sequence Attribute Name (ggg0,eee0) &gt; Item Attribute Name (ggg1,eee1)</emphasis>
              </para>
            </listitem>
          </itemizedlist>
          <para xml:id="para_5d205270-d5ad-4ae0-9b43-61b888504229">Data mapped from an HL7v2 field in the order for the study is identified by the Element Name and Segment Field identifier:</para>
          <itemizedlist mark="none">
            <listitem>
              <para xml:id="para_4d4bf67f-9fc1-4bfe-a06e-a234bc6fdb34">
                <emphasis role="italic">Element Name seg-n</emphasis>
              </para>
            </listitem>
          </itemizedlist>
          <para xml:id="para_fddc932a-7d0a-4527-912d-ec62304aa3b9">The mapping of the value typically requires a transformation from the DICOM Value Representation or the HL7v2 Data Type representation to the CDA Data Type encoding. For example, transforming a DICOM Code Sequence attribute or an HL7v2 CWE field to a CD or CE Data Type requires a look up of the Coding Scheme OID.</para>
        </section>
      </section>
      <section label="5.2.9" status="3" xml:id="sect_5.2.9">
        <title>Subsidiary Templates</title>
        <para xml:id="para_86f37200-4d15-4161-85ea-6f076183fc0c">A template may include subsidiary templates. Templates typically have one of two styles, a single parent element with child element structure, or a flat list of sibling elements.</para>
        <para xml:id="para_ff4def8a-33cb-45d1-a157-ffee92ae942f">The single parent element style is typical for the top level Document, Section, and Entry templates, and the parent element is of the HL7 v3 RIM act class. Inclusion of such a template therefore involves an actRelationship element; that actRelationship element is specified in the invoking template.</para>
        <para xml:id="para_29c3aa7e-3b6d-42cc-8841-04d2c988e2a1">The sibling elements style is typical for sets of elements and attributes aggregated for editorial convenience.</para>
        <para xml:id="para_de711bd8-7973-43aa-a1c3-c56d3c5b662a">Inclusion of a subsidiary template includes the name of included template and its templateID, specified in the Subsidiary Template column of the invoking template table.</para>
        <para xml:id="para_e5ae8243-7498-4330-9866-d8be80ff074d">For an included template of the single parent element style, the scoping business name and top level element are provided in italics in the invoking template table. This indicates this is data copied from the specification in the included template for the reader's convenience.</para>
        <section label="5.2.9.1" status="4" xml:id="sect_5.2.9.1">
          <title>Vocabulary Binding and Constraints</title>
          <para xml:id="para_ba2c59c9-2706-44a3-a389-b69daba36368">A template inclusion may provide Concept Domain Vocabulary Binding or other vocabulary constraints, e.g., limiting an element in the included template to a specific value from its defined Value Set. These vocabulary constraints are specified in tabular form, as shown in <xref linkend="table_5.2.9.1-1" xrefstyle="select: label"/>. The table is included in the additional requirements for the template, with a reference in the Value column of the template entry invoking the subsidiary template. The Value Conformance and Value specification columns are interpreted as in the templates tables.</para>
          <table frame="box" label="5.2.9.1-1" rules="all" xml:id="table_5.2.9.1-1">
            <caption>Vocabulary Binding Table Format</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2efaa67f-7d54-45c3-9643-adf1ab8ee2a9">Concept Domain or Element</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c8529d5e-5155-48de-825e-9a35e91950bf">Value Conf</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d557ce1f-470c-4e02-9e9e-305d075e3d19">Value</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8633e709-fe07-47fb-9dd3-5a5443b16f0b">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d4b0bed9-258c-48bd-9b49-8ad578bacf6f">...</para>
                </td>
                <td align="center" colspan="1" rowspan="1">
                  <para xml:id="para_02acce46-94bb-4ce5-abc6-e2d00f371872">...</para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      </section>
      <section label="5.2.10" status="3" xml:id="sect_5.2.10">
        <title>Additional Requirements</title>
        <para xml:id="para_a0395055-35d1-46a7-8109-8272a6e82548">Each template may be accompanied by additional requirements and usage explanations in narrative specification language.</para>
      </section>
    </section>
    <section label="5.3" status="2" xml:id="sect_5.3">
      <title>Encoding</title>
      <para xml:id="para_5e739151-4685-4491-b2ef-aa27c00645b0">A full discussion of the representation of data types and vocabulary is outside the scope of this document; for more information, see the HL7 V3 specifications on Abstract Data Types R1 and XML Data Types R1 referenced in the CDA Specification.</para>
      <note>
        <orderedlist>
          <listitem>
            <para xml:id="para_52b7cb9e-8914-40db-9f53-75dba7900981">Many Data Types encode their values in attributes, rather than character data. For example, the URL Data Type encodes its value in the
              <emphasis role="bold">value</emphasis>attribute within the element tag, e.g., &lt;reference value="http://xyz.org"&gt;. Within this specification, the attribute(s) that hold the value are not identified, except where specific constraints apply.</para>
          </listitem>
          <listitem>
            <para xml:id="para_4d7bf14a-32f3-4c17-95e9-72d90f27b1dd">The Consolidated CDA specification includes extensive examples of valid and invalid encodings, which may be useful for implementers.</para>
          </listitem>
          <listitem>
            <para xml:id="para_e763a4d9-2764-4aa2-8c5d-29b247169e62">The specification of a representation of Data Types for use in Business Name-based report production logic is outside the scope of this Standard.</para>
          </listitem>
        </orderedlist>
      </note>
      <section label="5.3.1" status="3" xml:id="sect_5.3.1">
        <title>Translation Code Element</title>
        <para xml:id="para_39e98445-4f1b-4f26-8bbb-0afc7dbc29f5">HL7 Data Types CD (Concept Descriptor) and CE (Coded With Equivalents) allow a translation code element, which allows the encoding of the same concept in an alternate coding system. This supports the encoding of both an originally entered (local) code, and a code specified for cross-system interoperability.</para>
        <para xml:id="para_a3ec0d78-0560-4fd5-8e69-a598d1792d88">This Part follows the convention used in the Consolidated CDA Implementation Guide specification, which specifies the standard interoperable code in the root, whether it is original or a translation. The HL7v3 Data Types R1 standard included by CDA formally specifies the original code (as initially entered in an information system application) to be placed in the root.</para>
        <note>
          <para xml:id="para_367809c4-d4af-4958-abca-69786d3b2709">This discrepancy is resolved in HL7v3 Data Types R2 to follow the convention used here, and the HL7 Structured Documents Working Group has approved the "pre-adoption" of the Data Types R2 approach in CDA implementations.</para>
        </note>
        <example label="5.3.1-1" xml:id="example_5.3.1-1">
          <title>Translation code example</title>
          <programlisting><![CDATA[<code code='206525008'
      displayName='neonatal necrotizing enterocolitis 'codeSystem='2.16.840.1.113883.6.96'
      codeSystemName='SNOMED CT'>
    <translation code='NEC-1'
      displayName='necrotizing enterocolitis'
      codeSystem='2.16.840.1.113883.19'/>
</code>]]></programlisting>
        </example>
      </section>
      <section label="5.3.2" status="3" xml:id="sect_5.3.2">
        <title>Null Flavor</title>
        <para xml:id="para_3c8b0a1d-5c9f-439e-8f89-df173e8cca7f">Information technology solutions store and manage data, but sometimes data are not available: an item may be unknown, not relevant, or not computable or measurable. In HL7 v3, a
          flavor of null, or nullFlavor, describes the reason for missing data.</para>
        <para xml:id="para_8fd8d3bc-d063-4c6b-8b28-0b6c94c60294">For example, if a patient arrives at an Emergency Department unconscious and with no identification, a null flavor is used to represent the lack of information. The patient's birth date could be represented with a null flavor of "NAV", which is the code for "temporarily unavailable". When the patient regains consciousness or a relative arrives, we expect to be able to obtain the patient's birth date.</para>
        <example label="5.3.2-1" xml:id="example_5.3.2-1">
          <title>nullFlavor example</title>
          <programlisting><![CDATA[<birthTime nullFlavor="NAV"/> <!--coding an unknown birthdate-->]]></programlisting>
        </example>
        <para xml:id="para_a692407d-41bf-4979-ad01-f538e67283f1">Use null flavors for unknown, required, or optional attributes:</para>
        <variablelist>
          <varlistentry>
            <term>NI</term>
            <listitem>
              <para xml:id="para_75c5c081-4075-4b74-8e52-fcd724314f33">No information. This is the most general and default null flavor.</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>NA</term>
            <listitem>
              <para xml:id="para_8afd55ae-59db-4c75-8138-b930bdcdaaac">Not applicable. Known to have no proper value (e.g., last menstrual period for a male).</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>UNK</term>
            <listitem>
              <para xml:id="para_59713173-4084-4fa8-957e-9c53bf0bf056">Unknown. A proper value is applicable, but is not known.</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>ASKU</term>
            <listitem>
              <para xml:id="para_a3792589-5ffe-49e8-b7e6-6b07ba14d4e1">Asked, but not known. Information was sought, but not found (e.g., the patient was asked but did not know).</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>NAV</term>
            <listitem>
              <para xml:id="para_0d01316f-ff12-4472-a230-b36b73c0db48">Temporarily unavailable. The information is not available, but is expected to be available later.</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>NASK</term>
            <listitem>
              <para xml:id="para_94339a0e-85cc-46be-8956-7a0541a35097">Not asked. The patient was not asked.</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>MSK</term>
            <listitem>
              <para xml:id="para_29e7206c-a942-4a2e-b06d-6f6f410423ec">Masked. There is information on this item available but it has not been provided by the sender due to security, privacy, or other reasons. There may be an alternate mechanism for gaining access to this information.</para>
            </listitem>
          </varlistentry>
          <varlistentry>
            <term>OTH</term>
            <listitem>
              <para xml:id="para_b112ef08-4d78-4915-bfaf-67ea399269be">Other. The actual value is not an element in the value domain of a variable. (e.g., concept not provided by required code system).</para>
            </listitem>
          </varlistentry>
        </variablelist>
        <para xml:id="para_5598aaf7-5f3a-4498-8ebc-e3d9ca79366b">The above null flavors are commonly used in clinical documents. For the full list and descriptions, see the nullFlavor vocabulary domain in the HL7 v3 Vocabulary referenced by the CDA specification.</para>
        <para xml:id="para_5104a8ca-1a58-4a79-886e-6a8c25e62927">Any SHALL conformance requirement on an element may use nullFlavor, unless nullFlavor is explicitly disallowed (as indicated by
          noNull, see <xref linkend="sect_5.2.7" xrefstyle="select: label title"/>). SHOULD and MAY conformance requirements may also use nullFlavor. nullFlavor does not apply to conformance requirements on attributes.</para>
        <para xml:id="para_ebd734f3-7d7c-4b8e-bcda-39b66d2f7651">The encoding of nullFlavor as an attribute of the data type element is not shown in the templates, hence there is no business name associated with the attribute.</para>
        <note>
          <para xml:id="para_672db73f-925b-4702-9dda-f07e84ad19d7">Production logic based on Business Names needs to provide a mechanism for assignment of a value to the nullFlavor attribute as an alternative for a value for the element. Specification of such production logic is outside the scope of this Standard.</para>
        </note>
        <example label="5.3.2-2" xml:id="example_5.3.2-2">
          <title>XML example of allowed nullFlavors when element is required</title>
          <programlisting><![CDATA[<entry>
    <observation classCode="OBS" moodCode="EVN">
        <id nullFlavor="NI"/>
        <code nullFlavor="OTH">
            <originalText>New Grading system</originalText>
        </code>
        <statusCode code="completed"/>
        <effectiveTime nullFlavor="UNK"/>
        <value xsi:type="CD" nullFlavor="NAV">
            <originalText>Spiculated mass grade 5</originalText>
        </value>
    </observation>
</entry>]]></programlisting>
        </example>
      </section>
      <section label="5.3.3" status="3" xml:id="sect_5.3.3">
        <title>Unknown Information</title>
        <para xml:id="para_9cff23e9-6e8e-4614-8a56-b573440255ea">If a document creator wants to state that a piece of information is unknown, the following principles apply:</para>
        <orderedlist>
          <listitem>
            <para xml:id="para_675e7ed0-34c5-4c0f-89e1-8e1affc285de">If the creator doesn't know an attribute of an act, that attribute can be null.</para>
            <example label="5.3.3-1" xml:id="example_5.3.3-1">
              <title>Unknown medication example</title>
              <programlisting><![CDATA[<text>patient was given a medication but I do not know what it was</text>
<entry>
    <substanceAdministration moodCode="EVN" classCode="SBADM">
        <consumable>
            <manufacturedProduct>
                <manufacturedLabeledDrug>
                    <code nullFlavor="NI"/>
                </manufacturedLabeledDrug>
            </manufacturedProduct>
        </consumable>
    </substanceAdministration>
</entry>]]></programlisting>
            </example>
          </listitem>
          <listitem>
            <para xml:id="para_76d28b6b-bbec-4c46-ac51-54b07e47ab57">If the creator doesn't know if an act occurred, the nullFlavor is on the act (detail could include specific allergy, drug, etc.).</para>
            <example label="5.3.3-2" xml:id="example_5.3.3-2">
              <title>Unknown medication use of anticoagulant drug example</title>
              <programlisting><![CDATA[<text>I do not know whether or not patient received an anticoagulant drug</text>
<entry></para>
    <substanceAdministration moodCode="EVN" classCode="SBADM" nullFlavor="NI">
        <consumable>
            <manufacturedProduct>
                <manufacturedLabeledDrug>
                    <code code="81839001" displayName="anticoagulant drug"
                        codeSystem="2.16.840.1.113883.6.96"
                        codeSystemName="SNOMED CT"/>
                </manufacturedLabeledDrug>
            </manufacturedProduct>
        </consumable>
    </substanceAdministration>
</entry>]]></programlisting>
            </example>
          </listitem>
          <listitem>
            <para xml:id="para_aae70fb4-93ac-42c4-86bd-8bbe65c386e4">If the sender wants to state 'no known', a negationInd can be used on the corresponding act (substanceAdministration, Procedure, etc.)</para>
            <example label="5.3.3-3" xml:id="example_5.3.3-3">
              <title>No known medications example</title>
              <programlisting><![CDATA[<text>No known medications</text>
<entry>
    <substanceAdministration moodCode="EVN" classCode="SBADM" negationInd="true">
        <consumable>
            <manufacturedProduct>
                <manufacturedLabeledDrug>
                    <code code="410942007" displayName="drug or medication"
                        codeSystem="2.16.840.1.113883.6.96"
                        codeSystemName="SNOMED CT"/>
                </manufacturedLabeledDrug>
            </manufacturedProduct>
        </consumable>
    </substanceAdministration>
</entry>]]></programlisting>
            </example>
          </listitem>
        </orderedlist>
        <para xml:id="para_8a66d697-4679-41f6-9944-b8e8a67a1be7">Other CDA implementation guides recommended using specific codes to assert no known content, for example SNOMED CT 160244002 "No known allergies" or 160245001 "No current problems or disability". Specific codes are still allowed; however, use of negationInd is an alternative, and the specific approach for each use will be specified in the associated template.</para>
      </section>
      <section label="5.3.4" status="3" xml:id="sect_5.3.4">
        <title>XML ID</title>
        <para xml:id="para_2e2054e1-a12e-4a7d-b901-d3b5d88803e2">The XML Specification allows a markup tag to have an attribute of type ID, whose value is unique within the document, that allows reference to that markup. The CDA schema defines such attributes with attribute name ID.</para>
        <note>
          <orderedlist>
            <listitem>
              <para xml:id="para_eca4ec3e-8dd0-4d9e-abe2-030d074a0d6a">Thus the attribute named ID is of XML attribute type ID. This must further be distinguished from the element named id of HL7v3 Data Type UID that is part of most RIM classes. The attribute name is always upper case, the element name is always lower case.</para>
            </listitem>
            <listitem>
              <para xml:id="para_791c66a3-d317-493a-b1e5-b2306467b675">The actual CDA schema specification uses the XML Schema Datatypes definition of XML ID (xs:ID). Readers may also be familiar with the xml:id specification, which is not formally used by CDA as it was published after the CDA specification.</para>
            </listitem>
          </orderedlist>
        </note>
        <para xml:id="para_94d6bc89-33b2-40d4-9449-02a7f9558472">In the CDA R2 Specification, the XML ID attribute capability is applied to the Section and observationMedia elements, and to various types of narrative block markup, and is used to provide linkage between structured entries and the corresponding narrative text (see <xref linkend="sect_9.1.1" xrefstyle="select: label title"/>).</para>
      </section>
    </section>
    <section label="5.4" status="2" xml:id="sect_5.4">
      <title>Extension and Namespace</title>
      <para xml:id="para_c9150d75-5360-4177-9323-df92154429d0">In accordance with CDA R2 (and HL7 v3 XML) extensibility rules, as described in CDA R2 Section 1.4, "locally-defined" XML markup may be specified where there is a need to communicate information for which there is no suitable representation in CDA R2. These extensions are described in the context of the templates where they are used. All such extensions use HL7 v3 Data Types used by CDA R2.</para>
      <note>
        <para xml:id="para_eb156fc4-ea7a-40ae-9fed-a7ae1f504f1a">The HL7 Structured Documents Working Group coordinates markup extensions that have been defined for implementation guides published by HL7, IHE, DICOM, and other organizations. See <link xl:href="http://wiki.hl7.org/index.php?title=CDA_R2_Extensions"/>
</para>
      </note>
      <para xml:id="para_78a64181-f87b-4ef0-ae5a-914466dedbe6">The namespace for extensions defined in this Standard is "urn:dicom-org:ps3-20", which is aliased in this Standard as "ps3-20". Extensions defined in this Standard are:</para>
      <itemizedlist>
        <listitem>
          <para xml:id="para_d0356100-c31a-4354-9c0c-db3fd777a5cb">ps3-20:accessionNumber - The accessionNumber extension allows for the clear identification of the imaging department identifier for a service request (order). While this identifier could be conveyed as another id for the inFulfillmentOf/Order element, there is no reliable way in that context to distinguish it from the Placer Order Number. As this is a primary management identifier in departmental workflows, a distinct local markup is defined. This extension uses the II Data Type.</para>    
        </listitem>
      </itemizedlist>
      <para xml:id="para_ab8ddd95-764a-484e-a521-c88e56d1b754">The namespace for extensions defined by HL7 is "urn:hl7-org:sdtc". which is aliased in this Standard as "sdtc". HL7 defined extensions used in this Standard are:</para>
      <itemizedlist>
        <listitem>
          <para xml:id="para_f5145fb7-2bd3-44fe-9a97-a58a79b9beea">sdtc:signatureText - Provides a location for a textual or multimedia depiction of the signature by which the participant endorses and accepts responsibility for his or her participation in the Act as specified in the Participation.typeCode. Details of the element content are described in the HL7 CDA Digital Signature Standard. This extension uses the ED Data Type.</para>
        </listitem>
      </itemizedlist>
    </section>
    <section label="5.5" status="2" xml:id="sect_5.5">
      <title>Serialization Order of Elements</title>
      <para xml:id="para_706c2f50-862b-438b-8c83-aef8098606eb">The CDA schema requires elements to be encoded in a specified order, which may be different than the order in which they are described in the templates. The serialization of elements shall be in accordance with the HL7 CDA Hierarchical Description. In particular, attention must be paid to the serialization order of elements defined in sibling templates (see <xref linkend="sect_5.1.2" xrefstyle="select: label"/>).</para>
      <note>
        <para xml:id="para_db4b595c-c0ac-4fff-9116-ef4607f3b119">For example, the various header templates are siblings, specifying sets of elements at the same hierarchical level. These elements of different templates must be encoded in their appropriate serialized order in the object instance - all templateID elements from the document template and all header templates first, followed by the elements of the clinicalDocument class in their prescribed order, followed by the participations in their prescribed order, followed by act relationships in their prescribed order.</para>
      </note>
    </section>
  </chapter>
  <chapter label="6" status="1" xml:id="chapter_6">
    <title>Conformance</title>
    <para xml:id="para_4a69aa4c-4fb6-4597-b18d-d866e3fb3acf">The CDA specification section 1.3 provides conformance requirements for Document Originators and Document Recipients.</para>
    <note>
      <orderedlist>
        <listitem>
          <para xml:id="para_5eea1a0d-c83c-4b03-9a9b-38ba6d05e4e9">Consolidated CDA Implementation Guide Section 2.8 includes recommended best practices for Document Recipients displaying CDA documents.</para>
        </listitem>
        <listitem>
          <para xml:id="para_fa04d6ce-cff7-4f42-a7ea-7b34195568c2">There may be other CDA-related standards to which an application may claim conformance. For example, IHE Patient Care Coordination Technical Framework specifies a Document Consumer actor with four options for conformance.</para>
        </listitem>
      </orderedlist>
    </note>
    <para xml:id="para_6201b038-4578-4916-b670-e3b7b4d85f83">A CDA document instance in accordance with this Standard asserts its conformance to a template by inclusion of the specified templateID elements in the document, sections, and entries.</para>
  </chapter>
  <chapter label="7" status="1" xml:id="chapter_7">
    <title>Document-level Templates</title>
    <para xml:id="para_de6e242e-a1c5-4e24-9070-25606a8e2647">Document-level templates describe the purpose and rules for constructing a conforming CDA document. Document templates include constraints on the CDA header and sections by referring to templates, and constraints on the vocabulary used in those templates.</para>
    <section label="7.1" status="2" xml:id="sect_7.1">
      <title>Imaging Report</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0d648aa-38fc-4654-af5e-ee6a3c01a6eb">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f277c7c-6e3f-4a6b-9c5e-e78da9ccf080">1.2.840.10008.9.1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_335b48e9-453e-4446-8427-38fdcc7019b2">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_abcfe6d2-239e-4cc2-8565-aa561aeb2bd9">Imaging Report</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_077aaa24-516d-4103-b97f-edba7b75bba6">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8ae636f8-491d-402b-a6c9-7204eeaa7f6c">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_882ab7b5-e55b-4489-8fea-bb345562ce79">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bbec86db-9111-4f9e-a169-c7f810707725">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_572194bd-67e9-4dfa-b06f-653a9f7a466d">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_177f4d81-ec94-4f7a-9285-12cad9318f00">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94f8ce56-613b-4b53-870b-17b3e8588949">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d09907e-8f6f-4b89-9e43-922b4993da92">This CDA Imaging Report document template defines the report content and technical constraints for top level elements, attributes, sections, and entries to be used in imaging report instances. This template may apply to screening, diagnostic, or therapeutic radiology, cardiology, or other imaging reports.</para>
              <para xml:id="para_a7d9ea85-2c9d-4524-83c1-6ec60796c038">The body of an Imaging Report may contain five main imaging report sections:</para>
              <itemizedlist>
                <listitem>
<para xml:id="para_6ff00939-65b5-47f0-9082-2448185ca350">Clinical information (optionally);</para>
</listitem>
                <listitem>
<para xml:id="para_e17bc3d1-f6f3-4661-9c23-24121aa58ef9">Current imaging procedure description;</para>
</listitem>
                <listitem>
<para xml:id="para_458d02d6-d594-4cdc-a37e-311c4c9db89b">Comparison studies (optionally);</para>
</listitem>
                <listitem>
<para xml:id="para_21427678-772f-4548-ae08-05f0b898e00a">Findings (optionally);</para>
</listitem>
                <listitem>
<para xml:id="para_9f5a4efc-8d06-4654-93cb-9cf1605cd8bb">Impression;</para>
</listitem>
                <listitem>
<para xml:id="para_f6275148-6cfd-47bc-b8f2-10664ef4c5b7">plus potentially an Addendum(s)</para>
</listitem>
              </itemizedlist>
              <para xml:id="para_3a09ecaf-3448-40f4-99a5-e4ed2b7183ea">The report templates sponsored by the RSNA Radiology Reporting Initiative (<link xl:href="http://www.radreport.org"/>) adhere to this general section outline.</para>
                <para xml:id="para_6c38c50a-df80-4578-9457-914784f993c0">The section and subsection structure of this template is also identified by LOINC panel code <link xl:href="http://loinc.org/87416-4/">87416-4</link>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe2ab1e4-68fc-4cdd-9796-d0a8fdea1c62">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1d92ba7-e3b6-48b0-848b-60477491514d">CDA Document Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6a0b430c-e8f0-4c16-aaf3-5c7853f1c0b8">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_876cc389-fbfc-4104-a2c1-e28d6af91cf4">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_db208460-9f07-4de2-a75a-abc65e3a83e2">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_332aaed1-4246-4d47-8899-179d0180871a">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2706c2c8-1349-4f1d-a649-8a3112635e54">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab35ceb0-a095-4dc5-9fa9-694da8d68787">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4e6b8ab-8072-4f98-9a4d-1cd81d99a6c6">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_722ed049-23a9-428d-bbcc-0b6e67f93dd0"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e3ee6bb-bd43-4fcb-a217-9e6d3df67778">Business Name</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_218dc191-f190-40ad-b51f-4981f1737516">Nest Level</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_3963e7b6-ca71-4344-bb1c-e1caac0044c0">Element/​Attribute</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_1b0c0687-8c0e-4b09-84bc-6f8974632091">Card</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab6ce317-a198-40f4-8024-95db82417bac">Elem/Attr Conf</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5e2b4d6-ce7c-4326-b385-cec3daedadee">Data Type</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_6251734c-2496-413a-98ca-4eb32fe68938">Value Conf</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d5f337e-8518-411d-90ba-cb0cd6f0f244">Value</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_f18d94bb-b439-4fbd-9a67-0b9322c6ee65">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9aef70b-56bd-431b-853e-8f7e327fbb06">
                <emphasis role="bold">Imaging​Report</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_000fcc4d-bd41-45f8-b457-9793d82853d8">Clinical​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1be7937a-2bb1-4558-a669-e07cc7c3fcfd">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a85ebaa1-14e3-41d7-a8b9-c0254446d616">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2352a27d-a941-4cc4-bee3-841f355165c0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ce1e988-2d9c-4113-8f25-9598ef43d37a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aea965fb-6af9-4bec-a17a-cab901cc56f6">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9423b0e9-e83b-4a51-8f77-8786a7fd6842">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8a0a1cc-3152-4f3d-9a67-5cdbd4d26dba">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9dc29657-4d31-4184-922e-357a26dac297">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c8df723-4d57-4118-b93f-f4eb013dc691">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07cc212e-1772-4550-96f5-4ac24cfbe623">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1af009e8-5b7d-4a53-a0e4-c14acb240f05">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_640909bc-feed-4b8f-b0f7-ad8ed720870e">1.2.840.10008.9.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cd3ee36-1b10-4e85-991f-ef51a81b670e">Doc​Type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6129867e-ee8c-4aa8-97b5-3c7b3e5c184b">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f50368da-b8f6-426e-af2c-c4c543357269">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73010460-1d27-4deb-b224-3c8f922ad586">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_68359d1f-430e-4ae7-b957-492f0b730f8c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9aae3ff6-d8db-4610-9b33-959ad0e8e663">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b67a1d09-1056-4c2f-99f5-a077359780ba">SHALL CWE noNull</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c34a634-6671-4ad4-a908-29330f534481">
<link xl:href="https://loinc.org/oids/1.3.6.1.4.1.12009.10.2.5/" xl:show="new">ValueSet LOINC Imaging Document Codes 1.3.6.1.4.1.12009.10.2.5</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab777798-da8b-4aa8-9342-37a09987d597">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19edbb71-8537-4f4e-bf92-876332e086cf">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8cfed71-01b1-409f-8883-fb005c5c37d3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2c16f246-4462-4e84-85f4-6e847a8fb55b">
<xref linkend="sect_8.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.20</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4001efb8-2ccf-40d8-bcb4-d9350a4e89d6">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c7e5e71-428e-4be0-80b1-2c977f8047d9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec5da52a-46d0-4c93-8857-4300d49bccb5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95db4bc4-cb39-42e2-b6f9-74a085d23683">
<xref linkend="sect_8.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.21</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_892b5be0-2de3-4a24-bb94-4fd03c23cd9b">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f51c2a61-3c47-44c1-a788-12997d5e9da5">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0061eda9-a4ef-453f-8565-2db6f6df51c1">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c93b0e7-49f6-48a8-80c7-112a7499d835">
<xref linkend="sect_8.3" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.22</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46a709aa-e1f1-4a4f-8b5e-58eea16c9d53">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed09186c-6115-4a58-918b-b41b6f0ab4d7">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce78d515-fcee-4b4d-9366-ec9cf56f4feb">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ed904a3-f738-4a17-9406-d88d791e426e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_99f397ba-0491-42fa-8195-d6f84f8405fd">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97749010-efab-47e1-b4f1-10d0904310ec">structured​Body</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff3ad869-be0b-456e-a33a-848c607601d2">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd89d2ac-8454-4867-886e-98ef7b6b3b29">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c9205b4-3875-40f8-bc84-7a66009a8838">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_315de026-375e-46a1-a4bc-12384eeb2c3b">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5bb1deb9-9cb6-431b-84c4-ef689322dc68">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9001d777-75d9-4906-90bf-eed14ee45b93">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_be4fba4b-b1c8-4f3b-ae64-bbfba60dc44d">
                <emphasis role="italic">Clinical​Information</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_01e27e11-5bcf-4c43-bc65-8d7579da3b6c">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d8b1740-5185-4b00-a24e-4929a44de64d">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83c77a92-521d-4d6c-84ff-f6db83eecada">
<xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.2</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7db39974-664d-4202-aa75-efa23ae1f3c0">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad8e9494-aef4-4eb0-b4c6-bdfc8fc35bec">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61a51a08-079a-4ebf-be56-da028fb311d7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94e04e66-c431-4fac-b92e-b358ece4e3a2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6cfe8faa-0c64-49d5-971c-ad5ae2f393d6">
                <emphasis role="italic">Procedure​Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2345f25-cb84-4d22-b2ee-3405be961189">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_554c224d-e512-47f8-82c8-85017396ca20">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e3fc41d-7449-4b6e-be0e-1da52451894e">
<xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.3</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71ae7a36-03ba-4446-af6a-7426a597166e">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e156ed96-4e7b-4e5b-b2df-7a3fe2f02488">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b1e59ae-37e3-426e-a734-9e54d8c7a2db">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e5042de-22ed-4049-9ef3-dea71aae0ac8">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0fd0e2d9-d34c-4542-9c83-0a9e87d6c344">
                <emphasis role="italic">Comparison​Study</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_524e4d92-b3d9-4b87-900a-2f78b836beb4">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06e3ec92-9c89-4ad1-8b86-24c7f7ad7afc">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_601ce563-d0d8-42cf-85cb-9fdaad1d714d">
<xref linkend="sect_9.4" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.4</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d9d87fcb-3d5e-49a9-b662-feec90f08db4">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1d1aa56-09be-4b2f-bae3-4ab2690824c3">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_44dc77d4-fa39-4df5-93d1-fb51845ed448">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_14ce9b7d-7e72-48d3-9e23-351b8f9233c4">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21e7b211-daed-404a-9729-d3bdee54361a">
                <emphasis role="italic">Findings</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ddf05500-89df-4ba7-bcc1-1c0199d5b850">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0bee936-3043-4744-9ea7-1718675deeef">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_12456135-54b0-4578-aaee-5ae975dce76d">
<xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.1.2</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c76068d4-c877-4ed7-af0a-ffe2e609f6d1">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5bb20748-3ef3-4d16-818a-9b1e3f66d3d8">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_33465a58-ec24-492f-8eca-e8b13993c78e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6a64402-276a-4a81-94a1-e0da874e32e5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6166dd4-0318-4d1d-836c-5b0564101600">
                <emphasis role="italic">Impression</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81ee9ad8-3d1f-42f6-8563-79e498558a3a">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0254dab6-c47b-472b-b285-e390122de179">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e70a468-723b-40e2-9d2c-1251c3a83208">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.5</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_544beefc-f6be-4f44-a3c2-d0f134c78451">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_966433b9-ca92-4f32-b28e-0676473ac60a">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_33876cb0-ed32-4338-bbfa-87cc8edf3994">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e26aa45-6d56-48fc-bf7f-feefe9afefc8">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b904a51-8253-4d17-8e27-bc9733e8016f">
                <emphasis role="italic">Addendum[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c068b5a6-4b72-438c-84a4-f3b311a33c10">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fef7f2ae-a068-4a37-b43c-2d3cdd340a1c">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b076f3d4-1d2d-4fa4-84f0-5ec66291b46b">
<xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.6</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <section label="7.1.1" status="3" xml:id="sect_7.1.1">
        <title>clinicalDocument/code</title>
        
        <para xml:id="para_834f63a5-f5b3-448f-8ccf-3d9af7205868">Most of the codes in Value Set LOINC Imaging Document Codes are pre-coordinated with the imaging modality, body part examined, and/or specific imaging method. When pre-coordinated codes are used, any coded values elsewhere in the document describing the modality, body part, etc., must be consistent with the document type code. Local codes used for report types may be included as a translation element in the code.</para>
        <note>
          <para xml:id="para_1cb0588f-2e21-4352-8ace-bf5a2b6c11d5">Use of Value Set LOINC Imaging Document Codes is harmonized with HL7 Consolidated CDA Templates for Clinical Notes, Release 2. DICOM <olink targetdoc="PS3.16" targetptr="sect_CID_7001" xrefstyle="select: labelnumber quotedtitle"/>, used in <olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber quotedtitle"/>, is a subset of the LOINC Imaging Document Codes.</para>
        </note>
        <example label="7.1.1-1" xml:id="example_7.1.1-1">
          <title>clinicalDocument/code example with translation element for local code</title>
          <programlisting><![CDATA[<code code="18748-4"
        displayName="Diagnostic Imaging Report"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" >
    <translation code="XRPEDS"
        displayName="Pediatric Radiography Report"
        codeSystem="2.16.840.1.123456.78.9" />
</code>]]></programlisting>
      </example>
      </section>
      <section label="7.1.2" status="3" xml:id="sect_7.1.2">
        <title>Addendum</title>
        <para xml:id="para_07f855f9-7668-4806-90f0-919fad989ea1">COND: If the header includes a relatedDocument element with typeCode RPLC, and the replaced document had a legalAuthenticator element (i.e., was signed), the component/structuredBody
          SHALL contain at least one <xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/>.</para>
      </section>
    </section>
    <section label="7.2" status="2" xml:id="sect_7.2">
      <title>Imaging Addendum Report</title>
      
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_885f2e61-66dd-428d-8ca9-d3c6dab5f0c1">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd60de9d-b718-4b4d-9e17-7f3872b3dd72">1.2.840.10008.​9.24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87e6e23e-2648-4cb7-947d-fd56e3c4c8b9">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b34a7dc-c25d-4acf-a273-53260d2e1313">Imaging Addendum Report</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_738f197c-d548-465e-9be8-5d4f3cc9dca0">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d0e50cf-40bb-4d5d-870c-287d03c38b6e">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7d9cf0c-d69f-443e-88d5-81c93f43937d">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c856bc8c-79af-48ad-ac6b-8ba1142b1f32">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c08d9b12-c43a-43e8-8d5a-aeecf50093e2">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70e1bebb-f16d-46a0-93a6-2ce9a4b55557">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa439572-dcef-4c21-9a9b-39775c2d14d0">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aaa27384-f8e2-40e0-be55-ac66aafe8772">Document structure for an Imaging Addendum Report, i.e., an appendage to an existing report document that contains supplemental information. The parent document content remains unaltered. The Addendum Report must be read together with its parent document for full context. Some institutions may have policies that forbid the use of Addendum Reports, and require revised reports with a complete restatement of the original documentation.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c221f2e-4299-4573-b642-099eef0d32db">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dcee9a8e-fec7-452a-abf5-8c988074f775">CDA Document Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_962295c4-8510-4366-8f52-904e209479e3">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_69713925-3a2b-47f3-8c37-fa41dc2547be">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_866e95be-86e7-4707-b5c8-dca71061f530">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_26fee426-e910-425e-8980-0318c4334be3">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_839f285b-397b-4c5c-8510-41965a60a042">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3dee2fde-bf22-4694-8bdf-450853b96cc9">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4d1a300-657e-4634-882d-6395047a20ae">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_17e1080c-c545-4d26-9701-2e904cdf7a1a"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_7fd48487-ac9c-4c7c-a223-2cd7d43518be">Business Name</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2c3f1d3-82ff-4183-8f80-f22d0b7729d1">Nest Level</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_d390523f-417a-45e5-812c-a26f0f5ca799">Element/​Attribute</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_0eace36e-0519-4a14-8ebc-78b3f657e0c7">Card</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ed585c0-02e3-4de7-a7f0-5e35f0ec2da5">Elem/Attr Conf</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_8db60bda-910c-4829-b2f8-4eb94020c85a">Data Type</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_55d428ed-bce9-42de-8ae9-c4ea924af14e">Value Conf</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_7dd89195-2953-4b0f-9ba5-906ecd4066a9">Value</para>
            </th>
            <th align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd5e897d-04e8-4453-ac3f-9a381c7cf80f">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e3a17ce-5f70-4190-9975-f83656948b03">
                <emphasis role="bold">Imaging​Addendum</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_443881d3-7d42-4d17-b8a4-a8ffab965e91">Clinical​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f7b9ed4-74b0-46d9-b61e-d38c2fff47e8">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d93447fb-3996-4e03-97f2-d4c149769af5">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_658c6a35-ac05-46bb-ab78-1d71bda9b317">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_72a8b590-a51a-496c-b032-faddce1a3dd4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd476eb7-91b6-4268-98fc-a202a07444b6">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad26b8a6-861c-4b89-9a0f-5a55cee2d87e">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4fbf111-bead-4343-88b6-908608c35a52">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b43a860-73dc-4205-980c-7ea7087ab121">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d84c19ef-b68b-42ef-9ead-1d9036ceb7c3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6bd93bc-62b6-4b48-ac9c-105dcea0ceb3">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17242ff0-a00b-45d4-a006-25cd05d816a2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9fa6c93-12d8-43b5-9daf-85e15451145d">1.2.840.10008.9.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7cc2181-41a4-4eac-ae80-b7c55d5605fc">Doc​Type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_75d206fd-bede-4f73-9f3a-022dcc155d80">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4353ab6-ed8e-41d0-af62-b55fe6e3b158">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c494bd09-602b-4838-a60a-3dba7eb5ebfc">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9dc80915-1c21-4881-97cb-94db0fd07d3e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1d33a1c-9b78-432e-8993-ea7895ab1bf6">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_69146ad8-6cf0-4031-9a29-eb9b9dbf37e8">SHALL CWE noNull</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e1d8a5a-0e75-42cf-b428-592c5082ecde">
<link xl:href="https://loinc.org/oids/1.3.6.1.4.1.12009.10.2.5/" xl:show="new">ValueSet LOINC Imaging Document Codes 1.3.6.1.4.1.12009.10.2.5</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_212e8122-58f7-4e54-b7c4-da91d445afd2">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_344c2158-3400-4f6c-adcc-f4890af60837">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_44fe1c7a-91ed-4d7a-99a9-c57ec6270f22">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e84b4e0d-5447-4ad8-8983-f89978a6ef7f">
<xref linkend="sect_8.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.20</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8207bca2-7110-476a-8e2f-98322f194228">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2d021d3-8e5c-4458-a98e-7c587ca89f3c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_86689078-d13b-4f23-8cc2-b9125c09e76e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4e0247e-e740-4347-b2f0-cc5504dba2b8">
<xref linkend="sect_8.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.21</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd09feff-a81f-4784-b462-ed454bb5d336">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c244bf8e-6436-474a-9c3f-b21c663f0483">related​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ae50b09d-9eb4-43f7-8346-6ddac3c17399">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a403af0a-3b86-4676-870f-15a7ffdf5bbe">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8915e0b5-e78a-49e9-9de5-5d163ef2f137">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d40e9b8a-cc51-4cc9-8afc-fe2e7e5c14af">@typecode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_acd64bad-7b53-49b1-9e68-c849fd814e94">1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81994d8b-8cdd-471d-8d06-6a76b3d9eaaf">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_45e199b4-5443-42a4-9139-ee865734d908">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b110591-3a25-4c1d-a397-19a8444a6bf3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c8aed2f-64c9-4fd0-9c58-5e6815c0c8f4">APND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f63e1cfb-41ec-4a0b-a677-834485d18902">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a41e195-9255-4882-8069-32a15328ae9d">parent​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_49ddb3a0-cd4a-4b50-914f-c13bc0c5dc34">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2f1775e7-b42c-4fa0-b27f-1d0d3a393f8f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f21cb79f-d4b8-4633-825e-9f3f41a7ce5f">Amended​Document​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2764a95f-9de9-4940-85a3-65e33f9a1618">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_961449c2-6cdb-4da8-9df3-8cbff489e53b">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dfb140a5-cf82-457d-83ac-2e08ac2fff2c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e751767-bc46-4c32-816c-90bfa3b163ed">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d87b1258-7a85-418c-a538-a9b7aac2696f">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bea13624-d891-4c98-9089-e8829c1a536d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31256be6-13dd-466a-aeca-8fddd8a81b79">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_319223d7-f89b-45f3-8dc2-5b1d8f21a843">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98721acd-f7ab-42f8-8cc9-2cd9b01932b1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ac8a34c-782a-44ac-9910-a231f916b34f">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e2ef02f8-99df-412c-8692-57b8e1738ca2">structured​Body</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_040f77ac-ee21-4fe6-a15b-359d39af1cc3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1167c820-6b4c-40e4-92a1-31ef0abb1e03">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c70111de-0f19-46fb-a257-084d332e961b">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d08a9721-db0e-42bf-90ef-d01628558f87">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c8ac302-77ed-4ca2-a037-d758c3d9d1c1">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20935c5c-d26e-4536-92b0-39e0e66aa66b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f287d899-d111-43db-81e0-ff7cfdcafdbe">
                <emphasis role="italic">Addendum[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d3dc9bb-da54-467e-8c46-0e860cd29e85">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a34b8da-be61-4a6d-884b-dc61cc835cf6">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9a6879d-0240-45c3-b981-7de167eabc39">
<xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.6</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
    </section>
  </chapter>
  <chapter label="8" status="1" xml:id="chapter_8">
    <title>Header Content Templates</title>
    <section label="8.1" status="2" xml:id="sect_8.1">
      <title>General Header</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a800724d-eb63-4364-99d8-fb00b6328308">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5926c395-f135-49d7-b829-f0c3faf92c4c">1.2.840.10008.9.20</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa8d343e-570d-4893-b527-0ca7a3ed55b3">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d161d678-2c07-4dbd-892f-10cada087619">General Header Elements</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94ad3d10-756f-4170-af92-20a2e51a4298">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_38e79674-53c1-42c3-a11e-5d0847b73a12">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cbdcb0ab-3c0f-4967-bd81-ffb2c4287e2b">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4308f87-45e5-4ed8-ab47-7bf249ad9b46">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_064fcddc-583e-4ff1-aa42-dd6469d92dbf">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15c236ad-3fa9-4bb3-962c-3514da3c2ff6">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f29a9a4e-7f84-4acd-8bf5-b310f74499f4">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a3d03b0-7b16-415e-b592-ae7bff5cc00d">CDA Header Elements for all documents, including primary participations</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_52e58642-5b35-49e3-a2a6-a52972187e4d">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c1c9377-ed80-4810-b085-ea52ca117764">CDA Header Elements</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bac1609a-e190-4112-8921-e250e340f88e">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6a055e9-8322-4eef-a78b-bf3d8f9e9ea6">Included in all document level templates</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ad94ef2-a8c9-4a60-996d-1bb30c355d6d">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d215126e-f99f-4745-9000-bf5ec80aa00f">sibling node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_354cfe35-8e59-4de4-a0b3-b25cbef7a772">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da25bab4-4164-4d5f-86db-316760634a16">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d5a59b82-ef91-430b-874a-96b5aeb1d1db">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3e31b39-e666-4bb1-80f0-1e20c8cc5fe6">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_25dcf3a9-8d9a-4379-82bb-8818bffc1708"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ccf344f3-a54b-4762-a523-4d57df4c7770">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_16a30197-944d-4855-b29e-81e7a49a1660">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3661cc18-0800-43b5-a430-5309bcd37a96">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c69670f6-7304-4020-bdaa-675b4b541280">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_60563478-fcac-4f68-a257-315bc33c2a85">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0c3cb56c-b345-4820-b15c-9b799c7b2594">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b230e24c-07d2-4b3a-a4b8-9ee126f2e8a6">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_44af94c2-b36a-4b36-9098-fe02666163d0">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f7dfbbe0-d942-4eb2-8ee1-2e16f0ecd1bc">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5701ca48-33fe-471c-bb89-350ea82a0165">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95d641a1-f9cf-4504-973f-99718aa9aab2">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89df2a1b-7b2e-4fd0-b3e1-94165f84567b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c735e36c-ff4d-49f3-88fa-b5cd3d35bf70">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe4754b6-0c7d-4dd3-a3cb-9fa12b70a807">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f37723b-0eac-420b-8261-bdfc0f3a3c16">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aa8c54c0-fb7f-45e4-9eef-ac3b3b6d4674">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_def01e29-1cf3-49eb-8ddc-282a10d506c2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f08f862-5c04-4f47-b31a-05ac8ec91e7a">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00e2f0a4-2074-4ef7-afd1-8eab50153627">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d610162-20a5-431d-9c25-2d25de4df455">1.2.840.10008.9.20</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_57fe3b26-0b7b-49e5-8ab6-c1fb1b0f4f15">Content​Template</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6922aa18-8534-4c3f-99a3-7b2e5528d29c">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5d4e4ba-1f3a-4db7-847c-0bf5e6b85be6">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5cd070de-9107-4748-ada5-4da37de445a3">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_493814d3-39c4-4776-8fe8-005846deed42">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a3d56aa5-d206-4520-b16e-e644f3aa4cbd">type​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97dca9d4-e4c0-44c4-b597-ebe61496f146">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70f8768c-dd26-4d23-879c-02d679a57864">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00cd63dd-60c1-4772-a0d0-06afc4759e59">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6fe80c30-2a67-465c-b66c-44d0ad6b0d9a">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4a33d72b-fd3d-46c4-9a25-e17d2555fac8">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_984ec92a-7e79-4f68-96ab-7f48b3984583">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f698998e-e9a9-42b1-adc4-554420ea504a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3496b9bf-70c2-4a46-951b-ad9da86057cf">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30644750-244b-4238-b6c0-f0e94f357594">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c4f866a8-b837-4d35-9e13-1c507adbb740">2.16.840.1.113883.​1.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5931615-b5b4-49e0-9abb-dbd0ffbbc945">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a92b27cc-bbd7-4d4c-b756-ab11d9e23b1c">@extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6139433-e138-4fc7-9812-92bc87a9b5a3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e24a63bb-0d9e-454c-b425-ecc5a291f5aa">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43036758-6056-4fb3-888a-ff61ce262c88">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cbc715b-b1a6-45e1-b2ce-4179133c1c4c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_205aee09-7cee-484b-9518-d7f076fbea6d">POCD_HD000040</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_104bbb72-9f2a-4445-b54a-5fd266d26ec6">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_607725ff-7b15-40ab-a631-1d4ac67f70e9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54092c6d-8953-4658-be13-5e3af71ed737">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ba88a4b-d86d-4315-84eb-29492abc9017">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ef11a944-0b4e-417a-8502-c1db9456171c">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c4e166f-f075-4faa-9a29-cdbea08ff703">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b585ac55-c04b-465c-81ca-3252088b351a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8540900d-aa3e-47a9-8f4a-504711fe70d5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_18baf7de-5867-42dd-96b7-534dba0cf027">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bdfb2fcf-8ef1-4613-b899-38882338c6d2">Creation​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5570ca4c-76cc-4f27-bc47-5d8ad1f7b3e8">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6a4411cc-b7e3-446b-a1cc-1f49322c0aad">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e904b65-e0ba-4859-b2d1-38c2c36fea92">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58806620-d7f5-43cb-8551-952c5ba99c11">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8383c58c-316e-4d75-bdf4-c59193dfd0cf">Confidentiality</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39908245-f2f9-4001-a977-dcbd4e190862">confidentiality​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08b4fc12-7cf0-41b0-80d4-ada07a77b93e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a5e9e0a4-27ec-44aa-bdff-75ad8ea2c86e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31210e12-9572-4255-8a8e-d2df6244adb2">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c91915a2-f586-4a18-949a-d4ebf3e7391c">SHALL CWE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7c9bacb6-cc28-4ee6-90cf-303f77cd167a">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.6" xrefstyle="select: title"/> 2.16.840.1.113883.11.16926</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63a3b3cd-445c-48f4-96de-5d9eafdedd22">Language​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_24eb7b68-24f9-4380-a2e4-4a49223e7347">language​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50280a99-0149-443a-82ba-408b92b88952">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c3a0083e-cf60-4239-ad3f-f0258d453c3c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_611e6628-dc82-4ca4-afb4-42c540eca503">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6c76faaa-1990-483a-862b-a723f70ac107">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d295dcde-947a-484f-9275-ec0d978df432">ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_5000" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1fc7f5ef-bf6a-4d28-b969-b71d841dfc28">Set​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8001d972-68aa-4730-b003-d3df457d3eab">set​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d4fe87e9-0522-43d4-a69d-51793deb349d">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ec6e149-957d-4d4c-9924-ecac42ef1fb5">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_232fce65-3f26-46e8-8de8-2dc297909576">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9013099f-1b68-4762-bd10-4737d081c91f">Version​Number</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4058225e-57aa-483b-b9dd-fced50d972f0">version​Number</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_85a6135b-cc9a-40ff-9844-54b7e9a95816">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f073c073-10ed-4479-bc4f-3bdeea29f7ef">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b6c3ce1-9ff1-45ae-a020-5a0a045ed001">INT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfc367cf-2cc1-41c6-824e-78158e5b1021">
                <emphasis role="bold">Patient[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2724aedf-3f0f-4dde-93fe-f0ba6e009e9d">record​Target</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0a0e3e5-2f80-4f6b-a12e-0b4c5afd998d">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a161208-2a8c-46a7-b42a-e236c4b84b53">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_af9cc117-bf6a-492e-9f79-52935d2a6ccf">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b8bc2a1-1934-46af-abba-e249d03d5ff2">patient​Role</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e6ea6898-9cad-4b50-acba-bdcc4e0a733a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ef6a48c3-b5f9-4346-a6e4-6c0187c16380">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8ed5ed26-4db1-41e2-b981-6c6be370268a">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd5e735b-3e75-4de0-9a2d-be5defd2d351">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5767dc28-f132-4c42-814f-e52bddc2c19a">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf75c743-b58a-43a8-894a-38e5710985ec">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_45ee6600-a714-47f8-9d52-36178518b21a">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bddcb167-818f-4b47-8c89-3b0b22454e89">IDIssuer</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d20a77c-700f-4bad-9a16-f113c816692e">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ac23d20-425c-445a-ab29-7fb84adecc88">root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4be375ec-946c-4a67-8b4e-313b00ad2466">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fc5d143-a4ea-47cf-ad6e-c7f7efeb64ee">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d2aec20-1597-435a-8ad7-15e13854ecad">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_873b3dbd-2f10-4278-bf77-24d805ef56e9">
                <emphasis role="italic">Issuer of Patient ID Qualifiers Sequence (0010,0024) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
              <para xml:id="para_655cbd2c-bb0a-47e1-8648-baddb16b60e4">
                <emphasis role="italic">Patient ID List PID-3.4.2</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_623a381b-b855-4d0b-ac8b-46a208309733">ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_180d6949-e3d5-41a8-83c4-dd532b4c4063">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a15d30c-5554-471f-9ff9-6418c8dcd171">extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fb41af2-eb4c-4b1a-b550-932ca90863e4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_edddac35-2f49-46a7-aa76-1ddd4ff3662a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b5e13394-0eb7-4c3e-add1-a842dd8e4a3d">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e109b2b7-0ff4-4258-bb1e-1681b271c7d2">
                <emphasis role="italic">Patient ID (0010,0020)</emphasis>
              </para>
              <para xml:id="para_6b16cdf9-fd2c-47c6-8fc1-2f8d33380074">
                <emphasis role="italic">Patient ID List PID-3.1</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a6b87357-a407-4075-bd86-e08782be25b5">Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64f6100a-107e-4496-bcb3-56c319a6c7fa">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff08b7f0-5d43-4ada-bb25-cb5e9b785ce3">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5383bac7-c757-43c8-8e73-26648df8e95d">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_555ec813-298d-41c9-b123-f53fcd4904bc">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0cee96e-e615-4329-bb11-ffb8bb60991f">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_210d3f16-32fd-44f5-9768-41ad10f3cc02">Tele</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c0460ef-3809-4a4b-a8be-87f51b3dd97d">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_508eccfb-3e06-437f-87f7-1ac2a499ea93">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_91541cc9-f6d6-42d4-b671-1407c235c38b">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_164e6a42-01ae-48e4-9f98-c1abde57e58a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b4e9069-7ca7-41be-af48-57b015a969f0">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a72ab8ec-566f-4e56-ad71-c7eb9e36e980">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a292644-a60b-46e1-be88-390aab607946">patient</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_093fbb0c-1fd1-483e-bd97-401f1b129ca6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8489916a-e828-479f-b3e5-bb1de50c52ba">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf67436c-2fd8-478e-8e54-fe7d8dbe59c1">Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_379383c3-7d8a-4289-8b0b-6a78cfebcbc2">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab7cc99f-32d4-4fff-9c14-4fecae3a7b7e">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c99cc25b-f213-49e6-9d19-eac585e22046">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c48082d-6615-417c-8546-910065ac8373">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61fce672-a61f-4a88-9dd7-59e6eb36c5ce">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7611651-a545-4f7a-a437-41e954f0beaf">
                <emphasis role="italic">Patient's Name (0010,0010)</emphasis>
              </para>
              <para xml:id="para_1a83430f-0c04-420a-9496-cb35440ec6f0">
                <emphasis role="italic">Patient Name PID-5</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60d04b1c-15a6-4a54-be0e-a10f158c0b2e">Gender</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89fd2f7d-06c3-4533-84c8-c251b5464b81">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c15cebd6-6178-48dc-911b-6deecfb49aad">administrative​Gender​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8dc7d70-0277-434e-ae0d-01a88b8ec138">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a0e4a04-6376-47a0-bea5-a1f4b28e445c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c9433a3-39cf-4f6f-a463-46b507e454b7">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_adee493d-97e5-452f-a71b-1b829b85ed07">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_69604667-5e75-4dca-8617-168c7e69de8d">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.2" xrefstyle="select: title"/> 2.16.840.1.113883.11.1</para>
              <para xml:id="para_68e8882d-f51d-4ae3-b93a-33c19652c522">
                <emphasis role="italic">Patient's Sex (0010,0040);</emphasis> [Map value "O" to nullFlavor UNK]</para>
              <para xml:id="para_82964924-a6a2-4d1e-9787-aa408d2a77b1">
                <emphasis role="italic">Administrative Sex PID-3.8</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cbadc8ea-de08-45f0-bbd8-8168d98bd177">Birth​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9cc4a2fb-c4c2-4a67-ae29-b15bedd91eed">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5edf0029-8227-413a-8efa-43bb912997e2">birth​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad988b65-dac9-4eb7-9b04-a525416cacc4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e058481f-4d44-4361-bc00-b6328245b539">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4736b173-5a7d-4914-accc-537094a3d81b">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8e78046-0147-4635-b62c-83fd8fd827bf">
                <emphasis role="italic">Patient's Birth Date (0010,0030) + Patient's Birth Time (0010,0032)</emphasis>
              </para>
              <para xml:id="para_f3f93502-9450-4983-9c35-cbee5e0bb46a">
                <emphasis role="italic">Date/Time of Birth PID-7</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60d055d5-90e8-4a1e-a12b-eddf5e4e9fc9">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aa4bb5a2-e779-4fdc-b64c-ef10d6f57618">provider​Organization</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac21f4fb-d796-4362-a0d0-485ce86310bc">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e9dd2ba-a7d9-4ac1-ad13-74dae5344420">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_222814d8-7b85-425c-b540-36d53ff4ef8b">Provider​Org​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4fc91ff7-3517-4861-8dbd-d7211d2fa729">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_09e63bcb-9548-456d-a1a5-6845b216f24a">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab00f09d-d674-4914-92bf-c0b3466c9262">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_faa323b3-4293-44bf-ad65-3ad35f08a7f9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f28d1ba-afb7-4af0-93e6-4005f8658375">ON</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba2ed0d2-8f39-48c7-b918-c754e215b195">
                <emphasis role="italic">Issuer of Patient ID (0010,0021)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77f3c73c-727a-4551-9ff4-17a4e3a5c052">Provider​Org​Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a9c7da3-be70-4226-ba23-f4b254439992">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71f39bb6-43a1-44b7-8beb-cf070d0d6fac">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e2d31df-c461-4f2c-bf26-ac6867eea4ba">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6744d15a-c7e2-4218-8989-76894bb98b59">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_93154553-a46d-4e39-b0d8-892a0b9b0dc0">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_620ecff5-52fe-4539-a574-dc12ec71ecc9">Provider​Org​Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f28e261c-1424-4cf1-a589-ab554926031e">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43f8b3a6-dcb2-4e29-884e-477ceeac4009">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_377feb64-6d31-413f-a720-c1c715bd5e2d">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bfc83957-4cba-449b-940f-f175dc4377f2">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0bd54af-557f-44fc-93fa-99a30e94767a">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e6d5440-b1f7-4324-a033-f34bf2c14a0a">legal​Authenticator</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ef0dd129-3753-478a-850e-a176f392e16c">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_96ec7b0f-bec0-4240-8d89-68e8e3dec5a4">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_498eaeaf-4b58-4869-ae57-4e5a2f5ec321">Signing​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84479075-7905-4937-ac3b-d5a45d50e1c7">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64488892-b5a0-4416-96a1-4b4f1d4dd7eb">time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f7645ba-9f3c-4cd0-a551-c2afe8ad2482">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6270697-7890-4489-8697-e22ab029e1ef">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ecde528c-970e-4fa9-94b5-a33e40cda2b1">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5bde96ee-7376-430f-8943-ce4bf2b40071">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4a39edf2-3947-496a-81ba-7d8bd733792d">signature​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cec8a363-be6c-43a6-b15b-d8a227e92091">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_baed8013-78af-4e1f-a331-6e6ab4648523">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f742812f-cbd1-4ce8-8ce5-a7a953c53489">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_90a89bbc-7da1-48dc-8d7e-ed1042e23568">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_27eae027-6246-447e-9ea1-3040f7ce5f12">S</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bebb56e9-d5b3-44ad-a0ca-e6ddde70a138">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_585f6996-afb9-4ed8-a188-71372c92bb49">assigned​Entity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9e8b89e-e198-4544-a3f5-0f43309c94d3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9377783-acf7-413f-bb53-8e239d740c1d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bc67da7f-87a4-4966-97bf-733536faae65">Signer​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_deaf08d8-5aac-4e66-b8a7-e6205068db51">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ee932cc-3240-43cb-93de-f0d9327fa680">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dfc87ec7-0367-4f6d-ae6d-1e012130f37b">1.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b917c166-7981-43a7-b975-f167d8814185">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd844564-2b33-4097-8685-95bf0c7d4411">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36392935-6c75-4421-99ac-86e4db817b98">Signer​Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0a0174d-adda-4c9e-9ff8-aa0fabd0cab5">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_891bb706-e6ad-4fcd-b9a4-7381050784da">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9193daf8-3e70-467a-8970-5ece33c3e6f6">1.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06665ee1-62f7-4f81-aa2e-826a2de34ef1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba0ebcda-cb26-407f-8b83-ae038ee7a2b3">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c82d0a2f-a406-46c8-ac7f-52b9ac11ae93">Signer​Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_96c3f2c0-b6e6-418b-95e3-0dae04080a7d">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f41100f-2249-4c1e-a790-2c365c674aed">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5aa774a0-2a75-43ff-912c-c4d237a687a2">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e24a495-5661-4e89-b839-58ab4dc26649">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f96e417e-2c20-4934-9a1b-00021f794909">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8faf6c6-03af-4c33-ba2d-630d53aea648">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_04272d16-c20a-4187-93eb-702eb7ffb76c">assigned​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97c3918b-dad1-4067-9aee-a128259ca31f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f4d8d6e-bd1d-4fbf-90d3-4c2fc040a977">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6066abc-5a0b-4ef4-95c6-b203d3f2138a">Signer​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ab4c385-f093-4fdc-8ba0-d26e45bbe81c">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b6649f8-9c01-45a9-a691-03803b9570ed">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_99289690-5f2e-486a-9967-8e30256311ac">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_809bb8b9-0d5a-43a1-beb5-966370294ed8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_600677b6-a8fd-436c-bab3-fb9333a67631">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f6f13599-665e-4ca2-afca-62e19215539d">Signature​Block</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8cb83955-ccd1-42ea-b714-26c4f54fc843">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a182abc9-a9e5-463e-84d3-9f556d4d9bda">sdtc:signatureText</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4aadfc82-06c8-4249-87d9-c67b3fd17c2e">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aa0efae6-f370-4ccb-a685-c634406f6b32">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a056c14-ec1d-4629-95de-54949e9d4773">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94dc15ff-09db-45ab-aa27-7fa9fe324984">
                <emphasis role="bold">Author[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_587074ab-6bdb-413b-8480-defa8e664731">author</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_706eec09-c0f5-4f66-b371-00e7e269b569">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3984038-46ff-4026-b857-2dbe6ff39e19">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c98700ea-e858-49c6-991e-67830d8f232d">Authoring​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_92c1892e-100a-4c14-b522-9bacf2caa363">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8542b168-03d4-4a71-98f2-8916fe05c31f">time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9e38751-b0d5-4336-bfab-98e048f43e2b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e995fad-7a64-45f6-bdfb-6502a145e5f9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6f329ab-e88c-4d10-b94b-51a941774a14">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9fd8862-7ef4-493e-af3c-9cab91d4962e">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53a2a26f-eb3a-46f3-be21-bf016c1994c8">assigned​Author</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_449a1930-84cd-41b2-9870-a8720c09aa9f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9bd26d2-93e4-472e-bd75-b5a4cd1075c6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23ff33e0-c189-471f-a00a-107d0b1c36f7">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cb6ff9ec-edce-4dde-902b-a4136bc02399">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ebed07c-a618-4b16-8f0b-c8b5d0823aa4">1.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94371561-1033-4ba2-810d-6505f4483367">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9de70a80-d3d5-4ff8-88f4-7e53a6504a07">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_faf4dc1e-e306-4e82-90ea-247d50536249">Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6880f2e8-0bbe-41a4-8091-02c7a18b6e8f">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eace953e-5973-4949-850d-6f0d049c1d5c">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7fbb76c-498d-4d1c-adf5-9ac4380cc08a">1.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_25bb4f35-4ba1-4496-9245-b7364b9272b9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7a37784-bd45-4199-b9c8-4567c157f231">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a26eb164-ef75-413a-a5c1-a641323041ea">Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e4ac3f3-d986-45de-aac4-8b305f4804e7">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74f6ac7c-286d-49aa-8b82-124fbe4c4263">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9cde16ea-c12d-427c-92bb-df3a1e3dbc70">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_18027901-4137-4ce3-ba72-6bff244db419">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58a59e00-7978-4b72-8209-75eaf3093791">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5595b818-894e-4fec-a6c3-8c29b01aecd9">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_587d634c-616b-450a-a4a0-5b06b721b5fb">assigned​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7c27a61-0c65-4928-92c0-a28fe801efa5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a97bdb7-dc37-498b-898c-6e87d59fbfea">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8e88c4e-2e3f-4f05-8176-1c855a9ec220">Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_45ba89d1-d70d-4b10-8cba-faea3da7c0b8">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_01fa4ef1-9754-4802-a461-03fb4c2b5225">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21410c32-004c-4304-a7ea-1db6f51c9ad1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8455038f-97f2-4d81-80bb-f181fbd3a447">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9eebd9d1-4b22-4f47-9c49-af0862eb1406">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a3ae8ecc-a983-46aa-8e5e-f7568c1d19f5">
                <emphasis role="bold">Recipient[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_469970c0-627f-45de-8c84-fb743760503b">information​Recipient</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f20a9fbc-f02c-4b2c-bf7e-167a78dfc789">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5fffdeee-d2eb-405e-b003-a19392ac0b03">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e553426-e34c-4807-9524-399e67da20f1">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61b2ba1f-4cc1-446b-8b0f-34b41effe4dc">intended​Recipient</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8095417-c7bd-4744-b3d8-b015536532fa">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3390df54-5ebc-4995-bfd6-1d8ad6d849b2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_91143445-cebd-4997-a906-93d8a75b74d9">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6477dfb-8b9e-4314-a9e5-d663cad9c2d8">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_38e8cafe-c4d6-4464-b319-7938c2681013">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6c755a9-c473-41c2-b9d5-cb34b366ea0c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c56feaf6-b804-4537-aeca-a73b407c113d">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_643aacbc-0f6f-41f8-816f-e981becdbbf5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8102ea84-d18d-4aee-a907-119d57ee1e9a">ASSIGNED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e385914e-4832-4cde-8c58-dd091979ed25">Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c8f3afa-da7b-40fa-ad90-ad67a0c6b43c">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb5b4593-3364-4c56-a24b-d4498fc4e1c2">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e74d5c6-706d-43b7-861b-3bc4303665fd">0.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2045d502-f209-4807-94b0-cc9b6165156a">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9274d878-7d23-4e09-8c1d-ff9b93a6cc2c">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9d90ace-7cd8-4aa8-8280-15736dfeb0c7">Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_38011561-8f67-411f-828f-c055271a83ac">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2248e548-070d-4ac4-9795-c4d34f8495c4">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a27ccd1-d88b-4d54-b5f8-caf9bbedfda5">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e9e64eb-ccc7-47b2-a915-2f7ebeae1e79">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f7770d28-035c-4a6c-8e0d-ff4dc30fb2b7">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5f177b0-433c-49f6-982e-d88fc57b0891">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c015316f-cf65-4838-9c89-b58d9ca449da">information​Recipient</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7e9998d2-466e-4995-9214-9aae75951e29">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46c23c5a-4963-4a47-90be-9a0b9eacadfe">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f6f9111-f120-4f2a-9023-b6e22ce0c734">Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ba7f806-9536-4c2e-b38e-0592b9a65079">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dcf981d3-12fe-4880-b39b-a5a21cc90484">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94ca30bc-0160-4e6e-84d8-144ca5ad3241">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b64d21fa-1149-473a-9f53-1133a52f07ff">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_240eac13-f220-49eb-a93e-37254311dcb9">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5745a9be-e670-4c79-bba5-9c47748d70e1">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bdf881e6-3f3a-41cf-8b76-822fca647e7f">received​Organization</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95a3646f-bbe3-461f-8ce6-36dca040957e">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a45ee94e-7875-47ab-9cf1-ef7bfbc5350f">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ab1a1f3-f7be-41e3-8281-c84f14606153">Org</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e1983451-3c31-49e8-ab15-b1e80c92c2fc">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7dfa3317-8647-445e-8db2-1feef7731394">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6632acc5-caee-4739-b033-762a2410e63b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aee16113-75cd-40b5-8e61-946ad75c384a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca0f282b-853f-467e-b9ae-14211e1982ff">ON</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_253199f3-e561-41e6-9cbd-d6c124b5e6c2">custodian</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70353a17-874a-413b-941f-48faa9bbba2b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb8321cb-26a0-41f1-bf1a-23cd0645d30b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5f5216f-b6a2-4f4a-b42a-65b32eba0e4c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d62d1950-1a49-46ca-af15-3c86ef909239">assigned​Custodian</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0bcdd2ff-b093-4db1-8c18-0e7dea5d4b93">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ca47eda-4440-448f-95d7-6b510b7d407c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b432a07-b81d-42bf-9645-12da5476c6af">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30518e60-290f-4e1f-bb29-bafdab784622">represented​Custodian​Organization</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2f6803ac-8a82-4d69-bd94-f6a5b65e0829">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74c05aaf-bdb5-408b-879a-eb74e31777a1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_409eb033-2224-4e4f-a3cb-d0c295cb0215">Custodian​Org​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e1103973-a1d0-41a5-acd0-10cb6e54ea2e">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c24dc26-5dae-4797-ba49-4d8022d5ff08">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c655bb6-9b72-4e7d-b5a7-ee230de23953">1.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_758fea90-f588-4772-81ab-7fdf3916acf8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a330c99-89ef-49b0-bd24-3c1f60bfc1ce">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2500eaa-2f93-44ac-9343-d834b194ba49">Custodian​Org​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2de17aee-c128-4103-ba28-64b816bafdf2">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f27080d2-f789-4512-9d82-43126259c907">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c623a72b-1640-4d8a-9983-09facf36c05d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39999381-3285-4e5a-9758-b9bbbef72fe6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f6c0760-d297-4449-af51-2fa94ef1a881">ON</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4980915-fa42-4bd2-87a4-988446047c4a">Custodian​Org​Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8c3c15c-eff7-4ad4-95f4-d2372228631f">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e0297fe-8e28-463d-b09f-6c07ec73186c">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8c39361-1fd1-4c36-b362-fc07f3052ba5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b952fd7d-6a57-416c-b7da-aacf63b428b6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a80e39e9-6c3f-42e4-b667-dddd25f6ca80">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4cf36f5d-5d45-4402-af16-163e57f3de84">Custodian​Org​Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c3fdedb-cb6b-44a8-b941-2031e5cf15e4">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5606d51-8441-46d8-b54e-cfec16088781">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_699975ef-86fb-4309-b065-9f26471b83a7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_27ec1acc-ac2e-445d-a09f-185387f67a7d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fbaaba3f-a70c-48ab-99e3-921d2ef9d426">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_8550e930-ea1d-489c-9d60-0c1be0e8ef1e">Note that there is no business name associated with this template. Rather, this template is an editorial convenience for template specification, and the Business Names for the elements of this template are logically part of the business name scope of the invoking template.</para>
      <section label="8.1.1" status="3" xml:id="sect_8.1.1">
        <title>templateId - contentTemplate</title>
        <para xml:id="para_200d9132-b956-4a75-91dd-04d0f48a0c89">This templateId may be used to identify the template(s) used to generate/constrain the content of the report. This element is in addition to the templateId of the document level template, and typically represents clinical sub-specialty requirements. See <xref linkend="sect_5.1.1" xrefstyle="select: label"/> on the structure and use of the templateId.</para>
        <note>
          <para xml:id="para_2655eb57-bc25-401f-bfa8-2f184fdb2f4b">The IHE MRRT profile defines a "dcterms.identifier" that may be used for this templateId.</para>
        </note>
      </section>
      <section label="8.1.2" status="3" xml:id="sect_8.1.2">
        <title>title</title>
        <para xml:id="para_bf0d6068-67ae-4be0-b21c-b3fa585da78d">The title may include the title of the report template used.</para>
        <note>
          <para xml:id="para_d38f61a0-e8fd-47a2-ada5-3ec1bc4a8175">The IHE MRRT profile defines a "dcterms.title" that may be used in this element.</para>
        </note>
      </section>
      <section label="8.1.3" status="3" xml:id="sect_8.1.3">
        <title>effectiveTime</title>
        <para xml:id="para_843e5067-8419-49bf-a7f9-0da819827cc7">The effectiveTime signifies the document creation time, when the document first came into being. Where the CDA document is a transform from an original document in some other format, the ClinicalDocument.effectiveTime is the time the original document is created. The time when the transform occurred is not represented in CDA.</para>
      </section>
      <section label="8.1.4" status="3" xml:id="sect_8.1.4">
        <title>setID and versionNumber</title>
        <para xml:id="para_3d73685e-dda9-4c96-9f4c-4d02263ec2c2">The setID and versionNumber elements may be used by the document creation system to manage document revisions, in accordance with the CDA specification sections 4.2.1.7 and 4.2.1.8.</para>
        <para xml:id="para_121913b8-924a-4344-ba33-8f126baf4900">COND: If and only if the setID element is present, the versionNumber element SHALL be present.</para>
      </section>
      <section label="8.1.5" status="3" xml:id="sect_8.1.5">
        <title>recordTarget/patientRole</title>
        <para xml:id="para_83d09e33-ee7a-4198-8286-11d766e0b6e8">The recordTarget records the patient whose health information is described by the clinical document; it must contain at least one patientRole element.</para>
        <para xml:id="para_7fd3670f-1b2f-42f6-b996-4ed3d581634f">Multiple recordTarget elements should be used only in the case of conjoined twins/triplets who are the subject of a single imaging procedure, or for special cases (e.g., pre-natal surgery, where a medical record has been established for the fetus).</para>
        <example label="8.1.5-1" xml:id="example_8.1.5-1">
          <title>Header example</title>
          <programlisting><![CDATA[<typeId root="2.16.840.1.113883.1.3" extension="POCD_HD000040"/>
<!-- DICOM Imaging Report Template -->
<templateId root="1.2.840.10008.9.1"/>
<!-- General Header Template -->
<templateId root="1.2.840.10008.9.20"/>
<id extension="999021" root="2.16.840.1.113883.19"/>
<code codeSystem="2.16.840.1.113883.6.1"
    codeSystemName="LOINC" code="18748-4"
    displayName="Diagnostic Imaging Report"/>
<title>Radiology Report</title>
<effectiveTime value="20150329171504+0500"/>
<confidentialityCode code="N" codeSystem="2.16.840.1.113883.5.25"/>
<languageCode code="en-US" codeSystem="2.16.840.1.113883.6.121"/>
<setId extension="111199021" root="2.16.840.1.113883.19"/>
<versionNumber value="1"/>]]></programlisting>
      </example>
      </section>
      <section label="8.1.6" status="3" xml:id="sect_8.1.6">
        <title>legalAuthenticator</title>
        <para xml:id="para_f94df5f3-0d92-482d-9343-79c760e7ead9">The legalAuthenticator identifies the single person legally responsible for the correctness of the content of the document and SHALL be present if the document has been legally authenticated. In the context of an imaging report, this means the radiologist, cardiologist, or other professional who signed or validated the report.</para>
        <note>
          <para xml:id="para_53d50bdc-c7b2-4020-a935-50daff43731d">Per the CDA Standard, the legal authenticator, if present, must be a person, and the authentication applies to the human-readable narrative in section/text and any renderMultiMedia referenced content. Structured entries and external images referenced through linkHtml are not attested by the legal authentication.</para>
        </note>
        <para xml:id="para_940b6b88-92cd-4d45-9ffd-f3d9b4b0c5ef">Based on local practice, clinical documents may be released before legal authentication. This implies that a clinical document that does not contain this element has not been legally authenticated.</para>
        <para xml:id="para_78606b69-02fa-4b6b-9bbf-a64cb84f35e8">The legalAuthenticator SHALL contain exactly one [1..1] time representing the time of signature.</para>
        <para xml:id="para_958de958-ff2e-461e-9021-fe6ff1cb3083">The legalAuthenticator MAY contain zero or one [0..1] sdtc:signatureText extension element. This provides a textual or multimedia depiction of the signature by which the participant endorses and accepts responsibility for his or her participation in the Act. The element is described in the HL7 CDA Digital Signature Standard.</para>
        <example label="8.1.6-1" xml:id="example_8.1.6-1">
          <title>legalAuthenticator example</title>
          <programlisting><![CDATA[<legalAuthenticator>
    <time value="20050329224411+0500"/>
    <signatureCode code="S"/>
    <assignedEntity>
        <id extension="KP00017" root="2.16.840.1.113883.19"/>
        <addr>
            <streetAddressLine>21 North Ave.</streetAddressLine>
            <city>Burlington</city>
            <state>MA</state>
            <postalCode>02368</postalCode>
            <country>US</country>
        </addr>
        <telecom use="WP" value="tel:(555) 555-1003"/>
        <assignedPerson>
            <name>
                <given>Henry</given>
                <family>Seven</family>
            </name>
        </assignedPerson>
    </assignedEntity>
</legalAuthenticator>]]></programlisting>
      </example>
      </section>
      <section label="8.1.7" status="3" xml:id="sect_8.1.7">
        <title>recordTarget/patientRole/Patient/birthTime</title>
        <para xml:id="para_16a52a85-7f7e-48bf-a2c6-c7ce0e97df90">Patient birthTime SHALL be precise to year, SHOULD be precise to day.</para>
        <example label="8.1.7-1" xml:id="example_8.1.7-1">
          <title>recordTarget example</title>
          <programlisting><![CDATA[<recordTarget>
    <patientRole>
        <id extension="12345" root="2.16.840.1.113883.19"/>
        <!-Example ID using fake assigning authority OID. ->
        <id extension="111-00-1234" root="2.16.840.1.118975.4.1"/>
        <!-Fake Social Security Number using the actual SSN OID. ->
        <addr use="HP">
            <!-HP is "primary home" from codeSystem 2.16.840.1.113883.5.1119 ->
            <streetAddressLine>17 Daws Rd.</streetAddressLine>
            <city>Blue Bell</city>
            <state>MA</state>
            <postalCode>02368</postalCode>
            <country>US</country>
            <!-US is "United States" from ISO 3166-1 Country Codes: 1.0.3166.1 ->
        </addr>
        <telecom value="tel:(781) 555-1212" use="HP"/>
        <!-HP is "primary home" from AddressUse 2.16.840.1.113883.5.1119 ->
        <patient>
            <name use="L">
                <!-L is "Legal" from EntityNameUse 2.16.840.1.113883.5.45 ->
                <prefix>Mr.</prefix>
                <given>Adam</given>
                <given qualifier="CL">Frankie</given>
                <!-CL is "Call me" from EntityNamePartQualifier 2.16.840.1.113883.5.43 ->
                <family>Everyman</family>
            </name>
            <administrativeGenderCode code="M"
                codeSystem="2.16.840.1.113883.5.1" displayName="Male"/>
            <birthTime value="19541125"/>
        </patient>
        <providerOrganization>
            <id root="2.16.840.1.113883.19"/>
            <name>Good Health Clinic</name>
            <telecom use="WP" value="tel:(781) 555-1212"/>
            <addr>
                <streetAddressLine>21 North Ave</streetAddressLine>
                <city>Burlington</city>
                <state>MA</state>
                <postalCode>02368</postalCode>
                <country>US</country>
            </addr>
        </providerOrganization>
    </patientRole>
</recordTarget>]]></programlisting>
      </example>
      </section>
      <section label="8.1.8" status="3" xml:id="sect_8.1.8">
        <title>author/assignedAuthor</title>
        <para xml:id="para_22a78686-872e-4d9c-9ee1-d6080490778c">The author element represents the creator of the clinical document. This template restricts the author to be a person.</para>
        <para xml:id="para_931f857e-ed40-4c60-87df-e5a7b59d795a">Such author SHALL contain exactly one [1..1] time representing the start time of the author's participation in the creation of the content of the clinical document.</para>
        <example label="8.1.8-1" xml:id="example_8.1.8-1">
          <title>Person author example</title>
          <programlisting><![CDATA[<author>
    <time value="20050329224411+0500"/>
    <assignedAuthor>
        <id extension="KP00017" root="2.16.840.1.113883.19.5"/>
        <addr>
            <streetAddressLine>21 North Ave.</streetAddressLine>
            <city>Burlington</city>
            <state>MA</state>
            <postalCode>02368</postalCode>
            <country>US</country>
        </addr>
        <telecom use="WP" value="tel:(555) 555-1003"/>
        <assignedPerson>
            <name>
                <given>Henry</given>
                <family>Seven</family>
            </name>
        </assignedPerson>
    </assignedAuthor>
</author>]]></programlisting>
      </example>
      </section>
      <section label="8.1.9" status="3" xml:id="sect_8.1.9">
        <title>InformationRecipient/intendedRecipient</title>
        <para xml:id="para_9412e33f-ba6a-4569-a1bf-d68c4b18e960">The informationRecipient participation elements record the intended recipients of the information at the time the document is created. An intended recipient may be a person (an informationRecipient entity), with or without an organization affiliation (receivedOrganization scoping entity), or simply an organization. If an organization, the document is expected to be incorporated into an information system of that organization (e.g., the electronic medical record for the patient).</para>
        <example label="8.1.9-1" xml:id="example_8.1.9-1">
          <title>informationRecipient example</title>
          <programlisting><![CDATA[<informationRecipient>
    <intendedRecipient classCode="ASSIGNED">
        <informationRecipient>
            <name>
                <given>Henry</given>
                <family>Seven</family>
            </name>
        </informationRecipient>
        <receivedOrganization>
            <name>Good Health Clinic</name>
        </receivedOrganization>
    </intendedRecipient>
</informationRecipient>]]></programlisting>
      </example>
      </section>
    </section>
    <section label="8.2" status="2" xml:id="sect_8.2">
      <title>Imaging Header</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f20ca597-a4bb-480c-b4ba-01625c2c2132">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8016461b-7284-415c-9fed-3688347c99c6">1.2.840.10008.9.21</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f920427-83e7-4d81-856f-b1edba002fa5">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_944bc657-ffe7-4107-b406-1131883e35b7">Imaging Header Elements</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_afb32cbc-dcac-4cf3-8aa9-82d86fb6bef9">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c6380d0-d7cb-45a4-ba31-80a1c194522c">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff0b2dd1-1207-4469-b8b6-ef04f95e6215">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9399e102-9806-4345-bc6b-696e7d0838e7">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_72351f25-e541-433a-83ff-822f5680ee98">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_75dd1e03-d934-4892-aa90-397f14221381">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_28f014ba-228d-4162-8ced-46dbda6d9cce">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b08afce-b46c-4a51-a396-8ceeed1568bb">CDA Header Elements for imaging reports, including encounter, order, and study context</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e7038dd-1fdc-40f3-9882-269423ff4842">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84606570-4b6e-4e78-891d-533c1767bccb">CDA Header Elements</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73692fd3-6f15-4cf0-baa3-c017c1a418b5">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fea5555-6ccd-4575-acc4-c4552ed492d9">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23346a7d-1671-4b2e-a07e-08a44db0e2bc">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7808d8bd-ad60-4777-b346-c0f10acd5723">sibling node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3505f3bc-4f81-4808-bd27-383aa8d42e2f">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb5ccfe9-401b-4567-9d32-97cd1f2db768">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad00d9c3-91a9-49b5-8e2b-025bd9bb7fc1">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_92eb91c5-fd93-40dc-a4af-638a78b3e774">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_6a635692-5063-483d-9f78-2b83f75f072a"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_dcf5cd25-9e1a-4418-90cc-4c38b744cd10">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4d7705b0-5f40-4ced-84ec-a78b16c6acae">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_39e25f9c-95e2-4bb7-91db-8ba3cd9c5f27">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_39cd42d0-d21b-4d1c-a1f2-bcae57729e1f">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_fd9b2754-7506-4784-aa99-0814e76161a3">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a56c0b63-29fd-4975-87b1-86f3a3831e82">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_62eacf75-b18c-46d7-afd0-37de41178bab">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4f3e459c-d375-451e-8afb-b9361bcfb3f9">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_75cffe67-8332-4e00-b9fc-1dcbce5ce1ce">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e5b08b1-5a2e-4809-a460-baaec61031b7">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa68a774-50c8-4d1b-8cfc-e4ad252e3f28">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9672d8f-f0f6-44ff-8389-421e37344cc0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eac9f03e-3f33-4ed8-8049-45cbbd01815e">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b1a9b697-b837-459e-84b0-8d77aae10b82">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f919d57-aaad-4fa4-9c47-137525e2b5de">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_44173f53-6cdb-4b26-aa2c-1ba4d14ddb75">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de322e78-776c-45d3-9cff-d9ff7a7a9331">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_86a09ed9-2473-44a0-9b76-07f0c8ec8761">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ba3ef46-9294-41f2-b998-e1ef56e57e9c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6a98bec-a981-4cac-823b-5de5bc370440">1.2.840.10008.9.21</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bdace87e-0bf3-49af-a742-51c92b3d2fea">component​Of</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfa0d8a9-7dd6-48b8-af6f-e61596eb8549">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ea06c94-c433-4030-ab24-0e262a5bfb6f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_471e1b6a-447f-43aa-b790-ba3067543b6f">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d72f34f1-1c3a-4a45-939f-d130fe46e490">encompassing​Encounter</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d906aa1-c6c6-4a5f-a6c5-9cd0c6415529">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17ad96bc-fe7c-4856-b9df-67b9e9d4b870">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ab4ae10-ff78-4fa3-b23f-d4fec4e3e461">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f72b2b5-f48d-416e-9250-2aa8fecd31df">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_26ca4f09-4d52-45ae-8b4f-fb71278a10ba">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06a91de9-00e6-45cc-9bf8-234a37644147">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_463555d5-a80f-461c-8da2-3bb45f8c1e3a">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40d7a133-f2cd-4945-90cc-0cb2341fa0c2">Encounter​IDIssuer</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cbcb6a92-5f1e-4a02-8f51-0b8802fcee07">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9304e125-c9bd-4167-933f-5b28eebb6ad8">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_216cb2f3-392f-48d2-8475-1a1e4fa98384">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c595ae20-5c09-48a0-b683-72b31361d966">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aed778cc-f367-4e40-adb8-d2175389d3e7">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6456659c-ddbd-48da-915d-54431243d063">
                <emphasis role="italic">Issuer of Admission ID Sequence (0038;0014) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
              <para xml:id="para_8d4d2da5-25bc-4069-8dc7-6926acf51efb">
                <emphasis role="italic">Visit Number PV1-19.4.2</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_203f187d-6dd8-4791-b686-dde0386664d0">Encounter​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3066ae6-916e-45e5-9cef-b839d403d65c">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ecf10e87-b810-47d9-9f09-21479cc68c7f">@extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b65b7a85-a1cd-4bbc-b9a2-a9ae666568c4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e930401-abc6-409c-a16b-073d77049f7e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_105771c0-edb9-4b50-89f7-058516e116ef">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e39f5562-e2b1-48be-8f27-cb134346ceec">
                <emphasis role="italic">Admission Id (0038,0010)</emphasis>
              </para>
              <para xml:id="para_57e3c5c8-94c2-4db9-8811-cc631c74e060">
                <emphasis role="italic">Visit Number PV1-19.1</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5f55499-e16a-47b7-88d7-23b42a611d8c">Encounter​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce5d9c95-d5c3-4c0b-b014-58c56c60d029">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f181e5d-47b0-4533-a29e-91181f65296e">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4af51216-f4ea-4c0d-ad0c-266d1aa06edc">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ff54670-c07a-4ef2-8c37-5444dd38042c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d48eedb9-95cd-46c3-bce0-3709b1412d73">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c471ceea-33c5-4c5b-97f2-46614dc1fe2f">location</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b183012b-6213-4008-9b90-c9b7a584d645">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87a8420d-1346-4f60-95de-e915a557087c">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bea91ff1-0d01-428e-a8a5-5a381afd0371">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e5da60f-2e0e-4f70-931b-8f63cfa9d54d">health​Care​Facility</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1125fe71-5f50-4857-8f01-00e836c785ea">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_76fc7106-be86-469e-8326-6fe7a18831ca">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9f6143d-adc4-448f-b316-6a8342fb458f">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4db908c-f7ab-4d60-9c3d-299cd171a81e">location</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aff92b6f-52ec-4a37-8729-f044ddb98d16">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00fc69d3-eb33-42b4-9983-ace7f7d7580e">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60018df6-a0be-4d6b-9b93-18ef5aeed2a5">Healthcare​Facility​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_780b7afc-e77e-4508-88e0-d218228faecd">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dfc29a1f-15f3-4c4c-96b3-681380251d14">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4f9ae89-3d2a-4e61-a680-171ce39966b0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5fc6ada7-b1c1-4d2c-a0d8-f62f1228fa1f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d027af8-32eb-44a8-899b-1145c4f3b840">EN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4fb09c84-4afc-45d6-88f6-7021a5acc763">Healthcare​Facility​Address</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_78abfc89-d331-4856-97ef-395366747f49">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_afd220f0-9840-4c3f-a1d4-4af4e7a31a9e">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c63cbde6-d6f5-4106-8d48-0fb673507153">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e331ae15-f67d-494d-868a-7dcb5ac9f453">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_391b57d4-1bef-4680-8275-0a0df8f1ea36">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f39700d3-0f80-4912-998a-1da70843f0e9">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0776e68-0cc4-42d1-bd12-c09ea8eb85b8">service​Provider​Organization</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6dd55fb-6f04-4d2d-a8ad-5f5be9b34ed3">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d81f4997-1fda-452e-bbfd-868f91f8bdeb">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_10199d49-eb31-456b-9f63-789dad1c742f">Healthcare​Provider​Organization​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d62e0d9-0dfd-4048-9c8c-ee9ddd84ff6e">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e9aa1ca-14e3-45da-9083-7118041f1f19">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a964d192-389a-404f-aa37-f786ade3ddb9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fcb28974-e602-4afa-b71e-a82e29d9f5c4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_af2f9804-8439-4051-bda3-721d0926b2c5">ON</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ac1e106-e2e0-4d06-a207-1218906725d1">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_34e1f5d3-5c03-4a8d-a601-85c727bc68ac">encounter​Participant</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc220cec-9696-462f-b783-3493df625167">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d126c1b9-49db-4c48-a0cc-8d32b0faf47c">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_590af682-b9a6-4067-bd39-c9d9fefabbfd">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_447cf5b6-83ac-4019-897b-59ea6d340094">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_266c737c-e740-4b09-9075-e3b94932c4dc">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5fabfb55-054e-48f5-92e1-29b0316c5af7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47b0227d-5a0f-4479-a679-fa25a9fa40b3">ATND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_686b641a-c3af-48d0-ba0c-9858aa26a2a1">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e678465-2d6d-4efd-81c8-59f41e8bfaef">assigned​Entity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d30547c4-e216-4f4d-82ca-b5c709f39fa6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d302803b-bd8b-4e6d-8709-ea626dae9f0a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b04a7dd-d822-4428-ac45-03f55d307109">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15b3d0cd-d8c2-4121-b780-4d430b7af840">assigned​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3de17ebf-5555-4133-9a95-29b72f208fb3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6cc3b7ba-6e2e-4da3-9205-574cd1523bcd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_82a29cf6-8bd1-452f-a349-1f255de2e9de">Attending​Physician​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_adea4189-da4c-4205-becb-ebed33eae788">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f8a3993c-aecb-4304-9ecc-c6169ac3488a">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0bbab4fd-08c0-45b2-a306-c607a2184ea5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7eabc8b-77c5-4c5c-8876-2bed5f9dff34">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22f15634-1016-4af1-b773-1fafa6f6864e">EN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2727b859-452a-45f3-bef3-cdbdb18fd8eb">in​Fulfillment​Of</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_453731bc-7d77-4600-8a99-f0180bcb1610">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c40f2661-3087-4d52-a961-4bb5f37762ac">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7a183f4-b499-4c7c-b538-ea73a182c4f9">
                <emphasis role="bold">Order[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_69f36478-cf7a-4494-948b-0d2e53cdd9d1">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c9e90e0-0503-4120-8556-950f864e1270">order</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0ffe3fa-ad41-431d-b329-b68bfe3fead2">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_acb329ec-d2e0-4680-9cab-79b00c86a3ac">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e48b7bc3-fd18-48c4-9c7c-3bb8442c4f02">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_afabb2d9-fa20-4fde-817f-100ac12b8764">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f15feec-bb91-4ffa-b493-085fd763c794">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9294e1ae-3c4f-40d0-b161-9d731bcf7545">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c61547d2-3502-4b05-89d9-730767787876">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e03f50dc-9d23-4bd9-97c1-ae83eecd8686">Order​Assigning​Authority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e38b9b56-9f82-4f56-951e-6c2b607eb151">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4517385e-307d-4071-8315-5eb8a9b2a160">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b8aeca59-dd01-4016-a59d-66e4f81f5fdc">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e0b595b-f76d-422c-83a6-7982411a7ae4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e2dd8d79-421a-4532-af9a-7d08ae66ccdf">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ba1b305-13b2-4417-b276-0f4ff64973fa">
                <emphasis role="italic">Order Placer Identifier Sequence (0040,0026) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
              <para xml:id="para_49f78395-1741-4d10-b75b-d196f8481aa1">
                <emphasis role="italic">Placer Order Number OBR-2.3</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_645585fd-e1e3-4e06-b9be-8ef3a9906cee">Order​Placer​Number</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1d3a058-636a-4abb-98c9-3ef4fcef4bc3">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_65d13dcb-5e66-4ff2-8677-80762462319a">@extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e8b479c-2aa0-4364-9db4-cc68afd5626e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d5e7ff3-4e56-4285-a981-2dfcc277a8a7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6c59ee99-3544-4eaf-a606-0bf2c7e19183">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89f2c3f9-2d25-4ec1-aba0-d1ca9f2fee95">
                <emphasis role="italic">Placer Order Number/Imaging Service Request (0040,2016)</emphasis>
              </para>
              <para xml:id="para_8d9e8f50-9a79-4dfb-9fa5-c44e692fb9b0">
                <emphasis role="italic">Placer Order Number OBR-2.1</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50f301b9-956b-496f-91bc-08526c749fd6">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c0e356e-0b29-4c7a-b1ca-d4935a19706a">ps3-20:accessionNumber</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d5ab92b7-b92a-4940-8cdc-e7036be70eba">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da0c52c1-6fe8-4f0e-b24f-bca4bda6dd01">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_218b5eab-a886-4cb9-b82d-24e8f70fc2c7">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_703aecb2-8b6c-47c2-bea9-27253814f666">Accession​Assigning​Authority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a328f0b1-ad58-4b96-9399-af1c97ee5339">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83b1ca67-1021-451a-a6fe-2c21f9d6d287">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_260dd833-0443-4f32-9ce2-76f250622104">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf54c0e2-53b3-411a-855f-b867a75ea181">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1dd26383-3536-4b6c-afc2-90f461cd9048">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f850ae1-5442-4569-a8d5-51f185daeabb">
                <emphasis role="italic">Issuer of Accession Number Sequence (0008,0051) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
              <para xml:id="para_f8beb62a-6483-4144-9822-e76c13b82524">
                <emphasis role="italic">Filler Order Number OBR-2.3</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f436c908-7835-4b70-8dcb-03c748397175">Accession​Number</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf661c34-6b83-42e3-a566-6b1106cdcc99">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6af694b9-6fbf-4eb5-a7f8-5ee37edee994">@extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6a533520-4cca-47cb-8ce4-8d4aec535ef4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7dd19b8-9ee8-4215-8272-01db806b1a7f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9302069-0258-4e8c-9dd5-da33cf45f00e">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4c2e132-1922-4fcd-9ee7-3ce6e02bf381">
                <emphasis role="italic">Accession Number (0008,0050)</emphasis>
              </para>
              <para xml:id="para_c4cbf4e1-94f3-42e6-b6cf-23d100c451d8">
                <emphasis role="italic">Filler Order Number OBR-2.1</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07cb1b0d-74d0-4778-aad4-c8871f1c003a">Ordered​Procedure​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_005b1c4f-4cea-4385-8b21-9e4915ba0a52">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed84ddb0-b90f-496d-804f-fb31eee1bb21">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c6f25ef-4465-471d-a04f-b6c2afd190c3">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e94f7b88-19e8-4eeb-948b-25530d9563e8">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d19f4238-6d23-41ff-bfa3-3945cc2c0afd">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9a4ef39-784b-49ae-89c2-a6324f6a20fa">
                <emphasis role="italic">Requested Procedure Code Sequence (0032,1064)</emphasis>
              </para>
              <para xml:id="para_b102afaa-63f0-4d96-963e-e81ea905a88d">
                <emphasis role="italic">Universal Service ID OBR-4</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2caa93c4-1233-4789-b5dc-0555277a89f6">Order​Priority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c21ca5cd-7594-4436-acbd-447fb158996a">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7135c61e-7b06-48bd-be60-90e83138148b">priority​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f62ca495-ba1c-4c60-9d68-28c6f34dc5ab">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_743cc95e-cfc6-48a9-bb51-c9c9565faf7d">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe9bc372-1531-479d-9f34-0cf47e6b0266">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7ebdfdf-d73e-4225-89bd-0a9112bdeecd">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.1" xrefstyle="select: title"/> 2.16.840.1.113883.11.16866</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4472d212-23f6-4712-a7d5-a2e9e0079baa">documentation​Of</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bc60ef4e-ae63-4a1f-b9ce-1f31ba8bf693">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b332014b-18a9-4f26-8362-a0698473d4cd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d08af2d8-93dc-4922-9038-a69be8c07af6">
                <emphasis role="bold">Study[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36234209-01cc-476d-9022-a032f0da71e3">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a982420e-971c-4a07-ad17-fc808e277d18">service​Event</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3acd504c-5274-4a19-8664-18cb15fd163f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0fed612-4ad1-4fcd-9d5d-da3f3e797ef1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e97ff6c-290c-4bcc-9edf-9c4479eeb597">Study​UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d2a8287-c3ce-4b04-a120-a8c6f7043fb5">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_609200d6-2864-4001-8a38-4bdf14cabd93">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22f9a990-b553-446f-895e-80a0dbd38c00">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4fd3975f-6902-48e7-8dca-b433b6b1eed9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5fb906d-1371-4b01-91a8-cf95dedeb290">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7feb0ef3-7d76-48aa-967b-78cae0dbd255">
                <emphasis role="italic">Study Instance UID (0020,000D)</emphasis>
              </para>
              <para xml:id="para_0da199a7-e83d-4908-8b1e-136c8abe7668">
                <emphasis role="italic">Study Instance UID IPC-3</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e8d1db9-06c1-4590-98de-2cbbaf4f8c19">Procedure​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f2c46f5-c485-42d3-befb-af25d767b770">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1355274e-f05d-4c72-8d66-439de0f47fbe">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aad07047-ede9-40da-8794-df8ca27f84b9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_09d73831-9745-45d7-92d1-3f9ceee8426c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff796e09-51ac-4c3a-95ca-56a841eb2e44">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08dd5f8f-c9b9-473e-999d-9fe7e0e5ea6f">
                <emphasis role="italic">Procedure Code Sequence (0008,1032)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_10156f58-cf60-4185-92d9-d86bed847bb2">Modality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97241d21-dacf-4f87-9784-dd2fe08fad8b">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81c0b199-30d3-4a10-ba4c-ade65158271b">translation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_945f37d9-1f10-4733-8804-b3274139fde0">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_386b2a87-f0cb-4d1b-a21e-411c13a44b5c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e1ec81e5-833a-4d64-8437-9ffc1afd4ae7">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7fd4a7c3-c671-4893-b3fb-b12f002a752b">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4672b6a-7ed4-4ba5-9154-aea39456ec3e">
                <emphasis role="italic">Modality (0008,0060)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07bdee42-4e83-41c3-b7b0-2647d39f2026">Anatomic​Region​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4491bbd5-6060-4de1-8873-46fe133a7a28">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_daca5b15-eda0-4e85-8cf0-fad4d0bc2876">translation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98223364-cedc-4079-b331-3ee3cd5096be">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aca505d5-6e80-4cc3-af28-faa3bb0e87f9">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_27a4cd1f-03a4-4c74-857f-ac6d86826027">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_be7f45e8-caef-4399-99be-fa8b344e60f4">Concept​Domain AnatomicRegion</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_14f5503f-efcd-46b3-b1da-157b6fbbf241">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_818c5152-66db-437a-8528-953432059f2d">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f797babd-1ec7-4407-a962-c38849ada319">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22fbb43e-2ef4-455a-9800-fb42b4239588">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19a1f9dd-7a74-4caa-b1cc-29938272ddd6">IVL &lt;TS&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4544ae65-6a47-4fd5-9536-24c95fe30d57">Study​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4c63f02-5836-4f34-9c51-2975fdc5c844">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06144975-942b-4ac7-95f4-204d94d5bece">low</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0943d94-7b41-4e93-9b11-3741a3de9fc9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cb05ac19-1fb1-47b1-85d5-79ec97e0df0d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00d4b77b-9a03-4221-96f5-571f9ee0d899">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab6ed0d5-0a48-4363-9037-415e58b47fcb">
                <emphasis role="italic">Study Date (0008,0020) + Study Time (0008,0030) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
              <para xml:id="para_479f4925-f63e-40c7-bfa5-2c08342626d3">
                <emphasis role="italic">Observation Date/Time OBR-7</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_daa41140-26af-4787-b5c5-166b64ef8f75">
                <emphasis role="bold">Performer[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c8fdce0-8548-4122-83e3-b6f0de561e23">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_416c3742-1571-4936-bdf7-1983cfcf5c44">performer</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a61ed7f5-f7f1-44db-acd4-449f0596d443">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3ec791f6-cc41-4200-9e66-dff0e3668ee1">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_664302d8-8fec-4943-9832-fd433ef07bfa">Type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d7e2627-376b-46e1-a8c6-50a017e4529d">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6897e43-2988-4789-a89d-b3acc04f43ec">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_04e1d861-6ecc-4a20-bae7-5f6568216d66">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4960b78c-1b80-4fae-b69d-dd400a58669c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e3e241b-7198-4832-b3a3-e5713555fea0">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a91ee4a8-e4de-4452-b500-09d6037dc2dd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_327ec4e4-207c-4fb6-8075-048e26d7fcdc">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.7" xrefstyle="select: title"/> 2.16.840.1.113883.11.19601</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_90786a57-3dc5-4c5e-a99e-d76ae11121ee">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4a536e06-26c9-4378-9516-9ef50596712d">assigned​Entity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a5db1d0-ea68-4120-bbfb-31658e6ca05f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21b0840d-3d5a-463a-9c66-3e8d9251ec80">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_78fca9b2-4344-4e87-b772-66840d41eaab">ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2cd4e755-00d3-4b90-896b-34605f3912ea">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cab6889-2549-417b-a42c-2181dbe91363">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7fdd4db9-4ea2-4a67-ac1b-22b4a75b56aa">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a1d44277-4d24-412f-931e-154c1fde7cb9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cbf61c16-9d4b-44fd-8024-63c923fbf802">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_004de78b-9c31-4fde-984b-34d78cb7c194">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e25c3761-1041-4f05-8428-9455b768e8c7">assigned​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fea6148c-f782-4987-9915-ea81c850110d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20009162-eac6-44bf-943a-4fa82797ba24">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2a8f37c-2c00-4350-8675-ce86cd3320a0">Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_054a3fe7-446b-4d67-815d-261f1bd54074">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0132ed13-155f-4579-8c8b-a7603c598761">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e77f991e-7ef6-4872-9aae-cd5e6a703b0e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7277b6c-a3fd-4443-bed0-4a5644905327">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a51c6f34-531f-4879-a4b7-0083dac20f30">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7034f2de-858d-4e14-9173-06bf718ea68a">participant</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_585c2225-4e9b-47d1-94af-c8fe60358786">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1aa084d5-aae4-40e8-a01a-d994c9aba208">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_11624db2-9631-4f70-a01c-172fed4101d9">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_edf5a9fe-fda1-4a6b-87f8-9ab2e357e950">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3fa6b971-4918-4e99-931e-eac2d6cb35f4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00efef01-e5d8-4b19-8126-5e2a5e027381">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7886055c-2ded-42f6-a403-b9909f1a02bf">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb172f05-b3cb-4325-9cbf-5c842ddee701">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0edcc8ca-96b6-4890-a37d-48fa5c5e61c8">REF</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6929f57f-79f8-4722-b7b4-346846ec91fe">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21642da2-7606-4d94-bdab-e4151d7cc1ec">associated​Entity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43a5b204-d747-44d4-a0b3-9664d30a5542">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0481a45d-1481-4c07-9503-ad02baf4d068">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4073593d-db60-4699-95d4-a8ec2f2e87fa">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f090240-6512-4b35-8395-e6c52c8188ba">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_56badc0c-9fd9-496b-a30c-2b354c322e9c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b57aa60-1609-471c-8b4a-12d45705b98b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_988e8557-28dc-4eaf-b064-e30f9a851f0e">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5920a993-6fab-47f9-9c31-96e4dc492b53">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1afdf9f0-2e98-4b43-975e-3bc731634c18">PROV</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ca86333-ed25-491c-a367-4d210f5307f6">Referrer​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53a43b91-d52a-45bc-bd56-958e6053ffd9">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7351acd-6188-400f-94e9-fb7314c6484d">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a568e6f1-d473-4c2e-b65a-86ff210145c3">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e081166-6627-40cb-895d-35a7dab54a7d">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_952d57db-f88a-4fc9-a6a6-6d4e94e596d3">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_424fb1b8-d0df-4b39-9811-de4ba83e9c84">
                <emphasis role="italic">Ordering Provider ORC-12.1 + ORC-12.9.2</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1837fc47-99f4-448f-a52e-b6f988b30a5c">Referrer​Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc1bffaf-6fdc-4777-9a1a-183570eab011">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe5b5d17-1336-451a-8022-4bae03a346ff">addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29e9e7ee-6745-4b70-a157-9423d6e894e6">0.*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2baea8b7-5b32-49ac-9c8e-e4cce8a801d0">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c38be24-2a49-40bd-b0e5-034445cc9f6e">AD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c81a296b-9ee9-4cdc-8b99-112981742711">
                <emphasis role="italic">Ordering Provider Address ORC-24</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aea7f8b4-44c7-4934-9de3-285d68557ecb">Referrer​Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dedb01e1-5d3a-4e15-bb1a-a3c3ed58f488">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e665e73-4a4d-4620-8cc0-6b97f90d8e1b">telecom</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_10a0218d-f8ee-4fc0-813c-477aa93938e2">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a58b1a4f-8977-4885-9671-a13b5ce1d03e">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f558a765-8173-49a6-bc7a-f40dac78d018">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_256d8b1b-a0a9-4f8e-8718-2fcfbbf05acf">
                <emphasis role="italic">Call Back Phone Number ORC-14</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f8383092-c701-46ec-93a7-91a34f986770">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d11b1c0-f990-4203-be98-79936f9d386b">associated​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d86a4557-7bc4-47bf-a745-ba77f88671e7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_311380c5-8024-48ff-a2c2-0b4566146c62">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b8d5f39-20d3-4fb7-9e64-cbe03ffc1b7e">Referrer​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b4e3c90-6d28-4dfb-a5ed-8c636eb56bb3">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7620e9e5-a409-453b-8f2d-df1ad29e9ced">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ceff863-5300-4e40-8ca2-19342ff78294">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_004d84ff-3cd4-4a24-8d8e-5453b0e6cd48">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa144f4f-d593-4345-9fcf-a11cd74761d8">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0905184-4eb2-434a-9282-4f2d5b8739ad">
                <emphasis role="italic">Referring Physician's Name (0008,0090)</emphasis>
              </para>
              <para xml:id="para_6fa221e7-dc63-4353-9c66-2867206a940c">
                <emphasis role="italic">Ordering Provider ORC-12</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41967d93-02dc-4ea4-97e5-f07180d5be10">data​Enterer</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1fd9c4c5-ecff-4ff3-aaac-03b1bc2b2535">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b9b7ce7-ba08-4d07-b800-b5536d89e482">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ab1dda8-cb73-4fb7-a134-dce0657be56a">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_59d623ec-0ccf-471a-82dc-4314aa8a1217">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4217683b-38c4-46fa-a5de-ca0b4d6a45c4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8186b0d3-09e6-4512-995d-7c5497b535d1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ceae75e1-05c4-42f1-b8ea-d39205be07ee">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58dd3711-34f4-434b-b8d0-1bf6d7ab473c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54f6d988-1a05-4d4e-b3cf-a071dcc3f934">ENT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe11c3ba-2980-4e37-9371-35a1e10b2616">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb4e86f9-40f7-433d-81cf-207dbbdce294">assigned​Entity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c843d4c4-c890-4e06-b043-968dccd8832c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c4fd644e-ef17-4b2d-9a53-870779271b63">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53628985-bf7b-42e8-a543-d90d06ed3e6c">Transcriptionist​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7ba457a-1ce1-4074-950d-cca4d51f87a5">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a639c83-b9c2-4e69-8174-f5ee4b790ff4">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd0d9299-7e98-4169-96b7-154c65c3c58f">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23b01492-2018-49be-9e21-1cd5fd3c7cfe">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_26cb7433-c559-45a8-8e5d-ecb275382c6b">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6347916-7c99-448b-8711-dc0e9d144460">
                <emphasis role="italic">Transcriptionist OBR-35.1.1</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_02d37bfa-1890-4e82-93ca-57049d4ed1db">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c994cc4d-738c-4d2b-839c-8ecca26abb62">assigned​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2f4eaf56-0a4c-4c25-9edf-dd7513a911e7">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_99e51063-feb6-4667-9b6a-ca12ddb4241f">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98d8576f-1514-4ce4-a340-785f088db5ae">Transcriptionist​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8037f72f-2549-4c21-ad89-0b002d8204db">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aefa0979-8139-47b6-b3d9-c609bc0d1445">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_66b658c1-8a1c-435e-a3ed-3bd7313b1d98">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee43d220-dddf-494c-a37d-840e01ccc67c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6af94046-4417-4435-b80e-67c776c12537">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d1a8368-95cf-421f-a71b-20070d1f71a6">
                <emphasis role="italic">Transcriptionist OBR-35.1</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_8b0dc495-e1ee-49f8-ba6e-0ad77c44671b">Note that there is no business name associated with this template. Rather, this template is an editorial convenience for template specification, and the Business Names for the elements of this template are logically part of the Business Name scope of the invoking template.</para>
      <section label="8.2.1" status="3" xml:id="sect_8.2.1">
        <title>componentOf/encompassingEncounter</title>
        <para xml:id="para_421709c7-09e9-498d-a3d7-3ac3293edad2">The id element of the encompassingEncounter represents the identifier for the encounter. When the diagnostic imaging procedure is performed in the context of a hospital stay or an outpatient visit for which there is an Encounter Number, Visit Number, or Admission ID, equivalent to DICOM attribute (0038,0010), that number should be present as the ID of the encompassingEncounter.</para>
        <para xml:id="para_9ed11495-f595-4339-9655-b384f4427044">The effectiveTime of the encompassingEncounter represents the time interval or point in time in which the encounter took place. The encompassing encounter might be that of the hospital or office visit in which the imaging procedure was performed. If the effective time is unknown, a nullFlavor attribute can be used.</para>
        <example label="8.2.1-1" xml:id="example_8.2.1-1">
          <title>componentOf example</title>
          <programlisting><![CDATA[<componentOf>
    <encompassingEncounter>
        <id extension="9937012" root="1.3.6.4.1.4.1.2835.12"/>
        <effectiveTime value="20060828170821"/>
        <encounterParticipant typeCode="ATND">
            <assignedEntity>
                <id extension="4" root="2.16.840.1.113883.19"/>
                <code code="208M00000X"
                    codeSystem="2.16.840.1.113883.6.101"
                    codeSystemName="NUCC"
                    displayName="Hospitalist"/>
                <addr nullFlavor="NI"/>
                <telecom nullFlavor="NI"/>
                <assignedPerson>
                    <name>
                        <prefix>Dr.</prefix>
                        <given>Fay </given>
                        <family>Family</family>
                    </name>
                </assignedPerson>
            </assignedEntity>
        </encounterParticipant>
    </encompassingEncounter>
</componentOf>]]></programlisting>
      </example>
      </section>
      <section label="8.2.2" status="3" xml:id="sect_8.2.2">
        <title>Physician of Record Participant</title>
        <para xml:id="para_cf26a76e-0f14-49a4-927f-0a8c2d8ca691">This encounterParticipant with typeCode="ATND" (Attender) is the attending physician and is usually different from the Physician Reading Study Performer defined in documentationOf/serviceEvent.</para>
        <example label="8.2.2-1" xml:id="example_8.2.2-1">
          <title>Physician of record participant example</title>
          <programlisting><![CDATA[<encounterParticipant typeCode="ATND">
    <assignedEntity>
        <id extension="44444444" root="2.16.840.1.113883.4.6"/>
        <code code="208M00000X"
            codeSystem="2.16.840.1.113883.6.101"
            codeSystemName="NUCC"
            displayName="Hospitalist"/>
        <addr nullFlavor="NI"/>
        <telecom nullFlavor="NI"/>
        <assignedPerson>
            <name>
                <prefix>Dr.</prefix>
                <given>Fay</given>
                <family>Family</family>
            </name>
        </assignedPerson>
    </assignedEntity>
</encounterParticipant>]]></programlisting>
      </example>
      </section>
      <section label="8.2.3" status="3" xml:id="sect_8.2.3">
        <title>inFulfillmentOf/Order and @ID</title>
        <para xml:id="para_7d606592-79c5-4387-9f5a-03100d7b6e03">An inFulfillmentOf element represents the Placer Order. There may be more than one inFulfillmentOf element in the case where a single report is fulfilling multiple orders. There SHALL be one inFulfillmentOf/order for each distinct Order associated with the report.</para>
        <para xml:id="para_988cfce3-2fd1-480c-b0a3-f6c88d3554a9">In each inFulfillmentOf/order there SHALL be one order/id for the Placer Order Number (0040,2016). There SHALL be one order/ps3-20:accessionNumber for the DICOM Accession Number (0008,0050) associated with the order. The ps3-20:accessionNumber SHALL be Data Type II; it SHALL have a UID root attribute identifying its assigning authority, and the DICOM Accession Number SHALL be in the extension attribute.</para>
        <example label="8.2.3-1" xml:id="example_8.2.3-1">
          <title>inFulfillmentOf example</title>
          <programlisting><![CDATA[<xs:schema ...
    xmlns:ps3-20="urn:dicom-org:ps3-20"
    ...
</xs:schema>
<inFulfillmentOf>
    <order>
        <id extension="089-927851" root="2.16.840.1.113883.19.4.33"/>
        <!-- {extension} =
          Placer Order Number/Imaging Service Request (0040,2016) {root} =
          Order Placer Identifier Sequence (0040,0026) > Universal Entity ID (0040,0032) -->
        <ps3-20:accessionNumber extension="10523475" root="2.16.840.1.113883.19.4.27" />
        <!-- {extension}=
          Accession Number (0008,0050) {root} =
          Issuer of Accession Number Sequence (0008,0051) > Universal Entity ID (0040,0032) -->
        <code code="RPID24"
            displayName="CT HEAD WITH IV CONTRAST"
            codeSystem="2.16.840.1.113883.6.256"
            codeSystemName="RadLex Playbook">
        <!-- Ordered Procedure Code is
          Requested Procedure Code Sequence (0032,1064) -->
    </order>
</inFulfillmentOf>]]></programlisting>
      </example>
      </section>
      <section label="8.2.4" status="3" xml:id="sect_8.2.4">
        <title>documentationOf/serviceEvent</title>
        <para xml:id="para_cafab180-21be-4395-b642-d628b34b0d25">Each documentationOf/serviceEvent indicates an imaging procedure that the provider describes and interprets in the content of the report. The main activity being described by this document is both the performance of the imaging procedure and the interpretation of the imaging procedure.</para>
        <para xml:id="para_f1f563c1-e6ea-4c4f-92a5-6987278c8d10">There may be more than one documentationOf/serviceEvent element if the report is interpreting multiple DICOM Studies. There may also be multiple reports for a single DICOM Study.</para>
        <para xml:id="para_09276d98-15ff-4984-8bfb-872b1e9b13c3">The serviceEvent/id element contains the DICOM Study Instance UID.</para>
        <para xml:id="para_1ff3d3c4-f792-488d-bef3-6d5a5f0be395">The date and time of the imaging procedure is indicated in the serviceEvent/effectiveTime element; the date and time of the interpretation is in the clinicalDocument/effectiveTime.</para>
        <note>
          <para xml:id="para_b6992d66-3655-48b7-9704-da20f4e55930">The serviceEvent/effectiveTime uses the IVL_TS data type with the low element required, for harmonization with Consolidated CDA release 1.1.</para>
        </note>
        <section label="8.2.4.1" status="4" xml:id="sect_8.2.4.1">
          <title>code and translation</title>
          <para xml:id="para_dc49a94d-b7c9-4bda-9870-52087353b121">Within each documentationOf element, there is one serviceEvent element. The type of imaging procedure may be further described in the serviceEvent/code element. This guide makes no specific recommendations about the primary vocabulary to use for describing this event, identified as Procedure Code.</para>
          <para xml:id="para_1e1789ed-3954-4886-b905-daf92454b6c9">The serviceEvent/code/translation elements include codes representing the primary image acquisition modality using DICOM (DCM) terminology, and target anatomic region (for which SNOMED terminology is recommended).</para>
          <note>
            <orderedlist>
              <listitem>
                <para xml:id="para_1aaa8dba-ecef-413a-9eb2-48014c6186c3">These codes may be used as health information exchange search metadata in accordance with the IHE Radiology Technical Framework Cross-Enterprise Document Sharing for Imaging (XDS-I) Profile.</para>
              </listitem>
              <listitem>
                <para xml:id="para_5ac84caf-df48-4d6b-9a00-f14a8b1bc1d5">Binding of the Concept Domains ProcedureCode and AnatomicRegion to specific Value Sets may be done in a further profiling of the use of this Template.</para>
              </listitem>
            </orderedlist>
          </note>
          <example label="8.2.4.1-1" xml:id="example_8.2.4.1-1">
            <title>documentationOf example</title>
            <programlisting><![CDATA[<documentationOf>
    <serviceEvent classCode="ACT" moodCode="EVN">
        <!-- study instance UID (0020,000D) -->
        <id root="1.2.840.113619.2.62.994044785528.114289542805"/>
        <!-- code is DICOM (Performed) Procedure Code Seq (0008,1032) -->
        <code code="71020"
            displayName="Radiologic examination, chest, two views, frontal and lateral"
            codeSystem="2.16.840.1.113883.6.12"
            codeSystemName="CPT4">
            <translation code="XR"
                displayName="XR"
                codeSystem="1.2.840.10008.2.16.4"
                codeSystemName="DCM"/>
        </code>
        <!-- translation code is Modality (0008,0060) -->
        <effectiveTime value="20060823222400+0800"/>
    </serviceEvent>
</documentationOf>]]></programlisting>
      </example>
        </section>
        <section label="8.2.4.2" status="4" xml:id="sect_8.2.4.2">
          <title>Performer</title>
          <para xml:id="para_58923d7e-3d03-4049-9820-e6cb52e5fe9e">The documentationOf/serviceEvent may include as a participant the physician reading the study, equivalent to DICOM attribute (0008,1060), and other healthcare professional participants in the procedure (e.g., the surgical performer in an interventional procedure).</para>
          <note>
            <para xml:id="para_22ce6e0a-c099-4f29-b76f-346686b9feba">In simple procedures, the physician reading the study is identified in the Author or LegalAuthenticator participation on the ClinicalDocument, and does not need to be re-identified in this element. The technologist performing the imaging may be identified in this element as a secondary performer, since the interpreting physician is the principal performer responsible for the service event.</para>
          </note>
          <example label="8.2.4.2-1" xml:id="example_8.2.4.2-1">
            <title>Physician reading study performer example</title>
            <programlisting><![CDATA[<performer typeCode="PRF">
    <assignedEntity>
        <id extension="111111111" root="2.16.840.1.113883.4.6"/>
        <code code="2085R0202X"
            codeSystem="2.16.840.1.113883.6.101"
            codeSystemName="NUCC"
            displayName="Diagnostic Radiology"/>
        <addr nullFlavor="NI"/>
        <telecom nullFlavor="NI"/>
        <assignedPerson>
            <name><given>Christine</given><family>Cure</family><suffix>MD</suffix></name>
        </assignedPerson>
    </assignedEntity>
</performer>]]></programlisting>
      </example>
          
          <example label="8.2.4.2-2" xml:id="example_8.2.4.2-2">
            <title>participant example</title>
            <programlisting><![CDATA[<participant typeCode="REF">
    <associatedEntity classCode="PROV">
        <id nullFlavor="NI"/>
        <addr nullFlavor="NI"/>
        <telecom nullFlavor="NI"/>
        <associatedPerson>
            <name><given>Amanda</given><family>Assigned</family><suffix>MD</suffix></name>
        </associatedPerson>
    </associatedEntity>
</participant>]]></programlisting>
          </example>

          <example label="8.2.4.2-3" xml:id="example_8.2.4.2-3">
            <title>dataEnterer example</title>
            <programlisting><![CDATA[<dataEnterer>
    <assignedEntity typeCode="ENT">
        <id root="2.16.840.1.113883.19.5" extension="43252"/>
        <addr>
            <streetAddressLine>21 North Ave.</streetAddressLine>
            <city>Burlington</city>
            <state>MA</state>
            <postalCode>02368</postalCode>
            <country>US</country>
        </addr>
        <telecom use="WP" value="tel:(555) 555-1003"/>
        <assignedPerson>
            <name><given>Henry</given><family>Seven</family></name>
        </assignedPerson>
    </assignedEntity>
</dataEnterer>]]></programlisting>
      </example>
        </section>
      </section>
    </section>
    <section label="8.3" status="2" xml:id="sect_8.3">
      <title>Parent Document</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ca98280-0a10-415b-bd80-f8e76b8b2ae2">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ae318da9-70e7-4407-9bc0-01134d343fdb">1.2.840.10008.9.22</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6be26d47-f48e-4510-ba63-0e9c6f9dca2d">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b91370c1-05be-440b-8258-f43f9223fea5">Parent Document Header Elements</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4284a33c-05ef-426a-b0bf-9ca0d46a3f6b">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23e5af83-5e50-458d-bbef-9fc398aa53ea">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2dfb915f-59a3-4157-8086-5aefc856acc9">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67c897bf-8be7-47e5-8a8f-f886f06dde50">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f796773f-88e6-4ea6-b28c-ea4983230708">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61916dd9-8981-4c95-9ea1-793bdd6e16a9">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_52fefa33-7b19-4023-9d64-fdd666bafc22">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8d52619-5c53-4c37-b55d-604c7b4905e0">CDA Header Elements describing relationship to prior/parent documents</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_12acc265-bd44-4810-b734-44e973a227ac">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7c93d0ca-6b94-4346-be70-4b3dd516fe0a">CDA Header Elements</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06da0862-f263-4ac7-ba9c-6414fefd15c3">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_665f0d1f-1447-4ed8-b6ad-3c6c8a5dc681">Included in all document level templates</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0993c16c-99f9-4c32-b002-cd612186c7ba">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e517cac1-155c-4e39-9fb7-413261d37d54">sibling node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0578606b-98af-4952-9a9f-96fdb6f79175">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ded0d7c-a0b3-4922-ac4c-10ae7edb4f60">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ccc1e3b-7a33-44a3-8509-c3462db24235">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd1fdd80-ca95-4156-89e4-15b743565cfe">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_d638b7d7-c312-48b3-a458-1a3f23f971ba"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9f58c255-6d16-4006-8dbf-38702f2b7bef">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_7fc5d13d-b9b3-49b5-b671-51c01a339d35">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_89ea6432-cedf-40eb-8fae-e18b2eeb1807">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_870937d2-0210-4cee-96ee-6bf257b4bf33">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9d0e8bd4-f5d1-4709-946b-c3b88ab70af4">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_73d04270-b88e-4b3c-98af-f63539387d9a">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_60c4b4f6-5f1a-439d-abf6-4043533e84ed">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_bc3c8064-3b3a-4e91-9c4a-7e4eef02a219">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_39bc4f1c-706a-4b40-866d-bc8beb5839f8">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_45e328b5-71e2-475b-9068-5085e5f393cc">related​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c3260160-5498-4329-9d95-d76ae696339b">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ecb5d04-5882-40e7-bc85-f45a5eed0c58">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7637320-19a9-4d33-8951-ffb3c6ab376b">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a33bea4-d16a-46e4-9fb3-d52a15e1716c">@typecode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_51de3e3e-4d31-4d29-8f26-ec16906447ce">1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6fd1441-9858-4751-96a9-0fd2b3ab7d79">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87a4e9e5-8176-4dcf-a9b6-76c94307b922">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6797721b-3d73-4458-8582-814134131703">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d2605a2c-5067-4c59-abfa-bef14f1b51ea">RPLC</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c3f3505-7102-49ad-8fd2-5a81edc92c00">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f25c6394-e60d-40e6-b9a7-b93832000128">parent​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb4ab043-d2dc-45d0-9259-62f378c18d42">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4c6bbb4-19b2-4718-b6dc-9efef8667c02">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_801f8ad3-798f-4ca9-a9ee-5bd10a8a5cd0">Replaced​Document​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c22a826-0acc-4299-aca9-823e866d0f07">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a7f40f7-a388-4569-9d54-f7c75c313e64">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1d4e4be-6617-436b-bc5f-68d81205f3cd">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b1e0be39-655e-4906-b2eb-d8914b8602b6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a75293ef-8302-4690-acb3-1e034fb0effc">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ce5a382-eb59-4c9e-9429-a44a26229b5d">Replaced​Document​Set​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec6b6551-9c74-4ce2-b26c-1fdf442d5b3f">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36f2d17d-4221-475f-acc9-6f02c8ac4f12">set​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9867c2f8-74bb-448a-b842-3ac0ff08ed20">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2048cd27-5660-41e2-93ac-ef140ec79420">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c7365d9-6c40-4a3f-bed0-2eedabfa775d">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95c6f6af-3ad1-4651-9103-13fe61624377">Replaced​Document​Version</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_86521730-caa5-427c-8e78-9f2d84ae2f0f">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9f09db1-616a-4e6c-87c7-88c740a6a578">version​Number</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d85407af-4c99-44ab-b4e5-9ed754be42c2">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f759cbf4-73fe-44bd-9c84-078abfc61ded">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de26622d-b3a6-4c07-9333-addc44ad973d">INT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e9d41e4-f075-4dc8-a62f-77f00a18961c">related​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_530738ee-b2b6-4aa4-b5fd-99e09e970084">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61ca5310-745c-4282-a925-75c00e0ffd9b">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9c7a6cf-f1c9-4617-9dfc-c782eb781d07">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec0590d9-44f3-4805-a82c-5af5cfaf2e34">@typecode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00fb2a5e-e019-4a14-94db-69703cbf660b">1.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e55d46f9-e89d-4f28-9221-e26876dbf937">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4763d4a-4474-4acc-8cf5-f5f05a6d9b13">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_91bebbaf-3d22-4bd5-9362-0075cfcd9393">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9799f409-ebce-491f-a23f-138e14831fdc">XFRM</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a84858d8-7f43-4641-85ae-5813721ddfc1">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc947edf-b7cb-4b32-acff-f12272d4d238">parent​Document</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c95023f-6038-444c-a666-d3572c3c6031">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9e404992-a0bd-4c89-8dce-24bddce4e806">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_edb2a7f6-6d19-4f25-9a88-ac5ac52880dc">Transformed​Document​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_24c6e2bf-815c-4751-bf35-5bab188f46e5">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce48f8ea-afb9-4f07-b26e-03484845a903">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63ec6a46-4155-4b4a-8042-19d0fb748dd9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a1cb0da4-82bb-46b7-b4fe-83ef02a83630">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_296098fd-db4a-4d68-81e4-b99e7d2ba7c1">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <section label="8.3.1" status="3" xml:id="sect_8.3.1">
        <title>relatedDocument</title>
        <para xml:id="para_fbd4f9e9-5a64-48f2-8313-9fa78fb1ff80">A document may have two types of parent document:</para>
        <itemizedlist>
          <listitem>
<para xml:id="para_df7d7de5-ec0a-44a0-9b11-da48af06806b">A superseded version that the present document wholly replaces (typeCode =RPLC). Documents may go through stages of revision prior to being legally authenticated. Such early stages may be drafts from transcription, those created by residents, or other preliminary versions. Policies not covered by this specification may govern requirements for retention of such earlier versions. Except for forensic purposes, the latest version in a chain of revisions represents the complete and current report.</para>
</listitem>
          <listitem>
<para xml:id="para_91741246-b334-4af2-9ac7-6dc70ba3d596">A source document from which the present document is transformed (typeCode = XFRM). A document may be created by transformation from a DICOM Structured Report (SR) document (see <xref linkend="chapter_C" xrefstyle="template:Annex %n"/>).</para>
</listitem>
        </itemizedlist>
        <para xml:id="para_359ccd7b-15cc-407d-83e2-a308321636bb">The CDA document management vocabulary includes a typeCode APND (append) relationship to a parent document. This relationship type is not supported in this specification; rather, append is effected by creating a replacement document with an <xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/>.</para>
      </section>
      <section label="8.3.2" status="3" xml:id="sect_8.3.2">
        <title>parentDocument/setId and versionNumber</title>
        <para xml:id="para_7fcd7985-3031-46b0-a5ba-35ad4ccf377a">COND: If and only if the setID element is present, the versionNumber element SHALL be present.</para>
        <example label="8.3.2-1" xml:id="example_8.3.2-1">
          <title>relatedDocument example</title>
          <programlisting><![CDATA[<!-- transformation of a DICOM SR -->
<relatedDocument typeCode="XFRM">
    <parentDocument>
        <id root="1.2.840.113619.2.62.994044785528.20060823.200608232232322.9"/>
        <!-- SOP Instance UID (0008,0018) of SR sample document -->
    </parentDocument>
</relatedDocument>]]></programlisting>
      </example>
      </section>
    </section>
  </chapter>
  <chapter label="9" status="1" xml:id="chapter_9">
    <title>Section-level Templates</title>
    <section label="9.1" status="2" xml:id="sect_9.1">
      <title>General Requirements For Sections</title>
      <section label="9.1.1" status="3" xml:id="sect_9.1.1">
        <title>Section Text</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47ba2b2d-ce49-4c8c-915a-dad75b09b008">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54bb6c60-a42c-44ba-9a05-804b3c73f4ba">1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98b9407d-f1d6-4e24-abad-a8e15c8a7781">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_feb90aa9-fa29-4778-b2bd-ff1576bf0c75">Section Text</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6cbfac8e-1a88-4dc3-8c47-988cebd2a670">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_68b49783-823f-41bf-b0ef-4ed62c0685c6">2022/01/16</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7514a568-56c7-4fc4-a473-bd016460c984">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9fcc092-4243-4afe-9523-f0af426778a3">DICOM-20220116</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2036903d-a252-47e2-a3b2-9c032adab7ed">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca5b8ca8-55ca-4897-b56f-4f40630ff1b8">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e09c8861-4f3f-4052-99cc-f747b17ad303">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a9b0f74-729f-42dd-9e4d-a85ab92b2b9e">This template specifies the common set of narrative block markup that may be included in a CDA imaging report section.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bd27c8d8-26a2-4bb2-b1f9-15998b09afb3">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_72319f4c-7d71-4b32-b13e-f728b9169bbc">CDA Element Set</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15408913-f7d5-48fe-b143-4b39d18e8a33">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a03baa54-3f2d-4ac5-80de-4d1cf8703964">Included in all sections</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_37ea1edc-7283-4420-9d7c-c47a81a1778a">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e7b8e60-c1a9-4883-a812-f1a5db7c87dc">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6be3eb34-090b-4158-80ad-c86948afbb5f">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20b5d057-c00a-41f4-b190-4bcfcd3dc869">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a45cdd5d-0e2f-4419-8e9f-3f3d7db8f516">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54267031-d559-4439-8e6f-97da8999bbcc">DICOM-20150324: Initial version</para>
                  <para xml:id="para_2c96b1c1-9c90-46c2-a8d8-947af16bf578">DICOM-20220116: Relax requirement on XML ID</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_3023ba0f-e71e-422c-9823-5a75eb74b880"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_20a3c6b8-6014-4935-aff7-671d75ca1f2a">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_602fac36-e1e9-4efb-9b53-a113a9e69564">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_15325ebb-7484-466d-a3eb-c9ecc6bac018">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_87d31107-f5c6-44a2-8dc0-72ac2ebbe187">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_fe5f4163-371f-426a-ae6f-3202d8c00bac">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3f7e8a50-0fd8-4efb-a456-1a99f47b0c93">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0ad4c292-7421-436c-91ea-be97a8addcce">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_98ea3edc-8649-42b8-ab9d-fb0eae88af05">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_32b1458c-a40c-4b05-b5e0-ddb97d8eed8a">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1ab375ff-33a7-4405-af4a-fb7aa2ffc3db">Text</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a877253-7166-4087-b6ed-39788494901f">text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08c88fce-d332-4f0a-bfdc-fbbbd5685116">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efe3ab61-7cb2-4bd1-bf1b-2e75d918d267">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ad93c57-f120-45aa-b945-e75056b5ea7c">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2c2a390-f473-4886-b759-933c1b936592">
                  <emphasis role="bold">Content[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8e96a58-b3a3-49e4-b658-1dcefd078820">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6240d60d-e4b1-48f8-a47c-d9edd159ddc2">content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_588fd2a1-9aef-427a-a6ae-a17311b5951d">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0a451cd-c45b-4c53-905e-7e51f7255093">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bedd0884-18ae-49c0-9ec8-997c54c2b910">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41eeb8d4-4d5b-468b-bb4c-9cecbe63add4">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0158e44c-e60c-44c3-b12b-e016553a6687">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1aa8f777-914b-4b7d-b930-a16aaf298630">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41181837-55fe-4b19-a3a8-238f3dd7f376">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f8dec74a-91db-445f-8cd9-fe5658471801">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f0da7626-dc9c-4665-b412-ba4088beee4e">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ad1f167-d514-4863-ad7d-ab27f51175e8">[See <xref linkend="sect_5.3.4" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b65b9d35-f359-4452-88ff-4be4c783c645">Style</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83d66713-72c6-4fd6-bb2e-1b93414b290a">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40d9b37f-6168-49ae-a567-0b3675fb4a1a">@styleCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fe913836-b1a7-4bd8-b2d3-aabf4fbda986">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0f6fbab7-134c-446a-9b53-bb2d7153ca9e">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5561e2a-f20d-4c2e-b9b8-7e0250b6ab3f">XML NMTOKENS</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c2648cdc-dcb4-4496-b0e0-01b019bb89e7">
<emphasis role="bold">Int​Ref[*]</emphasis>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e42a32b-0400-441b-a39d-899ea0fbc363">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6abea0b4-5693-4854-bf8c-3f445bab1bb2">link​Html</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3b57ed2-b135-4fc6-b04e-270a2e82de6b">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a55770a8-d012-4258-8560-c53ed55a55cd">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ab0f7e2-f867-4ba1-a0c7-c91ec9c50e2b">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_da89c296-1a53-44dc-8645-08bb248e8e5c">URL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d648f241-2ad4-4615-85b9-3b7615fc87b3">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6071d8e5-0f77-4c6a-822a-32168b795be6">@href</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f72a76bd-e9a0-4452-9173-37f71e76b94a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_df31b7b6-7a1c-4e5f-9609-af3e62e176ca">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8313de47-0d8d-493e-abc1-a8a1c6d3cdf1">ST (URL - XML IDREF)</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d43c50a-38ad-4aff-834c-f6adb9cf38f0">
                  <emphasis role="bold">Graphic​Ref[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_19c39693-191b-49c0-aae7-8a8cffc8c212">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c98fd9a1-e1a7-4aee-aefa-c03e1eadf937">render​Multi​Media</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28d34961-1591-4370-ae63-ce108c39de95">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6e2f3064-d8de-401c-9c24-fbda5aecc089">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85962250-098c-414f-9654-4e5793e3f08c">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dad096f5-39af-4c20-8da1-6a751b65278f">@referencedObject</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa2d7efc-1425-414a-8484-adce44ce94c4">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85e10364-fc34-4ce1-9270-41b6909e7692">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9fdd6adc-9267-4eaa-a09b-17a2f2286da7">XML IDREF</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5c9f6beb-e40d-4e85-8082-8ac41674e403">Caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca777d25-2d28-4aa8-b49b-bb7e55096c4d">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58fc1b0e-ad61-4c84-b5df-6f691acbbbf8">caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c137d472-2e86-4ce4-b433-3bc5aa35ca0d">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61f06564-4483-462b-8dff-0c629e2dd55c">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4990a2a-68b4-433f-a47e-0b716e899d19">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74e01f75-07bc-48e0-826d-c4e97e63a7e1">
                  <emphasis role="bold">Ext​Ref[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5dd9adb9-c9ea-4210-a944-11cd1f53ba76">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cffe598b-c3fe-494e-9fac-3d116ff0b7d3">link​Html</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be6b186b-244a-4dd1-bb01-4361e31de502">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b86e1e4d-e503-465e-ab99-52fe7d2fd1db">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9197f55c-450b-4437-a1ff-c20c6ce747fd">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ff2b24b-9707-412d-a835-f0b521791123">URL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_272c3a11-3585-4e1e-8a77-5e0ae5908878">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56eb9862-aba1-412d-8ff2-287f365faf74">@href</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3e83ffe5-c8f2-489f-9113-1b2ff6f480e4">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccc4d313-5623-47d2-a248-4532791a1a58">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3bf2440-2261-4ca1-a45a-a4ed8546539f">ST (URL)</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_121fa20e-7f53-4fd7-9446-b93ebe1dd42a">
                  <emphasis role="bold">Paragraph[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9e1a765-8ee1-416d-a889-de0d795fecb0">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ead56eb-566d-4800-bb5a-94b6a2949447">paragraph</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5987edcf-ee48-46da-a0ee-3a7a47d38c42">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59c85b32-437a-4586-8345-119dc6c9bf02">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_debf42af-c887-4210-ba2c-dfa0c567e957">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_24ae3bcf-42b1-49d8-a9bb-c85381d40345">Caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_11d8bebc-146c-4860-9ede-c23d65adedd1">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6631ffbd-77ea-4c52-81ff-62b7d91636d5">caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bd94fd98-af91-4870-914d-3f442ebfee29">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0ff54263-8b7d-4cab-a5a4-3dd2c88c61c1">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8cea7d30-aae6-43a2-881e-3598ec054127">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7bf0991-1370-42c0-95a8-cb2d11c394a8">
                  <emphasis role="bold">List[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52d44e60-6a9e-44e7-a7eb-cdd539fc8ed6">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f7c740ff-d119-4cf4-8fde-681b96872db4">list</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9b75ac7-63f4-4255-9f20-f78aabf34a5c">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ce8e1d8-9179-4f26-abe3-bafe13514b77">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_02bca4ba-1c7a-4ea3-a483-005827023075">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb09ed48-c4ae-4d3a-9158-1fbd27ab5a21">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af30661c-002d-40c8-bcf5-160e6462fb34">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fce3da5f-8434-424c-85c6-1799ae2c382f">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5c84843-d60f-4de5-893f-e8642e09cfae">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec695557-92df-47a5-b45b-feb88fe811fc">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1120b199-2adf-4f3c-893a-6f6c953ab7a5">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5175fa1-119a-4609-bdb4-6a9bd1fe904d">[See <xref linkend="sect_5.3.4" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0255f1b5-71f2-42fb-adbb-bb5ce54afca1">Ordered</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_21f1f3e7-8712-46f9-b68b-75d39637c4ab">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_95f97fe5-e9b1-41f4-944c-b6b92d06212d">@listType</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_735ab873-7bdf-4eee-b653-e73a0329c7a5">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_995c0972-7b94-4b1c-8b7d-d67ecf398310">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f92a07f1-ed2b-4711-bad2-8374f18a724b">XML NMTOKENS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b18e5622-34af-40cf-84a7-a35a3d8e4dbe">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b0350d43-c1ce-4dd3-9047-4dcbed2b0e48">ordered</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0860220e-d2cd-447d-9465-b537ec909679">Caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e6bd8048-861e-4c81-a251-edb3fb258255">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_284dc984-ca24-4b53-be8c-c24368e76f06">caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_86df241f-a114-4eae-8936-0d44db2c9980">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_14282ab9-d005-4809-91ae-628a6afe4405">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2617b296-dd8a-4501-abc7-4ed0a402c5eb">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57834fa5-77f2-4b1a-a27c-98e99addc602">Item[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_349d82ab-de62-4f64-a902-895fe0032103">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5595f527-1749-4282-aba4-eb0ad3794b10">item</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4fca2b4-8e5b-466a-ba90-4be569912cb1">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42c7f343-883b-470e-ab28-2356ea10cf5b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85f7c1b7-a1f1-4dbc-a76c-d14cea7ba4cb">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c50a681-207e-4777-98cc-cf7422c71410">*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d4da9a32-9b95-404a-8fdb-2a32ca41e10e">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9ea7c27-5175-414e-93e3-e8aa41b1086d">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_10204a14-e255-4e77-9a87-59f1f7a77e98">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76417039-716c-47e8-b6ca-f6ce1dd75eaa">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c057c9a0-96c2-4a61-8b3c-8412a584289e">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09e09e31-fcf7-4d37-830f-9dc428f699a1">[See <xref linkend="sect_5.3.4" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64d1d3ce-9099-40f5-b0eb-abea340648fe">
                  <emphasis role="bold">Table[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de3f4783-0dbb-4cca-bb94-9e60de66652e">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_383e7b13-2234-4b9d-8fb6-a07f857d1866">table</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5a61a89-d94b-4844-a15e-e2730e634117">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cdf276e8-35d3-46f6-8b3f-a72c18cbd56f">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4dd5bcd9-7892-458b-8cca-ac1d80d527fc">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_baa1baba-52aa-4001-a8f9-fb5a5887189f">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_196b6869-dd05-4fb3-a65a-84782b668476">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f00dd14-f851-4233-b2b0-c67263075910">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4dcb4006-1dac-4137-9c5d-60fcc92f96d4">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1fa0b6ff-122c-4110-ba02-5904ed807879">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdd2a981-15bb-4662-aedd-6c075c362ee0">[See <xref linkend="sect_5.3.4" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2dd27b93-1aae-4e97-9c8b-e7fff1aa793a">Caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59ac45ae-1da5-43a3-8816-75d07930d825">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bfab4853-ce78-42cf-8113-0a9349534f66">caption</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fbe33cb0-7a86-4ac8-ae9a-da18f391163c">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65b61f96-1164-4a5d-8838-b52b865ae721">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51588af0-6489-4dd9-8048-03d7aa90e296">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_00e495c1-8368-4cc1-8c65-8da61303d5c9">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a0a4112-1fe6-43c0-9bdd-397b962df6c2">Tr</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9c067fee-5e18-48a2-b471-ab30b784575d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8f47eb9e-b238-4a6d-94fd-c9bd0e3ab7b6">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_746a8847-77c2-484f-b525-cca4497d72d0">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_913a24ee-dc22-4add-8585-b3f812d843cc">@styleCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6758e9d6-4dda-4c00-831a-2eb5722bded2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d30c4a89-0a65-4d14-ab30-930f56d33585">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f66b67d6-90f7-4500-9923-fa9f6a319acb">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5729497b-6d6a-419f-a626-958492e32932">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6dab116c-582a-42e8-a410-6791dc350881">Bold</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0aa3b689-c14a-45b4-992f-47c2b20b9702">Column​Head[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6ffe5c5-274d-4400-b769-a64b19760660">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b210a07-3d89-4c81-b84f-c84b5473e489">Th</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e0a025b7-b7d9-4ed1-9ddb-6d74eac85e68">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e2a7b23-957a-493d-acd4-4aae85857f95">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6ca62db-9043-411b-9d18-d2f5d92d8d60">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91276a81-2977-4760-8be0-0a1a2e929364">
                  <emphasis role="bold">Row[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a648761-e353-4d67-934f-3c06e06d31af">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_33cd0ee7-2770-4c33-a4a3-c9b7f6a630a2">Tr</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15fb246a-615a-4bb1-bc5e-91f6b25a0850">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bdd158ca-19d4-4732-bc38-824c4177a4c2">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_746304d2-df97-44b1-bfd0-6b8c7431ce10">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8ce1fd0-8127-4861-b8c7-35afbcdd681e">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad7800bb-6ddf-441e-9f22-eb716313ef76">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d335fb9b-2fe1-46ed-b54d-b733da2726c4">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70b97022-f88f-4124-84e8-09cf634524e5">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b04946e9-3a8a-4536-90e6-b6b01a42f6c8">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fb274897-5d96-48b4-a0e0-1c9d6029cf2a">[See <xref linkend="sect_5.3.4" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9562ce5-2c3e-4ed1-aecd-0f99ecea2033">Cell[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a2648b8-4b3b-4d54-8fa6-046f035c6723">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa207b06-95ac-446a-a4b1-7a907d9dd04b">Td</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5946adf-bb71-4a0a-b02f-2fa462ca4e0c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_feee6923-6617-45fd-a6c6-a6bd61facb68">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d8ae86be-3788-4ac0-a8ba-0c10accf1ce2">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </informaltable>
        <para xml:id="para_654a6076-1460-474c-87fa-0368bb4bf684">The text element within the section stores the narrative to be rendered, as described in the CDA R2 specification, and is referred to as the CDA narrative block.</para>
        <para xml:id="para_0b81c113-ad23-4639-8858-6521b9b46e27">COND: The text element SHALL be present if the section content is not completely represented by subsections.</para>
        <para xml:id="para_1a44d46d-dee3-4502-a824-b7a5be2ae36f">As noted in the CDA R2 specification, the document originator is responsible for ensuring that the narrative block contains the complete, human readable, attested content of the section. Structured entries support computer processing and computation, and are not a replacement for the attestable, human-readable content of the CDA narrative block.</para>
        <para xml:id="para_7d1f6ec5-640e-43c3-a5be-b1166b4d86ee">Additional specification information for the CDA narrative block can be found in the CDA R2 specification in sections 1.2.1, 1.2.3, 1.3, 1.3.1, 1.3.2, 4.3.4.2, and 6.</para>
        <para xml:id="para_506e4e07-ebbd-4772-b0cb-d75927f44e7f">The narrative block allows a variety of markup. The markup that implements various types of internal and external linkage is shown in the table, and is included in the conformance specifications for each section narrative block that invokes this template. The markup elements may occur in any order and at any point within the narrative block text as allowed by the CDA R2 specification.</para>
        <section label="9.1.1.1" status="4" xml:id="sect_9.1.1.1">
          <title>&lt;content&gt; Markup and Links From Entries</title>
          <para xml:id="para_fd7e8ccd-4494-4f2c-b336-355ddc0d41f4">The CDA narrative block may contain the &lt;content&gt; markup element to wrap a block of text so that it can be explicitly identified using its XML ID attribute, and referenced from elsewhere in the document. Specifically, structured entries may link to their equivalent narrative rendering in a content block using the XML ID (see CDA R2 Specification, section 4.3.5.1).</para>
          <para xml:id="para_2564602f-ca12-4d07-b1ff-c012caaccb0b">COND: The XML ID attribute SHALL be present if the content is referenced from elsewhere in the document.</para>
          <para xml:id="para_d1519fd8-b33d-494a-b0a7-d8f888ba7426">Additionally, a content block may include a styleCode attribute to suggest rendering (see CDA R2 Specification, section 4.3.5.11). For example, Bold could also be used to highlight actionable findings in the text of the <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> and/or <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> sections.</para>
        </section>
        <section label="9.1.1.2" status="4" xml:id="sect_9.1.1.2">
          <title>&lt;linkHtml&gt; Markup and Internal References</title>
          <para xml:id="para_96d2f0ed-f035-4bcb-b082-39de9c708ab6">The CDA narrative block MAY contain the &lt;linkHtml&gt; markup to provide a link between narrative text in one section and a content block in another section (see CDA R2 specification section 4.3.5.2). The XML ID of the target content block is used in the linkHtml href attribute, with a prefixed '#' to indicate the reference is in the current document.</para>
          <para xml:id="para_c188b3cd-4d38-4421-9cb4-ecfc26cb60df">For example, a linkHtml reference could be used to link an actionable finding in the <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> section to the specific, detailed measurement evidencing a problem that was identified in the text of the <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> section.</para>
        </section>
        <section label="9.1.1.3" status="4" xml:id="sect_9.1.1.3">
          <title>&lt;renderMultiMedia&gt; Markup and Graphical Content</title>
          <para xml:id="para_ba402049-89c0-4b5c-b217-b8757114690a">The CDA narrative block MAY contain the &lt;renderMultiMedia&gt; markup element to include graphical content,
              e.g., a coronary tree diagram or myocardial wall motion "bulls-eye chart".
              The renderMultiMedia element SHALL link to an observationMedia structured entry using the XML ID of that entry
              (see <xref linkend="sect_10.3.1" xrefstyle="select: label"/> and CDA R2 Specification, section 4.3.5.6).</para>
            <note>
                <para xml:id="para_13a15816-aac3-4848-aaac-05054e66a993">Production logic using the Business Name GraphicRef should assign the associated value to the referencedObject attribute.</para>
            </note>
        </section>
        <section label="9.1.1.4" status="4" xml:id="sect_9.1.1.4">
          <title>&lt;linkHtml&gt; Markup and External References</title>
          <para xml:id="para_b448944a-500d-46cf-ae77-4db76307ae96">The CDA narrative block MAY contain the &lt;linkHtml&gt; markup to provide a link between narrative text and an external (non-attested) resource (see CDA R2 specification section 4.3.5.2).</para>
          <note>
            <para xml:id="para_aa262f44-c1c2-4604-a605-8d9ace9c266d">For radiology reports, this capability may be used to tag concepts in the narrative block to concepts defined in the RadLex terminology (<link xl:href="http://www.radlex.org"/>), developed by the Radiological Society of North America. The RadLex coded vocabulary is a useful tool for indexing report content for data mining purposes. It is not intended to be a complete grammar for expression of clinical statements, but rather a lexicon for tagging concepts of interest.</para>
          </note>
          <para xml:id="para_6ae44efb-c48d-43f9-bcf2-55475c43e412">Within the report section narrative blocks, RadLex codes may be included using the &lt;linkHtml&gt; element and a URI pointing to the RadLex resource. &lt;linkHtml&gt; elements may be embedded in the text at the location of the concept (within the scope of a content tag), or may be provided in a list at the end of the narrative block.</para>
          <example label="9.1.1.4-1" xml:id="example_9.1.1.4-1">
            <title>Example - linkHtml references at point of use for RadLex</title>
            <programlisting><![CDATA[<section>
    ...
    <text>
        ...
        <content ID="find1">There is focal opacity
            <linkHtml href="http://www.radlex.org/RID/RID28530"/>
            at the right lung
            <linkHtml href="http://www.radlex.org/RID/RID1302"/>
            base most likely representing right lower lobe atelectasis
            <linkHtml href="http://www.radlex.org/RID/RID28493"/>.
        </content>
        <content ID="find2">The mediastinum ...</content>
    </text>
    ...
</section>]]></programlisting>
      </example>
          <example label="9.1.1.4-2" xml:id="example_9.1.1.4-2">
            <title>Example- linkHtml references at end of narrative block for RadLex</title>
            <programlisting><![CDATA[<section>
    <title>Findings</title>
    <text>
        <content ID="find1">Pleura normal... </content>
        <linkHtml href="http://www.radlex.org/RID/RID1362"/>
    </text>
</section>]]></programlisting>
      </example>
        </section>
        <section label="9.1.1.5" status="4" xml:id="sect_9.1.1.5">
          <title>&lt;linkHtml&gt; Markup and Image References</title>
          <para xml:id="para_bd3927a3-c221-4fb0-9d59-b2f75d6b2245">The text elements (and their children) MAY contain Web Access to DICOM Persistent Object (WADO) references to DICOM objects by including a linkHtml element where @href is a valid WADO URL. The text content of linkHtml MAY be either the visible text of the hyperlink, or a descriptor or identifier of the image.</para>
          <para xml:id="para_643664d6-e632-4878-956d-708edb85b66e">The linkHtml may be associated with a &lt;renderMultiMedia&gt; markup element to specify a (limited resolution) copy of the image to be rendered in the narrative (e.g., a thumbnail); the renderMultiMedia element SHALL link to an observationMedia structured entry using the XML ID of that entry. As CDA does not allow use of an image as the linkHtml displayable hyper-linked content, the linkHtml should immediately follow the renderMultiMedia for the thumbnail.</para>
          <example label="9.1.1.5-1" xml:id="example_9.1.1.5-1">
            <title>Example linkHtml reference for WADO image access</title>
            <programlisting><![CDATA[<text>
    ...
    <paragraph>
        <caption>Source of Measurement</caption>
        <renderMultiMedia referencedObject="#thumb1"/>
        <linkHtml href="http://www.example.org/wado?requestType=WADO
            &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
            &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823223142485051
            &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232232322.3
            &amp;contentType=application/dicom">Chest_PA</linkHtml>
    </paragraph>
    ...
</text>]]></programlisting>
      </example>
        </section>
        <section label="9.1.1.6" status="4" xml:id="sect_9.1.1.6">
          <title>list</title>
          <para xml:id="para_9dfd0121-15ff-4750-aa01-0bc0e6561cdc">This template specifies a structure and Business Names for list markup in the narrative text, as described in the CDA Specification section 4.3.5.8. Inclusion of the listType="ordered" attribute specifies a numbered list of items.</para>
          <para xml:id="para_d5788ef3-a298-46d2-b8f1-d84974ed60a5">COND: The XML ID attribute SHALL be present for a list or a list item if the content is referenced from elsewhere in the document.</para>
          <para xml:id="para_488f8730-46e5-4349-98eb-d102f2e0c7ec">The list items contain human readable displayable text using any of the narrative text structures permitted in section/text, including internal, external, or image references, or graphics. Processable structured data may be encoded in <xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> or <xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> entries in the
            <emphasis role="italic">section</emphasis>. Such observation entries SHOULD be linked to the corresponding item through the ID attribute of the item (see <xref linkend="sect_10.1.2" xrefstyle="select: label"/> and <xref linkend="sect_10.5.1" xrefstyle="select: label"/>).</para>
        </section>
        <section label="9.1.1.7" status="4" xml:id="sect_9.1.1.7">
          <title>table</title>
          <para xml:id="para_2dfc60e5-bfe4-4927-9c4c-907781e92ea4">This template specifies a structure and Business Names for table markup in the narrative text, as described in the CDA Specification section 4.3.5.9, typically used for a table of measurements. The table may be of arbitrary size.</para>
          <note>
            <para xml:id="para_cc4b328b-e6bc-4c6d-8d4e-ae3bb008b4d0">See Travis, A., et al., "Preferences for Structured Reporting of Measurement Data", <emphasis role="italic">JAcadRadiology</emphasis> 21:6 DOI:10.1016/j.acra.2014.02.008</para>
          </note>
          <para xml:id="para_39de2a49-cf5b-4840-9291-d9eca0b54691">COND: The XML ID attribute SHALL be present for a table or a table row if the content is referenced from elsewhere in the document.</para>
          <para xml:id="para_200b5e78-fb47-468b-8eaf-609815873849">The table cells contain human readable displayable text using any of the narrative text structures permitted in section/text, including internal, external, or image references, or graphics.
            Processable structured data may be encoded in <xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> or <xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> entries in the <emphasis role="italic">section</emphasis>.
            Such observation entries SHOULD be linked to the corresponding row through the ID attribute of the row (see <xref linkend="sect_10.1.2" xrefstyle="select: label"/> and <xref linkend="sect_10.5.1" xrefstyle="select: label"/>).</para>
          <example label="9.1.1.7-1" xml:id="example_9.1.1.7-1">
            <title>Measurements Table Example 1</title>
            <para xml:id="para_430caaf6-8ba8-43a5-8886-c3de59d60e59">As displayed</para>
            <table frame="box" label="" rules="all" xml:id="table_9.1.1.7-1">
            <caption>Cardiac Measurements</caption>
            <thead>
              <tr valign="top">
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d8cd08bc-b5cf-4b7e-a3f5-b5a688470bd3">Measurement name</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d6115b0b-799b-4480-a502-a9a26a988a07">Value</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_07028e89-18a4-457f-b56d-5a2bf0bbe573">Flag</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9dece34c-89d2-4447-a532-c45eda4c613f">Left ventricular ejection fraction</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5429a40a-34dd-4ca4-bdf6-4169ea5fb5cf">40 %</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fe396c3c-43b3-47c5-80e6-4a5d6b7ae3cb">
                    <emphasis role="bold">LOW</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_abd759fb-4f4c-47e7-b3c3-6e820eb376ec">Left ventricle end diastolic volume</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cd9fa0f4-6e65-4f77-b065-e7cd8f5f67c0">120 ml</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_75d99139-70fd-4b07-b961-537e1b5fbacf">Left ventricle end systolic volume</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_64735e8a-d4b5-4e11-9d85-93fae8be067c">72 ml</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_fbe27e0d-a738-4596-9e32-d92dfd946b45">As encoded in CDA instance</para>
            <programlisting><![CDATA[<text>
	<table ID="T-C">
		<caption>Cardiac Measurements</caption>
		<tr styleCode="Bold">
			<th>Measurement name</th>
			<th>Value</th>
			<th>Flag</th>
		</tr>
		<tr ID="Q1">
			<td>Left ventricular ejection fraction</td>
			<td>40 %</td>
			<td styleCode="Bold">LOW</td>
		</tr>
		<tr ID="Q2">
			<td>Left ventricle end diastolic volume</td>
			<td>120 ml</td>
		</tr>
		<tr ID="Q3">
			<td>Left ventricle end systolic volume</td>
			<td>72 ml</td>
		</tr>
	</table>
</text>
<entry>
	<observation classCode="OBS" moodCode="EVN">
		<templateId root="2.16.840.1.113883.10.20.6.2.14"/>
		<id root="1.2.840.10213.2.62.7044234.11652014"/>
		<code code="10230-1" codeSystem="2.16.840.1.113883.6.1"
		    codeSystemName="LOINC" displayName="LVEF" />
		<text><reference value="#Q1"/></text>
		<statusCode code="completed"/>
		<effectiveTime value="20140913223912"/>
		<value xsi:type="PQ" unit="%" value="40" />
		<interpretationCode code="L" codeSystem="2.16.840.1.113883.6.83"
		    codeSystemName="ObservationInterpretation" displayName="Low" />
	</observation>
</entry>
<entry>
	<observation classCode="OBS" moodCode="EVN">
		<templateId root="2.16.840.1.113883.10.20.6.2.14"/>
		<id root="1.2.840.10213.2.62.7044234.11652014"/>
		<code code="8821-1" codeSystem="2.16.840.1.113883.6.1"
		    codeSystemName="LOINC" displayName="LVEDV" />
		<text><reference value="#Q2"/></text>
		<statusCode code="completed"/>
		<effectiveTime value="20140913223912"/>
		<value xsi:type="PQ" unit="ml" value="120" />
	</observation>
</entry>
<entry>
	<observation classCode="OBS" moodCode="EVN">
		<templateId root="2.16.840.1.113883.10.20.6.2.14"/>
		<id root="1.2.840.10213.2.62.7044234.11652014"/>
		<code code="8823-7" codeSystem="2.16.840.1.113883.6.1"
		    codeSystemName="LOINC" displayName="LVESV" />
		<text><reference value="#Q3"/></text>
		<statusCode code="completed"/>
		<effectiveTime value="20140913223912"/>
		<value xsi:type="PQ" unit="ml" value="72" />
	</observation>
</entry>]]></programlisting>
          </example>              
              
          <example label="9.1.1.7-2" xml:id="example_9.1.1.7-2">
            <title>Measurements Table Example 2</title>
            <para xml:id="para_7401d381-8eaa-4cb2-b159-ffc6bc7d7c61">As displayed</para>
            <table frame="box" label="" rules="all" xml:id="table_9.1.1.7-2">
            <caption>Current Lesion Sizes with Comparison to Exam on 2014/11/16</caption>
            <thead>
              <tr valign="top">
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0724aba6-a30d-490b-9a89-6ebeac7b7b48">Ref</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_67d90d1b-2b5c-478f-987b-5149678a700f">Measurement name</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f3ab6960-e908-4b5d-83d3-35859712b5e9">Current Value</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6d2870d6-da19-4587-8324-4f4d9a37d478">Prior Value</para>
                </th>
                <th align="left" colspan="1" rowspan="1">
                  <para xml:id="para_40aef5b9-ffc1-48b0-ba3c-8c5790240daa">Image Reference</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3e1953dd-579e-4acf-a02c-95318d51b90b">L1</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1700a195-4bd7-4e18-9251-b5b9bb8dd77f">Left periaortic lymph node size (mm)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a9b988aa-bcd9-4500-afee-b034ae43bda8">12 x 8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_26a5ce44-e237-4db0-acd3-b22e7d8718eb">15 x 10</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_45582e59-eaa6-470d-a052-876377e194eb">Ser:3, Img:67</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b5af2497-f8f5-45ab-873e-b5cc351d453e">L2</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7edc0c86-0d5d-427d-9bf7-82ca1314d186">Segment 2 left lobe lesion size (mm)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_66eb3ad7-5837-4675-87b5-40199d1b98c4">6 x 8</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_75f22b3e-a8ab-4236-a2e4-9990c75449d6">10 x 9</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d08be022-36d9-4c37-85b0-6327a50e2d92">Ser:3, Img:79</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9f4a1d4d-2af5-4e0c-a319-a8c7fb8e4895">L3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b0f9d544-862e-4847-8dbb-27000517e6fd">Left common iliac lymph node size (mm)</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9382cc59-7b34-4c69-a77a-6c8f69c4aa43">12 x 3</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b722a28d-2805-4a7c-b93f-5f0b72c5d642">16 x 5</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d0945399-140c-4fec-9fa3-b98a3d413c87">Ser:3, Img:139</para>
                </td>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_bd9fcfb0-9b41-41da-8c3b-eded1bdad823">As encoded in CDA instance</para>
          <programlisting><![CDATA[<text>
	<table ID="Table2">
		<caption>Current Lesion Sizes with Comparison to Exam on 2014/11/16</caption>
		<tr styleCode="Bold">
			<td>Ref</td>
			<td>Measurement name</td>
			<td>Current Value</td>
			<td>Prior Value</td>
			<td>Image Reference</td>
		</tr>
		<tr ID="lesRow1">
			<td>L1</td>
			<td>Left periaortic lymph node size (mm) </td>
			<td>12 x 8</td>
			<td>15 x 10</td>
			<td><linkHtml href="http://wado.pacs.guh.org/..." >Ser:3, Img:67</linkHtml></td>
		</tr>
		...
	</table>
</text>]]></programlisting>
          </example> 
        </section>
      </section>
      <section label="9.1.2" status="3" xml:id="sect_9.1.2">
        <title>General Section Entries</title>
        
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d5dfdae-259d-42f4-b406-f1000830e3d0">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25365d57-6e63-4fbc-8f24-dd5d45378668">1.2.840.10008.9.23</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3c9b02d-7753-48da-9959-3dc887f831e3">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_379baeff-7ef5-4814-bf79-feee33f9db4b">General Section Entries</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_582e4dab-b066-4777-8da6-26888b6d3e8d">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6adc28bd-093e-4444-9e28-ffcf3a034a78">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6bbfe64e-d725-466e-bb38-2dbb039cb53f">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1fe5ab7-20a6-4a54-94a0-831b223516bc">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e66861c9-f026-416e-90c9-b4c7106891ce">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_367a6496-ef88-4255-9051-3d7cf4b1584e">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6d859b3-a59a-4824-8e8a-ab491949243b">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dac065ff-4879-406c-a87b-bd83918b68e2">This template specifies the common set of structured entries that may be included in a CDA imaging report section, and an author participation for the section.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_902abf2a-6754-41c6-bdee-c3c2946cf58a">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a53bfb8-7dca-4998-ad6f-eb0d523b66e9">CDA Element Set</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98a83761-44b6-475d-826a-4e8de5f13649">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4229e4d-78c7-49ad-86ee-269b3255b1ba">Included in <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> section and its sub-sections, <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> and other sections</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6e981c6f-6616-47fe-af44-abe806dc96c0">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1df11f5-9c25-400a-a25e-3b4bd02aa674">sibling node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a6b6142-e18a-41cf-97c6-b28e24945979">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e0f9d90f-8600-433c-aac9-57e748df02a6">open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28c90652-2994-4b21-b5bb-b192727af09c">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0814b53a-fd84-4a81-875f-2dd6cdffd094">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_74ac2c4a-2c12-435a-8295-c60a4c412f6f"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_eed1bead-4924-4a68-b2b4-b77eceff6b49">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9793b1fd-7a38-437c-a4aa-7ac007029cdd">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_eb0b82a1-468a-4c3c-91c7-d5016f755eb9">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0d870130-1ed4-436b-84fa-a6501cfd5695">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0fa28d4a-12fc-40db-853a-ef931230d860">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9e1d24ee-eca5-48b0-a62b-a3ad221973da">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ac5e431e-b7a7-4612-a22e-9e358478ced6">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_76ebd523-81c0-48a1-80e9-5fccd99c3ab8">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_fe30338a-60f5-46dc-b15b-44a7b835d217">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91975e99-e3b3-4fea-a506-26d85632482d">Content​Template</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ae0e44c2-b6c3-4980-9706-fe3d633fef50">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a55488d5-fcb3-4fbf-a28e-0332b04f486a">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a065ca31-e567-42af-8f32-0a3502cfc100">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ac35157-c154-4424-bbdf-cf1f3be136d2">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d60abc36-d373-4278-8c23-ad71be661aa9">author</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f00d824c-d8e7-44c6-a709-55e88049ca35">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f349e6cd-8fa7-4dd1-979f-88f3eb1fde2f">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ac1cd36-a900-40ae-9fc4-c4159f7ba042">Authoring​Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e479e7af-4159-4585-b5c6-9d614a1f0f86">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efcc4afc-f8b3-4af7-a99f-33b89f98c8a1">time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98322346-da86-4733-be00-db29cf19228c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f0aa112-ae63-4cb2-a7ee-1ea63ad1197b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7da822a-1e98-4f2d-b275-59133a9c2a85">TS</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a1133e4-d556-4666-bb9d-389f3047a738">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b60d8353-6b66-4cee-a559-6922e9e74952">assigned​Author</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5244e469-3cc5-4243-b6af-79b2a2c6933e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20cd1100-3c93-488f-a429-0784c97d977c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_33ba1bd2-9431-4530-8ffd-df159e11840c">Author​ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_49efc58e-3b30-49b7-bbfb-b07b2b493aef">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_384b050f-e535-4dc5-8a3c-eedece8b015f">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45db2d90-d81b-4bc7-9a18-9c3644cc73a2">1.*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7632ae49-44e7-4a80-9ddd-59e6e118350e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_acc72b99-7470-4d5f-bd44-15b487b48cd2">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_864d5c3b-e6d0-4721-aa61-fb21c33787ca">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c7a4688-7aaa-47e8-a387-3b66d3c24ed8">assigned​Person</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_108fada6-7a08-45bf-90ac-f3534c443e3d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ebe6d1c6-7f72-4897-bbed-d92308e5675d">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f83d8d2-3c95-450c-bf58-1a004449348a">Author​Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8e4dc08c-7194-4ff9-b676-b378bf33cca8">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be871a52-2c42-4afa-b697-cda4af5946a2">name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ee60389b-4051-4e2c-9147-6d3a333b9701">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_888b62a9-f11d-4c23-bbf2-950d8ee1731d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8fa98688-d0d6-491f-9720-476eabf32548">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e661fa1b-0c8e-487f-9b80-aea88b315f4d">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_879a5040-1e38-4c82-9b98-bd2e3bd5389f">assigned​Authoring​Device</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf983fa9-4d54-4eb1-bf89-9a9f8eac9ffd">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a80fe283-b542-4114-92cd-544172f07882">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d3b4552-3c15-4528-8f0b-6b9fc77b663a">Author​Device​Model</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6378f065-f01d-4412-a9a1-12f2177bfa3d">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de5b2650-e833-4f7e-91c5-ad2bb15d9dca">manufacturer​Model​Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c937d267-2d1e-4571-9f5b-0e77398dc1c7">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccdacdb7-44a3-404d-b88a-1d909fe17cad">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bea3b9bb-bdc8-46d7-9b58-30b5e076fdc9">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b211a6c5-a02b-468b-85cc-6c4e8bd52d7d">Author​Software</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82cf34bc-013c-4222-a7c3-47ac75585810">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7e0f9ba-587a-4ad3-a77a-e86fc0998876">software​Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1ea298bf-554a-435c-bc5c-65d21260a348">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_faff9601-db6b-4ecc-91e4-f99ba70f9760">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91d78f5f-1a17-4bc4-b66b-e7d7c1bf2fef">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56720355-6b12-4f9a-9aca-db34163a35f9">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfb46047-ab5a-41af-8e5d-f1db1f01b68e">represented​Organization</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fe97f97c-9a39-4cec-a6e4-31c1e36f5330">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_049f0d82-f4b5-4603-91c0-f48599ebb252">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0f708365-6053-47df-94b2-bf5b5fa7f774">Author​Organization</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a42ac3fa-fc84-4ad9-9159-9b3b318c4f6d">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74ee7a74-91a2-43d0-856a-bcc32ba87320">name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e6a83c1-c573-40bb-9518-cafd9b3e5b02">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f7e06878-1af8-4781-87d6-ead769ed8c00">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1da6ab2c-4144-4d61-bc3f-815dd2f2d4b3">ON</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a210fd8-731f-4231-8314-b5d39fb157ba">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8caf886d-d631-4324-bde2-8e30d8500658">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a8e99f2-82fc-4b45-bf44-78e06a0cae34">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4354d099-d1cd-4e58-87b3-f5733f33eb1a">
                  <emphasis role="italic">Coded​Observation[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44152fae-cc93-4b48-adfe-b5f85d13ccc6">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d73621f-3338-4564-8487-e0b1b98f93f4">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_844a6faa-9aef-490c-a06f-f209777e1958">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a058e88-db36-4509-9dcc-6280e80c9614">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cd04cd84-bcc9-40de-b008-0d8b694c83d8">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8e4f91f3-efcd-4565-a88b-ba5b7ecc1eab">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_18fdb5b1-dfa4-4265-bbc5-93baf8ef3683">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e58ea870-288a-4902-b62d-e0e987104b6c">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7dfcc454-f7eb-4063-8101-868ee364e284">
                  <emphasis role="italic">Quantity​Measurement[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc9d8c6f-fc3b-4edb-9fe7-e2350f640cc8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f100b090-4689-4fbb-b09c-18b7a27f31ed">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_13f31fbd-758f-4a89-b8e2-90b9deb4d235">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6fbf5216-178f-4c47-b39d-a5d9f726ba74">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_576ba613-7508-45d1-ad8e-616aea25186d">
<xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.14</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35cfb4dc-4aa6-47c5-a109-3e17ac41a2c3">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4007a23d-b843-4977-805b-81086a319b7b">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_852d86df-432c-434c-8a05-ea8f466c2e5c">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efbfd3ff-8dd5-40db-9e36-29baa8e6f36b">
                  <emphasis role="italic">Graphic[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b302d33b-f469-4060-a016-2c826ea32121">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdb6399a-847c-454e-bccd-ce6d56cae449">
                  <emphasis role="italic">observation​Media</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08484268-ea01-479a-a56d-fea243fedd03">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43ea0ecd-9eaa-422a-adcc-7bac650ab740">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_822fae24-1c08-4cc1-8d86-f35b085ddc1d">
<xref linkend="sect_10.3" xrefstyle="select: labelnumber title"/> 1.3.6.1.4.1.19376.​1.4.1.4.7</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eceb2f83-edcb-4eb3-8975-781f04a46494">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d829bcd-5ade-4014-a7c1-aa7b820714c7">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_457865a8-2292-4042-b2aa-ec2e557ccc95">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_037406c5-2179-4c51-9407-cf95d1fa9841">
                  <emphasis role="italic">SOPInstance[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a13e75c-7e5d-4079-98bd-df66022a16f8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_973f0828-06c4-4e98-adfd-5578bef1c3aa">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d410171a-b8b7-4bf6-b612-f85a8217649a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccb85b31-bfa5-4a29-b906-87d6d50bedc4">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f4a2640-f063-423b-b5d8-08ad70cf9230">
<xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.18</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55f0e26d-b315-4329-9e90-141a6ed2ee1e">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30d9fef9-6ef3-408f-8424-670f3e3a7419">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91b864a0-32ce-444d-bcd8-1258a0113e3b">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_975e71ca-b8d7-4f69-a8a7-8ff2a1f3cf24">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_519e43ec-490d-45b3-94ca-6ee91704d1fc">region​Of​Interest</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61cbe7a5-807d-4c1e-b868-215e6ea6b085">0..0</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e232a7e0-62b1-45f2-ab37-41037a29278d">SHALL NOT</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </informaltable>
        <para xml:id="para_c998eaa2-8c21-4858-a470-cbfcf176967f">Note that there is no business name associated with this template. Rather, this template is an editorial convenience for template specification, and the Business Names for the elements of this template are logically part of the Business Name scope of the invoking template.</para>
        <para xml:id="para_0d1ebf6d-8a7a-48b0-b5f4-799ce84cfaf8">Also, the ID of this template is not represented in a templateID element. Rather, the templateID of the invoking template implicitly includes the elements specified by this template.</para>
        <section label="9.1.2.1" status="4" xml:id="sect_9.1.2.1">
          <title>templateId</title>
          <para xml:id="para_847d994d-748e-4f91-b20c-9d0dc63b8445">This templateId may be used to identify the template(s) used to generate/constrain the content of the section. This is in addition to the templateId of the section level template, and typically represents clinical sub-specialty requirements. See <xref linkend="sect_5.1.1" xrefstyle="select: label"/> on the structure and use of the templateId.</para>
        </section>
        <section label="9.1.2.2" status="4" xml:id="sect_9.1.2.2">
          <title>author</title>
          <para xml:id="para_77d3e986-237a-4ebc-aff6-d4673b0bd391">The author participation allows the recording of an author for a section, equivalent to the <olink targetdoc="PS3.16" targetptr="sect_TID_1002" xrefstyle="select: labelnumber quotedtitle"/>. Either a person or a device may be identified as the author for a section or subsection.</para>
          <para xml:id="para_7658ad2b-34fb-4cd1-a473-aa7a84c6ed7c">COND: Either the assignedPerson or assignedAuthoringDevice element SHALL be present.</para>
          <example label="9.1.2.2-1" xml:id="example_9.1.2.2-1">
            <title>Author example</title>
            <programlisting><![CDATA[<author>
    <assignedAuthor>
        <id extension="121008" root="2.16.840.1.113883.19.5"/>
        <assignedPerson>
            <name>
                <given>John</given>
                <family>Blitz</family>
                <suffix>MD</suffix>
            </name>
        </assignedPerson>
    </assignedAuthor>
</author>]]></programlisting>
      </example>
        </section>
        <section label="9.1.2.3" status="4" xml:id="sect_9.1.2.3">
          <title>section/entry</title>
          <para xml:id="para_6be9a205-3f5e-4eb2-ba63-c73ab8d42655">A section may contain CDA entries that represent clinical statements in coded form (using the <xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> template), numeric measurements (using the <xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> template), images to be displayed in the narrative block (using the <xref linkend="sect_10.3" xrefstyle="select: labelnumber title"/> template, and invoked from a renderMultiMedia element), or references to external images or annotated images (using the <xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> template).</para>
          <para xml:id="para_714918d3-eefb-4dcc-b0d3-386ad03ee07d">These entries may appear in any order.</para>
        </section>
        <section label="9.1.2.4" status="4" xml:id="sect_9.1.2.4">
          <title>regionOfInterest</title>
          <para xml:id="para_2df757b4-9197-4872-8988-0d197d6e5eca">Section templates defined in this Implementation guide SHALL NOT use the CDA Region of Interest Overlay entry (classCode="ROIOVL"). If it is desired to show images with graphical annotations, the annotations SHOULD be encoded in DICOM Presentation State objects that reference the image. Report applications that display referenced images and annotation may retrieve a rendered image using a WADO reference in accordance with PS3.18, including the image and Presentation State, or other DICOM retrieval and rendering methods. This approach avoids the risks of errors in registering a CDA region of interest annotation with DICOM images, and places all image rendering within the scope of the DICOM Standard, including the full range of 2D and 3D presentations defined in DICOM.</para>
        </section>
      </section>
    </section>
    <section label="9.2" status="2" xml:id="sect_9.2">
      <title>Clinical Information</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_257f5737-8535-4af7-afc4-72e94fe10e4d">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d6dcb28-0e6e-4c6d-b883-fd5e53449ad5">1.2.840.10008.9.2</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7143e055-0dc8-467e-9e7c-3550bd7c5003">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dda75efb-2ca9-450a-96c5-095f4cc68777">Clinical Information</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9b9fc26-2a1b-4fc4-96fe-c07ba0e9ac1d">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a99ad9fe-737f-42b3-890e-13b9b4501c01">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88be070e-367b-4a7a-8c93-66938ae10b70">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77a1a784-e7d5-483c-98b8-6d5cedbd7882">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0d8e64e-1b4c-4d93-9d5f-e28a4a36f42f">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7514204-5ca3-4c48-aff9-9a46e97b4e6f">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f243942-11e0-4eda-8d89-539d8e0e1e3c">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb905734-a942-4c6c-85ce-e41cf47c6b93">Clinical details about the case such as presenting signs and symptoms, past clinical history, the overall condition of the patient, etc.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2521052a-c62e-4767-ace8-3ac96004a19a">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f6f2ad9-7437-4d3b-9948-07a188419be9">CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b8f433b-b649-42df-aef3-9e77259e7d00">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ecc2616-e006-4571-b164-59bd6b3ef6ff">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_18a1a861-8162-4eb1-ab2f-b87f5d457b40">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd291011-f3c3-470c-8740-aae7b4cb66d8">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_16d8702f-0650-44c4-b7a3-f07f9bd837cd">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_efc82db6-be61-4870-b497-23589fb50d4b">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5e960d3-7e04-41c1-bbc4-de7f67451aa4">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c33f714c-f663-4837-b66a-0354503f788a">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_8e7ae2f7-9715-45e0-a91d-a319d4bf2f92"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_cf945a9e-7cfa-416f-b2b7-5f98350b27c7">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_78bc839d-df33-4d1a-bce1-da481b55f9c5">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1b68a8f8-21ad-4f45-89db-e1141ae2ecfc">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_d636edcb-e515-4dc2-884f-de582573a943">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_7020419b-67da-47c3-a5ee-2783af0200c0">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_98a0ca78-ec3c-41be-b70c-81166cc05a37">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_62667e36-e1dd-4690-945f-a75285f3e28a">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ad64e988-dee9-4e37-bb76-76279380d989">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9cdea53f-4839-417e-aadc-d56e342729aa">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aa065682-3096-4a40-be81-531305b3a4dd">
                <emphasis role="bold">Clinical​Information</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b882688-2c38-4a8d-8f0c-907a998f93bd">section</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4e19857-523d-49f9-b7c1-8386ad59e7e0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30717d11-0fba-42a7-badb-fa28e8464797">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d3b072c-8781-48af-b1b4-1d84cb18c4ad">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_638357ab-123d-4cf6-8e8e-981b2925d4ce">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b2f4af6-3103-442b-b04c-398b06abdf69">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_26a42bfd-125a-4418-a3de-b43d4aa14c04">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f57ccb30-90e9-4d75-95db-7b1aeef971a4">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4f9c968-0784-4a23-9edc-782073df7cc5">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c4fd4bf8-1920-466f-bf01-3a3ead0fecaf">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c6f22d9-8380-4ce3-b81f-05958bd806b6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06dcf81d-f618-438e-bd71-5b74c3b8abfb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c19a5be3-1b47-45d3-b807-e6f79d4815ff">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b12aeb6-c1ff-4167-8ccc-175df37153e4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f736aac-5cbd-48d7-b54d-8e1a807e03f2">1.2.840.10008.9.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f890f67-1c93-4b97-93f9-0b410543ea76">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5d0135f-0682-49f3-ab5f-12d5ef5fa50b">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e100f861-05a9-4c66-8075-86e20928f7d1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d9c59c4-8da9-4f12-adfb-ccb3c743ffa0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e762329-9c03-4ad9-8f57-67097a6f69db">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f122ff91-359b-4e5e-a4c4-566844449479">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c17cf67f-aefc-4062-8d48-f20173cd4c4a">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_822668cd-0edc-44ae-8e23-abfaedfd6d49">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0980a13b-d18e-4574-9aff-5c3a3e03bc25">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b36dfee-8664-45bd-b99b-37ffaf7972ee">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fcf7d57c-96d9-404a-96df-0b58fca88aba">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0d9ed75-354f-4687-ba4e-35a649df5951">
<link xl:href="http://loinc.org/55752-0/" xl:show="new">
<link xl:href="http://loinc.org/55752-0/" xl:show="new">
<link xl:href="http://loinc.org/55752-0/" xl:show="new">(55752-0, LOINC, "Clinical Information")</link>
</link>
</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca2b9943-bfc5-42b5-973a-7042f695a03c">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fcee0043-d0f4-420c-884d-ea3efeffc38d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8250953b-cd10-4128-bf42-654f13e0fd1a">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_38b15a39-5792-4737-9802-6cc5f71fd14c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f7d054b-c560-4be9-994e-7957fd4b0d55">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d3d89c6-6a82-412f-848c-c364bdc8d8e6">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_16d4ec97-3c15-4a81-9282-d217d6b4bc93">
                <emphasis role="italic">Text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e765b54b-aeee-4388-aa79-7940cb84a937">
                <emphasis role="italic">&gt;</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8282a127-3a7b-4f33-9700-7cc5f66c8dcb">
                <emphasis role="italic">text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_888f8915-c3da-46cf-91ce-aa72928b7acf">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fbd8a510-d72e-42b4-8c04-b86529131295">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8a19974-bde7-43cd-b2b5-e97fff116d62">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fd19604-a7f1-4545-bbf3-1dcf9e1ee510">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_952b3a82-5611-421c-917f-a80014de982a">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8312554d-ae6d-49f0-bf13-ae6e85729fe4">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_55369204-3d9a-4935-8bec-42f175b82413">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_936a2256-fcd6-43be-a26a-ccc26769079e">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46b56ad7-7dc3-46dd-ba44-3ec86496c7c3">
                <emphasis role="italic">Request</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad2d721a-22d7-4a1a-8d69-71c27d7451da">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dcb34a62-d3bb-4d12-befb-9090b6963651">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b594a38b-a223-4b69-8909-c8ec76fe05ca">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aadd6e8b-94f4-4c4f-b10f-cefb064cb088">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39cc156c-0fbf-4386-a694-f37821749f47">
<xref linkend="sect_9.8.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.7</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_404b18ba-f64b-4fad-ad95-431d0953ba7d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7e9e4f6c-4822-4fe1-89c3-4bed85054c69">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b46765c-4860-47fb-9328-d1b14ab5d60c">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_177978cb-0eea-479b-abaa-7b6234d1e16f">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a025fa63-3bf6-4829-97f3-fab422d489f9">
                <emphasis role="italic">Procedure​Indications</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_65604511-1e9c-4b70-885c-452bf1c28bd5">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a91f1c05-39bb-4fc8-ba05-bc290f3f6865">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f68400d4-1722-4a84-b913-184a511317b7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4955e509-2229-4fec-aec3-ebbaf346a53f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d05b51f-6688-4db5-90d4-6461776a32d5">
<xref linkend="sect_9.8.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.22</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab7527f2-3df2-4126-80a2-d9d7e096fc4a">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c5b3bf9-4758-41fa-be0e-d8123fa199d9">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98d98b7c-0390-4fed-8d1a-69b4adf66087">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_528d87e0-be91-4d59-8564-3f50d7febbb5">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e2b3858-79cd-4b0e-a347-f7a37af6a0f9">
                <emphasis role="italic">History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a9cccd3-c3f2-41d6-93eb-bb81bfeaed08">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7b5fcab-1b51-4fce-90e2-5bf4902fae11">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_699c122c-95e6-4cec-8b1c-8d0c8763ca3b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_79fcee8c-0b71-48a0-ba25-286bb1c24a50">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ede86662-d05d-49b1-8a7b-bcc019039622">
<xref linkend="sect_9.8.3" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.22.2.39</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa66a871-47ef-46c4-bda7-9514cae6c87c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b34281ed-31dc-4d8d-9bab-33bd45961b00">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4bf60eab-2608-42e9-a5b9-cba6849940fd">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_01a1a246-aa3a-44bd-b8a8-6755232222bc">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <example label="9.2-1" xml:id="example_9.2-1">
        <title>Clinical Information section example</title>
        <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.2"/>
    <id root="1.2.840.10213.2.62.994044785528.114289542805"/>
    <code code="55752-0" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Clinical Information"/>
    <title>Clinical Information</title>
    <text>The patient was referred for evaluation of suspected pulmonary embolism.</text>
    <!---see examples for other sections/entries -->
</section>]]></programlisting>
      </example>
    </section>
    <section label="9.3" status="2" xml:id="sect_9.3">
      <title>Imaging Procedure Description</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b604971-aac3-4bb1-ba2d-2fec29b048bd">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_42a31b01-323d-453f-b413-8ed242a715b5">1.2.840.10008.9.3</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ae5722e4-8432-4a03-aaea-ec66e5f5a1d0">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1a1063e-5cbc-4d96-9034-e459f145d2ab">Imaging Procedure Description</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a08621bd-64c8-4e9d-aeec-ec696731d07c">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_decdde26-3e07-44ee-bfb6-e313ba5928c3">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1aa8332c-046d-461b-9ba6-c73171b3ebfb">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff37860a-b887-47e0-a6a7-4c7bfa93ba7b">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d725331-55b8-4242-a882-608856dac283">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dcd75b24-1f82-4cdc-bb4a-83555c9f9b6a">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e20033ea-2afc-4e82-9aad-1249b227ec48">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_456245e9-01e0-45fd-b063-6e8dd119b1cb">The Imaging Procedure Description section records the technical details of the procedure and may include information about protocol, imaging device, contrast, radiation dose, medications administered (sedation, stress agents), etc.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6cb25bf-737e-4b31-b5ea-394e5fb2fde6">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3ca4e324-a722-4567-8ed2-7832dd097606">CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a09def97-6270-4c89-acaf-e1284e536896">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d4cffa11-a862-4fec-92f6-7666785a6a9b">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39df4354-7138-4f77-a6f7-4483d423f667">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_decee911-58c9-440c-9933-d32d9a4157c1">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac4395b4-486e-41a3-bd4f-9b47ba978495">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62d26003-c821-462f-b95f-ebe8d0bc3a7c">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a06da1c-6a15-4538-8e19-3eb8e83928c4">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70550802-0bd6-4d9b-b86f-46980be47a16">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_15688891-bc65-48f0-93b7-6d78222123f9"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c22f5257-89e3-4121-bd20-f1943c45c886">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e9ade4fa-0efb-451a-a3e6-4970cc4368df">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_6d57b631-0f8c-450f-a1b1-447ee16b9c58">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_21165477-d6c5-403e-8a4d-72b1d980297b">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a5c546ce-4edf-4380-92f6-758e2d784277">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_50e74de5-f429-44a3-862a-2ae83c5bd210">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_da1cc418-08c2-467f-8c31-27ebfa571f17">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_7633804b-4924-4425-a2ae-7a64c9269c6b">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b423189d-0f39-4453-9c45-b50d8f78bec8">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f347c571-a27d-4a8e-bb24-d05a861d9cf9">
                <emphasis role="bold">Procedure​Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f661f40-124e-470e-99d2-b9393a761a5c">section</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36cacd00-2586-43a6-9dcd-83b620bd775f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e9295f5-c933-470a-bf19-674231302deb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf8c1d5f-6fcc-4ecc-ab55-971fc2cac5d0">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a24f883f-0f38-4f2f-bfac-263928f57d1c">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_878382a5-810b-49ef-b37a-e533f8ebca25">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ccda7922-34e8-4042-ae67-f694aa70858e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c8141bc-e3fd-4e8d-9af9-169b48f269e4">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7f0d4bf-82a0-4f33-8b4a-2fcd89b1350b">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00101ca8-1a95-4ea1-a126-a105b0d6690f">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d95ee314-9575-4047-b669-e235635a587c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7bb01931-712e-48ec-8427-b8e50e685993">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_025642c9-8c2c-4402-bd24-85488f553e79">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_572d1a6b-29d9-4800-85f5-510ddf8a4e34">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f9c785f-1f16-4856-af60-91832b77ca7b">1.2.840.10008.9.3</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eae7b898-f2ad-4573-b7ca-ed6b2f003998">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54522ff7-d84f-4999-bc22-d177d86f7bfc">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a3449a7e-bfd6-408b-97a7-dc95b162ddbd">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9aba8632-c888-4ea2-8513-261537eecd9b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_be073adc-23d4-46d4-b906-2d58facf57ce">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfa7ecae-ed79-42cb-89df-d60bf3c5be36">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4733620-ec46-4ba4-8090-60ef91a5fb40">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67e77513-016b-4a25-9002-affe3ce43691">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06a3f722-1026-4353-8e0d-b40cf6e67350">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_66e1cb59-76dc-4544-a314-d7ca74598f10">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_337e2903-ff8d-4c8b-9aa9-11e5bf45181a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_611e3d8f-9616-4b06-b777-9f6b8b057ad9">
<link xl:href="http://loinc.org/55111-9/" xl:show="new">
<link xl:href="http://loinc.org/55111-9/" xl:show="new">(55111-9, LOINC, "Current Imaging Procedure Description")</link>
</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf7ac6b2-146e-4a99-bc1d-013acb39868d">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7254857-20da-4b23-8535-ec3c1bc6bae3">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0626c15-2913-4914-9365-44008fdc51ab">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bdca3afe-df0e-4fc8-8a57-9dca931ac25d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4456ce38-3c19-4ee1-b1c9-1e52f6471b1c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5fe0fd3-04a1-41e5-b724-7d143934bfa2">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e16c658-1f07-402a-8a3f-cc2e9cc14f42">
                <emphasis role="italic">Text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a6741de8-0b8c-4445-b1b0-a76e114334db">
                <emphasis role="italic">&gt;</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d21694ad-9352-4777-8fc4-772a2f84bf83">
                <emphasis role="italic">text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31617aee-7917-4dc6-859f-b1e4bb373a87">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_655bd18f-34a0-494e-82a8-7f2b6c36f65d">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b0593fb-e472-477a-b596-bacbb99a1558">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3e55d00-d1ab-4d15-bf55-1f75108925ec">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_240376b5-806c-49de-999b-c45610495e2c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df0299cc-963f-4e54-9542-f00cb31e15a2">entry</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_49b6c450-2b77-4f61-a6d3-37906c1bc44d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_487c853e-d783-417e-bc15-04b79b890ec3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_963d6679-b14b-4cc8-800d-7a89b44626b4">
                <emphasis role="italic">Procedure​Technique</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_34fe5b13-6898-4ad4-b6a6-63bd5e8bf10c">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71dbd166-a3b8-4162-9856-f2b1e0ff7762">
                <emphasis role="italic">procedure</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bea6a91e-711f-4d04-9128-6f455da1470b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7e220074-9225-4b01-8bff-c0e0f7b2c470">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8efe99b3-fcf3-4770-a651-7cb75dc8d4c2">
<xref linkend="sect_10.4" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.14</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_51850015-93c4-4d3e-b73e-755f307700f9">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_52c0a540-968e-4f34-9e12-eef39e0a15f6">entry</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2639872c-0181-4688-98e0-271825060c86">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4941154c-bd8e-45bc-bd68-ced84d5ac102">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfc33e45-08bd-43bd-96d2-9076ac11c7d7">
                <emphasis role="italic">Procedural​Medication[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b0e0e59-b91c-4ff0-a56a-12abe2d1d921">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b8b9627-beb5-41cf-b0ac-807bfe328694">
                <emphasis role="italic">substance​Administration</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce7a22f6-c648-43c8-a2d9-9f78c59913e1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ebbe8e0-9832-4978-a99c-6f52a2b854eb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f42625a-9cdf-4741-9568-33ee28ae728a">
<xref linkend="sect_10.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.13</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ffa73052-112b-4761-a27f-fdc825d96cfc">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b79def5-2241-4fee-bfe1-fd8c11a50314">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df72c91c-7016-487b-8010-35c53ed408db">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7182845-1509-44f0-b33e-ff67730951af">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b170617-5c21-4860-8c86-3965566e30c0">
                <emphasis role="italic">Complications</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fde9805a-ae19-4814-8f7c-81f6e960a0bc">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8d978f4-549d-45b0-b04b-1101302507d6">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e005922-dfbe-480f-b9f9-53db42dce7df">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_774e59d9-f5f6-4fb1-8fbe-281b3f7f5ab3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9887ec62-e526-4e70-b60c-8dc1bb1dafea">
<xref linkend="sect_9.8.4" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.22.2.37</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_42a5ab2b-d6d3-4ee3-b228-dc9df3866632">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa84aff4-380c-4c3d-a7d7-b3b7908525c3">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9db945d-b45e-4103-b740-44e1a7645922">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b36a489c-2be3-45ac-a69c-509795965020">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f89cc199-cf08-4c95-ab07-68cabb379677">
                <emphasis role="italic">Radiation​Exposure</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c258c6b-77c7-42d1-994b-6d559c4cf925">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_093645ee-d982-4a54-a1a4-00f5927dae2a">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_143c5507-e1c4-41e3-ac72-511f68a3646f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d136461-c6b8-4fd0-9509-d3542cc08110">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e9da84a-1ce1-4c03-ad17-4b499fd6dde3">
<xref linkend="sect_9.8.5" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.8</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36a8b408-edb6-402c-9ec6-d117850979a9">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53493722-6003-4026-a50e-57a33f2f7d43">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fdd5643a-b603-4569-90d7-81d711670d9c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d9fa5d1f-c0e2-4c64-ad16-9361a90d6773">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7f34bc4-e57a-413c-9b0e-58ca0b6b6962">
                <emphasis role="italic">DICOMObject​Catalog</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b44ae8a3-c174-4f22-a182-3be5eddb06b1">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc20fd48-9ad8-49e6-8742-c5c2dc62b459">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70797c9e-4e92-4007-9cea-7e05b9f533e4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_569662a9-67b1-478e-a08f-e165436e2933">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63915460-110a-4ba1-bfce-a047f2b80599">
<xref linkend="sect_9.8.7" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.1.1</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fcdddbd-6a4b-460c-8973-3d07c359cc63">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61b2581b-223c-48e6-bbbe-5cbc5eb5e98d">entry</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_24625114-8c50-448a-a83f-4838d1844bf1">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_980d414b-4d6a-4c3b-a935-a327a2b73e48">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fcb98a9d-05f4-43af-95bd-ab83dc3292f4">
                <emphasis role="italic">Image​Quality</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_078698f9-f206-403e-97c6-63c26b81cbea">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_790ba0d1-0d20-4bfd-99dd-8d5b17b232cf">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_efd4422a-3d80-407d-8580-095426d32da2">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ed9ea1c-c6da-4298-978a-fef9872a1377">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d9004f48-12c2-4987-9e22-703c05cf0bce">
<xref linkend="sect_10.9" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.15</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <example label="9.3-1" xml:id="example_9.3-1">
        <title>Current Imaging Procedure description section example</title>
        <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.3"/>
    <id root="1.2.840.10213.2.62.9940434234785528.11428954534542805"/>
    <code code="55111-9" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Current Imaging Procedure Description"/>
    <title>Imaging Procedure Description</title>
    <text>A CT study was acquired with 2.5 mm images of the abdomen and pelvis with 140 ml of... </text>
    <!-- See Procedure Technique template example - required here -->
    <!-- See DICOM Imaging Catalog template example - required here -->
    <!---see examples for other sections/entries -->
</section>]]></programlisting>
      </example>
      <section label="9.3.1" status="3" xml:id="sect_9.3.1">
        <title>component/section Radiation Exposure and Protection Information</title>
        <para xml:id="para_e9958ed8-8e95-41ef-9899-9ea96c5f67a6">COND: If the documented service utilizes ionizing radiation, a <xref linkend="sect_9.8.5" xrefstyle="select: labelnumber title"/> section MAY be present.</para>
      </section>
    </section>
    <section label="9.4" status="2" xml:id="sect_9.4">
      <title>Comparison Study</title>
      <informaltable frame="box" rules="all">        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_284cdbcb-0873-4c49-8d16-0bd209cfd68a">
                <emphasis role="bold">Template​ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f7390fe2-c1e8-4b95-a14f-0a1f1be363ac">1.2.840.10008.9.4</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9e7e030b-dc17-44a5-9c37-ecd8749f456c">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eff74bf7-0119-454e-9d72-bc8916526bc4">Comparison Study</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_af607f51-47b9-4e52-aaf3-c2afeb7e72f4">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_439f6095-e71b-41cc-9ec2-e00df760561b">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00baa383-f48c-484c-a251-351341ead4e5">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd4ab415-185e-4aea-bb14-0485009264c8">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f169ea03-50c1-4d1f-984f-0275aec1b30e">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c11d1e7d-2032-4912-be5b-ab21f37a6f00">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48341d4b-bdf5-4dec-9964-2cb2d0b51ed8">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5d70805-454c-4bfc-90ee-ef1806329c95">Documentation of a prior Imaging Procedure to which the current images were compared</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d18fd8b-2322-4757-859a-3a124cb6f09d">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_52307a88-aed3-4755-9f13-9735f4702fbe">CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba1db3aa-4aa3-4ccd-a24a-7071045500b1">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89022da8-3f5c-4b5e-bc5f-b6751846831a">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_318e5d47-7872-4812-b34f-d73639a30560">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad9a9442-b33a-412f-9242-69070d99596d">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_acf86ff0-8d88-4a29-a8a9-393e07ea5fcb">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b42313c-5761-457b-9bc5-414286c437fa">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_502a699b-6255-46ce-b177-f26db64a206c">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74c4dbaa-65d8-40ba-b499-7f556b0eb4bd">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_830d5dc0-2358-4ff4-8192-c5f0bc3d039e"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_16d82902-25c5-4b6f-8755-506ff0549403">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0b145324-7dbe-4f70-97d5-a3d1f9f7b995">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ea7aa5ea-d778-448c-8409-1a274b642f9f">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ba13e043-2ddc-4456-98d0-83de55df3ec0">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_6a875489-0952-4fa9-aa94-a271666d8bc6">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_70bd7ebf-b5c6-4c29-8b2c-95a0d1d3eee2">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_459a992b-1ac9-44f1-897e-bc754bc78d33">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_dba8c8e2-ffdd-4254-8893-b8d406e1e45c">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5a776af6-c5bf-475d-b698-69e271958c35">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e6510019-58d9-4708-af88-935b95cc1bd6">
                <emphasis role="bold">Comparison​Study</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48413db1-c6c9-42eb-9e35-0ebb5a4ce630">section</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8a8f8ff-93d2-4e1a-910a-ac27d23afb5d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0fc3d11-1150-41dd-813d-c9029f1f1b86">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d180af2f-9192-4f2e-8115-e172529a58ae">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9664d460-3734-48d3-8270-53e68673c576">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8b64924-df49-4517-952f-35510b3d0a2e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_390c9c6c-7de3-4284-9e52-5d6632b6f1e8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_34e65db1-b0d6-4778-ae5e-317d4d52bfee">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d68cad3-9b8e-4ef5-a525-fd46336a1ed2">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2346007-e418-4864-9a89-e7a570cc97c3">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ceae72f-046c-481b-a4be-2dcb036932e1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_acb8f729-240e-45e7-9daa-ae03fae4d856">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e65d4bf1-6271-47d7-8448-379268549517">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36d2ba0c-914c-4054-9b7d-68b6734ed834">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1855ad40-ec1d-45fa-9c09-b9a73165443d">1.2.840.10008.9.4</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_190d26e4-c418-44a5-9c18-ad93193194a1">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8f9fff9-a749-419f-a9ac-16cd96905868">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f4ffa89-a21c-47a9-94bc-6e28150515db">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2059457-a7a2-4697-a6e2-4081f56aa33b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a6d3d4f4-618e-43fe-8b47-a5f8dc543666">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e2a5b5c1-abbe-49b4-ba9b-9a766cdede02">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89ddc6b4-8b99-4add-8699-ac6c5ad1fb00">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3908144-2b84-435b-997f-976b744bdf40">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc65e767-85f9-4bfe-b2b9-a74340403e86">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_884e66fd-f84b-4db6-ae98-c6340d8553c4">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b641265-a6ec-4aa3-a96d-1094dd9747c6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4a09c68-fd63-4e64-8760-07fd611f2166">
<link xl:href="http://loinc.org/18834-2/" xl:show="new">
<link xl:href="http://loinc.org/18834-2/" xl:show="new">(18834-2, LOINC, "Radiology Comparison study")</link>
</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_18eb448a-907d-40f6-b236-37d7c03ad007">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d70f456a-4321-4872-96b8-9faf13cb6c2c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac933927-5e0c-4513-9978-66b815749ce2">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e5758d0-0466-46a3-b579-8ccf4f03d059">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e479a60-16b2-49c0-8624-ce56611d15cb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b0cad53-cb37-4432-8604-e7d771072b8e">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec91c63d-280b-4f42-85cb-b67816ff62ff">
                <emphasis role="italic">Text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43951fed-19ca-49e5-b202-0dbf25c7ba48">
                <emphasis role="italic">&gt;</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9dca972c-cb7f-4a41-a890-90e867e0d2eb">
                <emphasis role="italic">text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da87a70d-66eb-4af6-a18f-8511f8d97de3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_702e5216-6416-4eac-bfc7-cf63329f3d10">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_941b2465-c59a-4b1d-9d97-feca763c8b91">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c977e90c-e967-423c-a14b-9f3547f570ea">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc09b976-0b0a-4041-82c9-4e5ed5693975">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c06906c6-d492-4d95-9170-5623bd93448b">entry</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d30131ba-cee6-4966-b0ae-d1a595dbe5b1">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_490393ab-503e-4023-a86b-ee22ed21725b">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_49f2ebbd-d4e1-4025-a844-a8a9c0685ae4">
                <emphasis role="italic">Procedure​Technique</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b865ec5-5727-4023-80ab-4069f1a618be">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_722996dd-9081-4e8a-a8a5-ab3d98dfce69">
                <emphasis role="italic">procedure</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ffbe7b3-17ab-40ba-8ebc-4a67e407b1c8">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63ddb50b-38d9-407f-8321-9ac7311f9682">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc6c66ca-2dda-49da-ab0e-5077b9bc7c16">
<xref linkend="sect_10.4" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.14</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97271890-acfb-4527-a823-f47e1e6f9956">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_caa0625f-926f-4fbd-8173-5625244e5aa6">entry</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_862a0fbd-3cef-459b-a60a-317e85576751">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b12d89a-0286-4999-8c57-5e1a96d4fd78">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d407626-3ba4-49e3-9920-c49df158c7e8">
                <emphasis role="italic">Study[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5646a7f5-7a08-4339-a03f-ad75f064a397">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd6d258b-8f0d-4bdc-b29b-689bdcf5374c">
                <emphasis role="italic">act</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec864322-2657-434d-8658-7d3abe6d55b7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_993bae7d-d40c-429c-9d59-a0b4c7aef1e7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17e6187c-e27b-45bd-996e-40359e5307a6">
<xref linkend="sect_10.6" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.16</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_152ed240-f949-48a2-b974-b32f65d13160">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a89773ed-946b-4b5b-a637-7914ac996ddf">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9671e960-3ce4-477b-a917-d1e3fc099bc4">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5067dfb9-dffb-4a4d-9864-0d73095fbaca">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <example label="9.4-1" xml:id="example_9.4-1">
        <title>Comparison study section example</title>
        <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.4"/>
    <id root="1.2.840.10213.2.62.994056444785528.1142893564536542805"/>
    <code code="18834-2" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Radiology Comparison Study"/>
    <title>Comparison Study</title>
    <text>A prior CT with contrast performed on May 7, 2012, showed that ...</text>
    <! ---see examples for other sections/entries />
</section>]]></programlisting>
      </example>
    </section>
    <section label="9.5" status="2" xml:id="sect_9.5">
      <title>Findings</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ce1865e-57bb-416c-ba9f-12c3e21e1cb8">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d2975e2-1871-4215-aa4c-2aa383cc41c4">2.16.840.1.113883.​10.20.6.1.2</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_377320df-afe9-4829-84c7-2875c8059ec5">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_771d018f-7345-4528-aa06-14abb59c9e54">Findings</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf876659-2685-4ebd-bced-7dfac48b96b9">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_32b68fb2-5732-414c-b9a2-05a31d74d09e">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d404a32a-e732-4523-ac25-a9333e128f95">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39737fd2-9cad-42f3-9f82-7722a44caf4b">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_42c02029-bceb-4215-8df6-977c288be347">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce24ff6c-4f50-48a4-bc15-278dc073e818">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81b99581-dd48-40c6-8142-21a0f8ebed9b">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d4675f7f-26df-43d4-b939-bcb246689c50">Records clinically significant observations confirmed or discovered during the procedure.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8932baa2-4fc6-4b33-81d5-1f24090dc0ba">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_11d4e9cf-72a3-4f4a-adbc-7f38e0653fdc">CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a542270-d8be-49c5-b982-47f802a7de29">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40e490a0-d03a-4663-9a21-e2c54674b3fc">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6253e14-9aad-45d2-b759-5baf4c7c7558">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_91b66232-42a1-4692-a4d3-bb17aed493be">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d86af7e-efe0-4c56-80d2-07f99eee6fd3">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b74934ea-c599-422b-b79d-3b747dff6dc7">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c50c0ed-3c2b-440a-ae72-234186056995">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46e8544d-d78a-4ace-9b96-7e6585ea0e19">From Consolidated CDA r1.1</para>
              <para xml:id="para_a9809ab6-8d32-494e-9a7d-947adb033589">DICOM-20150324: Added optional subsections and entries</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_805a5208-7fcc-4106-adb9-56fec7b38132"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_672847ce-ae4b-4f4c-b2ce-d5cce3cb2f5b">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a489329d-00a5-4303-9921-ccd9304f6db2">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8e19f2aa-5c79-4573-89ee-3efa868cbbd7">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b71bc46d-d685-48d7-8379-c1aed5b722bb">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_714558a3-4774-4f04-9d91-59364f83f509">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b5d9493e-f7fe-4408-b508-33c41367bf55">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4b817ecb-e4ff-410f-bb03-713fd388196a">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a1e65b11-d586-46af-9da4-ceef26f2b96f">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_dcd43e30-7e17-4e00-8a76-368764100c87">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73c884bf-f4ee-4e59-b40b-2f514a15e577">
                <emphasis role="bold">Findings</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95fe6d44-ec60-4687-ac0a-14f4a6610226">section</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d112d93-56b3-4281-ac78-d8715c9a2879">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a7baae9-9851-4aef-aff4-26aa9450196d">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95051430-3fd8-480f-8c6a-d1a45201c2bb">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cff56224-44fd-4cfa-bb2d-ac21a787fb47">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7e1fe2f0-4804-40a2-a8e0-b80c16a85363">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d479082b-f1c2-49d7-bafe-daeb205abc14">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba5f69ee-97ba-4f4a-907e-8ccb46400453">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3035294b-2e06-4c6e-abe8-e7dc501f5be9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48ff6209-9354-4567-b1b4-0669bf387916">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ff85e52-5d37-430c-a330-acaeaca533b1">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_debda58c-1ca6-4223-87cd-7bda2fc6fe14">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_695f818a-5349-454c-ba1a-0883a4d3faf7">2.16.840.1.113883.​10.20.6.1.2</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_422b091c-c7d7-44d5-8f83-05cf74faee6d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_242b25fe-649b-41ff-b0e3-0fbc482240a4">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_767c6a0a-cd16-47cf-9848-c17ec1401734">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6f6aa6e-15a1-4998-899a-ea4891f68efd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98ed370f-cbf1-4318-ab1b-b2904f2e43ea">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa6cb1f6-e747-4c5a-a5d5-b68e08b84fd2">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22b50017-0ee0-4999-a789-36483019dd39">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d2814f7-1d50-4d94-8646-beebf5713c12">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a02229e9-7fcb-4a8a-88f0-1eac734aa197">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b830953-6989-4d3d-b531-70313c4ecc8b">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84833a94-cfde-4d48-b86e-971b35a62aa5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_178f958d-c314-47fc-9bd0-1e20609ff6c2">
<link xl:href="http://loinc.org/59776-5/" xl:show="new">
<link xl:href="http://loinc.org/59776-5/" xl:show="new">(59776-5, LOINC, "Procedure Findings")</link>
</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf1b02b3-bd27-4836-a630-acb2522f753f">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d88b8824-ec3d-41ed-9873-34aba363c587">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fb913b9-663b-48bf-aaca-476457b49569">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c24e01b-7254-4420-92ae-d89bfe651f27">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_01f77ee1-43fb-4dec-b66d-22da5f102519">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_75b99b93-4c22-4893-8f90-f82744a95b1a">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8772d894-5f9c-4a46-9ec6-10ed66d130b1">
                <emphasis role="italic">Text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b18c7d16-889a-4cb5-84f1-484d52278808">
                <emphasis role="italic">&gt;</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bfb0a60a-adca-4668-83ad-d4d2f87cd81b">
                <emphasis role="italic">text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a15d91e6-b92d-4155-96b0-bebd0b5a94ea">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2909d91-2028-4ad0-814a-f6302c7b2e06">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d4187d44-360f-4c34-b502-19d91a26b09d">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fd30474-99d2-4ca5-a19a-5d8f3a3d23aa">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b72d14b-056b-4c84-af29-761c16ee72f8">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41d00f43-8622-43f3-9159-0ef73d29856b">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b75bf7df-1e2d-4748-a168-5925a65ca60b">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98444a31-0106-4d7f-837e-b2169188b88a">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64b0b29c-c336-4f3c-bf01-d7795071c43d">
                <emphasis role="italic">Fetus​Findings[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_690e4b3a-f590-4835-ac8c-5a7e2171e2aa">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_016e9dac-d36f-45f0-b092-2b8205c335fa">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_199ede6e-4f30-4f2b-9f3a-8285f1f61c6a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ac11223-1619-49d3-9f7b-95761403d497">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_931c6b2a-055e-4913-b1d6-0cc616354026">
<xref linkend="sect_9.8.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.9</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab4395c8-a4c6-4e7f-8551-d50e54386989">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d4c9688-7576-40c0-9707-4e3bf94dd7c8">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6022843-5687-48dc-9a99-7a45386cfe57">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_280d3ac4-023e-4b7b-b1ce-bb8e0d974852">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c978415d-64b7-4b29-8e62-d851ab1d1836">
                <emphasis role="italic">Subsection[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d8dfeb7-ec14-479a-b996-ae3b302db256">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2de00a8c-fd47-442b-89fb-f0e4766c026e">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7fc3d935-cdd4-4214-8e32-0dc18109824c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d4a3b99-210f-494c-bb8b-1ac57fceb053">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bcde49e5-6db3-4cca-8761-7183be2d18eb">
<xref linkend="sect_9.8.9" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.10</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f6b80b2b-2763-467c-864f-ea197f43b44d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ce856e0-8f07-493b-8115-ed6af91f49e8">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e584b57-f07e-4911-bfcb-31d2fb8ea47a">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e41b3c19-e073-46a1-9f60-4e14ea1e8629">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <section label="9.5.1" status="3" xml:id="sect_9.5.1">
        <title>text</title>
        <para xml:id="para_83b8059d-a17d-4309-9768-781c36898638">If entries are present, the section/text SHALL represent faithfully all such statements and MAY contain additional text.</para>
        <para xml:id="para_f2b60460-b14a-439e-88e9-54503968c483">The narrative text associated with an actionable finding SHOULD be highlighted using styleCode Bold. See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
        <para xml:id="para_67d3bb19-7bf6-45ae-baab-f838e3c1b254">Actionable findings that require a specific follow-up action or procedure SHOULD be referenced from a recommendation in the <xref linkend="sect_9.8.11" xrefstyle="select: labelnumber title"/> section.</para>
        <para xml:id="para_8e8efdfe-65a7-4b2b-b7a2-277b182d91a7">Communication of actionable findings SHOULD be documented in the <xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/> section.</para>
        <example label="9.5.1-1" xml:id="example_9.5.1-1">
          <title>Findings section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.6.1.2"/>
    <id root="1.2.840.10213.2.62.941494044785528.114289542452452805"/>
    <code code="59776-5" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Procedure Findings"/>
    <title>Findings</title>
    <text>
        <paragraph>
            <caption>Finding</caption>
            <content ID="Fndng2">The cardiomediastinum is... </content>
        </paragraph>
        <paragraph>
            <caption>Diameter</caption>
            <content ID="Diam2">45mm</content>
        </paragraph>
        ...
    </text>
    <entry>
        <templateId root="2.16.840.1.113883.10.20.6.2.12"/>
        ...
    </entry>
    <!-- see examples for other sections/entries -->
</section>]]></programlisting>
      </example>
      </section>
    </section>
    <section label="9.6" status="2" xml:id="sect_9.6">
      <title>Impression</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50fab504-2559-44cf-8fa7-571725297ae3">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_38443404-1e7b-44f5-ad6b-e02b4b68311f">1.2.840.10008.9.5</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1bb5acf8-99e5-4246-8ff4-d60435cd030c">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_11f8c580-3b00-4fdf-8ca5-de359b3984f6">Impression</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8eaf7d79-5bbf-46f1-81c2-231ff8333610">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a4cb6d4-fd38-4daf-a33a-5a9a7d7b4a4f">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_12c80f94-3776-4d0d-948d-ade2f005a136">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aefc066b-d143-479a-b366-2cee402f5d1c">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88d60e5f-7db7-4e78-8154-449d2e51cff3">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ed6ffb6-ebe4-401c-b847-3c0c75831df1">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc204a92-7745-41c4-9788-3b8d0d5eae74">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c745f61-07a1-48da-a8d0-c242a3efbd48">The most important diagnoses or other clinical conclusions that can be made from the imaging observations and other clinical information are recorded here. This section may include recommendations for additional imaging tests or other actions, as well as global assessments, such as BI-RADS Categories or the equivalent.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20c26673-ff81-4ae3-bb4e-8a6a3bcff190">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d2f9d7f6-cc1e-40ad-81c2-e3c7fc59e3fe">CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77fa4606-e362-455b-acd1-024c14205cd4">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d041cef7-b047-4564-b0ec-e141756bb799">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_db9fefde-53a6-4a43-b8d5-d2f52fdaf243">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ce0c43e-a552-45a0-966a-46c54067a659">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2867d28d-f0d1-4998-8d10-ba8a29b632c3">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88f0a3a8-9282-4707-b849-76159eebc32f">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a67ad65-14e5-49af-970f-c90c016fc3de">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b96f451d-b68e-48eb-b8a8-84467e9c55ee">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_234a1b19-8f33-4a3d-a654-63843d31728f"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_5a98dfd0-daa1-46f5-9dd2-7da54b0d6d0b">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_088e54d9-f7ba-4fe0-920e-4eb5e581f82e">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8bc29cc9-9663-4270-951c-a3e56b674c49">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c643d80e-234b-4798-a490-8e923aea95fb">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_950c4fd6-73b7-471f-9bda-4ab43e1b39e2">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_37203d45-5925-4a5c-ad7d-3babcc743d13">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1c3c1666-a4a5-4c47-89bf-ba90f39c3ac7">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0b041658-ccb1-4a44-acca-1ddbc5bd5d04">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c726d7db-7d93-4efc-b16a-aa0fac4ec281">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4374b12-db22-4389-b4f8-b2910e61ad52">
                <emphasis role="bold">Impression</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_27b288b7-224b-4ca8-bae0-7b24fff0cb7a">section</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6edc04c2-ea6c-4c22-926d-2633bb1743c3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f78ed24-4a3e-4f9a-bd31-a1ad2bcbd484">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_368ec998-2338-4310-8688-6c6a005cb660">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2967a18b-5bf6-4b98-b4be-5f0eed84cb1e">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4f5fba7-ec8f-4681-acac-cc26558f2761">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ec234ec-34c4-41fb-b34b-e428d5fd2dd6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e43e5693-0f9d-40c5-a014-1662ea09a7cb">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f04e9342-ce8d-40b5-a00a-7bdb6b5990cb">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c14c07f1-87f0-4d7b-a81b-0e5c53a5555c">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7965d49-5866-4f98-bde5-5487b251f84c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d89f1ee-8bf0-4128-aa64-75859d09a559">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23700373-1b9d-440b-804e-6128f2c4dbb2">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50202a6c-8bd8-4d31-a01e-6b36437468c4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b62e592-57a7-4136-8d9f-382249c218d6">1.2.840.10008.9.5</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b75b1079-27c8-4c08-9bc5-daba5ba558b1">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62a57955-8950-45a5-9334-613be3171eac">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b166fc00-7f55-492d-aa84-eb772c37c53e">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_674738c9-0059-4fc5-b237-3554eb8261b3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d3fb1d5-73da-4b9a-935a-67d70e5d73e2">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2005d507-93ff-4c42-8ab9-e6cca4abc794">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a04d8bfe-a426-4221-b275-72bf18e625b0">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74b4ad2a-0bf8-4a4c-9260-23403c9a0fd2">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da881560-3b90-4445-a95c-8fa35a6fee29">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17a8a955-1bb1-4125-831c-0070c20a6b20">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98a47724-8584-4a73-a569-cb9553564260">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_790d4d6b-247a-4eb0-b2f5-cfefb331754a">
<link xl:href="http://loinc.org/19005-8/" xl:show="new">
<link xl:href="http://loinc.org/19005-8/" xl:show="new">(19005-8, LOINC, "Impressions")</link>
</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4afd005f-5260-4ccf-bb6c-62a99859470a">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c75a4628-0165-4aca-8eea-89c624cec595">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_35615a5a-94a0-475c-9fcd-0f418ffcfd61">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_268e6e10-2a6e-458c-8a63-a62de0f30372">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_528ea589-1469-44b2-9a8e-eeb78e2358ca">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50c43c92-2955-462b-b74a-2482f9136d6a">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08dcbf37-91e3-4d1e-adb4-19a0a22aad30">
                <emphasis role="italic">Text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5349fa4-d006-492b-8b41-7b0675196e48">
                <emphasis role="italic">&gt;</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2314687f-d372-4177-8b09-876df5ebf699">
                <emphasis role="italic">text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7cdd40e-dd64-4007-9c90-6c4c8bd4ef1d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9984d33d-4a49-454f-b34a-0c3801f6a105">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd335fbc-0b13-4b9c-bd6a-8717a4be45b8">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1673733-b4e6-4095-b14f-9d4b1f3573a3">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_189c8735-58f9-4e34-b2b1-ff6f0b94a654">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e076c520-2640-4bc1-a803-7b2ff4368c3b">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0d8a631-aa88-42a7-b5fd-8dfd8b025c67">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_052b0c6e-757a-4ae1-8a76-816ca430e2b7">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b8baa709-1b8e-4556-a10e-0a5b830b9352">
                <emphasis role="italic">Communication​Of​Actionable​Findings</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9564d8b8-7da9-4d7e-9d4b-4c0e70bf6346">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36b5c006-dd06-439d-8809-a728253e8278">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f1eebff-543c-4c31-8577-5222cbaa6b69">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ae2c5845-9477-4230-9b2b-20286484273d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b1a03c6-2430-469d-b77d-04dc15d3c7f9">
<xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.11</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc7f297f-41cf-44b9-af74-acd5748927b0">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_852af1d4-b134-44d1-8e0e-902560e2fcbe">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e104369-e62c-4a9d-b6b0-1aa118b90bb5">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4979746b-7647-470a-b677-72d47aa67724">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8665cd07-a433-4168-8f99-00c9592f8ecd">
                <emphasis role="italic">Key​Images</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a098c44-deb0-40e9-b707-287b28846b82">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b1e192e-4957-4f68-9614-fffbbd63dde9">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43546e24-17e5-4e99-81cf-502147639343">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4915c2b-d8cc-4360-a1bd-d4e34a353dfa">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8c2732b-4312-402f-8dd4-d1dd895d703e">
<xref linkend="sect_9.8.6" xrefstyle="select: labelnumber title"/> 1.3.6.1.4.1.19376.​1.4.1.2.14</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c97755b-c529-4a94-aa65-2664361cefab">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4d073d1-0579-460e-8c4d-2f074c9dec65">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_940403d6-8b05-4f2a-9aba-bf0228112983">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a51c9c4a-d60f-48e2-a4af-9504dd5de8c2">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a26814d5-2366-46f4-8e3a-6e1cf1f54b2c">
                <emphasis role="italic">Recommendation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98810695-afb4-453c-bc09-e9163b5b8ee1">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07489a67-9143-48ee-80a2-fa0836291dc8">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e1809355-f5c9-4cf5-bebe-6ead53d04f14">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b7127c9-e7bf-4a2a-8800-98347c797043">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0816d2e4-ca83-4ca5-b277-bc56dd4f057e">
<xref linkend="sect_9.8.11" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.12</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1763acb-3880-4def-90f2-bdb9ff65d1f1">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3cfa342c-671e-4edf-a5b5-e88e22671a3f">entry</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9aecae8-9bb8-409a-9a07-1461c087e70a">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_474d2910-db78-4515-ab6f-5da6a409cef2">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64f1deef-05f4-4e68-ac44-6fef2fb3ae3c">
                <emphasis role="italic">Coded​Observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb978d6f-f6ff-4ee1-ba70-3ad71b143d00">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7650d34d-0e14-4bf3-af8c-f06c8c87a07d">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0418091-4c79-4c31-b8d3-f9ceef8ea854">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b6abc4c-26f7-4114-8493-a82d0d67af2b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c4526dbb-a7f9-4724-97f0-c7692c96728b">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87e1361b-6531-44ea-8e5a-7ec5419ea942">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <example label="9.6-1" xml:id="example_9.6-1">
        <title>Impression section example</title>
        <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.22.2.27"/>
    <id root="1.2.840.10213.2.62.994948294044785528.11422458954285205"/>
    <code code="19005-8"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC"
        displayName="Impressions"/>
    <title>Impression</title>
    <text>This exam identified ...</text>
    <!-- other sections and entries here -->
</section>]]></programlisting>
      </example>
    </section>
    <section label="9.7" status="2" xml:id="sect_9.7">
      <title>Addendum</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5ab8c02-81c9-4973-9c5a-7699621fb45e">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d7353de-be74-46e8-b4f3-6afcd35e71cd">1.2.840.10008.9.6</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f72d32ea-7f72-4107-8d8c-3c0e7c4da24d">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc0b3541-0b19-4ae2-991b-1364aad2852c">Addendum</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f8d0460b-2ea7-45a7-84fc-0d365bf7be78">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e7ad19d-92de-4ab1-aefb-1813f1af20a3">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d151e516-d7a3-4bcb-82a6-041b5f84d7ce">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b9ef586-89a1-428f-8dca-b9a63ae848b1">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7947e2d-e57b-4475-80d1-9711e23bb9fa">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_513cd462-a88b-49b6-a5e8-47d11f8f0e19">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c5b78e3-8046-4faf-a535-929fba96be73">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e8213cf-a151-4aab-93dc-b21709ca21ef">Addendum section for imaging report includes supplemental information added to the original document contents..</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d72a937-0e6d-44ed-b1d1-4d9a447e0eee">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_68f6988b-10f9-463b-b83c-977387293881">CDA Section Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4bd0739-6ff9-4b3f-beea-017b3a9e4dee">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2209eba-cedf-48d7-ae97-611d8912c1f5">Included in <xref linkend="sect_7.1" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9dcac74d-90b1-4ab0-96b9-e8868de812ae">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b04d9817-5888-4f18-97b7-9db163ef4465">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64bc863b-4850-4c9d-b934-dcb340765ccf">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_25181d5b-cdc4-4ea0-8441-8196b5cf0f10">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d3f3df2-1594-4888-9929-798a2800cd89">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81771f4a-d354-496f-ba69-14e393093751">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_b2fe6a62-12f9-42a4-88d2-15dee671667c"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2f3e45b1-41d3-40c5-87ab-7614b1782e02">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f76d4920-8d91-4c19-ab0e-c3a87dc247c3">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e5ae72cc-cfcf-41ec-b05e-e1bdc9c2b881">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e7ba0f8b-0516-4ce7-9e95-0636dc70d9a1">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f0109029-58e7-4042-bbd9-71394a8e3808">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_d87c20c4-ffc1-4883-a16c-afb2ec3fb855">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_33cd987a-b932-4f69-a4bd-d291f562a80f">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a757543e-79fb-44a1-9841-9a5a59f65e2e">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_d0509a21-a528-45ca-afc5-47e246213f97">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_660cc34c-40a8-4495-99b3-f8c7c1b788a1">
                <emphasis role="bold">Addendum[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0a5825d-945e-410f-8141-c59f6d01349c">section</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a81a4ced-1e59-40ab-8120-0516a52ce751">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8486d2cb-115a-4d5b-af1e-dfac656a3a71">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2147e9c-3781-4764-a25d-1fa4ee61fbda">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_65914d5f-5cfd-423a-84c8-6974a327aab7">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5b1be45-8e3b-495d-bab1-4b9a8bea18e6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c600e5a4-f439-454b-80a8-c8106c304f9d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f812246-e552-4b57-9940-94a7f9cbfeab">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_151bb70a-f795-437c-be81-68b9440628bb">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_68faea5a-50e6-4032-bf6d-7012c07d16e3">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_27c4a410-6603-4cf0-954f-13986de1923b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5cdb9d4f-8b75-4290-88ab-0a46f2c611eb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e89b891f-9054-4fa6-90d2-c8e94868ed24">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_78973937-bb27-40c2-a033-7dccf28bbbd5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_679dfb84-504e-400a-a082-1895f7a76b0b">1.2.840.10008.9.6</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0f48ca9-2100-46cc-af41-6e7adb78bc0e">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3f6377a-e668-446c-9050-f212abc8f680">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_310d5983-4340-469a-b1f2-3eaeb5413d1a">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fdeb71a0-8af7-4992-96f7-7069dad5754e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bed3144d-66dd-4f84-9c77-1e29d0397884">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_564a9eb5-8088-4af1-92be-f6c2f62cd5c8">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1be3860a-21c5-44cb-b075-42e7e3da5d25">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bdf48f23-79bf-4158-88fc-f1120a2faf89">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f6b702d0-5fac-4e21-970a-4b893b19cf5d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f60dbd8-8b5b-42dd-8b4c-205bcb6cb230">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3868ac4-0c0d-4252-9086-206544096482">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_03524b91-399f-47c7-bfcd-02967b058d64">
<link xl:href="http://loinc.org/55107-7/" xl:show="new">(55107-7, LOINC, "Addendum")</link>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_02df204c-43fb-4657-ae04-fe829d026790">Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4cee949-fd4d-4360-a9e6-dbb85e5995df">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67384acf-1f36-4f0e-bbb3-020344b67782">title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_446f261e-1bf7-40ce-b3b8-3da4e5c412d6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dee1a455-169b-4227-b866-43644289a772">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4dc643ab-3486-4c22-b203-7d47209605e4">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f701cd4-8d0c-4250-b356-2eb3bcaaabf9">
                <emphasis role="italic">Text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b435cc89-63d4-47e3-a3c2-2640e23764ad">
                <emphasis role="italic">&gt;</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf44b946-8ef5-4d10-9cf0-afb7b654ca7f">
                <emphasis role="italic">text</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8068b7e-087e-4bde-b6c8-4c0ddf246054">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_11bdf912-8847-430a-b56f-e39e463982d9">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e6740572-9248-451a-aef4-38a0d6be3557">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20f20613-a850-4ab7-9bdb-1beae280578f">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3626fb18-9981-4620-8285-44ed452939d3">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_722fafd5-ed28-4837-9b46-2677f0f0ab34">author</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b057b2a9-b709-426d-b2f5-2c1aa819d613">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0f236be-9418-4e51-bebf-a3f083a68e24">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_28b6bb72-69aa-49bc-a3b4-f6083f525aac">Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d6e4c4a-c747-429b-8703-7f86ef753a4d">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_34819e63-12ab-422c-ab62-88d6a8b315cb">time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f63fad4c-c05e-40dd-a3c1-47a9163f0687">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f697b3d-d759-40bc-8b50-b8db2d302774">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_01145148-bfd6-4462-9bd5-1c658b57757a">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_614950db-7efc-4afc-bd6c-bdde6a2f7126">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_586ebfb0-bcf7-4aa3-8842-7d35167be58a">assigned​Author</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9977293-c4a0-4904-a4cf-919846aa24bc">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d520f538-25a4-4036-b683-183661478d44">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_99e225f4-bbcc-4936-aae0-01644ba38c85">Author​ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cb481892-6b87-47e3-a55d-9c0a9152c3dd">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8c3f23d-db00-457e-b028-a6151dabf94a">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c8f9f92c-85f9-4dc7-8942-1c81f0ee9e39">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cbdc43c2-cda2-4da4-b419-d1901bac82eb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b5ec0d01-8b59-4e37-a471-0bd67826d42b">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95cf1be8-d403-4b65-a82c-b23611769ef1">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ba8aca1-8204-4b3a-bcb1-f4107fc6f3ac">assigned​Person</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ffbbb33-db1d-4fc0-8f90-e883797501aa">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73046455-3ec3-4d87-b0e6-92cb268d3777">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c402f954-bafa-43b5-87fc-3d7a54663e0a">Author​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_127a8bcd-23f9-440b-852f-cf228f68dbe9">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17b60056-74c5-4a37-9b46-39f38ec6d5e7">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8450bc0-a150-44de-b8b3-a02689e1ddbe">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15ed8621-05df-43a0-98b3-f00661fb135e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9d73013-a1b7-4bd0-a41b-d30b996dda12">PN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_583385df-2c69-4251-aed6-889722cd2a51">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f9a5173-90d7-4577-8f0a-693d6241d7c0">component</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5316ea11-8a4d-41be-9ef8-ecedc709f427">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d40bda2-a414-401a-91a9-aa71f694e1ee">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_234da3e9-8a01-44b2-badb-36c84123cb46">
                <emphasis role="italic">Communication​Of​Actionable​Findings</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d15a9f3f-53d1-49a2-8b3f-6b8d64eb8441">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_370de0b5-54a2-4186-a76f-de889e3d11f0">
                <emphasis role="italic">section</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9df3559-477c-46ed-a195-334434f9e015">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17b38c4d-5473-4d6e-9828-97958fd5a876">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3b41bc9-fa6b-464e-aa47-7702bc478473">
<xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.11</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17b57057-d106-449f-9f03-e4f7c3164e99">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15c07f35-5794-4bc0-a522-d868f2d6360a">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_858f6a71-4651-4886-ad85-512c489da513">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62a0f857-3fdb-4017-9699-5681fc019013">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <section label="9.7.1" status="3" xml:id="sect_9.7.1">
        <title>author</title>
        <para xml:id="para_68d09505-6f63-4fbc-950b-6b7f7d860eb6">Note that the Author identified in the document header is the author of the original report, as that participation sets the default authoring context for the report. The Author participation in this section shall be present, and identifies the author of the addendum, even if the same as the author of the original report.</para>
      </section>
      <section label="9.7.2" status="3" xml:id="sect_9.7.2">
        <title>component/section - Communication of Actionable Findings</title>
        <para xml:id="para_4cf360ee-0434-4973-bd32-eaee2987a016">It is possible for an imaging report to be legally signed (authenticated) prior to the Actionable Findings being properly communicated. In this event, an addendum to the imaging report is often created to document the communication of the actionable findings. This can be included in the section/text of the <xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/>, or using the <xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/> subsection.</para>
        <example label="9.7.2-1" xml:id="example_9.7.2-1">
          <title>Addendum section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN" ID="Adndm">
    <templateId root="1.2.840.10008.9.6"/>
    <id root="1.2.840.10213.2.62.7906994044785528.1142895428068506"/>
    <code code="55107-7" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName=" Addendum"/>
    <title>Addendum</title>
    <text>The supplemental information added to the original document...</text>
    <author>
        <time value="20140605143000+0500"/>
        <assignedAuthor>
            <id extension="23454345" root="2.16.840.1.113883.19.5"/>
            <assignedPerson>
            <name><given>Henry</given> <family>Radiologist</family></name>
            </assignedPerson>
        </assignedAuthor>
    </author>
</section>]]></programlisting>
      </example>
      </section>
    </section>
    <section label="9.8" status="2" xml:id="sect_9.8">
      <title>Sub-sections</title>
      <section label="9.8.1" status="3" xml:id="sect_9.8.1">
        <title>Request</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15c660d0-5ca5-44b3-b2e0-0b81359661fc">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0033a237-cf17-42c4-9c88-8ad96a8e92bd">1.2.840.10008.9.7</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da8f1b76-f8b4-4e62-8ffb-7b635cfca906">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_471a8a82-89a0-40fa-9743-c868039c79f3">Request</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_742e9077-5ab0-4947-a89f-d53ca19005c2">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5a7a472-bea9-4a10-96c0-9a8c5b98ba2a">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_443d8cf2-9ffc-452d-9c72-551013013f21">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7516d08-a7ee-46e6-953b-14b92440ec9a">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71232b59-ee6f-4534-891e-8d473a31e819">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca0aede7-84eb-4e9b-b8dd-4552a379dc18">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d2df1f4-038e-4544-8414-2fe226e6560d">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f6137c7-a7b6-4f50-8e7e-126cc4e1caa0">Information about the requested imaging studies and associated tests. It may include information on the reason for the request, and on any validation of the request by clinical decision support against relevant appropriateness criteria.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_895cf676-bb4d-48bc-bf12-2a9b3b24c28e">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5efd40f8-eef4-4fc6-ad1d-89a901bce430">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccfb046d-ccc3-484c-9fd6-a63e9b8feee4">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87d3a0e7-55bd-4655-9484-b25098b3bc3c">Included in <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d547de5-232b-4309-968b-75d08e316b1a">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b005de4d-827e-44ff-91cc-9223b7c2e0a4">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c61edec0-6649-400d-b8a5-379a9e29acea">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4789244-4f78-48a8-b7e5-22c184f10254">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b49a58a-7140-4bcc-8976-cea8670829e6">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f3098be7-7739-4cdd-9183-1880d916016b">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_715ab10e-9428-4d17-859e-daeaa5461c5e"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d330f836-b0f2-467d-912c-e7e072c0b65e">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_97736fdc-a5af-4803-a58f-ae15541c6029">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_eae38f63-360d-4264-ae25-99ff0b61aacc">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_453de20a-0057-4202-ae98-d4a72d05e817">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_f9841d9d-fa20-41f0-8217-ea875496ddad">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_01ad3e21-484c-43f3-8f11-9553ac90fa0d">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b96f7eda-a47a-4b5e-afe5-49ccd6ade79f">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_da73949d-c940-4880-a8e5-4c65ec377b7b">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_7155ce46-fad0-4fe9-9634-3465091600cf">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f11e51ab-6024-4417-97fd-c4960ab1323d">
                  <emphasis role="bold">Request</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_481a4912-e94c-4f72-8537-20e6c173106a">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ecaa91fe-a142-41b1-8464-ae4f62fac3da">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9717a4b8-d9ab-4f25-8afa-b4814f25505c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d1fb23a-c2c6-4851-a20e-b8504fc85e1d">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d013d70-7aa4-430f-9407-74b155d04ce1">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d5ef1ec-57d3-48b0-8662-3d38a66cfdcf">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3c1fdd1-31a6-4a49-a0e2-6ec7b23026f4">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b5053fab-c16d-48a5-973f-6f028babf037">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d85ba565-df6f-4643-be03-4e36b810615c">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c7f4d63-789b-4f2d-b3d6-3ad8f3257cec">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6ed1985-e667-4457-a897-6d5617959aff">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03de1675-d8ae-471d-8c3d-60c73c02ba22">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35bf3b27-3701-4383-8da5-8500b7700338">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_06074b59-170f-4ed1-8f9e-0ee77aa7cbbd">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66157a43-5c7b-4833-9656-a32410c541c8">1.2.840.10008.9.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b65b9ec-da69-4013-9d63-32be6843be34">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1ad116d5-5650-4aea-972f-21c7e2dccf21">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ddff67dd-26e9-40cb-9dd6-3a303bf11bc2">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_53c8402f-00cf-4350-8be2-f6d4a9ce8f71">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c17d199-ce6a-4c6c-ad4c-0e92102855dd">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_773d22b0-6c42-4e8b-bda4-3bbbeb15b4d4">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_699e7bf7-e684-4efd-ab28-8a4622a9e5ed">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed8059b9-ddd6-4951-93f2-31b4f6524fce">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77023ef1-9e8f-4916-84c6-2cc89df73e72">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65228774-868f-48fa-9f6a-35c7e5a1e3dc">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5cf19d7-4279-48f4-958c-7e3f7cf6a37e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_31ddb46e-e4dd-4b77-9054-ab4a4488b981">
<link xl:href="http://loinc.org/55115-0/" xl:show="new">
<link xl:href="http://loinc.org/55115-0/" xl:show="new">(55115-0, LOINC, "Request")</link>
</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67fdaf8b-3bb9-4ec4-b1d1-5b7320f11473">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e7d1f70-4e26-45c0-923b-c56793702dca">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_39d67fe4-da8b-43ac-a991-5fdfb351864f">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_674b6f6e-1482-413c-a8ab-aa2b8a9b001d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b0e0394f-ab44-4ce4-8fe8-97702cf8ec7e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8234b5a2-0148-43ca-9a92-eaefb7eca244">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1028a3df-6338-478d-b277-aa4d36a47a2f">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66069d81-b33e-4daa-9522-b4dec5bdc466">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4f6e518-eb95-4a0e-a450-00bd9d263492">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3456e0d-6611-461f-a6bd-5ea6f7fdb9db">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_655a6009-1d66-4fe8-88e7-e10c9d0a4554">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a73748a-1f8d-4f75-8967-803bbd37f797">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1520bb93-807b-4e51-86a8-813a0fc8f399">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f62aa35d-c8c6-4be7-823c-db8b3aadb40d">CDSRecord​Text[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab3b0a98-c5a0-40b0-be7b-be80e937909a">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e79beafc-d420-45a0-83a9-041d1b354dc5">content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ccd24f3d-d0d3-4be0-9911-f8df8bd843c7">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0dbb14bb-0c65-44d2-a07a-c7d1e2ed50d2">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c32108b8-8aff-4d02-8ce1-a7c9cee38431">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_235e6f7f-3482-4ab4-bd82-d2ec2732508d">*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b35f9eb-19e3-4ac5-a6d0-b323b71016e1">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90e8cc70-88bc-4853-99e1-7d166c81f223">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_381ac1aa-28c1-40cd-912b-45f43635e7f1">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a648564e-fe24-4af7-9e41-4fe3c669d342">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74fb3aa5-023b-4ea5-8fb6-c346295a8e55">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f9d2a48-ef06-4751-b96a-69593aa48475">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_21da3f0e-ab0e-4d03-ba80-5535a688b3c2">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed5c048c-363f-4319-ab3f-1cf9d6f96a7f">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c89038c1-8d97-42ed-8fef-1d7c03a1d459">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.1.1" status="4" xml:id="sect_9.8.1.1">
          <title>text/content and @ID – CDS Record</title>
          <para xml:id="para_6d9c00ad-035f-4f08-bc26-414bec690da0">The Request section narrative text block MAY include content blocks recording clinical decision support assessments of the request with respect to the indications, patient characteristics, and relevant guidelines. Each such text/content SHALL include an XML ID attribute that serves as the business name discriminator associated with an instantiation of the element. Even if only one content block is instantiated, the ID attribute shall be present.</para>
          <example label="9.8.1.1-1" xml:id="example_9.8.1.1-1">
            <title>Request section example</title>
            <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.7" />
    <id root="1.2.840.10213.2.62.7906994785528.114289506"/>
    <code code="55115-0"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC"
        displayName="Request" />
    <title>Request</title>
    <text>PTA (Iliac Angioplasty) for treatment of symptomatic
        atherosclerotic disease in both iliac arteries.
        <content ID="CDS001">Procedure ordered by Pat Smith, MD, NPI:8740944987.
        Classified APPROPRIATE by RadCDS based on ACR Select criteria
        at 2015-07-21 10:52:31 CDT</content>
    </text>
</section>]]></programlisting>
      </example>
        </section>
      </section>
      <section label="9.8.2" status="3" xml:id="sect_9.8.2">
        <title>Procedure Indications</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9b2318e-4e10-4d7d-ab71-033dfd15c796">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d21d3f79-4aae-42d8-9d53-223951986763">2.16.840.1.113883.​10.20.22.2.29</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88a52567-8dce-44f1-a7fa-114a304a97b2">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_844f4410-ff63-4f49-af9b-ce420eee9d0a">Procedure Indications</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_00e07df1-cfcc-4af6-8a47-59656fa09fd0">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a67e03a3-1865-4c98-add3-8468dc0fbafa">2012-07</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa9abe88-efdb-46aa-b0da-d645eb5a3f56">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_473643df-14bb-4d4f-9004-84696a79b6da">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_37cddf66-311b-4e55-b728-66c27c6423e5">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_29da686c-f383-4dbf-8308-aab1b9e103fd">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_924abc06-6ed6-4b32-aa32-f61a0cc266ec">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7a8fea5-5460-4146-96c0-74d806a8917a">Records details about the reason for the procedure. This section may include the pre-procedure diagnosis or diagnoses as well as one or more symptoms that contribute to the reason the procedure is being performed.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87c92cd6-d731-4e33-b986-82c66efc0f25">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8245dfb8-4c12-4a27-8094-3248746c5b25">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5191b161-2ee3-45b0-a78e-66124786ea80">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e1fd14d-3bc3-40fa-8941-52bd9f4745f1">Included in <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_14103e76-957c-4354-9f20-22e058f0394b">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d72fb072-749a-4623-b9ab-d1ebbcef72a8">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2703646-10b5-4cc3-871d-713ce194bb09">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed3bc980-ed09-4d6e-830a-5cfa5779299b">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57d8e9e9-9549-4228-9d49-6944d2577733">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1726b467-a438-4662-af7a-93cfbe78892f">From Consolidated CDA r1.1</para>
                <para xml:id="para_99c087b5-cb17-4b51-a548-2543cdc7fe82">DICOM-20150324: adapted to use optional Coded Observation entry rather than optional Indication entry</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_b5cd5e81-278b-4bbd-bb66-fb3cb49e224c"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d22747fd-b522-4f36-81fe-a29b02f9a6e6">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_c0a97ee9-4d03-4235-adac-aefbd7267d3f">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_91aae475-09be-4746-97a1-2a8738c2cb39">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ac0bd806-0939-4a29-b4cc-2177c8a4d546">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3d02d27d-5dd3-49da-8f00-1a9cd79770a0">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_dee8e0c2-a8cb-4a7d-9af2-9bea520886d5">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d723556c-7c13-409e-b2c7-785a3c569388">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_186ec587-432e-495b-90b8-469623222f35">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_64cece6d-3a62-49d2-8f6e-3f45799ba617">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b37851bd-0b4c-47fc-9618-734c5e889274">
                  <emphasis role="bold">Procedure​Indications</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa278699-a9e5-45de-9a69-21148f96db90">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c82a762b-8715-418c-ae31-6a63bafee328">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdd2bcc5-d030-4b52-8319-f5daaca90113">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9991d78-14c6-41eb-ae03-2e54ca71a3bd">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_23204cd6-11a9-4f3c-a427-5d06f1b5d2b3">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_664fba03-3c0e-475e-8875-abdf2c9fefd2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7100193f-f671-422a-8722-4a411680b3a1">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fc9a1692-7dc3-42fc-a797-0a9888780992">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a574fafb-919c-4df1-9f64-f6ee006c9521">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fe389bca-c304-4380-aca3-ef6fb3fe117e">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3745d04-edd7-4bc7-a6f0-4a693067feb6">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2d43731-f20f-4346-8e41-ec94cff14be8">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85c6116e-ce2c-48e4-a7bf-130122c268b0">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb6f7d16-d517-491c-8d44-699a8c5d124d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9923ac11-934d-4d58-a6b9-61d7f0d8e04e">2.16.840.1.113883.​10.20.22.2.29</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dad3afb5-4cfa-4910-bd32-e8951137f2f3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90de0c3b-fd98-490c-8b98-7fa728ef963f">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1720718d-9a24-4c25-bfdb-d27266340ba5">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a820b69-98ba-4643-993e-b16bc1d6a26d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b71a0dc-9559-430f-847d-059d615f33a1">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bdb3d036-5616-4f32-b703-6b040f37a04b">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f72c14bc-78df-4752-b868-b909413947f6">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a77a0d8-2832-4439-8e23-5d7529563093">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea8d7903-0929-4fb6-8036-2431b9ae0c63">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_035e8737-2dbc-4d52-bfad-be6988599c13">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9382dc5-3d39-4f94-be15-9cd7912de0f2">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb2c4ec1-c249-473b-9afd-b0d501aad5ae">
<link xl:href="http://loinc.org/59768-2/" xl:show="new">
<link xl:href="http://loinc.org/59768-2/" xl:show="new">(59768-2, LOINC, "Procedure Indications")</link>
</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d6cf2f1b-386a-46aa-a6cc-1d508306fae2">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc9d891a-203d-43c4-855f-650e217ea48c">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f1cadf0-f57e-4a4c-b214-2cb0d7fecb59">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36871787-fdbd-4e9b-8653-0a493423ce2c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec9a7f28-f559-44ad-b405-1214031ddaad">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3618da7-f1fe-4fdf-a542-8662b0ef35c9">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52d53993-195a-4dc6-ae2a-8d8b1f17106e">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b12f3473-c289-4c58-a2de-d0d469b57b94">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1e2d46a2-73fd-43a6-80b8-6886e3e64aca">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c20db4d3-0e05-4d2e-a60e-cde4d56962e6">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82a43e53-cb42-4236-867e-69b3a1531455">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d65c2a79-7585-414a-9751-b441254e2e7f">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b38ee3c6-1730-46cd-be61-bec13c4317e9">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb101b62-badb-4167-8f73-84521b259882">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f57f08b-7985-4ac6-8fc2-b68b39352d91">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88f4db7e-5ff8-428b-8c91-7e2d35acdbf2">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1448145-11ec-4480-926e-c501346b1b1f">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85b7de26-fe70-4e09-982a-1b46298ca75b">
                  <emphasis role="italic">Coded​Observation[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81879f3c-7e09-4fb2-8d88-c3d840b58ec8">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_389acde6-0b90-47a1-895c-f51cf9a3993b">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cdeab232-7ec5-4bc9-8cfb-6900801f4476">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d795bd67-3422-43c2-b8d9-45c95062b2d5">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fd0673b7-cafa-4cd8-8e77-4b41c723da11">See <xref linkend="sect_9.8.2.1" xrefstyle="select: labelnumber title"/>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_907b6a2b-7c17-4b2a-bbaf-acd786eae58c">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.2.1" status="4" xml:id="sect_9.8.2.1">
          <title>entry/observation</title>
          <para xml:id="para_f2633e56-006c-47a2-aee2-57b04f12bb59">The binding to the Coded Observation concept domains is:</para>
          <informaltable frame="box" rules="all">
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_aa010b6f-3372-4b39-8c2f-f963aab83186">Concept Domain or Element</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_057c84f9-3efe-4d70-adec-115157e0e574">Value Conf</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9cc92fbe-7a31-4ab5-8540-49224d5a22c0">Value</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8a7c8d60-c7a1-41d6-8b26-553e1cbfce17">Observation​Type</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d66fd1e1-b401-40c8-b270-e99537751ffc">SHOULD</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_24051109-c9ee-4052-9477-4c7de351905b">(432678004, SNOMED, "Indication for procedure")</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bc8996b0-f81c-49aa-bf1d-f24f7f1f5e6e">Other concept domains</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_63eab246-06d7-4d12-bd82-70b094b66589">unspecified</para>
                </td>
              </tr>
            </tbody>
          </informaltable>
          <note>
            <para xml:id="para_aea86e19-ee05-4b70-a413-c450aea8c595">In Consolidated CDA r1.1 the binding to the observationType is to Value Set Problem Type (2.16.840.1.113883.3.88.12.3221.7.2) with conformance SHOULD. Values from that Value Set are acceptable here as well.</para>
          </note>
          <example label="9.8.2.1-1" xml:id="example_9.8.2.1-1">
            <title>Procedure indications section example</title>
            <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.22.2.29"/>
    <id root="1.2.840.10213.2.62.044785528.1142895426"/>
    <code code="59768-2"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC"
        displayName="Procedure Indications"/>
    <title>Procedure Indications</title>
    <text>The procedure is performed as a follow-up for abnormal screening result.</text>
</section>]]></programlisting>
      </example>
        </section>
      </section>
      <section label="9.8.3" status="3" xml:id="sect_9.8.3">
        <title>Medical (General) History</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c2fc88be-4a89-4686-98f0-ee5af4d327dd">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7ad567c-46a1-431e-acc3-95b6eb188ef1">2.16.840.1.113883.​10.20.22.2.39</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_affcfb6d-88cd-41c0-aab9-cec3c7f9beb3">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55788247-7df4-4663-a04b-b2192c564f08">Medical (General) History</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb996254-a673-4447-8316-14684db75a43">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b052e1e-a462-4b9e-ae99-c8bb4b212f33">2012-07</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f63f737e-055d-4b4b-8348-4d42362efbd9">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c28c9db5-15ee-4d8a-8bb3-304fc704ce35">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e57f155f-4594-4c6d-8167-6192039b49ee">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a949681d-0567-4f55-a8a5-609b50265ac7">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61fde50b-8752-4fb9-8199-114e0e970846">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_334f19e9-c963-4f61-938d-6ffe67b6a992">History general describes all aspects of medical history of the patient even if not pertinent to the current procedure, and may include chief complaint, past medical history, social history, family history, surgical or procedure history, medication history, and other history information. The history may be limited to information pertinent to the current procedure or may be more comprehensive. It may also be reported as a collection of random clinical statements or it may be reported categorically. Categorical report formats may be divided into multiple subsections, including Past Medical History and Social History.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2162a759-5479-486d-bbb0-a3639034a07b">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54856b79-b9e5-406f-b8cc-3e5490cfa709">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_31691606-03b5-433d-b6ec-c306f0ce87d5">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16fe8ea9-6231-4238-ba3f-1654f523b72e">Included in <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1603ce37-3cf9-481a-9a0b-d8e7f1ccb4eb">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6188f4d4-a285-4142-948f-71cfa7622df9">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa006143-5793-4871-bcaf-415a9e49a940">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_000ec84d-285d-48ce-8682-d5d83c171365">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cfac98b1-547a-4b5d-b220-78df547d5857">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58d7aa19-a82c-4c74-ba5a-eb6a739018e5">From Consolidated CDA r1.1</para>
                <para xml:id="para_51614a8e-7582-4472-b771-ec105839068a">DICOM-20150324: Addition of optional entries; C-CDA templateID retained</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_61c3a16f-710b-4efc-8bef-916915d7efe9"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_efeae83e-0dad-4516-995e-fdec3865e20b">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_bf58fee9-2b73-422a-956a-5c808a5d1f9a">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d8e10598-83c5-4f89-a415-004371e0fa71">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4449678f-366f-4956-939b-1dba717d63be">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_bf2a921f-2ee4-4fd7-adcf-acd0942ebf45">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_406bdec4-51ca-44a1-bfcf-fa709858a166">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ce623aba-e5a7-42b7-85d9-6ef38a244b59">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_66b84a58-debb-427f-8166-f9c7bec4a455">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d36ebc3d-6421-4c84-aed9-c5978d45bd27">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ddce6a35-6a6f-4360-91e7-39eb1842eb45">
                  <emphasis role="bold">History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a2df966-1b17-499d-af54-79ab460d18c3">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c362ee64-7cb5-408e-a190-5c0be5598722">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ed80040-f5d2-4077-ad77-3cc014301692">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9fa187ae-80e1-4411-8507-91dc48274ad4">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_786a3eca-252b-4724-934a-76e641cd98f3">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_349b59e9-8c16-493a-8ef1-f60e978e02cf">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85a28d44-bd05-4aae-b7ca-c136774dd127">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0c7b383-5fa4-4662-a277-f0d0fa42ba78">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_502398ef-df2d-4135-b586-463dcf968a9c">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_482593e2-a79a-4330-95c1-485eafc78909">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3a6e25a-8db1-42dd-8cde-3e720ac838a6">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a9be9af-2a13-4b1e-80ae-3e8e07550a91">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b641d7cc-deba-42b3-ad88-501a0d0f0286">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_39b2fdbb-1bac-4adf-aaa5-c4133c0cae95">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_499080af-0c46-46c4-83a9-4e53ef4afa4c">2.16.840.1.113883.​10.20.22.2.39</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c66113a-f8b7-4bee-bbd5-7870082489f3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c04f14c2-649f-459f-a87e-b3c24188770e">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44795256-69b7-4da2-bfbb-6de708c08e54">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db4c76be-7bb5-42d0-8ebb-fca9ebb9156e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75b36e6d-3522-4b7c-8006-768117114314">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d71fe19f-5e14-4281-b735-f78fb607b713">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98dd187c-8ac9-4ec5-91d7-d6794b372a90">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_97c6c860-2ba5-475f-aa26-7b9aedcd48e0">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c2ea22e4-75dd-4782-b0c3-ddec8a21d393">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a266d1f2-9558-4b20-8228-425f4b605072">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09fb7ba8-9a2d-470b-a7c6-a766263110e3">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4720602-f297-43f9-81a9-c1f0613b044e">
<link xl:href="http://loinc.org/11329-0/" xl:show="new">
<link xl:href="http://loinc.org/11329-0/" xl:show="new">(11329-0, LOINC, "History General")</link>
</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87a80390-0a6d-4218-988b-b0a6b39c7423">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f42f04c-a426-4035-bfa7-bc2a45958d5c">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_901f5aef-af6a-48a0-90e6-d17ca1636733">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f11dc9c4-0c41-4d0c-aeea-be406b6a423e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba969bdd-d2f3-48e6-ac8e-bb01c3ad9fc6">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52048658-646a-4a4e-89c2-8fde5ebb64e0">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd32cdb6-b557-46eb-a7a7-9d4944a69606">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1e7a9b0-f658-433f-9272-8c043c9f9387">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_24f0b3da-30e5-4501-80d7-77284dd89c6b">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b02f54c1-a09c-4f0d-beaa-770a9de3fb0c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_27b262b4-a09c-4ee3-a115-060f2664efc5">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9d6450f-65f9-42b5-b241-52cfcef30460">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_980b9e51-38c6-4e1e-af43-a6a873706c32">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_04e5d816-5a45-4a98-a8da-625e48d70d2e">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7000d71e-6111-4559-a798-f1f5272ad60c">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_34c67887-7045-420c-b51b-90f60480309b">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8319caa4-ff0e-4ce8-8d40-da814f9a0569">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.3.1" status="4" xml:id="sect_9.8.3.1">
          <title>section/text</title>
          <para xml:id="para_edd4b749-1680-442d-b1a9-0df44ed4182c">In the context of an Imaging Report, the section/text should document any contraindications to contrast administration or other procedure techniques that affected the selection or performance of the protocol.</para>
          <example label="9.8.3.1-1" xml:id="example_9.8.3.1-1">
            <title>Medical (General) History section example</title>
            <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.22.2.39"/>
    <id root="1.2.840.10213.2.62.7044785528.114289875"/>
    <code code="11329-0"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC"
        displayName="History General"/>
    <title>Relevant Medical History</title>
    <text>
        <list>
            <item>Patient reported adverse reaction to iodine.</item>
            <item>Patient is smoker (1 pack daily).</item>
        </list>
    </text>
</section>]]></programlisting>
      </example>
        </section>
      </section>
      <section label="9.8.4" status="3" xml:id="sect_9.8.4">
        <title>Complications</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb93255e-9e76-4036-971c-808aa95b19f8">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5780e33-0173-4419-891b-56f86a288796">2.16.840.1.113883.​10.20.22.2.37</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_607dd17c-d480-4f89-912f-9b803dcbae89">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_436e3439-f5da-41d6-8be5-10715e4f5210">Complications</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76dd28ea-75b7-438a-a0af-78babddcd04c">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_782443b5-4797-4ca0-a5eb-2501f1e5d9dd">2012-07</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1697dad0-aced-49c8-bc76-56fbd94b8a38">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6de113d7-c72e-4ca2-a00c-ab8859d5c4ea">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_578b5924-5520-485a-98ed-0aaec8d68fa9">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f6ed58b-6469-4679-a0eb-8ef30aafceca">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25e360d9-32f5-4199-b6dc-7008eaefaac0">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_63d326e0-1a0d-4e08-8b44-697646408965">The Complications section records problems that occurred during the procedure or other activity. The complications may have been known risks or unanticipated problems.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_273a4e42-80ea-4538-8a30-05f161c77565">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e670e38c-8370-4ab4-b265-65e938b7ba64">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d5408c0-bc40-40f7-ad2b-88b7e2c22fa6">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8500783-bc3c-4e8d-90d4-1e4882ad9438">Included in <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ecba5935-baa3-4506-ab10-432b0cd4e342">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_941b37ca-3d25-4264-b11a-897cf38c3e58">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0fb73947-6602-4ac7-aa6e-80b20783c3f1">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50626113-0d8f-4489-87aa-6482c7384af9">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_daa74be6-9261-42d9-b160-1733115565ad">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43370ab9-2e69-4652-8e88-36058757edea">From Consolidated CDA r1.1</para>
                <para xml:id="para_bd81a1de-10b6-4a32-ac48-384ff1347c8c">DICOM-20150324: Addition of optional entries</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_ebd8f4e9-c57a-4a1a-a521-1152a7cd057d"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_1a2f3d0f-005c-4305-9fe3-e28caf802c8a">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b7bf20ae-4bd6-47c1-991f-a32e3ad8677f">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_8350df08-0ab3-4b2c-9ec9-e22cb3001523">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_767bdbe5-b8f8-4c71-bade-8dafc4dfdb1e">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_719639be-6fd0-4aea-94c4-c4d9fbe708e7">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9045cae8-18ea-4db4-8de5-b7201f04743d">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_c450d653-fedb-4c17-a65a-d88a557c4a29">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_970fb05f-cf3d-4e7c-b7e6-35231410c39f">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d81b4806-8588-4837-91d4-f661fd51eda7">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0bd7ac6e-6a8a-47f5-871b-2bb7735e0f39">
                  <emphasis role="bold">Complications</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_338fa55c-9495-40da-9a53-629c51c159fa">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3716c467-4c70-4376-98a4-3566bbce91a4">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c695225-0a92-4748-95eb-fbd1f9b86638">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc131f4d-de0a-46f1-9028-cbb2f9dc72de">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfad3758-caa0-4746-b274-f71ca8de80bb">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f195cc6b-f013-403c-b9b2-160e67bc05f5">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12ba0091-bf8b-40c3-bce4-8a049d339040">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ee3b8625-d4ba-49ec-8e53-93a7e60f99b8">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9b3b28c-370a-4d57-9b07-e914dfeadf57">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a476da0-e17a-4d7d-bdf4-c0c792794ea3">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42c55359-38c6-44a3-ad42-828a5dbe9ae2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb8aa531-0331-4dcf-a2e5-8a39d72ea6dd">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cda996e3-5b67-406b-932e-7b315761195a">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25a85f76-b817-43ca-b2b6-09b28ea8e13d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7baef70-726f-4454-860d-c7b0f4def275">2.16.840.1.113883.​10.20.22.2.37</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_738d70d0-fc0f-411c-878d-bf295eb4fb23">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55fe157f-e5c8-4a63-851e-d777d0f72f43">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eece3465-82cd-48b6-b91f-ffa5ae82f1dd">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0ab3d01b-3b41-4fff-844f-5d9b16242f21">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16c4f34e-f89d-4cff-8e94-2e0848719b0e">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ba1fa2f-939d-4595-ab5f-f1a1e0fceaf3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aea164ae-8684-49a5-a0c9-763e8f0b320e">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3f57f05-5516-4593-b12e-1f7246c61c18">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_626b614b-d1b3-4c02-8bef-1886235474b2">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb8486f3-c39b-46cc-8522-556899c6e94a">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d36d8e1a-c120-415f-90d1-0706f27dd58e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4cdbabfe-31b1-42f5-a476-5d6630d365e8">
<link xl:href="http://loinc.org/55109-3/" xl:show="new">(55109-3, LOINC, "Complications")</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4d989e7e-78bc-437f-a7b0-0ae6cf19a70b">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6dc4d259-d94c-4da0-bd57-67beae6da392">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96658f88-be5a-4e75-8440-2480f75a461a">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12b9aaa1-1ec9-4b4f-91dc-0ebf14182128">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17d2173b-3bf4-4e6a-8b66-b41cb26483d8">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e217ff61-afb9-4a8a-b7e5-b18abed3c612">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4d3aa4d6-c2cd-4f94-97e1-bc3807e02aa1">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5b12ba1-0150-4ba8-8c06-15a8f99d2262">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bda92d75-cf48-47c7-b331-8efd90e1334e">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35940d00-4f2b-4843-a58b-529dbc5ed6ae">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47852fbd-7d66-4e03-b0b4-27b832a3a35a">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb7c0a81-a6ca-4c33-9f6f-7828e69aba95">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42c50a0c-093a-40e5-aea7-440e7d1416a0">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a881e57f-d6f5-4db0-91d4-a4efb652aeda">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a6a24aba-61dc-4f01-8d79-e913dfcadebe">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa63d5ce-5b90-400a-a921-b1085d0158ab">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a641d475-a9d5-48a8-b214-ab0fc0daf329">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_071a7f1c-dddb-4cac-b702-8ab4c4086551">
                  <emphasis role="italic">Coded​Observation[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f20de06-ae1d-4d70-af57-23c76a160580">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_34897b5d-4fae-4b31-85f3-f504d680a78c">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1557717-cfb5-4f2c-80d4-6bbf7584cd34">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <example label="9.8.4-1" xml:id="example_9.8.4-1">
          <title>Complications section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.22.2.37"/>
    <id root="1.2.840.10213.2.62.70444786655528.11428987524546666"/>
    <code code="55109-3"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC"
        displayName="Complications"/>
    <title>Complications</title>
    <text>Immediately following IV contrast injection, the patient reporting itching "all over".
        Dr. Smith examined the patient and found multiple urticaria. The patient denied difficulty
        breathing or swallowing. The patient was given Benadryl 50 mg PO and was followed for
        30 minutes, during which time the symptoms subsided.</text>
    <entry>
        <observation classCode="OBS" moodCode="EVN">
            <templateId root="2.16.840.1.113883.10.20.6.2.13"/>
            <!-- Coded Observation -->
            ...
        </observation>
    </entry>
</section>]]></programlisting>
      </example>
      </section>
      <section label="9.8.5" status="3" xml:id="sect_9.8.5">
        <title>Radiation Exposure and Protection Information</title>
        <informaltable frame="box" rules="all">
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8de85f6f-eddf-4bce-ab33-de9736dcf831">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a18db70-fccd-4192-a5e3-27e438a18a0f">1.2.840.10008.9.8</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6a5f218-cb9e-4630-8075-e657734a2a4f">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4437634d-0ea0-4a92-9a05-a2ae1c534a5b">Radiation Exposure and Protection Information</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9ede7115-d3f8-4dbb-833e-725c41448eed">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db0e675e-9f47-406c-81a6-268a2f2cbb90">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_99072607-0c07-488c-8daa-a5f4f164d9ff">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_761075ce-6894-4a60-a011-6c9725974719">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b068f41-2f1e-475b-b213-2e268f10cb29">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ee7bb9d-07c6-4b01-98a5-bf6a688f605d">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a2dbdee4-7c79-408e-9f60-1eb13bd59746">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8435d997-6bfb-4ec7-a10b-5ca6eb754960">Contains information related to the radiation exposure and protection of the patient, as may be required by national or local legal requirements or standards.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5478b202-d63c-4228-9914-5c93354cec50">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6d1ae4b-908e-4040-8223-0aa5b4705ba9">CDA Section and Entry Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9f59fc8c-f4e3-4c42-8d40-a914b1b09bfa">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1eb68bd0-4838-4656-9380-a4dff0d675ab">Included in <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c22dede-7d0f-4faf-9eca-8367fb509a2a">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85a29c7a-1ad1-4b9d-9b0b-32869c935a62">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09c5dfbe-40a1-4e73-8fa5-e94f302b0617">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ab06542-8f89-4412-9d04-fddc5378b8a0">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6606067f-231e-4538-a7b7-9516894128b4">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6950d3dc-09d7-42f4-bf18-e9591960a855">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_d098e1d7-f363-4d00-a3ee-6ecd267d85ff"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e7e4c9a7-f0c9-47eb-a66e-1ef4f99d880e">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d8194211-3697-45e4-88ce-cbbbc50cc9d5">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_88549df5-dc5f-4ca1-98ac-72edf078d933">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_16031d96-8d6b-4dd6-a2b2-06b72e9bac27">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_119d86b1-329a-4c92-8a3a-cff798518c73">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0725c743-b09b-4b04-854b-962911d800a9">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_28a019f5-e4df-45b6-beba-6c9c9850f1ba">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e126873f-d7ad-4053-a677-1901322291bb">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a8648f29-c54e-4748-97a9-a8a268c47e55">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e4f00b1c-7faa-4243-910f-f92b38a4b056">
                  <emphasis role="bold">Radiation​Exposure</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a9f21c8-620d-48f2-99d7-ec4cfc8b4f72">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcbe3099-9c49-4eb6-8dcd-7d38860cfe2e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e386898f-bbd7-408b-bb8e-adddbfb3eed7">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_26b5a1de-e76e-49b3-9ee6-777c171fbf08">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f67cc31-5ea6-4dbd-81f8-38659f858dde">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cef95382-389d-4849-a260-2a74c7ac91fe">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2b9d8ea-28b8-4122-a3e0-dea9ec68fe1f">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3fce8f0-42fa-4197-9e2b-3c4b2498a15c">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85606e72-461c-4792-9f6c-b7356834cd1c">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c7ae4ef4-4dad-40cb-8fd9-ee5696ad12fd">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_999253d2-8a93-4a0a-8e99-b4ccff11308d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a63ee9f-9ae1-4b75-b407-a31a8a675c99">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_602f0a4d-9f7e-4825-afe2-7d750e9cfc92">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8fcae365-5e70-479a-bbe2-8e756ec74ee1">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b02c2c7e-2057-4c24-8357-88579890cd26">1.2.840.10008.9.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b757b772-cc57-4a5f-9c44-cb22903376dc">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed39cf10-10ee-4e5b-8b60-c58378234976">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7ccb329-1a51-4157-8714-b26308531af0">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8838acf-5796-4146-ab13-c92a200f1d4b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d0fb0fa7-9541-4b37-909a-cd47a0b3ecf5">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be68beff-972f-4c40-9c9f-7ea32214d1ed">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ee171c2-02a8-4e77-8b87-b0085300ccb1">
<link xl:href="http://loinc.org/73569-6/" xl:show="new">(73569-6, LOINC, "Radiation exposure and protection information")</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_19d5ff70-25c1-4bb2-afaa-5b65712d9bcd">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d82e7fe-6d24-42a0-b233-3823d929d0b7">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_780c46cf-7d9b-46f2-9ba5-0bd3233328fc">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5bfd9a3-348c-4f7f-b481-3323c1635c17">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5954fbba-8b77-4589-b422-c29ad216bc83">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15532524-c43d-42a1-9929-91e00053ba56">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b98b1b34-2b46-4157-bd06-a5d77f5bd87a">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb10c934-53ab-4bd2-92e7-ab5860af8d6b">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16e98c0f-c389-4ad9-895f-a7b47967eea3">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30557b5d-8380-462f-911f-2764ae875a01">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_93486590-ff58-4ce5-a8ae-3bb8bfe23aa1">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17555be8-32db-46b6-a208-dedc1b2f7242">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03e36651-af28-4174-b55b-843a0164b6fc">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d719742-6988-4ce4-b2ba-cb30d7db46c4">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c867ba52-fa8b-4662-b58c-3658611131a8">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40ce4020-8347-4f3f-9739-692fcb9ffc7c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec3f1bbc-a9fa-4edb-a634-7606e9b78c5a">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6712250b-975b-4c15-a29c-185d48839dca">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a79695a8-5a4b-4651-ba76-1418747eaaab">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75714239-dea8-4a56-a28a-583036e8e1d9">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b459081-8d32-4112-880a-de1a51252b8b">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ade5b11b-e3e8-4fcf-aed7-e8b441321f3f">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20da4c71-06ff-4b32-8d57-9c2c6a309b37">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b78c54ff-52ab-4580-88d6-81f0476fa8dd">procedure</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f2d6fe1-0b10-4345-9d04-f69913ac22d3">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e253f7b0-265a-4b43-b08a-49db61e554c0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a6e046f4-6e05-4d28-b3ae-7030d0f24a08">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2b47a6ea-4a83-4a1c-ac55-24cfbded61e3">@classCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69583477-7d76-4f55-95f0-b13c04b0f41c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f75809a-a626-4a3b-acc7-72f145ab750b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a99d1520-e774-4a97-af8b-1e4f0df7ef29">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f7a4f300-567f-40ed-bedf-695bd8c81e34">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9dcb21df-c8eb-41b9-aaa0-8eff0e8183c0">PROC</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8f88eb63-00e6-4dee-917b-029f4c58a613">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92badd82-cf8a-4260-9987-68a092e1a004">@moodCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8ed45e9-4d83-4b6a-b4e4-44ef36354005">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a73a5f29-c4df-4971-a6ed-e10d617fa358">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9dd4583c-e2e5-4d63-a4cc-dfe53ee3f424">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a27f4cf3-4c3a-461c-998f-e3310229f83a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d06ee02-28ac-4aa1-8d8c-d586664c34d6">EVN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0be509ff-72ae-4934-ba76-baaa78ceda17">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9cc1b800-adf4-4233-8680-1a8af833be38">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44391246-b87e-4453-ac7a-b01f4b19a3ed">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38d061f7-d1bd-4307-b7f4-c6254b2c6ac9">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3ca1cd5-ad40-40e6-8fbf-bc3b985a34e0">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d03afe66-148d-44d6-a239-cf9b6aeb2b7c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f1d356a-c3f9-4862-ba0c-7cb33bb087d9">
<olink targetdoc="PS3.16" targetptr="DCM_121290">(121290, DCM, "Patient exposure to ionizing radiation")</olink>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81acc73f-e4ca-493b-a15e-35fc84b6f5b3">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e751899c-c081-4072-9864-8099834f93ef">participant</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef8b290b-7f1f-4d26-a59b-fa3896ca9bef">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed5fa998-46de-49c3-8a64-090e2476a61c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec7a1646-569c-4622-b0cd-51b3055e0da1">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6747031f-3424-45dd-8c44-51bea8c37af5">@typeCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77d0ec48-e04e-4daa-9383-beda65f83c11">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f58ca5fc-b652-40d4-ab33-3b5d9e30eb3f">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_edbc8ff5-9e4d-402b-80ba-c1d535382658">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b273a35-7bb2-4cd1-9cb7-65e60cd766b0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c8783a5c-04be-499b-8a16-1407558ac4f5">RESP</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca9ccbc3-728b-47b8-8707-c7bc1ece5d0d">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d6be714b-02f4-4471-9509-ae619a192521">participant​Role</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_960a62b9-d31d-4b3b-8447-6974bf6cbeb9">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7dca478-51dc-49f5-b6bb-3c333d36f703">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5c956f21-d47d-496b-bb50-48b0f17afb67">Irradiation​Authorizing​ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e6ab77c-f26d-45ba-9923-8fc07851f075">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_368fb1e4-b3cf-483b-bab2-55cdade5f72b">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82f37dd3-ca15-4b75-bbe2-283e4269882b">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71c84509-0df1-4530-a068-6a70800af6fa">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9e88b5c2-ef88-4c6a-9d23-3649cb9dbc3a">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1863c18d-2687-47b5-8639-77ff773abafe">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a9b7cb6-cd96-42b1-bab3-9a868c0c9e04">function​Code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa92a450-ddb6-4c1a-a355-bab4c73a7a61">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aecca667-541d-4874-a144-874c66dd723a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42af0f27-973f-41a6-8d6f-717992fdbe3a">CE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_515d9803-0f74-4c3f-ae42-ecb162cf258d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d24a8736-c893-481a-a80c-0a401a1452ef">
<olink targetdoc="PS3.16" targetptr="DCM_113850">(113850, DCM, "Irradiation Authorizing")</olink>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5fbd1736-38a0-4d9c-bfd8-358bf2c111d2">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1c15451-8787-402b-a6c6-4f2abb63588b">playing​Entity</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8cbcb4b-563a-4397-9dc8-46579b56ae3a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15233d93-aed5-4a0b-bbf8-b82ee12677ed">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4d786bd-a293-4315-ba45-7554395aaf76">Irradiation​Authorizing​Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab4c9ed1-eace-472a-b47c-7552e22ec69e">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4c79df2-8d77-48d2-863c-156d6c3cc660">name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9f869204-6b5f-46c4-9b73-382bacac75d7">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6e8374f5-900b-4cd3-8237-f6a222a56919">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d23b705d-dfe4-463a-9e69-0b744299848c">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b43828e0-bc57-4cf0-b332-a8aab760a360">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_556ac97a-6313-4441-a164-98265578539e">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9a9138a5-a771-4b10-b471-a42d6990de03">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5945621-4108-48d3-89fc-74695ba5b2ae">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e75bc86c-e62e-4e07-a1b6-1733fd819606">
                  <emphasis role="italic">SOPInstance[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9aa119d1-b451-4606-9471-20a9e15b77be">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b800d64-9226-4975-8776-40e3851cf8c0">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1fec88c-b34d-4550-b954-aa29ed1aef73">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c7aaf314-78be-44f2-9b78-7edd95092abb">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0677d5bb-453f-464f-86ae-c6f9e1776921">
<xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.18</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_18ab998a-04b8-4019-91bb-6450a83e9bab">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50bce47e-7f36-495d-ac50-6c1c8f1f9399">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8cbfae4-b57a-4128-9a76-914c9cc9e3fc">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5866f74f-3eba-45d5-b077-6bfd2671b86d">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c9ec160-763d-4c9e-96cb-8b3c325fb0a7">
                  <emphasis role="italic">Coded​Observation​[pregnancy]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d9c777a4-9a04-4883-bfcf-9141a255ef18">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea6a03d7-79e2-4a23-93a7-4820d8d9cd07">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f952ad29-e738-435f-aab3-996541327f24">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad43a6fb-9607-4739-ad89-666961d8825b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d8b142a-e660-42b5-810e-75cf9d9c1eb4">[See <xref linkend="sect_9.8.5.4" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac3927ba-e445-4397-aff3-12e68f202c14">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c5318b0-94fb-4578-a39e-16c41d6eb638">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84d7c564-52c4-474d-9665-9eca5d0bf4b9">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61062692-9b00-4034-9650-5659a47dd537">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc41c16c-4ebc-4689-9b7e-fea92da5130d">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fcca72bb-d0a1-4e90-988d-b3fd769caae5">
                  <emphasis role="italic">Coded​Observation​[indication]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_26c083e3-e492-4a4b-a4d3-485b2dfb2c44">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9f33672c-c414-4caf-8879-62d2b5bbbb50">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7fb9576d-e9c1-4333-97f9-8112a083e711">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e066b28e-51be-47f7-8266-621e2cd87753">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a2f4e59f-bf5a-408a-8bd4-05337fbc0048">[See <xref linkend="sect_9.8.5.5" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c42f2594-192b-4c93-8c55-e47d68938485">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6c6cde6-f132-40b3-aea9-ad10e0b5472d">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_732bc35d-ba90-4a58-b371-46488ff843b8">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_739dce2b-b294-4886-89d0-5f6796f21007">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88290703-e30a-4aee-b19c-e8d48c8f2236">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2907a8b6-4636-4961-8e07-e338b7b26638">
                  <emphasis role="italic">Quantity​Measurement[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8307ac52-fafe-44ce-ae32-70e2fdb483c1">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eac29b5b-8352-40cf-bfb3-749112824f4a">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d450aaf-820a-47a0-a913-2e61c5168b3e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1150e24-1762-4a9e-9893-e7838d366bbd">[See <xref linkend="sect_9.8.5.6" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84fbcc19-6b4c-4088-a05c-23678f1ca4a5">
<xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.14</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b1a6776-ef57-41c5-81da-8b2710fd97a7">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_916b2f53-5727-4248-bc23-5ca9efbec039">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed99fc90-754f-4c41-9a99-7fbe800b6964">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16cd4ff2-14eb-4c0e-9210-63fdd864364f">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad04097a-588d-4546-ba22-4ac7724fb00b">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ea96dc6-9718-4425-ad90-9262cb42f75e">substance​Administration</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57407ba0-b8f2-441e-9015-0975df4ec5df">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91e6a8bc-adaf-4641-b08c-202a481cb5b8">@classCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d2feb38d-f990-4d9b-b2dc-57a3f75c75c9">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5c5c76f5-5518-4c7e-880b-9f532315c07d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d309815a-f8c9-4447-82a4-aefcf64141bd">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_53230308-9d3a-4bb4-905e-462b2f084bfb">SBADM</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40a6ff55-3192-4081-bfbc-035bc444a495">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77a8723e-f623-4f1b-9c69-4e38890b1d16">@moodCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9b99807-3417-4c56-9601-799d2882735b">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_967dbb19-093a-48e0-abe7-baf236d54178">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e624b2c8-d53f-4588-8a74-3219eb48b3b7">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fea24206-75a3-4f88-9284-5579ada62f24">EVN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e480e094-0fb8-4fa3-92b3-d2def3c77939">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e0eafe6f-f72a-4f82-82c7-daacb92ff51f">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74405f64-5640-4571-9a3c-f11429495e1b">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1e756cf-1f21-4482-9f5f-e43b97d78c0a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef7df6ee-e23b-48f0-a6bf-fe0a47ec80bd">(440252007, SNOMED, "Administration of radiopharmaceutical")</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b4c500e-184a-44b7-bbf7-0590f63e348f">Radioactivity​Dose</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2840055c-5415-4948-8588-396760b30361">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a39a0c83-7b01-438f-a081-0f2258baeeb7">dose​Quantity</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8851ab2-56d2-4b24-ba10-63a83b06248f">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e827c4d7-edf7-4e15-8761-f7e2fb1f6977">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_375c965d-682e-4360-bc86-7232d3ffe7b2">PQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa6edc35-0e5e-4807-8836-50c22b4387eb">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed53dea5-6fb7-49f3-bb1a-eaaa9325ebf7">consumable</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_560f716e-bf03-4704-9036-0225bf37b2ca">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5e44366-ec94-4698-9f12-8b0994ee3cac">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73a93265-5355-4c1b-b0d1-bb76189d1122">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_95c11608-2038-4d67-8dae-c1ee608c3a21">manufactured​Product</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c7109c4-dfcc-4a9e-9077-278066ffd363">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_63e7ee57-ec59-4d2c-a24d-f1d32e8f0b7a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_23372bd9-2d87-4659-a43d-746c6530106f">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcbd77dd-f578-4b9e-a3db-16949a5b7b32">material</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2ee0218-aecd-4410-976d-b214075b30b2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db04d7b0-e55d-4090-ae91-1d6954d5cc8e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62ece74a-5dbb-4388-9b97-056740f1a8f0">Radio​pharmaceutical</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8821b459-d694-4bcf-933e-8bfdb27a3e56">&gt;&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9d3e10f-d51e-482e-b973-eb9948b71d20">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_419c5bfd-67b1-4117-ae72-6de8034f5881">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_abf6acb3-a8ee-40dc-bd7f-608e29b7b10a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff6af4c2-a296-4891-949d-f0aa129d27c9">CE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ce3aee78-2864-4937-81f3-d5550f9070ab">SHOULD</para>
                <para xml:id="para_031e8cbc-6a44-4754-9037-57835c5bc070">CWE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7f6d4b24-dcbd-48b6-b3dd-5a37f0f2c0a3">
                  ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_25" xrefstyle="select: labelnumber quotedtitle"/>, or <olink targetdoc="PS3.16" targetptr="sect_CID_4021" xrefstyle="select: labelnumber quotedtitle"/>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_887d8f86-d871-4993-a594-f8b27395db0a">Free​Text​Radio​pharmaceutical</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_660eabd6-680d-4b29-a515-318e303c2e5d">&gt;&gt;&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_732aa7ad-1910-4c8e-bb8c-4d472af7c57d">original Text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb535d26-e154-4f3c-911a-aa46061f4b93">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_13ff0b4a-7c06-44f9-a82a-b89e1cccb62b">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74b1f416-11c4-459f-a0fd-7c3a570b04f5">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.5.1" status="4" xml:id="sect_9.8.5.1">
          <title>text</title>
          <para xml:id="para_c419b5c2-c29d-432d-a921-5dd748da5b7d">The section text SHALL contain information related to the radiation exposure and protection of the patient, as is required by state/national legal requirements or standards, for example:</para>
          <itemizedlist>
            <listitem>
              <para xml:id="para_7165f5cf-427f-42ab-8da5-74b4c0ad18e0">information on the indications for the procedure</para>
            </listitem>
            <listitem>
              <para xml:id="para_6a6aebb3-314a-47bf-b600-825220126ee8">the name of the "Irradiation Authorizing" person who is the clinician responsible for determining that the irradiating procedure was appropriate for the indications.</para>
            </listitem>
            <listitem>
              <para xml:id="para_cfa67bc8-dae0-47fe-a17a-12de0e504b11">summary information on radiation exposure if ionizing is applied in the context of the current procedure (detailed specification of exposure is out of the scope of this textual summary).</para>
            </listitem>
            <listitem>
              <para xml:id="para_7e7c253a-8953-4af4-b342-f4f348dfe79d">information on the radioactive substance administered if radioactive substance is administered in the context of the current procedure.</para>
            </listitem>
          </itemizedlist>
          <note>
            <para xml:id="para_12f31bd2-50fc-47d0-86c4-adca9b001f77">Compare to <olink targetdoc="PS3.16" targetptr="sect_TID_2008" xrefstyle="select: labelnumber quotedtitle"/>.</para>
          </note>
        </section>
        <section label="9.8.5.2" status="4" xml:id="sect_9.8.5.2">
          <title>entry/procedure Patient Exposure</title>
          <para xml:id="para_80b5584c-749d-47d1-a11e-403d45169636">COND: If modality is CT, MG, NM, PT, XR, XA, or XF, the section SHOULD contain a procedure entry for the exposure of the patient to ionizing radiation.</para>
          <para xml:id="para_8e8404be-f816-4dbc-9322-d03a7c1e883e">This entry SHALL have a participant, the irradiation authorizing person who is the clinician responsible for determining that the irradiating procedure was appropriate for the indications.</para>
          <note>
            <para xml:id="para_690ec801-d9eb-4e2c-9fc9-160cb2ad9328">This may be the same person as the performing physician identified in the header.</para>
          </note>
        </section>
        <section label="9.8.5.3" status="4" xml:id="sect_9.8.5.3">
          <title>entry/observation SOP Instance</title>
          <para xml:id="para_418915d1-35dd-4995-ace6-ab9c8e236509">The section may include reference to one or more DICOM Dose Report SOP Instances that provides a detailed record of exposure.</para>
        </section>
        <section label="9.8.5.4" status="4" xml:id="sect_9.8.5.4">
          <title>entry/observation Pregnancy</title>
          <para xml:id="para_2987d45b-9eb4-41a3-8590-e856b4bbe68c">COND: A coded observation entry SHALL be present if the patient is female and child-bearing age.</para>
          <para xml:id="para_be65a043-a762-4500-b4b5-38d3dd0b1806">The binding to the Coded Observation concept domains is:</para>
          <informaltable frame="box" rules="all">
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9e18aa5c-0c28-4a75-b944-38237e88a3d9">Concept Domain or Element</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_7aabdfd0-40f0-4c57-905d-faca044be25d">Value Conf</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_bd231bb4-325f-41a6-9a3a-f7a5352bf585">Value</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f0e376d5-5d0c-4e49-bdc2-ef3547ec1ab8">Observation​Type</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c5f6b76f-4399-466d-8308-2b2916d76b92">SHALL</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_af807a68-f6df-4c87-bff1-edd6392f6242">(364320009, SNOMED, "Pregnancy observable")</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_189c742f-4fb8-42a0-878a-722c20cd1e8c">Observation​Value</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4bf35ef6-27fc-4e6d-925b-8129bec421a0">SHALL CNE</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cda28563-ae52-4b49-9090-9ec4b2f0f52d">Value​Set <olink targetdoc="PS3.16" targetptr="sect_CID_6096" xrefstyle="select: labelnumber title"/>
</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_09e719c1-3a1e-4cca-b813-57acc333e1b6">Other concept domains</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6982959d-e59b-4e7b-8ab8-eea2d82d77a2">unspecified</para>
                </td>
              </tr>
            </tbody>
          </informaltable>
        </section>
        <section label="9.8.5.5" status="4" xml:id="sect_9.8.5.5">
          <title>entry/observation Indication</title>
          <para xml:id="para_e0f793c2-b88e-466f-9de3-5420a8ecfe27">An indication for procedure recorded in this section should be consistent with any indications identified in the <xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> and/or <xref linkend="sect_9.8.2" xrefstyle="select: labelnumber title"/> section. It is included here for conformance with regulatory requirements in some jurisdictions for the indications to be specified in the context of the radiation exposure information.</para>
          <para xml:id="para_d696cb99-4345-4672-bc9b-ac6ec132db58">The binding to the Coded Observation concept domains is:</para>
          <informaltable frame="box" rules="all">
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2809dc67-1de2-47ae-b6ab-6b0645492d8d">Concept Domain or Element</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_3d70f600-88af-47f4-9fc7-b52c327f7dbd">Value Conf</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_01956649-aacf-4fb5-9389-5117903c8782">Value</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c0dd6b93-8f92-4299-a96d-7d7f705207e2">Observation​Type</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b2353f32-14a5-483d-a86c-83741e18a099">SHALL</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a25ac00d-f2b9-4387-b1bc-7ac7d4c1217c">(432678004, SNOMED, "Indication for procedure")</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_da422bb6-bb12-45e8-aa7c-6938697a35b4">Other concept domains</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_be561928-06be-4bc2-ad19-94ad59442b0a">unspecified</para>
                </td>
              </tr>
            </tbody>
          </informaltable>
        </section>
        <section label="9.8.5.6" status="4" xml:id="sect_9.8.5.6">
          <title>entry/observation Dose Measurements</title>
          <para xml:id="para_6262142c-0e61-4de4-b663-bd33872db499">The section may include multiple dose measurements. The binding to the Quantity Measurement concept domains is:</para>
          <informaltable frame="box" rules="all">
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_01d38b05-8c61-450c-961e-38a872a6abfb">Concept Domain or Element</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_6907d9e0-b221-422c-b146-015569dafaa1">Value Conf</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_12ea916f-8795-4f65-ba37-dcbb9f1d4e01">Value</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_881c25c0-039a-478a-81c7-d8a19ef563b5">Observation​Type</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_070a1e4e-3df9-4266-9896-8aeff4756fcf">SHALL CWE</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_23fa9ec0-6936-4202-b386-2b99bf9f4ba9">Value​Set <olink targetdoc="PS3.16" targetptr="sect_CID_10050" xrefstyle="select: labelnumber quotedtitle"/>
</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_da060a29-82d6-4dd0-bceb-d687abde6837">Other concept domains</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_402a7714-aaa7-41c8-b95f-636170b64a79">unspecified</para>
                </td>
              </tr>
            </tbody>
          </informaltable>
          <example label="9.8.5.6-1" xml:id="example_9.8.5.6-1">
            <title>Radiation Exposure and Protection section example</title>
            <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
	<templateId root="1.2.840.10008.9.8" />
	<id root="1.2.840.10213.2.62.704478559484.11428372623"/>
	<code code="73569-6"
		codeSystem="2.16.840.1.113883.6.1"
		codeSystemName="LOINC"
		displayName="Radiation Exposure And Protection Information"/>
	<title> Radiation Exposure and Protection Information</title>
	<text>A dosage of... </text>
	<entry>
		<procedure classCode="PROC" moodCode="EVN">
			<code code="121290"
			    codeSystem="2.16.840.1.113883.6.1"
			    codeSystemName="DCM"
			    displayName="Patient exposure to ionizing radiation"/>
			<participant typeCode="RESP">
				<participantRole>
					<id root="2.16.840.1.113883.4.6" extension="980003719"/>
					<code code="113850"
					    codeSystem="2.16.840.1.113883.6.1"
					    codeSystemName="DCM"
					    displayName="Irradiation Authorizing"/>
					<playingEntity>
						<name>
							<given>Martha</given>
							<family>Radiologist</family>
						</name>
					</playingEntity>
				</participantRole>
			</participant>
		</procedure>
	</entry>
	<entry>
		<observation classCode="OBS" moodCode="EVN" ID="pregnancy">
			<templateId root="2.16.840.1.113883.10.20.6.2.13"/>
			<id root="1.2.840.10213.2.62.7044779.114265201"/>
			<code code="364320009"
			    codeSystem="2.16.840.1.113883.6.96"
			    codeSystemName="SNOMED CT"
			    displayName="Pregnancy observable"/>
			<statusCode code="completed"/>
			<value xsi:type="CD" code="60001007"
			    codeSystem="2.16.840.1.113883.6.96"
			    codeSystemName="SNOMED CT"
			    displayName="not pregnant"/>
			<effectiveTime value="20140914171504+0500"/>
		</observation>
	</entry>
</section>]]></programlisting>
      </example>
        </section>
      </section>
      <section label="9.8.6" status="3" xml:id="sect_9.8.6">
        <title>Key Images</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d24c857f-cb9d-4a2b-8d0b-e8b20299d474">
                  <emphasis role="bold">ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eeae153c-7005-4068-b9c0-b1a93643a7d7">1.3.6.1.4.1.19376.​1.4.1.2.14</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35348859-4bbd-4215-87f1-f62eaf2e0674">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ffe0cf97-e348-4b42-a836-ab6ff8cbb616">Key Images</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09c204dc-ec02-4ab2-996c-ebd0cb09a9c9">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b070d34-a3cb-46e5-8e70-5b51b5db5bc0">2011-07</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f0d278f-46b2-4cd2-b738-1cf15c6680ba">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0db25711-6db9-484f-9109-658c5da41b3b">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41813eae-85f1-4408-b203-82f9bec8f860">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_551f0d6f-8a40-4aa6-bc0f-9c756472ab17">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8e15481-df29-45fc-9d20-4aac5316b61d">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_804e5c39-8e2e-4da3-abaf-9d15cb6b6eb8">The Key Images section contains narrative description of and references to DICOM Image Information Objects that illustrate the findings of the procedure reported.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ca19343-82b8-4401-b24c-80f77406fa01">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1972c67-f122-4da9-979e-af01d3879aa2">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_86a74c15-912f-4198-babb-e131701b7fc3">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_423a4c29-0197-4bac-9940-65c2e4faf9ce">Included in <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91e924aa-9579-46c6-8c01-58d9db5f9817">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6164fc16-d308-4bae-a19b-f99f551060b2">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c8818da0-ecff-4691-93a5-929d7b85d3f1">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a61587de-bd0b-46d6-91a6-4293f826411b">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56c286b4-09c6-4b12-8072-0df6c4f7e1ac">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_49c748e0-524e-40e5-825d-d49d92fb53a5">From IHE Cardiac Imaging Report Content</para>
                <para xml:id="para_1a40d3f8-ea89-427a-a647-da995ae73f79">DICOM-20150324: Addition of optional inline image (observationMedia)</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_610e7c01-2bac-41f9-90fa-42a7af971871"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_7b44c912-1d9d-4be3-af1f-cc302d86fd11">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_c902ffce-da99-4d98-a5f9-44760096f8e6">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9695a601-e7e9-47ab-b804-5040a541c262">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_1fe9fab6-0fa8-4276-9af6-62f8de9e5940">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_992047f0-a4a8-4eb0-81f2-103ef31d505c">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b7f89e43-0d4b-47d6-949c-589316d9bb84">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_7aab008b-09c6-487c-bbc7-77eaad7ea4a0">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0c5af2f2-8c60-4b60-bbe4-437674d5bde9">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_2ec35bec-4669-4187-9fe1-cb2cea459db5">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54bf39f0-bd4d-4c00-8934-63a2a7d9f9ae">
                  <emphasis role="bold">Key​Images</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a15ca964-49e1-41b4-8977-6e1e03540dbc">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_80a5e4cc-7eb6-4242-90d5-612de519163a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_034efa0a-d5c9-4c0b-8650-6d7f83788d1c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e19ca2b4-e63c-4d25-b605-3953478fa913">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96e7460b-280a-4d53-a697-0d2e6b4e4f95">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa285fa1-fd31-424e-8eb7-b054ce7b590d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4a8c321-d847-4062-963b-c411307b0764">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ce5cbeee-17dd-4a74-8a05-1145f9e11763">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d76cb4a1-7594-406e-8227-991fdd9b09b2">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57276208-7e4b-44ab-b04a-cc37c34d25f9">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62962cc3-648f-42dd-b01e-00bfe3ece0c2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff9192e3-2724-4bbd-a645-4b8013f02e46">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35c8fcda-7c81-4834-a7df-98e7f5ab32b5">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_18056745-6782-4256-94a5-df9fb4bed10e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2fa2551-f153-477e-bc9d-7318c2344a1d">1.3.6.1.4.1.19376.​1.4.1.2.14</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb7d4efb-0a90-4e12-82d2-b7f34e546692">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de3eeac8-de35-4dae-99bb-584df353ca4d">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6468f6de-828e-4428-9ceb-69077c915ac9">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85208b44-edb1-41d6-bf35-ec73efeeb472">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64fbf164-3916-47e7-a724-2ae23728ee43">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66ed5bcb-14a7-462e-856b-14475e80aad3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a63fef1-adb9-4659-83c2-38466fc12518">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db6862a2-7feb-4083-af79-c7b4cb6e0d73">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d2609ce7-ca49-4a3b-9621-aa4eccc4ffc0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b150014-20ff-4278-8f3c-7a3753247e68">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_534ec88a-459c-4e82-ad47-dee99875d58c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8927609-3df4-4062-b7c2-3fb6a0b76669">
<link xl:href="http://loinc.org/55113-5/" xl:show="new">(55113-5, LOINC, "Key Images")</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c053fc49-a6d2-4eed-87f9-2cd897b68031">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7289b9a5-302b-400c-bce9-f7d7982a6251">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d46dab22-7278-4cdd-bfd3-6489f1ec9576">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1376b7b4-342d-4571-a91a-2e5873bcceae">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f6cc2d5c-9447-47d7-ae3c-f2337bead8a7">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc6123be-9d44-40c3-82f9-745f09e4ebd3">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47c888ed-1a1d-45d2-9683-07b210ec503a">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1e88bbb8-230d-430e-9cf5-7da609bbc080">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f257576b-1b97-42f0-ae86-472044b3c63c">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50fa6217-2769-4dd2-8337-e5f48181a29c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b2f3572-caea-450f-9575-51c0191bc0dd">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5ca31c9-b673-45b3-9fc8-07beb82fdb3f">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eeb546d5-8be6-41e9-af88-54db93d91629">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_809b793c-84e3-4acc-a9ea-17706855ada9">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f08dd974-cf0d-45ca-bd85-a395b5a5c1bb">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3da41f32-e593-4c75-985e-f107386382bc">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6549d68c-3524-40e1-9541-e2d8ea7e7559">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d7e4251-4fbc-4b71-85d2-c63c87c28d08">
                  <emphasis role="italic">SOPInstance[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b2e6b5e-ba96-425f-8c4a-b435c81f278a">
                  <emphasis role="italic">&gt;&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be7a0870-9059-4ffe-a72e-751270537633">
                  <emphasis role="italic">observation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cf39f402-d22f-4e58-88f7-d239de421a27">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_707b029a-f8d8-426e-9d07-4af982ffb250">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d85577c8-b5c9-400f-ab23-ed1898d6b0c4">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_161ab052-9da6-4be8-92c6-dbf48df297e8">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b53d22d0-a0c4-495e-ad94-8726029fbe1d">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b57c8b1-2cf2-4130-85d0-fc2d69c15954">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4ce86d60-b745-4a19-96f3-eeb9f4a5ec53">
                  <emphasis role="italic">Graphic[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6745e450-d84f-4163-8d02-b81773de5009">
                  <emphasis role="italic">&gt;&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_896cbb8f-93f3-47c8-a911-8214c8b1a29d">
                  <emphasis role="italic">observation​Media</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f270070-60b5-41c6-acf4-a9dcec7df5de">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3111c239-d323-4ee9-a691-4a105d5305ad">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2b73da8d-6c5c-417a-ab6f-efbd421c6be7">
<xref linkend="sect_10.3" xrefstyle="select: labelnumber title"/> 1.3.6.1.4.1.19376.​1.4.1.4.7</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.6.1" status="4" xml:id="sect_9.8.6.1">
          <title>Section/text</title>
          <para xml:id="para_0444b240-c475-4a7f-a9b9-d7bb67f537a3">The Key Images section text
            SHALL contain image references using linkHtml elements, where @href is a valid Web Access to DICOM Persistent Object (WADO) URL. See <xref linkend="sect_9.1.1.5" xrefstyle="select: label"/>. The text content of linkHtml should be either visible text of the hyperlink, or a descriptor or identifier of the image; it may be associated with a (limited resolution) copy of the image (see <xref linkend="sect_9.8.6.3" xrefstyle="select: label"/>).</para>
        </section>
        <section label="9.8.6.2" status="4" xml:id="sect_9.8.6.2">
          <title>SOP Instance Observation</title>
          <para xml:id="para_83ef5e7d-5149-4b27-97aa-b0fd01f7dd20">The Key Images section SHOULD include <xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> entries equivalent to the linkHtml image references.</para>
        </section>
        <section label="9.8.6.3" status="4" xml:id="sect_9.8.6.3">
          <title>observationMedia</title>
          <para xml:id="para_13e433d6-e08c-4e64-986a-5963f90cff18">The Key Images section MAY include observationMedia entries with in-line encoded copies of the referenced images, linked into the narrative block using the renderMultiMedia markup. See <xref linkend="sect_9.1.1.3" xrefstyle="select: label"/>. These in-line encoded images may have limited resolution and lossy compression as appropriate for inclusion in a report.</para>
        </section>
        <example label="9.8.6-1" xml:id="example_9.8.6-1">
          <title>Key Images section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">>
	<templateId root="1.3.6.1.4.1.19376.1.4.1.2.14" />
	<id root="1.2.840.10213.2.62.704478559484.11428372623" />
	<code code="55113-5"
		codeSystem="2.16.840.1.113883.6.1"
		codeSystemName="LOINC"
		displayName="Key Images"/>
	<title>Key Images</title>
	<text>Maximum extent of tumor is shown in
		<linkHtml href="http://www.example.org/wado?requestType=WADO&amp;...">image 1</linkHtml>
		<renderMultiMedia referencedObject="refimag1"/>
	</text>
	<entry>
		<!-- SOP Instance reference -->
		<observation classCode=DGIMG moodCode=EVN ID="SOP1-2"/>
	</entry>
	<entry>
		<!-- inline rendered image -->
		<observationMedia ID="refimag1">
			<value representation=B64 mediaType="image/jpeg">
				Bgd3fsET4g...
			</value>
		</observationMedia>
	</entry>
</section>]]></programlisting>
        </example>
      </section>
      <section label="9.8.7" status="3" xml:id="sect_9.8.7">
        <title>DICOM Object Catalog</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3bed04d-1e79-4ab1-98d0-723aef9de34d">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f76693e-3178-45a2-9718-9334ac7c6343">2.16.840.1.113883.​10.20.6.1.1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35fd5526-07bc-48b0-9fee-7ae621ff91aa">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8b6a462-38f0-42fc-a5f5-c1b63475dac3">DICOM Object Catalog Section</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d98e002-4e99-494e-bda8-64bb4e4c89a1">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7491dcd-db5e-4a2a-8a07-f3cbf230ff02">2012-07</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_93b531ba-c4db-456c-becb-77fd3fb4d90c">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_960f8bc6-fe03-4913-9c48-ea23b95563e5">CCDA-1.1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_927ab6bd-f3b8-4e5d-a0b8-115deed9e1e0">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3a12bdd-b6a7-4418-bf62-29f70538fabb">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7757333e-000c-46b3-a6c3-0e2bd98e65ac">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_683bd695-aded-4572-a87f-c60411624c3f">DICOM Object Catalog lists all referenced objects and their parent Series and Studies, plus other DICOM attributes required for retrieving the objects. The DICOM Object Catalog section is not intended for viewing and may contain empty section text.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90b501a4-0e18-413b-8d45-7c3398126f8d">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eeefc770-2a33-46d3-877c-27119ec2dc73">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6b80bb3-e270-4201-b441-b48253df9086">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5442a3b6-ee3e-4d5e-afe0-b9376e1028cc">Included in <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a26e9212-2e09-48d6-b44d-b02577bf8dd7">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58c11574-bdfa-4d57-9b22-0170e50b4727">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9f4ee40-f5e4-409c-bf93-6c8742fbcf5d">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_468515e1-8390-47bf-9892-52021b8a2703">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_df0272e1-34d2-4bed-b0dd-a590f228bd7d">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b2b185e-53e6-441f-9981-91e233cc08a4">From Consolidated CDA r1.1</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_a926c295-f4fd-49d0-bb7d-35890698d692"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ff3e33a5-afee-4343-b3f3-5867b59a7815">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_80feaaeb-cdaa-452f-b881-938538cb6cdd">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b1388f7c-06b3-446c-9b06-a9a253523e8a">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_3b460b50-b1ce-42f3-979d-3d807b636b8d">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_db129c64-fa46-441a-a7da-2cc1be316a82">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_77d07187-03ef-4cee-b612-a98e90d80fb6">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ef014054-2f55-4050-ac3a-3ecd51118519">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_872b5941-69a6-425c-a7ec-d20a89452e69">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ff9478ac-51f9-48a1-a713-a434387fdf1c">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f331e5c5-0117-418f-9780-3ac466598a99">
                  <emphasis role="bold">DICOMCatalog</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0edbf296-d267-4603-898b-86e59aaaf47a">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ab0b5a2-9dcc-4303-a623-21660c7a045f">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd332156-cad5-489a-930a-de52192e43e5">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08602c08-ff54-4d15-9ddb-e0c59d96e697">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bd1c8991-3e0b-4bf3-9807-2f8a284d1e2e">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_39333e0d-10c9-49d7-b12d-e2fef572e64a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50368aa9-d560-4f1f-bc34-f3d70fe81b95">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9b0fe498-69c8-443e-bfbb-76ead33abe74">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eeafd743-5c94-406a-b0b2-94a0f730b14d">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_412b23a7-fe19-41e8-af7a-3cbcb73795ca">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7bec2336-4b55-4b3e-82b0-25d43c7f4433">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41bd11db-7861-472d-bd5f-52675939371f">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_869feea2-d74f-46b0-a7f8-fedddd73c90e">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6c7b301d-4ee5-4b79-a5f5-d9b9038be268">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8cb7e550-15e3-4aaa-86a0-170af688f416">2.16.840.1.113883.​10.20.6.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83118a00-745e-42e9-99b1-f38b05264761">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74dd5756-f077-4c28-a9a6-007561b5db5b">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_581b3458-6b3c-472b-b33b-d6f723d2c111">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed4cc23e-dff3-4c01-8c12-1e3641d2bd6e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2be6c848-8f1c-4e08-aae5-7c59b93a7a68">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_95400add-433c-41d1-8fb2-ac32a3217b55">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_168dc8ed-ceff-4c29-be28-a1847ff9ba0f">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d78a5610-3902-4d09-991f-574bb8a2eb14">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e512c8dd-74f5-4e30-ba90-b7b4a1f6e4db">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea6449bd-7964-4498-82be-8afa15932763">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b840887-88d1-407b-b412-f7999de1bb41">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_755f3b13-402d-4add-9e6e-a7a8b48200fd">
<olink targetdoc="PS3.16" targetptr="DCM_121181">(121181, DCM, "Dicom Object Catalog")</olink>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b8aaa94-3052-4f61-b023-85973ce2d3eb">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5efde344-58bf-4fa5-b425-aff63c2aa9cd">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a239e715-9b85-4ea6-8851-5a937d2e0935">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_469aef94-105b-4f12-90a1-a93273e7348c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35a588f8-e241-414e-b3d4-dd94641633ff">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_27f904f1-d696-49ce-a820-db3103a0f904">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a48dda2b-58e7-47b4-a8a2-4b3a35c5fa43">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c8e85f0e-8465-4912-ad5e-c3a5b8b81e83">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dbcdc607-4da4-4e58-99f8-3ac773767286">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a87ed339-4a4e-45e9-a71e-25f02d67cb10">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e48905c3-ad4d-4f3a-b90a-8726963cc85b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5916519f-27db-40e2-a5e6-89652c904bd4">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7fda3ad-1b61-47c1-a817-a1d088dfc32b">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_139cc0ae-f54c-4e56-b525-fa1e380b2083">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c561b77-6ae1-4381-906a-6b2b0b1725cf">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8b4ff2b-54fb-4815-bcd3-c3f4ad47ad3f">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d8a023dd-204f-41ee-ab11-bc8e031827d9">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d0526c69-768a-4d9c-b667-8bb0f2d6ed45">
                  <emphasis role="italic">Study[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e11a6b6c-2b05-43f1-b25b-55f566f49ff2">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6c8eb763-174c-4903-8cd3-3600b6e86bea">
                  <emphasis role="italic">act</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_97dc2f7f-8565-45b8-b758-46db97808adc">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3cdd59d1-4adf-4672-a6a9-ccf4e4ac1bd0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08144444-a2dc-43f4-a9ec-03f5cbc5116d">
<xref linkend="sect_10.6" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.6</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <example label="9.8.7-1" xml:id="example_9.8.7-1">
          <title>DICOM object catalog section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
	<templateId root="2.16.840.1.113883.10.20.6.1.1"/>
	<id root="1.2.840.10213.2.62.70447834679.11429737"/>
	<code code="121181"
		codeSystem="1.2.840.10008.2.16.4"
		codeSystemName="DCM"
		displayName="DICOM Object Catalog"/>
	<entry>
        <!-- **** Study Act **** -->
        <act classCode="ACT" moodCode="EVN">
			<templateId root="2.16.840.1.113883.10.20.6.2.6"/>
			<id root="1.2.840.113619.2.62.994044785528.114289542805"/>
			<code code="113014" codeSystem="1.2.840.10008.2.16.4"
				codeSystemName="DCM" displayName="Study"/>
			<!-- **** Series Act ****-->
			<entryRelationship typeCode="COMP">
				<act classCode="ACT" moodCode="EVN">
					<id root="1.2.840.113619.2.62.994044785528.20060823223142485051"/>
					<code code="113015" codeSystem="1.2.840.10008.2.16.4"
						codeSystemName="DCM" displayName="Series">
						...
					</code>
					<!-- **** SOP Instance UID *** -->
					<!-- 2 References -->
					<entryRelationship typeCode="COMP">
						<observation classCode="DGIMG" moodCode="EVN">
							<templateId root="2.16.840.1.113883.10.20.6.2.8"/>
							...
						</observation>
					</entryRelationship>
					<entryRelationship typeCode="COMP">
						<observation classCode="DGIMG" moodCode="EVN">
							<templateId root="2.16.840.1.113883.10.20.6.2.8"/>
							...
						</observation>
					</entryRelationship>
				</act>
			</entryRelationship>
        </act>
	</entry>]]></programlisting>
      </example>
      </section>
      <section label="9.8.8" status="3" xml:id="sect_9.8.8">
        <title>Fetus Findings</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e6d6d305-04e7-4228-bc30-0ec2eef537b8">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d25e3b59-9b9c-4cca-b111-47044d75b04e">1.2.840.10008.9.9</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_572f773b-1b15-48aa-b29d-748395c5c492">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5082607-1bfc-48b2-aad0-a0910d0e9d5d">Fetus Findings</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_93d7669e-8ade-46fe-a189-5d1557a0f6bf">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bfabb4d0-0752-4026-9705-d68dad660fd9">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c648fd3-61cc-4d93-ad25-8a70be3ef7f0">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8d73c72-78d9-4d3f-94e6-7b1c79374060">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7727b47a-cae5-4293-9f42-3467f594cf86">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_68cb4eec-bb16-4d8a-a4e5-0de0fa64ebb1">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7be86d2e-dd69-4adb-b6f8-0ef6793c9f36">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_94d389b3-36a4-4ae0-b819-e7beab81c45d">Records observations related to a fetus confirmed or discovered during an imaging procedure.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3a81322-71ad-4066-b0bb-6e1de7b3fa50">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d24d10a-aa33-4650-916c-c260f9775404">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e586221c-56c1-4982-a25f-b6bde095985f">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f7898bd-bfdb-4a73-97e1-1f49d09cc618">Included in <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15fab523-7e22-4aa8-adcd-837fb7fd1f2b">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e0fcd476-6179-4ad3-a59c-c270aca67db1">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b36b7ca-b227-47ed-a997-6f4445b12879">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4573c78b-4a47-4b90-a7cd-0bbc7fced471">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_744f44be-d390-4d06-b61d-0f3eeaf288ed">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5879340-a74a-4e48-80ac-2bf1cfe0d794">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_534dd434-5b7c-4ecd-8fa6-1e359c7be6c6"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_42e571d0-6870-4bfd-b9e0-d1ee1e81e56b">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_2f2811ff-ca47-4c8c-b5a0-4f4270d682d2">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_346da168-8737-4237-b106-6673c545173d">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_68b93ae2-c0d3-4b67-a1c9-fb2cc87d7a22">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d178bb89-1edd-4de0-ab52-216e840d6272">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b59355c9-5496-4c87-8fbd-8eefbe5449b5">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_efd01527-8d94-477c-8dc0-d81a525f75dc">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_80128915-f5ed-4cea-97b6-e72f5ade252a">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_dbe291c1-e485-44b7-a54c-c39e74ec8843">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2fd86cfc-909a-40d4-86ee-8260d1409dd4">
                  <emphasis role="bold">Fetus​Findings[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76245251-d7de-4b69-abf7-8bd93f1e9da8">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_033b3e26-2554-4ccf-bb50-1214b69b9d2e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d608443c-17ce-4431-8570-a287c64d958c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d5434dda-c97e-43d1-aea2-9acf5413bc96">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac3b5a26-c275-4882-9480-7b26be68881c">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a85590ec-00bb-4690-97f0-f084de6a897c">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_94c3728e-83f0-4678-a12f-f876dc952b3e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0175bec9-3f8a-4804-84b2-14c384dd6c39">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b97b1b24-c292-4ad4-ae3a-2f2cb1295489">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb32f14f-3b2f-4dcc-8e7d-5fada79cd2d9">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7f2216bb-5076-48f8-b0e9-4948149d76fa">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_624861e7-e9fd-4f2b-8cb0-f2746102bd5f">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d15cd94e-125b-4f44-8f1b-f56d062e2763">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2494f696-3847-484b-a06b-ea1c89ff2150">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70cd769c-bf39-4462-92dd-755c44ec4a8e">1.2.840.10008.9.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a718c0b-771d-476a-92c1-a7158fa4f6c0">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_68c0840a-db95-4117-a322-925114d6b0b8">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58ecb9eb-2174-4943-a728-7437049fccdf">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c80c8d6f-65a5-499f-83cf-3f31237a393f">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ce3eab6e-de5e-49ff-90a3-71d1fb4a11c7">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3861e69-a80d-4c9a-ac58-ccd3a9ed72ac">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4694cb0-812f-4f19-9269-bc1bf08b4172">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d7b5e60a-ec5b-4a1b-91e7-55252eb3ed3e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41d3b17d-dc59-4b20-82a6-3140a8e27073">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30651b32-9bea-4e2f-b118-bb0f1524e889">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bd5580e9-bebd-4c1b-8a7c-30d173ee961e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61951a81-5511-474a-8ce7-34cd8da6c7fa">
<link xl:href="http://loinc.org/76514-9/" xl:show="new">(76514-9, LOINC, "Fetal Study observation")</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f477b11f-d3a8-42a7-95bc-5670f26d9302">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f0c5eee-c943-4524-a2e4-9a5a8208a210">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1425bdd-a59c-432e-bad1-e8f7eaedc44b">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e16753e1-0e64-4396-a03b-f1e12912d51d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_caea5332-f096-48e6-b96f-b2f3e1e82737">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e840c8b-1ffb-4ed1-8a80-86b4f1195ec5">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfc5d441-6a38-44e8-9df1-c5987935f411">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5a81882-5e80-4f57-beb1-bd602df3a33e">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_11fece51-3b5a-4f30-9e7f-f8211cb2610f">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38bcf1f2-e2f4-41bf-89d9-f7376602c83e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d8f5e70-9dc1-4331-9346-f762896d6d4f">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_34304729-2011-416a-8791-a18feee0a395">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75ea210f-ae04-43a6-9327-be127e1ec4d9">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40f5ac0b-c74e-475f-9149-5338aaecd201">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2762dc06-3fe2-4e88-953c-e22fdc0b2054">subject</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b54fd85-c0ab-4a2f-941e-1a0eae33876d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fbdcfedf-d972-4b7b-91db-f865952e10cc">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cf9bde32-720e-40fb-8f3c-0092669ef4ad">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61bf7c4c-4a85-44b6-8b91-03148e557fcf">related​Subject</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57122483-0253-4e46-8185-9dd23afb164d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c9f32d25-bc59-4064-a257-dd365b9c12a0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c6a4e3c-a8f0-41a0-843f-23d5f7a9b4f9">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f1aa743-2ef3-405b-8431-15e479e53bcc">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_11b193d9-29e4-4318-9caa-90beded746d3">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d18c836e-dfda-4404-bcba-4d6f0019b5d2">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a5fcbd9-65f0-4bed-be4b-b3a2e53587fa">CE</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_786b7591-5720-4d1b-abe6-31c8e9bce6cb">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f32398e3-ff87-4c0f-9d43-6a7b857fbe13">
<olink targetdoc="PS3.16" targetptr="DCM_121026">(121026, DCM, "Fetus")</olink>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81406563-8291-4364-b6c2-5abaf4154a3a">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78244330-8542-41b5-bb84-d604d2b220a8">subject</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3febe85-a54b-4d5f-a641-aecead573d7f">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a541e71-a696-49b0-9786-b06c934099ee">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca02b1c5-1b1e-4dca-90a7-f28df8525ee6">Fetus​ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e7cc6fb-2006-4206-a8c7-5568fa8c8206">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12eb6368-5aa0-4d57-90b8-e41f07a38a17">name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_abe79836-21ee-4337-a741-6768183acf9a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4520fccb-9195-48a8-8af4-002851f06853">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad3665a9-7725-48ca-a33d-8437ee02e125">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3993eb81-c9c9-4c22-8eba-76b8608e466f">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f63ca79a-8e38-4135-a8aa-3dd8c38eb68e">component</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab58758e-60ee-4618-a087-72f82acd7de5">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f0092a50-7c1d-4b30-b349-8c3acaee7be8">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b65399ac-9c81-4d3f-9f63-2bcf821c4176">
                  <emphasis role="italic">Subsection[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59349b68-1c0d-49a5-ad4f-ac7fadcd95c9">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7afecdb5-dad5-4484-9951-ec218d877222">
                  <emphasis role="italic">section</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5dff4ee1-7abf-4b16-9224-8eadfecbc218">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4d8fe935-13a3-463c-a9a9-c6fcd6987b45">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a92691c7-ec88-4fd2-9aaf-e5b4412d5833">
<xref linkend="sect_9.8.9" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.10</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c636b90f-fb99-42c7-848e-26788e59509c">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_895a6ad6-41a7-42bb-955d-8fbf9df61fae">0.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3a7f45d-c2c2-4f12-821a-cac13ac2bf34">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_48907c34-f34e-47ac-8e30-9036904c1e52">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <para xml:id="para_14973c93-eca6-4df0-ab18-ad44e5667a8d">For reports on mothers and their fetus(es), information on a mother is mapped to recordTarget/PatientRole/Patient in the CDA header. Information on the fetus is mapped to subject/relatedSubject/SubjectPerson at the CDA section level. Both context information on the mother and fetus must be included in the document if observations on fetus(es) are contained in the document.</para>
        <section label="9.8.8.1" status="4" xml:id="sect_9.8.8.1">
          <title>name - FetusID</title>
          <para xml:id="para_03e81e1d-6944-45d9-83e1-1fc0a827f7fd">The subject/relatedSubject/subject/name element is used to store the fetus ID, typically a pseudonym such as "fetus A". This shall be present even if only one fetus is identified in the document.</para>
        </section>
        <example label="9.8.8-1" xml:id="example_9.8.8-1">
          <title>Fetus Findings section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.22.2.27" />
    <id root="1.2.840.10213.2.62.70447834679.11429737"/>
    <code code="76514-9" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Fetal Study observation" />
    <title>Fetus #1</title>
    <text>Estimated gestational age of 27 weeks... </text>
    <relatedSubject>
        <code code="121026" codeSystem="1.2.840.10008.2.16.4" displayName="Fetus"/>
        <subject>
            <name>Fetus 1</name>
        </subject>
    </relatedSubject>
</section>
]]></programlisting>
      </example>
      </section>
      <section label="9.8.9" status="3" xml:id="sect_9.8.9">
        <title>Labeled Subsection</title>
        
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b11b83e-537e-490a-ac89-1960f88e45c9">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_93f63d16-9b79-4576-9023-ecf97820b1f2">1.2.840.10008.9.10</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_865d9bdf-595b-423a-9e42-26d4a694440c">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e41a2ffa-2720-4d15-9096-c32986314399">Labeled Subsection</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c45d4980-4c3e-4f0c-8cc4-7fa29a4384ad">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a68ae439-4a3c-4ac6-8ddb-232c942e5af9">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d53191f-a62a-4e6c-b108-b0d0a3fb5d15">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cde1193b-2918-4200-b1c1-8776ddd927dc">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c9a45aae-56d1-463a-a81b-631dba481781">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b9f58ff-08e0-43fa-83d2-00c9771c4f34">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1ab52dc5-898e-4f50-848a-1bd636d2588f">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6714848f-142a-46d0-bf15-36d8bc8d5c79">Narrative or coded subsection that allows organization of content for a labeled topic (a particular organ or anatomic feature, a lesion, a tumor, etc.). The section.code shall be absent, but the section.title shall be present.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8f562ed-276a-4abf-a2f6-936ebdc5298f">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43f34201-7844-48b5-9901-8a4b82763066">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36c2e7bf-5c1a-41d9-969b-647fc61e1ad6">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a2864ed3-1d24-4198-84d6-44cccda3cda5">Included in <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d36c565d-ada1-4d6f-802c-4c5a74db3f7b">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc61a742-489f-473b-9321-5c7145054cbd">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_325dccb6-3099-472f-b3b9-cd5ba219212d">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_760120ca-e3fa-4824-bfeb-ffa84c48b836">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75914fe4-c196-4c56-ad72-613a5a03fd04">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c02f2f19-86a7-4da2-9a50-1950f728328e">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_01b82f1e-2d71-43f4-af23-606981f7cfce"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_7ecbe53d-b2d6-4ac3-93ba-96f55986e238">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_82d68437-63ae-4de4-ba99-583214a08c6f">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a8072a39-8bf3-41c7-81ad-2d5cfc13a464">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_8d96addc-ba49-4fb7-9956-dd58c8f44705">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_c14d8340-e88e-4911-99c3-971e87a95500">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_73848946-4863-41c2-bcf9-f18f2b2fd6e5">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_4e61a39e-f828-43c9-b916-bd0d45f264a2">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d2540d34-9624-466b-b95f-7e75353c69e8">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_0427be81-8061-44c4-98f9-34e5a1f7186a">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e89202f-5a91-473c-8b9f-1c44894cafa0">
                  <emphasis role="bold">Subsection[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_95fa6745-1efe-4d78-94ff-95bd17f248f2">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6bdb84a8-8b7a-4439-ae95-d2a36c19b3ef">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e348f0d0-583a-47e8-899e-9b578f332e6a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5064e554-0cd6-445e-bd32-e58b9b7888eb">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c2c388a9-1ada-4cbe-9fce-0c1e1975bd46">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_49c09c0e-6b41-4d7c-9f02-8f12c6e38c5d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1bc9c0f6-0125-4a5e-ad71-2cc178b093c9">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d8129132-050d-4e06-8303-5e1ba24928ca">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67b58e02-af32-421b-815d-cef3ac045a87">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_803da9be-1d30-4671-9fd9-e308e9e8086a">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9cd5baff-672a-4281-97f7-ebd10e406a5b">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bfca714e-efbc-49cc-a8bf-89086e8beeae">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b97e697-52a3-4c07-82a6-41f70e307d5d">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a26a6df-624d-4779-bb0c-f6ce4fef6691">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9dbbb7c-635b-44fd-96d7-b171b52a453a">1.2.840.10008.9.10</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8886fb56-26ac-4464-866f-febd2557095d">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d91d5dc5-ad5a-4c34-8ad4-313248b3e119">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1113597-e928-4464-9d5a-1806e97783da">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b391b26-5e38-4fe4-acda-d908b6d1ae2a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0fa4a92f-7d61-4f8c-9df0-13277ec91390">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7314a6c1-a856-4677-a6e3-96c2b620ce45">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b99c4ed4-984e-455f-9169-ffaa0bc9fdf1">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1383a05-c462-47cd-b8b0-8fc50f3aad3c">0..0</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e8b58a4-bf07-4743-bcd2-4923e3fdf2ca">SHALL NOT</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d3f2abd-15d0-4e47-8439-096484fa78e0">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50a463ff-6536-4466-ae9d-666d025f90bf">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f1b0106e-03b3-4b7d-9df1-89ec1db34101">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1f6378a-8e0e-4cf4-ac45-f849c8b27f64">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdd38071-9d8e-437d-8f88-d224cde57d31">SHALL</para>
                <para xml:id="para_f7aa5545-b2d8-4aac-afaa-facb08ca07f1">no​Null</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f30134a9-729f-4c54-b246-08fc024d3c6b">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf78bc26-b8d0-4a39-acd8-d279b9661038">[See <xref linkend="sect_9.8.9.1" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc0799f2-f617-468b-bc66-c9adfd06bf7a">
                  <emphasis role="italic">Text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d685e24-6dce-4dd6-b29d-c970648c5e74">
                  <emphasis role="italic">&gt;</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9538b456-3506-4aff-9776-fc7325d3674a">
                  <emphasis role="italic">text</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f6c9a8d9-c5c6-488f-aa88-7981949385c4">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_848ca2d4-269a-4bb9-a3d2-c80aab387fe9">COND</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1ccbe61-5841-4c9b-aa0b-1e05c60ab7a1">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7033b513-0379-4f99-9706-98cd6aa4002e">
<xref linkend="sect_9.1.1" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.19</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96e4583e-b39d-47d8-b173-db8f74c4ba74">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e5cbf8a-1e0a-48a6-a31d-5994a9f14312">component</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84d00a01-1a7e-4046-88d1-c8da518cca24">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c24bf5d1-c8e0-4229-af7c-8d9528e1f46a">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5a41a09-6115-4caf-8156-29715ad6a56c">
                  <emphasis role="italic">Subsection[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fcb10dea-f2b9-493f-93da-199219ff1e43">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12053a64-70bc-4228-a89a-aab6586e5faa">
                  <emphasis role="italic">section</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9984e65a-7890-4852-a8e0-7d1621d007e0">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2eb54bea-472c-4520-a0e3-707441e298ef">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_364dd462-442c-4ab7-ab6c-8cafac19be87">
<xref linkend="sect_9.8.9" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.10</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e976ba8-9222-4ae3-97d1-f47cde8523d1">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af1a4d8f-c6aa-413c-a750-9f2161498a73">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bef766c0-ebfa-41bb-900f-2b96af912fbd">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8a1e9fe-446d-4c0e-9118-3b1e468eafef">
<xref linkend="sect_9.1.2" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.23</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.9.1" status="4" xml:id="sect_9.8.9.1">
          <title>title</title>
          <para xml:id="para_db72c77d-c9cc-4216-8577-f55406c8a393">The title element is used to identify the topic (specific organ or anatomic feature, abnormality, lesion, etc.) as the subject of the sub-section findings in the human readable document. As there is no section.code, this is the required mechanism to represent the section purpose as free text.</para>
        </section>
        <section label="9.8.9.2" status="4" xml:id="sect_9.8.9.2">
          <title>component/section Labeled Subsection</title>
          <para xml:id="para_9064db31-e28e-41fd-98fb-485982dc4920">This template invokes itself recursively to allow arbitrarily deep nested subsections.</para>
          <example label="9.8.9.2-1" xml:id="example_9.8.9.2-1">
            <title>Labeled sub-section example</title>
            <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.10" />
    <id root="1.2.840.10213.2.62.7044794679.114296787"/>
    <title>Liver</title>
    <text>No evidence of cirrhosis, nodular regeneration, or ... </text>
</section>]]></programlisting>
      </example>
        </section>
      </section>
      <section label="9.8.10" status="3" xml:id="sect_9.8.10">
        <title>Communication of Actionable Findings</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0f2246ed-5cad-437f-b2f4-bf9899f17196">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2fa91a7f-b7c8-4ff9-9a56-b114879bacec">1.2.840.10008.9.11</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81c6f166-67ec-4f0d-bdb4-68ef7c9d3450">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c60c551-2163-4635-aa1c-7205f422382b">Communication of Actionable Findings</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c385cfd-9644-48c9-ab50-f6caf6d98b67">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b133b026-e851-41f2-9be7-f132d84928e6">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de16a5f2-353b-4771-b929-fd5d1556a6ec">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f8ea947-d166-4363-bedd-f25580ce55cc">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b0c1e58-4b65-4d53-9346-2b9db60c0d84">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ee7bfac5-7cc7-42b3-9c5b-5a430787c870">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5c26f3ce-0640-4180-af96-b8db36252da5">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_afe0aa88-2264-4dea-86f4-d791b6517441">A section that documents the notification of an actionable finding to a provider or other person responsible for patient care. The documentation in narrative text, and optionally in a coded entry, includes by whom, to whom, and at what date/time.</para>
                <para xml:id="para_7a3d86e2-ae48-4c28-9f4b-b3607fa39eff">Specific findings, including actionable (aka. critical) findings documented in text or as coded entries, are typically found in the <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/>. The actionable findings may be duplicated in the <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> in either text or as coded entries. The actionable findings may be new (critical) or a change to a previous report/diagnosis (discrepant).</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_23ecd122-f727-4bdf-94ee-5881dbe1a445">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e83d72ae-1070-4245-bc9d-2f7a6f728674">CDA Section and Entry Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b247a8f0-81b1-4b62-a5d1-f8862b15495a">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9241d4c4-1682-4afe-a8f4-937118fef32f">Included in <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> and <xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_463ed89e-df7f-41aa-83a2-b0f47c1ee73e">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_772c5f9f-377e-462c-8e95-3a96dd1eab6a">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e600205-2e18-495e-b09a-af133344ee72">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_07a3dd0c-2b6b-45b5-b455-34a364ffd1c1">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92849a86-1b92-41b7-8e03-af79be61f35c">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2c0896c-e24b-452d-acec-c116a46cf71a">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_2e4ab34d-eef9-4e81-aadb-6a9549dd5230"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_245f05aa-2d5a-45a9-8634-692821d57d56">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_fd3e5caf-fd89-4b7f-8c57-0292b0da108f">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_437d9cc0-5fbe-4728-a183-05f441ce0a74">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_58d594e6-aa6b-4ff5-9a63-569083aadf1e">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_776e1bda-d00a-4826-b2d2-b52a8a93bf78">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b9ea4bce-4fdb-405d-ab3e-3dc9b4324e2a">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9c4cb498-7066-4dd2-8e01-98d0060dcd3c">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_5fd9308b-669a-499a-99c7-5ee734388512">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_7f171b3d-e8c6-4d2b-a644-655c105e7311">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54d9ac3a-1fa5-4962-9e8d-adac956359d7">
                  <emphasis role="bold">Actionable​Findings</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d6c0b158-b6c5-4898-8f58-a3bb78956270">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c2bdbd4-b0c3-4a2e-a954-5f9437143ac7">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5fe42dce-fd0c-42dc-9019-f1cb4c4c9e4a">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd0588f1-e26a-4c65-a486-e1f4468615e3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f5a0ccc-000e-4e4f-a046-1fb000e89198">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_948d87eb-922c-4ce7-b0d0-4220e8ebf6d1">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff1a5be3-6b76-4f12-895a-a5a83657dbd6">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf996b12-ebb0-4640-83bc-33147bdd1c83">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfdbddd4-6237-4979-9abf-d77eece61a5d">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25df6d6d-985f-40e0-9788-8faa9bab19a5">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c696e449-adb8-419f-b279-94438a431910">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1dde24fe-883d-412f-838b-9c0c3e6bda5c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_63369b67-6b95-4563-bc23-bff7db43e018">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2accd504-626f-4967-b03b-fea1c6fc4a13">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_95520300-35d2-4650-b5d8-93b4b9d51650">1.2.840.10008.9.11</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a9124e1-49fc-4978-aab5-b1c9eb628f33">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_312d58c8-9267-4948-8046-54b0b3f36edd">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0203888-1cfb-48dc-bb10-829685e81173">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8de81176-31bd-497f-94e2-63bbda47c1ab">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_999794a3-fecc-4a08-ae3c-9fe687e3cde7">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa28fc68-b76b-4c29-ad62-d4d1a9e5eb73">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_692bcb87-8c16-426a-ab33-b655910afe04">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8fe2b39c-c611-423a-b8c8-aa55d76aaf28">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed58f428-b292-41bd-b132-d4a8dc3b80c2">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_712129af-b930-419a-aa29-1e4835eb191a">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b2edc17-75f0-4112-9997-3f5005590c56">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30ca27ed-b518-4cae-8940-781b3956816c">
<link xl:href="http://loinc.org/73568-8/" xl:show="new">(73568-8, LOINC, "Communication of Critical Results")</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c7871d0-dc7a-4574-a01b-a9414f9d1f03">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8449882-9d2c-43cf-9486-7dcf6ad6a459">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a29f2c98-89df-45f9-b063-2918e41d2843">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28d79036-039c-412e-8373-b6d0b4e10cb2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e48e0dfe-e518-4980-b90e-8850112b6601">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81645c82-a38b-40a5-a9c0-e2e2735b1572">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51287e66-d05c-4ecf-96f2-1e36bec100b7">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36e7717d-1e42-4f76-944c-701d95cbcb14">text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f56830ef-3659-4f15-970f-1fe1ffdb3cc3">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_156d8b5b-f15d-4157-87df-14bd10fa4f2e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f952335-5389-4ce8-8f6b-f12ab3ece968">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab5cd306-9311-42d1-8557-fe3f34daf2d7">
                  <emphasis role="bold">Content[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51355bd6-94a2-43f8-ae71-402ddcaa40a2">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61effd34-f546-433a-a361-9a3738166e69">content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44d6eebb-445c-4e74-b83e-bb12d25f7c0c">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75261b97-e0a7-454a-9b94-e4d0245d25b8">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e184befc-1fe3-455d-b794-3e05bfa5c74d">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a62b2dd-23f7-4980-a210-4ea6c7b2231f">[See <xref linkend="sect_9.8.10.1" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2b186000-0ace-40f6-aeb9-be24f2b3bbc2">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e851f8d5-86b5-4ab6-8951-6b06e4371ab7">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a4e3aa7-6155-4247-a213-e8f64b939b99">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1b8f0115-2a1a-4211-8813-4e0ac17c86cd">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_016c3ab1-7259-45eb-9307-461207e8ab39">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d91d385-5df8-48e3-9759-d3f29e43591b">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d347bb2-34cf-4068-8547-46ff6fcd215c">Finding​Ref</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45651e79-38c4-4a80-a39b-50800db54996">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b53fbbb9-2c3d-47ad-b3d7-84c0b840b4f5">link​Html</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a7c18ae-ee07-46b4-8217-f3a06c5aba69">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50480bf9-7080-44d0-aa3a-c61e98e02d2a">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3ec7b7d-b856-4403-b265-971ba1e8c593">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12a86d6a-2676-41c5-ba77-ed0e65efb4ac">Finding​URI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ceaedde-9329-4efc-be3e-19f246e2b109">&gt;&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16c5e71d-4d2e-4d79-be99-fff7e96cc04c">@href</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a72fed2-cec2-479f-9575-4e7e85d71ed6">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_26db588d-49e1-49d6-a5ad-0f779e6f1f97">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9404b1ed-cdf3-46ab-bbea-f9add1a5fd23">URL (XML IDREF)</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ee9bde18-f33e-497f-813d-d43ee94c719d">
                  <emphasis role="italic">#findingRef</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bb96ac87-782d-4ac2-b9a1-6f1c06e5db0c">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30c72b4b-cc8e-48fe-8678-90b0cf1d81ab">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59ec2245-493b-4e56-8b30-2d0a70344da0">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b17bd798-4b74-460d-b7db-ffbeb30a5908">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9e458978-eb8a-4107-ae02-afb908c216b1">
                  <emphasis role="bold">Communication[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a592ca0a-4545-422d-bf00-32a1b4726678">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c888ecb2-d4c1-45e7-9180-5570ee0b93d4">act</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef327489-cec0-4dc7-a8d6-a82c6c3593cf">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_22290efc-6124-4956-a779-4568c2be6fd6">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_79e8d014-cd14-4be1-97fa-aacfec0d6cb5">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96d2b7e8-f3f4-4c9e-b96f-87a52b5a54c9">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f86040e6-e6c7-45e4-8c5f-cce7489f03fe">@classCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25bde086-b54d-4f20-8292-17f2e6712cd8">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b00740d-24ee-46bf-be6a-742498edab16">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e67d1a24-7dfd-4097-8fda-0d3a46d51672">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4f59942-eb03-4d59-a9ec-9f95671f3741">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62a6cec6-d2d9-4d00-bb2b-b844ecb56b92">ACT</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_348eac9c-c585-43ba-bbee-5963e8d21d15">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f36ce77-59ba-4efd-b8b2-87e0deb5ebc8">@moodCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9a46db5-5d63-4371-9175-18ef13352c83">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43d5b0d9-d87a-4c35-86a9-fcd06fd3e3b3">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4b8d303-892c-4afd-b30b-9db2382d5604">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71ae4555-8266-4f67-93f5-0ca0a5f8f393">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_501ad4a4-1c15-4495-ba15-4bcd6b0d2799">EVN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0438b08f-dff5-4673-903d-1463d70a5954">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81a31a11-4c4f-4b82-92ca-4f421c56e201">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6357040f-e43e-4fa4-aed7-48c7da18c89c">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b5464447-427b-4b70-aa35-bca564a9ebc2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7b4b2bc-5267-4b9c-81c4-47fa18415e78">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b12a976e-06bc-467c-8626-f77a3e846edc">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfb1033a-adca-45e7-a957-6cf351478574">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73e72ee3-ea5e-4ebd-b4ea-1433f6d6c17a">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7defea19-b06f-4e99-98d9-4fd0a4124aee">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db2d8d7f-63a0-4770-a33a-5dc6ad150c36">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd5fe70c-cfd2-46fc-8b06-7be9a4c31508">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f1c884e3-38a6-48e1-abd2-3ffde1375ff0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c0fd793-05a8-49bf-9b7c-6ce5a341c7a6">
<olink targetdoc="PS3.16" targetptr="DCM_121291">(121291, DCM, "Results communicated")</olink>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54054bb7-28c3-4644-9950-dcae9faff05d">Comm​Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1790694c-49b6-4142-a3d8-a5fb975f7730">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77fd633f-acc2-45fe-9256-438406d15d63">effective​Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b1c90d0-b152-4e3d-a538-6797f5ce6582">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_213f42fd-392b-463c-af2c-b9861ddce732">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70d38595-0369-4805-8853-abba1bc2cd6c">TS</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_477341e8-6705-4347-9d7e-c638aa51b995">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3cf2b80a-e667-40e0-80eb-ce30aaab6468">text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52f35f9d-0b36-4a8c-b42f-104ad6dc4d8b">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4cd3b78e-dd3d-4412-9c87-745f2cc0e92e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_829eb3a5-e607-428a-91d8-b18841c1992c">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9ed7a5b5-e958-48b2-86c8-6a42f15524c4">Ref</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f607a2a-4fff-47a8-bc2b-ed1ff425e986">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3e14990-bf5c-4f59-b6fb-75c52537f7ac">reference</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5731fd85-b9bc-446f-8adb-06e5d8ee135d">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8551d871-be77-462f-89aa-76dc67bbfd96">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d53d5aa5-34f2-4786-be56-382ef925b318">URL (XML IDREF)</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91be6cd9-99e4-4d6e-ada3-28b490ea750b">#
                  <emphasis role="italic">content​Ref</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7554079e-fcec-442f-bc80-5ed544938fbf">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47dc4fe5-e668-483d-b4fd-feda0963d556">performer</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c08bc8c1-9c07-4421-9f88-b1b870d1e2dc">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_630c300e-c9bc-4316-ba47-656b9175b21d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2bcd7c9e-8569-4957-86de-2a621486d8c4">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17ad3e23-659d-4544-8ce7-c7b1224f6261">assigned​Entity</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8125f49-ee5d-497e-ada1-1d70b77e7b8a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41be60f8-cf91-4fea-826a-b2f7a9cde724">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bccac5b7-64dc-49e4-9744-0fe1b2e2fdad">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d896cad5-8024-45fd-b8c4-849b03d494b2">assignedPerson</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5cd2f16-3270-4e0e-b415-0802b1039655">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a50ebca9-a7f9-4cdb-8a86-fd2189e5a18e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e2e7088-d406-4fb3-ae72-920d3ece712d">Reporting​Physician​Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89ffb4eb-3dc1-4b36-9de9-a22bc7e829f5">&gt;&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_06d78b97-aca1-4e7a-9a01-3d47ae07f224">name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64e75ae1-fe6e-4805-9bd7-d5a572b0e873">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb9b0e1c-4fa5-4e1d-8733-ae2b12c66704">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64d10937-d108-4f62-88ee-4a669faa00f5">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0fb6f51e-9db2-495d-8110-d6ee41515897">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2b912c2-edf6-4cd4-a006-4ec05d6e50fa">participant</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3cdf6ae-8157-4368-a0a3-18774f18633f">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58aec3a5-d67b-43cc-87a8-70db292b2636">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d76a9aaf-6762-4d28-911b-ad0d18993229">&gt;&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e943c4c-d8ca-4c71-947b-f90cb5b48d76">@typeCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8f0acef-1ff5-48e9-9d42-2e1c843e9c6e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e483ce34-af7e-47ac-892e-063d57f3594d">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4740c27-0134-4d1d-a1f0-187cba1ec183">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70201a9c-017c-4797-90bc-989126e73515">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65fda57a-e2e9-4c8b-8103-e06450216ec2">NOT</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e6f68b63-42c6-4533-ba75-5acdae89d1f4">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6ccb380-ce7d-406a-a852-2d4fe8629844">participant​Role</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9c04f38b-5946-4f73-874c-cb8498a3ee4a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2cf6efbd-baaf-4160-9a91-b11b051ed9fb">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f0bd979b-5900-42d0-8516-b793bc2a0fd8">Notification​Contact​Telecom</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_abe00b32-8877-428a-b78c-424dceb52539">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf0f6df8-f559-40ce-951d-d9f75ab43209">telecom</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50cd48e5-f156-4d55-bfb6-aa7ce281d139">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4585bdc-0020-4e87-9b57-1fb33cd603f2">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8cd324fb-45ba-48fe-894d-714fa8370ca9">TEL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4178f35b-ac9e-467e-a948-a504032bcc1e">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_377394f0-cdd4-4e9a-8693-72c63e1cb801">playing​Entity</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_37ff9670-be84-44d6-b79f-84219cdb5bf7">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_721d352e-eb5a-4f01-878d-9edc523d7971">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f27e5716-7dca-48a6-9c7e-7c160f8312c4">Notification​Contact​Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e40a6e3-52b9-49a9-8737-5ffc47f84e9a">&gt;&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de5656e9-9996-460f-b00e-8c687fd44f53">name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8d5814b-aa00-44da-b877-496d135f601a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7714a1f4-57e3-4642-9671-1310d27f2c26">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5724e7f-5067-417a-9787-30684dbf36ed">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.10.1" status="4" xml:id="sect_9.8.10.1">
          <title>section/text/content - narrative</title>
          <para xml:id="para_ae7149c3-b263-45c8-84e6-b5bd0334a048">Each documented act of communication of actionable findings SHALL be included as narrative in a section/text/content element, labeled with an XML ID (see <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>).</para>
          <note>
            <para xml:id="para_cf12f399-a30f-455a-b286-306a64eef0b3">The following text content for such a block is specified in the RSNA Radiology Reporting Templates, Template 297: Communication of Actionable Finding (<link xl:href="http://radreport.org/txt-mrrt/0000297"/>):</para>
            <para xml:id="para_3896a7aa-88f2-42c6-b40e-51f27a78bff2">method [discussed directly | discussed by telephone | described in message]</para>
            <para xml:id="para_2f0eb08b-3499-4fef-8c7b-8dab93b1ecab">by [person]</para>
            <para xml:id="para_20305e54-107f-4985-80ab-bec18bf4d963">to [person]</para>
            <para xml:id="para_83aeac50-f440-492f-ac0f-303d5162eb58">on [&lt;date&gt;] at [&lt;time&gt;]</para>
          </note>
          <para xml:id="para_3ac07863-5104-41a1-8a21-90229fd28a92">The documentation may also provide a linkHtml reference to the actionable finding narrative elsewhere in the report, e.g., in the <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> or <xref linkend="sect_9.8.4" xrefstyle="select: labelnumber title"/> section (see <xref linkend="sect_9.1.1.2" xrefstyle="select: label"/>).</para>
        </section>
        <section label="9.8.10.2" status="4" xml:id="sect_9.8.10.2">
          <title>entry/act</title>
          <para xml:id="para_e2991117-e1d5-41b6-8a65-592272f3f800">A structured entry representation of the act of communication MAY be included in the section. This entry does not necessarily represent the entirety of the act as described in the narrative text, e.g., the communication method and actual content of the communication is not represented, nor whether the receiver acknowledged the communication ("read-back"). The act/text/reference element SHALL include an XML IDREF value pointing to the associated narrative content block.</para>
        </section>
        <section label="9.8.10.3" status="4" xml:id="sect_9.8.10.3">
          <title>entry/act/effectiveTime</title>
          <para xml:id="para_e8958333-928e-4b35-843b-b2de78e2c086">The entry/act/effectiveTime element represents the date and time that actionable findings were communicated. The time that the findings were first observed is recorded in the effectiveTime element of the original observation, as linked through the section/text/content/linkHtml element.</para>
        </section>
        <section label="9.8.10.4" status="4" xml:id="sect_9.8.10.4">
          <title>entry/act/participant</title>
          <para xml:id="para_f0271e72-3ecd-42be-b13d-4604bf7245d7">The entry/act/participant element represents the notified party (@typecode = "NOT"). This could be the patient.</para>
        </section>

        <example label="9.8.10-1" xml:id="example_9.8.10-1">
          <title>Communication of Actionable Results section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.11"/>
    <id root="1.2.840.10213.2.62.7044794679.114296787"/>
    <code code="73568-8"
        codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC"
        displayName="Communication of Critical Results"/>
    <title>Communication of Actionable Results</title>
    <text>
        <content ID=CR1>Dr. Smith was phoned at 262-966-0120 at 3:14pm on
        Wednesday, June 4, 2014, and the 4mm lung nodule was discussed directly
        with Dr. Smith to explain the follow-up recommendation of ...</content>
    </text>
    <entry>
        <act classCode="ACT" moodCode="EVN">
            <code code="121291"
                codeSystem="1.2.840.10008.2.16.4"
                codeSystemName="DCM"
                displayName="Results Communicated"/>
            <text>
                <reference value="#CR1"/>
            </text>
            <effectiveTime value="20140604221400-0700"/>
            <performer>
                <assignedEntity>
                    <id root="1.2.840.10213.2.62.7044794679.114298686"/>
                    <assignedPerson>
                        <name>Jane Doctor</name>
                    </assignedPerson>
                </assignedEntity>
            </performer>
            <participant typeCode="NOT">
                <participantRole>
                    <telecom value="tel:262-966-0120"/>
                    <playingEntity>
                        <name>Dr. Smith</name>
                    </playingEntity>
                </participantRole>
            </participant>
        </act>
    </entry>
</section>]]></programlisting>
      </example>
      </section>
      <section label="9.8.11" status="3" xml:id="sect_9.8.11">
        <title>Recommendation</title>
        <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5c07b48-6ef0-43d1-a4e7-eb68dab3b4e7">
                  <emphasis role="bold">Template ID</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c9df37fe-4e57-41b2-8b10-f31a1a752b25">1.2.840.10008.9.12</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1e902dc1-4828-4278-85db-3f83bdea1d52">
                  <emphasis role="bold">Name</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc31db4c-1da3-4e49-abd9-e8441a6f5d33">Recommendation</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f915bd3a-e3c4-4d66-807e-3309baa51eb0">
                  <emphasis role="bold">Effective Date</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba7e1e5d-74ac-4d8f-a0e5-2421f0d16b5f">2015/03/24</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ae709d11-e9e3-470f-8ce5-53b7ee5405ab">
                  <emphasis role="bold">Version Label</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03d3aca8-06d2-49fe-973e-f6aedf697818">DICOM-20150324</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f0274d0-63d9-4880-b8db-595d919316b8">
                  <emphasis role="bold">Status</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ae85fbdb-e247-40a2-ae1e-cbe53dfa3102">Active</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88475526-9a37-414e-9eb6-476b20ea376a">
                  <emphasis role="bold">Description</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f86b143f-f961-4311-8c9c-2d141fa73418">This section provides a separate section to describe the study interpreter's recommendations for follow-up studies or procedures.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c1de643-d1aa-457d-b142-f714e30080bd">
                  <emphasis role="bold">Classification</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c876a543-e4bf-46ee-990d-a1876baa53db">CDA Section Level</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66a19f22-979b-4bf2-82fa-d353e48544b8">
                  <emphasis role="bold">Relationships</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c95392e-9bb5-4d11-b7ce-7f29288e2755">Included in <xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfaf74ad-1675-4f25-90ce-d57e600d3bbc">
                  <emphasis role="bold">Context</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2bb157f6-4fcf-43ca-a131-4abc501717b0">parent node</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b68b2182-6fc8-4946-aa9c-292a9085cc5b">
                  <emphasis role="bold">Open/Closed</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c35d77c-7b31-4327-80b3-629d22bc9a27">Open</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6adf85c4-df86-447c-8fb3-df64a761bab9">
                  <emphasis role="bold">Revision History</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a2e3266-e079-4fb1-afe6-cf18a2d69014">DICOM-20150324: Initial version</para>
              </td>
            </tr>
          </tbody>
        </informaltable>
      <para xml:id="para_42b86bf1-890f-4a2c-8e7a-17292046c7e3"/>
      <informaltable frame="box" rules="all">
        <tbody>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9d2f11d2-b337-4dd1-b5d7-a8f92d9b532b">
                  <emphasis role="bold">Business Name</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_d630f28b-9fb1-4645-b789-0037b219df71">
                  <emphasis role="bold">Nest Level</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_11416d7e-2f04-4076-887d-b600c6cfa7d5">
                  <emphasis role="bold">Element/​Attribute</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_55c8e12f-b2f2-4f1d-a6b0-a6646d32a5e6">
                  <emphasis role="bold">Card</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_b2357140-12bf-4983-8f93-27b8c57dbbee">
                  <emphasis role="bold">Elem/Attr Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_ebfedc46-13ab-4ecb-9e41-5fa190d17dd8">
                  <emphasis role="bold">Data Type</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_72ad08b5-6601-41bf-85ac-5ad8484c6a41">
                  <emphasis role="bold">Value Conf</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_9372e4be-c7c4-47d8-9869-d5d81982ceb0">
                  <emphasis role="bold">Value</emphasis>
                </para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_45044766-ce59-4948-922c-66b0e4c431ce">
                  <emphasis role="bold">Subsidiary Template</emphasis>
                </para>
              </th>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3dd01533-200a-4db3-9302-39cc5402cc24">
                  <emphasis role="bold">Recommendation</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aba8bed2-edce-4fb3-9034-cb18c259e1d8">section</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8cc8e395-827d-4c83-979a-c774bdb2ca39">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74b0c9a3-b5e5-47c8-9cb1-981ac86f4b8d">template​Id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c25314f8-ff58-4146-9a1e-50764337977e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b87ad17f-7739-422a-8052-5b80370d36a9">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f159d84d-0261-4df4-9b32-b9560e604304">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2fe5386e-81f4-479e-885b-13299d23154e">&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f43db0a1-039d-4d14-8b6e-ae4141b0af66">@root</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f2534b13-8c05-4b02-a7ad-3af600e26f30">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7cf1992f-b55e-437e-8ff2-9ac1fd214f93">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e6276df7-923c-43e8-a16d-ae1a4351d99d">UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0f5466cf-0939-484c-86cd-ab62f2a9e2d9">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5389932a-20fb-444a-bfcf-e77e877d3ead">1.2.840.10008.9.12</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc194b1d-7fbf-4e3f-9af8-1288ed90f6e7">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_72ddd483-d2a9-416e-85ce-57f511bc1d25">id</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_49ba7428-2826-4a39-9613-4c3a04bd2ece">1..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d7a0ded-a1ac-4adc-a023-6a43663bab79">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb413547-b952-4575-8f04-07fe59b06b11">II</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f30ac53d-b698-4da8-8062-2148374789eb">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b1bec75-71c5-4095-97d7-670be9d89d4e">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a8fe5ff-c1e6-4c09-8c61-e0379e96ccd2">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7736c2e-3590-4f90-86f5-d553cfb30bd0">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ff9e8c0-81b4-46de-aeb8-2cf0b6f91106">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28448e2e-d646-4061-86cd-f34e97fdc7ab">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e3689a9-9d8d-4cc3-b14e-80e82203ae13">
<link xl:href="http://loinc.org/18783-1/" xl:show="new">(18783-1, LOINC, "Study recommendation")</link>
</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92765c53-063a-4382-bcec-3a4f2755dc21">Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_14b8cdeb-7890-4327-a856-a8ff98f69921">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba4cf9a4-ac3c-45e2-bb5a-732651dc2014">title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d24af622-2c97-491f-99a7-c3732434c60c">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67cab5a3-58fe-4dd9-bcef-a2ba8030e35d">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43fce51d-233c-42b0-be76-bec8242dce64">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d4374651-6e76-4071-8615-a17009b7a8ec">Text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eed9d1ef-2c8d-492f-90c1-0da50b944d1b">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8649b0da-cbe7-4dc5-bc85-8a396456529d">text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b56618e9-e402-4fff-8ec5-fe968cd6d17f">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a94719f4-f2a2-4688-ab9a-120dc64a6b09">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac17dcc9-52d5-4416-8ff9-c74a94a8369f">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a4082ec-8c76-4a9f-a6ee-26a2ab27570f">
                  <emphasis role="bold">Content[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fec86eeb-8558-4e3a-8ad6-1306003c97a2">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8484f377-85f9-41b0-b993-764f33cd78c1">content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d86de1f-2357-4feb-b8e1-f34719631a27">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb0e5557-7865-4077-9133-5b8ef33f7208">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90051702-4555-437c-93f0-eb6ebbc4fec1">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb166769-090d-48d7-986f-8861d529740f">[See <xref linkend="sect_9.8.11.1" xrefstyle="select: labelnumber title"/>]</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f06b4389-a561-4e76-9d6d-3462a94f8ea2">
                  <emphasis role="bold">*</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41791d88-968c-442a-9d05-67211fb3b4f9">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d51f6278-aaf2-46de-8b21-66e39eb62773">@ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_367c911a-0016-407b-b747-24b4089f23b8">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d9a81ca-1a62-49f6-a83e-08219f044cce">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b123b845-a748-4f07-88be-352180209930">XML ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec92dd85-153b-4ba1-b5b5-eccfe91f1126">Guideline​Ref</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_732129e1-adc3-4966-b1c7-f8660f974eef">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3a4b9a6-3429-423a-8082-63f2bed2825b">link​Html</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70382a3d-8771-467b-b97d-a631122f0f69">0..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c283f27-3434-4a50-8621-19f9a6e38562">MAY</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f0c0876-7ba8-43af-9e20-91b3b8d74c60">ST</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0cb2dde-cb58-4aed-a465-5798b949de9f">Guideline​URI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12dd40d5-9166-4680-8cee-6f3055286b5b">&gt;&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_68d3f1d8-591d-4e62-8251-3338b238684f">@href</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3783a46-efb9-4b37-bfb4-1ef140baa92e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_661bef5a-49b6-42e1-a77d-513172c2ab5c">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d061442-c090-4338-b731-7a44e581e63d">URI</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_387127d0-7a19-4631-a806-b35dad3362bc">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_32f902f7-ce0a-4cd6-a3f9-8acdd19b45e8">entry</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7bcb3aa4-ac91-4df1-a5b8-bbb4544bb698">0..*</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fd8e7d16-2bd1-4d1e-be04-801e113f56dc">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b75d69a5-647b-4597-8140-d6fb25bdd6fe">
                  <emphasis role="bold">Followup​Procedure[*]</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8453e24-814f-4cf2-a3f7-03ffa54a3399">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4d4efc63-edd4-42fb-b243-c81b59581b4d">procedure</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42cd3615-ce70-4e22-a47d-923eb9c8b754">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59732d1a-5864-495e-8494-a45ced68002b">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56ff7888-f429-4c95-a533-1fcafc628cb3">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa77fc47-62b4-44ff-8a5a-e94c21cceff2">@classCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a5ed67c-c8c9-4cae-97a2-a58a90c6f7dd">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0659e686-5346-4072-a7b2-3b5d616568a7">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ee765a5-9ad8-477a-a65f-6396fa48d7bf">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0372007-31a0-4181-b335-e307942813a3">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4902277-ac20-456d-b3bb-810ddfbd8c31">PROC</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a84cb611-fe90-48b7-a0b7-10796649c2d8">&gt;&gt;@</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_056bafea-7055-4632-86e9-3a77618f6e08">@moodCode</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_890f97d7-c17a-40bc-80ec-e2e291ae9ad5">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75812dd9-3b68-4191-a16a-9ebfcd396728">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64b34b60-d602-4966-8a43-154561eb6514">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb6ee83b-6e80-4612-91e3-26a38ae6622e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac99bb00-d9be-400c-bf53-ccdda7716b5c">PRP</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89fc69f7-c237-46dc-a905-ea9812c3529e">Procedure​Code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5fc5cdbb-d937-4f67-9f00-2fe653fd57d0">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77e5bdc2-9c7e-49b6-a9fa-7accc8436490">code</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fadb8371-88f7-4f8b-b3c9-956460a73231">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9626f87e-9397-4ff7-9de7-fb149498dfb3">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e833995-bde8-4c22-a397-41a2d65b01d9">CD</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71da87bb-9fd5-4811-85d7-c086851fd1fd">Concept​Domain Recommended Follow-up</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e73bb77-71e1-41da-a97d-718df4402d82">When</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6468035f-4180-4a99-b50d-7a3f8c1ff95c">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28a33a0c-a382-4628-8eed-8be3c68f76a6">effective​Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7fdb860a-81bb-412a-82da-775c648bcf1a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4a293fd-e316-449d-a4c5-75a9be1e7c1d">SHOULD</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70cb6f62-995c-4bc8-a8e6-3ca6828b2feb">IVL &lt;TS&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eaad2a4d-fd3b-4a29-bf6c-e7fa63195ad7">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_df6f4028-c6a8-4f2f-96a7-b40a03628475">text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af027d23-1942-4138-81a0-0dda1f79212a">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55c422c8-4874-4a1e-9072-8dfcd0fb8a05">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e585c719-19d1-4c97-aea8-d5f40490b9a4">ED</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4282baa-37e2-4241-b2f8-a5f73e3d9789">Ref</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_01240997-b2c8-4316-baca-23806970ba87">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2505529c-c610-4b8c-994a-0682a772fe19">reference</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f00dc3b-28a3-48a0-b9d1-e7643a67bf3e">1..1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0170b630-4b29-4578-890b-ca811045d35e">SHALL</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d4e0dcc-e53e-47d7-a46b-cbc508ecda2c">URL (XML IDREF)</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_378c6211-448b-4b60-9c68-68638e85b51a">#
                  <emphasis role="italic">content​Ref</emphasis>
                </para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </informaltable>
        <section label="9.8.11.1" status="4" xml:id="sect_9.8.11.1">
          <title>text/content</title>
          <para xml:id="para_faa08167-a9f5-42fc-9c7e-8c885ecd6360">Each documented recommendation SHALL be included as narrative in a content element, labeled with an XML ID (see <xref linkend="sect_9.1.1.1" xrefstyle="select: label title"/>). The content element NEED NOT be top level markup within the section/text element; it MAY be wrapped in another allowed narrative block markup, such as paragraph, list/item, or table/row/cell.</para>
          <para xml:id="para_3cb4b533-5320-405e-b919-5cae090deefa">If the recommendation is based on a clinical guideline, a reference to that guideline MAY be included in a linkHtml element.</para>
          <para xml:id="para_752fe309-8650-47f1-8813-680fa9c43b1a">Each recommendation SHOULD have a corresponding structured entry.</para>
        </section>
        <section label="9.8.11.2" status="4" xml:id="sect_9.8.11.2">
          <title>entry/procedure</title>
          <para xml:id="para_3bb6ac5f-e8d0-4506-acc6-439c13b7a15a">The Recommendation section SHOULD include entries for recommended follow-up actions or procedures.</para>
          <note>
            <para xml:id="para_efb927a1-e77f-4218-aaa0-a988f7bd95cd">While this entry may be a trigger for a tracking system for ensuring follow up on recommendations, the imaging study report only conveys the interpreting physician's recommendations.</para>
          </note>
        </section>
        <section label="9.8.11.3" status="4" xml:id="sect_9.8.11.3">
          <title>entry/procedure/code</title>
          <para xml:id="para_f6464e28-b1a9-4a0e-be87-ba54efef3d57">Vocabulary binding for Concept Domain Recommended Follow-up may be further profiled in sub-specialty guidelines.</para>
          <note>
            <para xml:id="para_6cdb2f40-3a0b-48b0-b26e-5038c7f4cfba">An example would be Value Set <olink targetdoc="PS3.16" targetptr="sect_CID_6028" xrefstyle="select: labelnumber quotedtitle"/>, incorporating concepts from ACR BI-RADS <superscript>®</superscript>.</para>
          </note>
        </section>
        <section label="9.8.11.4" status="4" xml:id="sect_9.8.11.4">
          <title>entry/procedure/effectiveTime</title>
          <para xml:id="para_258478d1-1a7d-4d16-a53f-0d2c74a7bd14">The HL7v3 IVL &lt;TS&gt; Data Type used for effectiveTime requires the specification of absolute dates, rather than a date relative to the date of the report.</para>
          <note>
            <para xml:id="para_99d9ea05-f9a1-424f-8a5d-db9811f06dd0">Thus the concept "follow-up within one year" needs to be encoded as a IVL &lt;TS&gt; with an effectiveTime/high element value one year after the date of the report.</para>
          </note>
        </section>
        <section label="9.8.11.5" status="4" xml:id="sect_9.8.11.5">
          <title>entry/procedure/text/reference</title>
          <para xml:id="para_78e5c67b-3fc9-4621-a753-ad8983ea3844">The procedure entry SHALL include a text/reference element, whose value attribute SHALL begin with a '#' and SHALL point to its corresponding narrative content block. See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
        </section>
        <example label="9.8.11-1" xml:id="example_9.8.11-1">
          <title>Radiology recommendation section example</title>
          <programlisting><![CDATA[<section classCode="DOCSECT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.12" />
    <id root="1.2.840.10213.2.62.7044779.114265201"/>
    <code code="18783-1" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Study Recommendation"/>
    <title>Radiology Recommendation</title>
    <text>
        <content ID="rec01">Biopsy should be considered. Follow-up at 3 month interval.
        </content>
        <linkHtml href="http://pubs.rsna.org/doi/abs/10.1148/radiol.2372041887"/>
    </text>
    <entry>
        <procedure ID="RadRec1" classCode="PROC" moodCode="PRP"/>
        <!-- local coding scheme -->
        <code code="9191919" codeSystem="2.16.840.1.56789.6.1"
            codeSystemName="My Hospital Coding System"
            displayName="3 month follow-up"/>
        <text><reference value="#rec01"/></text>
        <effectiveTime value="20141213"/>
    </entry>
</section>]]></programlisting>
      </example>
      </section>
    </section>
  </chapter>
  <chapter label="10" status="1" xml:id="chapter_10">
    <title>Entry-level Templates</title>
    
    <section label="10.1" status="2" xml:id="sect_10.1">
      <title>Coded Observation</title>
      
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_642e3559-40d5-43cf-96d0-f5cf324dd4e2">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9e55c117-6995-4df9-ae60-4d861f3543fb">2.16.840.1.113883.​10.20.6.2.13</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c61bf305-3316-4661-a85f-98379e1bb701">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c9d7eb0-1875-4454-af1b-4a55074e8804">Coded Observation</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6852d2d-bc99-478e-b68f-d857567dcad1">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac6ad710-7a3e-40f4-b292-3ac2a4e0a525">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e55b465a-9949-40ae-93d7-af3a763eea04">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a965acb7-1326-48fd-a83a-d29c1c49c823">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2175c19e-6c3c-4371-9080-630f506bdba2">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b58d5246-7b40-49b2-b30d-236a1a6814b8">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_22cedff0-73b3-4b1c-8175-ff24d96f9b4e">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b5972503-e5d5-42dc-8d6c-a0851575af37">Qualitative or categorical observation using a value of type CD.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_770459e4-de88-4f9f-bf29-22701f515cc1">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a1a334f-d762-491c-9ab3-dad6a01d0cfa">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cde7aba0-7d99-4872-98e3-edc1339395ac">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94cb32f7-2e8f-4c54-bce6-5e5993d5e09d">Included in all sections</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf18221b-bec7-45ab-bf07-301d4ff1a9d4">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb2b1f6c-5de7-4ab7-859e-1ae585c1ad43">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a2fb45f-c2c4-49fd-8855-101e3528e40d">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88a8f52a-b181-41c2-a751-828e9fe8e19c">open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_364de7a7-d631-4bc1-bbfe-9c94cc66b34f">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8422ce8-e8b8-4187-8419-1a1d15490b7e">From Consolidated CDA r1.1</para>
              <para xml:id="para_c81ba395-f25b-49e9-802f-832a86eb5b96">DICOM-20150324: Added optional negationInd, interpretationCode, targetSiteCode, and methodCode with Business Names; added optional subject Coded Observation</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_882412b9-4bf6-486d-9146-906266bbec83"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_eb5895f1-7958-4936-9e90-0021c4ecacf7">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_43cecf1e-2689-4b6a-bcd2-c9a4fd47c478">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_cbfdbd7a-e873-41e5-a5cc-7c89cd7fd872">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_bca4c45d-0ddd-46e2-aa4b-92a7ac0d5730">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3e0f9fe6-b1ff-4eb0-96cd-2569df9c1f56">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3598729b-2d73-4789-91a1-f0527e086891">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0d9743d5-455d-4486-ae2c-abc18b74785c">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e37672d2-347f-4782-9944-1c8df7881f45">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_eaa2b5a8-0286-4a5f-9b76-14a89ddae67a">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98513e61-0116-4285-aeb2-84c51ef1fc23">
                <emphasis role="bold">Coded​Observation[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7cbf794a-fe7c-4577-8094-55b63b3d021e">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d65cc01c-e7e5-4320-9516-b191f91886e9">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77983c41-3c97-44c7-aa8a-d0351be3da8e">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb89847d-5890-4528-8eb8-8d4134fbdcd4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d7dc53c-31b5-4730-9677-41d223f34d32">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd404a9a-58c1-463d-813d-4b8191f9601c">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_72245157-fa89-4eba-9074-1afc4edec2b6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8814861f-2274-4c4e-8b03-48826f01ed3f">OBS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1c96e5c-4d42-48e3-b302-1bbb3b4a2be1">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0595d19b-c5ac-4748-a72a-0ede62d239f6">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4a097119-dbb5-4de5-9dd3-64362c25296a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e14d18a-6a40-4434-b5c9-614cf8e298e0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7e44e7b-53d6-44a9-afea-8dbbeeadb9c6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89089f52-228a-4089-9d33-7070f4e35ccd">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9025c87c-2824-410c-b06d-3726df6f6aca">Not</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d9ab5619-02f0-4aa7-9ed9-eb8edd21fe85">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a296399-56e2-46cd-9f88-7641a55c5613">@negationInd</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0173af3-4788-4165-8324-cdd45a6ec7dc">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b55b1a3-1194-4dab-a5cc-f3bef2e59971">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a43da26-64eb-4372-971f-dd776bc1011f">BL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a026a9f-bbdc-45aa-aa59-84585999e642">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4876dd2-9639-4cd6-8e71-05dffb43c75b">true</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_371d7532-0b2d-494d-8f71-6559b81fd86d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8bf3a06-2809-4322-8dd4-43b03ff89689">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d12c06e8-8a9c-48ff-bf68-7a378981ba9b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_505e2411-5c3a-41c8-940b-b304133177c6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_477339e5-34ed-4292-b15c-87f236201d46">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a7e3057-f7b9-45dc-b01f-3b3612652372">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_daef8cb4-5ca8-485b-96f1-6703c7b7253e">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d23260d5-a2c5-4274-8f49-437a78a3e17d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7d7923a-0200-47f7-9d50-617008670995">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30bde6e8-f63f-4b5e-93cc-386a687044b1">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e31354f-2808-4b5c-b64a-c76678e6d2d2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_45f7de77-0d3a-410a-bfe5-58993e420302">2.16.840.1.113883.​10.20.6.2.13</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0810a81b-ffa0-4005-aaf3-834ebec02382">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dfcb7867-4ddf-4a3e-aecf-67f05fb6ee95">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0d660c6-d7dc-4ce2-a758-f45663903ddf">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23464aa9-2d54-4bbd-a97e-b73765a4f1d5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b0074e7-14c2-45b8-9bcc-27888a3cf819">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f05fd79a-ede4-4717-a943-3d6873ffe965">Obs​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd6b93c4-9089-43fb-b144-fabc650957fa">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d9a0a71-c1ca-4413-9fff-a34fa6dbd5b9">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98da5d5d-157b-4275-90f1-6341a3eef336">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_599ce0e0-0e06-42ea-93c8-1b6579ff58b2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb8220c3-04e6-4901-b2c5-6cc10e7a8d1e">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fca7e64-10fd-4dbc-abba-078ee4e01b2c">Concept​Domain ObservationType</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_add0346a-253c-4b4a-9b36-4cb0c61609d2">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60b4be28-67ee-4f78-932e-0499707edecc">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b960990a-3512-4255-b243-de064dbe2f61">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d8ac76f-3121-48bb-a6f0-1010c08ca7e2">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_509c5c35-faa7-4da7-950b-72baa59b208c">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_99c3c048-c617-411f-acf9-02b2f791991a">Ref</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4991088-e8c3-4ce8-baf2-7278e29cd2a9">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e5586b5-aaf9-4941-b28e-9f3c79f88bde">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f78c65c9-80a7-402a-a504-cd5783a7093c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3c01254-4b01-4ee5-86b8-26120a1495b5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_93c4e2c4-5996-4940-a1a1-e410b0bc9d74">URL (XML IDREF)</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f610e7e9-ac48-4426-8ccb-07d5274383a2">#
                <emphasis role="italic">content​Ref</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_104e0b59-3310-4e77-8b6a-2cbcc19811b6">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_907bc817-5c1e-4d7b-b346-1416a3afdd97">status​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f6f6564-2e6c-4fec-a7e4-0c7c1d549e79">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9585a692-860d-4d40-88d5-0d96e57dcd9b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3362f181-9051-4db3-9bbd-bc40e723edef">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb737386-6812-47e7-922f-7e7ddf337067">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63ddb1fc-575f-4c2a-8899-af97d279347f">COMPLETED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8efc1156-5c07-4b7c-b5f3-d87907998eca">Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d82eb25-b765-4bac-b2fe-9b165b647f3d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84c0496b-d74e-4373-a2b5-f8fd38a6738d">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb0cb135-fe81-4f6f-a004-6d501cbb4d5a">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9c559b6-59aa-43aa-8f71-18c38280a0f4">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6269f005-7dca-4584-8508-95370f0fca91">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_593216f5-cc1f-4f49-846e-de3629616ef3">Obs​Value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aeb44429-1b98-490d-bde7-199a9324fd90">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cdd02ef6-040d-440a-a499-e7e89e1c9636">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d0e3075-fa1b-45f3-84d4-c3e7f37587a1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_600cb316-1ce7-4d5e-a3dc-9cd7b562dff5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71bd0584-5617-4621-bf6d-de983ba0e8ad">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1be7d5aa-f9ed-48a2-8183-91b4de6cdac7">Concept​Domain ObservationValue</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f128b0d-dad0-4c2a-8e91-7410fc3bfd43">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_32c7aecc-1c46-42ae-9b9e-6ebabc15a065">@xsi:type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7bd15be-175c-4ee6-9b83-e144c421b684">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9efc2bba-41ce-42eb-982f-ff4ffec24cae">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c6bc1a5-8b4e-405a-b4b0-fdfaf29b8536">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_80a8f956-ab3a-4eee-b494-74662d992ce6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff8535c8-3794-4894-94df-ea406296c9ab">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f13b324a-d392-4aba-a9aa-edfc237e6a36">Interpretation​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e54dd8e1-9d5e-42f4-8e0b-cbcbc4f1e746">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ac7fc28-e2cf-468a-9fce-3d82e4a859ab">interpretation​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d4cd35b-61d0-4cf0-8608-3efa2d2340ff">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2abea570-8ed4-41f8-8016-c2a43b76ca4e">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f2aae59-372a-43f6-a9ed-1377efecc020">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c62d3f4-fa33-46ad-8b2a-50a4f3b54da4">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_244ee062-3e2d-4cb5-bed8-6058883e63ea">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.5" xrefstyle="select: title"/> 2.16.840.1.113883.11.78</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_060ac8f2-3ce8-4959-b0d5-87eb6a92c476">Actionable​Priority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5cf269a5-30ae-443f-a92b-4f4cffe95b70">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ef7c5f3-0112-40b5-852d-cf79371465b6">translation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3ea5903-318f-4819-8a1b-0bd2cff6873d">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9f70ee7-a9be-4f11-8832-8171cbbfddb2">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6aea97c-e57d-4cd8-b684-269860ce1df0">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ddbc5aa7-27c4-4bd9-80a6-c732909b59df">MAY CWE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54d8ecf3-0fcd-4dc9-a756-53bb92c343db">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_7035" xrefstyle="select: labelnumber quotedtitle"/>
</para>
              <para xml:id="para_b2068488-0cd2-4f52-a6fd-66c5a03168d5">[See <xref linkend="sect_10.1.3" xrefstyle="select: labelnumber title"/>]</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_daf427a8-f1b3-4c67-9393-b79df2eb8749">Target​Site</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9900bb0-5e6d-4851-ad57-11c18431869b">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e96148c7-0f69-4231-8765-1b06cd64394c">target​Site​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4db19f0c-c7d8-4e99-8c4c-e16afcd26b08">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eacf436b-f544-459d-b988-e35fba1b4231">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f92277b-6b12-427b-b362-20aa6d8fb301">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cb3ccbae-9e91-4ff3-87da-545c8a8c1e4c">Concept​Domain ObservationSite</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_809d6698-dacd-4485-aa1e-39bfa266885c">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0812966-a947-40f6-9343-5035ac9fa7db">qualifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21046007-6416-405f-85cb-475e78811476">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c1f9c7f-a3e8-4970-99cd-fabfa30d9b7d">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b50d7b9-f90f-4d75-bc64-87b2dc12fae9">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0623eee1-7cb0-4707-ae89-bc937b6b39ab">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd8135ae-42e9-4da7-ae95-d413ed9649aa">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d550c9b-987d-43ee-8387-547a5620e3ff">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7bd6abb5-d056-415a-bab8-5b130755d529">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21c6f8cc-acac-4601-bbae-f850ac8767b2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ea2450e3-48b8-4259-89c0-2949567abad8">(272741003, SNOMED CT, "laterality")</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0f57c77d-e851-4e8b-9a5b-4e0532008bfb">Laterality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf9d1afb-cd36-4c21-a0c4-21db4e07abad">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d672cc8-f0c7-468c-9e8d-209aa28d4a31">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd70adcf-9fbd-416c-8498-ba58dd589d92">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29b72dd9-fd8b-4f79-9691-3f137ec57ff5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6aa1fac2-0e46-4fd9-9eef-9cde37e21d3f">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07633eb3-59a9-4116-af6f-942375bfd28b">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b248e2eb-fc9c-4ec4-ac1e-9b7617b27db5">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ead1a6a-436e-4766-8ce5-d7e8bfb0487a">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_910148e3-db29-4530-b0ab-20a8f7dddfd1">qualifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bbc8e66e-2cc3-4008-8542-874b2cb09c82">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6737378a-4803-4319-9364-7d8dcc59ba65">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f21398f-de79-49a0-8793-72d9f0c33b84">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f023251-b7a7-4fb1-a71c-1889a42ce670">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9bd9ea25-b12e-4789-beca-e0aa487620b3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_374cc29e-e188-4f69-a142-93dbffd82ffd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_efe12bcd-03db-4210-adf7-c00991f3b425">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b4ffc4e-e611-4de9-ad0a-e8c9f958b19c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f3217b9-ff79-4811-927a-0681a9f85abf">(106233006, SNOMED CT, "topographical modifier")</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_76c19788-b447-410b-9bbd-13f5ba8b8576">Topo​Modifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_264fa8f5-dd46-4079-8515-a4dbb9062829">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9e526831-c2f9-43c5-8690-40226e507563">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0028f2b-0d7f-43e2-af32-8a6f3a45924e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f72bee19-02a8-4bb1-b634-364dc8e1b57a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd6bb5c2-f8e1-4c75-bc4b-69c057d5dd86">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50a40e81-08b4-43da-afdd-eb058d535071">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_175450cb-8621-496b-8a55-6df858bcf8ea">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_16699790-d7dc-427c-80b3-b9be72312cac">Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_484481e7-8a47-4c26-b145-d71cd706f162">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3cc695df-b209-4291-8a31-2920a77eed35">method​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f44955ce-47f4-4bf3-bacf-3d80bd74f70c">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29b25537-0fcc-465f-bd28-504572c46a7b">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7468cd09-d9d2-4ea5-bd89-3efb2b921f5f">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5921857a-f370-4db8-a9e3-d155355ed087">Concept​Domain ObservationMethod</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ede099b-bc5c-4b59-96dc-f4c9132c09b6">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a410516-374c-4c0e-8e43-8cf5409beb0b">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9e42577-e9aa-4cf4-983d-916aa4752ad3">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_574c3cdf-eaad-4da7-8c86-46580f963bfc">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73e44522-4865-4797-98fd-4106e046983a">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fbd992f-e081-4894-8cd2-403770425839">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_80b5d0d5-c5ab-4815-afc2-af28190f73d1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c44697e-45bb-4d28-8e47-7a7c00b5c5f7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ca33e24-b9a2-4fde-9cf9-4f8680cdd412">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8af69059-363f-427b-8fcc-0e41150d8e30">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9958c1cc-a312-41b3-b14d-65d4059fc56c">SPRT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e7c47a2-237d-4eec-af13-b45c5a279742">
                <emphasis role="italic">SOPInstance[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8acad3ff-c93a-4e38-b56d-f6afeb327fc6">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e773253-be50-4ce4-b1a1-6388e5db38f5">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5cb6e62-0930-4107-9a27-51a8f875263a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_053a2e8f-7be9-4982-b4a3-e3717c38d992">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3acf64aa-e66a-498e-8abf-c4c72c19f01c">
<xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.18</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7ee8944-a016-4dd6-b0f5-e7c879dfb60f">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c480ef06-76d4-4b43-838c-da9eab443d1a">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9735364a-1b29-4bab-aef2-20406187c7ad">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b03623d-9b53-48d8-bab8-8462dd634683">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e01e68e9-6a56-45d4-a777-04bfdd644152">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_940a1cdc-dfc0-4497-81bf-473fc679aaea">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ada0316-453f-45bd-ab55-34fbd9fae16f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6966b6ca-2e88-410f-973b-7787ea088e06">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9a73c43c-5e5c-4dac-8993-1c5307f50261">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_adc2300b-cdd7-47be-866c-571635e4b37e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df5fc1a5-7107-434d-8dde-74319e5349a3">SPRT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d24dd7b1-718e-40a4-8c56-57a82e5d1625">
                <emphasis role="italic">Quantity​Measurement[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_56ff5ed7-342e-4bcb-b3b0-9bf0981aee5f">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c752759b-8942-449e-a796-0eb2f4a82b07">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d410554b-d5c0-449c-9a0c-a3e8e19aa472">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_977812c8-c72e-429a-b24b-431db33d2085">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5fbcbad1-9376-45f9-8424-2b6aaeb7a938">
<xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.14</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eec5d83c-76bf-4395-b3fa-5adc8f269e1c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_24eff890-4e44-48ee-b857-61d727871069">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_80496505-a0bc-4526-82ba-bf9e4d9e3226">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3f6faf9-57b8-4be6-8381-d496454ad2b1">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d970c27-193d-448e-aa41-4520f3075556">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c5d625d-896c-4ade-9e0b-caeac34585e6">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da890761-3c0b-45df-90b8-30e26cf843e1">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfe73531-999e-483d-ab78-12cf28917ca8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1911d237-f4e5-4672-8307-6e9cdcd100ab">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_822e2957-148f-47eb-b7a5-857b252c8949">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3ed05a25-8e6b-4988-b85c-dcb58ba60d86">SUBJ</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97f02864-7178-418a-bf63-31738eda0bdf">
                <emphasis role="italic">Coded​Observation[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19512740-7e5c-44a0-b77b-f345458b484e">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc3058ad-43b8-4e92-8635-eeed564df355">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_55d3ab98-8557-4603-88d2-72a7822d3521">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d47781d4-89c3-4c6b-b4e2-120795c51a36">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0ed26f3-e873-4b33-a6ad-ea4575023722">
<xref linkend="sect_10.1" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.13</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.1.1" status="3" xml:id="sect_10.1.1">
        <title>code and @negationInd</title>
        <para xml:id="para_b0b6b836-6941-45a3-9046-a3f8ae6c549e">The Observation code element has an associated Concept Domain ObservationType. A representative binding for this Concept Domain is to the value (ASSERTION, actcode[2.16.840.1.113883.5.4], "Assertion"), providing an assertion of a finding concept in the value element.</para>
        <para xml:id="para_81261737-b62b-4e6e-a6d9-e9e73697ef71">The Observation may have @negationInd attribute "true", which together with the code "ASSERTION" indicates that the finding was not observed, e.g., to represent "No finding of stroke".</para>
        <note>
          <para xml:id="para_0b964e8f-7200-44d3-a92d-7fb6feac678c">This is the pattern used in Consolidated CDA for negative findings.</para>
        </note>
      </section>
      <section label="10.1.2" status="3" xml:id="sect_10.1.2">
        <title>text/reference and Related Narrative Block Markup</title>
        <para xml:id="para_3e565c9b-883a-43a6-a23a-d267619ab0a1">The Observation entry SHOULD include a text/reference element, whose value attribute (not to be confused with the value element of the Observation class) SHALL begin with a '#' and SHALL point to its corresponding narrative in the parent section (using the approach defined in CDA Release 2, section 4.3.5.1). See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
      </section>
      <section label="10.1.3" status="3" xml:id="sect_10.1.3">
        <title>interpretationCode and translation For Actionable Findings</title>
        <para xml:id="para_26d0d7a8-375b-4f1f-a805-d0f684b6776c">When an observation is unexpected or "actionable" (one type of which is denoted a "critical finding"), it may be flagged using the interpretationCode. For very abnormal findings the interpretationCode element SHALL be set to (AA, ObservationInterpretation, "abnormal alert"). Unexpected normal findings, e.g., no findings of disease when patient treatment had been planned on the presumption of disease, may also be flagged using interpretationCode (N, ObservationInterpretation, "normal").</para>
        <para xml:id="para_2f556763-32c2-4034-9f55-85a5a735a95a">The translation element of the interpretationCode may be used to provide a further classification as defined in a regionally- or professionally-specified value set. This template identifies an optional value set for the ACR Actionable Finding categories 1, 2, and 3, as defined by: Larson PA, et al. J Am Coll Radiol 2014; published online. DOI 10.1016/j.jacr.2013.12.016.</para>
        <para xml:id="para_7dc07f33-7717-4ba6-8918-f42e02014946">The narrative text associated with the actionable finding SHOULD be highlighted using styleCode Bold. See <xref linkend="sect_9.5.1" xrefstyle="select: label"/> and <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
        <para xml:id="para_d88c5100-fe21-4704-b935-37eea4e4eb87">Actionable findings that require a specific follow-up action or procedure SHOULD be referenced from a recommendation in the <xref linkend="sect_9.8.11" xrefstyle="select: labelnumber title"/> section.</para>
        <para xml:id="para_85440c15-2c05-492e-ab19-7486dffd0d33">Communication of actionable findings SHOULD be documented in the <xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/> section.</para>
      </section>
      <section label="10.1.4" status="3" xml:id="sect_10.1.4">
        <title>targetSiteCode</title>
        <para xml:id="para_9a0652b4-e071-476e-bb7e-ee5bfe55c306">Each observation needs to fully specify its site/location.</para>
        <para xml:id="para_0a9ef472-cc06-4b6c-bc89-b94e4880b8ab">COND: If the observation site is not pre-coordinated in the observation/code or observation/value, it SHALL be specified in the observation/targetSiteCode.</para>
        <para xml:id="para_5760d687-f9f6-498d-866d-721ba4309c53">COND: The qualifier element for laterality SHALL be present if the targetSiteCode represents a paired body part and laterality is not pre-coordinated in the targetSiteCode.</para>
        <para xml:id="para_e11971b0-c541-4a3b-bd67-10cc8ea63a4e">Note that inclusion in a labeled subsection (see <xref linkend="sect_9.8.9" xrefstyle="select: label"/>) does not imply a finding site for the observation from the title. The title is not semantically part of the post-coordination.</para>
      </section>
      <section label="10.1.5" status="3" xml:id="sect_10.1.5">
        <title>entryRelationship/@typeCode=SUBJ/observation - Coded</title>
        <para xml:id="para_41b9b5c6-be07-48f2-a8a2-f91593da6bda">The Coded Observation entry MAY include an actRelationship of type SUBJ (has subject) to a subsidiary Coded Observation (recursively invoking this same template). This allows the constructions of complex clinical statements.</para>
      </section>
      <example label="10.1-1" xml:id="example_10.1-1">
        <title>Coded observation example</title>
        <programlisting><![CDATA[<text>
    ...
    <content ID="fnd-1"> ...finding of a right hilar mass (abnormal - class 1) ...</content>
</text>
...
<entry>
    <observation classCode="OBS" moodCode="EVN">
        <templateId root="2.16.840.1.113883.10.20.6.2.13"/>
        <id root="1.2.840.10213.2.62.7044779.114265201"/>
        <code code="ASSERTION" codeSystem="2.16.840.1.113883.5.4"
            codeSystemName="actCode"
            displayName="Assertion"/>
        <text><reference value="#fnd-1"/></text>
        <statusCode code="completed"/>
        <effectiveTime value="20140914171504+0500"/>
        <value xsi:type="CD" code="309530007"
            codeSystem="2.16.840.1.113883.6.96"
            codeSystemName="SNOMED CT"
            displayName="Hilar mass"/>
        <interpretationCode code = "AA" codeSystem="2.16.840.1.113883.5.83"
                codeSystemName="ObservationInterpretation"
                displayName="Abnormal Alert">
            <translation code="RID49480" codeSystem="2.16.840.1.113883.6.256"
                codeSystemName="RADLEX"
                displayName="ACR Category 1 Actionable Finding"/>
        </interpretationCode>
        <!-- although "hilar mass" is by definition in the lung, the observation.value
             does not describe right or left lung, so targetSite is required -->
        <targetSiteCode code="3341006"
            codeSystem="2.16.840.1.113883.6.96" codeSystemName="SNOMED CT"
            displayName="right lung">
        </targetSiteCode>
        <!-- entryRelationship elements referring to SOP Instance Observations
             or Quantity Measurement Observations may appear here -->
    </observation>
</entry>]]></programlisting>
      </example>
    </section>
    <section label="10.2" status="2" xml:id="sect_10.2">
      <title>Procedural Medication</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_494a73ac-03a5-4205-99bc-dd1ffd26b74c">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_baf7c934-c490-429d-8d91-ce6e8d52a14f">1.2.840.10008.9.13</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20bb560a-5fba-4d90-a79d-2bcb4619444c">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3037c8d4-0559-4dc9-917a-b4199048b310">Procedural Medication</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3bea93a7-7691-488f-9cca-083a671afc5d">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5691f9f4-e9b5-4cc6-8071-8f0ae7c7b8e1">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a804797-6b72-49e5-b306-44d2f9ed31f3">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6de7a336-c32a-47c0-a8e4-c0c66c4ac017">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a5ae15ce-8184-4685-b12c-79931aa3bb5f">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_466b23b2-89ce-4bea-a4f9-5b735b964d57">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_152f526e-0821-46fc-8693-37e753c15fb8">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7843e140-2da7-4603-b009-e33d6d31ce35">Procedural medication describes a substance administration that has actually occurred prior to or during a procedure (e.g., imaging contrast/agents, anti-histamines, anti-anxiety, beta blockers to control heart rate during procedure, etc.).</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0983ff92-12ce-4d2a-88e9-ebc7bc0098ab">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_24a8bd66-1fac-4442-90f6-18e84cf90917">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d2898173-47dc-4d28-9700-7f9ac6fa6176">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6bb000c1-d2b6-40ce-b4cb-a90a4d095985">Included in <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_056fc3db-2f23-4222-888b-6d54d578d858">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6923ec0-7622-4b79-b1ff-306e38fea121">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_499bf351-4f50-4606-b865-2ca8e5b90d9e">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0f49d2f-3ad0-4c8e-bc22-fcd3d54994dc">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bac025c1-1368-404c-bb81-4c28f9c8f85b">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f398566-247a-4d36-8389-4d6f123c7433">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_51997b7d-b49c-4369-a5ad-9b874f44a3b6"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_511050fe-1ec5-4f57-b837-ea0e3cc4e4f5">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a3afbc12-ed08-4fed-9f31-735615126953">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_62579a9e-5c0b-412d-9ef8-b123e79cc8bd">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2434c95e-6c1a-4c1a-bf9c-93314276ea47">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9fc62ae3-8c66-45a0-82ac-66a112d56302">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2ce34c4a-ee23-4c21-b26f-9095b561be41">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9ce575db-3c70-4bbc-8898-34912a8ced0d">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_273848c6-7892-4317-bfb2-6963b4fc60cf">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1dd4b0d5-2c87-4a3e-b934-4e644d353fc9">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_051a50dc-4be3-4f92-8bbe-f978226ae0c4">
                <emphasis role="bold">Procedural​Medication[*]</emphasis> or
                <emphasis role="bold">Contrast[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7276764d-3a54-4593-a47d-0a2cf41530d2">substance​Administration</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f23ae84-0361-485c-86e6-4be5a57cd1b7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dfdc20a5-a806-4ca9-97fd-7ca730e79459">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_375e5c8d-f53c-4719-85c7-adea43af99f7">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c443cb1-5b51-4886-a807-1c3e7401bb1a">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d6e1909-cd23-4e11-8bff-267b71bfe41a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_593ea569-48ec-4e9b-b998-1daf50b042b4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30c644f3-1873-48ff-991a-3fe18aa5d4c7">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_073ceb4a-e655-451a-b81f-cf4aac492ea6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b68291b2-c2cf-483b-9b08-0aa6a39d799b">SBADM</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a580ef10-6e11-498f-a8d1-3302f5e3e732">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b4d69412-cbdd-4a0e-9ae3-f7e0bb481dc4">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd3757a8-2a9f-4e49-aeee-be38f8916c6a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_10770d82-71ac-45ac-9a42-6ab6ee95faae">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_49b14b2f-f1ac-4895-b5f8-97995c5893f9">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30b7fe98-5426-45b9-be49-f6140e379deb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad2004c1-d36d-4cc7-aa39-538fd017b559">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee7b0112-94a3-4cb2-9df9-7ae557159d94">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cc36975b-62bf-4bb0-9fd0-510f4ac0d178">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aa5cadc7-e2d7-4037-8bb2-491b3471635c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a49de7db-bb74-4975-bb03-2e0b375af9b9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9be71592-0f76-465b-85d8-72870bca6fe7">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_adca253d-bf17-4ea5-b03d-904a69deaaaa">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4180810-6d7a-43cc-98b3-405a39940c8b">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd9b7392-616f-42d0-a790-967cb2e4d646">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95cbe93f-26be-4599-b1a3-bd1b48e44556">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2e10f10-1ada-49d2-998b-af26c60ea535">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca000df5-ca95-41b5-8a98-d0c4fbba9ed9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e91d6af-e093-410f-acbe-cad6ce8aa38c">1.2.840.10008.9.13</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2780f16f-de7a-4185-8daa-007a49cdb406">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f1108ea-9161-4e8f-aed4-c5d93b60bec2">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_406c2a67-5db5-4fd1-8fd8-956bf2c60dfe">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd19dd29-9f4a-48ba-9351-f5dfab3a62a7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f196843f-31a8-468d-9aa4-7dede6255602">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2efe0fd4-0912-42a1-999b-b729dedcd52d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2095e3b1-8001-4375-895e-586fdba96505">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_28fa8787-d406-4482-b02b-f9ff37cda948">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_16d685cd-10c4-4011-8f1f-c35f467efc3c">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_493d08b9-8d23-41b8-8ca5-a3dd7791500a">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f2b9b2f5-d461-4212-88d4-0f4d9fae3c3f">Ref</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21c03c5a-8373-46c2-9ef3-1ec0922c1edf">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87377876-0bb2-46e3-88d3-e4d3fe617348">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_324743a1-1610-4606-85de-e3196f6d15d8">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_530fb3d4-3d78-40b1-99cc-8f1a7423e2bb">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0c41aaf-f18a-48dd-aa00-e31bfdda6b89">URL (XML IDREF)</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d1b7423c-a006-4e9e-a9e6-05e9e8ffc809">
                <emphasis role="italic">#contentRef</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b580ca35-3963-4db9-ab84-8bf0c9961c18">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ef904153-dad2-4ed1-9aed-3481170e2ffc">status​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2cb40d63-9e94-406f-87db-4252f6972e75">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7cf85d2b-9891-40f7-be1d-65ca62c346c6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_215c7da8-e123-45ce-9a50-5eb2af9fcb26">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71d55070-c934-43a8-beff-4b53e0f894f8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_35fe700d-6d61-4452-9d0c-4fc5f4fcb8fb">COMPLETED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e4c71e1-9030-412f-bcf4-d45944cae7f5">Route</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_72b62640-9fc7-4984-8c7e-fff58cd87566">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a1d7855a-8fdd-4a8f-9d56-c0de1f099b17">route​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6ec00108-cbd5-495c-b94d-2b572fd385cc">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_efbdcde5-3e51-484f-a0f5-b964e2551e6f">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de67b745-3643-4666-861e-92d76d99aabf">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e062516a-b645-433d-ad3b-e83104ea3523">SHOULD CWE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f4ebb88-060d-4171-b540-1852d275f295">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_11" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a06d37bd-7374-4463-9bc6-82366a7b38fb">Dose</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58bfe47d-b072-43f9-91c4-63b3d647cd3d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_febc2913-c134-41c4-b0b6-95a07bf27119">dose​Quantity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0325d782-acc1-4937-9aad-12164111555e">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bfce6094-8209-43ab-9903-4d35ff5f5fa5">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_716c0441-71bf-47fe-9a2b-2d7face29948">PQ</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dab2c4ef-c21c-445f-91c6-3e0a115fb83b">Dose​Unit</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f9cc1bb-2d84-4c89-bba2-be39c1d0c2ef">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c41a56f3-5a4f-424b-86ff-2fd2d24bfabb">@unit</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_332dffb8-0398-4bb4-aec3-09ca3b0993c8">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1961fef1-a6df-4ec7-b127-c864165a9632">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_57967c66-251c-4483-8c4c-9407711d723f">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4db7648-3f0d-4cdb-bcae-79f695cbb187">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_82" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9aab2180-9753-4e9b-a70a-27bc3426e824">Rate</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cb9428ed-f11a-4c06-b300-44e999bbb9d9">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1aaa0c08-2e9b-419f-9872-c026a459e088">rate​Quantity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5d37829-f572-46c2-a2c6-0c3445333d01">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84a20cee-5e13-4e4e-9616-473cc68f87ac">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6aef9ca-969c-419f-9f22-99f2ee846d1d">PQ</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43caf7f2-6fd3-48db-8ee8-810481f93d8a">Rate​Unit</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_850400ac-5ef5-4369-915a-a0867b5e3c8c">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_447f2247-3601-4933-bef8-f97afb05c29b">@unit</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7a4525b-2007-49c7-be5a-f1baf9e3edee">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1743f1c9-1b7e-4eb1-b428-65f5b6155d5a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9321f1cd-2295-4f31-bb6f-d187a1b5679e">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e485ab47-9e42-4aa8-8c97-29d3297a2aa8">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a91d46d5-b6dd-4889-b0e8-09229a71771d">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_82" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ea797c3d-4c5d-419f-87e7-458e12a3b8e0">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7818b8b8-cbeb-4d18-a546-b2040d2316bf">consumable</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a3667d7-97c5-4025-92cc-554be5b77e02">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f16c6b08-2446-4ea5-a1ba-798b0b88ba30">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ffd8ff5-35a0-456a-877b-28fde9d6d259">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31ea88e3-40b5-4c0c-8423-4ddc0af0cea7">manufactured​Product</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f11f6004-6961-4489-bd00-bf4af89d22d6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d25cf72e-448e-42b9-921c-29f83c98e7c5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0aee5cff-17bd-4aa6-9707-987b8c0f45a3">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb3017b5-af4b-496a-b0a3-240f1cbf6f01">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7b4bb51-dc60-4f07-9a7c-288c0fa7b73c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0f56ba7-dec8-4d03-8a49-f9bb0ed1889f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_515d5850-f976-4819-b977-d4b21031617f">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9bd61aff-b4c7-466f-922f-c9050385f2f9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ecd64ba1-df19-4738-bea9-ff71dcad4701">MANU</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_844e4d0d-71bb-4f23-9e35-038b43ebd946">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_758b035d-9837-45c4-b87d-907b15e84569">manufactured​Material</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d5bf4fc1-d22e-4b2b-b9c8-c2f0ddabbeca">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6d99229-df22-40a8-8832-e867be6e2bc0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3efdc707-6711-445d-98a3-e613b2018fa0">Coded​Product​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a55fb48f-6d17-450a-8542-ca2effa7103b">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_abd49105-94aa-4d6b-9212-c8ffe21ca8e1">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_90c04944-74a9-41b3-8d0a-243a74028c07">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20554a68-ee43-47d6-8118-0b60636fc4b6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_760935bc-265d-421b-9021-610ae124e75f">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17a658d0-68d0-40cb-866a-8c61634a7032">Concept​Domain</para>
              <para xml:id="para_83c2c4c0-fa80-4f79-89a5-5f22551378a2">Med​Contrast​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d87c5867-6f98-4263-873b-cd15a5bc25d3">Free​Text​Product​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4270d688-2d48-403c-8140-e8027682997d">&gt;&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_afbc31a5-553d-4fb4-96d0-53e260258031">original Text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_939b15ad-7175-4fed-8e83-27f3601113fe">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_636de8ca-8924-4922-ae10-6e6e74cd5fe5">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f457611b-b8ea-4352-a566-e509b6c2d688">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.2.1" status="3" xml:id="sect_10.2.1">
        <title>Business Name Alias</title>
        <para xml:id="para_c729a885-9464-4c9d-90bf-e51056cd0991">This template defines a primary scoping business name "ProceduralMedication" and an alias "Contrast". This allows production logic to use either term, although the structure is identical.</para>
      </section>
      <section label="10.2.2" status="3" xml:id="sect_10.2.2">
        <title>text/reference and Related Narrative Block Markup</title>
        <para xml:id="para_da325513-01ae-45bc-827c-72a0bd49c847">The substanceAdministration entry SHOULD include a text/reference element, whose value attribute SHALL begin with a '#' and SHALL point to its corresponding narrative in the parent section (using the approach defined in CDA Release 2, section 4.3.5.1). See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
      </section>
      <section label="10.2.3" status="3" xml:id="sect_10.2.3">
        <title>doseQuantity</title>
        <itemizedlist>
          <listitem>
            <para xml:id="para_6d434621-ca85-4988-8005-4a6c4d180ec0">Pre-coordinated consumable: If the consumable code is a pre-coordinated unit dose (e.g., "metoprolol 25mg tablet") then doseQuantity is a unitless number that indicates the number of products given per administration (e.g., "2", meaning 2 x "metoprolol 25mg tablet").</para>
          </listitem>
          <listitem>
            <para xml:id="para_4360112a-27d1-4aee-9b8c-0412f63cace5">Not pre-coordinated consumable: If the consumable code is not pre-coordinated (e.g., is simply "metoprolol"), then doseQuantity must represent a physical quantity with @unit, e.g., "25" and "mg", specifying the amount of product given per administration.</para>
          </listitem>
        </itemizedlist>
      </section>
      <example label="10.2-1" xml:id="example_10.2-1">
        <title>Procedural Medication activity example</title>
        <programlisting><![CDATA[<substanceAdministration classCode="SBADM" moodCode="EVN">
    <templateId root="1.2.840.10008.9.13"/>
    <id root="cdbd33f0-6cde-11db-9fe1-0800200c9a66"/>
    <text>
        <reference value="#med1"/>
    </text>
    <statusCode code="completed"/>
    <routeCode code="47625008" codeSystem="2.16.840.1.113883.6.96"
        codeSystemName="SNOMED CT" displayName="intravenous route"/>
    <doseQuantity value="100" unit="ml"/>
    <consumable>
        <manufacturedProduct classCode="MANU">
            <templateId root="2.16.840.1.113883.10.20.22.4.23"/>
            <id/>
            <manufacturedMaterial>
                <code code="412372002"
                        codeSystem="2.16.840.1.113883.6.96"
                        codeSystemName="SNOMED CT"
                        displayName="Meglumine Diatrizoate">
                    <originalText>
                        <reference value="#manmat1"/>
                    </originalText>
                    <translation code="3320"
                        codeSystem="2.16.840.1.113883.6.88" codeSystemName="RxNorm"
                        displayName="Diatrizoate Meglumine"/>
                </code>
            </manufacturedMaterial>
        </manufacturedProduct>
    </consumable>
</substanceAdministration>]]></programlisting>
      </example>
    </section>
    <section label="10.3" status="2" xml:id="sect_10.3">
      <title>observationMedia</title>
      
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b89b9c05-00c7-4b51-ad55-f99bd28e00eb">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_34912959-74e4-4cd9-b386-912596bee9fa">1.3.6.1.4.1.19376.​1.4.1.4.7</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee28ead2-6ae1-4a6a-b8cc-9739eb452d21">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_be5a3180-4ab8-4ad6-b06c-388623416595">observation​Media Entry</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5fd8054-2315-4c75-ac95-8e3db31d37c8">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_409ff3d4-2229-43db-92b4-ad702fdc5500">2011-07</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb164793-03bd-444b-b5aa-3ec547d60a3b">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da5f458b-33c6-4ed4-84d1-266d441eb595">IHECIRC-TI</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_722ee6ef-98b2-4588-8ae6-6143e1250e40">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e49ea8c-59dc-48ee-a05b-94ecdc0328dc">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f4f02a0-dedd-424e-9598-33773c2e3035">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7d89016-266c-464b-9342-30ec7522f7c0">The observationMedia Entry provides an in-line graphic depiction of the section findings. It is referenced by a &lt;renderMultiMedia&gt; element in the section text. Typical uses are for graphic representation of findings (e.g., arterial tree diagrams) or in-line representations of key images.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d8acd70-25b7-49a9-8db4-4de3d8ac9e15">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3b4c0a1-d921-4d42-82a0-d928997c828e">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4a84814a-e2d2-4958-a6da-ec11608a8b34">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e510b451-f83a-4303-9f69-32eff9563a21">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7dc13d5-9e02-48cf-8c8d-282a41af14fd">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_deae44e0-b6ab-43b2-8f5c-36179e594bf3">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c87104da-a688-491a-9f20-cbdba8a24e5d">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6473e5a-a990-40cd-9742-6f5decb22447">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1a47d59-2aa5-4aae-aa7a-a0595ca890bd">From IHE Cardiac Imaging Report Content Profile Supplement for Trial Implementation</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_1cf16fcb-c301-42e5-b16c-660a528db97d"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9d675423-88d7-4d4f-8cf0-cba821b6251d">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4beea218-45dc-4a6d-9f1f-0537a94f951e">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f8cdfdd1-4c28-416c-819f-1435272692ea">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1237a42f-a910-475e-8209-464e2a451810">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_31f8d29b-937e-49e9-a2bd-a8cded204382">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_277277f0-ef1b-4f17-9f1b-fa82419d16ea">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_02a3d4c9-767c-4314-8e62-ccae8b76d864">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_31d4dc14-7a3d-44ca-9709-b5a3289ea29d">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_fb4bf513-2ef6-4c61-8341-74c9e900e89b">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_09365bc5-f957-4aef-b3b4-8f3d12e0a01e">
                <emphasis role="bold">Graphic[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d98da1eb-bf5f-4f23-8fad-c25d48724bc4">observation​Media</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_925249e6-6fc5-426f-aac4-af604bede561">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cabfe59e-3328-4c2b-92b0-c2779efbce4e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_db8aa4fc-c963-4cb0-b049-6d42653bf762">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06127a74-169b-4c33-a1f1-2b58553fa6f0">class​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_166c0e77-6181-41cd-a4b5-9f0624214490">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6ffd16f7-6626-4084-bbbb-ba8a01fd9fc4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb083b0a-4f2d-4c32-bbdb-9d75fac2b3b4">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b09136c8-f1ef-4a64-9aee-bbea8c52a84e">OBS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e7b206ac-05c4-4668-96d8-220158ddd3fd">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5762f104-09e2-4a1b-8877-ac084b51f8c1">mood​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6fadb3b6-9c44-484a-b3ab-7aa22cf966f7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_af9076f1-f009-4e89-a0fa-9a9c0259b202">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff1cd8a5-8814-445a-bc06-082c728702b8">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d9f41a9b-4739-44d4-9e1c-062052d3dde7">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f314c686-18da-4615-b8ed-6ce93f82bdcf">
                <emphasis role="bold">*</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62d66ba1-55a8-4273-a6ad-2bbbb16302a6">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b537dae9-cf7a-4aae-8848-cc43dbf1f78d">@ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5f1fdc6-a8e2-4d1f-b0ad-cb0a3a048917">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e89f5f43-3fc2-480d-be10-51f28be4a000">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a3eadfa7-4599-4f41-8184-2e2277125c33">XML ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0dbea0e-ee14-475e-a0df-760aa0f272af">[See <xref linkend="sect_5.3.4" xrefstyle="select: labelnumber title"/>]</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3959f388-2f2e-42e0-b825-a28619445b36">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb92cb08-82ad-41f3-8240-b9f0544fae83">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_036dee13-c0b1-46ec-8ce7-fc879bd4bebd">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e820b8dc-c956-4c11-9f50-5805d962585b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_832a0922-5f53-48b6-9fb5-a167faab66f8">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e538eccf-0176-440b-817a-1adf6ff9e6b9">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53a3f632-36ec-455a-92f8-25de4037ab51">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1d054c8e-371b-44fb-bd15-4d48113f01b3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_13f76312-3cf4-4949-9bc7-af72d4396ad1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5db8e458-e6b2-4d57-91a5-161d253d4702">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e4709cb-5bbe-4883-b173-e50f44c9e25b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0cc8c8b3-b2f2-4dad-b675-6866bc79174d">1.3.6.1.4.1.19376.​1.4.1.4.7</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_132b2c07-11ba-43e3-8e74-4ae5822b2568">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_28e63480-6858-4ba8-a115-a883fde37fc4">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dcacc2c8-31d2-4947-a413-34b720c70cdf">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec42b7c7-ca92-43f9-b656-d35af0dcd710">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd6c911c-6614-4bc3-b831-087bf23c794f">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0e9c0e9-0aa1-4510-a838-ad2d598ce6ea">Image</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d459c6c-6480-458f-8faa-57e8ee399d94">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f37426af-429f-406b-9cfd-433c112d9ae3">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0ca8bc4-a67c-4da1-bfb6-90ad51b6cc5e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0481ecd0-e48c-452d-b36a-4a18eeb99c7b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7311ff4-6648-4bf5-b2e0-0d2ad882cb3c">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aeefc04b-e563-4167-afca-aa35bdc3f2c5">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b42adc5-7d94-4ba0-9cd4-a2a041d4693e">@representation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6ea875cb-e580-4eb5-8b71-96fe0eccb534">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_342ed533-aa5f-47d0-b218-6657b45b24ba">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cafbfef-ffcf-4bf0-b854-c8f41f664a2b">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d43ccc8e-a010-418d-a6f8-801511c38b4b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2308c0a-ec86-4027-b93a-a203ab607d3f">B64</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb4c1182-d31f-419a-ac45-4a4cce71af24">Media​Type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b5eb5360-9580-4fa6-a5ae-0040f18e08f1">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b2da4e8-2d8e-44a4-8798-9ec3ee12d5e5">@mediaType</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bbd153f3-37e5-49f2-9eb6-3a0a0b979f65">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f5eb710-2b9b-4607-842c-32aebfbf182f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40ff211a-e57a-40fb-8be1-aac436c8c841">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_330f9753-dca0-4db2-a920-eb82d753880c">SHALL</para>
              <para xml:id="para_3a8f77c2-caac-4efb-b57b-8dbc3efe4de1">CNE</para>
              <para xml:id="para_9d9ee132-cee7-4723-a1a6-170f9eaa1f69">STATIC</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_094a1af3-a6d6-4e78-9899-5bbf00c3ea2b">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.3" xrefstyle="select: title"/> 2.16.840.1.113883.11.14839</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8cdf3d1e-109f-4bce-ad97-c84ed20a883c">Image​URI</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_59f05f14-3b27-48f2-aec5-7f4c77eca618">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f94ce764-726b-4fd4-914e-77711f85d876">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4fa2fb57-526c-4783-a8af-7c65ad30c93b">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_49cb5039-a033-441e-a732-9e3fc656e886">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e61c8a17-9668-44c9-8172-e04d23b76cfb">TEL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.3.1" status="3" xml:id="sect_10.3.1">
        <title>observationMedia/@ID and Related Narrative Block Markup</title>
        <para xml:id="para_ff8968f1-3824-48e8-97d4-6dd333b6b6ad">The ObservationMedia entry SHALL include an XML ID attribute (not to be confused with the id element of the act class) used as a target of a&lt;renderMultiMedia&gt; element in the section/text narrative block of the parent section. See <xref linkend="sect_9.1.1.3" xrefstyle="select: label"/>.</para>
      </section>
      <section label="10.3.2" status="3" xml:id="sect_10.3.2">
        <title>value and Reference</title>
        <para xml:id="para_efe1eae5-8a8c-4d5a-9374-402ba19b931b">The value of type ED SHALL contain an in-line encoding of a graphic using base64. The &lt;reference&gt; element, if present, SHALL reference a URI for the same image as included in-line.</para>
      </section>
        <example label="10.3-1" xml:id="example10.3-1">
          <title>Observation Media activity example</title>
          <programlisting><![CDATA[<observationMedia classCode="SBADM" moodCode="EVN" ID="obsMedia-1">
    <templateId root="1.3.6.1.4.1.19376.1.4.1.4.7"/>
    <id root="1.2.840.19432234.2342342.23232232"/>
    <value representation="B64" mediaType="image/jpeg">
        Bgd3fsET4g...
    </value>
</observationMedia>]]></programlisting>
      </example>

    </section>
    <section label="10.4" status="2" xml:id="sect_10.4">
      <title>Procedure Technique</title>
      
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e0340ae-1b0b-4508-801b-5425617f2b39">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88cc9d31-1acb-4029-b548-a9319aff9718">1.2.840.10008.9.14</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a67f482-f4ca-4a33-8c0d-36b677a5ff42">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3358ab16-24e8-4b5f-91e8-ff884bf14b41">Procedure Technique</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_996f2dd0-c327-446b-a934-7126e9b30b8d">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19eab1c8-3af6-4489-9418-98e0985ee7a2">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0180005c-abb0-408f-93be-3af9fccf1322">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df91d8ae-6655-4747-9b00-0d3c73c1f178">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d48db791-7dce-4a17-8156-54d2fcde11b5">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b2cd058-2389-421d-add5-f516c30e913f">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfe6f29f-fc32-4223-bf3b-ae73eccda194">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f49bda34-0c56-4484-984f-7d17e84107c9">The Procedure Technique entry allows the encoding of various parameters of the image acquisition. Other details may be found in other entries (e.g., procedural medication).</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_621fce8b-da4e-4025-b8a3-d28b7eab9e0a">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6bc90630-9213-46bc-bdd6-f8d7996c90ab">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e086095-5d52-4780-9b84-603f4fc4f71e">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_459a044c-cb64-4f97-98c7-26b006154923">Included in <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/> and <xref linkend="sect_9.4" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d76286c-db48-4cc3-a90f-e47c8c7e62af">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2c2ae054-e465-4d12-825f-27f9aa74d7f0">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00ced364-df41-4046-a2a4-48faed1933ef">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c4f53c0a-7496-47c6-9c3d-3703e2f8ced5">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd701ce0-6927-4696-9153-747d491c3bd1">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed086374-71fa-446c-9275-a5dc903e3c01">DICOM-20150324: Initial version</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_a23914fa-9886-4b8c-826f-6f35d77b2cd9"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1878261a-f392-4538-a247-3661f93c7ec7">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b739c529-5a24-43c7-abbc-a616846de4b8">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2e95a672-4cf6-42ba-b2d2-5145ffed92f1">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_858cee19-db08-4450-97e8-088d24b93947">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_6967255c-92ab-4d5c-89dd-0c0055dfcd32">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_6a13fba6-6824-40df-b117-9fd8c1688739">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b600e968-137a-4500-ac85-0d14e20dc254">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_ed59bbec-a621-498f-a94e-b2a0750e444d">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_9d99ee0c-b0fb-4cc5-84ed-aad3147f60a9">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8ee66d27-812c-4775-9959-dc38495c4386">
                <emphasis role="bold">Procedure​Technique</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_13423859-90cf-4156-95cf-9841e94c74cc">procedure</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_327d16c9-c570-481e-b762-dd8c85c90708">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b10c026d-1161-4499-b7bf-74729b3923e7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9816784f-2667-4778-bbc2-2e44c5fb678f">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3214e701-c037-42f7-bdae-449043949a12">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0939b78a-a4f7-4970-864f-f211c3fd8b15">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c1718e1-d9db-4f48-bb50-23fe10da1f4d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_891c3a46-cc1b-47e9-b8a0-e25599e217a7">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cde2732b-02b3-4695-b82b-c4d963a646ef">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46dca261-d682-479d-85ba-ca4104fbd0ba">PROC</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29edd0c0-7144-4be5-8c01-8f1f46464519">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c306142-d06e-4e27-83fd-10eee1d824aa">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_520dccb4-c19b-4888-a662-d8e1f15a4b3c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b0b6cc0-f6e5-44d4-8ef6-63f4f23dbfb4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f8dbf9a3-60b7-4e6c-874e-a51061b7e2ab">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf4d9b2e-b6af-42c8-9705-eb8783918445">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a4dbc3f-d4ef-4d52-ba4d-55273499da44">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30149884-e1d9-40fc-ba2b-c7c891a089fd">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b52e61ac-6e80-4e56-9fab-626a9f40fe47">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9bb53e4-232d-4fab-b052-a0c0f18e7bed">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b8cb4b25-6a7f-4287-ab05-16ddd2228369">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64d08e22-f45e-4eba-ac61-3ccd98985087">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58da3ac3-7f47-4a7b-8ab9-55b5b9b50334">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_909e848a-68d5-4b04-ae51-08b513e6b951">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c3f04c63-e73f-4b8e-b3c3-ced78e44c173">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b35bee3-cb06-4bd2-9492-6695a7c87b76">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c22ff9dc-9a9f-4cd2-8e22-f241717be1bb">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_65bce6be-0da9-4127-a91e-4ebd54103070">1.2.840.10008.9.14</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a16422fd-78f1-462c-ace7-b783d5feb284">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4dd15181-9b46-4f8b-8bcf-cdd05946a600">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd2ca42d-6cbf-4911-9222-5b270e5a038a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fdbacb2-ad60-4e4f-8f93-9730e6907cb3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39830739-575a-492b-ac27-6ffcd7709d9e">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_02a1b960-091e-41ca-9f70-bdcb7c176d1d">Procedure​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b1d4f51d-4cb0-4a99-afb8-2313031d5fa4">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f25b801-fa13-491e-9243-b7d3011fddc4">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_340530de-da25-479a-bdbf-d0d638f0320c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b785a7e-8a1f-4173-b8d0-f500f86881bd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40705b17-2292-4e26-83cc-614b82cac79a">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f790734d-bc43-4290-906c-260b3359e5de">Concept​Domain ProcedureCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_299eaf4c-3be6-4a41-8871-58d22d6d4607">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f24311d6-64d1-47bb-870c-6db34786b92c">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50744c71-ad70-4d68-8777-84bbb29c9f92">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89365eee-9e3b-4df0-ba99-ebc10965434b">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b5a0aff-cafc-41de-a368-0c581f116d00">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_11b60d19-17b0-4a4b-a5b2-8f81b4dc048f">Ref</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b8a3a2af-f906-4977-b481-40feb38e7ed0">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6ff934a-6d3a-4802-a2f8-1dcf9c334f8b">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c08ea5d-9799-464b-aff5-32457267bb0b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3533cb61-41fc-4c83-9cca-a50f67268944">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e02e2d82-4a6d-4dbc-909d-4b36ae36cf3d">URL (XML IDREF)</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0bbf6b2e-c963-4b81-a3b7-8a8247def50f">#
                <emphasis role="italic">content​Ref</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_873b022b-2238-42ae-b23c-44b016b5f6f8">Effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c50b7ce-df5f-4b21-9fd2-ab3e33b8cb34">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_33ae3825-26a7-4769-8d89-344db4090e20">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a51aec7c-5b5f-456f-a981-90419cd4c1bb">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f075f770-92ff-42db-9f06-d348e2e188a4">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0e6ec01-6859-4f01-ae11-5454f62284c5">IVL &lt;TS&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bac7309b-a206-44f9-ab2c-10e8ab5d082c">Modality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac428635-ec2f-4dcb-a47a-f648ae5958c8">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac8d3f74-f3fb-4dbe-a110-ac9f3288246d">method​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6a3510d0-0411-4f94-bf7d-070995ef0520">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e38866e-bdd2-47b1-b74b-eee86d6bf94a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81d4bf9a-2acc-4f49-a06c-749708d6e115">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87551eb1-2c6c-4bfd-99ec-0438332d38c3">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_481529b9-6611-4775-acd9-bc1152390881">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_29" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74c0d2af-32a6-4e39-a4f0-89278588d86c">Method​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_84db0462-619d-49b3-9ac7-77724f482b43">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c92893bd-f5f2-42f4-96c9-2c5c3bc9393b">method​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca06bff7-281a-4c8e-9a9e-b3a45969ee06">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_707f5fb7-8211-4610-8ec0-50ab1afc3c23">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4c19a56-d6da-4545-a4be-595ac5ac6d90">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6599f89-e85e-4083-a7e0-da5c26b3cdaf">Concept​Domain ImagingTechnique</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_783133ab-7ca7-417d-aa11-3b01c7cf81c0">Target​Site</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb0cd1a7-8119-4506-be7e-ee2a142905fd">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_575845c8-bd4d-4cfd-b7eb-1bc9ea73aee3">target​Site​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d674dfd9-cca4-45d1-92b7-9407ccb89af1">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6c6e3d6-1331-4cca-858e-f47bb5402d81">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_baed2ba9-e79c-441d-8735-b7e543c1773d">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b954d457-4379-422b-a31f-fcf4bfd464f0">Concept​Domain TargetSite</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19feb28f-366d-481c-98a7-c6b41ecad43b">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d320397-1727-4060-a3b1-153153275123">qualifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7a342d0-f7a6-4b59-b8c1-8d06d2478e1b">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b13ee658-f16d-4973-a3ca-e1e60254bd3e">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd9c1e74-3878-439a-afa3-5d61b21cc8fa">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20535824-5056-475b-8dd6-b242d139452f">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_324228e0-0994-4920-8602-be89b7e9ab11">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_191660f1-0ef4-43a2-9294-70fda2b83d5f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a20b0d0a-d84b-483f-9eab-02b79738c512">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_57917035-0fc5-4b0a-9c99-e184464af300">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce6bd0e6-0b1d-4ec8-9b5a-8c32a7742f66">(272741003, SNOMED CT, "laterality")</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17e4dbe9-b603-445a-886e-7251920a629b">Laterality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_55013d7c-b8fd-45cc-8fec-5bfd987f7b83">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da02135b-1b73-4b30-a3b5-0758dc717ad1">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c2d6ecd-e031-4322-86b5-9ce6edf9581d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c663e816-1715-4740-82da-48c66c1ec2e8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_52215917-cd2f-4b36-8996-d0538b291c30">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6ec5b09b-446e-4f53-9337-2aa2e9cf4f9a">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_37fce0fb-017f-4e32-82b7-ca74541d7098">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d372d72-a2f0-4371-b5a5-5b06b27f0392">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_021cd346-89ac-41b4-bb40-7ee502d38cfa">participation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3dc2f4f8-66ad-48a8-bd66-e5c8b6b3214b">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b67b8d2b-a9cf-431a-bec1-b9a12593a802">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d520c76-a99f-4abd-bec7-289c705c745c">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5930a27-cf10-4321-8c08-27594faf39d6">@typecode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58ebf57a-d13d-4d70-94ff-5c8726067154">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd256da1-bd10-4563-9e8f-1d04a9aaab88">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2699fd9f-84ad-44c5-9670-d1146d3cdf13">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c543c981-7bb5-4625-a699-acb23458b15d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c55a61b9-7397-4daa-975d-64689ea87be9">LOC</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f2e768e-1c21-49a2-9f52-58f94a47abd4">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e67a3b7b-4620-41f0-b498-05cbbec53d68">participant​Role</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ccec3825-fd9f-470e-9df4-2db73628f99b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8416a879-3b7a-43a9-84b7-51dcf1444b20">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5150132c-ce95-4b84-958b-8f3c1eae084a">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0aab571d-c458-4e47-ab92-840f714eb39d">class​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a91c22af-b2e7-42ed-89d4-0e775680a432">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c82c47aa-b290-44bf-aa0c-a152c3816ff7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_335bad91-89a6-4e18-9edb-883edc3adb39">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1780d900-53f5-4e8b-8e48-4f59e0ae7618">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b50a32c-8d08-4934-be6b-265b39b34df9">SDLOC</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_59e6d825-7868-4012-9ba0-0868f7a2399d">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4a585b4-77ec-4381-8596-fbf29a9bfd9e">scoping​Entity</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a1e8d20-2148-478c-ab7d-7ce4392b06b0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a6889763-2a65-495e-bd0b-3527800a7342">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_071f5fdc-3d20-455d-baac-4dbe12d80e8e">Provider​Organization</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ae8c4a2d-c88c-4f9b-af86-94f0ffdf9f99">&gt;&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_96e8d8af-53db-4809-9cac-17cc39e23573">desc</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b0b561a-b549-4071-950f-7b1b5689573d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d7c5180-cf2c-4694-b4c8-5d1a801e41c1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9dd19956-89ac-4160-ba16-fba29c8ddf74">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.4.1" status="3" xml:id="sect_10.4.1">
      <title>id</title>
      <para xml:id="para_257f782a-09d1-48e6-8518-d6231b13a642">procedure/id does not correspond to any DICOM UID, but is an arbitrary identifier for this entry.</para>
      </section>
      <section label="10.4.2" status="3" xml:id="sect_10.4.2">
      <title>code</title>
      <para xml:id="para_7faa4343-ce87-4dfa-af13-a05decdbfbb4">When invoked from the (current) <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>, procedure/code SHALL be identical to documentationOf/serviceEvent/code in the CDA header.</para>
      </section>
      <section label="10.4.3" status="3" xml:id="sect_10.4.3">
        <title>text/reference and Related Narrative Block Markup</title>
        <para xml:id="para_633f700c-2234-48e4-8ed7-316b1e4eb1cd">The Procedure entry SHOULD include a text/reference element, whose value attribute SHALL begin with a '#' and SHALL point to its corresponding narrative in the parent section (using the approach defined in CDA Release 2, section 4.3.5.1). See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
      </section>
      <section label="10.4.4" status="3" xml:id="sect_10.4.4">
        <title>methodCode - Modality</title>
        <para xml:id="para_88c5ef06-727e-4cbd-84fb-499de88f1ffc">When invoked from the (current) <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>, procedure/methodCode used for modality SHALL be identical to documentationOf/serviceEvent/code/translation used for modality in the CDA header (see <xref linkend="sect_8.2.4.1" xrefstyle="select: label"/>).</para>
      </section>
      <section label="10.4.5" status="3" xml:id="sect_10.4.5">
        <title>methodCode - Other Parameters</title>
        <para xml:id="para_6b40ee58-1ce5-48a8-b5e7-fe4fa2365a75">method​Code may be used to encode study type, contrast use, challenge, views, positioning (<olink targetdoc="PS3.16" targetptr="sect_CID_91" xrefstyle="select: labelnumber quotedtitle"/>, <olink targetdoc="PS3.16" targetptr="sect_CID_92" xrefstyle="select: labelnumber quotedtitle"/>, <olink targetdoc="PS3.16" targetptr="sect_CID_93" xrefstyle="select: labelnumber quotedtitle"/>, <olink targetdoc="PS3.16" targetptr="sect_CID_94" xrefstyle="select: labelnumber quotedtitle"/>), etc.</para>
      </section>
      <section label="10.4.6" status="3" xml:id="sect_10.4.6">
        <title>targetSiteCode and Laterality</title>
        <para xml:id="para_91b25919-807c-44d9-b4d0-5d5591332b0f">procedure/targetSiteCode may be used to encode the specific anatomic focus, and is not necessarily identical to documentationOf/serviceEvent/code/translation used for anatomic region in the CDA header. This may be derived from
          <emphasis role="italic">Body Part Examined (0018,0015)</emphasis>, as mapped to SNOMED codes in <olink targetdoc="PS3.16" targetptr="chapter_L" xrefstyle="template:Annex %n “%t” in PS3.16"/>, or from
          <emphasis role="italic">Anatomic Region Sequence (0008,2218)</emphasis>.</para>
        <para xml:id="para_fc096a1d-7be0-48e3-a90b-a1eaaf6cee01">COND: The qualifier element for laterality SHALL be present if the targetSiteCode represents a paired body part and laterality is not pre-coordinated in the targetSiteCode.</para>
      </section>
      <section label="10.4.7" status="3" xml:id="sect_10.4.7">
        <title>participation - Location</title>
        <para xml:id="para_deb105f7-786e-481a-a003-56b2c4cb07e5">COND: If this template is invoked from the Comparison Study section, procedure/participation MAY be used to identify the location (provider organization) at which the Comparison Study was performed.</para>
      </section>
      <example label="10.4-1" xml:id="example_10.4-1">
        <title>Procedure Technique template example</title>
        <programlisting><![CDATA[<procedure moodCode="EVN" classCode="PROC">
    <templateId root="1.2.840.10008.9.14"/>
    <id root="1.2.840.6544.33.9100653988998717.997527582345600170"/>
    <code code="RPID465"
        displayName="MR NECK ANGIOGRAPHY"
        codeSystem="2.16.840.1.113883.6.256"
        codeSystemName="RadLex"/>
    <text><reference value="#proc"/></text>
    <effectiveTime value="20140913222400"/>
    <methodCode code="MR"
        displayName="Magnetic Resonance"
        codeSystem="1.2.840.10008.2.16.4" codeSystemName="DCM"/>
    <targetSiteCode code="45048000"
        codeSystem="2.16.840.1.113883.6.96" codeSystemName="SNOMED CT"
        displayName="Neck (structure)">
    </targetSiteCode>
</procedure>]]></programlisting>
      </example>
    </section>
    <section label="10.5" status="2" xml:id="sect_10.5">
      <title>Quantity Measurement</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a28a3854-99db-4391-90a7-93c6462a3971">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a572c411-6534-48f6-a0d6-a294fa0aea81">2.16.840.1.113883.​10.20.6.2.14</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1a60bad8-71a0-4331-97b4-85d8c72d8abb">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0c7f7f8-4649-4940-b352-d555d646fe7a">Quantity Measurement</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1002a53c-a120-4dfb-99c7-ab99c22d9da2">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_103d5e7e-ec7e-496a-b237-07b485a9fb9f">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb6de3ef-2f7a-4f0c-b6e2-5d2df6cbee44">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d9db4f5-00c2-45b8-b03c-8e3bc2a09be9">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40067776-8e22-44ac-a0ac-3b0775e46958">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0518ed60-3679-477a-95c5-0698bcb777a7">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15906b52-3427-4dcd-83ee-1ae6960c3f0b">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73e18c1f-9112-4a14-b63b-89888d52b313">A Quantity Measurement records quantitative measurements such as linear, area, volume, and numeric measurements. If based on image data, a reference to the image may be present.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_21cdfe7e-3a88-447e-ba4e-ac11fe6e3739">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_34ad9f0c-c705-42b9-a01f-5ba43bf574a6">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1411c3a6-87f0-4365-b7ff-1c94a90a2d02">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88f3c65a-afa7-41a6-aa07-0e241f98f7b0">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6a15deaf-8ed0-4b07-bc44-5b27c3a5c00a">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a86d51f-c134-4001-8e40-c584cab9b91a">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e5a21b0-0036-4bf5-aa6a-a6e4c3fc71ef">open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_396ab2b8-63a7-4f19-bfe5-2ecbdd3c6b27">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d41564b-59ad-467e-8d7d-a956f08b40b8">DICOM-20150324: Initial publication, derived from template originally published in DIR r1-2009, revised in Consolidated CDA r1-2011 as 2.16.840.1.113883.10.20.6.2.14. This derivation includes Units of Measure specified with DICOM value set for UCUM (<olink targetdoc="PS3.16" targetptr="sect_CID_82" xrefstyle="select: labelnumber quotedtitle"/>), equivalent to C-CDA specified value set (UCUM Units of Measure (case sensitive) 2.16.840.1.113883.11.12839); addition of optional interpretationCode and actionable priority</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_9f49d929-7d6c-4fe4-bda1-8dcd691776b5"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_af866694-12e4-4ba0-ae1d-ea39123318f6">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b95838d4-a144-4201-86b6-3664f69c51af">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_60762b46-fc28-4eed-9f34-2989820af421">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_95d44ab7-a086-4bcb-a994-b399d7431e5a">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_1da4ff01-0bfd-4309-909b-f105d6345527">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4eb78d91-39aa-4de0-a618-a8010d3cd206">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b65a908a-a4de-47b6-9714-e034209b81a5">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_51c6cda9-223f-4bf4-a575-7d1548b0468d">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4cfcdd5f-d27d-40c3-9d70-0d4d4a0b1c9c">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43405f92-f569-4091-aa9a-21e32974d5e1">
                <emphasis role="bold">Quantity​Measurement[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c17938b-551b-4460-aca6-76295983fad4">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f23bcb7-ff31-42c0-9465-c1ab0cf7a5d0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_585b8433-a9fa-45c0-bf54-00e39f6b154c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06ef6b0f-fc7c-42cf-96a4-ac8fabe2d5a9">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c55a73dd-824f-4ad7-8f41-00d4c1abb066">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_665267bd-9e0d-43cb-b5bd-12345d8b10d4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e4401f9-53eb-46a0-8bf4-cb7020a262b3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d524b3a4-e19a-42d1-ba40-41f27d653ccc">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd57bd92-33bf-41d1-b924-7f75617a8d98">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00e66b69-5e81-4c4b-a55f-3076ad214e96">OBS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3eb22e35-9068-4b1c-9d98-29bf57af7b7a">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_beb639f6-cf35-4cf7-86e5-ddccf08f7847">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3619cee3-2308-424e-9d3f-4a37c9579d20">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f302978e-e9e3-4c9e-8162-5848f8bca961">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fca5cd8b-1095-4b61-99d6-476c7a527a35">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6fbaee3f-26ae-4ed2-879d-8093b2b34b7f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3ba3bf41-6d05-4707-8df6-e42df5e6475b">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f2a1e20-bee3-437b-93db-5688478626cf">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4513a3b-9184-4c71-b9cb-31515c585f41">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19f0876a-f38c-4aa9-9a8e-6d91e5b0db36">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5850ab38-aee8-42ef-ae45-c264439e228b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3503cdb-2755-41ab-8412-504f58fc965b">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41e81a7a-986e-4733-9cbf-b18610b0a5aa">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_14a37377-edcc-4b79-ace9-f94eda26f7c2">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0619e78-1082-4de7-9719-a9e3e2240614">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb75cbfe-e066-4417-a2cc-92893f9bdc0f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff351cc7-54ff-4f89-b51e-4d2806bf3488">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1b03c226-751e-404c-a6d5-8e1a7eeec63f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_435e08f2-8f73-49b9-a8cb-0944d7a3cce4">2.16.840.1.113883.​10.20.6.2.14</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9987a3c-daa3-42a5-93eb-ba211356d768">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_815e9042-6972-4a03-b68b-8514d19084da">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4d7a5cd-5ef1-4f71-861b-9167609f0794">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9bc32f5f-d2f9-4193-9879-b3b4c9919c91">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d5bd0c8-0580-4069-990e-f9318b47125a">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df9de9ad-8993-41f5-ab32-80ceff66232c">Measurement​Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88ba9bcc-d6e9-4568-9601-f3ffc2e09bdf">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67e47894-9052-45e2-ac9f-6aaec99a8f8f">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd4cdd92-1641-4be1-98b9-e6405cd982c5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9a96c047-6a52-49fb-886f-2e45a635e03b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_494584dd-42e0-4070-ada6-2f2b62a16b01">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1ec083e-5e7c-4fd7-aed5-d9d60d20d288">Concept​Domain ObservationType</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23c9996e-516c-41ec-8eea-109acd0743a9">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a77ac45-5f02-4feb-bd39-91e449dc0e7f">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ad98fae-8d2d-4f51-bb29-db0977b8ab8f">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e132961a-3610-48d4-9c19-19abbd5a61c7">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc433f2d-4ae9-4fc8-9489-ac037b454d3e">Ref</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5b59507-a381-4c1c-86d2-d623103ef350">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98fa3bec-af03-4e15-b912-a8f9f37c414c">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f728e781-823f-4f24-bba0-f9c6573dd520">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f9b6a0c-f95a-414e-abe5-5b5a9cd46280">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70dc67f4-6831-48a1-898c-93667d33b219">URL (XML IDREF)</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_278ba8a6-e96a-44f9-b67d-0fcb2c066cc4">#
                <emphasis role="italic">content​Ref</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c2294c2-b940-46ef-9096-85e107848c2d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_613497d9-2a0d-4369-9b07-ae2c6875e8d8">status​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_86fe09b9-84d1-4fe3-8d9a-414629b40c2e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8cc5a3fe-1547-4997-92a2-09bab84a8975">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec10b55b-7d46-4f16-971d-6c949775b9d1">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07482467-f4fa-4136-8d70-5f761a7d636a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d49d57fe-c157-4c15-94de-1c364a3d3883">COMPLETED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a7e7dad-589b-48fb-9d3f-3fd751abff8f">Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d33b5ecf-9917-4f06-ad8e-ff0bdb79a3eb">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d17faae-7b02-4851-bde1-1ee724fc337b">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c690ac27-729c-4576-a4b9-d507ddc8824e">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7c4e90b1-dc4c-4acf-9afa-d436e204d48e">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6d2fc79-8268-493c-823a-a997f8900acf">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed1d30d5-83f8-46bd-8f23-bcdce1255f8f">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_671bd53c-1ed4-4f7d-a243-7fc50545b816">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1bc72fc5-645d-4907-b232-b6cdd72d9e46">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d5c4359-7a22-43f9-8bd7-f38ab5fc1a87">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfab70da-9457-4cf6-a9d1-5cbe41b43948">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d45522c-ce9d-40ca-86e4-0913c96412dc">@xsi:type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b28de12-bc00-46df-9603-2a0b744ba7f3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5803f203-77f3-4ec8-be46-31baf144d1a3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d5f1335-0977-40f2-99c6-20db12f3499c">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d5b93df2-b2b4-4b2b-9c6a-be73f69e1c74">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6fa33452-f1eb-40c3-a56f-05ffd843d59c">PQ</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0227c0c1-eb64-4fb5-b6d2-d345e3f2c149">Measurement​Value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f2af8c0-645b-49ce-8784-92eea1670ec7">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_518c8244-0853-49cf-bc3a-83614accc148">@value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5b38e2c-06f7-4c9a-9576-0b78ac077269">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd02d9e2-bdac-475a-8540-19916864f4a2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8642388-136a-4b07-8a11-ff864128d052">REAL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a586e9b5-56ce-47e2-a9e0-04b199c33458">Measurement​Units</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6ee66310-e0ff-408f-9d49-976e6e9165a1">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29563721-8460-4233-97ee-87b0c183dae8">@unit</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c1787e1e-b8fe-4fdf-9f32-d7ed7af749b7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c268ca43-f4f8-491a-9d92-95172caec6ad">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf5b7303-b881-472d-8f03-1fc8021cf459">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b01aed2-ce4d-4da2-8e50-122b92c36fa9">SHALL</para>
              <para xml:id="para_52555c4a-1d4c-4479-8bd6-bb1a064f0b25">CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50073e83-a59e-4d54-82f0-c1cb6b0e0af3">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_82" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba412447-faa8-4600-847c-5ddeda9cf97c">Interpretation​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a599582d-428e-453a-b0aa-aa7328194d61">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ca78cba-d977-4b88-a10a-3ddb6c977c6e">interpretation​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b2e096e-eb48-468b-be99-4c109fe5dd8a">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6fb73b21-84d3-4a0f-bd2f-710355ae1cf9">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a759add1-bcfd-4f06-afae-5324f58cc273">CE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a4d9ee00-4cda-43e4-af2b-2a9799f71e63">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fdf45a23-0772-4ad6-a0c7-af4e3f712008">ValueSet <olink targetdoc="PS3.16" targetptr="sect_N.1.5" xrefstyle="select: title"/> 2.16.840.1.113883.11.78</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c6b8d35-212d-40a2-8e5b-5af7a798ed91">Actionable​Priority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9599bf7b-b0a7-422b-a2cf-b6aa6f4859f7">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f3f037f3-16f7-40dc-85d0-67e459292aa2">translation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_af271667-d9f7-49f5-8f76-6b378e7b0d7a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9a325cc-8ae8-4d93-9bff-444da51bf551">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29865896-b913-4860-8e35-d6f0d19444b6">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b043b3e-fd50-4490-8b1b-14bff8e93e24">MAY CWE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_56044f32-3187-4238-83db-05f30487fb81">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_7035" xrefstyle="select: labelnumber quotedtitle"/>
</para>
              <para xml:id="para_89364472-1279-4ed4-8095-74afd9535399">[See <xref linkend="sect_10.1.3" xrefstyle="select: labelnumber title"/>]</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4bd1a2c-f6eb-4656-a3da-0592326118bb">Target​Site</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de94fcac-4bfd-4841-a3c1-1ea42bdebc44">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aaca2581-5acc-482e-9079-74be99aa42ef">target​Site​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a49856dd-5ce3-46ca-bf93-c776d0fc34f7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b028f587-5189-431a-bcec-c7817f6fc6a0">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0bde2250-a21f-429c-9a43-3779e34a2e35">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a6ef33f-6e6b-4d23-a0f0-f8ab342ef757">Concept​Domain ObservationSite</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c98a3091-7825-457b-85cb-ec8a541e0fa0">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a31325b1-6e80-4856-a78f-efd77b986f12">qualifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3bc0f0f5-ca96-491c-aa3d-054867bcd2cb">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd99581a-cc14-419d-acf2-3df8ea824a25">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_607058af-512f-4695-a404-2b03f0b03d1f">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d46ba701-9eeb-49af-b849-ab46de06e02e">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2f25876-572c-4ca0-86d2-2cc2ef59e3cd">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4138ed77-768f-461c-a0c4-130806182c51">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_473eadc8-59c0-4b92-9dbc-61c3067256cd">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bbc82e1b-d617-4b58-b27a-ef88541ed00d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f8215ea-c740-44ab-80d5-9c0158b0557c">(272741003, SNOMED CT, "laterality")</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e27ce4a1-78fd-4f79-979e-108ff4d27f3a">Laterality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8ba3a01e-15d6-4ac9-b92c-fe01eb48cf4b">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73c9c6e0-310a-4dad-ac0f-60bdf96597ea">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ad664e1-445f-43ed-83a4-3abe7641d756">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9caec66-bc68-43f4-8f4c-dd46ffc98cd0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5f2f779-6120-4200-aab0-ea59b9a70233">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a82fc3a-f767-4b1d-bae4-28cf9f8adca7">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0938f1b-e8eb-4042-a285-47a009f54aa7">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_244" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b032a48b-057a-40d2-99b0-3c057f18e33e">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b31502b1-7811-49a6-b704-0423eca98c81">qualifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_396883d8-4286-411b-bda1-77c3273869cf">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ea2561a8-cc07-41b3-a015-89701380b2c3">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8830aaa-4dc8-4f06-ab05-9a6e38b2d598">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_218bdc25-2d20-4b7c-8b58-932f53feb547">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_85ae55e5-6662-401f-b576-db0721c827b4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf9cc542-c43a-4ac5-a498-7a9335a5eb75">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a57fd85-5250-4fa7-9066-1109d68c7206">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bac5443b-cd9b-47e2-9253-c480aa9721ac">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ddf408b-a6ff-48a7-a96e-8fcd512e6d12">(106233006, SNOMED CT, "Topographical modifier")</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_16f22495-1dc4-4bd0-a25e-a6884bccf568">Topo​Modifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cebadd79-6238-4d08-bca4-36973f1b6edc">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a9e4c827-b15a-4c9b-9443-acea234840aa">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a3d6c2dd-ddf2-45a6-95db-74f3affbea44">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9a6cc68a-9ebc-4c81-8a3e-344b5eb009ac">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2fb1b227-c8b4-4bec-a9e3-0e7473d01ce1">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0205238d-74f0-4715-b9b8-8361796fa9a8">SHALL CNE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e41dad1a-9128-40be-8b84-7f7c8ebc138b">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_2" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_835ee384-e8f0-497e-97f2-434c13e1f370">Method</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_add563a3-20d6-4d3e-ba81-1a5845c3ea9f">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2c0c527-1bbc-4cc5-a7ea-1a20c0a9416d">method​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_536e2211-5d30-48d1-9c70-effcbd783830">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62662270-50dd-48a9-a765-b59172aff123">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f08542a9-2797-4574-80d2-7c8e9060c169">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_896d870c-fb57-4f02-8d0e-759d1b4c4be2">Concept​Domain ObservationMethod</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_feff08f5-9a48-4cd4-b5ce-b6c8dbe8dde3">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77d8aa1b-551f-46b9-b6a0-11f1341ed359">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4960660c-0c50-4c20-81d5-02daf6005d16">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_787bacb1-2f59-4225-8ec9-306c5311ef1a">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc8bfb2b-8b4a-4879-b0a5-3da78c9e8361">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dddb302c-d622-460a-9701-2b030e7388c9">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_efebd5cc-8ce2-4218-a813-54eab9c58863">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4e7eae2-37a5-4d06-bcb9-6f1089ce2641">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ce67272-89d1-40e2-b574-8d19d68fc17e">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_627f906a-042d-416e-9f7d-fd2a0bf4370b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1b313958-b9ff-498d-8f06-a4ff07bd60c6">SPRT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1099aeba-e59a-4acc-95d4-f7a171908682">
                <emphasis role="italic">SOPInstance[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed5d2624-1ae8-4594-af2a-f4a2c9ba04f6">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77a0e6c3-ef1f-4f01-a9de-99ab2711bd90">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_067de28e-f3bf-4bc7-9df9-b57a55e1916e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_25508984-947c-475a-8f3c-cffdda6c03d7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d6e31d88-2d72-458c-84bf-c34013993304">
<xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.18</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ac4d64e-bd77-4fe9-a1b1-4db1898f0e7d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_837059d5-f210-4afd-8130-d9e39bcb9b07">
                <emphasis role="italic">entry​Relationship</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7820c7a-c18f-4774-a53e-2f2e1049df67">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_14948c17-3202-4476-91ae-02c001060a67">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7dd6195f-d1fa-4096-8758-c130995427a5">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0940d569-e4b6-4e61-a1a9-00908e93e228">
                <emphasis role="italic">@typeCode</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8e180d18-40ba-4bbc-8982-315fd2012a0d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1872e609-0c08-4607-8a93-1639385d828f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73e9ad66-ae3c-4146-92b7-dae6d3fe2cb7">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed7d3d29-cacc-4deb-8422-60366b4ebf14">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60fceb89-8d94-45db-8867-68fa60dc6c48">SPRT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f413d98-d7ee-411e-90d7-102f91178790">
                <emphasis role="italic">Quantity​Measurement[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e923c2ca-95f8-45c0-a7fa-76639d0ed0f4">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ffb2b774-8ed8-43e9-931e-75bac3d67785">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0cb57af-adbb-4c61-8bbb-45e42ac392e9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc8c861c-c047-4084-9f00-251de11c71b8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3a3068a4-511d-4124-95f4-c683ea643181">
<xref linkend="sect_10.5" xrefstyle="select: labelnumber title"/> 2.16.840.1.113883.​10.20.6.2.14</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.5.1" status="3" xml:id="sect_10.5.1">
        <title>text/reference and Related Narrative Block Markup</title>
        <para xml:id="para_3a0fc8fb-7576-4a33-a194-1bc9543c722e">The Observation entry SHOULD include a text/reference element, whose
          <emphasis role="bold">value</emphasis> attribute (not to be confused with the
          <emphasis role="bold">value</emphasis> element of the Observation class) SHALL begin with a '#' and SHALL point to its corresponding narrative in the parent section (using the approach defined in CDA Release 2, section 4.3.5.1). See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
      </section>
      <section label="10.5.2" status="3" xml:id="sect_10.5.2">
        <title>interpretationCode and Translation For Actionable Findings</title>
        <para xml:id="para_2e5f3c53-38d3-47a3-aed9-1f96c66026b5">When a measurement is out of normal range, it may be flagged using the interpretationCode. Very abnormal values, often denoted as exceeding "panic limits", or as "actionable" or "critical findings", may have values such as (LL, ObservationInterpretation, "low alert"), (HH, ObservationInterpretation, "high alert"), or (AA, ObservationInterpretation, "abnormal alert").</para>
        <para xml:id="para_a19ea0dd-2f49-48f9-9627-9dbdfb8c8e67">The translation element of the interpretationCode may be used to provide a further classification as defined in a regionally- or professionally-specified value set. This template identifies an optional value set for the ACR Actionable Finding categories 1, 2, and 3, as defined by: Larson PA, et al. J Am Coll Radiol 2014; published online. DOI 10.1016/j.jacr.2013.12.016.</para>
        <para xml:id="para_9e978fd7-761f-4802-a5e2-bfbb9084185d">The narrative text associated with the actionable finding SHOULD be highlighted using styleCode Bold. See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
        <para xml:id="para_7ad8be36-1503-48b5-b66b-d7cdc2c4eb76">Actionable findings that require a specific follow-up action or procedure SHOULD be referenced from a recommendation in the <xref linkend="sect_9.8.11" xrefstyle="select: labelnumber title"/> Section.</para>
        <para xml:id="para_9cc98957-4843-43a9-8346-87de961ed811">Communication of actionable findings SHOULD be documented in the <xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/> Section.</para>
      </section>
      <section label="10.5.3" status="3" xml:id="sect_10.5.3">
        <title>targetSiteCode</title>
        <para xml:id="para_f6e74496-9799-406d-be9c-6fb3f047d118">Each observation needs to fully specify its site/location.</para>
        <para xml:id="para_0ab531bc-fd98-40b5-925a-5de9b9da4829">COND: If the observation site is not pre-coordinated in the observation/code, it SHALL be specified in the observation/targetSiteCode.</para>
        <para xml:id="para_e42566c0-f6f7-41ed-972e-c42bc8684013">COND: The qualifier element for laterality SHALL be present if the targetSiteCode represents a paired body part and laterality is not pre-coordinated in the targetSiteCode.</para>
        <para xml:id="para_7fbe8992-a801-4342-b4c8-e6e61d45a649">COND: The qualifier element for topographical modifier SHALL be present if the targetSiteCode does not fully specify the observation location in sufficient detail.</para>
        <note>
          <para xml:id="para_e327d49f-fbb7-46a0-8bea-0dc67f4637f4">Inclusion of a site name in a labeled subsection title (see <xref linkend="sect_9.8.9" xrefstyle="select: label"/>) does not imply a finding site for observations within that subsection. The title is not semantically part of the post-coordination, and target sites must be explicitly identified.</para>
          <para xml:id="para_61db8673-ee67-4bb3-b81f-b5c685fce5ed">See <xref linkend="example_10.5-2" xrefstyle="select: label quotedtitle"/>, an example of a measurement using a topographical modifier qualifier.</para>
        </note>
      </section>

      <example label="10.5-1" xml:id="example_10.5-1">
        <title>Quantity measurement observation example 1</title>
        <programlisting><![CDATA[<text> ...
    <content ID="Q21" styleCode="Bold">Calcium score (Agatston) : 817 [HIGH - ACR Cat3]</content>
    ...
</text>
<entry>
    <observation classCode="OBS" moodCode="EVN">
        <templateId root="2.16.840.1.113883.10.20.6.2.14"/>
        <id root="1.2.840.10213.2.62.7044234.11652014"/>
        <code code="112058" codeSystem="1.2.840.10008.2.16.4"
            codeSystemName="DCM" displayName="Calcium score" />
        <text><reference value="#Q21"/></text>
        <statusCode code="COMPLETED"/>
        <effectiveTime value="20140913223912"/>
        <value xsi:type="PQ" unit="[arb'U]" value="817" />
        <interpretationCode code="HH" codeSystem="2.16.840.1.113883.5.83"
                codeSystemName="ObservationInterpretation" displayName="High alert">
            <translation code="RID49482" codeSystem="2.16.840.1.113883.6.256"
                codeSystemName="RADLEX" displayName="ACR Category 3 Actionable Finding" />
        </interpretationCode>
        <methodCode code="112055" codeSystem="1.2.840.10008.2.16.4"
            codeSystemName="DCM" displayName="Agatston" />
        <!-- entryRelationships to SOP Instance Observations may go here -->
    </observation>
</entry>]]></programlisting>
      </example>
      
      <example label="10.5-2" xml:id="example_10.5-2">
        <title>Quantity measurement observation example 2</title>
        <programlisting><![CDATA[<section>
    <title>Left femoral artery</title>
    <text>
        ...
        <content ID="M10">Distal lumen stenosis: 75%</content>
        ...
    </text>
    <entry>
        <observation classCode="OBS" moodCode="EVN">
            <templateId root="2.16.840.1.113883.10.20.6.2.14"/>
            <id root="1.2.840.10213.2.62.7044234.988810005"/>
            <code code="408714007" codeSystem="2.16.840.1.113883.6.96"
                codeSystemName=" SNOMED CT"
                displayName="Vessel lumen diameter reduction" />
            <text><reference value="#M10"/></text>
            <statusCode code="COMPLETED"/>
            <effectiveTime value="20140913223912"/>
            <value xsi:type="PQ" unit="%" value="75" />
            <targetSiteCode code="113270003"
                    codeSystem="2.16.840.1.113883.6.96" codeSystemName="SNOMED CT"
                    displayName="Left femoral artery">
                <qualifier>
                    <name code="106233006" codeSystem="2.16.840.1.113883.6.96"
                        codeSystemName="SNOMED CT" displayName="Topographical modifier" />
                    <value code="46053002" codeSystem="2.16.840.1.113883.6.96"
                        codeSystemName="SNOMED CT" displayName="Distal" />
                </qualifier>
            </targetSiteCode>
        </observation>
    </entry>
</section>]]></programlisting>
      </example>
    </section>
    <section label="10.6" status="2" xml:id="sect_10.6">
      <title>Study Act</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d44f318b-9350-42b1-9652-7ff7d8def1d0">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46179dec-86e3-44ea-8af0-1c4df5fee7c3">1.2.840.10008.9.16</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_37f8007e-50c7-4d07-9d20-3524bf61c9d0">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de74fd02-70f1-4349-8e1c-a61ed1ec98f5">Study Act</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8ccbafd0-d56a-4fca-8ae3-2d05ecc5489b">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d56e450d-c462-458b-8fca-902acc4275ee">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c227044-5a53-4112-b317-70cfb5020047">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5878b26-d863-4c34-b86a-15c535d5c4e6">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8c2167b-e02a-44d9-92ce-72919b914497">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_daf777c3-9322-4500-b81d-9014f2570fab">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b951287-37a1-4a59-be42-6a626b1e6016">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_620d7e39-6f4c-45ff-8009-c83cfc8f1b99">A Study Act contains the DICOM study information that defines the characteristics of an imaging study performed on a patient. An imaging study is a collection of one or more series of medical images, presentation states, SR documents, overlays, and/or curves that are logically related for the purpose of diagnosing a patient. Each study is associated with exactly one patient. A study may include composite instances that are created by a single modality, multiple modalities, or by multiple devices of the same modality. The study information is modality-independent.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf95de24-006e-41dc-b461-90bb2129740f">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8917427e-43e3-41e1-a874-68ca6d8b5d15">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c6d66b75-2110-4b58-aa28-90326a0bff15">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0a6150cd-dd43-4685-a877-e7d2fbb3d278">Included in <xref linkend="sect_9.8.7" xrefstyle="select: labelnumber title"/> and <xref linkend="sect_9.4" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_58bc2bbb-ffb9-4d1a-8a97-df90765398db">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8fb8d46-5741-481e-a960-4b72eb146882">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e85124d-7e1f-43db-85d7-810ef0af59d3">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3e6d006-452f-411e-91c6-637e987980b7">Open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9ad27e6f-5fb4-455f-8da5-78161a4c6634">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8323963-f631-4206-b612-e14c29b7d02e">DICOM-20150324: Initial publication, derived from template originally published in DIR r1-2009, revised in Consolidated CDA r1-2011 as 2.16.840.1.113883.10.20.6.2.6. This derivation makes Series conditional (required for Object Catalog) to support use in Comparison Study reference, and uses DICOM-20150324 Series Act subsidiary template.</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_f5a68a50-09df-48eb-b3e2-98889f1d2a9b"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3adcc039-d24a-48da-a7d4-4001d940ca49">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_37ce87e7-22e2-479c-b776-db2da958c7e3">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a902903b-7510-4b0e-b286-5fe826b4c52c">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_02f9cc69-8c0f-4bc6-8d65-5d7aa0825466">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c6582ede-8118-41f0-b34d-eb8c196b7871">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2bba7393-67d5-4ba6-821d-1603acf86f5a">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a2d7f474-f35c-42de-a341-e6e7289b6f9d">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_fb9a7d60-4bf4-47ab-bd55-4c77fc340274">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_84ba0893-efcf-4504-b3a4-51867f7ff9fb">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5461fe97-3bc2-4e94-8c1b-7a8742e6c681">
                <emphasis role="bold">Study[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e03ccfc-8511-418b-ba9a-7caaf1c97104">act</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_28f87462-1608-4f23-93dc-33a8239a0464">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b16cbf3a-46b6-417d-8c11-dd9214279e0b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a6cf1f4-0b04-4197-82a0-af0ced139cd3">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e967feb-06a6-4323-8c14-3e2f6936fbeb">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61d15b8a-a552-4594-b938-387268bb32a8">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_86ad0bfb-0305-4263-81f3-debb0e517605">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48df81e3-c434-4f18-82dd-ab3dfda32115">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bc8c7569-6db4-4862-b678-2e6cedde7a70">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3fa09cad-8c08-4fe8-b5f4-77c29a70485a">ACT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_69b4f6be-0d78-48d9-a64f-a61a8d59d4a6">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2c4e039e-fc11-4185-b0b8-08250c8edffb">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1c7c37f-f179-490e-b816-7c2a2e7e88df">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_203c7832-f214-410d-94d7-a450a05caef0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4164d7a0-05f9-44b4-aa40-a791d502907d">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97fb2063-ef36-4cbf-bec9-274d4205c686">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8896b534-bcce-4890-bea7-b5c30fb54bb2">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_92467010-c608-490a-9eec-f513cf3df18d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a0bc047-9fb9-4ad8-b28e-004d3d331ff8">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9c9a442d-1229-4c23-a58e-67e7c8a5a442">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ce015b7-a04d-45b9-897f-a293c7f582c7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a5c89086-1aed-4b97-8c20-7feaefd15a74">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b7469f2-45ce-4728-a218-3bfc542cdd1e">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ea71352-f9c0-4b24-b363-7254e30a3af8">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd2f2231-7c13-4fff-a754-3d56dc1b074e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48e935ad-46a6-475a-96e8-4c60b09ac4a0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15214e39-5c73-4583-a1a7-4b34a261b6fc">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3f584c2-21ae-4cd7-8d0d-283278c69080">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_357915c0-04bd-486a-8fb0-5dfa7d8fcb8c">1.2.840.10008.9.16</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2c014c6-62a0-4a08-b033-6f13382136b5">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60154115-ed51-4792-bd9e-df36c14070a4">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ae26ce37-7ffc-42eb-b160-7476d1976879">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_227faad5-96c6-4bb5-9b9b-15bbb6e7f70c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_18b33311-2480-47f9-a259-b836c4a28eee">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc29fe54-4281-481a-81f9-e4742ab69e29">Study​UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_55675f90-d971-48f4-8827-ece82431764c">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_57b673d1-7c9e-4214-9c31-c74b1cfda9d4">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_476fd011-aa9b-4614-85ca-ef62f64494c3">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c611df34-1d7d-4338-aafe-9039b00eda8d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f8b42e5a-b0f1-480e-aa56-20b350665164">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_32c773be-8c7a-424c-93b5-0c20cc9dc6ff">
                <emphasis role="italic">Study Instance UID (0020,000D)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b425a17-b62e-491f-8113-cbb3b56e3004">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46c19168-6c31-4ea2-b133-f3fc4d18b470">@extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_799747bf-7f30-4da7-b7f2-4831173e77bd">0..0</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3b365aa-0a68-447d-88e5-89f879d16070">SHALL NOT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7808c054-3fb0-415d-8278-5ee6e49c37b3">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9cbdce48-c505-4a4d-83f8-e438a29190e1">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3949d015-beb2-4b5c-95bc-b317ce20f1d6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6b85261a-0336-4038-b0e9-68ba83bd7dae">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40a481b6-e54b-47cd-b719-851bc3c6c6f3">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc695935-f1c5-45df-8c45-e8a4641e4e16">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06d8b5ac-22e6-4efa-8ba7-ee815024fb37">
<olink targetdoc="PS3.16" targetptr="DCM_113014">(113014, DCM, "Study")</olink>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8243c02d-2ff5-4621-a5b3-bd7b92314596">Description</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee03becd-7416-471e-abb7-55c0daf2e400">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5289b383-2c25-445d-9ae7-39214d5c23e5">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce6762e0-3c3e-4062-ac3c-50ec0ca63670">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_845cc307-7087-4bf1-8a8c-bead5503806f">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1341d99c-6d50-49ab-bccd-2eeac274a1ec">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d867729e-2c8f-4ac9-8563-d47e00ad4419">Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ea3c63ed-40c4-46f1-8a1e-25ccad690273">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e4fe3a5e-d4a6-467b-bd2f-39238db6d60f">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5478178c-846e-4781-a8a5-bee7a1139026">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_736f5abe-4f72-4f1a-9c8c-564d0d8d398b">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2dd351d0-a360-48af-84e9-77670a8729c4">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b8346605-363e-4f15-aa05-c7a74abf2c33">
                <emphasis role="italic">Study Date (0008,0020) + Study Time (0008,0030) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d19c3f64-e3e8-4d1e-b84c-d4492177a957">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43404c3d-e1f2-4387-a410-c86e1b47e3ef">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_06514fd9-a6d4-4cc2-ab48-c5eff092c776">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_832fd425-56ee-4fa1-82cf-ed797161a411">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfdf257b-5bde-43cd-b416-28a53047d851">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba455308-810e-4a4c-90de-5b8a1a9aee59">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_441322fe-da2f-43ed-96fa-014305fda92b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_689cc6c7-f68a-4c44-9c45-84c42ec09ee2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17d5186f-9f40-4169-816c-9952afe4042a">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4417a36d-606a-4f9a-b473-d251ab48f1ea">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d8d5d5c2-679e-4e97-b40b-63b271f3628b">COMP</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63988dbe-a370-46f2-80a7-ae3c891dfac3">
                <emphasis role="italic">Series[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4cadba6f-22c6-4dd0-b6e2-d600876c44f4">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_390785cc-2896-4a71-bea6-f86b53593625">
                <emphasis role="italic">act</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48e3898a-bcd6-4554-bd08-1ed5a1b3cca4">
<xref linkend="sect_10.7" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.17</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.6.1" status="3" xml:id="sect_10.6.1">
        <title>entryRelationship/act - Series</title>
        <para xml:id="para_37572c02-b8cc-43b1-ab72-ac97ecad38e9">COND: If this template is invoked by the <xref linkend="sect_9.8.7" xrefstyle="select: labelnumber title"/>, the entryRelationship to the Series act SHALL be present, otherwise it MAY be present.</para>
      </section>
      <example label="10.6-1" xml:id="example_10.6-1">
        <title>Study act example</title>
        <programlisting><![CDATA[<act classCode="ACT" moodCode="EVN">
    <templateId root="2.16.840.1.113883.10.20.6.2.6"/>
    <id root="1.2.840.113619.2.62.994044785528.114289542805"/>
    <code code="113014" codeSystem="1.2.840.10008.2.16.4"
        codeSystemName="DCM" displayName="Study"/>
    <effectiveTime value="20060823223232"/>
    <!-- **** Series ****-->
    <entryRelationship typeCode="COMP">
        <act classCode="ACT" moodCode="EVN">
            ...
        </act>
    </entryRelationship>
</act>]]></programlisting>
      </example>
    </section>
    <section label="10.7" status="2" xml:id="sect_10.7">
      <title>Series Act</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f90b89fe-4e60-402d-a3ff-27e75401bd1e">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5a6f6378-530f-49e2-87ed-ce73ad1851df">1.2.840.10008.9.17</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0c13893-b9d7-4903-bcb3-cb5bf2ece18b">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0b3f488-927f-46ce-bb57-afd2e5b1b56c">Series Act</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ca4a6cf-203a-4ee1-b194-93e6896ea623">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54d03cbf-ca5b-4a29-89e7-bbcc166d55c6">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_161e1960-b220-4e5a-80a7-b0ebd2980066">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83753a04-cce2-40ec-8f88-6ceff304bb28">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f652ccc-fb04-42be-a74c-73e741cb465c">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d11a5dc-0d4b-4480-a13e-bae3fe96684c">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ff9d2415-6bf2-47ae-82df-a4db05cb4b7b">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20fdc2d4-0bc7-43fb-812d-9ff4a751a27e">A Series Act contains the DICOM series information for referenced DICOM composite objects. The series information defines the attributes that are used to group composite instances into distinct logical sets. Each series is associated with exactly one study. Series Act clinical statements are only instantiated in the <xref linkend="sect_9.8.7" xrefstyle="select: labelnumber title"/> section inside a <xref linkend="sect_10.6" xrefstyle="select: labelnumber title"/>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ffc6b69-516c-4bb6-b7ad-d96b95bd9792">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9da83dad-c40a-487a-a56e-f0252bf22872">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4b73e645-44df-4eca-9b08-fcc0d618a928">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3c5d8a0-6109-44ee-8c16-d3f8199df281">Included in <xref linkend="sect_10.6" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e61dcd9-ed43-48ec-9bd7-3aaef697b2bb">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b9e0509-2f40-4cf5-bae1-ddc1cac1f437">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7985d6eb-20e8-462c-864e-78e132b699f4">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2bbd692b-bc25-4725-990d-7b58e21d0123">open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f87b67e8-c280-4156-b575-d940e039b1ef">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c7acdd9b-15bc-470f-bc71-d5e2afcf936e">DICOM-20150324: Initial publication, derived from template originally published in DIR r1-2009, revised in Consolidated CDA r1-2011 as 2.16.840.1.113883.10.20.22.4.63. This derivation uses DICOM-20150324 SOP Instance subsidiary template.</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_bc19ea2e-886d-4165-a84b-816513963a71"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_f9897516-56df-4266-81fd-4bab4e273421">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b64b45f9-498a-431c-8420-680fdc739bb4">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_75d253a9-5242-4626-975a-143a95d940a0">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e3a23b20-731d-403a-bcbb-7f1815deba1f">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b56ae4f1-b1da-4c4e-9c4f-3b817bb0d9cd">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_27e96add-a6e1-425f-a2d3-16736f2b694f">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_95390e3e-2508-4fe4-bd13-a37c30aab99d">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_8d3523ad-645e-4b21-8fc5-9cf82c329f33">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c25305c3-161c-4518-97fc-178b67f4198f">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dfc24681-937b-4545-b924-6e84a3a39863">
                <emphasis role="bold">Series[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3fe889f7-f7eb-495b-84b4-4520dcd7556c">act</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_febc45b6-7bd5-43be-9ce4-944423b6f6f4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_974491a2-34b6-42be-8932-24d0369ad840">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_432e8a14-8b1c-4b59-b27a-71a5bec3dcb9">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e525da48-dd6c-47a2-ae45-e4c71d60c5f1">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_383ea080-4a1e-4bc7-86d9-38c72f57e3e0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_69da81b5-a894-4208-8dcd-02528bd25788">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a9285a4-a648-4f77-9f1c-ad2886a6cf26">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e9be760f-e354-46d6-b50f-40ed8f8a4481">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5c958a2f-41d8-49e4-8bcd-27e4cf41e9da">ACT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d216da3-f37e-4d93-babb-f7f2ea10be69">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8dc110b7-e4f8-4f82-8b58-43d2c923e0fd">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0701f8bc-4e40-471e-b85f-6e330b967b87">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a68a199-f50e-4334-9b8e-a7930f72e093">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08b75eef-1596-4f89-9a83-45b7108f5fd2">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e06684cd-8d0b-43ba-b585-665867bad49b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31ab0a72-578c-43e6-be88-7a82b3996210">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ed7dcff-80f0-4ae1-bdf3-395d7b5372ec">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c375f07b-aec8-41f2-bcec-52625bd07be1">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83c2a83d-b08b-4687-b75a-8508817d1403">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41334f35-e574-4135-bb81-e8e062044bf1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3e6cf9b5-bfe3-4acb-b3ef-0a5a94e30b4a">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe7bb43b-5d31-4831-bd35-99d2866ee358">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a550a0d3-d1d7-422a-ad30-f3503e92188d">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_814c3e14-4e42-484d-ad8f-81f32558c8d7">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2ea4fd91-b473-4b16-934e-8027df3b22c9">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cca89f89-29f1-47dd-bbdb-66425fa41656">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31f5f380-db1b-4e72-898e-125af9169374">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fbec2e63-75b3-4402-a590-3ee3457043ba">1.2.840.10008.9.17</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0fd448b9-7b17-4267-a0a3-0d18f417549d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_847d8776-5c1d-4c5f-b756-8b80a36bb7d8">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dd45f140-2878-4b18-a59d-4c574a9f49cd">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0b2dcd1-a9b9-457f-89c2-4e32ab72c867">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8727f117-1648-49b3-89cf-308db70e2496">Series​UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e6cc2015-ba1d-43c9-81f9-e3015e0e37ec">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2412399f-ef46-4c96-a5b8-7578ce607736">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_65d7b99c-720c-44f0-92d7-7866ffc4010d">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62683127-06bc-459b-83dc-26824f08472c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89e3ab70-0910-494b-929c-b65662e3b2ea">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_054ae337-2039-4151-976f-5d6a5dd09306">
                <emphasis role="italic">Series Instance UID (0020,000E)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c2a154e-aeff-4eba-b703-17e2d78354b6">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b1bca843-8f0d-400c-98c5-cf2bb4397e85">@extension</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad71ff38-b487-40bf-ace5-ec432e32db41">0..0</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1326ac11-ccee-48f4-8840-32354b89bb4f">SHALL NOT</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ed8191e-e269-408d-9ce7-378fcc75f150">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_390128b2-5712-42c8-8d9e-f6a42e054988">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_665132ed-5a06-424b-b2f7-8a0585c4ac20">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_520cbb1a-e666-456a-8c7a-9359c7dff52a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_638d3673-06fe-4bcc-a102-2756bdf0c9d1">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c82280ca-ba50-40d7-b61f-cc8715f68b2f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d6c4b3f-361c-4f2f-b208-1851da71fdbb">
<olink targetdoc="PS3.16" targetptr="DCM_113015">(113015, DCM, "Series")</olink>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_579dde02-4a32-4943-be06-1cc3cc8e1596">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_23189d56-0e2b-4104-9702-19fae4ba802e">qualifier</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f1859e4-b1cf-4e82-a636-d143887d55ef">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e28c2676-d4f3-4109-9a91-e21e8b84e504">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f4d11ccb-d1c2-4c76-b896-bd94c787a2f8">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca079c04-d302-4d0b-b963-bb38352a7b7d">name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_001449ab-6586-4618-a980-cf4c584b86bb">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d1a28cd-1bac-49e1-be82-0e3f3636617b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71ca2310-24e0-407d-ae9f-eff7c3c22a42">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2bd3a0df-0068-496b-bacb-89114a8e77d6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_933dd6a4-8d87-43ea-ac86-98dc29bc6b8f">
<olink targetdoc="PS3.16" targetptr="DCM_121139">(121139, DCM, "Modality")</olink>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_93d3d004-91d2-421b-8f4e-2c6e0ba1e5ee">Modality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81c37623-179c-4a87-bec0-1dcfd0fa7087">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a8bec13b-0c56-4014-aff3-f1e37b16abfd">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3672fa56-2b93-45af-a14a-c00de29cb3f5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_32916672-0715-43b2-8ab0-75062448e487">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81a0b9f6-defa-479f-8a3c-df7fd978bbf1">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2426a209-ebb5-4099-b27f-95779750520e">
                <emphasis role="italic">Modality (0008,0060)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f865ac7d-6d34-4de7-beef-2ac50a881fee">Description</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98132e61-51da-47e3-b0c4-5774ee5a36d8">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9bf1510b-78af-45db-bd3d-ae640805ab9d">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bc924385-1f2f-4c3b-8118-59a4206e51a3">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3021b2bb-7a66-4358-8754-c959070e0d3e">MAY</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b576d37a-1cfa-4160-b590-55ee5f2d8e18">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b0095c1d-e81d-4fa9-9309-c250b55cd24f">Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_60d6fd01-c485-47dd-89e3-3eba530ff06c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b8e0f2e-f006-4e32-b021-2109776680ca">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0c870a0-89d1-4a1a-bb09-ea949040b2ca">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2bcd0cc6-dd14-437e-8414-83b25d3c6c4f">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5abe05ce-5d71-49a2-8382-4cf3acfa5630">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4219261f-dcaa-4adf-b8e3-b576944e1e74">
                <emphasis role="italic">Series Date (0008,0021) + Series Time (0008,0031) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_59cba98d-1616-48aa-acd5-7d56c195d89e">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74f6bed5-af2f-4450-8c91-da073bc0fa6b">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4624dc15-1a12-4144-85e8-3b37a3d707a3">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ecee447d-76d9-4b2d-a757-084be1ca7bc0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_da773953-1631-4c20-88f9-e1b6fc1f40be">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_73f625f0-1c7c-4309-b8ef-963f4e622ce9">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_75ce5b6d-ef55-44f8-af35-30d12efd0064">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b7e4eb48-7644-4b4a-a373-7c42c4fc199f">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_09619a6a-c451-4012-ba23-dde3fa72a430">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b21837ee-ad03-4692-855a-05c70b3472fe">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_31449fc9-9f7e-4eda-b6c3-fe274c4e4314">COMP</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9e4b9ae0-1659-42a6-8303-9b1fd016b261">
                <emphasis role="italic">SOPInstance[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_96ea3a68-0819-471f-a084-13b4224c01b9">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3dfa406-cf71-4236-9628-2df402e558e2">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53df010f-f12f-46a2-ac68-00cfdbef519f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0592b3d8-aecb-4e9f-80f8-b751db873266">
<xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.18</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <example label="10.7-1" xml:id="example_10.7-1">
        <title>Series act example</title>
        <programlisting><![CDATA[<act classCode="ACT" moodCode="EVN">
    <templateId root="1.2.840.10008.9.17"/>
    <id root="1.2.840.113619.2.62.994044785528.20060823223142485051"/>
    <code code="113015" codeSystem="1.2.840.10008.2.16.4"
            codeSystemName="DCM" displayName="Series">
        <qualifier>
            <name code="121139" codeSystem="1.2.840.10008.2.16.4"
                codeSystemName="DCM"
                displayName="Modality"/>
            <value code="CR" codeSystem="1.2.840.10008.2.16.4"
                codeSystemName="DCM"
                displayName="Computed Radiography"/>
        </qualifier>
    </code>
    <!-- **** SOP Instance UID *** -->
    <entryRelationship typeCode="COMP">
        <observation classCode="DGIMG" moodCode="EVN">
            <templateId root="2.16.840.1.113883.10.20.6.2.8"/>
            ...
        </observation>
    </entryRelationship>
</act>]]></programlisting>
      </example>
    </section>
    <section label="10.8" status="2" xml:id="sect_10.8">
      <title>SOP Instance Observation</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dbcb8cb2-e1af-4935-8990-8c71b60d1266">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_70f2552c-acdc-4fe0-93c2-cdecbffd81ec">1.2.840.10008.9.18</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b83d8def-e9fb-48b5-8f5c-8db56ddb7a43">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed49a78c-21e1-4df2-aec9-bc88b64951b5">SOP Instance Observation</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ebb53941-8dea-464b-8748-8c0bf9b5a8d7">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ffc9c7f2-17ba-471c-afa5-9beabdf3de94">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_15c20c43-7ec5-40ca-9ca7-682edb20e202">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e4acd30-6725-472a-afe0-b2eb79f3dcbb">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f38a3aa-e275-4248-abe2-4effa92cefa2">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77d0f199-c116-4703-83c4-534e50fc0f89">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_912ff7dc-eed0-4c8a-ad47-0f226877341d">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fbfb6619-1ad2-4625-b925-87764df1e7a7">A SOP Instance Observation contains the DICOM Service Object Pair (SOP) Instance information for referenced DICOM composite objects. The SOP Instance act class is used to reference both image and non-image DICOM instances. The text attribute contains the DICOM WADO reference.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5081edf-12b4-4060-bfd7-d0582fbe8319">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb44c063-0233-4d69-b762-069f6f671c38">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08119a6d-83dc-4409-b096-d448c6411e80">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98fe82ae-e861-4b5d-a5a4-f35995df7cc9">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8bd79851-43f6-4dd1-afea-321489858dde">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1ff8a223-a5ec-4b53-be2f-2754e461f695">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_947db92c-b49a-4a08-af70-1c52a328df85">open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d1a7bfb-7046-4b8d-99ac-93a3c85e309d">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5571e93c-9582-495d-9e37-0a597ae36bc3">DICOM-20150324: Initial publication, derived from template originally published in DIR r1-2009, revised in Consolidated CDA r1-2011 as 2.16.840.1.113883.10.20.6.2.8</para>
              <para xml:id="para_bb6de03e-754c-47a1-bfb9-d9c7d434116b">This derivation includes Purpose of Reference value set specified with DICOM CID 7003; directly incorporates descendant templates Purpose of Reference Observation, Referenced Frames, and Boundary Observation</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_f53b9af9-e0d6-4157-98e3-4890063f8cab"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_e503c292-c1bf-4cad-bd10-094968c68a2c">Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_976734d1-3e3e-436a-b769-64ae169fef42">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_69df80af-8a3a-4bcd-9fd9-01a70b0a1f28">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_a721e5bc-4ee7-4eb5-a38c-089832f0a20b">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_cbcdfadd-811a-41fe-82e3-2a6bf5a88dd4">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_c4b2d075-2352-4537-97da-183d34ef5248">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_742885eb-6e73-446f-abca-b5672bbc6d35">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_872cc14d-b1c6-4187-ac19-cee9f588f6a9">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_7e1f15c3-02f6-4e95-a803-25dfe2b8b304">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_324a98d9-de4b-47c2-ba4e-af77bfbf67bf">
                <emphasis role="bold">SOPInstance[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a7353746-3dbf-41b8-a7a2-763f0eb8d73b">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_48748df9-f739-48fb-bad8-91a54b265f6e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2dad9aa-a256-46a5-a550-1e0ed4ba30a6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3690f1a-3673-401d-b2fc-ae3fbf7189ea">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8844ea71-ed92-460b-99ff-c62d99c428fb">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a41050b5-4297-41d4-a98b-e295fcf56139">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b91c7755-5872-45ee-9fa7-0267f3a88757">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_220090f2-8449-43c0-b7f7-79591d546ade">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61082848-5af7-4df0-8bed-1af1e881079e">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_145561b2-8775-4f19-be51-7bff06aaeaea">DGIMG</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_04c9ab61-526b-4117-94be-642a41f2fabf">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_994c17c5-c499-40e7-b9d3-95f34e98b96a">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee24d9df-0e46-4ca9-9092-4e6ea2c0062a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5fa8cb31-af91-4e0b-941e-6e9febaeb979">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_13d97113-0d17-4f93-afb3-817ea4423339">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5f92309-5770-4e77-93ff-33fb4e0c55fa">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9b1bd19-b305-4f82-a5a4-db303f649bc0">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e46e0780-eee3-48f6-81d3-4ab43969efed">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_036bc7ea-13c3-4da7-936a-830925984ef9">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6dc90274-2515-440b-9b71-21f1cf1a742c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3ccb8d4-710c-4ad9-a915-61237e4e43e4">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc394f84-7118-4237-ae3d-0973d53c13dc">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e351827-a7b1-4376-aac4-b9475136ca07">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20c2db26-e7c4-46f8-90d8-21f7aea035bc">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_714f7d07-8eab-43a0-8a95-10b5ce3ea12b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19404eae-cdd1-446f-9275-ad7d97331358">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a628dcee-40ba-4ad6-a05b-e0ea0f5da88f">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_defd0b1f-10ac-4a0d-a648-aebf8548a440">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bc15f1ab-096b-4d36-ae47-49ee9d35dec3">1.2.840.10008.9.18</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f56579d5-3343-49ac-b777-c6dc8813bd71">SOPInstance​UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f14d9a63-19ab-4618-8872-7e35c830b2cd">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_80e53455-882e-4cc3-bae0-4565deca4ff9">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64089bfb-2cba-4233-89c5-060f89b0751f">1..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_83625544-bcfa-4cc4-be23-ae78b6034450">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_63b68640-81cc-46b4-908e-b932e5a51230">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a2a423b-04de-462d-b8fa-6d10e69bd502">
                <emphasis role="italic">SOP Instance UID (0008,0018)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_91b6c0d9-9979-4e9d-8d2f-7308244afb20">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_65ee8be5-c641-4a5b-ba02-d1072fe3c4ec">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5da7a7e0-3de0-48c5-bbd6-88bc1d919bcb">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d69069f5-4fc5-4b96-8533-b08ff2712b56">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1931a146-2cbe-494c-8422-561dc9359d32">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6c23aac8-8f9f-4d4f-953f-b5ae4b52065e">SOPClass​UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_55605414-54a9-4891-b27d-c97b03081f37">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_98f6f554-cbb3-4f90-a2ea-ffd1e99f7a09">@code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36ecf695-53ae-4a6d-971a-fa155874dfa8">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_61ac0205-2560-43f0-9c7a-81b6f1298d22">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36b9f0af-a519-4a67-9b8a-dfa8d2b92ab8">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_85365527-21c6-43dc-a558-b54aeafb8a1c">
                <emphasis role="italic">SOP Class UID (0008,0016)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de84c964-f7d4-44ef-a403-8e2e4f95ecf5">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0826d60-96c3-491d-b005-72dc2df68230">@codeSystem</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62976525-4b4f-46dd-b867-0b5819852fb8">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53994a78-afc1-4411-b4ac-f0bf6b443a94">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36be5560-7d22-4a3f-89db-e761b09fc1d8">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a30b138e-d37f-4e37-9ee3-971473a994a3">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_68e602fd-7d5e-4cfb-bd2b-76ca7b6dc314">1.2.840.10008.2.6.1</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fe639e9-8d60-4fcd-86a6-675724a225ed">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94ecf240-d75b-4348-8920-653b42d4cf9b">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc2770ff-c5de-4dfb-88ad-aca84fc2c88c">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_810ca510-991c-4afe-a225-754106c85db4">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0d5db9f3-d403-4bf5-8946-daa91bca73c6">ED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8dfaa4ab-b3b2-489e-b53f-04e02d1dd7a1">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ca75bbfb-a8f5-486a-97f9-8a4a74b1ef30">@mediaType</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a079c07-ee42-4952-8918-25675941abbe">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f828d7b-43c4-48be-a22c-2581ce354d50">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2356039e-512c-4588-a5f1-e9d2926ce067">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4667a05f-f80b-40f4-b0ca-609ed0406841">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed3182af-4595-4858-96e7-e4414bfb1f76">application/dicom</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_43edfa85-8cf1-4c05-af19-6dd72ce674ac">WADOReference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5e5a00c7-f039-4f13-ad65-8ec108372a7c">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_46dd5723-a02e-48e6-9457-363c275e39e4">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_234185dc-d06e-4db1-b2ea-6a2c5d59cb47">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53f7badd-c7c0-4cb6-b972-5f1efa4f5b6a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62421747-e7b4-49e5-8a5a-5792ce2cee27">URL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ead44f3-0ed6-4f11-bbce-2bf5c204307f">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9052e0d-f2a1-4133-94d6-301e483cd43c">effective​Time</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b3808da6-eecc-495b-b7e2-0368b689d448">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f3a2d1ba-20ba-47e3-91e0-fbe0cc585959">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_19a50327-217f-4d93-aedf-ada497651507">TS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d36583b-474e-4f51-9cbe-56bb8c4ab544">
                <emphasis role="italic">Instance Creation Date (0008,0012) + Instance Creation Time (0008,0013) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f1c3e76-56fe-4161-9bb3-9050c8ea2521">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d96edd6-f163-49d6-b5bd-b0f83092cc51">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c4b7fe6-95e7-46e7-ac42-e4b3a355fef0">0..*</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6f3bc9fb-f796-4713-a5a2-547052f8e72c">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c012c29b-00ef-407e-8dab-1ff1939d0c3f">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a68f74a-1fb3-40d9-8c4e-05189be71d41">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_07a77757-1b67-4a30-9787-e3246c9e0d8c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_735e4450-7a55-4cbb-9fb4-ab0065253b40">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac371a08-f347-44d8-89b4-29fd7fd5dab6">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f741331-7bd7-46c4-b7ed-f9c0591d9b36">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_20b0389f-8bfe-4208-877d-1ed47694b1e7">SUBJ</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5078dc3b-25ec-4e00-9ff1-f533f4f8e298">
                <emphasis role="italic">SOPInstance[*]</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6800d191-0973-4c66-9975-0331f2742099">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_321fd8ed-b901-4a90-a216-78b858566f62">
                <emphasis role="italic">observation</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9ade162-3ac6-4842-b472-f3885beef8b5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f2936a94-9f70-4235-a3fb-f87f98575123">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ecd02c6e-207f-4f9f-adce-3c50756483d1">
<xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> 1.2.840.10008.9.18</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0122837b-af86-48c3-bc09-f84c1141cb80">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f9f1fc05-fd54-4238-a185-7aab3cf29a69">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2c2d140-7ea0-447f-8ae8-380abe34eeae">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_beb5a726-4bf1-4350-a70c-f5935147a4fb">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f61e8ee9-e77a-4feb-858e-7dec73791ab2">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_27058745-7ee4-400a-972a-d3f18923d526">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fb0df606-37fe-4c73-9ee6-d78f2f5278f9">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3ce47df-8eb8-4eaa-ac0a-d90412fa88dd">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dc562eb8-e933-438f-afd4-55f4a7fb6db9">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_faa2f4e3-969f-4e17-aeba-894a41dc5101">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47d08f22-1f83-4a46-b114-f838f090a0a9">RSON</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e00f968e-d6ba-461f-bf2a-66e9b7732606">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ef15d9a6-853a-4196-9632-42f302cd6538">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6129e66a-2145-42a4-be32-7848344403a6">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d769b28e-bb6a-487c-aeb8-cb74568e5f58">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_33dc2e2b-50b3-4fbc-93fc-4ba70ca8f7de">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2da891f-9462-4c86-ab21-791508f3b69e">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6608b6ab-ba3b-4567-844f-0644f87d4b2e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cacaf7d1-528d-4f1d-84d3-0e02f928749d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f5d774f-6c80-4d36-9184-3b070a50f8fc">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_520781e3-3bce-4086-8f49-8c8bb03d0ebb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71e6b640-02b5-4307-a947-31b04a967954">OBS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c311c47f-ff8c-4578-8952-730830840a89">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_beb30c8d-17f9-44df-a26b-f3daa47b6cf2">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_96c47827-e43f-46b0-90c8-8429246fef8b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_71ef2f20-6ac4-4e16-960c-da4998b0e6d1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_29665d94-01ba-4cdd-b8b4-518f570af1f7">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_401a3a9e-f2a8-4924-92b8-32cf03e5fa96">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ef9a93da-297f-48be-91d2-ea756755139c">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f1bb4ad-4133-437c-b08e-3b0da0a7a366">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_10d5564c-24c0-4c8c-b847-12604e349387">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7ebcd1e7-67fd-4a24-b37f-3b7e69111e83">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e643bc36-f6b0-4fab-992f-a21ecb572755">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40bbf109-b0e8-4d3a-9db5-36372e0aa85a">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7e41ff9d-4c9e-4b11-b67c-e7d0c2f3049c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3b428b4-cf8e-4e5e-9cec-324c19c0cba5">(ASSERTION, ActCode [2.16.840.1.113883.5.4], "Assertion")</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41313254-65c0-4bfc-8bfe-b3087cc9b102">Purpose​Of​Reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b9ee29b5-9ed4-4fe4-a922-8921550ac0af">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_930552ee-b123-4fbf-b381-3a73b071780e">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_42f86020-8a92-4431-9099-f2738a7d2071">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d546aa3d-1b28-4e5f-85ee-a92fee38e290">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67b6d72b-3466-4eb8-86e3-18d0fb6183ad">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_14981cd9-b066-49de-ac64-9e45039df983">SHALL</para>
              <para xml:id="para_4efb69de-8fd4-4842-9ab2-9d9c81c5478d">CWE</para>
              <para xml:id="para_aabc9f37-9f03-40c7-ac0e-09a79a9cae40">DYNAMIC</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a0e62444-e71e-445c-a39e-65f944f4221f">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_7003" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8dbd589a-175b-4843-a4fe-5a3abf3963ad">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7f1888f0-42cb-4a00-93fb-3d1dd953c156">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dbf9673d-fe52-4346-9846-2ff71c19cfe7">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89d24dc8-0087-48b7-b582-56439c3cec4f">COND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54add960-3fb3-4b1e-9a05-34ece3c6a1cd">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97423ae4-7dfc-4f70-b0eb-9185ed2d5bcd">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_590f8d21-1cbb-4f39-a63f-047dfa021fa5">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a2f8c6bf-089d-4ce2-901c-dd0582604721">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_04e1c1f6-0127-43a6-ab58-b4d675de8bff">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_93918df2-7b43-43fd-8ea7-c2638e759936">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9eeafc89-3afe-4a27-af82-c94b94f48f23">COMP</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47d94fa0-c235-4ba5-9cf1-581dd5dda66e">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b194a61-b7ff-4873-9a7b-344c633d8332">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0bb28183-ee7b-4827-b30b-c0a764c1034e">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee1b11aa-0abc-4fab-843a-57b5ff4c5764">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a80788b2-cdd1-4273-b897-db9078690f97">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_42a6e962-1740-49de-99cc-aba7917e5d6f">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d00c7cd2-b8b9-4760-b8d0-40e30c4a1233">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_489d61d2-5dca-4b88-8ada-2202eafe8626">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d3e8ac5b-1614-4e41-a1cc-ae61fcc815a0">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_49e78660-e65d-4530-9257-606a6b4b3275">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9f687576-3e16-4d1e-ac47-a4f512c78194">ROIBND</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_91308499-03f2-43a4-b300-bc13f165b2f6">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6278b442-f241-4e13-9ab1-0e04f3340014">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3839237b-b91f-4b12-a9fe-204be0df691c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2494a910-960d-4e52-a216-92bb51fa8ecb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4f7b2951-f5a8-492e-bd25-5515db3b9c4a">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8523d969-5ad6-4b02-bc9d-499278f625f2">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67fb08ad-b7db-4487-b8c9-c483a3bc365c">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_62ad9e8c-665a-4e56-b02c-64d8b24f23c7">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e30dacd4-eb2d-44cc-83ae-1cc69b3c53f1">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0fc1810c-fd06-4183-abdd-8bbf5f344d39">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_50934c60-c9ec-4261-9d1e-679e04036348">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_498a7637-1766-41fb-ba53-d82e9a1acc46">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7d3c1735-ff36-41ed-bb19-4f09c9694f37">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a1f5b02-2eb3-4ebb-9c27-9ad8eb7997ee">
<olink targetdoc="PS3.16" targetptr="DCM_121190">(121190, DCM, "Referenced Frames")</olink>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_28a8576d-067c-4a74-9260-8f064e09f745">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9299c7d7-8194-4bfa-9b38-d61e97457aaf">entry​Relationship</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_33bd2e52-cd4d-4fe6-83fc-08a136fd3bc4">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2b162e18-a38c-428a-b5f1-af6dbb5ce8f5">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87e8a1c5-9ca2-4156-b522-3e0452baa962">&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_475b7a26-4908-4456-a81e-fa1293d0fd7e">@typeCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4ad78550-b55c-405d-b105-8a58bd0d6bf0">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_30ecbf3d-4454-48d5-a494-631da0a4bf82">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d4c0c9c4-cd3a-48b7-967f-b9b2ee9902ae">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aa1e3cb1-379a-4924-b8ec-e036f266fca7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bebb7ddc-4dab-4cf6-b620-dcbe2af2c626">COMP</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_febdca13-5dc4-4049-a170-34c05596a7dd">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0b1fa515-13bb-4338-96e0-827089e8e464">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_85cb34c8-6a75-4727-84ce-9b42bbad9c48">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d9ab66b-6d38-42de-ac69-acc8a2fa508b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3ebc931b-2da0-46c0-aefe-b3731b3840f1">&gt;&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9805e927-ddf5-410a-b068-cb6ffc2833fd">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fe19c10-5c91-491b-81da-40ed0854402b">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7a656625-c5cb-4ad5-8b2c-b87a20bdb841">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77186c10-afdb-469f-a946-34ca014a2820">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bb3a08fc-876f-4d4b-9a67-9c9bd1fb165a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c915b0a-bf5f-4065-a593-9c18462e2c7c">OBS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d7b145f1-401e-46b0-b645-742ad86affac">&gt;&gt;&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c5f7888-8bbd-4ff6-a17a-936dfee0930f">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5ac3f040-a17b-4351-9212-fd59687b074a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e0e8831a-f152-4ac6-a128-4227983f9279">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a14206ad-4983-492c-9976-56b0b1e901f9">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3bbf6fd9-6c9d-4861-83dd-5f4464796870">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_198fc9d3-2cc5-498a-94db-d4c626ba9ebc">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6ecb0c34-02df-4f73-ab14-fe8b62c5195a">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81acd3f4-5570-41dd-ad7c-20e0674a9484">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1fc39fc0-ebfd-405a-9a77-3615ce22f1d8">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_075716ed-fe54-4679-994e-6e9d59b45afc">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41721889-e87e-4b1a-bf98-2391aeec85f5">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_77d3338b-638f-4425-9d80-dcc0025c17e8">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_89228340-1306-4069-bbb1-8d3400aeb476">
<olink targetdoc="PS3.16" targetptr="DCM_113036">(113036, DCM, "Frames for Display")</olink>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_82ff13a0-d396-40ed-93b7-b1f372ed8d44">Referenced​Frames</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cfd039a-ae28-41ff-910a-739e52ec752b">&gt;&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_88410748-7b1c-4b69-8690-ca2e9cf28839">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c503c83-3793-4da0-b6a3-bbefe28ff045">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b583557d-198e-4dde-95ea-d098a089f73b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dccdba49-2f0e-4822-97ff-6b7864f0d4e0">LIST &lt;INT&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.8.1" status="3" xml:id="sect_10.8.1">
        <title>entryRelationship</title>
        <para xml:id="para_8ab5f3af-e1c6-41c4-9fcb-9659d96bf18e">COND: entryRelationship SHALL NOT be present in a <xref linkend="sect_10.8" xrefstyle="select: labelnumber title"/> included within a <xref linkend="sect_9.8.7" xrefstyle="select: labelnumber title"/> section, and MAY be present otherwise.</para>
        <section label="10.8.1.1" status="4" xml:id="sect_10.8.1.1">
          <title>entryRelationship/@typeCode=SUBJ (SOP Instance)</title>
          <para xml:id="para_306ae8b0-8657-4518-913b-4d360da6031b">This template recursively invokes itself to allow a Presentation State SOP Instance reference to identify the target Image SOP Instances, or for a derived Image to reference its source Image, or similar linkages between instances.</para>
          <note>
            <para xml:id="para_2a7cb375-03ce-430e-b5ef-5717c981049d">This is generally not required, as the DICOM SOP Instance itself identifies relationships to the relevant other SOP Instances.</para>
          </note>
        </section>
        <section label="10.8.1.2" status="4" xml:id="sect_10.8.1.2">
          <title>entryRelationship/@typeCode=RSON (Purpose of Reference)</title>
          <para xml:id="para_519dc781-6d21-4723-a1c7-7cc6294dc09a">A Purpose of Reference Observation describes the purpose of the DICOM composite object reference. Appropriate codes, such as externally defined DICOM codes, may be used to specify the semantics of the purpose of reference. When this observation is absent, it implies that the reason for the reference is unknown.</para>
          <note>
            <para xml:id="para_66f96de9-9f34-4f01-9310-cd5e4732eec0">In Consolidated CDA r1.1, this was defined using a separate "Purpose of Reference Observation" template, which is included directly in this template specification.</para>
          </note>
        </section>
        <section label="10.8.1.3" status="4" xml:id="sect_10.8.1.3">
          <title>entryRelationship/@typeCode=COMP (Referenced Frames)</title>
          <para xml:id="para_0aee7d91-4e3a-411b-bd8e-c52af5a9e194">A Referenced Frames Observation contains a list of integer values for the referenced frames of a DICOM multi-frame image SOP instance. It identifies the frame numbers within the referenced SOP instance to which the reference applies. The observation identifies frames using the same convention as DICOM, with the first frame in the referenced object being Frame 1. A Referenced Frames Observation must be used if a referenced DICOM SOP instance is a multi-frame image and the reference does not apply to all frames.</para>
          <note>
            <para xml:id="para_cfb75547-f782-4404-8b62-61d9400fc8bf">In Consolidated CDA r1.1, this was defined using separate "Referenced Frames Observation" and "Boundary Observation" templates, which are included directly in this template specification.</para>
          </note>
        </section>
      </section>

      <example label="10.8-1" xml:id="example_10.8-1">
        <title>SOP instance observation example with purpose of reference</title>
        <programlisting><![CDATA[<observation classCode="DGIMG" moodCode="EVN">
    <templateId root="1.2.840.10008.9.18"/>
    <id root="1.2.840.113619.2.62.994044785528.20060823.200608232232322.3"/>
    <code code="1.2.840.10008.5.1.4.1.1.1"
        codeSystem="1.2.840.10008.2.6.1" codeSystemName="DCMUID"
        displayName="Computed Radiography Image">
    </code>
    <text mediaType="application/dicom">
        <reference value="http://www.example.org/wado?requestType=WADO
            &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
            &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823223142485051
            &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232232322.3
            &amp;contentType=application/dicom"/>
        <!--reference to image 1 (PA) -->
    </text>
    <effectiveTime value="20060823223232"/>
    <entryRelationship typeCode="RSON">
        <observation classCode="OBS" moodCode="EVN">
            <templateId root="2.16.840.1.113883.10.20.6.2.9"/>
            <code code="ASSERTION" codeSystem="2.16.840.1.113883.5.4"/>
            <value xsi:type="CD" code="121112"
                codeSystem="1.2.840.10008.2.16.4"
                codeSystemName="DCM"
                displayName="Source of Measurement"/>
        </observation>
    </entryRelationship>
</observation>]]></programlisting>
      </example>
    </section>
    <section label="10.9" status="2" xml:id="sect_10.9">
      <title>Image Quality</title>
      <informaltable frame="box" rules="all">
         <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_059876e7-0013-4a6d-b303-3185ebeb93d9">
                <emphasis role="bold">Template ID</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dbf00f65-59f5-4ed8-acc8-b2985668f48a">1.2.840.10008.9.15</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f7a2a213-a7d8-430d-84cb-71836aa8df4b">
                <emphasis role="bold">Name</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4d36b1bf-0526-41d6-a0e8-fdf0c2a0495c">Image Quality</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_17a7974a-21e5-4c10-bfba-73963d07fafb">
                <emphasis role="bold">Effective Date</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_888ee886-32b6-4efb-90ff-e61eec1881c4">2015/03/24</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_059ca964-1072-4bd1-acee-c5beb1af6ea9">
                <emphasis role="bold">Version Label</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3da633d7-c7fd-4352-8a1c-10dfa4be00ce">DICOM-20150324</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d0936b3a-40bf-45b9-9827-b621c28d78b3">
                <emphasis role="bold">Status</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ed2fd7f0-39e8-45bf-a1c0-602043903196">Active</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_35e08f9a-613d-4d80-a875-e30fe7b376ce">
                <emphasis role="bold">Description</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f37f9041-64d3-46c6-b824-096b531b2c97">Provides a quality assessment for the image set identified by the invoking section. By default unless otherwise identified, applies to the image set interpreted by the document (typically a Study). If the quality rating applies to only a subset of the Study (e.g., a Series, or a specific Image), that subset shall be identified in the invoking section.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ab9daeda-8066-4bae-b2ac-36508ac2c2e7">
                <emphasis role="bold">Classification</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b6b587f8-1204-4d22-8f8d-c13f4b23cbdd">CDA Entry Level</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b126170d-2035-4f39-b350-e38bc917d98a">
                <emphasis role="bold">Relationships</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e369bf8e-031f-48e6-a258-ac3ef6949ff7">Included in <xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>
</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fcedabd0-2a89-4918-8e42-899ce03e80d8">
                <emphasis role="bold">Context</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cf45ba32-e90e-4b98-8a00-db18ee8c71cb">parent node</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_823c667c-af34-4c97-9c03-38a169482c63">
                <emphasis role="bold">Open/Closed</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c5bd0e20-1d13-4552-ab6f-430337d33791">open</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1cd92a04-7043-4163-81fc-7b5de4c84639">
                <emphasis role="bold">Revision History</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_41fec01e-44bc-4a6f-988f-14646b133c85">DICOM-20150324: Initial version Derived from Coded Observation</para>
            </td>
          </tr>
        </tbody>
      </informaltable>
      <para xml:id="para_dd1fe38a-59b6-43fa-ab82-29b27e7c0576"/>
      <informaltable frame="box" rules="all">
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_07d4151f-4c54-420e-bbb1-b57c725819ce">Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_12efface-d39e-4a2a-b88d-9b6087e380ad">Nest Level</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_92a2cbc0-dd2f-42e4-877a-f1a487ce7697">Element/​Attribute</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_4110ec6b-7c7f-42d3-8308-94d992bbf91c">Card</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_0fb8e4bc-697a-4e9c-8e40-e2ab80d18d8c">Elem/Attr Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_dc6e2e2b-fdbd-4dc8-abb4-f5e7a7170b1a">Data Type</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_d9a8226c-edca-42d3-81f8-a1183a557233">Value Conf</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_81cd0f40-38ad-4eaa-9248-e4f2d0ce3697">Value</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_2d71fbaf-af4e-43ff-8105-8b66f71177f1">Subsidiary Template</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_df37b87e-c93c-4375-9518-37906f1c74a1">
                <emphasis role="bold">Image​Quality</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cd7386eb-652b-4ed7-ab9e-f0923848ede5">observation</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fab58519-4781-4b1d-b977-22d9f9297558">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fd19a1c-a94c-475f-b1e3-9fb9459d2b5b">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d192117-0ed3-497c-a2f7-4a057760f5bb">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1dcb7210-65e2-453a-bc3c-8b0845ecd03d">@classCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1e024946-382b-4132-88e2-58b92922310f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_356fec71-3f0d-4513-a605-68d9021a47eb">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aef38cad-bd98-4ea1-a962-4916d8f0372e">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_eb41059d-d30e-4447-abc0-1e80bd962079">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5495fb01-84c9-484f-82b6-eeeadcf72f98">OBS</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_87f56596-8c8f-4c40-bb6c-5e012f33a3c6">@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74930686-125f-497b-b2cf-ff1e1eccc62f">@moodCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de400362-d430-4493-8814-4311b8dcf82f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5515f193-4321-4c60-81ff-af5a5ab0ccd0">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2d1a9955-ef86-4fe2-9d0d-1de8538033c6">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0cb928f0-1940-4eee-b84b-17df6bba9092">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_acccd433-e5b8-4b0b-a5ab-f8a7b2310311">EVN</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fc57de57-252b-491c-81ff-09e696fec79c">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_385dffc7-a4d3-426d-8060-526a5de91877">template​Id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_cfc64e95-0bc2-4c49-abd0-b2c42c2aa63a">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f86791ba-dd89-456b-9856-746f0b6da219">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_11635331-9139-4afb-8701-fd9512987744">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe49852c-c63f-4311-8bff-a65767059b1c">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4e868445-5acf-46e7-aaaa-aabba7a43d1e">@root</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf2b60bb-d0b0-4930-a6a5-42c1645ab46f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_64d7c63b-fb7d-497f-9c5b-57ae312a2017">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54fce639-8596-4e22-ac92-aeb983cb305a">UID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d28652d-d2c2-4f86-a239-0c1c4ed8ba0c">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0400415f-187f-434c-934b-2d7a1672e5a4">1.2.840.10008.9.15</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0ba60ce7-f0af-429d-b300-67c193d072ac">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_169d20e5-4df5-42d7-a4ba-a88ede674302">id</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_466472cf-5b0d-444f-b13e-cff620a61474">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_190a6c84-593c-48eb-8d5b-342abed678d1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0dae981a-0e47-453e-8632-c81c3d7fd8a2">II</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e5a16da-3424-47d2-9716-ad0b68cc7b1d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_516ef9ae-eb73-46fc-a541-c6a0e50a42b4">code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d9a53e49-1342-4af0-a731-5f69b605476f">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b1cfb783-e4a2-46c6-9321-aae7d94437d1">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3c4fc256-8b4f-48e9-b17f-4f7bbe395523">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1f0df39d-b33d-426f-a22e-52b4415a2f21">
<olink targetdoc="PS3.16" targetptr="DCM_111050">(111050, DCM, "Image Quality Assessment")</olink>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_81f93dd3-10a5-43f5-8266-87bf0129c03d">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_56ce1a4c-cc37-4ba3-b8e0-92be2db40af3">text</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7e591be8-bfcc-4e15-b942-83572be3744f">0..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2734e31-114d-4c2a-8cd8-6ee59a292e6e">SHOULD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec7adb68-30da-4571-930f-114cf288eaca">Ref</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0e52f27-9144-4882-a28d-ad3af522ddfb">&gt;&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8b2f5b2b-82ca-4e49-81ed-582634ab71ca">reference</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d38ba326-488c-4b2e-90c6-54a8d2c73312">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_de45c109-6ba0-417f-b9d3-f855f5d83b99">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74e4bfb4-dece-4866-b96b-a643339f1ec8">URL (XML IDREF)</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6487cf49-d08c-41d2-b2f6-d93f49de8a26">#
                <emphasis role="italic">content​Ref</emphasis>
              </para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8a6d6bc2-d9c9-4622-aa26-ee12a0a057db">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0adcdbcb-d087-4dac-9f64-c6968a0e01e4">status​Code</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e905601b-a704-4de0-8d80-7c8e8c7995eb">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_253194da-064b-4edd-ae2f-d8355b2d5422">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c2889b9e-6c37-4f3d-97d4-32c5e3b08cc2">CS</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9e2f1d55-5c7f-4e11-9e31-95c1308ed33d">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d2d4478e-9991-4123-bf92-8d9c54f70e2f">COMPLETED</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e5e77305-c7ce-4719-9a28-deacf6ff8426">Rating</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c3ad44fc-66ad-45e8-a65e-d0036588aa86">&gt;</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b97ce867-b69a-4d46-a10e-bc096fba2a7e">value</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ee426a21-4ba0-4e5e-b298-145dc1741904">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e8dba809-9526-4895-b792-95ad8e9dd24a">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c3f65655-17e6-4331-8e45-35fd5852b33e">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2c86f6ea-1944-4a08-a517-d081ad4aa209">SHOULD CWE</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_061b5cf0-bd0f-4512-b434-a509eb8f43aa">
                ValueSet <olink targetdoc="PS3.16" targetptr="sect_CID_7036" xrefstyle="select: labelnumber quotedtitle"/>
</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1"/>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c71e03ef-ce02-491d-b91b-1edfeea712a1">&gt;@</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f351c26c-1a7d-4fa7-9d16-758c6dddfc63">@xsi:type</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f0a5c447-8481-4858-8311-266e7da2b58c">1..1</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_daa823c8-30ba-4bf8-9c53-4d146fde92e6">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ea0d1a2f-7dcc-45a9-ad71-6a40372dfec8">ST</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d81d69d9-d35d-4366-b0aa-8d7c8e6d66e7">SHALL</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ec7a03e1-1b08-487e-9107-39c1a36bba3f">CD</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
        </tbody>
      </informaltable>
      <section label="10.9.1" status="3" xml:id="sect_10.9.1">
        <title>text/reference and Related Narrative Block Markup</title>
        <para xml:id="para_dfdf061d-b5dd-4d0c-b4d6-2fafdbdd8dec">The Observation entry SHOULD include a text/reference element, whose value attribute (not to be confused with the value element of the Observation class) SHALL begin with a '#' and SHALL point to its corresponding narrative in the parent section (using the approach defined in CDA Release 2, section 4.3.5.1). See <xref linkend="sect_9.1.1.1" xrefstyle="select: label"/>.</para>
      </section>
      <example label="10.9-1" xml:id="example_10.9-1">
        <title>Image Quality example</title>
        <programlisting><![CDATA[<observation classCode="OBS" moodCode="EVN">
    <templateId root="1.2.840.10008.9.15"/>
    <id root="1.2.840.113619.2.62.994044785528.20060823.200608232232322.3"/>
    <code code="111050" codeSystem="1.2.840.10008.2.6.1"
        codeSystemName="DCM"
        displayName="Image Quality Assessment"/>
    <text>
        <reference value="#Q9"/>
    </text>
    <statusCode code="completed"/>
    <value xsi:type="CD" code="RID12"
        codeSystem="2.16.840.1.113883.6.256"
        codeSystemName="RADLEX"
        displayName="Diagnostic quality"/>
</observation>]]></programlisting>
      </example>
    </section>
  </chapter>
  <chapter label="A" status="1" xml:id="chapter_A">
    <title>SR Diagnostic Imaging Report Transformation Guide</title>
    <para xml:id="para_a139f670-24fa-4073-bc11-8cb1c6a0c78f">Retired. See PS3.20-2015.</para>
    <note>
      <para xml:id="para_55cadd03-e814-48dd-addd-a259061d1aa5">This Annex provided a transformation of SR documents based on <olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber title"/> to HL7 CDA Release 2 Imaging Reports based on the HL7 Diagnostic Imaging Reports (DIR) Release 1.0 Informative specification Template 2.16.840.1.113883.10.20.6.</para>
    </note>
  </chapter>
  <chapter label="B" status="1" xml:id="chapter_B">
    <title>SR Diagnostic Imaging Report Transformation Guide</title>
    <para xml:id="para_799dc8d5-b810-44b2-a809-abe82c75b8cf">Retired. See PS3.20-2015.</para>
    <note>
      <para xml:id="para_5290bcab-f22a-4a83-93bd-30200bb1f322">This Annex provided a transformation of SR documents based on <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber title"/> to HL7 CDA Release 2 Imaging Reports based on the HL7 Diagnostic Imaging Reports (DIR) Release 1.0 Informative specification Template 2.16.840.1.113883.10.20.6.</para>
    </note>
  </chapter>
  <chapter label="C" status="1" xml:id="chapter_C">
    <title>SR to CDA Imaging Report Transformation Guide</title>
    <para xml:id="para_2567f92f-fc22-4090-86b9-5550cc6bbccf">Constrained DICOM SR documents based on Imaging Report templates can be mapped to HL7 CDA Release 2 Imaging Reports based on Template 1.2.840.10008.9.1, as specified in <xref linkend="sect_7.1" xrefstyle="select: label"/>. The SR report templates to which this transformation applies include:</para>
    <itemizedlist>
      <listitem>
<para xml:id="para_22491787-7856-41f6-84d9-8ac0f2c4fd21">
<olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber title"/>
</para>
</listitem>
      <listitem>
<para xml:id="para_9bf4b36e-29bf-4f2a-9739-b6319208f207">
<olink targetdoc="PS3.16" targetptr="sect_TID_2005" xrefstyle="select: labelnumber title"/>
</para>
</listitem>
      <listitem>
<para xml:id="para_ae0c70d5-6afc-4a6a-b15e-772b3d68db2a">
<olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber title"/>
</para>
</listitem>
    </itemizedlist>
    <para xml:id="para_cf70ce4c-dc2a-4b42-a89e-8a86c52ad208">SR instances based on other templates may also be able to be mapped using the transformations in this Annex.</para>
    <para xml:id="para_aba4933c-18f3-41fb-9c60-72ed932cb62a">SR documents can be thought of as consisting of a document header and a document body, corresponding to a CDA document header and body. The header includes the modules related to the Patient, Study, Series, and Equipment Information Entities, plus the SR Document General Module, as specified in <olink targetdoc="PS3.3" targetptr="PS3.3" xrefstyle="select: labelnumber"/>. The SR Document Content Module contains the content tree (structured content) of the document body. Note, however, that DICOM SR considers the root content item, including the coded report title, and some context-setting content items as part of the document body content tree, but these constitute part of the CDA header. See <xref linkend="figure_C-1" xrefstyle="select: label"/>.</para>
    <section label="C.1" status="2" xml:id="sect_C.1">
      <title>Constraints</title>
      <para xml:id="para_097bbcdd-bbc2-42dc-91a6-9c976cea8c59">This Annex defines the transformation of an Enhanced SR SOP Instance to a CDA instance. The following constraints apply to such SOP Instances:</para>
      <itemizedlist>
        <listitem>
          <para xml:id="para_3b01f3bf-28fe-4c64-9d4b-e95272fe4978">Observation Context: The mapping does not support changing the observation context for the report as a whole from its default context, as specified in the Patient, Study, and Document Information Entities (see <olink targetdoc="PS3.3" targetptr="sect_C.17.5" xrefstyle="template:PS3.3 Section %n “%t”"/>)</para>
          <note>
            <para xml:id="para_835726fb-2b9a-4086-a2de-70c44268e99e">
<olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.16" targetptr="sect_TID_2005" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber"/> specify inclusion of <olink targetdoc="PS3.16" targetptr="sect_TID_1001" xrefstyle="select: labelnumber title"/> as Mandatory, but <olink targetdoc="PS3.16" targetptr="sect_TID_1001" xrefstyle="select: labelnumber"/> has no content if all aspects of context are inherited, as under this constraint.</para>
          </note>
        </listitem>
        <listitem>
<para xml:id="para_6b115712-7727-4e49-b63e-fed45cd4b2d2">Subject Context: The mapping does not support the subject of any of the report sections to be a specimen <olink targetdoc="PS3.16" targetptr="sect_TID_1009" xrefstyle="select: labelnumber"/>), a device (<olink targetdoc="PS3.16" targetptr="sect_TID_1010" xrefstyle="select: labelnumber"/>), or a non-human subject. Only a fetus subject context is supported for a Findings section.</para>
</listitem>
        <listitem>
<para xml:id="para_1c8f9317-eb37-401f-8377-a3693ca7d3f4">Procedure Context: The mapping allows identification of a different procedure than the procedure identified in the SR Study IE only as context for a Prior Procedure Descriptions section.</para>
</listitem>
        <listitem>
<para xml:id="para_b1883b77-5585-4276-aca5-4576587aebd0">De-identified Documents: There is no CDA implementation guidance from HL7 for de-identified documents, other than general rules for using the MSK null flavor (see <xref linkend="sect_5.3.2" xrefstyle="select: label"/>). There is no CDA capability equivalent to the Encrypted Attributes Sequence (see <olink targetdoc="PS3.3" targetptr="sect_C.12.1.1.4.1" xrefstyle="template:PS3.3 Section %n “%t”"/>) for carrying encrypted re-identification data.</para>
</listitem>
        <listitem>
<para xml:id="para_0a695233-139f-4831-983c-5dc8ba068974">Patient Study Module: Medical or clinical characteristics of the patient specified in the Patient Study Module are not mapped (see <olink targetdoc="PS3.3" targetptr="sect_C.17.5" xrefstyle="template:PS3.3 Section %n “%t”"/>)</para>
</listitem>
        <listitem>
<para xml:id="para_ec51613d-586c-4e24-939d-4bbcbf205325">Clinical Trials: Template 1.2.840.10008.9.1 does not define attributes for clinical trials equivalent to those of the Patient, Study, and Series IEs (Clinical Trial Subject Module, Clinical Trial Study Module, Clinical Trial Series Module).</para>
</listitem>
        <listitem>
<para xml:id="para_3c10a295-0cb7-48ce-9158-bd8ffc9589c5">Spatial Coordinates: The mapping does not support SCOORD observations. As CDA documents are principally for human reading, detailed ROI data is presumed to reside in the DICOM SOP Instances of the study, or in images ready for rendering with a Presentation State, not in the CDA report. Template 1.2.840.10008.9.1 does not support the CDA Region of Interest Overlay entry class (see <xref linkend="sect_9.1.2.4" xrefstyle="select: label"/>).</para>
</listitem>
      </itemizedlist>
      <para xml:id="para_d53297ed-a134-4bce-b717-2a9a871648a1">
        <figure label="C-1" pgwide="1" xml:id="figure_C-1">
          <title>TID 2000 Structure Summarized from PS3.16, and mapping to CDA</title>
          <mediaobject>
            <imageobject>
              <imagedata fileref="figures/PS3.20_C-1.svg"/>
              <!--<imagedata fileref="sup155_fromword_files/image005.png"/>-->
            </imageobject>
          </mediaobject>
        </figure>
      </para>
    </section>
    <section label="C.2" status="2" xml:id="sect_C.2">
      <title>Conventions</title>
      <para xml:id="para_78d76b50-a5fc-4ed2-ac91-6425f06c7215">Literal values to be encoded in CDA elements are represented in the mapping tables in normal font, as a string, or as a coded value triplet:</para>
      <itemizedlist mark="none">
        <listitem>
<para xml:id="para_43a22772-b028-42e2-a67b-78fcd6bb6101">"NI"</para>
</listitem>
        <listitem>
<para xml:id="para_ca6ebeed-dd49-482e-bc01-69c7ae0a074f">(codeValue, codingScheme, codeMeaning)</para>
</listitem>
      </itemizedlist>
      <para xml:id="para_d12cc04d-6a96-4f85-a1b5-362e660d15dd">Conventions for mapping from DICOM attributes in the transformed SR are described in <xref linkend="sect_5.2.8" xrefstyle="select: label"/>.</para>
      <para xml:id="para_a70c878a-f89c-475f-bd8d-e99715ad52c2">Data mapped from an SR Content Item is identified by the Concept Name of the Content Item, represented in the mapping tables as a triplet in italic font:</para>
      <itemizedlist mark="none">
        <listitem>
<para xml:id="para_8cf5e0d9-f2e2-4fa0-9b83-e0df40aacc6a">
<emphasis role="italic">(codeValue, codingScheme, codeMeaning)</emphasis>
</para>
</listitem>
      </itemizedlist>
      <para xml:id="para_fd95ed1a-94a5-4100-bbef-99231ee46be3">Data mapped from a specific Attribute in an SR Content Item uses the triplet to identify the Content Item, with the &gt; character and the specific attribute name and tag:</para>
      <itemizedlist mark="none">
        <listitem>
<para xml:id="para_e7b1f40c-dc07-408d-8192-efd4e2e9f4ed">
<emphasis role="italic">(codeValue, codingScheme, codeMeaning) &gt; Attribute Name (gggg,eeee)</emphasis>
</para>
</listitem>
      </itemizedlist>
      <para xml:id="para_5557a834-f956-4124-a00f-98c872e556fa">Additional notes are within square brackets:</para>
      <itemizedlist mark="none">
        <listitem>
<para xml:id="para_7b51038a-9a12-4c48-81b4-8cdf7d392b95">[Note]</para>
</listitem>
      </itemizedlist>
      <para xml:id="para_65b7c60e-bc3f-4489-808b-0c6a8c0a8602">Mandatory CDA elements for which there is no corresponding source data in the SR SOP Instance may be coded with a nullFlavor attribute (see <xref linkend="sect_5.3.2" xrefstyle="select: label"/>).</para>
    </section>
    <section label="C.3" status="2" xml:id="sect_C.3">
      <title>Header Transformation</title>
      <para xml:id="para_b4d41170-df8a-43f5-b0e4-7c280802740b">For transformation of the SR content into the CDA header, the target elements of the CDA instance are listed in <xref linkend="table_C.3-1" xrefstyle="select: label"/> by their Business Names, together with the recommended source in an SR instance. This allows the transforming application to "pull" the relevant information from the SR to populate the CDA header.</para>
      <table frame="box" label="C.3-1" rules="all" xml:id="table_C.3-1">
        <caption>CDA Header content from SR</caption>
        <thead>
          <tr valign="top">
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_3b39436c-e27b-4359-9330-b1b75a5811c8">CDA Business Name</para>
            </th>
            <th align="center" colspan="1" rowspan="1">
              <para xml:id="para_b01f1559-cc48-43bd-8762-0278cf5447ed">DICOM SR</para>
            </th>
          </tr>
        </thead>
        <tbody>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_08b0ee8c-e4ea-4e76-a658-6e24285d1118">ImagingReport: DocType</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9bd220dd-23ac-40c2-a750-04d0d8b23731">
                <emphasis role="italic">Concept Name Code Sequence (0040,A043)</emphasis> [of the root content item]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_54b93f5b-e553-4de2-b9e0-0323fc3cf69d">ImagingReport: ContentTemplate</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ccf2ea56-b027-46ab-ba88-f6c7ce7d2030">ImagingReport: DocumentID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5edbdadd-00d2-41b5-8154-8a3303e029ae">ImagingReport: Title</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9aa030d8-221b-4941-ac3c-37f5a235ceac">
                <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121050">(121050, DCM, "Equivalent Meaning of Concept Name")</olink> &gt; Concept Code Sequence (0040,A168) &gt; Code Meaning (0008,0104)</emphasis> if present; otherwise
                <emphasis role="italic">Concept Name Code Sequence (0040,A043) &gt; Code Meaning (0008,0104)</emphasis> [of the root content item].</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b2d643da-a27a-4482-8a8f-2b3c546379e3">ImagingReport: CreationTime</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_72f73e5b-ae2e-4ca7-be09-e84302a5ea22">
                <emphasis role="italic">Content Date (0008,0023) + Content Time (0008,0033) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_acb6b308-f837-4bed-b537-ee09b2430b16">ImagingReport: Confidentiality</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_00725f78-9da0-4114-873d-eb250207515b">ImagingReport: LanguageCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_42c0a940-e48d-424b-8b0e-542316f1f368">
                <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121049">(121049, DCM, "Language of Content Item and Descendants")</olink>
</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_075d30a6-4aed-4b65-84a0-ffced930dca7">ImagingReport: SetId</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_67c03870-928f-4e93-a701-6c477b1f4274">ImagingReport: VersionNumber</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f7257f62-cfaf-430a-b654-f1115744ff36">ImagingReport: Patient:ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8264788c-8c0f-45f9-9876-fd17cad9bf34">
                <emphasis role="italic">Patient ID (0010,0020)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1f49b1c-58d2-4621-bb5c-eb3d681e68c2">ImagingReport: Patient:IDIssuer</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_332c3abc-cb5c-4348-b9e8-66d420817963">
                <emphasis role="italic">Issuer of Patient ID Qualifiers Sequence (0010,0024) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4c274e67-7fda-492d-8767-98021601eb04">ImagingReport: Patient:Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f1d801b9-4618-44f2-b2af-360a308325e0">
                <emphasis role="italic">Patient's Address (0010,1040)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b1c1f717-b4c6-44be-8ef8-0ab9d544531d">ImagingReport: Patient:Tele</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_939e37e2-0b27-4b1e-9a43-5d8d53df8611">
                <emphasis role="italic">Patient's Telephone Numbers (0010,2154)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e31bc9b-1d14-4dca-84e7-bc7b3e717a01">ImagingReport: Patient:Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2252067c-2a4c-4d6b-a99a-3773e4d36ca8">
                <emphasis role="italic">Patient's Name (0010,0010)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0e5a5503-b921-4197-b2ea-d6fc0230f352">ImagingReport: Patient:Gender</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d8f3f6b-98b7-4f68-979e-0d0f132b4052">
                <emphasis role="italic">Patient's Sex (0010,0040)</emphasis>
              </para>
              <para xml:id="para_beff3318-4cc4-40f6-a69f-30703830a84d">[Map value "O" to nullFlavor UNK]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_95af8718-4098-4dc5-bb8a-95b9f358b6bf">ImagingReport: Patient:BirthTime</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_368a8f64-e977-42b8-a5fa-a2bc560b5cfa">
                <emphasis role="italic">Patient's Birth Date (0010,0030) + Patient's Birth Time (0010,0032)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7c683f3e-36e2-4f5d-a94d-0aa1bdb12aa3">ImagingReport: Patient:ProviderOrgName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fd43e64e-c44e-4c89-a6df-af6f49570178">
                <emphasis role="italic">Issuer of Patient ID (0010,0021)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_16d0e9c9-bd85-4db5-8cf4-278ca0d8ca8d">ImagingReport: Patient:ProviderOrgTel</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c57a69c1-2616-4668-9b36-436fb097093f">ImagingReport: Patient:ProviderOrgAddr</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5f1cd6d1-cdc8-4b24-83c9-88894b8b6cb4">ImagingReport: SigningTime</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_643823f6-01eb-48f1-a267-4f3883a68218">
                <emphasis role="italic">Verifying Observer Sequence (0040,A073) &gt; Verification DateTime (0040,A030)</emphasis>.</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_40c5ed3c-562e-46b5-a5f9-1f7e56a7a021">ImagingReport: SignerID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_66a1ff24-c413-4609-9b2e-a5cbc59e8412">
                <emphasis role="italic">Verifying Observer Sequence (0040,A073) &gt; Verifying Observer Identification Code Sequence (0040,A088)</emphasis> [code value as identifier]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2e39bd30-d2d1-4050-a7e7-757271d63046">ImagingReport: SignerAddr</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_7b2dee90-1c74-4163-ba7e-ea4bb8655d90">ImagingReport: SignerTel</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bf8fc897-54f9-432e-afd9-7c712da1d3ed">ImagingReport: SignerName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f31bf7e2-ab51-4373-b0a0-f66ffbeb8f67">
                <emphasis role="italic">Verifying Observer Sequence (0040,A073) &gt; Verifying Observer Name (0040,A075)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_bd24a4ff-1602-45e6-a0c6-914cf1c087e2">ImagingReport: SignatureBlock</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c0fe4e82-ad9c-45c8-87f3-cfb19c891ea3">ImagingReport: Author:AuthoringTime</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6e3a06e0-ea8e-4624-93ae-1bc856ac92ad">
                <emphasis role="italic">Content Date (0008,0023) + Content Time (0008,0033) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6d48d60d-ee32-4b80-b09b-081a652a8cfe">ImagingReport: Author:ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_85c7934d-24a1-4d4e-93e7-78d786b0c8bd">
                <emphasis role="italic">Author Observer Sequence (0040,A078) &gt; Person Identification Code Sequence (0040,1101)</emphasis> [code value as identifier]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad059ce9-075e-435b-a239-a8ac5e52ea05">ImagingReport: Author:Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac54a168-7e8e-480b-a016-3e1908d02923">ImagingReport: Author:Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9a5485e6-b177-46b7-9b86-f9e55ed954de">ImagingReport: Author:Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5b522eda-abc4-42f3-90cf-adff79898580">
                <emphasis role="italic">Author Observer Sequence (0040,A078) &gt; Person Name (0040,A123)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_aac3497e-82d7-48f4-8b7a-c4993a9818c6">ImagingReport: Recipient:Addr</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_be12a739-3f66-47e2-80c7-5ec714919c70">ImagingReport: Recipient:Tel</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c67ffdf9-a5cd-4f09-82aa-4c389c88281d">ImagingReport: Recipient:Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8c382913-3dec-4b64-b733-ef2e927fcbf6">ImagingReport: Recipient:Org</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d319e7a-d5fe-43e2-a6d7-a74f198db000">ImagingReport: CustodianOrgID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_94b4b827-49c6-4415-b99e-d5cf4bfbd535">
                <emphasis role="italic">Custodial Organization Sequence (0040,A07C) &gt; Institution Code Sequence (0008,0082)</emphasis> [code value as identifier]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_f5e0a5a7-203e-4cd5-a41e-84d8aaffa03c">ImagingReport: CustodianOrgName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_faf0a480-c204-4f80-9043-db4d8fa01bef">
                <emphasis role="italic">Custodial Organization Sequence (0040,A07C) &gt; Institution Name (0008,0080)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9b4e3842-e81b-44da-8d28-a5c22cf26adc">ImagingReport: CustodianOrgAddr</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3b5d4de0-e77e-4b2c-aad2-087aacadc30b">ImagingReport: CustodianOrgTel</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2a1f410e-c637-4849-a053-3c5076ae9861">ImagingReport: EncounterID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_332da269-74ad-4fb7-9d3e-0d99a95742e7">
                <emphasis role="italic">Admission Id (0038,0010)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8fccfab4-8657-4790-ab77-8cbf883ed129">ImagingReport: EncounterIDIssuer</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_1c1e1865-2de9-4679-8052-18b67b132fb8">
                <emphasis role="italic">Issuer of Admission ID Sequence (0038;0014) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_211d4673-d38a-4de6-9556-dc63707df037">ImagingReport: EncounterTime</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_0c022bae-07c4-4354-ad95-6f547f705b76">ImagingReport: HealthcareFacilityName</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c482d361-e589-4bf6-ac7c-0521d414cbe1">ImagingReport: HealthcareFacilityAddress</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_665ae786-cb6e-496b-b88f-e112c722554e">
                <emphasis role="italic">Institution Address (0008,0081)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_d80766b1-25d0-43eb-9871-cc82eea73e71">ImagingReport:HealthcareProviderOrganizationName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_24ee37b1-6340-4115-8861-ec0dbe595302">
                <emphasis role="italic">Institution Name (0008,0080)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9aab2aae-c727-4a21-bc40-54a8c3cd9c4d">ImagingReport:AttendingPhysicianName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ce8348c7-091f-4865-aaf8-481a82baa2b3">
                <emphasis role="italic">Physician(s) of Record (0008,1048)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d4bfe65-79e5-4988-8317-838c1d979f9a">ImagingReport:OrderPlacerNumber</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_47e0b2b3-b931-45c8-9912-34c0b19b1892">
                <emphasis role="italic">Referenced Request Sequence (0040,A370) &gt; Placer Order Number/Imaging Service Request (0040,2016)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_4fe406d0-5b56-4ca1-91ab-348105e95188">ImagingReport:OrderAssigningAuthority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5146a55f-6d56-4590-8409-8dee7d110c60">
                <emphasis role="italic">Referenced Request Sequence (0040,A370) &gt; Order Placer Identifier Sequence (0040,0026) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9fc3bafa-c3dc-4084-aea7-aa8252a94f71">ImagingReport:AccessionNumber</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8016218d-0748-44ea-b5ea-0e9dfe70d1ea">
                <emphasis role="italic">Accession Number (0008,0050)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fe4219c1-312c-4a7e-af48-cdfc26ee6f2c">ImagingReport:AccessionAssigningAuthority</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b66d6d2a-55bc-4946-a482-2cfcf417eb22">
                <emphasis role="italic">Issuer of Accession Number Sequence (0008,0051) &gt; Universal Entity ID (0040,0032)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_2394eb3e-475e-4c0a-8891-6fa09de50f70">ImagingReport:OrderedProcedureCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a54a68ba-7f8e-4f86-9e57-5629a3d4daac">
                <emphasis role="italic">Referenced Request Sequence (0040,A370) &gt; Requested Procedure Code Sequence (0032,1064)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ba8b5a68-9af8-47a9-b494-1fecfb78fe4e">ImagingReport: OrderPriority</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_e3c7e0cf-e04f-40b5-a38a-b784d8cd226e">ImagingReport:Study:StudyUID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_36d196cd-5ad6-4f39-aa4f-17bceaf93f24">
                <emphasis role="italic">Study Instance UID (0020,000D)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8f4d6154-f34c-45bf-9b54-900ecc87c09a">ImagingReport:Study:ProcedureCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_02e9f320-78f9-410c-b1c2-acf0fbd375c4">
                <emphasis role="italic">Procedure Code Sequence (0008,1032)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3ee62c79-c907-4dc2-b8be-4f911a303f88">ImagingReport:Study:Modality</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3f643b88-0c7f-4fb3-8427-a11e92231bbf">
                <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_122142">(122142, DCM, "Acquisition Device Type")</olink>
</emphasis> or
                <emphasis role="italic">
<link xl:href="http://loinc.org/55111-9/" xl:show="new">
<link xl:href="http://loinc.org/55111-9/" xl:show="new">(55111-9, LN, "Current Procedure Descriptions")</link>
</link> &gt; <olink targetdoc="PS3.16" targetptr="DCM_122142">(122142, DCM, "Acquisition Device Type")</olink>
</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ac5609e9-5db9-4eb8-88e1-3fd864a8b839">ImagingReport:Study:AnatomicRegionCode</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_af5f9910-799d-419f-bd7f-56b287a6431f">
                <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_123014">(123014, DCM, "Target Region")</olink>
</emphasis> or
                <emphasis role="italic">
<link xl:href="http://loinc.org/55111-9/" xl:show="new">
<link xl:href="http://loinc.org/55111-9/" xl:show="new">(55111-9, LN, "Current Procedure Descriptions")</link>
</link> &gt; <olink targetdoc="PS3.16" targetptr="DCM_123014">(123014, DCM, "Target Region")</olink>
</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_ad5de2e4-fceb-413d-b5bf-b9247bd8a344">ImagingReport:Study:StudyTime</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_3d41c9b3-1fc6-447b-bc06-652f97b10d05">
                <emphasis role="italic">Study Date (0008,0020) + Study Time (0008,0030) + Timezone Offset From UTC (0008,0201)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_53eb0f7e-6074-42ec-a3e7-c80c8ed4a45c">ImagingReport: Performer:Type</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_39d29b7c-359b-4b15-863c-5d62a17710eb">ImagingReport: Performer:ID</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_6145f58d-780c-4e8e-b054-ee9b5256b0cb">ImagingReport: Performer:Name</para>
            </td>
            <td align="left" colspan="1" rowspan="1"/>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_876b21a1-93da-49c2-a656-90b2256a54e2">ImagingReport: ReferrerAddr</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_74d4d498-6b22-4899-9fd8-3a2159610c0e">
                <emphasis role="italic">Referring Physician Identification Sequence (0008,0096) &gt; Person's Address (0040,1102)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_fa296209-7b7d-47e6-b155-2811dbc8a771">ImagingReport: ReferrerTel</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_b63208ee-39c9-4eb6-b0aa-3c00e62756bc">
                <emphasis role="italic">Referring Physician Identification Sequence (0008,0096) &gt; Person's Telephone Numbers (0040,1103)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_c9759a60-29fa-4b0a-a4c1-0dc8c2a7f2c2">ImagingReport: ReferrerName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_97a4fbd6-6001-4548-9b87-c72db8af1893">
                <emphasis role="italic">Referring Physician's Name (0008,0090)</emphasis>
              </para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_5d3c3161-9afa-4bb1-b798-960b9b8d170a">ImagingReport: TranscriptionistID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_a281e5ef-8c89-40ef-a55b-3f4d7c6aaed2">
                <emphasis role="italic">Participant Sequence (0040,A07A) &gt; Person Identification Code Sequence (0040,1101)</emphasis> , [where Participation Type (0040,A080) equals "ENT" (Data Enterer); code value as identifier]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_9d264d01-52e5-4878-bf37-10799d1a8314">ImagingReport: TranscriptionistName</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_beeab96f-fbb3-419c-995a-e313c9d2a578">
                <emphasis role="italic">Participant Sequence (0040,A07A) Person Name (0040,A123)</emphasis> [where Participation Type (0040,A080) equals "ENT" (Data Enterer) ]</para>
            </td>
          </tr>
          <tr valign="top">
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_dca1c1a2-f680-4df1-adf2-b1b36ee4b284">ImagingReport: TransformedDocumentID</para>
            </td>
            <td align="left" colspan="1" rowspan="1">
              <para xml:id="para_8d134da6-90d1-436a-96e3-15f5de379b6c">
                <emphasis role="italic">SOP Instance UID (0008,0018)</emphasis>
              </para>
            </td>
          </tr>
        </tbody>
      </table>
      <para xml:id="para_1bf1afce-6600-4e3c-b616-8ea308896ab2">ImagingReport:Study:Modality and ImagingReport:Study:AnatomicRegionCode may be mapped from attributes in the root CONTAINER, if present there as in <olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber"/>, or in the Current Procedure Descriptions section CONTAINER, if present there as in <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber"/>.</para>
    </section>
    <section label="C.4" status="2" xml:id="sect_C.4">
      <title>Body Transformation</title>
      <para xml:id="para_d87f1194-a38f-4e1b-a434-62d090e3806d">For transformation of the body, this Section maps the SR content items to their target CDA elements. This allows the transforming application to traverse the SR content tree and construct equivalent CDA content.</para>
      <section label="C.4.1" status="3" xml:id="sect_C.4.1">
        <title>Section Mapping</title>
        <para xml:id="para_894b185b-0c71-488f-b1d2-dfd9634aa04d">SR <olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.16" targetptr="sect_TID_2005" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber"/> specify that imaging report elements are contained in sections, represented as CONTAINERs with concept name codes from <olink targetdoc="PS3.16" targetptr="sect_CID_7001" xrefstyle="select: labelnumber"/>.</para>
        <para xml:id="para_d208ecac-1439-4259-b4f0-2a4ef44f06aa">Each CONTAINER immediately subsidiary to the root CONTAINER shall be mapped to the section or subsection as specified in <xref linkend="table_C.4-1" xrefstyle="select: label"/>. Note that some SR document sections are mapped to subsections under CDA Template 1.2.840.10008.9.1.</para>
        <table frame="box" label="C.4-1" rules="all" xml:id="table_C.4-1">
          <caption>SR Section mapping to CDA</caption>
          <thead>
            <tr valign="top">
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_d65d201d-45f1-449c-b631-8e8679398933">Coding Scheme Designator</para>
              </th>
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_7504d911-1571-4fc0-8e0e-7b4969be262e">Code Value</para>
              </th>
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_06f1d2f5-e42e-4c81-b05d-02b4afbe6708">Code Meaning</para>
              </th>
              <th align="left" colspan="1" rowspan="1">
                <para xml:id="para_f42e7c2a-3c9a-4904-b6e9-812c0c60f6b5">Map to Template Section/Subsection</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_019bdd04-672f-49ac-a56b-e6a65ff457b3">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_72ab5b0c-8fea-4103-9a95-5b8b651174d8">11329-0</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71835fa2-3f93-4716-b790-06ffa2e3a885">History</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9cb950fd-e4f5-48f1-b6a5-188d624da8f1">
<xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.3" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_afe4f175-ce65-4db4-a6f6-c65e9a4f18d4">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c80c424-34a1-4b44-99b8-251ca155dbed">55115-0</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45c1db7f-f62d-4225-9241-55b89ac98b04">Request</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b141f7cb-d989-4b8c-86bd-21075214a51d">
<xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.1" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ad30604-2e9c-4c2a-9df3-6f765fd539a1">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1433397f-d90e-434a-a566-9c2e7facbbee">55111-9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1cd72c80-0e40-4c86-8e06-d8120e619236">Current Procedure Descriptions</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1bb893f0-0e59-4d6c-8364-137972660ca1">
<xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_48bf4660-128b-4791-8e76-add9cc24bf75">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed5acde8-2ecb-4e68-84b7-2b4d4800b326">55114-3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7f505a05-337f-4670-ab3f-36cf6ab340e7">Prior Procedure Descriptions</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d84ffeb0-aefb-42fa-84f4-13413cb4635b">
<xref linkend="sect_9.4" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_355757ed-fc53-4f5d-8fc0-64e78eab5875">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_caf32c54-8e63-4ac9-b2c9-def35d0d2317">18834-2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_069d8ebd-2ddf-45fc-9548-edbce08c9d0e">Previous Findings</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75448fa0-d3f8-4800-b747-8e3618cd759e">
<xref linkend="sect_9.4" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_857511e2-7974-4855-9665-c49ac4be6c6d">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30eae55c-5a9f-4864-909b-88e2f6ac3be3">18782-3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d35b17ec-562b-499c-b5a0-d9120c5994cc">Findings (Study Observation)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf80d9d1-8977-4842-b420-0ecf3c66d233">
<xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> or <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.8" xrefstyle="select: labelnumber title"/> (see <xref linkend="sect_C.4.1.3" xrefstyle="select: labelnumber"/>)</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_179b7f6d-e8b6-46a1-8db4-f90c504b7d47">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a6b679a-d660-4912-af5e-027c71c4d838">59776-5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_728ed6e8-91c2-4d15-85b4-24de19b73cca">Findings</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c09db0bc-2850-429e-a61d-d13919f4c29a">
<xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> or <xref linkend="sect_9.5" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.8" xrefstyle="select: labelnumber title"/> (see <xref linkend="sect_C.4.1.3" xrefstyle="select: labelnumber"/>)</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08887e3e-b8b3-4f03-9046-cd45b28c5ee4">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b3801ec-8b5b-467b-a591-b878ccb75633">19005-8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_112be464-2cae-4695-b0f3-9b4a4e9c964e">Impressions</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1b139baf-2817-499a-a9ff-3ea399e4ef00">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55030c64-e0d0-40c5-a871-5d483084dce0">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62a29c14-3719-40cb-b75c-8ee23bb26975">18783-1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aefdc364-f4c2-4d5f-99d0-dd4e61ad6584">Recommendations</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87ad8dc0-e9c8-4bd6-85c9-9d774906ef36">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.11" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_577692d3-c2c1-4d6e-81dd-5304b94768fb">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75b6a930-26b9-4646-92b9-8634901fb550">55110-1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d38a6fe4-05e0-413d-bafc-21fbda348fb4">Conclusions</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a792ca46-f0e0-4811-acda-02c478396a02">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96dd32a0-c949-4516-a248-47bb2d17d670">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cf1a4ea6-f1a0-4317-9e84-ebc089a00f2d">55107-7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a19cb757-08fa-4998-8255-128247803a51">Addendum</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac2abd0a-5c92-4ecc-8c33-e4306db48072">
<xref linkend="sect_9.7" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7579828d-dd17-48b2-b61e-745df827f672">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ced71b7a-131e-4948-a08c-03f10c068a73">18785-6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5c4733e-5e32-4e08-b4f8-60dddf9d6f04">Indications for Procedure</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_faae038a-c653-4fb2-8c55-6048ce99305e">
<xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_02b7c55a-6998-447c-99df-964e9d8d95f8">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3407ce68-63ef-4065-8169-6a36f54322b1">55108-5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a2648d6-9d25-4ab7-bae8-128bc9bf5718">Patient Presentation</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0ad283bf-e7ba-45f6-b8f5-d23f5d8a7729">
<xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ce9049f9-8355-4f7f-bebb-fd21a0626720">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5350060f-1484-4bb8-bd77-d3b4ec877352">55109-3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9f807514-a4ec-4a28-9e38-9ad74cfa3c8c">Complications</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89bb1c05-67c5-4ed9-9dac-0ced5fcaa166">
<xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.4" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c400412-7f75-4725-b9ec-23d189000533">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_32e8ee4d-5aa6-4e4d-af60-209f48eada0d">55112-7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d28c61f0-caf7-4107-a1dc-d794633583e7">Summary</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_981911e1-348e-4d00-9844-23b3676fdcba">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba1b4559-9f77-40f7-beaf-122bdfbeae7c">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ddabb173-22cc-409e-82cb-e4fef70f7e51">55113-5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0023aa1d-0336-4fb6-ad2a-8c9a9de115e3">Key Images</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c810a2e5-4e28-4bc4-a712-6153977666f0">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.6" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bdd664f0-f468-4200-b4a7-7e2756149713">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_364e6234-a7dc-4863-ade3-bb48f12a68db">73569-6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d733a31-cdf4-4a8a-ab64-32e9d9fa98b3">Radiation Exposure and Protection Information</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f7667ac6-2a82-4d9c-a0aa-b6a8e04e8d81">
<xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.5" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5539d949-ea45-4baf-887c-e0213c7f836d">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78dc7afb-2fed-4d03-8cef-6b0250d57119">55752-0</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d2b3562f-5eda-4cb3-bffb-3709c76b012d">Clinical Information</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_caf7f5b8-ff93-4295-b265-4b4cc80edc9c">
<xref linkend="sect_9.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7bb2bed0-6f0d-4b61-8d23-3dd1766bda63">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efbdc8ec-0d4d-4257-adac-b1ab54535c13">29549-3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_13a7a64f-26e7-4f35-b716-8f3d15d15bad">Medications Administered</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc334da0-b46d-4f1a-9f5f-fa1f1f68813b">
<xref linkend="sect_9.3" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_10.2" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5c7b4616-3d5b-42a9-8867-66225c6c371a">LN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_011eaf81-d35c-4ddb-a9c3-76fc7ddcc85a">73568-8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eef186e0-652d-4ae2-bb77-e38640bb99e7">Communication of Critical Results</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d15053b3-29ec-4743-9151-4680a3c8f998">
<xref linkend="sect_9.6" xrefstyle="select: labelnumber title"/> / <xref linkend="sect_9.8.10" xrefstyle="select: labelnumber title"/>
</para>
              </td>
            </tr>
          </tbody>
        </table>
        <para xml:id="para_afd52985-3d14-43bc-b051-18d24fb92537">CDA Template 1.2.840.10008.9.1 requires a minimum of an Imaging Procedure Description section and an Impression section.</para>
        <para xml:id="para_d0111d29-b9a8-437e-87d9-d12c87babdc1">The section/code element shall be populated in accordance with the relevant CDA template; note that the code might not be the same as the Concept Name code of the SR section CONTAINER. The title element of each CDA section shall be populated as shown in <xref linkend="table_C.4-2" xrefstyle="select: label"/>.</para>
        <table frame="box" label="C.4-2" rules="all" xml:id="table_C.4-2">
          <caption>CDA Section mapping from SR</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_23273bff-d6d1-4381-83d8-dee9c5d35831">CDA Business Name</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_8a686942-1cb0-4750-872d-d2a3d57dc7c4">DICOM SR</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_373f00b0-912a-4b47-8044-50e8f94e9340">&lt;section&gt;: Title</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f38cb921-55ec-416a-a052-5c53624fe944">
                  <emphasis role="italic">Concept Name Code Sequence (0040,A043) &gt; Code Meaning (0008,0104)</emphasis> [of the section CONTAINER content item]</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a61a9384-fd76-4e48-9e67-6617b5d4cdb1">&lt;section&gt;: Text</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b234db3-099b-45b5-bc8d-2236849057b8">[See <xref linkend="sect_C.4.2" xrefstyle="select: labelnumber"/>]</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8142d05-83ae-4cfb-8223-e109ee5a7128">&lt;section&gt;: CodedObservation[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6c5d2ff-5236-4b17-993d-a3c63577954f">[See <xref linkend="sect_C.4.3.1" xrefstyle="select: labelnumber"/> and <xref linkend="sect_C.4.3.2" xrefstyle="select: labelnumber"/>]</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b9d7425-7511-4f9b-9674-543832ec5ae9">&lt;section&gt;: QuantityMeasurement[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da1e3a72-302b-412f-9d53-5784866dc4dc">[See <xref linkend="sect_C.4.3.4" xrefstyle="select: labelnumber"/>]</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_416301af-0296-403a-ad4e-854a08a959ec">&lt;section&gt;: SOPInstance[*]</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba5a86eb-6cbc-42ce-a971-83151b47df60">[See <xref linkend="sect_C.4.3.3" xrefstyle="select: labelnumber"/>]</para>
              </td>
            </tr>
          </tbody>
        </table>
        <para xml:id="para_afad45ec-4cc5-4262-b4ec-89c35a954d81">SR allows sections to be qualified by observation context, using <olink targetdoc="PS3.16" targetptr="sect_TID_1001" xrefstyle="select: labelnumber"/> and its subsidiary templates. This capability is constrained in this mapping.</para>
        <section label="C.4.1.1" status="4" xml:id="sect_C.4.1.1">
          <title>Section Observer Context</title>
          <para xml:id="para_2fc421e9-8961-4d48-a3fb-7216da3e31f5">
<olink targetdoc="PS3.16" targetptr="sect_TID_1002" xrefstyle="select: labelnumber title"/> allows identification of a human or device author.</para>
          <table frame="box" label="C.4-3" rules="all" xml:id="table_C.4-3">
            <caption>CDA Section author mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_8dfe0592-cfa9-4e8e-b2e1-2f88a47bd014">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_bdd60899-6932-487c-b5bb-b66de5e38c6f">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_778dc4c2-e594-4e05-b9b9-956761355687">&lt;section&gt;: AuthorID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_188df97e-b89e-4f91-a70e-c5d9f5e83dd6">If
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121005">(121005, DCM, "Observer Type")</olink>
</emphasis> = <olink targetdoc="PS3.16" targetptr="DCM_121007">(121007, DCM, "Device")</olink>, then
                    <emphasis role="italic"/>
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121012">(121012, DCM, "Device Observer UID")</olink>
</emphasis>
                  </para>
                  <para xml:id="para_83d40c88-08ef-402a-b856-d831ebeaa115">ID for human observer not represented in SR; use nullFlavor="UNK"</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_db4d82f2-be30-40b7-a90b-81045480ba8b">&lt;section&gt;: AuthorName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bb144204-2c21-4c78-9c34-56fa7044d5fd">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121008">(121008, DCM, "Person Observer Name")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5fdeb1b7-a47f-4850-9088-56015a773ae3">&lt;section&gt;: AuthorOrganization</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_dc69481a-8649-4e0f-8f0b-b543cf7bf466">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121009">(121009, DCM, "Person Observer's Organization Name")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0575ce50-2131-4339-a57a-94ec1d4d61c1">&lt;section&gt;: AuthorDeviceModel</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_11af308b-d47e-4d86-aef1-7a4647f82a56">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121015">(121015, DCM, "Device Observer Model Name")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9a404ab3-70a6-491f-a0ea-49a4b59cf79f">&lt;section&gt;: AuthorSoftware</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cb200898-cfdf-4046-a314-06e92df4d22d">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121013">(121013, DCM, "Device Observer Name")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.4.1.2" status="4" xml:id="sect_C.4.1.2">
          <title>Comparison Study Procedure Context</title>
          <para xml:id="para_0d7c4717-2e7d-4068-8913-cdea609267a8">
<olink targetdoc="PS3.16" targetptr="sect_TID_1005" xrefstyle="select: labelnumber title"/> allows identification of a different procedure than the procedure identified in the SR Study IE as the context for the section observations. In the transformations of this Annex, only an identified comparison procedure is supported as Procedure Context, the SR section being transformed must be either Prior Procedure Descriptions or Previous Findings, and the CDA section shall be in accordance with the Comparison Study section Template 1.2.840.10008.9.4.</para>
          <para xml:id="para_ba0e8c61-4acb-4f66-ad30-cc283135f543">SR Instances using <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber"/> have additional attributes of a comparison procedure specified using <olink targetdoc="PS3.16" targetptr="sect_TID_2007" xrefstyle="select: labelnumber"/>, which is used in the Prior Procedure Descriptions section. The attributes of both <olink targetdoc="PS3.16" targetptr="sect_TID_1005" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="sect_TID_2007" xrefstyle="select: labelnumber"/> are source data in the <xref linkend="table_C.4-4" xrefstyle="select: label"/> mapping.</para>
          <table frame="box" label="C.4-4" rules="all" xml:id="table_C.4-4">
            <caption>Comparison Study mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_354b989b-656c-455a-b810-d79213234dfc">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f8732746-180f-42f5-952e-b8521445ef1d">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2bc57851-5dde-41fd-ada4-a8fe694f830d">ComparisonStudy: ProcedureTechnique: ProcedureCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5f6f826f-5112-4ee9-9522-2dc94a1e4e06">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121023">(121023, DCM, "Procedure Code")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ee2ca962-c14f-49c3-8320-a733aa789ab0">ComparisonStudy: ProcedureTechnique: EffectiveTime</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_44e7ef70-9829-4155-ae04-6a6b60d8f362">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_111060">(111060, DCM, "Study Date")</olink> + <olink targetdoc="PS3.16" targetptr="DCM_111061">(111061, DCM, "Study Time")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8f051b28-99be-43a4-acca-e6b4371bd851">ComparisonStudy: ProcedureTechnique: Modality</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_682ad076-7ddf-4646-b2be-3b575197d260">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_122142">(122142, DCM, "Acquisition Device Type")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ae75c932-eba7-48e8-a099-14b939879069">ComparisonStudy: ProcedureTechnique: MethodCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c400edf7-d7ce-4a82-93a1-9563a4f6ffaa">ComparisonStudy: ProcedureTechnique: TargetSite</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3ccd7444-a84c-48e2-9b7a-03932ddafa29">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_123014">(123014, DCM, "Target Region")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_12a93589-5b4b-4c79-941b-d8f7c3edf1de">ComparisonStudy: ProcedureTechnique: Laterality:</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_6cae14ee-d9a1-4891-915d-af42d7586dd8">ComparisonStudy: ProcedureTechnique: Ref:</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9be87c34-6a8c-4ca1-bef4-8c326befb2b0">ComparisonStudy: ProcedureTechnique: ProviderOrganization</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7d2e6565-b617-4cd9-867d-6e885d1d9e3e">ComparisonStudy: Study[*]: StudyUID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_eb7ba60f-e0e6-48ed-a032-3d976af6e526">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121018">(121018, DCM, "Procedure Study Instance UID")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3c8ebb85-1655-46e7-b260-8e16813dfd96">ComparisonStudy: Study[*]: Description</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f43a777e-18e5-4491-a613-3656340ab2dd">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121065">(121065, DCM, "Procedure Description")</olink>
</emphasis> , if present, or
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121023">(121023, DCM, "Procedure Code")</olink> &gt; Code Meaning (0008,0104)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e28775ef-f141-414d-b66b-c59c3b9261ba">ComparisonStudy: Study[*]: Time</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7b43e1dd-133d-4ddb-9c1b-2aa0e8628d01">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_111060">(111060, DCM, "Study Date")</olink> + <olink targetdoc="PS3.16" targetptr="DCM_111061">(111061, DCM, "Study Time")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.4.1.3" status="4" xml:id="sect_C.4.1.3">
          <title>Fetus Subject Context</title>
          <para xml:id="para_8482a277-c5a4-43bc-bc96-2b6f654d329d">
<olink targetdoc="PS3.16" targetptr="sect_TID_1006" xrefstyle="select: labelnumber title"/> allows identification of a different subject than the patient identified in the SR Patient IE. In the transformations of this Annex, only an identified fetus subject is supported as Subject Context for a Findings section. An SR section with a fetus subject context shall be mapped to a CDA section shall be in accordance with the Fetus Findings subsection Template 1.2.840.10008.9.9. This section is subsidiary to the top level Findings section; multiple SR fetus findings sections may be mapped to separate CDA Fetus Findings subsections.</para>
          <table frame="box" label="C.4-5" rules="all" xml:id="table_C.4-5">
            <caption>CDA Fetus subject mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_76d37db4-bfc4-4aa3-bf52-0b1b459e58f1">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_f1667301-ab67-41cf-9fdc-c363fbe7bd21">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e473ae3e-b372-4fcc-b3a5-4da54b1d2bee">Findings: FetusFindings[*]: FetusID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_bedc06fa-b398-4720-9287-bbb6c118d766">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_121030">(121030, DCM, "Subject ID")</olink>
</emphasis> or
                    <emphasis role="italic">
<link xl:href="http://loinc.org/11951-1/" xl:show="new">(11951-1, LN, "Fetus ID")</link>
</emphasis>
                  </para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
      </section>
      <section label="C.4.2" status="3" xml:id="sect_C.4.2">
        <title>Section/text</title>
        <para xml:id="para_cce84d7b-09b4-4e31-9ecf-ed6494d3a787">DICOM <olink targetdoc="PS3.16" targetptr="sect_TID_2002" xrefstyle="select: labelnumber title"/> specifies that sections contain imaging report elements of type CODE, TEXT, IMAGE, or NUM.</para>
        <para xml:id="para_40d92c58-912f-4cdd-a949-8c937fa20c0d">Section/text in the CDA document contains the narrative text (attested content) of the document. Section/text shall be generated from all the Content Items subsidiary to a section CONTAINER of the SR document, such that the full meaning is be conveyed in an unambiguous manner in the narrative block.</para>
        <para xml:id="para_c5cc848a-bf25-4508-bb78-bdef6f855679">The narrative rendered from each Content Item shall be encapsulated in a &lt;content&gt; element of the narrative block, allowing the associated entry to reference it.</para>
      </section>
      <section label="C.4.3" status="3" xml:id="sect_C.4.3">
        <title>Content Item Mapping</title>
        <para xml:id="para_70a05c77-d252-4295-98d0-0478fd9a9ffc">Each Content Item immediately subsidiary to a section CONTAINER shall be mapped to the corresponding entry level template, and shall be included subsidiary to the associated CDA section or subsection. This is in addition to its rendering in the section/text narrative block.</para>
        <para xml:id="para_a5755408-a0cf-4254-82fc-00afa28cf514">Coded concepts that are encoded in the SR using with the Coding Scheme Designator "SRT" shall be mapped to the equivalent SNOMED CT code. Mappings for value sets invoked in both SR and CDA are provided in <olink targetdoc="PS3.16" targetptr="PS3.16" xrefstyle="select: labelnumber"/>.</para>
        <section label="C.4.3.1" status="4" xml:id="sect_C.4.3.1">
          <title>Coded Observations</title>
          <para xml:id="para_b23cc69d-40a4-450f-844a-e070bf2962a1">SR CODE Content Items shall be mapped to Coded Observation entries.</para>
          <table frame="box" label="C.4-6" rules="all" xml:id="table_C.4-6">
            <caption>CDA Coded Observation mapping from SR CODE</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_73023d2e-3d17-4913-b261-9338c988f873">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d24e267d-e9c1-4ef0-819e-4a54b636b67e">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9d30da4c-1070-4903-8be1-b0af9c126d53">CodedObservation[*]: ObsName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_fb2a1b25-5ed1-422a-9527-4a65889de15b">
                    <emphasis role="italic">Concept Name Code Sequence (0040,A043)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a4014d9d-0563-497e-9bdd-2e4fcbb6ebf6">CodedObservation[*]: ObsValue</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a167821d-4174-4f2d-9139-72f063e7963f">
                    <emphasis role="italic">Concept Code Sequence (0040,A168)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c8874e10-f744-4d6b-9ea7-435a343a86b7">CodedObservation[*]: Time</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e03bf6cd-f066-4195-99e9-f42f58ff336f">
                    <emphasis role="italic">Observation DateTime (0040,A032)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_068bc310-fd85-4e33-840c-3362e5780ee5">CodedObservation[*]: InterpretationCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9a55372e-0e63-40c9-82e4-19d051c6c744">CodedObservation[*]: ActionableFindingCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1a9cd162-6854-46df-8f77-7c456e1057ef">CodedObservation[*]: TargetSite</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1128e56e-032a-4a6e-8e71-e2e574b46c7d">
                    <emphasis role="italic">
<link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_50d930ac-2dbb-4707-88c8-2f1ac2958312">CodedObservation[*]:Laterality</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9aeb4907-940a-4731-b42e-d2e64ab51d4e">
                    <emphasis role="italic">
<link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/272741003" xl:show="new">(272741003, SCT, "Laterality")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ec905400-4818-4f78-9738-2fdc75e3f03f">CodedObservation[*]:TopoModifier</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_740b38d8-86ac-48df-9e43-b0b40b74313b">CodedObservation[*]:Method</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f45f623a-d749-402b-b2f4-cb9f3315bfb9">CodedObservation[*]: SOPInstance</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2af65abd-abaf-4c9f-a629-8be9307b2639">[See <xref linkend="sect_C.4.3.3" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7894b777-481f-49b1-a8e6-84fd587804ca">CodedObservation[*]: QuantityMeasurement</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a0df70fe-d417-4358-992b-80017418d28f">[See <xref linkend="sect_C.4.3.4" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0784ff35-a2fd-48f5-9853-07e8e07b8222">CodedObservation[*]: CodedObservation</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_7574f732-5e15-45d6-8d04-cb023d98a834">The CODE observations in <olink targetdoc="PS3.16" targetptr="sect_TID_2002" xrefstyle="select: labelnumber"/> do not specifically include finding site, laterality, and topographical modifiers, but these modifiers are not forbidden in the template, and may be present in a SR SOP Instance being transformed to CDA.</para>
        </section>
        <section label="C.4.3.2" status="4" xml:id="sect_C.4.3.2">
          <title>Text Observations</title>
          <para xml:id="para_2e710a90-9a7f-41fb-9fdd-ba76cd04409d">SR TEXT Content Items are mapped to Coded Observation entries, but the code is a nullFlavor with the text content in originalText.</para>
          <table frame="box" label="C.4-7" rules="all" xml:id="table_C.4-7">
            <caption>CDA Coded Observation mapping from SR TEXT</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2fee3f7f-a202-4a19-91fb-e82a17394873">CDA Business Name or XPath</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_5edd8692-4539-478a-b7ad-2eab8db75d39">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3a3ffc50-5d36-4216-81c4-38b6d7debfda">CodedObservation[*]: ObsName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8f4da85b-7774-409f-ab99-3f0674b47427">
                    <emphasis role="italic">Concept Name Code Sequence (0040,A043)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_8d4873cd-6b5d-4fdb-94fb-e7b0aa0a24e7">observation/value/@nullFlavor</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_54bf6996-835e-45cb-9474-92f63ee75a36">"NI"</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e473944e-1aa5-4c4a-b6b1-22a55808acc5">observation/value/originalText</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a40ee7f3-d83a-4a35-b0ee-8bccd02383d6">
                    <emphasis role="italic">Text Value (0040,A160)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_26005f87-e614-4216-930d-e0ef3650ae6e">CodedObservation[*]: Time</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_214b7d4a-9faf-455f-a18a-6935427cfa5f">
                    <emphasis role="italic">Observation DateTime (0040,A032)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7fd6ab89-91f8-436b-8756-6e2d7e1601f2">CodedObservation[*]: InterpretationCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_807a64da-0587-4ecd-864e-d9b48b6d964a">CodedObservation[*]: ActionableFindingCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_603f0c92-7881-4760-a690-e7b1f97eadad">CodedObservation[*]: TargetSite</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c496f7ce-4e0b-4ebb-aa53-a0189861407b">CodedObservation[*]:Laterality</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7be18647-4b15-4811-a9ae-55325fad3cf8">CodedObservation[*]:TopoModifier</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b3f0ec0f-cb86-4d58-9bf4-f0afe77c0a10">CodedObservation[*]:Method</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ed1d1634-ab9f-485a-94f2-e26fdc854da2">CodedObservation[*]: SOPInstance</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_64853a0d-95b7-45c5-b8b2-59615029ee31">[See <xref linkend="sect_C.4.3.3" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_61e37ba5-4a1f-4808-bcd6-dd19cb5338c0">CodedObservation[*]: QuantityMeasurement</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_27bbddf4-431a-45d1-8382-f10c9b70cc0f">[See <xref linkend="sect_C.4.3.4" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_25075934-ceb7-4162-a4e3-e751272ef5bf">CodedObservation[*]: CodedObservation</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.4.3.3" status="4" xml:id="sect_C.4.3.3">
          <title>Image Observations</title>
          <para xml:id="para_32fe1f4e-dab8-4ca3-b4d7-f8fb1724e5a0">SR IMAGE Content Items shall be mapped to SOP Instance Observation entries.</para>
          <table frame="box" label="C.4-8" rules="all" xml:id="table_C.4-8">
            <caption>CDA SOP Instance Observation mapping from SR IMAGE</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2fdc3e4a-fdc2-4191-950f-9014c2c11b5f">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a11363be-d150-4e7d-8a4c-317ea2790226">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b41fcf17-ff5b-4898-a6ac-cb618d07724b">SOPInstance[*]:SOPInstanceUID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_57884ce9-9ec1-483d-9dd8-9c46813bad42">
                    <emphasis role="italic">Referenced SOP Sequence (0008,1199) &gt; Referenced SOP Instance UID (0008,1155)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0870886a-6be9-4cd3-8b80-6f7003824801">SOPInstance[*]:SOPClassUID</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0ac1fb21-9bfa-4b8a-83b3-003a1a04eeda">
                    <emphasis role="italic">Referenced SOP Sequence (0008,1199) &gt; Referenced SOP Class UID (0008,1150)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3683cf6d-9d7d-428d-bfda-5db79cb7fced">SOPInstance[*]:WADOReference</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2b8918d9-c185-4b85-9714-ac4d630eb1d8">[WADO link constructed from image reference; also used in linkHtml in narrative block]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_0435cfc9-b6f4-4f30-84a2-d6514ea0f6c9">SOPInstance[*]:PurposeOfReference</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5a04ccab-8333-4044-8f7a-9d2b26d1b3e7">
                    <emphasis role="italic">Concept Name Code Sequence (0040,A043)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_329734dc-712b-4b2e-b845-a0d7d524b328">SOPInstance[*]:ReferencedFrames</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_cc1876e2-7a52-468f-8ec6-802ec25b8faf">
                    <emphasis role="italic">Referenced SOP Sequence (0008,1199) &gt; Referenced Frame Number (0008,1160)</emphasis>
                  </para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.4.3.4" status="4" xml:id="sect_C.4.3.4">
          <title>Numeric Observations</title>
          <para xml:id="para_ee372192-b358-4edb-9ffa-fbf51022a44f">SR NUM Content Items shall be mapped to Quantity Measurement entries.</para>
          <table frame="box" label="C.4-9" rules="all" xml:id="table_C.4-9">
            <caption>CDA Quantity Measurement mapping from SR NUM</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_fde0826e-b8f7-47ac-b175-4a2c9e243ae0">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_1be253a4-ca00-491b-b47f-a3f44650c301">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a19bdd5c-6abb-4047-be81-66a241b20637">QuantityMeasurement[*]: MeasurementName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b2b98e22-234b-4201-8b29-2721938a1cfd">
                    <emphasis role="italic">Concept Name Code Sequence (0040,A043)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5ed6b745-1c5d-476f-aca6-596ad4074a87">QuantityMeasurement[*]: MeasurementValue</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f10367f2-b5c7-4f71-963a-d5a4fd7af174">
                    <emphasis role="italic">Measured Value Sequence (0040,A300) &gt; Numeric Value (0040,A30A)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ca0d610e-c46b-4804-b1f6-5e85dfd97c88">QuantityMeasurement[*]: MeasurementUnits</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d7466411-19aa-436e-b7ce-2ccb24a86465">
                    <emphasis role="italic">Measured Value Sequence (0040,A300) &gt; Measurement Units Code Sequence (0040,08EA) &gt; Code Value (0008,0100)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_88496a2f-ab4b-4377-af18-f02afd62cdba">QuantityMeasurement[*]: Time</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d6954c3f-912a-4b62-9a8e-567031e759d6">
                    <emphasis role="italic">Observation DateTime (0040,A032)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a068cef8-8146-4786-b971-c828befd1849">QuantityMeasurement[*]: InterpretationCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_526d0108-4e42-4dd5-848b-f25268b7cc1b">QuantityMeasurement[*]: ActionableFindingCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_576f6af7-2909-458e-b2fa-af4f81fecadd">QuantityMeasurement[*]: TargetSite</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_b4e54a81-30bc-4d76-95fb-40c59ad55ac0">
                    <emphasis role="italic">
<link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_54f33625-44f1-4777-abfa-b9b27565bacc">QuantityMeasurement[*]:Laterality</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a718e1e6-d125-4fcf-b838-c2358770ac16">
                    <emphasis role="italic">
<link xl:href="http://snomed.info/id/363698007" xl:show="new">(363698007, SCT, "Finding Site")</link> &gt; <link xl:href="http://snomed.info/id/272741003" xl:show="new">(272741003, SCT, "Laterality")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_17eedc21-26f8-464c-abc5-73feae05ccf0">QuantityMeasurement[*]:Method</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_eef04666-4de9-44c5-82ab-833372a5b3aa">
                    <emphasis role="italic">
<link xl:href="http://snomed.info/id/370129005" xl:show="new">(370129005, SCT, "Measurement Method")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a08e2345-699e-4493-89fd-0161a7513fed">QuantityMeasurement[*]:TopoModifier</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2eff334e-9123-4558-a9ee-e5100c7d866d">
                    <emphasis role="italic">
<link xl:href="http://snomed.info/id/106233006" xl:show="new">(106233006, SCT, "Topographical modifier")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_e5c50115-bb0d-4f0c-a7ca-4393f6113be0">QuantityMeasurement[*]: SOPInstance</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ae3c2ba4-96b0-4c5c-9acd-0cbbb9d34f21">[See <xref linkend="sect_C.4.3.3" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5cadb65f-e88c-46f8-8090-19888135c983">QuantityMeasurement[*]: QuantityMeasurement</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_35830100-647c-4b59-b6df-fb4e386bacec">[See <xref linkend="sect_C.4.3.4" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_bda0b69b-d516-4a9c-a636-409302d283ce">The SR templates invoked for NUM measurements from <olink targetdoc="PS3.16" targetptr="sect_TID_2000" xrefstyle="select: labelnumber"/> do not specifically include finding site, laterality, and topographical modifiers, but these modifiers are not forbidden in the template, they are used in many other NUM value templates (e.g., <olink targetdoc="PS3.16" targetptr="sect_TID_300" xrefstyle="select: labelnumber title"/>), and may be present in a SR SOP Instance being transformed to CDA.</para>
        </section>
        <section label="C.4.3.5" status="4" xml:id="sect_C.4.3.5">
          <title>Inferred From Image Observations</title>
          <para xml:id="para_d9b60f71-9b72-4721-ab29-20f696a655c6">SR <olink targetdoc="PS3.16" targetptr="sect_TID_2001" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="sect_TID_2002" xrefstyle="select: labelnumber"/> allow Content Items to be INFERRED FROM IMAGE observations. The INFERRED FROM relationship is mapped to the entryRelationship with typeCode=SPRT, and the IMAGE observation is mapped to a CDA SOP Instance Observation entry subsidiary to its parent CDA Coded Observation or Quantity Measurement entry. This entryRelationship is shown in the Coded Observation and Quantity Measurement CDA Templates.</para>
        </section>
        <section label="C.4.3.6" status="4" xml:id="sect_C.4.3.6">
          <title>Inferred From Numeric Observations</title>
          <para xml:id="para_2e7e8977-1d2e-475f-a282-03d51a74c82a">SR <olink targetdoc="PS3.16" targetptr="sect_TID_2001" xrefstyle="select: labelnumber"/> and <olink targetdoc="PS3.16" targetptr="sect_TID_2002" xrefstyle="select: labelnumber"/> allow Content Items to be INFERRED FROM NUM observations. The INFERRED FROM relationship is mapped to the entryRelationship with typeCode=SPRT, and the NUM observation is mapped to CDA Quantity Measurement entry subsidiary to its parent CDA Coded Observation or Quantity Measurement entry. This entryRelationship is shown in the Coded Observation and Quantity Measurement CDA Templates.</para>
        </section>
        <section label="C.4.3.7" status="4" xml:id="sect_C.4.3.7">
          <title>Inferred From Spatial Coordinates Observations</title>
          <para xml:id="para_0ec0bfe3-fc0a-45b1-b400-21194534df1a">SR <olink targetdoc="PS3.16" targetptr="sect_TID_1400" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.16" targetptr="sect_TID_1401" xrefstyle="select: labelnumber"/>, <olink targetdoc="PS3.16" targetptr="sect_TID_1402" xrefstyle="select: labelnumber"/>, and <olink targetdoc="PS3.16" targetptr="sect_TID_1404" xrefstyle="select: labelnumber"/> allow NUM Content Items to be INFERRED FROM SCOORD observations, which are SELECTED FROM IMAGE observations. This Annex does not specify the transformation for SCOORD observations; these would use the CDA Region Of Interest entry, which PS3.20 forbids (see <xref linkend="sect_9.1.2.4" xrefstyle="select: label"/>).</para>
        </section>
      </section>
      <section label="C.4.4" status="3" xml:id="sect_C.4.4">
        <title>Specific Section Content Mapping</title>
        <para xml:id="para_3b0686f0-952d-4af6-80fa-961591d71a5d">Certain sections in a CDA Imaging Report have specific mappings from the DICOM SR header, or from specialized templates with content for particular uses.</para>
        <section label="C.4.4.1" status="4" xml:id="sect_C.4.4.1">
          <title>Procedure Indications</title>
          <para xml:id="para_a9a1d2ca-8066-45c9-b229-a8ca4a855da5">The DICOM SR Document General Module may specify the Reason for the Requested Procedure as either free text in attribute (0040,1002), and/or as multiple coded values in attribute (0040,100A). These are mapped to the Procedure Indications subsection of the Clinical Information section of the CDA Imaging Report.</para>
          <note>
            <para xml:id="para_8c47292d-27ea-4aef-8281-6b5330946baf">Procedure indications may also be specified as SR content items in the <link xl:href="http://loinc.org/18785-6/" xl:show="new">(18785-6, LN, "Indications for Procedure")</link> CONTAINER, which may be mapped to the CDA instance in accordance with <xref linkend="sect_C.4.3" xrefstyle="select: label"/>. It is an implementation decision how to handle multiple representations of indications in the SR document.</para>
          </note>
          <table frame="box" label="C.4-10" rules="all" xml:id="table_C.4-10">
            <caption>Clinical Information Procedure Indications mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_9e8b81f3-0580-4c5d-a680-57f3f9c4aeff">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_572d140d-864e-4e48-ad31-81fd47ace664">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_39c60e39-6435-4921-a9a6-d65fb4e198f1">ClinicalInformation: ProcedureIndications: Text</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_33dd8dce-216d-422a-816e-916f5be41fe7">
                    <emphasis role="italic">Referenced Request Sequence (0040,A370) &gt; Reason for the Requested Procedure (0040,1002)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d8ce3f7e-a95f-4877-91c6-8e8ef36a10d4">ClinicalInformation: ProcedureIndications: CodedObservation[*]: ObsName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_edc9981f-dc79-4a6e-9d0d-5340361f9114">(432678004, SNOMED, "Indication for procedure")</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2417a276-c710-4a65-bb19-263645b78848">ClinicalInformation: ProcedureIndications: CodedObservation[*]: ObsValue</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_791bae72-9cd8-47fa-87a3-af879219fc06">
                    <emphasis role="italic">Referenced Request Sequence (0040,A370) &gt; Reason for the Requested Procedure Code Sequence (0040,100A)</emphasis>
                  </para>
                </td>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.4.4.2" status="4" xml:id="sect_C.4.4.2">
          <title>Current Procedure Descriptions</title>
          <para xml:id="para_c85173a7-cee0-46ae-98a0-09bc82ad3395">SR Instances using <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber"/> have a Current Procedure Descriptions section specified using <olink targetdoc="PS3.16" targetptr="sect_TID_2007" xrefstyle="select: labelnumber"/>. Source data in that template and from the General Study Module is mapped into the CDA Procedure Description section.</para>
          <table frame="box" label="C.4-11" rules="all" xml:id="table_C.4-11">
            <caption>Current Procedure Description mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_a1e0d414-f73f-4a0f-8236-31fc86306e84">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_d29f4ed4-192e-4e7c-a436-2e580c8957a0">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_675f19fe-4d94-4776-8c80-8c80edba4c3b">ProcedureDescription: ProcedureTechnique: ProcedureCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_16be0b95-89de-42c0-9c0d-20aabed61d86">
                    <emphasis role="italic">Procedure Code Sequence (0008,1032)</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_15c0d20c-a1ae-4d6d-82b3-17d37fab7f92">ProcedureDescription: ProcedureTechnique: EffectiveTime</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4235e6e1-47eb-490e-bf38-f2c992bfb7f3">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_111060">(111060, DCM, "Study Date")</olink> + <olink targetdoc="PS3.16" targetptr="DCM_111061">(111061, DCM, "Study Time")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_140a33fb-b60b-4c99-b30d-5c4552b36f5e">ProcedureDescription: ProcedureTechnique: Modality</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_569573a6-78f5-4d28-9a0f-ad2e60c327be">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_122142">(122142, DCM, "Acquisition Device Type")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5c2f2b5b-eaff-4bd2-b190-78b5c2277f10">ProcedureDescription: ProcedureTechnique: MethodCode</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_f7ed2de2-1f23-4041-8b55-0a5be8e41a86">ProcedureDescription: ProcedureTechnique: TargetSite</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_5070ee02-6ab0-4e8c-9b44-c0971d1ef408">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_123014">(123014, DCM, "Target Region")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_995df438-415e-4a4e-8045-a78a482d8786">ProcedureDescription: ProcedureTechnique: Laterality</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_3f380fe1-812c-47c3-956a-605f80a7aea2">ProcedureDescription: ProcedureTechnique: Ref</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
            </tbody>
          </table>
        </section>
        <section label="C.4.4.3" status="4" xml:id="sect_C.4.4.3">
          <title>Radiation Exposure and Protection Information</title>
          <para xml:id="para_856915fc-0060-4353-affe-be19314f1dbb">The Radiation Exposure and Protection Information section defined in SR <olink targetdoc="PS3.16" targetptr="sect_TID_2006" xrefstyle="select: labelnumber"/> is specified using <olink targetdoc="PS3.16" targetptr="sect_TID_2008" xrefstyle="select: labelnumber"/>, which provides additional source data for mapping into the equivalent CDA subsection of the Imaging Procedure Description section.</para>
          <table frame="box" label="C.4-12" rules="all" xml:id="table_C.4-12">
            <caption>CDA Radiation Exposure and Protection Information mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_2ece6148-a97b-41ae-a634-be844df5c52d">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_19d4497c-5c4f-4a1c-87d9-3076c473df74">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_54b27d28-d78a-49f8-a4e8-8bba3b9baa25">RadiationExposure: IrradiationAuthorizingID</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_985be25c-66d4-485b-9f91-a2efd7b4cfa1">RadiationExposure: IrradiationAuthorizingName</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_05e92bd4-480a-47d6-ad52-9c009854bdf6">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_113850">(113850, DCM, "Irradiation Authorizing ")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1278abf9-6077-4559-98b5-d245b6de38bb">RadiationExposure:SOPInstance[doseReport]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_2fdcd61b-a347-4fc4-b747-edca3d92dd35">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_113701">(113701, DCM, "X-Ray Radiation Dose Report")</olink>
</emphasis> [from Current Procedure Description section]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_43c8384c-cf38-4c47-b317-1ecaed8ba5f0">RadiationExposure:CodedObservation[pregnancy]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_498f079b-a83c-4ce8-b576-9f49150e7bf0">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_111532">(111532, DCM, "Pregnancy Status")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_4ebddf69-49c7-402a-8152-eed852a27333">RadiationExposure:CodedObservation[indication]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_11c90025-3d3b-47bd-abbd-01010b64d74e">
                    <emphasis role="italic">
<link xl:href="http://loinc.org/18785-6/" xl:show="new">(18785-6, LN, "Indications for Procedure")</link>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_9cfd160d-b411-4cff-8637-215062756cba">RadiationExposure:CodedObservation[exposure]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_87084928-3dff-4a33-91cb-78275b10d03d">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_113921">(113921, DCM, "Radiation Exposure")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1c2a3880-48a4-4c67-a957-1009d2e4a450">RadiationExposure:QuantityMeasurement</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ae98e23c-7100-449c-a675-58d968ec8450">RadiationExposure: RadioactivityDose</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_a59e6a4c-8153-4d97-b15a-eee4e3e2a732">RadiationExposure: Radiopharmaceutical</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_17e4a88e-f611-4c2f-8357-7e4125e89d7c">RadiationExposure: FreeTextRadiopharmaceutical</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_55a3158c-b5fa-4789-9643-23ac313eb826">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_113922">(113922, DCM, "Radioactive Substance Administered")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
            </tbody>
          </table>
          <para xml:id="para_06ad9f9a-d09a-44be-8c71-5c914fe9b01f">The Radiation Exposure Content Item in <olink targetdoc="PS3.16" targetptr="sect_TID_2008" xrefstyle="select: labelnumber"/> uses Value Type TEXT, not NUM, and is therefore mapped to a Coded Observation entry in accordance with <xref linkend="sect_C.4.3.2" xrefstyle="select: label"/>.</para>
        </section>
        <section label="C.4.4.4" status="4" xml:id="sect_C.4.4.4">
          <title>Key Images</title>
          <para xml:id="para_d9b87175-3586-4f94-9c63-0f853f8bbe1d">
<olink targetdoc="PS3.16" targetptr="sect_TID_2005" xrefstyle="select: labelnumber title"/> specifies a section structure for the Key Images section of an SR, which allows mapping into the equivalent CDA subsection of the Impression section.</para>
          <table frame="box" label="C.4-13" rules="all" xml:id="table_C.4-13">
            <caption>Key Image mapping from SR</caption>
            <thead>
              <tr valign="top">
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_c28243d3-693f-41fb-b7ad-2a8c37f2e280">CDA Business Name</para>
                </th>
                <th align="center" colspan="1" rowspan="1">
                  <para xml:id="para_ab4d33be-1f7d-4647-a75a-0bc90c438410">DICOM SR</para>
                </th>
              </tr>
            </thead>
            <tbody>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_05ca05a7-85a0-44fa-a881-af4442f237c8">KeyImages: Title</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_148b1e70-85af-4362-8c2d-a1fe58c68f45">"Key Images" [or equivalent in local language]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_960ebb8c-e3e9-45aa-962d-09041a6fcf3e">KeyImages: Text</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_aebbdda6-1223-4ac5-81a1-56a2bcb70fd4">
                    <emphasis role="italic">
<olink targetdoc="PS3.16" targetptr="DCM_113012">(113012, DCM, "Key Object Description")</olink>
</emphasis>
                  </para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_24f47ed1-1d9d-4a42-b977-51d4c0902ca6">KeyImages: Text: GraphicRef[*]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_de21a9ff-8343-4cea-afe9-7c5e9ec8c76e">[Reference to ObservationMedia entry]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_d2e9b336-2964-4f35-aaac-91a660b005d5">KeyImages: Text: ExtRef[*]: URL</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_ccb1e55d-11b9-4d41-bc31-63f20bc1d99e">[WADO link constructed from image reference]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_16bf05ac-4574-472d-b873-dc3e583fba19">KeyImages: SOPInstance[*]</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_043a99df-60f3-4eeb-983c-748d293a2968">[See <xref linkend="sect_C.4.3.3" xrefstyle="select: labelnumber"/>]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c1da9c95-4de7-47a5-99f8-d2eb711b930d">KeyImages: Graphic[*]: Image</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_7488a3fe-20a0-4478-b893-276d70f7654c">[Thumbnail constructed from referenced image]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_c916ab21-0e3a-4abd-ab1b-f463fc73640a">KeyImages: Graphic[*]: MediaType</para>
                </td>
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_005383ba-ca30-4058-bcd9-af6d00a463dd">[recommended "image/jpeg"]</para>
                </td>
              </tr>
              <tr valign="top">
                <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_52e8cfe4-b878-4520-94a7-07143bf17e59">KeyImages: Graphic[*]: ImageURI</para>
                </td>
                <td align="left" colspan="1" rowspan="1"/>
              </tr>
            </tbody>
          </table>
        </section>
      </section>
    </section>
    <section label="C.5" status="2" xml:id="sect_C.5">
      <title>Example</title>
      <section label="C.5.1" status="3" xml:id="sect_C.5.1">
        <title>DICOM SR "Basic Diagnostic Imaging Report" (TID 2000)</title>
        <para xml:id="para_e76456eb-d2aa-4a59-87d2-a94ea2fd7cdd">The SR sample document encoding includes information on the SR document body tree depth (column 1: SR Tree Depth), nesting level for nested artifacts such as sequences and sequence items (column 2: Nesting), DICOM attribute names (column 3: Attribute), DICOM tag (column 4: Tag), the DICOM attribute value representation (Column 5: VR as specified in PS3.5), the hexadecimal value of value length (column 6: VL (hex)) and the sample document attribute values (column 7: Value).</para>
        <table frame="box" label="C.5-1" rules="all" xml:id="table_C.5-1">
          <caption>Sample document encoding</caption>
          <thead>
            <tr valign="top">
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_70e09c0b-d4b8-4294-8836-ebcabbb0e0d5">SR Tree Depth</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_fad12460-8194-4b7b-bd72-09199982e97b">Nesting</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_f8d8b090-3e22-41db-aafa-866d9bdb1182">Attribute</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_e7db6db0-9fce-4e7e-9bdf-9dfb8db93c79">Tag</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_70377c13-8d79-49cc-bb48-c5e6c12c0882">VR</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_dbbc7929-4348-4c8f-86f7-764a95c8d29c">VL (hex)</para>
              </th>
              <th align="center" colspan="1" rowspan="1">
                <para xml:id="para_a08a285a-7b7b-4eec-ace4-7048b8d2e2bd">Value</para>
              </th>
            </tr>
          </thead>
          <tbody>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b59e7fe4-8173-4e19-8c56-92d374753917">Instance Creation Date</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05cf1398-02a2-4722-be99-bcaef73d5a90">(0008,0012)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25262b9d-a218-4356-b77a-6ce9b93c2f88">DA</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e207e94b-b22a-463b-b149-38ba1ec739a8">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bd48b669-da8c-423c-9e87-37cceb585cc4">20060827</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7aa6e8f-a4a6-4091-bc8a-e391eaf1f34c">Instance Creation Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa9adc5a-3700-4ccc-b28b-2c661029f8fd">(0008,0013)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8036670-02da-4b70-bd97-20edb1b2da65">TM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4857ab06-dfd9-40cc-8718-288fea0a90a3">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db1ff324-2b18-42ab-848d-c6e7e7ca6e11">224157</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_477aa345-40b9-4d6b-b3cb-dcf9373f19c7">Instance Creator UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa984345-ab3b-4b5c-9567-b9b4bdda1e3d">(0008,0014)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d02f4f15-618f-4a1d-8e31-196db79ee542">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_afdbc46f-e0f7-4030-bb17-e357537e0542">001c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41ace169-8510-4385-93d7-f391e718a5d6">1.2.276.0.7230010.3.0.3.5.4</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_43a80f47-65b0-430b-ab18-5a16bb03dce0">SOP Class UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c5a1308-7540-4506-8673-986a2c99614e">(0008,0016)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8f1deb90-c4a7-4943-9d28-233ed6d12f88">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5633d481-d506-4546-8632-fe41aea044ec">001e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7748c808-15dd-4135-b57b-4a831a5fff57">1.2.840.10008.5.1.4.1.1.88.22</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1da2f06-da2a-4f33-891a-70a23b555b57">SOP Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5ecca0cd-a17e-43df-970a-0cfa412299be">(0008,0018)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1f54c52-250a-45c4-9aaf-64c14b790517">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7627dd13-4a62-491a-924e-7a9837752af5">003c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b14784d3-af69-401f-99cc-a0aa63140a37">1.2.840.113619.2.62.994044785528.20060823.200608232232322.9</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e378ab1-810a-4926-be17-7c0bb4ef7c72">Study Date</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb9141e2-4e1c-477d-a672-79f68d0d28ff">(0008,0020)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0f14750-221a-4b04-ba79-b32bfe8b6d00">DA</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_837e8d17-63e2-47c8-ae39-43cd7c115e13">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5159ea7-4d5a-47d2-9a0c-ca5e9c1f0844">20060823</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f0d99e9-0fc3-4160-bce2-2f95edaf46eb">Content Date</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_22dc8573-1595-4598-a264-1c9789c87010">(0008,0023)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_313cc8e0-3ac3-465b-b378-e8faa13f3e63">DA</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9c55ed74-bc50-4baf-b530-108834e665b1">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3dfdc90-404e-4583-a253-f1e54b7d67d1">20060823</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0864c598-9cda-446c-a190-11d1a64b9660">Study Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c43b04c7-595e-4dd3-8ced-9dfb9aa83b08">(0008,0030)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c9d699d-5a7c-4383-8d5d-9f37acec6d8d">TM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eaa410e9-9e70-430a-9f8f-5cb1ed6015bf">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5804351f-6544-4af7-8b1b-91ca46ea91b1">222400</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38fbc8aa-44e6-42b9-9544-e9abfe915229">Content Time</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5eefbac4-713a-4f6a-8e19-ac14ede91818">(0008,0033)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd804232-be18-4020-bcb9-a59cb48ecc76">TM</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f263048c-9b39-429e-b159-2994af8c842f">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_24e593f8-8f4a-4bc8-851d-d3a4dfc6a9a2">224352</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b78506b8-8b27-4508-bcd9-7acc799990fe">Accession Number</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9778e99-a011-4bdd-8443-e8f54f1d62cf">(0008,0050)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77b2b6cc-56c8-4bf9-b945-7957d88646b1">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ecb7bc73-b559-4828-8dc6-b14151bbb8d9">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08452b48-10a3-4595-bb32-0c205c48b793">10523475</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac9a4e8e-88c2-43b3-8227-71683b507055">Issuer of Accession Number Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51cd7a09-c4f3-4cbd-b63c-d863ea221d6f">(0008,0051)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_27b04160-2086-49ee-8728-6ca77fbeedd2">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a65c77b3-c531-40ac-a61d-6702fb669874">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a59db2d9-594c-47d5-8bc7-b0bcb3ed0114">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45d0e808-f5ad-4709-a639-3b74b28bec6a">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78db21f2-3430-4d6f-a761-569037e8c2f6">Local Namespace Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8915339a-186f-4f3e-9882-ebe9672ff5af">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1c59769-3c29-408e-a576-74a49b1a2333">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3e470ff9-82dc-43ed-b294-b4e3942cc9bd">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d8b6373-dc5b-4f5b-a0fa-d35783ced3be">WUH-RIS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4e4c89c7-ac59-4bea-9c82-25f7c6a70ca5">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f2f5f68-be1d-4bde-8243-2fe3aaf1c83d">Universal Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc0b1dd8-6f7c-418e-b01c-ed0d4e954fe7">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a7173a6-d77f-4657-b9bd-6352227fba18">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_318d4fed-a6cd-4a82-a4a2-61b23feab9e9">0024</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b579720c-7efd-4391-9b40-5397d2574e89">1.2.840.113619.2.62.994044785528.27</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4d2030d5-d2a9-4cb9-975a-6a8bf518cc65">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8866915e-c3ca-4b67-8d8b-7a5d5d412b8c">Universal Entity ID Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a78b6e15-11f8-44df-a56c-46f18836c4be">(0040,0033)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_afc17b79-736d-43e7-9855-05a53484846a">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5ad9bfc-8d45-4efc-b815-ec4e0f745368">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17f07874-69c3-4808-a206-c1d30f1c30c0">ISO</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d4beb03-50be-4ff6-ad6e-c34924ce0365">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0d4b5e6-dfa0-417b-b59d-0af03b63704a">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_49ac9ad3-5160-4e88-a303-43580620b409">Modality</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8991a54b-a355-4dd4-bb18-cedbdf74b723">(0008,0060)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca7bae1f-0ee6-4b2b-b485-5d1b5552c6b4">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e199232b-e661-46b0-a86c-ca1343ce9065">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4c42603-9415-42f3-ab1c-0bfb7706b484">SR</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e887c40-6281-4c50-9461-0da65edbb1c1">Manufacturer</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b284d620-465f-44d4-aa78-cdebcf6881bf">(0008,0070)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56a4207f-d1e3-43bf-a035-7e208fe8e366">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_727cf5fc-0dd3-4900-b73a-b7f58731ab05">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1539a09a-36df-47d8-a8ef-39eca8fff2ef">DicomWg20</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b7eebdf-f278-46c6-a956-0481cecaa87b">Referring Physician's Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d311ceb0-3d2e-466d-8561-a3b1f71f02fe">(0008,0090)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1140a04d-0f11-44ad-a8a9-b4ea6f5937ab">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efd242c5-ae8a-4f39-994b-642ad2bf1503">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff9d17b9-bac1-4c59-ac67-5b898a0bdfbe">Smith^John^^^MD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a4bbc11-51af-4131-b56e-160fcf0f37ef">Procedure Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac0a82c3-e812-49e0-880d-6b4224b9dfa4">(0008,1032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_368df37d-6da8-46c9-892c-6ca7303534ee">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1214ab01-1b95-4c8b-ba64-9e1c8d44ecd5">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_333a7b33-493b-4eb9-8aa7-bd6427a836d3">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4ff99e5a-5c89-4eb9-bee6-388e449ac9f8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_398f92da-f6ae-409b-8462-d1ec8d8b762e">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85509a2b-f35b-4ba5-a9c5-c56b4ec5fc89">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_893a1c04-9800-43a3-ae0f-4565177c2b3a">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b93e2c00-8bc7-45b1-8b13-1deed44a66d5">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f65ab11-3a62-436f-a841-a7992b4cf028">11123</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2110c862-7f42-4a02-905f-e7f984dd3136">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7075e56d-284e-4448-b076-2ef8389d5933">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45ee7f97-0906-4deb-ae63-04c52b4a1979">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb4a6cf1-1671-4a37-96e6-f3294e474f5d">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75e1055d-d3a2-4928-893d-c7d8dba52f09">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9815c1d8-59e0-4533-9352-b39de124baed">99WUHID</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f29b631e-9740-4164-8a85-50d38a818fc1">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_207e2eb4-6742-4846-83af-9ef5ac94161e">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1dd286e2-e7e1-48b1-b2d6-b673e8f307a6">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71a6b375-edce-491d-9076-d4f04380983e">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1eb8f7ab-2179-4966-bf86-9cdc6ab2bd3c">000c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9a12880a-a413-47b2-aa0d-7ea201d35cde">X-Ray Study</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3bccd670-2e90-4974-8434-8ec64128ea01">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_63232c54-3ac0-4c41-b182-a4eccfd80afd">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b5adb5d-d3a9-40d7-87ab-86064623ce2d">Referenced Performed Procedure Step Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6c5a09d7-e5e9-43aa-8734-fc6a6219276d">(0008,1111)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b02fc26c-9841-404e-a465-8c4c881d01dd">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1e1c46b-9cab-4f86-9f71-faf598aa5162">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f66a8aa3-639b-49e3-a58b-eaf5efe06d40">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54176a70-a1a8-456e-892c-b03f056642d8">Patient's Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51e59bfb-db86-49be-b9c2-0918a33b9f41">(0010,0010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d9684cb-6488-4818-aaf5-a4f574cc3909">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4682e90e-a6f8-46bd-b2c0-284d4fc4fa85">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a08ae969-1d38-4749-a95f-3c7bd804a4a3">Doe^John</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8bfb62d3-6a8f-4d6d-9150-b508ed7bc54b">Patient ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96e984a8-a74e-44e7-b23d-f77cca2ab9ac">(0010,0020)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_505a22ca-5194-4e3a-a503-6b4d316801f0">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9da83ad2-16f2-494e-9e6e-100e2bba147e">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1df5546-59a3-4bec-a472-97c2d124e8e1">0000680029</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6deb4dd0-036e-4d57-8142-f65ddd0746ba">Issuer of Patient ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a6ca269-a897-48c0-96d7-52bedde97df2">(0010,0021)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3611e2e8-d93c-4afe-a9d7-5b5ca4835df3">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a60a663b-2737-4030-bc3f-e42afb4a582f">001a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdbba7d9-101d-4d25-a4ef-30a61fc0b283">World University Hospital</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76f9d8a7-5809-4e96-b347-790f4a7d30ee">Issuer of Patient ID Qualifiers Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a6f7a37-96a5-4a6d-8a83-2968b811fe93">(0010,0024)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb4ad4bd-6f64-4c14-989e-420b68a4741c">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1353354f-15b3-4757-b5aa-403815687a3e">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30d89cba-80c2-40b7-924c-5379fd430e5d">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5913bee-1b24-498e-bb68-9107c1236707">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d22cdb3-e08b-4f4d-8df3-47497ecd5e76">Universal Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_04dabf14-6110-4efa-b229-70925e88e531">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a8dcfc4-050b-44af-b7e6-aefeb06e3343">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88269d11-ca50-4546-8e63-e25ed62171eb">0024</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_adb48079-3ef9-4ed4-8450-14f949ab1caf">1.2.840.113619.2.62.994044785528.10</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_85d27352-957b-42d9-9c4a-4713c775720e">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a399f39-7657-4575-a7ba-ab34b9be45d1">Universal Entity ID Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9a9dd503-212f-4f78-9442-b12b106151f0">(0040,0033)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b08ee37-a4a7-4736-bf88-25a309d2b9fb">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_97458763-fdff-48ce-ba31-a6ce27c505f2">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c287cd1d-07da-49f1-9e5c-0c3c32ad32a5">ISO</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1ae7f59-addd-4c6a-bbbe-a04fae8d9784">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ff31d89-4fa9-43f0-b80f-83a3880fb616">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_491a0e8f-39b3-4cb3-89e4-8ba6315b4ff0">Patient's Birth Date</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bde07a02-439f-4703-a780-71cfbe924d75">(0010,0030)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83c96869-22ae-4e00-b2c7-4d20137485d2">DA</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81c9792a-85c3-41bf-910e-38557f10f314">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a649c51-d378-491f-999d-7a6291cc41bf">19641128</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c25060ab-8ece-4a68-81a8-06d920d2e38d">Patient's Sex</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a422f6f-ada3-4b9d-82e2-17f9e1415533">(0010,0040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2994b713-3f10-49bc-b70c-ef4b1b31b933">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6553bdf6-303c-4437-82fe-b7acf4501210">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5212f178-110a-453f-bab6-b5ea300a7f02">M</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4d74a12-4fa1-49c5-88ce-5b2b8b0013d2">Study Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f3b2b629-4df9-4414-8328-c2e9afa526b0">(0020,000d)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_97dfe980-0c16-453b-94da-4bf9470747d5">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_918d88d7-df0a-45a7-997a-a2118a5617ba">002e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6416818a-1bb8-4aba-9673-5e39e5b933af">1.2.840.113619.2.62.994044785528.114289542805</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4899c42c-d60e-49d5-b26f-f99e2289efab">Series Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77d1764a-2a8a-4705-bd9b-d22284b865dd">(0020,000e)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5bb72177-689e-4634-9907-df89c71da36e">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8cc8960b-a587-4e17-8234-20ed7afa2f4a">0036</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d573d9ed-fed7-4c38-818b-cc0171902ec5">1.2.840.113619.2.62.994044785528.20060823223142485052</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6f90fc8-be0e-4385-95e0-dc4572fc43a0">Study ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c52d29c2-e1f1-4324-9f0c-30a9624ff92d">(0020,0010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_80bf40b7-dcc6-45a2-8b15-e89cdf65f6f1">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58c8143e-4763-4045-8cb7-dd2e806001ab">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3197ac47-3360-4e53-aa04-d8188506e229">10523475</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4ae2de29-2b41-463c-ac71-19c665f9427d">Series Number</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58248607-518b-49e5-aa60-cbccfdd4d561">(0020,0011)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_473f09f0-f65d-49ee-8368-f812a64f8400">IS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_24663f1d-8763-44e2-b55d-799bbf73b795">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dca262da-ad4c-4c35-bb83-185283015fb1">560</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1234c2d8-65e8-491a-964d-25863380b5d5">Instance Number</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_453522e5-bd87-4c92-8c7e-d1f02b5b19c8">(0020,0013)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f5c663c-8a7b-49a2-8219-93afc29c8381">IS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2126b9af-a131-4a1c-999f-8f5cc8667d91">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_80ff1cc9-19a5-403d-b20e-c481760fd301">07851</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4c111d4-ef8d-45e5-b426-c5760543a5b0">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc30104a-c7b6-48f5-9017-d960b5af0e06">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa288bab-47b9-43e5-a46f-aeceb048e2de">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9c6ea6db-c468-4104-806c-9b02f14fb665">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ad050c5-20d3-4e37-b15f-741be8c1eed4">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f5bedfa-ece2-4f1b-82fe-d91eae3fc40f">CONTAINER</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a02ad82-c38f-4db9-bb16-945c2ca61554">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38bc65a7-ffc5-4ee8-ad23-1259f88d4d73">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da6157d2-aba7-42cc-8264-61c1ea9671a9">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9eec70da-4ba3-4b21-a5c6-cc44643d0d29">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65949d31-43e3-4578-8471-06dea8d87756">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71e2b56e-48d8-454e-8d4d-907e92d4ff5e">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6cc2a329-9295-4f3f-9b1a-3548bc3a5063">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65518862-6396-491f-b24e-562f6d350417">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c2863ffd-e2b4-4f49-b33e-d0346679a301">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62357c4e-f675-46a0-b2b3-78ba66ccb092">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_475d5600-e480-450c-9a6b-114a5935cb67">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03877e8b-f938-45bb-9b7d-b2e094358e29">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa914ed4-09d6-4b94-8759-faa4583fc9da">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a388fd01-c1d8-4cfa-a743-93824f3e3f27">18782-3</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7f25e8c6-15d8-4dd4-8778-f71581f2e5eb">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_360a181b-8bda-4667-83ef-3a0fdb9ab23f">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c43f57f-e83c-41e4-b0b0-6520ec238b51">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1dddb91-3bbd-4c7f-85da-e3b87ac5c4ba">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87d038a1-64bf-43de-8898-767f99af8e69">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9598f57-eef0-4c2c-8840-3aa1ab6064a7">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d4e57380-b065-4213-9093-7d9a79d8bdd3">LN</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f133f7bd-767d-47cc-8698-0485b05ae60e">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcaef0ec-81ed-450b-aec8-fe88364a18f8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28b7816b-f8b6-424e-a546-398057cdbe1d">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7f7ae61d-eba2-44fb-83a4-660cd4dc3b1f">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_174ca6b3-ba1d-46d8-a1c2-687965ae99e6">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9ed162bb-ac8f-41e9-bf88-b83b5dbd6faf">000c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_958c93c5-e6ca-4b0b-b5aa-b6247e59f0b8">X-Ray Report</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fcde1b3e-af72-4b63-827f-57506e567dd7">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08dc87f8-b2fd-4fff-b073-b6789b846c82">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8980461-7df5-4ddf-ad97-7680dfda0308">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab98cc14-86c1-441f-a11c-39219886a7b2">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_504b6ba5-591d-401e-95df-037d9fb80eee">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcf44b2d-428c-4cac-98cc-cbd003444d5c">Continuity Of Content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d146305-247d-402d-8ff2-b304f7fccca9">(0040,a050)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83e240c4-30a3-42cb-ab25-2575271d4b8a">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_209b0fbb-31ff-4473-9a4a-db2a648d12c3">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a1be881-a7c0-4b3a-b3cc-96b2886e3c35">SEPARATE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2afe655f-d162-4f89-89c5-34a070416c9f">Verifying Observer Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be156db9-b29e-49e8-b475-5b2599aea56d">(0040,a073)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_642e782e-ed27-4c87-8f9c-db61c3c8e84d">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c2392c8-69ee-4d07-a3ee-06e310357525">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6dc313dc-a28a-4737-bf13-ed393f65c4ae">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1cd3bff0-a1a5-4a80-b0c3-f0876c17ef77">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03096064-5fdc-4ee4-a5da-3547e807c110">Verifying Organization</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9a2f3ff5-9e33-4b38-9ea5-850a7532ff02">(0040,a027)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a5f687be-e438-4469-a094-3c7d70c20435">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_427c3129-c601-4003-baf0-1528023e646b">001a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c84e6fe2-f1ef-45c4-8c31-7161b3a26fbd">World University Hospital</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76ce018a-401e-4c7d-9169-2b40fa00df61">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f69ee31d-970a-46cc-bad9-228bb16c4769">Verification DateTime</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_984ba184-2a6a-4c7f-8b30-a3ea30b4af0a">(0040,a030)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98358bb2-cf29-4461-9c18-b7f17ed1f0c7">DT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44692dd2-d854-4ae2-af02-db308d9161b9">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_80d5b96f-10c3-4556-ac12-7da450eb2fe4">20060827141500</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a7a7b3f-81dc-4114-84d5-2213404719f1">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c7c93d79-5753-449a-9d47-392fe0354d2d">Verifying Observer Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f772d16b-fc8e-4827-9690-884569a22344">(0040,a075)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36b4237b-b174-4d18-aa6d-60fbdf047420">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_27c5d970-2bda-4add-91af-976b13cbc543">0012</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcb0bd50-6967-445d-8c06-de84dd48b3df">Blitz^Richard^^^MD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84f401c0-de96-4d6c-a4cf-a3f4f4f79f87">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4a6bf61-a07d-499d-9241-be44ecdb680a">Verifying Observer Identification Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_667acd2d-87e9-4af9-a1c4-635cd0c84eda">(0040,a088)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5842c0b8-e4b5-4465-803a-ceca52bab6ef">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3c196bf-2582-4647-aa5b-0e59e6b255c3">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d7e8b6bc-d55d-43e9-9ac2-8b431ddbd62a">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5ff4efa2-e235-4a19-b11d-759af4ff0f21">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a97580a1-a8f1-44c4-b4a8-90de68275c85">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f575232c-c83e-4732-9a35-63741eb4037a">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41d0ce17-24d2-49ce-9075-b06860206e0a">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e9b24f2-989b-49bd-81b2-05607e029aec">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f1d6a3bf-92d1-43cb-92b8-a84c0fe5892f">08150000</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_97ea5404-f907-471f-9477-0f7792a9c417">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fd1ea5eb-8246-493a-8c7e-cfedafc5def4">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_68e8d3bb-a587-4588-85f0-8918df67393e">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9020a30b-91ed-4fb6-81d4-54ce0a4949ab">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_39c3d961-c38f-4973-bfd5-46988241dd66">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c807a573-ad2d-4bac-b79e-48f017893f0f">99WUHID</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08d54555-900b-46f1-af7e-ca81e75ad305">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77b8aff3-4c23-4fec-8002-54b6f8a802b0">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a32112b9-9c42-4684-a1e5-4de256e495f4">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8473ac4f-0b43-4d0f-a8dd-1232e109e633">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d1d10fe-5d1e-4e84-b9e5-6b5b06c13381">0016</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_01218a04-83bd-4ff0-973f-c4a69bd6f83e">Verifying Observer ID</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bb0f9d9a-f72a-44d4-b548-da3a21349ded">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f52bfa08-bd24-4016-a75f-317bf0ac27cb">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a6641ce-6a4a-4d6a-b72f-6056efe50a4f">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4724562-2973-417d-9c77-b811c7104a9c">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5186d1d6-384b-433e-bf45-c49525d92e74">Referenced Request Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52212a66-b6d5-4551-9cb2-f8c3a55e0e41">(0040,a370)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e58a9d6e-13bb-44a9-9603-06cb5379f9ca">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d11f661d-eca2-48e6-9921-4f857c3ee8ba">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_667450f7-9cc5-4648-a145-a6e3e530d7e2">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4af91bb2-3b49-4393-b60a-bffc2fa6308a">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58dcacb3-78b3-4c52-8d01-10d6c25e3641">Accession Number</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9084e86d-98b9-4fdd-9198-3927af8cfe46">(0008,0050)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6fcf7626-3b08-4d8a-a943-801d9f7b4a6d">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4850cfde-3840-4fe0-aa65-3630391f1d2e">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d874ca7-392c-4011-acc6-d28bec735d91">10523475</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67e4c9ec-9c1a-44d3-b5c7-8b40fd7b547c">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5931e6ca-9da6-41dd-953e-b285bbc3cf4e">Issuer of Accession Number Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7551bf39-9e56-4f97-9e3b-3c9ef78c0247">(0008,0051)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0ca7571c-cb70-4f45-a350-231239515674">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_72393df2-e01f-4e1b-9dd7-146bb0a2d7e4">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_281a105b-338e-43ca-8a66-04ec93b9e622">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b2a8038-014d-4655-8bff-027b7026a19b">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2bffbdfd-0e76-4b39-a06e-00b1a28e2a84">Local Namespace Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38b897c5-bb97-4bc1-9513-b0ab1da3ad58">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69c43c12-dca7-4a87-8245-056053988f6f">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_957b603d-799c-41aa-a8ec-3c43ad18c448">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da4df859-e30f-4771-8027-163d9f6cea60">WUH-RIS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d45176ce-8b9f-47a5-aabe-6ddc01243ef1">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb73a467-ad0f-4c99-9d0f-c9fe90c6e7b4">Universal Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73d3c4d9-c84d-4058-9c55-e0e919248543">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d8aaab2-2664-484a-ad8a-1ef89fc2fdca">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db4059c6-5cb3-4e93-8120-ee77fc13d049">0024</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e180a07b-54f7-425a-95ce-4e8ebd213f55">1.2.840.113619.2.62.994044785528.27</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92d2e779-1dce-4bbb-9997-1384fda4c232">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_df785bf8-6d83-41c3-9bfd-ea4d971470ac">Universal Entity ID Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1019a2fb-45dc-4cf6-b02e-638f6cb39963">(0040,0033)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cfe8d2d5-107b-4b3f-805a-459c10697a65">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d348f78c-cbb3-4827-822c-c5eb01cb0358">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_29e66bce-9ae8-48b2-934b-298fcda60195">ISO</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f3b03b56-26a2-4f57-942f-46472abc9ce1">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff56c19d-1b2d-4d17-81eb-ab326a695e99">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_00d51e17-de5f-452b-89bb-08bb5c719107">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_365a125a-0e81-4775-a903-b945d6277bab">Referenced Study Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e12b3769-9ff9-4630-84e3-1d9000dafa87">(0008,1110)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c9d6b8ca-23fe-4321-b721-01f285be16d9">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9b69a789-a786-4633-a830-2f73110096e4">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f72b175-5e01-479a-acfa-da0b1efced89">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4c591c6-8038-46e9-9f63-e5f9b01beed2">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a77687b7-4e4b-455a-88aa-431bae5bbcee">Referenced SOP Class UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a1d27a9-f7fa-48be-a5bb-baaedeefd318">(0008,1150)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e491812-753f-45a8-8f6a-2ac0e17433f0">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1a03a5e-6eb8-4514-bdcf-4e6f055c3ffa">001a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3139403a-183f-4a1f-9e11-9850669f41b9">1.2.840.10008.5.1.4.1.1.1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf9e162b-5faa-4e66-8c83-d904931acbfb">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1989f292-a926-4d20-89cc-179824401636">Referenced SOP Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c6c8038-6f8d-4751-bd1e-4519254c3269">(0008,1155)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c3ba5fa-4dcb-4685-8493-ecc81b8d697f">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4232f1df-b0b9-4930-896b-37034151ae1d">003c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65ff63c7-f15e-4093-9f48-08e2a494c422">1.2.840.113619.2.62.994044785528.20060823.200608232232322.3</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0feda6d2-ccbe-47b1-8516-15ff482c0c97">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3caf6d7b-da90-4549-a7ae-5485a99909e1">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a2770d53-03ef-4c1f-bb08-014ecf486843">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_847e2749-72cc-4268-bc1c-23434d6d7825">Study Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28979f9a-ab0a-4cca-adf4-6ecca8917bbd">(0020,000d)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_baf3050d-75d6-4f2c-8c1d-3d7351ba1194">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1574652-7c20-4345-86c7-32beed00648e">002e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d7ee662-dcb6-46bc-8973-97a7828ab3ba">1.2.840.113619.2.62.994044785528.114289542805</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b0d5ed2-cb74-42ff-8da9-a4c79406f8c3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8f3e47db-9602-4fb8-85cf-61f5d939f691">Requested Procedure Description</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4d855eb-75cc-48e8-8b73-1c6752b51594">(0032,1060)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9a3072d-d389-46a4-96d6-39dd7f41795d">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25b74a4f-23bf-4b0d-9f9a-03c9d8fa59cd">0020</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d422ae3d-4023-43f9-b45c-43a04085829c">CHEST TWO VIEWS, PA AND LATERAL</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a7b3308-687a-4eab-8be7-1364a694b4b2">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec068f89-a0c1-4001-8635-4d908e541f0f">Requested Procedure Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_adddae98-107e-4fb2-8439-cde10f17deb2">(0032,1064)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76406f43-b7a5-43fc-a506-8dffa3a0e220">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e1100ff-982d-4e03-bd92-0c346720a6c7">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0eb97eba-30d6-4563-9979-29816c0b7e3a">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d35963b4-0275-4350-b237-669379797906">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b52eedda-e7c5-47e2-b0c0-57b4c12f4893">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6fd9de1e-b131-4de5-b574-28758e2fefbc">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30744aba-7cea-4c37-bdbd-4eb51c2f7fde">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88332139-a525-46a3-84cd-9290f6f419ad">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff018429-e887-4e46-869e-78dc398472fb">11123</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4575514-3b69-4bf7-bf0b-a921a97b11ee">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8e17a3e6-d624-4f4a-acfa-af4d1c34c111">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f341bffd-bdb0-4b46-8631-cc9b9f4257ac">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_214abb2b-5247-434c-84b5-23568048cfec">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_160f81f5-fc84-4469-a436-25ff308c4eba">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab10663c-9431-48cc-b2d6-52cc40aedcd5">99WUHID</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_28ef55b7-55ec-4f73-9202-20b63be94198">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51875999-32f9-421c-9a86-d4b4eec88aea">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c0c48b3-275e-4c9e-abab-8a4462c38b0c">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_00d74ae3-b861-46cb-83c8-6df65d8f442c">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cad4d820-073e-421c-b926-c91ce09bae46">000c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6c92c4ed-45bc-4693-8172-5f77bee9f208">X-Ray Study</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc44c2c6-c219-4b21-8693-bf17b4ef5bac">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_de9dab31-89f0-4150-8994-954d078222c9">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f820344-6578-4ebc-bdf1-8dbf3129ed32">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4de6ada7-af62-47c6-8dc8-4f010a02f2ea">Order Placer Identifier Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_11bc60aa-3633-467d-bf27-4360e7202501">(0040,0026)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9b8117ea-0879-4215-97d3-9c6e06f373a0">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f14396c2-23ac-4f68-a535-d7acc0e43d70">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_31fcac62-1d71-499a-8afb-605d45d0b964">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ae2097f-6e0c-43b3-b3b7-cb18dbcb7c1a">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d55ae8ba-1bd5-4ac9-9734-503e7118df9b">Local Namespace Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a69c1f88-33a2-4997-8c08-76647e4e65c6">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aee59601-16e5-48e7-8a98-1c7dcf443a5a">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62f00876-65a7-49fb-8d3c-d9893791d9b5">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f347c0c-6caa-4e16-9964-f31d72aeb380">WUH-CPOE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd09237a-d84d-4042-a80f-36f1c2cf7952">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77167000-0a04-4b6c-9f90-d6ad6ba52c32">Universal Entity ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a81c8b3b-8cc5-4f50-88e5-171de0ccd3f9">(0040,0032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b4b496d-4979-49f7-ae81-c0126d356fec">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5ea77aa6-8093-4290-8182-694b1b28cec8">0024</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3cbeca66-a899-4fc9-9b3b-7d0f79c33e8f">1.2.840.113619.2.62.994044785528.29</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87864b86-f878-4117-9b74-c7025a253c35">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4c19080-7d7a-4228-ba87-6b616d610108">Universal Entity ID Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1b0ca619-e206-460c-8621-6b28c54e8bc7">(0040,0033)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47785a74-4399-4ce4-8b20-8a2837bda379">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4ed62c3a-a0ab-4092-9466-3e7723bea92c">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a47c4c4f-648b-4ab4-be66-f2880e0b99c5">ISO</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96847afe-34ef-4438-bffc-d5c07c54c839">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a876dc60-55d3-4b3a-9d69-5437331820df">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0ee000b-bfaa-49ed-ba2c-6d3209b58f1a">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9272fe92-7de9-406c-84c2-65ae709c7b9f">Requested Procedure ID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d376dcab-0f82-4d12-a37d-675c3b63e469">(0040,1001)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44035c1f-2bdd-4b0a-afae-d8600d4dcd75">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83979fdf-5d61-4c89-8c09-3f4303a85469">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c58bb37-ac92-4901-b90f-a99e2139ef50">123453</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8eb67a71-b24b-4136-9d99-22e7b66ef86f">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_afb939b3-9530-47bb-a9f5-202284b6c68f">Reason for the Requested Procedure</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69078b4c-f5b4-435f-9c96-a624071e5c13">(0040,1002)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b0e56db5-5d07-4e7a-bd0a-048cf9324bfd">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_11ecdcfe-7633-47d8-8460-31b54ecb71d5">0014</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c0cb18b-54cf-43d2-81a9-f353cc45dfa6">Suspected lung tumor</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdd78575-3f98-4eed-8a25-520f90eabd44">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c90707f-206f-46ac-8375-9ab3b049bc90">Placer Order Number/Imaging Service Request</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1af77e8-48ef-417c-adee-ccdb09701e59">(0040,2016)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45c34858-6ea8-4887-8f20-ad59bea54786">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eca0d41c-bc8f-4032-b334-29f73f1605c1">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_982ec149-6161-4739-87ba-9cd996ce86e1">123451</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac48885c-5ceb-43c0-8c07-2901242b1de9">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d43ed14b-c115-41ba-8f98-527dd0e8eaae">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b0d0a2e2-8769-49f7-8e79-d6bb8f001ce5">Performed Procedure Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9cc2fab8-8fc8-4518-8692-e134805d4d72">(0040,a372)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d6c1245c-c174-4ace-a50f-7121aa942bf9">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_392b728e-a989-4362-90df-af9a03f0b236">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b525235-8978-4c3a-afa5-f4d788ebf6c2">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_584be38f-9768-4413-9072-87804d2d4777">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d73abd82-bc03-4ff4-81b5-5cd0f169f68b">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f01b49b1-3917-40ea-91ec-9bdf8cd6633e">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8007378c-d777-4861-b049-e77144ccefdd">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_434872c3-6127-4677-a014-467dcc499766">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_292407d6-b9c0-4779-8d9a-731587692036">11123</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d8d15055-5178-48c5-b72b-f4e9d6f9582d">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af41b3f8-c8dc-4458-8104-e4ec4a27102a">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69e77a2f-0165-45f4-8420-4d91f809a3aa">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2df74a12-58fc-4768-882e-ed198289fe7d">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_beb1c7de-a67d-47f5-8967-006a44ef04bc">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16be62d4-1b31-42c3-bb09-694612950acc">99WUHID</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fddb1ada-3d31-4a69-a039-dcd74a6e87d7">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af74abec-40c1-4bff-82db-2c6894428766">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab28e78a-8fef-4bb2-9fc2-ebaf0c9b3f47">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef2a080b-0ed5-46a6-89be-9fbbede013a4">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52b3a312-66de-49ec-a025-691223f936cc">000c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_94ae8214-d4ed-4dba-924f-5bf9f204f266">X-Ray Study</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ba11f00-df06-46be-a54c-62c62d718204">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51df04ad-f593-45dd-a18c-d26d5d36e1c4">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e4f0d5f6-23fc-4427-85d0-91135f639d51">Current Requested Procedure Evidence Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75098c52-b2cf-4c6d-8b34-1bc7e607f338">(0040,a375)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_56f80129-f989-4484-a261-b0b9272a42eb">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a3a35d6-a91d-436c-8679-84a731d37440">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f86d33b5-02bc-4a3e-a4de-1ed98b0b010b">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73351915-e37a-4f2f-88c9-cdc93aae5149">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc43f8e6-8bc5-481d-b02d-73f295623a16">Referenced Series Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b52739be-fbcd-4d3d-a992-82bc919bdf2b">(0008,1115)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a0b18192-3e9f-4b26-afa0-693500aec535">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_530262df-697f-44a9-b637-a0ecacfeb063">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_235629ca-44f0-4f2f-a7e9-42dacbd14e1c">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0faa2a48-6a45-49a5-97e8-e742955be22d">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98f11a09-0904-4f92-8acc-66851427df41">Referenced SOP Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12943bc7-99b1-4718-87a3-14ea4d33ca12">(0008,1199)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef325c16-d028-4a29-bf77-25b153db0792">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcf3bea4-33f4-4610-bf35-33b8c0432cbd">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf7ad373-9801-4c50-8dc8-f435e59f3cab">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d7a8494-3ac7-46e1-b2a3-22f734e4ef24">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ebf29ad-322f-4b26-a7ac-f68bd60c27a6">Referenced SOP Class UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51049141-2287-4a1f-9415-794f2c5f1feb">(0008,1150)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_138708fe-c58f-4714-831a-cdd70d275a1e">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84851513-09fb-41ed-a502-c031afc3617d">001a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2023742-52b9-4c26-9a98-71bd6042abfb">1.2.840.10008.5.1.4.1.1.1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84c4b74a-8868-45b4-993b-839362d1a55e">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16fbdf30-a499-431c-9fe3-220aa057ceb6">Referenced SOP Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a503e44f-1b92-42a9-b431-b2643a4fe912">(0008,1155)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b97cf1c-b037-4e92-8cc8-3bd0187dd4ce">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f58a15e5-7b44-433c-b9b7-8c8712a0b5a7">003c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf3c3934-bb91-465f-855f-fe9696460129">1.2.840.113619.2.62.994044785528.20060823.200608232232322.3</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5da085ff-86bc-4b96-985b-569f97c70469">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c1062ce-a4c0-4228-94c0-51599adbeddd">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be6313d3-ce09-4acf-bf86-2fa4bffd7475">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b66202b2-e93b-4c39-9e76-abc55c138c4d">Referenced SOP Class UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c2e5a7f-bc27-414e-8495-bef871063267">(0008,1150)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1d8755a-7c1f-4c91-81d8-ecb479f7db2e">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7767e70-c372-472a-b955-08f89132a3e7">001a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a046ae48-6957-4b87-8375-7024ff0ae33a">1.2.840.10008.5.1.4.1.1.1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05b0afed-94b6-4710-b71a-57a51b8b7ed4">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f1459b2c-0a56-438e-85fb-743db1eddb9d">Referenced SOP Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d848d74-d06f-440f-9c8a-2dd53197f6b4">(0008,1155)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9924e08-c9cd-40a4-bca4-f532aedab458">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b834d42e-8859-4d44-850d-97aa3ddbd87f">003c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d4c12a8-c6d7-498f-a280-ed8b27e41474">1.2.840.113619.2.62.994044785528.20060823.200608232231422.3</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_500acb7e-3156-4ee0-9c3e-f76cdb1797fc">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1e82b6f3-19cd-4e4d-a872-fdd016fc501f">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42c5fa52-571c-4a0b-8803-dac10efe06c7">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cf025965-b118-47bb-b1bd-00814c73096f">Series Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9471a4f-d423-46ba-ac9b-c1a5131759fd">(0020,000e)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_049b193a-a933-4724-8241-1bce5cf8c618">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_94975a8d-3f42-4479-8755-5c8248df0e94">0036</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8598cd91-9dcd-4dce-9cf7-baaefbcf62ff">1.2.840.113619.2.62.994044785528.20060823223142485051</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88b61505-004b-4d95-accb-dab13cb2dc33">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9b15c9d7-fd5d-45f2-8771-2536151b2e4a">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6c06a336-c6d7-42cf-a601-63ded207da65">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8e74b2b8-6af0-4537-acdd-83663534217b">Study Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b4380527-7005-4257-9006-d8a0de072d9e">(0020,000d)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4170b77-0444-4b65-aa78-d94c2884e9f0">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ebc009db-5efa-4275-834c-44e6230f53a0">002e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_251d762b-387a-49c2-a36e-3cea5fa48503">1.2.840.113619.2.62.994044785528.114289542805</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_76b546da-d88c-4cac-a5b7-72357d1959f3">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7275e304-fc14-4819-be88-c32545fa378d">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8597e03c-1b1f-4842-a81e-93900343c4ad">Completion Flag</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81612577-301d-46d0-b656-acff87a517a9">(0040,a491)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9da5f71a-a429-46fb-80e2-33064406469e">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7dcc7d9-77c4-49a4-9ebe-0fd4ec5cad2f">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6cae6afc-87b2-4110-ac18-0dd2332fc396">COMPLETE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b49b5f3-d710-456e-b716-6e97fc534751">Verification Flag</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12b4b3e3-0a8b-435b-b4b2-aca1a109d459">(0040,a493)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d458fa41-ff40-418f-9613-d98bc7ff2845">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7472e04-18d9-4b43-979a-e87a1676dc4a">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67e8141a-90af-40af-96cd-4a59439360ae">VERIFIED</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_203f8c4f-4c79-43ba-9e45-c49800edbfd9">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d40d67c-f1da-4828-9761-6e7328bcec6b">Content Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c09b9e10-0079-4f25-98c9-475d8413a78b">(0040,a730)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bac11253-1def-4259-ad8b-2f351c33ec45">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7aa706d-ae1d-4b6d-b96c-58c85c4fbe1f">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1cee41e5-3ce0-42fc-b3ab-70bcb96712bd">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad073c85-ca21-4f3c-9e0b-6ee3382f9567">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40e9b16d-5153-475e-87a2-a9ff707037c0">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0fc23a84-39ea-4d30-90d8-f6681d00b815">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1e82742d-b988-4fef-8c4d-e524d2bf9696">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_149d796c-7690-434e-a953-d6f9cb1e5ccb">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1da2668c-b742-4aef-b7f6-77d34a6c6364">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f24f1dff-e26d-4f45-bded-4e136e6d96e7">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_603f5cae-f89d-47f8-ab13-a0e3a75ee7ab">HAS CONCEPT MOD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2d48f9d-1d0d-4030-aab4-7f4ed818a940">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c38fe023-7a79-46f8-a0fa-76813d760c12">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ebef1b29-fb7e-47d5-8a6f-c9da675ec4e2">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83dd912c-9131-49d4-9ba9-e6a204537e16">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3e34839-0868-4aff-bc1f-7d2400b8e7e9">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8bf8a585-933d-4730-8e11-69247dfa4630">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_beb73ce8-49ad-4659-abea-652de65a0661">CODE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_167f1719-33b1-4bd6-a417-8cc50dfa03f2">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdeff450-8b1f-4471-9db9-45652eb41fd5">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bfba1dca-1a46-499e-b175-9cde4f4e970a">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5290d6d5-8577-416c-89a8-8859845d3234">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_963d958e-c16d-4897-8945-98ad254bab07">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6ab7f353-74a4-4bb3-9f46-0f41b65a8c4a">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_79eb9471-2375-405c-8c26-637ff17e8ebe">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_338608a4-4e45-41b4-844f-d74a452aa86e">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_68998639-bdc6-4af4-975d-ed247b214ca0">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_729e9d4a-0f7b-44f5-8caf-15d5007e72ca">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_478f39ea-b9a1-418d-8c9d-cdc8405b9190">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_743c42c1-f467-4809-86ab-6fb83d134b40">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59472ad8-7f2f-498d-b497-4532639e0da2">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bea0240b-57bf-4bea-ab5b-d526406f9c45">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d16e73da-a539-46a7-a924-b4a30a3166a7">122142</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be6246cb-0179-4fda-a094-d28dfa9f28f1">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16a1f0ce-b067-4d6e-87f6-ae0ad10a56dd">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8a9f8be-f665-4170-ab0e-5bf26fdf9c91">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58a2605f-51f6-4f3a-a71d-de1019b65954">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9660262-01a1-440b-98b9-49b435d33511">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_93895ad0-e87b-4942-bb22-46012b44c006">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8349650-49a3-4e69-a536-fc68c4571de2">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59439b7d-fa15-4158-b67e-93593c48e753">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9f600dd-22d0-42a7-b780-3a9297d1128f">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_81e8b680-be9e-480c-9098-10394cb553ad">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c0b92e6b-2315-4ad9-a2b8-dce4f855b6c3">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c2be5fb-9d67-48ee-b9bf-e1f3186c9c33">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d9a6029a-7d8b-4333-9821-a1d75cdb672b">0018</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36c4e197-caf2-4a83-9028-01a9979d2c43">Acquisition Device Type</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7eab84db-c7bb-4ba6-b9d0-3f38f1eef60a">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec8dc4a1-c335-4fd1-ad40-9d113510c8c1">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ae521e0b-4d58-4a04-bfff-ef4564ca59d2">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba898b14-8237-4e82-a83b-1ba4bb7468e7">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b224d586-8687-4914-9be0-bc6b5e35ff20">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_922eb80c-6543-4d1c-9078-a5b4a3673dc0">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed8da38c-f1bb-4c48-bfff-8cc539968273">Concept Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca2b9d37-482f-4f83-a80e-b29a01d25f0a">(0040,a168)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdf1ac90-66c5-463d-ac30-f2bc50fa4154">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a9ef8b1-2622-4d09-a6fc-cca9a529cd37">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1b2ed43b-87f6-48f5-9aef-d141104a648e">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96b48785-f156-4628-8a4a-4f3f9333102d">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9901e213-f954-4feb-99f3-f9f0885ef87c">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82aa9d5c-610e-404f-a5f8-998dd5e4b590">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b11e0eef-a0f2-43d6-9535-e47567243481">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_097e547b-913b-48be-bd67-4243bb60a41f">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6589f693-a33a-424a-aad4-636ac5e55e93">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ae335746-1f73-412c-b875-f607b9d4fd6d">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ebe86fbb-2f96-457d-8cee-d9163f378b19">XR</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_172d07e7-d2c2-4055-94ce-dccd05739449">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a71bfd8d-fb03-4541-b193-7a6d5d6a41a5">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15640eb8-b5db-4187-bf3d-59af9759397b">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8abc2a55-460a-494a-8d00-5eb65a319c01">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_739230ce-aefc-4b0a-bbbd-f9f6e08b5ba3">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d3ee5fd-589c-42ca-8a61-de45bbb966ab">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b24969d-ecd8-4021-8043-0e71fd2fff91">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_892faced-d0aa-44c2-8546-acf444c80bf5">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4dcc37e3-212a-4fc4-9bc0-dd478a43ead5">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7850137a-7492-45d1-a2e7-0f4d7ed3c7a4">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1434c150-6cca-4715-8bc0-b5dc87b5e851">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8beeb65a-f6d3-4682-95c3-5c0c197113c1">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47fa9d4d-2582-45c9-959d-5f9fbbc36434">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a478332e-2dc2-4e2a-84b3-74db0429fa28">XR</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5fff557-b99e-4ca9-bc5d-c94bcc22a8d4">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4178511c-64c9-485e-99c3-505091c1e4c2">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f14bc804-ee6c-4191-aaa9-e7e9240c9c4b">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_21b819fd-88c5-445c-9ed4-b82c7437e2bc">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab19ecb8-5d1d-4e68-9ca6-582ed68a66a5">1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7b78526-7224-44e9-9d7c-3ce68b230b45">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_def4e2cb-4e0f-4226-957b-e7d7db507d1b">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_24ef5ca1-0ec1-4298-806b-20a3b98ce370">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20fa0239-aa61-4c01-8b56-763db6886d0a">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3e42b5e6-a128-4c1f-b3cf-09424ee0d9ec">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_54140cea-67ab-45cf-a7d8-f18104628f46">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9302bf64-04c1-4512-a24e-ed58a98f4121">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ea31512-ec18-4674-a285-eced7d49968a">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_608ba423-7cb4-4a5e-af87-725bbc776326">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50267ddf-95e9-4a3f-bf41-495956e0d4a7">HAS CONCEPT MOD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2bd0ad15-dc35-4481-b8a6-09bc84979e1e">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9c77a4f-bd33-4bdc-acc0-2ade1995da74">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bbd660e9-68cc-40e9-9dcc-fa8c600be3ce">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6220ab11-02b5-471a-aa6e-c2aa58151527">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65799704-74c4-472f-97dc-78945b7b906a">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4f00ae95-703d-4c96-b7a4-13b72f227ec6">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c8a7e27b-eee5-44bc-941e-bb11ec417eff">CODE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78488b92-3bbd-4621-b6ad-b2fee35f52c4">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35927090-cab0-43be-984c-657cd9e47b26">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d643d4ec-20f8-4065-9380-4d51d1c1f033">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e6a2f4c-fbe4-44ba-bf3a-f732aa47190b">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_33fbe79d-e4ed-4444-8811-e750a8007bd2">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db0abbd9-9a6b-443c-90cf-0ea0cd4baf5a">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a044de71-eb52-4946-9a46-a73c897acf78">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4eb3d56a-1f6c-4cdd-8178-990fe4ab81c9">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e5360aa-5a16-4168-8c70-2b1cfc0b93b7">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71ed10ce-443d-49b6-8d02-d4d41703aebe">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ceb679c4-8d1c-45ce-a424-46b0e0c8d835">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a39146fc-a040-4810-a849-288f5df628d2">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ac7acc6-b491-4e7b-a6fb-a6ef04652015">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_920a95dd-a0f7-4abb-8930-2cd879879da3">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f680ca17-6bbc-49d9-8cb2-d437f2879411">123014</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b8373478-eef5-41a2-9f4c-fe5e7014f3e9">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_857b4aa5-cd87-4117-84aa-da32ce89f9a8">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da2aa1bb-3970-433f-a1f7-a6dcdc396569">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c83e55f-bfa4-4473-81e0-ba29131b0024">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b30a01cd-fd01-468b-b29e-e0a630d87be9">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d05dd88d-b868-4fee-a306-ddcf0de1984e">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea54ec9b-be32-474a-98bd-9d84d968af93">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cddf7c4e-8eb4-41c2-8a86-036eb79be16d">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b2ecb35-004c-48bc-b237-57250bd8c98e">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a97af1c8-07a9-4283-98c6-04d00114d25a">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b89d46cb-21cd-452c-8867-c1b09e41e6b8">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5c156969-b80d-4530-b86d-9a04f90df801">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc9522fc-f9ec-40f9-96ec-4e9dd6d06632">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4aea380-ad04-4200-8cfe-c9bd1400bbdc">Target Region</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1340f2e-1776-45ee-920d-b6705541d7c6">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7cbf47e-6fee-4d1a-980c-07081281bb72">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f25573b4-ebbb-485a-a43f-38065b8241c9">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac5e61c7-5012-4637-bb67-ab19766bb1a6">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8dadfc7e-84f4-4ed0-b436-dedab886652f">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d88835a-4457-4cb4-8dcd-eb415e3a645b">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d305d831-442d-4ce1-90bf-916d41c8a632">Concept Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50b87655-fbdd-4fd5-af88-13e9f1276283">(0040,a168)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f2dfc34-f922-45bb-a2fa-a01bdd511e0f">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4229e2c2-42af-4e41-9644-af8329e44f56">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75515449-7096-4c49-a855-ab1b1a445e90">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e6b4acec-a5b5-446b-ba61-a54864abc8d0">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_32fb6a1e-c9f5-4920-8030-868416e827b2">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_531bd3c0-54c8-454e-9b40-a710574c14e1">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_817dab2e-34c2-4845-8dc3-1b5607ea9c4a">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a50b456e-25b5-47cb-829d-816ad89698c8">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_34f4bce8-9fcb-4607-b623-ef983630c9e0">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b749f39d-28bf-4df7-bd9f-82d4020682bc">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1aa817c5-e8bc-4989-981e-381f1780d2cc">51185008</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ecef73be-ea9c-4260-8267-d4f78ff74dea">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_02ed3883-fced-41c5-b5dc-f45431bbaf21">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c68a080f-0eae-4c83-b0ba-316f73c9bce4">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55a5eec8-c7e5-4e77-9ef5-29bdd4e2b62f">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb34cec3-af13-4d3e-b1cb-5efe9432bfe2">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3f25111-12a7-4bf6-b7b3-ab0581b3da70">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_355dd396-3c46-4d7c-bfac-1586023bf4e7">SCT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da224190-9517-46f9-bf56-fbebc4d2ba9f">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_886eb523-499f-45ba-86f1-b593fc4b3455">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9254039-6b5c-43ab-8bf1-18977ed8b1f2">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8890a7dc-5f03-48db-9a89-401c46cc566b">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_283a70db-98ba-4c3f-9f32-8928e5ce64b6">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83c128ea-4329-4a8f-b683-621dad207a7e">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efaf384e-5de1-416d-bf71-4c4bb028c2ef">Chest</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b416636f-590e-45b1-9e97-76b903a265bf">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_25414e68-f1b8-4f5e-a4eb-70f3cca104fb">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f94cd01-bbad-4641-90c2-2ff8e7b9edf8">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4245beab-9a9a-49e1-bb30-340acff6e627">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9e0458ff-6c63-40e9-8396-ac6df53980b7">1.2</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74c367da-eaa0-47d1-aad8-ba30f825083c">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d253d50-d8e4-4889-a680-26223d1bd9ff">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_509d0e67-45e8-4c44-8f38-6a28a558648b">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_293a38c6-7af3-4df1-857e-6f9d75fd0378">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6629eea6-4090-49a0-9349-c8eab3ae5542">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_29cd1de1-0ec2-4ba4-a177-8e5eebe7d441">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1cd818af-b884-4ffd-8674-9607d735e589">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_39807230-12fe-465e-93c4-ba4a27dc0084">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_07f7c22d-605e-478a-b08e-dafd0133529d">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1918ae19-451c-49b1-aab3-356636145186">HAS CONCEPT MOD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af6b81e1-421e-4740-837f-f7de18f32e7a">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e6c63eb-61f0-45bf-a094-64e00fa5ec03">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2b10487a-d2bb-4115-a8e2-9f4529bd6d0a">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d74a2ca3-3c7a-45af-8db6-68f306bed910">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8490fb71-e38b-4ba6-b310-c0bc0d0ed4c5">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d69f48f-b5e3-4f72-b67f-fb7b1babf74e">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_edf41bab-406e-4ecc-ae02-7f33c389d95d">CODE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eecb8f20-3238-4b71-8b50-885ab852fffa">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aebfd031-ac6a-4b0e-99ea-4041072f8d58">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f5b1006-d94a-46c9-b640-d620161bdddd">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_91fa79f4-3c35-4e6e-bf6c-582a50418f9c">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0adbe9fa-22f4-4c93-ba88-84e743229d9c">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4035fdf-834e-4afc-824d-01a55bd6a7c2">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7193888a-9d4a-41d0-b4c2-70515210fd15">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a0b7e4b6-2114-455c-8e80-bd5e3c5df1ec">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_10de060f-ab7e-4934-aaf9-d6184b5949f2">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d39135fc-010d-4125-bc72-0ca9d3f67533">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_48d71e27-4d06-4ace-8445-040863829dd0">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a50bb678-310b-4152-b183-cb91523d67a2">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_078062b3-c41e-492c-b712-3bb173473056">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8bd898b3-0bc2-408e-8fe1-c822a8390ab0">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_804b90ee-b27f-4cbb-8d11-04a1fe08be06">121049</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3274d49-6699-4dec-bde1-7da659bfd977">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_efed2828-0ab6-44f9-a45a-43e26a439134">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b0e90239-4fdd-4f75-b5de-fcaaadd8a6f5">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55decc59-cd98-4164-821b-e0666dcc86e8">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_223a205c-f7c3-4201-87c3-4d63aea641a3">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2f430c4-d6f8-4c72-b507-3015ca227422">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4729229a-334a-4afd-9da8-df380ae66cbb">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05e10d71-f829-46c3-987b-2ee99c1baee6">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2725798-97e1-49cd-8784-f05b78c6a67e">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa0e59cd-7c28-4887-9fce-f08358d208aa">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82f8ce1d-e0e1-4802-ae79-0c6dce3486e3">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e96c55d7-71b2-47ee-91ce-97515fb9f942">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9279c4bd-da86-4af9-acb3-f57df441c023">0028</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05b0a7f2-2eb3-4e08-8690-49fcec4a2d29">Language of Content Item and Descendants</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb75abbb-a40d-46bf-9dc4-91f11ab04100">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77e27cbf-a259-4935-b3e9-a4d0c1b3fc30">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aebd78f9-e2d0-48c5-9411-e410b6e6ed50">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7926a5b8-716b-470c-8a4f-d1daeb231667">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa9381fa-f8ab-4ca8-83fa-92099a46e583">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_52bf9ca6-3be6-43eb-837d-6aa06a012855">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b610faaf-e635-4c46-a6e8-02db62d4e196">Concept Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b19d153-7aae-4241-8192-699336fde559">(0040,a168)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59ad0620-dc35-4e8f-99e4-5304c215a714">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65ed0b56-6118-4586-97d1-50ea45cbd114">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b7fbfe3-739c-4e46-98af-6a750498da8b">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_14af4701-a781-49d1-aa0a-12dadcdf9ce4">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9928550-7971-4a4a-a836-53c852fa6377">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd350f92-0862-4db7-afaf-8509f4a0114a">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c2e7b73-bb4e-4faa-8fc6-12be255a3906">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d6d1900-b77b-45bc-9791-542d2e4c9ed0">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3e53d34c-f21a-4dbe-b048-95d880ab23d2">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5daa544-2bb4-4c72-8bc7-23d4700fcb35">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c979a82-7787-4d28-a06a-a3d002628df5">en-US</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ab3e362-2f2b-4f5b-ae71-272131824922">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e468b44-f9b0-4468-8466-8dfc5798a300">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f18fef6f-6e57-41f7-999b-fb8b7c960119">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5db5d7e2-0b71-4b6f-9e77-8312a0031372">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_015995d0-d2b6-44e5-80ff-6b1d42cfde33">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec49c5df-d34b-463d-ad6c-4b1110cccd1c">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1d5975e9-d8de-4846-8d7f-d708d30f70fa">ISO639_1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a717db3f-384d-4b9c-8f5e-2696ba790036">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ca8107c-a905-47e0-8411-658ef504555a">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05f713b4-8f49-491f-aa97-270e91d8cb6b">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9cf9694-cf87-4184-ae0c-dd28c7a8e9c1">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5b9fc82b-bc32-4cde-ab1f-ef55f686078a">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8eb62bdd-c7b1-4b5d-93c8-6c0f9e9e753c">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef7508be-9ba6-43a6-a032-31e11341a476">English (U.S.)</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cfb13b57-91aa-49fe-8dd8-35879ab0507e">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9bdb7f81-8d6b-4c7c-bd9d-7a56f95ef5cc">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_94904dcb-37b4-487c-950b-72095d7e3708">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_412e01b4-7016-4c60-a861-4702f69adfec">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_04fadee3-9b55-4c79-bd07-835537995b8d">1.3</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_493cf101-c7f4-42b2-85ae-1b6f03e7ef40">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c47df6d-a21a-42e0-9c85-e757a2d45eac">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cfdb63d7-2075-4580-9909-c8bd3bd44e64">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2308f41-c4e6-47bf-88de-91a48f5df65f">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c57dfe2-a300-4282-998c-856ccc956aef">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03d28d45-c729-4968-ac5f-fcf5b7002ea2">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40214c3c-dffb-40b5-8432-cbf5c8daac38">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_496ba532-dd26-47bd-9ee3-75c61eaef418">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_79f47175-92bd-414e-a831-d1bacc577c5c">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d028cc04-1122-420c-85ed-4272a6240694">HAS CONCEPT MOD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_63303cf6-7c55-4ce9-934e-7a37bf06edfb">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eaff5288-e43f-46a0-9275-f2f5cc2f8724">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_887c6da8-5f26-45ef-86b7-9ce0b077f9b2">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec54e0c0-72d3-4363-8381-463528b752b9">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09c88f8c-9951-4099-aec5-86c328b8acc1">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3540a87-7985-4b6c-b202-7e0db7f68e09">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38147212-42cb-4a57-8545-c38d8a5591c5">TEXT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_844a4bc7-838c-4aa1-a584-159cd64e3423">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c2d4d1ed-43a3-4179-9d7d-cd8eaa40a140">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5441aea4-892f-4809-a505-e93a357ef593">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9e744471-04e7-40f7-a06b-a4bc677d18e5">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_224dde4c-1d12-4395-a011-c1503d882b98">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bf45471c-f300-4572-9799-61f9257f3fe3">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1538d5a-908c-4df8-8e06-c1aa48e50834">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d46850b4-a322-4a96-91dd-31c2e8c1c195">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b41be51-70a3-4cf3-82cb-fb32a4a77326">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b736d18e-ddd7-4579-abe8-8bfb994830b4">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6c6a7205-959e-4a69-9b79-326387d80142">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b46f30b0-7eaa-42f4-a7c9-a07937a48552">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b94a6a8-193c-4729-9d77-be499383d16f">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_018f5e21-1e41-4633-a672-c5b15827d20b">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_540fcc00-c5b1-4a77-bcd6-7493f0a6da6f">121050</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f142a4c1-657c-42f7-94dc-f006fbf83e3b">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71444106-b099-4d34-be22-60afcb05b386">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71631746-6056-403c-b019-18620223e4b6">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_02066950-b310-4249-a1cf-dca1d922ac00">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8fc373e-797a-4cf2-8d15-cebb25f57a67">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f36321ba-0df6-4eb6-b29f-afa50a0016a3">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3789f2b-499e-46e1-8fbe-f8f50e029135">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84cfe92f-fae5-477e-8606-7c09cbde5bf3">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_847ad73f-4f45-4df8-ad83-c53643916088">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2c78259-f999-4f09-8116-7eacde860528">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_41653148-07bf-4601-85cf-2db37064f4e5">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f38d7243-9260-4b35-8a23-f71fe457a621">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_456ddf50-df36-4fce-8d2e-744e0fc1276c">0022</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c5aa967-43fc-4166-9004-3de2fab9a174">Equivalent Meaning of Concept Name</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e107c1c-9fc7-40ce-9aa6-6558193f5794">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f786d32f-d952-49b6-98d4-290df9fd7829">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8de4381b-b747-4b24-873f-dc4a16f6ced3">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5302a750-fcaa-4f56-8ed5-58ab4469600a">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fdf371f3-dcc7-41b7-914b-ccab2f5ef30e">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87f684b1-f52c-4769-b110-3e171a1f62d8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67c9feaf-019c-4bb6-8116-8ab401ce8d4c">Text Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f6492a62-3a74-49c4-84c0-c91cb2a02fd8">(0040,a160)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a08d78a-39b3-4947-bbed-aef216a643f0">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fd8e8d61-aa91-49d9-a681-e21aa685c41f">001c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fe92a08c-6bab-4063-90f5-ea70e068eb80">Chest X-Ray, PA and LAT View</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82f1b71d-5240-4a11-9570-843c80c4c504">1.4</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_463a9517-ef0c-4d33-97d6-fc6c19fbc51c">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea7f3a7d-e0e9-41e3-9523-4b04f7578b0e">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1262712c-7a38-44d2-b94c-90075352e481">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d27a04b-94ff-4dc5-a94c-5b7f659e6b33">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_26382863-c203-49fc-94b5-6833e7a89308">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7575206e-917f-4ac8-aa72-b09ed0e3c2db">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44ca36a8-b876-4c7d-adc2-3c63fec85e50">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6d6f7369-1f72-4207-b194-d7f3492e68ac">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a5f1b8a-3851-44f4-ada0-6967fc9e24b2">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f31bb106-b5e8-4fde-b2cc-b7135e632a35">HAS OBS CONTEXT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_246cdd6d-958a-4476-bf15-905452b6db04">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6238789a-b0ae-4b9c-b0d9-a3c7a32e5e41">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7334cc5e-ac14-4357-ab1e-7dfd9ed6ac68">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_01a3cd24-42f9-4f13-9c0e-eb0cd64834b6">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2110902-98b8-4e94-b7b0-adac732bd6cf">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_796cf850-6573-4877-9bc6-1b67a6fd259c">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_86e908b9-8705-4557-983c-7abf0884ee24">CODE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fc434288-b6e7-4e48-90b7-fd66c08cd319">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7484930-69a4-4c24-b5b8-2b7b89885ec3">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8817c6ab-0abf-409c-8d51-ab2d94c5b17a">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09c4d0e0-6549-4d95-81de-abe16576fdd0">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a76413d-c1c1-4027-affb-7c153bab6b01">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cadad4bc-9260-4357-aee0-87db4a26d102">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a79143e-3f1e-425f-96a8-620393c4253e">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1452dc7c-7bfc-4f8a-8176-4519c4c3b096">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b087aaf-d4c7-42c7-9895-cd7bf1e02195">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9c67f1bc-4e89-4b24-ba75-c2267470de81">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89a956a6-c025-4809-bb14-7f72bed3dee5">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3e60499-ce98-4ea7-bf95-ae4d9f329324">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b14bdb15-4b6b-4b61-bd95-b66418ca3504">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bd7c1410-7cc8-4c1e-b166-c7e960e172c1">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_26c3ebe1-02fc-46c8-b2f5-5c30f4e692fe">121005</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_df5a3578-4790-4dfa-be7f-5b7a36d3a6af">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ddfdcc56-6035-45b0-8559-2cf32643295b">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12be4684-b0d4-4d58-a431-436134143a5d">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4019f9bb-e1e2-4209-bc58-7bfbb272a27a">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d368528-f6cf-45d3-8149-0f06b90e5faa">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_75a68ade-b5c6-4cac-a95b-194ffc6b3d2d">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3ceca56-b5d6-4aea-b0f3-736b139d4e3c">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_99227e70-ac35-4904-b390-cfa58da08fab">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0288544c-466f-437f-8c1e-bc17cefe6798">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcddaaa2-fd07-417f-8f60-c3daad220b12">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fafb02ac-f095-4042-a770-2c0ba4672968">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_129a9555-b2a7-418f-af68-ddd5cf5024e7">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f2851c9e-7c97-4549-93ce-54236af1858d">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c0aa8c0-b88d-43f1-b021-2fc87dd08e8e">Observer Type</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dc2b64b3-7c10-4527-9464-a7221e5a69e6">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7fba79f8-93a4-4426-a151-088093999e3e">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_768acbd1-3c8a-4bc7-bf89-6596982d0d13">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3540843f-1f3e-4586-818c-c9a9b13b2534">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69efeab2-12b8-4188-9c84-326f2deaa866">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_241f904a-82cb-43e5-9834-3098345e75ae">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0d9920ec-42d4-46ac-8eef-039a55ab51d3">Concept Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a9febb5-e166-480d-9613-315027070805">(0040,a168)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc32fad8-c572-482d-b530-25bc160b5101">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5ed7b8c-b056-4fd2-9412-e764aed187bd">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dcd21feb-1eed-4e25-86fb-b0bcb511cd91">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9e66015-0ccb-4ff9-8c5e-3129c0e0e0ff">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa08de7d-8654-4ffa-a3ba-fd9077c1a356">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2cfbe81f-d537-41da-b725-2fda6b6ac19e">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9552bbf1-064c-471c-b40a-5f3318aa6870">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed241acf-198c-4071-9aef-9da8408cc58c">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_49ce79d1-35dc-43d5-b05b-00e2fc395758">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12ccfa0f-1d3e-4aa4-b718-57d003830c78">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2b1f8cc6-3f98-4440-a27f-bcc72e6bcdde">121006</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_07bf44dc-50d9-424d-8305-7726d5e6b9e6">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a0366dae-ad2e-450b-92d5-a79b09912a81">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90bfd4a7-c811-4959-8f98-df8356f853db">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92ab2774-e6e7-4e94-8d11-d8ef11324ea1">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e02590a-56df-4aae-976e-1be0977354da">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2f826e8c-92ae-4018-a430-23d4c21b3338">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f7ef4cd-d092-430b-8b7b-d7edfdc1d53b">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9bcb4783-8491-42da-aa7e-f72dca70989d">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_feea4c21-0699-4050-9b3a-769b2083fe10">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c215f039-f716-4070-8077-9e2613ab5cdf">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f08c8b5b-1c35-4b20-8224-e57d55931963">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2655f8a2-f8a4-4d77-8d6e-b9f98c7bc03b">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15bd0237-aa02-4f3a-a50b-179fd209d955">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_544b3ebb-788f-49c7-afc2-c3a8a7c80484">Person</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_98aee245-1654-4116-a0fa-cd2b36dd4ef9">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_915bb5cb-6418-4405-95ca-ba6c7d36b95e">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fcc60352-f8d5-4167-874e-ad1e0aaf21a0">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73ded6f5-0043-45e8-803b-7d391333fbbe">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_550489c8-ff4f-40f5-94d6-d3336a67636a">1.5</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_827967bc-eb45-4edc-82ed-7e842afbd7df">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5ad8b7c7-e9e8-43ff-b8b8-27e5ff1ae23e">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db3c0e91-22ba-4f95-af99-d60184cddaa0">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_79c7570c-8c7d-4cb4-8332-aac795dc14d5">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9c75b75-b616-4cc8-9ef8-9dd1a86b238f">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7ffcbd4-b8f6-493b-92d3-5b8f56f14738">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_63209955-926b-4d42-9f50-03368314c6b5">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2beff7e5-438d-4d52-8d11-8e5ee2467e0e">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_494fb69c-443f-406f-b215-e18ec5fb8223">0010</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47df233c-e222-4a31-8473-bb99aec99ccf">HAS OBS CONTEXT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8fcb2cde-0ea4-4d41-9c31-2fb3e5008d7e">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cea89c21-8f47-455b-9d76-2ab69305ec93">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_04c56337-d06d-4d9a-98dc-0bf2afff5231">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_87800e01-1fc5-4ea1-ae1c-109c0e7d4bc8">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3fb74a4b-2c02-4ecd-9a53-2f861b6c10c7">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_86f23860-5519-4f99-b4b6-be1aba8d7168">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fcfb135f-e21e-4d6f-9a01-201db3a07c5e">PNAME</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_420aac5c-36e0-4982-a9a9-660eb333fe65">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70926444-6523-44e8-af3d-080f34b8631d">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78c27df8-7949-4fbd-868a-074bc36173a5">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8e901907-4335-44a3-a7a7-34ad742c80ef">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90aaba14-919c-4039-a4a1-8fd12d90cbd9">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f77a25fe-c999-4b50-b57a-12cd5df331d2">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c7788c6f-5b5b-4135-bebc-37c34639d3f4">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bca4ded2-b29e-47d0-90d6-b4864b7464d9">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d810e2fc-fc23-4351-a0d0-9fc363038447">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_71902437-85a6-4167-b700-8265c7b7fc7d">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb5eedde-edfa-4b63-ae56-2690b138c8ac">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6df137f3-d9de-42f0-ae93-709d865476af">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_761d24b7-6693-414d-913d-90603fa20fa0">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_712e2b16-464e-439c-a92b-20e5acb414e1">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd5def68-0ac9-4656-9791-a991638e6d26">121008</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3547b0fa-dded-40b7-a545-ab6158f2c4f1">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_368adfb3-75f9-4797-a48c-3907eb2ed90e">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d4bd421b-2ac0-409a-a5bf-a7f27ff92b46">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc81358f-c013-4231-9057-e7b8452702d9">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef103bb1-448b-4a35-b398-b5d9db1774bb">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d80e4922-df73-46ad-85de-4d564fbf7a54">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_22edc884-060c-48f2-8fa6-0c191e5be9e4">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a85e400-767e-49e4-bd45-8293f4a5c365">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5da3069-cff2-4fe7-a94d-ade0663b8344">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_60880a1c-77d4-42c1-802a-25bea8708767">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a323dba-7393-4ce6-9ed9-2a0a1be7341b">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d9e268d7-f740-49d5-9e2c-95b565cb47b4">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e80d299-b28f-46be-85aa-59bdad151b18">0014</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba52a400-06de-4815-bd02-49e8a7c32c6c">Person Observer Name</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_10445c5f-e021-4470-a2de-751e970ad313">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d61f027-24ac-41a4-a24b-528deaf5cc2d">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66228dd5-34a8-47c0-bab7-0fbc15dd38b1">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17febe89-0035-4c0a-aef2-afb0c69694fb">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bbff99da-bd40-4f6c-bd42-ea6c1deafc6b">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9a7bb23-ccfc-41b7-b0c5-2d30ee3943a6">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_deddd807-e72d-4830-bec2-b267a70091d1">Person Name</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40d0f176-df41-453e-b1ec-4b09c07aff6c">(0040,a123)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3cb5f0c1-f1fb-4291-a897-00fb25abb194">PN</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7ed3ce4c-7648-4b45-846b-18db5d8d300f">0012</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_702f4d23-430c-43a2-ac9e-68d24ba6df7a">Blitz^Richard^^^MD</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b5529c5-51cf-42f1-b0f6-43274aa885df">1.6</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a65a683-b394-4738-8cae-15ae31286180">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_112ca469-26c5-41d4-906e-c5531c4fdfb2">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7130a327-3fb2-464c-a65e-85c674e8f862">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c8c9c70-89e6-4181-92a2-54791cf0c028">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b92fbbe-eb44-4f5f-85d5-766a5d1568c8">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_44f61616-9908-47c0-9114-29b1f027e280">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_849693c9-c476-4962-99b6-2b45201c64d5">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1cd231f4-3194-4712-89ad-212a2c9710e6">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1dc7146b-1324-4493-b12b-a235394c0781">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_538463f3-33dc-433d-b718-0c756c704bcb">CONTAINS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c77f5303-ca31-4af4-b5ce-b51c8d868374">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96ffc66e-036a-4625-9da5-68282f7281c5">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dfc4801e-6727-42e6-860d-ca042f8c9d35">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad0418c3-ef8d-4666-9ece-153e38d01560">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca3a2d82-9559-4adb-8536-170895ddfac6">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c1f6bff8-7aa0-4203-83fa-b53cb07bb480">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f0a81d98-ab72-4e36-a6d5-86d1c5b1ea57">CONTAINER</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c9522b25-5295-4bfe-bb5e-eb814818b5d1">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_636dd53c-a135-4866-95e7-0a1fdde36773">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b5661029-2518-4339-9d4a-3d9ff63c0206">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f08e0c05-724f-4dd4-8947-f6eba43e7def">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1afc330-bea8-45e3-85e5-6ad74237abd7">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f75210bc-a424-4a49-b4b9-ba4f1bc93601">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d3603e54-0744-42ef-8c43-125dfabe5659">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a394556-0bc1-47e2-9a15-ea9535af63f7">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_442fbada-6e4c-4f92-970d-f8e3f2503038">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_983f3e7e-b549-486f-8079-87d79759f418">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3913bee8-c2bb-48d8-97da-3aac633fa535">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea5b7077-249f-4a3f-b7fa-ace3e6b3df22">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5bfd6d4-5cd4-4453-9bff-dd97f2ad56bb">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d6077e3c-0453-4d0a-ba8b-2b442632bc06">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f246d45d-e8a7-4c96-9b0c-5de5c66e11d0">121060</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_316cba80-1e5c-43fe-b33b-b6906addb6ea">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a04abc9-9caa-4a5c-ab5e-990b8fdcacc1">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5983be7-81a9-446b-bcd1-8c3e508612e4">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_433eb05b-b8b9-47f7-8f51-ee404cc8553e">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e0e54509-917b-42f4-bdcc-95204ae684e5">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea7aa4e4-9867-4a2b-a85a-7a816c1cf697">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2871779d-8478-4b87-a94c-5da5cca510e6">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_18ee0fcd-4273-42eb-99ab-f5428eafcf6a">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5da2d96a-22f8-4303-b092-a261d74d12d7">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f5fa86be-69ff-403e-aa69-40d432a34e00">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c4020da9-4958-4658-b76d-77e24494b6f0">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_18aa373d-42e0-4b2a-8623-b07797c24a1a">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69176e9b-fd22-4b8e-a29c-d5a41524748c">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e2c89b54-7a1d-4701-89d5-647058737ad0">History</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_60ec4530-4777-4c45-b256-df24ef815900">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_df4881cc-4627-42b5-83d6-a31c27a662f0">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_be5a541e-7d22-475f-9b87-4e14384dbeb3">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1890be3-5f41-447f-97a3-1f8078ae0078">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_854482d0-a502-457b-b5c7-82e5f90d3e37">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_798a2720-ab0a-48ad-82ac-d2330c3a8e74">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6332656c-02bc-4eaf-8157-d4eb545fa22b">Continuity Of Content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c3f808e-5a6a-4452-a2cb-33cfde4d5043">(0040,a050)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a392210c-a7b4-4a60-8776-d30ceee373a6">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c5e2783-7eee-41ca-bcfe-d62ea6520b2d">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_072db11c-9c1a-41d4-aed9-25021f1f4a54">SEPARATE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8cf4beaf-4fc4-4bc3-b0d7-be07083a20fa">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_77b6411d-898e-4eca-9d11-fec088875797">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d042bc8c-e1a2-4c7d-a3f5-2cd0460dbe6f">Content Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3225eccc-fde0-42a3-8051-c94b7834a896">(0040,a730)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a0c4f55d-44db-42e8-bc41-53627503274f">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8c62362-4ade-4be6-bf11-b8e7db15a418">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac861aa8-553b-474e-93fd-1131953010e0">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_18391df9-a0e7-407d-a22c-8e89bb4e3c2c">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fe1e7c5b-1f60-4aed-849d-07ad0be1f9d2">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_239ba4fd-b21e-40ee-88c5-b456745879b8">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba1aca96-d38f-4476-bab2-d618890f1b67">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_346b9631-89a7-494c-bd3e-915d99af2209">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_272ef2f8-364b-449e-a751-fcac0f1b477d">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cb5e3d02-24ae-4e31-849d-e502eec2f5c1">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3dbe3a1a-e428-4378-a565-3cdbd71425c7">CONTAINS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7b22d37-df99-4c60-8659-e1a8c0db05c8">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec6084e0-3ccb-44f5-bcd7-9a1104a262b9">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42285daf-532c-44f6-9479-6391f8f19a04">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e40dda6c-b0f4-4a7c-818f-d487e3a815f0">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_865709b2-112b-4090-be27-d338f336e8dd">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d1f925d-b45e-4f9a-bbe7-dfbdfd63a7fb">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea70272e-7120-481b-881e-417e6ce368e3">TEXT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7826f1c-881c-4dee-8974-4e9a4fd1aafd">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_22b4ee9d-d085-4ab5-acbb-b349256d1aa9">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_359487de-cae3-4bff-985d-dd32c757b2f2">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f897a656-a33f-495d-9acd-a38fdc26faa8">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d394bfb-07e2-4546-98cf-40a545842dac">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b26b4a02-f8e7-42f0-a5c1-a26cedae4455">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a17e684-3b1a-4370-8331-52a6ac68e6a7">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89fb46ed-dbca-492d-962e-d64ab8691254">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f44c0c1-a686-4eae-b518-f48469efc071">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b81e4f6-b877-42ee-b042-3de140a2f9e6">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c5164c14-d4d7-42d6-8e4f-7f8180226744">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f6b4c015-32b1-4688-a66a-ccfaae14fe4d">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1880503-1c47-40e7-80d9-f0c7beb23f74">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2cd00ed5-639d-4120-9ea6-732e7db8f24b">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66eca9b6-8c07-4d72-b8a2-d9f277685fa0">121060</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73c031d0-e962-4c2c-8830-f06656f3ed36">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f22f51b0-feca-4c83-8ede-70201478366a">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4c160e05-02fd-48ab-9cd0-1b07603402d0">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_94dbf2b1-d41d-40ed-b1be-15b27e043e73">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0ceb7b6b-0cc3-4582-ac57-26398c8c80f3">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b51a6e1a-93e9-41bc-9d89-bf86a1b1e80a">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_207ea69d-0e6b-4281-8e2e-cad364ea573b">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_13ebffa2-ecb2-47bf-ba8c-c4ecb0f31769">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_83ca1fcd-7152-4d51-92ae-60065a2f3c2c">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c87e9a9-930c-419f-9af4-51374bb547cd">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b83e99b-c557-4127-8301-3ae67f83779e">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9ef6fe0-9292-4416-befe-44914f8d743c">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f41ae18c-8c75-4aee-bb27-0bd5f754beb4">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7d02049-dad9-42a0-aade-1c2a8f1c3f88">History</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8a9c9a7-7ec0-407f-b109-78365ff153e6">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_40ae13ea-9f7e-412b-83c0-32dbacf095e6">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2144646b-3570-4045-ac95-bfdaef44b1f9">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b7816fb0-c94b-4675-b7e9-22cca6fb9968">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa3e34d3-55df-4861-8fcd-e60ee1d23379">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bbff213c-673d-4e58-b0fe-87c17cfbf0f6">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_48049b47-969d-4379-b558-91f6a2d287dc">Text Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca174d70-dc0d-46de-9919-22c0c700c1eb">(0040,a160)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_128f86ac-598e-49aa-9559-020effde1f37">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b184856c-45dc-4a52-8aee-8dcd50838a82">000c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0eeead35-e6e5-40b9-acdc-334bdea40511">Sore throat.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a6c3bde-a20e-454e-9b8e-567b7179d87e">1.7.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8c9f71cb-35be-49a8-8d2b-4c236428eba0">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bcfb4531-d45e-4e77-b768-d446d3158fdf">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_930c3d75-bb8e-4a12-8ce5-3b12d446a174">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5cd6f10e-0de8-46ac-8658-2c44dea029c4">1.7</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa907945-545d-42a1-a87b-3eb4c2a25b7d">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a53446e5-f7f1-4361-98d1-f1afb4d05eb5">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f3b7547-e2c0-4c4d-8539-023936ca3f8a">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d6752f6-6279-45ee-b7b7-f9d4eafbf310">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6303a35a-55b5-4f7d-a6a0-43400a94dbec">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12792e5c-fbf0-44ba-bc48-ba5e29df6317">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d9facdf9-4e04-4194-a3e7-1046406ca815">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_74b69f34-64ac-40e8-aae7-0ff9fa987a6b">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a32bc651-478b-4722-ac35-e7c9a843bdfe">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c72ca08-07a8-43a3-817d-a9d8b5a8590c">CONTAINS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_73188cea-286c-47d8-9383-3c9492fbdc2a">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9354349f-8670-448a-850f-b57fd39863ae">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3ad7a177-a5a0-4c69-aaf4-7c0d447a93fe">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fca10f23-6cc6-4aca-9a52-7029ef24076c">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42bd01fb-b6a8-4612-a08f-cd23e2621263">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a94456c5-6bcb-4d62-a80a-c6e1998e2604">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_20053bc9-677e-49f4-b4dc-bce6cca9da14">CONTAINER</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70630944-7bf7-47d9-a649-88cec27228c5">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f6dafb2-f28f-44d4-8a36-feae5e416985">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7a93edb0-7d1a-40b4-a534-df23380fb973">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_820eb2c6-afca-4a63-8c85-b302719c222a">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e74468af-33ad-4c06-b57e-371abbc58d05">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d050507b-c6c6-402b-8d4d-183ca1794284">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a952906-fa3b-471f-afc5-7c8dedf0eeb5">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_10cfbe8b-ba5e-4d5c-a335-999d5dfbcd10">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3866843-629f-4b2b-b0f9-1d555169e94a">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50dba201-f202-4ed3-8ec2-81da180cf271">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_240e7e1c-583f-40f9-886e-dafc2b3f0dfc">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9b4cf6ee-0ec7-421c-815a-50a482f8560b">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2ae67ecd-b39a-4eff-a215-4ac6a54e0983">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67967f70-7b9c-4e44-9263-982db85ac269">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_36aff7f1-05c6-47d1-bcda-b3a1394b85d9">121070</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac24a85c-fbc1-4118-9cea-1e220bfb4c7e">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_726ef4f2-47fb-4419-9d19-71e1ff21a61c">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b269e416-72f2-4b29-bbe6-f94d44630dc9">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a284beb3-66f3-463b-a106-2476a7b48bde">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b38469f2-525a-4654-9531-1906cf1e1ffc">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_99d8c7be-c88b-409d-9e97-bf3a171bc95e">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_21588f66-5507-476e-b7b5-3bcc6a033e39">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_498e0b2c-fee8-4a11-8d19-71bb148d3d8c">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff84f554-c98f-4d9b-bd47-2407ec1b2dcc">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f069947-a9bf-49cf-8d74-26381e634626">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6e024ce-d08e-4c13-88bc-0ff25d635482">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_93307210-087b-4724-9ded-4277fc3fd830">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b41cdc5-718e-42f3-b877-249710b515c2">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1c9a0402-dc27-4a29-9636-4c99ab85401e">Findings</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84a011ad-9731-4d5e-87ba-f441cd0a94ac">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_289ad7a0-593e-4158-8afa-d12e82915d19">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b957b209-21e5-4537-8c32-c28ea4d40b5c">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_506c19c8-d8c6-43aa-873d-bbb3078ee3c8">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ab7cbe2b-2d09-4201-9917-ed9adb06bc3f">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3c40b37e-bd6e-4276-ace0-dde2e070f239">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96b78fc9-8970-4433-bce8-9a724bda411f">Continuity Of Content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5ceac20-3835-49ae-b260-852288f2b38e">(0040,a050)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a42b118a-0f1a-4370-9005-c8392f914533">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3bb5e867-defd-4c75-98b5-62ad4337100d">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7d3793a-2bf9-4fed-b359-578a6e77cbfe">SEPARATE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2bb3520f-a746-41b8-b8f1-4c4bc8a29b89">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_301175ac-6001-4010-9a46-1e93d7c9c5fe">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e5cc5ebf-61b6-49a6-aef2-2a6bf514d725">Content Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a49f061d-d265-4830-8b5b-584de7909d40">(0040,a730)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_67630813-9403-4e9b-ad6e-96c2dec04828">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3e7b5846-83e9-476c-ba68-9c6642fa1df7">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0522c3f5-a0ae-404d-af65-5abeca1ad26a">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_39aae791-7a96-4d61-974d-807b380c66be">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_21d2eae6-a3a9-4f11-a433-33653ed2217f">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cfb9310b-2566-46c9-9803-dc17f22e23be">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_119b44ad-5108-4f9c-be4c-b9725376daba">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_522cdab6-ee1b-4454-8705-67a89ec2e4e7">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f522910-fafc-40fa-a713-6ea467a3d919">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1ee2d80c-93ae-49e2-b186-5d2561eaec3d">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_763d184f-a820-45d7-9550-5e2829617e1e">CONTAINS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_124f0b3b-a35b-4447-9df3-b1d906d44c37">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50d76246-f3a7-458b-a794-c6af0ddc35fa">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d6c262f7-3bf3-444a-855a-52a5b6c2e466">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0bbb9be5-49a3-49a1-bd01-3a95bcd3726e">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0cf95e6e-1b03-4526-b4b2-f6d927b55b4b">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7b248688-2ee6-411e-bbbe-84f25798ebeb">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_294e4403-07e3-4bcc-8f13-faab6097125d">TEXT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f94d405d-45e7-4295-bca4-72e77a0cb45e">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e3ef6705-7d8c-4f65-bb02-c6763c90e202">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b87067b-79cc-4ac3-b413-b418d86d9999">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_137eb190-1086-45ff-9b26-120a88c4c6dd">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a8759774-7ac7-40f1-ad4c-30a40716badb">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_605c4c42-22cd-4edd-b4e4-dc8f67227deb">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6326872-fb34-4240-93ec-7a358688220a">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_45cfa8df-9973-4607-a205-81ea2fb38fb2">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ec4146cc-6e2b-4b33-912d-cd86235b4508">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d28f57fd-e448-4aa4-9c42-a42f2ade41ae">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8b3c35dc-99ed-4277-85e3-9d00d5c652b5">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b51cc31-8191-443f-9a0c-22bcefe355ed">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e09ceae9-95fc-472b-baf0-632a1a879080">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fedca509-7952-4984-aa09-cd937cf7a966">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_31954586-c84b-41d4-b4f4-8257d96bbd5f">121071</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_51d68698-824a-4f62-98fe-477b10a438aa">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d34b34c-29d9-4db3-bb2d-db9d677efabf">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f8596c65-65e2-4aed-969c-71b2b0d40105">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d805de99-3fe4-4ec9-9ad0-2c57bd339b4d">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0135d5a3-ab95-4c2b-8e3f-d1f4b0e49d06">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7d1c88d-e74a-4cdd-aca7-2fb10bdd05b3">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3e16917-a1e0-4150-bd66-53bc2a02efdc">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_044a481c-3b1e-421b-9fcd-e529d9542c18">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fb4db870-c7c0-4015-ab95-284f0d3f1aa8">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a291b1ed-b184-4c5c-a651-90497c61df79">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3fd902fa-8427-4b07-8e18-3ad7460b0fe2">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e05b3e47-6bda-40f1-a754-0a408618f975">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_01260453-87d2-4ee2-8bd3-a695823d55c5">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b5858a44-2573-4d73-a130-bec9f404f3c5">Finding</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5fa67ead-9b65-4fa0-9202-d9d31581e510">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f55dd18-09b3-4211-94b1-951d1ce8c43f">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7af457a6-217a-4123-a550-cdf2155363c7">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e50a7476-9559-4579-af1c-32ad38320c37">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6281cb3b-55b4-4109-b41c-e61225701710">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f7444b5c-0057-4bb7-ba6f-be2863490d23">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a98c48eb-692e-4d78-9058-f74e9572806f">Text Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f69388b-f3d6-47aa-870a-2f9f6cd52bfb">(0040,a160)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78533ebb-271a-4c3e-8eba-e0825d7233a2">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_05e36ec4-55b3-43a4-a0e8-1b04ddd4b662">01ae</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5a122a8f-5f85-4df9-a6d2-4b676741180a">The cardiomediastinum is within normal limits. The trachea is midline. The previously described opacity at the medial right lung base has cleared. There are no new infiltrates. There is a new round density at the left hilus, superiorly (diameter about 45mm). A CT scan is recommended for further evaluation. The pleural spaces are clear. The visualized musculoskeletal structures and the upper abdomen are stable and unremarkable.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_612cefff-b5d9-47b0-a49d-2fe1bff2c2d0">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a9540888-7e0d-486c-b305-046a98323835">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1407dd9b-42d3-41e7-8784-955e4512b7c9">Content Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f0c91ffe-3f09-4c30-9bd6-38aa99b98eb9">(0040,a730)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a94d648-e339-4030-8472-419c3971fd30">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c2a0f2c-4927-4b70-911f-c438c0148b59">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d3de636-e88b-47fa-8463-6d90fac7ed32">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_edfddd79-4aca-4d50-9f74-def52388f18d">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_14e786a8-45c6-46fa-ba4d-da964cfdf8ab">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2d8ede5e-e5dd-4c34-9829-8d6763fb86ef">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e1248d4-f30a-41e2-a323-2d824519138b">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a38ffb2-09c6-4ad6-bdf8-ae245e72e266">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_46c6a0ae-9d55-494c-a66d-3ddc73f92797">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac025497-aeef-4c47-b56c-eb7c09d9739a">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a46bca65-0220-4539-9869-e11d205ed336">INFERRED FROM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bdbdf652-e1bc-4823-b29c-c1eb37714fc5">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_09e2224e-fba1-4198-a911-8f1e211d4ed9">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f4d54756-0f67-42c8-b389-8bb6523316c3">Observation DateTime</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed05e37d-b8d1-4163-ad50-88f56b671857">(0040,a032)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a457d62a-d56d-412d-a6f3-33de339dbae6">DT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0b6b96d5-5e2e-4262-9937-2bd98c20ecf9">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ebfa3c58-12bc-475e-b302-39a787fa9a4b">20060823223912</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a00b83c-ee76-4001-9cf0-552144962508">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0068a60b-cc74-4a50-9d13-cd4e67cdcac5">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e6c69426-4713-4efb-9a98-4eecf7c538c8">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_35aa2c26-98b7-4893-955f-1c2c165262b1">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b51b6de4-8823-4060-a9ff-97eb91b4edc0">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fac4d8d3-2e47-47e3-9976-ed67a7162c94">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_14effa18-9ec1-4527-8402-0a3d3abfd8fa">NUM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61e5ecfe-547c-47e2-8c29-01f13d1d62d4">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_897c3810-ebea-4a9b-954c-aec29ecc1dd6">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7d3b8d10-b7cd-4328-ad6a-5d20c12bc67f">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_daba70e2-ffcf-435f-b5ad-01f73a354eb0">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_13b92840-f961-4539-a6da-b6f4f6b43e26">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ba737329-3fab-4996-b08d-5c7c8c3a527c">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17175e87-c45a-4d4f-9616-8c6750fd7e74">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b3123fbc-16ad-46c5-94bc-91dd200726c9">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1fb5b3f5-d98c-4aa3-a7f0-303519effd30">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca6e0b33-df62-4808-a571-5ae44c28a439">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4424b8c0-dd50-4982-8847-3586acff46c2">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_143623a7-6421-40d3-9a99-08614a6ea646">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4e5e6a8-916e-480d-ab23-bed0a9ef0184">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c9a5394e-d7d6-45c1-affc-4bf380a3f312">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                  <para xml:id="para_1bcffd15-ab38-4031-ade1-26af8cddb1ff">81827009</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5931619d-303a-49b5-b9c5-0fafeabb5a3b">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_08e2f40d-e829-44c5-a60f-7f9733e932d2">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fcbfc16a-7e7e-4fa1-ba7d-1e6da6a0a1a0">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_21106179-ac0f-4def-90ec-2b0570856209">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1ef55a6-20fe-475c-a20c-430960ff07dd">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a7f04953-9fc9-4455-b2e2-c14c8e23a772">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b585be17-7ab7-410a-9779-70ab8efd139d">SCT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0664d278-acd5-41eb-9994-12a8cca821ac">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_575b0740-200e-48cf-82d1-6d172b951913">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_84049e58-fa89-41ef-aea3-35038722a0a4">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5be45dc9-a6b5-469a-95bd-4c248eb2acf6">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bae039d3-a4e5-435a-8f92-4e8490753204">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_887fd7e3-a1d4-48ea-b69e-8ee0fe6942b6">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a48a5ce7-478c-463d-9c2e-01a18cfb393e">Diameter</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a6f0013-2dd2-4365-bb05-b8faeb50aa4a">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42b3265d-254a-4237-bb59-b7fcf3a6538b">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b85ccf3b-7a42-4c6e-b381-3a035a206e83">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ee2955d2-09fe-430f-b7d0-c48063a5e7e6">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6a492df-d7b7-43a1-9912-fcb768c9b4de">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69dc87b7-1ebe-4475-9400-5b2ec5d72737">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_398b5b7f-d315-45ec-a6cf-14fba56ea417">Measured Value Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac785b7a-476c-41fa-b899-207b3309821f">(0040,a300)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_248d0219-b1e8-4a9b-8223-cebc91ba3eda">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_78113983-4c42-47ac-a7f3-87e8250fb221">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a4ba5810-4547-4385-a19d-1e606b0805ee">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_777cf009-6b66-4ff5-b114-090dd68cb1f6">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d4f36494-1294-49c2-949e-07a9b72dc3f5">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2a364890-d33e-450f-a388-6e3249835a47">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b6308bc1-b703-464b-a61a-7d3198af6b15">Measurement Units Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d4bd6dbc-3d08-467f-97d9-d12384ac2888">(0040,08ea)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9629f61d-7153-4d24-9f76-39f075115301">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_62b7d8a7-5493-49a5-8f77-76d1985118f9">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8af0c7a8-e5db-4a65-a334-d96c21d2576c">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af6f260c-b200-456f-953a-19b56610f618">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3589cdf-0ea8-4bf4-9ec9-5396d1d62912">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f3e8c338-3620-4dd9-b894-b284000cde45">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d61f9f1-822e-481f-a5e1-c1ccd5861718">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd6d70b2-ad93-4003-b48d-8e72e82c185c">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1d9ca8b-a42c-439c-ad03-375a606ec423">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9b914f64-fc40-47a4-b220-0f12c8d8ddf9">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7abdd8d2-edd5-47c0-a794-56ee54353f18">mm</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8ef2c386-15b3-41d8-9ff3-aaafe6fd5b11">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_397cf80c-208d-4a55-b747-3bd452a8a3e8">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_998210c1-5a17-4d9b-b148-a7239bc162f8">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4fc98c44-d9b0-4139-84de-d9c52252e002">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b366be06-0e98-40ad-8b25-86685ad69d75">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5d2944b8-8b8a-4121-838c-c2ba22778424">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4b2a1a67-d7c2-440b-b88a-730681e10a27">UCUM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_50e6ec0d-033d-43cf-b6c0-48b931f63b50">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3155234-c3c4-4811-9bd1-d755d7a72539">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6e3c4567-5467-48a8-be17-cf84b4cbb16b">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15154c0b-a8ff-4643-bced-0caa8fa16044">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1f00ce68-9208-4067-a025-e440ae4ccb93">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1bc3b106-f0fe-4ba1-8294-bbeee2cc1f97">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1a96e641-16d9-41f6-9653-bab08d6e1321">mm</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8509fb2d-4874-4e31-ab8a-317c27e2908e">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f08f8c93-1a00-487e-9d06-abcf3c8e125f">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c6069427-9743-4689-99a2-0cab9c3d28b8">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d9b7dce3-4626-4430-910a-f902b5080120">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fea12482-b694-4093-af36-4165d4bd9dfa">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_29035854-8c8d-40a6-9558-4f30cf00b411">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_116a80c0-a011-4b91-8faf-34357724a469">Numeric Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa914865-500c-467b-81ad-49421ad949aa">(0040,a30a)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f9cef8e2-8626-40a8-95db-414dcebb88bb">DS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c500bb52-2713-4778-beca-77d230523cb2">0002</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f33f4e2d-96e5-4ab0-b500-3a38f91e220b">45</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_427d5368-292f-4630-b3e7-8fea43c4064a">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d1661ada-a4cd-4dbd-83cc-3d74eb905534">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dad52943-b391-4d4b-ba17-57116d7ce49b">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8689a11e-dcaf-4416-9780-31821acfbb9c">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_889ded1a-26aa-416d-a31e-8c12f46c2747">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b5900242-7d4f-48c3-a625-50801688cf9a">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c17649e6-20c8-470b-8093-a4a57fd8b6ee">Content Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0c767f55-b6c5-47da-a871-ccf9b97c495c">(0040,a730)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2bcfaac1-04fd-429a-83c7-dce9c66b71b3">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9e97d3b1-7307-46e8-8b1b-b198a043a736">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b58e4759-8550-40f9-9444-125c77307956">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3bed6a7a-89b8-4f15-bdc1-f41ec6f433d8">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7c721fbd-6ae1-4911-b375-612e1db65fbf">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92168e28-979d-4292-92d0-f7b98d865955">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cc569459-5920-4db9-9ea1-6f6af8f67c75">Referenced SOP Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dd54c5fb-cee3-4db1-8a64-8fc4fa9eac43">(0008,1199)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c02f682f-fe7e-41fd-aa3d-13698b4665bd">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c073257e-a617-4f38-9569-39f7f7b05a62">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3f569e16-0e92-4ee2-b672-cdc14c48d01d">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3cf38e4c-cfc3-424b-bf0c-c0fb5722e0d8">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_416eea4b-a737-4012-ad75-2f00e6d55ba0">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6067a692-3c4b-4a79-94c6-6c05a0b221f5">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9c6eadce-08a1-47b9-9b27-1d055c888293">Referenced SOP Class UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_65be5e13-2b98-4c7f-b652-2c1157693c6b">(0008,1150)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8bfcb33a-5a63-4c89-adfb-d8a6206aba81">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2aa3227c-459a-4952-bf8d-f2f77dcf4ff8">001a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ca5bc90c-f6a9-4a5f-bfa2-6c22ddde2ebe">1.2.840.10008.5.1.4.1.1.1</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d8b4378-331b-45df-b05b-6bbe9e74f9a1">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0dbe1601-961e-4a14-926a-ff1c588f15f5">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_07ea6d38-a96e-483d-97bf-4d2db6afa341">Referenced SOP Instance UID</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_229c2186-b616-46e3-a02c-50a5097415f9">(0008,1155)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_47356fff-e1f7-41f6-9683-ca08c410df6f">UI</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9bcfad75-b273-4afa-aee0-81a38dd77343">003c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e9572e97-785b-4d2d-9a19-ef6d80dc3664">1.2.840.113619.2.62.994044785528.20060823.200608232232322.3</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c29f0969-11ba-4111-a88d-761f4caa9fdf">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_59220108-4c06-4e23-8639-2aa9f885153b">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5df7e9af-960a-4138-a399-1a5007712e05">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b0a4a459-7089-4934-b36b-a827a1c701eb">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c8e12f43-e6cb-4234-a01a-fcbeb11ddf91">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1feaa67b-7466-4de2-ac1c-6bcfdf45d78e">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f7a87332-ea9b-4ae1-aba9-51810af7c88f">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7e2b439e-a739-42f5-933c-c8be1dac3fd8">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f6df915e-b97c-4e2c-ac94-db0d7476c37c">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96bd4ae1-9f9c-4f86-90f2-aa9dedd95813">000e</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f1d4c9e9-a5a8-402b-a640-fe8e34ba6dbd">INFERRED FROM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2103b61e-457c-4088-8fef-66858ba2f571">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_783b3e62-a7a3-475a-aefc-8143baabbeb6">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b9ca1d61-f58d-4170-91be-db74e6a5ad8a">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_90f58ff6-127d-492d-be83-26ea65a8ef52">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_16a25d9c-1822-4887-a06d-dd5f7acf147e">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2c312ce-d975-4581-bf1d-c5cc07138832">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e45e41bf-f69d-4862-a91a-6133e957d004">IMAGE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3d804809-079e-4798-ab95-e6ef2513b344">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_148d79b5-4b74-4a0a-af75-49668df83db7">&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dae62133-2d61-4cf2-93ec-3e0c3986c600">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7514bcc0-27c5-4f0f-ab22-1b805338c699">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03722c96-ee84-4bc1-b6a8-28148155a230">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_46debfc1-3505-470c-acc3-6c233297cd36">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_eb3f8743-945f-4077-9583-9d7544091570">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3124f49-60cc-4539-8d60-13c1c677d526">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e1b93cb8-a986-4b1d-aeb3-93bd8fc390a5">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_82215e7a-3e8a-46b6-be68-8e3aed642914">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2346cd44-d972-4bf1-9577-7c0a5d1cef44">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d53e62c6-bd33-43ce-97ff-e1ac7e36f985">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7dae54ce-acaa-4949-9422-c1e586530a11">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_03c4cc0d-d2b0-4b21-9416-5dc532db6708">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_42335a9d-8928-4b23-81d4-4f65314d13f1">121112</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89955c09-7c45-4b85-8d94-e63a35bd8b59">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc14305d-fe47-4b7b-b716-7f432d3be86c">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_af71e483-02ae-45ab-976b-3f953da72633">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_95c3c7d1-d6dc-4aaf-9563-f3aeecb4ca2f">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_101dc977-cb92-4895-9065-bc43d7967daa">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70f251f3-7ad1-48a3-99a9-e5ce1ee32404">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3bbda43-7510-4a2a-8588-a81836929830">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_66597668-a072-46a6-a164-8759199c01bf">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b4fafd0-eb11-4ff9-9da5-afb7dc8ee542">&gt;&gt;&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dde2273f-959b-494e-bf0a-7de925f696fc">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0fbe6bc4-2504-48e0-aa0b-47a84558c0d4">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d12fa65f-822c-475e-ad59-0c936d68f969">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6db55742-eaa7-4cfd-9709-8c2cba8f356c">0016</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ea3fc4d1-effa-48ee-beee-09d4186c137d">Source of Measurement</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_420b6890-4032-4883-928e-c0532c2232a0">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e7659e60-aa83-47d5-aaf9-9060c557c445">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1becc7d4-ae4e-405b-9c5a-33420a4dcbd5">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e18ed8d7-9d3e-4656-805b-ce38e92568f0">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_294de3ee-bdd3-4cff-920c-7c3f282be619">1.8.1.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0cbdffc9-cdf2-4d2c-aa54-865a4ff96f03">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0a6629c2-e501-4cc7-82ef-3fbe93740e88">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7f67cba7-d39e-4af3-a6ab-d1cc37ed4e32">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_34bd73eb-4797-45f7-9b11-0d5929aea7bb">1.8.1.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_55b724a2-3c02-468d-8fa1-de3cee572321">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_616542a4-338b-49e0-8775-7052e09bfed8">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_19007ee8-f493-4a57-8be9-3c44869e4aec">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_13c9dddc-73cd-4ccf-a1d2-437a5142677d">1.8.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ef74707a-d91a-4373-a0c1-2be61ae3b4c8">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f36b2084-17f2-4548-8123-5effefb831c0">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e8ecffe0-9330-4712-9cc6-3bdd76effb20">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_01de3f4a-7a24-4a08-9d0d-25794e38a2c6">1.8</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57861aaf-b2ab-4316-9ffc-7a9fdeb9d5ff">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_fa41d770-4f95-4c05-a0bf-4aad3ac30655">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_376c6a5d-6bcb-44a7-b7b9-5e723aac7c59">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_34248ee0-c00c-4b9f-b6ea-8902eb394fea">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4de135e7-f602-4de4-b376-995e95ddb9b0">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3e46bdcf-82dd-4411-bf18-5fdaa60b866c">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1b951752-820d-4687-8911-cf2cd0299668">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_53f36981-25f3-4431-b3fe-499378619398">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_aa3b2305-e3d7-4919-812c-cbc04fe1c3d8">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2c495f47-9ecc-4c9e-b542-4f5ead8dc201">CONTAINS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_60e59a8e-2ffe-428e-a594-e7889725ef86">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7dd72a5c-4844-4a22-af2e-cd6f7227288b">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad61e305-9646-4371-aab1-b4df958201ad">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9f213cd7-5178-463d-9d1b-6215944ecc8e">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4eca78d1-c3ec-40b5-9a3f-7f9d33626aec">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_10d250db-eaab-4201-846f-faaf0f0a31ac">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ebd372d3-e81a-494e-b957-6142d1c50376">CONTAINER</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cd08bce8-3ad3-4d0b-9ad8-0e21433856fe">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_da2cbde6-1a72-4660-83c0-45ab1bd42f6e">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a73efb43-58cf-459f-9004-471245a63ddd">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_baa5826a-9618-4884-9bb5-7142ac80bd21">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64467cc6-4670-4074-a8b3-c58ba1524eed">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c3994e0c-77ca-458d-9105-be3b47575ea0">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_88787210-3a1a-497e-b9f6-c3c526399326">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_23dfa57f-fede-46df-8cf5-3e6571a18850">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6b741fa8-1190-4232-bf43-a20b64f81c40">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_543c9b5b-739d-407a-935e-3a8e32663687">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db06ccdb-8a74-46f8-a50e-bef1e367c690">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_12373025-a0a2-4f46-90f7-1d54f4de866b">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2006eab6-0f9c-4479-81ea-5c812fe158ad">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6de859c4-0a89-4dac-92a6-9227181b5835">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8eb5697a-9152-450d-b4a1-bb1a782af590">121072</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c7652ca4-f5db-4243-9eee-dd4eb208996a">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_849c7c9e-3382-4be7-bd8f-b0f987cf3348">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1be04f9c-7e60-48dc-865b-d1647136cb6e">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0f3a3928-eae7-4fe8-a7bb-2b298262e636">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_db871171-355d-4e0c-befd-0bb87ae51eaf">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_53e5d0e2-af55-4f2d-bf27-af24c1fba94e">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1aa138df-556a-4ca5-b4d1-b639efe11a2a">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8d8fc2e3-b8a2-441b-9e17-57e3255cc698">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69eb8670-5257-49ac-8495-1a0b28977795">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69f4b2b6-e5d0-4dc9-a837-ff489d1a4cfc">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b81f09b9-8a26-44c2-8140-e80c812a7775">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9df22e4d-32fc-44b8-a0b5-e8a622fdabd3">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8a5fbe1e-6035-4b3f-9f59-8b30f00a94eb">000c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_792cfc58-ddcf-4585-ac7f-e503cd95d80f">Impressions</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_96c47dc9-68af-4e09-a8a9-57e7bca172ec">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_f3dfaec2-ff73-4573-9784-f7809f4cbae5">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a332d8d7-e130-48d1-a6f9-99e106dcd2df">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_098fae27-31bb-48cd-bc56-6adae0cd3cdc">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_30e7e995-750c-451f-9d84-01c501798f41">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_64ce686d-6b8a-44bc-bfc9-cd5a636365d2">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d37359a0-4b1f-4a8b-af27-e7f8bfc34d65">Continuity Of Content</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_79e9db9f-dbe1-4423-80e7-e6d7eb44ca14">(0040,a050)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_89445a8c-5dd0-4f04-955a-f195c89981c0">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_931adeea-a77c-4aa5-be5f-0234859ff3c4">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6a4ec159-99a3-467b-b67d-0895ce317144">SEPARATE</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1316249c-da0a-4a39-9786-38d30525a1ed">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_006f744f-9055-4854-ad58-cd4073b4229a">&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a932668c-4463-4657-8cb6-d332b5be4685">Content Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_962c6ff2-9732-41ad-b225-4e6af99f4724">(0040,a730)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_632ef4e0-b768-4a89-aa8e-b5b15eaeb430">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0e0f0a71-fd0b-4508-aefd-837af237946b">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c7727761-b25f-4b6c-8fe0-a007d45bb775">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_590ed21f-0762-4b49-8b19-86f4cc434dee">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc6ad513-84a3-4373-ad60-dade902d3b9d">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3b61ace8-1472-461b-b1a2-9d2ad5c2c6cc">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e731e5ed-eb8c-4460-a806-ea0d603a44c0">Relationship Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6cff5f96-7be4-44c9-959b-8a3fb2fc0b78">(0040,a010)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_dff62fdc-4b4f-40dc-a062-997d7f768a32">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_38ce9edc-f050-4ee0-9ee4-0c0efdc6bc83">0008</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1fe7418a-11c5-46a0-a54b-c0a0ece4528f">CONTAINS</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4e8a6ddd-92e1-4b04-a9ae-85b8de0ea426">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ff94bf91-b5d0-4b71-94b7-dfbd1cd70cf0">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57274222-e4ad-4f19-9c5a-f21443645e19">Value Type</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_636eaf7a-9f72-4f51-a768-b4d2dfd8a715">(0040,a040)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_3a38c863-153e-4bd2-ac4e-655d9437be7e">CS</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_836113e8-7cde-4e75-8ff2-72f3d4794b1c">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a3504034-8270-4d46-8999-9f34a2bfa7f0">TEXT</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_57bb4ab1-c16d-446a-ac4d-ebf098528b8e">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_61f9bd1b-4d41-44bf-9942-297f8195e17a">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_29329c72-1870-459c-ab98-96c5f87132d1">Concept Name Code Sequence</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_15cc8e5a-086f-45fe-804c-59ef29077d30">(0040,a043)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f59e9c3-2836-404f-b7ae-e71e7eedee7a">SQ</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6f246d3a-76a8-437b-b9c4-1a4d686f0b0c">ffffffff</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0db44276-7701-4bcc-a893-6996c20f5bdb">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1834937d-0b05-4fb5-beef-e732e84343c7">%item</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_17399ac3-7025-4cd3-bdbf-fab6537f56bb">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_cfce15bb-c4eb-4430-9cf2-42b328a18a0a">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_70e8bfda-c485-4702-b0a9-c22ced3e43a4">Code Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_0bed0efb-4044-4436-a8d3-770dd488bd41">(0008,0100)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_4a8f10a3-d2ed-439c-9f64-dc431a58a50f">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ad544d95-5645-4bf8-818e-392c7c12e7be">0006</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ed354294-7b6c-4e7e-9679-957fc02f6e22">121073</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_517c850c-102f-4f9b-9111-12fa689e865e">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9a67a84d-1e4a-4d73-ada7-03519938ee1d">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2667ee3f-fd8b-402a-9d99-07e9e0a565ad">Coding Scheme Designator</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_6cded657-cd96-42a5-b8c3-344b95c87198">(0008,0102)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e64e7f9b-bcad-4561-87a9-e3a7a1a0e0de">SH</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_9d76df9f-c1f7-48f2-a6c3-f80672df77be">0004</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_e678b7ca-63a9-4202-94a1-e6cc507cbcab">DCM</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_ac5215ea-4496-46c4-bfb6-76a60b6e62ab">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b1808c9f-27f6-487a-9648-c30dc5d1c8ef">&gt;&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d64fbbdb-d5b2-451e-bb7b-764fee8999b0">Code Meaning</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_d220e8dd-20d2-45d2-bdcb-aeb9238ebb43">(0008,0104)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_69109aa9-4a4b-42da-8aa6-ac6d2206545f">LO</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_900424a6-ae50-420a-9f3d-0434200d8539">000a</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_b2b6da37-2af0-4adf-9f27-861a8f10fc6a">Impression</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_073c94e1-736e-4d17-8ee9-2984c078e676">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_7fca133f-b028-4f13-aae6-b7f911425518">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5f08fdf5-26e4-434d-82ac-d0f98cd2ca81">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_00926c0a-0856-440b-870f-72e99470bf3f">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c289a66f-9eea-4f7c-9c71-dd8680e4c6ac">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bdac742a-fc95-4b34-964e-afd1fcdd129e">&gt;&gt;</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_58954b22-fd75-4896-a93d-62f2b92a469d">Text Value</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_2e1a6eaf-f6ee-45d1-875f-2bd3e35f0eeb">(0040,a160)</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_5e9872ed-04ab-4dd5-a1ee-c4cc40e7ad18">UT</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_294051a3-d631-4a60-a89d-8fbba79e729e">009c</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_23794d56-5dac-4e82-93cb-54f2a6ae0a9c">No acute cardiopulmonary process. Round density in left superior hilus, further evaluation with CT is recommended as underlying malignancy is not excluded.</para>
              </td>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_92a39d4d-9218-43bf-ac6c-db53d773fc78">1.9.1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_8680a43d-5aeb-46e7-94c0-9ad7b71a3a25">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1a87f663-e5b3-49e8-9a46-cd34325b89a7">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_c505f7b4-2a95-44ae-9702-73af6c1f427b">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_00841910-bca0-43e4-92f2-4dfb83c0b65a">1.9</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_bc8d555c-2646-46d8-bb92-1408aa2d8059">%enditem</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
            <tr valign="top">
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_a1e37bc4-11c0-49aa-83ac-e3815fbe2008">1</para>
              </td>
              <td align="left" colspan="1" rowspan="1">
                <para xml:id="para_1206297e-d7e7-4588-836c-cd836cae54b0">%endseq</para>
              </td>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
              <td align="left" colspan="1" rowspan="1"/>
            </tr>
          </tbody>
        </table>
      </section>
      <section label="C.5.2" status="3" xml:id="sect_C.5.2">
        <title>Transcoded HL7 CDA Release 2 Imaging Report</title>
        <programlisting><![CDATA[<?xml version="1.0" encoding="utf-8"?>
<?xml-stylesheet type="text/xsl" href="CDA-DIR.xsl"?>
<ClinicalDocument xmlns="urn:hl7-org:v3"
        xmlns:voc="urn:hl7-org:v3/voc"
        xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance"
        xmlns:ps3-20="urn:dicom-org:ps3-20"
        xsi:schemaLocation="urn:hl7-org:v3 CDA.xsd">
    <realmCode code="UV"/>
    <typeId root="2.16.840.1.113883.1.3" extension="POCD_HD000040"/>
    <templateId root="1.2.840.10008.9.1"/>
    <templateId root="1.2.840.10008.9.20"/>
    <templateId root="1.2.840.10008.9.21"/>
    <templateId root="1.2.840.10008.9.22"/>
    <id root="1.2.840.113619.2.62.994044785528.12"extension="20060828170821659"/>
    <code code="18748-4" codeSystem="2.16.840.1.113883.6.1"
        codeSystemName="LOINC" displayName="Diagnostic Imaging Report"/>
    <title>Chest X-Ray, PA and LAT View</title>
    <effectiveTime value="20060828170821"/>
    <confidentialityCode code="N" codeSystem="2.16.840.1.113883.5.25"/>
    <languageCode code="en-US"/>
    <recordTarget>
        <patientRole>
            <id root="1.2.840.113619.2.62.994044785528.10" extension="0000680029"/>
            <addr nullFlavor="NI"/>
            <telecom nullFlavor="NI"/>
            <patient>
                <name>
                    <given>John</given>
                    <family>Doe</family>
                </name>
                <administrativeGenderCode codeSystem="2.16.840.1.113883.5.1"code="M"/>
                <birthTime value="19641128"/>
            </patient>
        </patientRole>
    </recordTarget>
    <author>
        <time value="20060823224352"/>
        <assignedAuthor>
            <id extension="121008" root="2.16.840.1.113883.19.5"/>
            <addr nullFlavor="NI"/>
            <telecom nullFlavor="NI"/>
            <assignedPerson>
                <name>
                    <given>Richard</given>
                    <family>Blitz</family>
                    <suffix>MD</suffix>
                </name>
            </assignedPerson>
        </assignedAuthor>
    </author>
    <custodian>
        <!-- custodian values have been added based on organizational policy (int his
         case they are not mapped from the SR sample document) -->
        <assignedCustodian>
            <representedCustodianOrganization>
                <id root="2.16.840.1.113883.19.5"/>
                <name>World University Hospital</name>
                <telecom nullFlavor="NI"/>
                <addr nullFlavor="NI"/>
            </representedCustodianOrganization>
        </assignedCustodian>
    </custodian>
    <!-- legal authenticator present in sample, document is VERIFIED -->
    <legalAuthenticator>
        <time value="20060827141500"/>
        <!-- Verification DateTime (0040,A030) -->
        <signatureCode code="S"/>
        <assignedEntity>
            <id extension="08150000" root="1.2.840.113619.2.62.994044785528.33"/>
            <addr nullFlavor="NI"/>
            <telecom nullFlavor="NI"/>
            <assignedPerson>
                <name>
                    <given>Richard</given>
                    <family>Blitz</family>
                    <suffix>MD</suffix>
                </name>
            </assignedPerson>
        </assignedEntity>
    </legalAuthenticator>
    <!-- Mapped from Referring Physician's Name (0008,0090) SR sample document -->
    <participant typeCode="REF">
        <associatedEntity classCode="PROV">
            <id nullFlavor="NI"/>
            <addr nullFlavor="NI"/>
            <telecom nullFlavor="NI"/>
            <associatedPerson>
                <name>
                    <given>John</given>
                    <family>Smith</family>
                    <suffix>MD</suffix>
                </name>
            </associatedPerson>
        </associatedEntity>
    </participant>
    <inFulfillmentOf>
        <order>
            <id extension="123451" root="1.2.840.113619.2.62.994044785528.29"/>
            <ps3-20:accessionNumber extension="10523475"root="1.2.840.113619.2.62.994044785528.27"/>
        </order>
    </inFulfillmentOf>
    <documentationOf>
        <serviceEvent classCode="ACT">
            <id root="1.2.840.113619.2.62.994044785528.114289542805"/>
            <!-- Study Instance UID -->
            <code code="11123" codeSystem="1.2.840.113619.2.62.5661"
                codeSystemName="99WUHID" displayName="X-Ray Study"/>
            <translation code="XR" displayName="XR"
                codeSystem="1.2.840.10008.2.16.4" codeSystemName="DCM"/>
            <translation code="51185008" displayName="Chest"
                codeSystem="2.16.840.1.113883.6.96" codeSystemName="SNOMED CT"/>
            <!-- anatomy code mapped from old style SNOMED in SR to new -->
        </code>
    </code>
    <effectiveTime>
        <low value="20060823222400"/>
    </effectiveTime>
</serviceEvent>
</documentationOf>
<!-- transformation of a DICOM SR -->
<relatedDocument typeCode="XFRM">
    <parentDocument>
        <id root="1.2.840.113619.2.62.994044785528.20060823.200608232232322.9"/>
        <!-- SOP Instance UID (0008,0018) of SR sample document-->
    </parentDocument>
</relatedDocument>
<component>
    <structuredBody>
        <component>
            <!--**************** Clinical Information Section *****************-->
            <section>
                <templateId root="1.2.840.10008.9.2"/>
                <code code="55752-0" codeSystem="2.16.840.1.113883.6.1"codeSystemName="LOINC" displayName="Clinical Information"/>
                <title>Clinical Information</title>
                <component>
                    <!--**************** Procedure Indications Subsection *****************
                        Section text mapped from "Reason for the Requested Procedure" (0040,1002)
                        within the Referenced Request Sequence (0040,A370) of the SR header, under
                        the assumption that the header attribute value has been displayed to, and
                        accepted by, the legal authenticator.-->
                    <section>
                        <templateId root="2.16.840.1.113883.10.20.22.2.29"/>
                        <id root="1.2.840.10213.2.62.044785528.1142895426"/>
                        <code code="59768-2" codeSystem="2.16.840.1.113883.6.1"
                        codeSystemName="LOINC" displayName="Procedure Indications"/>
                        <title>Indications for Procedure</title>
                        <text>Suspected lung tumor</text>
                    </section>
                    <!--**************** End of Procedure Indications Subsection *****************-->
                </component>
                <component>
                    <!--**************** History Subsection *****************-->
                    <section>
                        <templateId root="2.16.840.1.113883.10.20.22.2.39"/>
                        <id root="1.2.840.10213.2.62.7044785528.114289875"/>
                        <code code="11329-0" codeSystem="2.16.840.1.113883.6.1"
                        codeSystemName="LOINC" displayName="History General"/>
                        <title>History</title>
                        <text>
                            <paragraph>
                                <caption>History</caption>
                                <content ID="Fndng1">Sore throat.</content>
                            </paragraph>
                        </text>
                        <entry>
                            <!-- History report element (TEXT) -->
                            <observation classCode="OBS" moodCode="EVN">
                                <templateId root="2.16.840.1.113883.10.20.6.2.12"/>
                                <code code="121060" codeSystem="1.2.840.10008.2.16.4"
                                    codeSystemName="DCM" displayName="History"/>
                                <value xsi:type="ED">
                                    <reference value="#Fndng1"/>
                                </value>
                            </observation>
                        </entry>
                    </section>
                    <!--**************** End of History Subsection *****************-->
                </component>
                <!--**************** End of Clinical Information Section *****************-->
            </component>
            <component>
                <!--**************** Imaging Procedure Description Section *****************-->
                <section classCode="DOCSECT" moodCode="EVN">
                    <templateId root="1.2.840.10008.9.3"/>
                    <id root="1.2.840.10213.2.62.9940434234785528.11428954534542805"/>
                    <code code="55111-9" codeSystem="2.16.840.1.113883.6.1"
                    codeSystemName="LOINC" displayName="Current Imaging Procedure Description"/>
                    <title>Imaging Procedure Description</title>
                    <text> </text>
                    <entry>
                        <procedure moodCode="EVN" classCode="PROC">
                            <templateId root="1.2.840.10008.9.14"/>
                            <id root="1.2.840.6544.33.9100653988998717.997527582345600170"/>
                            <code code="11123" displayName="X-Ray Study"
                            codeSystem="1.2.840.113619.2.62.5661" codeSystemName="99WUHID"/>
                            <effectiveTime value="20060823222400"/>
                            <methodCode code="XR" displayName="XR"
                            codeSystem="1.2.840.10008.2.16.4" codeSystemName="DCM"/>
                            <targetSiteCode code="51185008" displayName="Chest"
                            codeSystem="2.16.840.1.113883.6.96" codeSystemName="SNOMED CT"/>
                        </procedure>
                    </entry>
                    <component>
                        <!--**************** DICOM Object Catalog Sub-section *****************-->
                        <section classCode="DOCSECT" moodCode="EVN">
                            <templateId root="2.16.840.1.113883.10.20.6.1.1"/>
                            <code code="121181" codeSystem="1.2.840.10008.2.16.4"
                                codeSystemName="DCM" displayName="DICOM Object Catalog"/>
                            <entry>
                                <!--**************** Study *****************-->
                                <act classCode="ACT" moodCode="EVN">
                                    <templateId root="2.16.840.1.113883.10.20.6.2.6"/>
                                    <id root="1.2.840.113619.2.62.994044785528.114289542805"/>
                                    <code code="113014" codeSystem="1.2.840.10008.2.16.4"
                                        codeSystemName="DCM" displayName="Study"/>
                                    <!--**************** Series (Parent SR Document) *****************-->
                                    <entryRelationship typeCode="COMP">
                                        <act classCode="ACT" moodCode="EVN">
                                            <id root="1.2.840.113619.2.62.994044785528.20060823222132232023"/>
                                            <code code="113015" codeSystem="1.2.840.10008.2.16.4"
                                                    codeSystemName="DCM" displayName="Series">
                                                <qualifier>
                                                    <name code="121139" codeSystem="1.2.840.10008.2.16.4"
                                                        codeSystemName="DCM" displayName="Modality">
                                                    </name>
                                                    <value code="CR" codeSystem="1.2.840.10008.2.16.4"
                                                        codeSystemName="DCM" displayName="SR Document">
                                                    </value>
                                                </qualifier>
                                            </code>
                                            <!--**************** SOP Instance UID *****************-->
                                            <!-- Reference to SR Document -->
                                            <entryRelationship typeCode="COMP">
                                                <observation classCode="DGIMG" moodCode="EVN">
                                                    <templateId root="2.16.840.1.113883.10.20.6.2.8"/>
                                                    <id
                                                    root="1.2.840.113619.2.62.994044785528.20060823.200608242334312.3"/>
                                                    <code code="1.2.840.10008.5.1.4.1.1.88.22"codeSystem="1.2.840.10008.2.6.1"
                                                        codeSystemName="DCMUID"displayName="Enhanced SR">
                                                    </code>
                                                    <text mediaType="application/dicom">
                                                        <reference value="http://www.example.org/wado?requestType=WADO
                                                            &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
                                                            &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823222132232023
                                                            &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232232322.9
                                                            &amp;contentType=application/dicom"/>
                                                        <!--reference to image 1 (PA) -->
                                                    </text>
                                                    <effectiveTime value="20060823223232"/>
                                                </observation>
                                            </entryRelationship>
                                        </act>
                                    </entryRelationship>
                                    <!--**************** Series (CR Images) *****************-->
                                    <entryRelationship typeCode="COMP">
                                        <act classCode="ACT"
                                            moodCode="EVN">
                                            <id root="1.2.840.113619.2.62.994044785528.20060823223142485051"/>
                                            <code code="113015" codeSystem="1.2.840.10008.2.16.4"
                                                    codeSystemName="DCM" displayName="Series">
                                                <qualifier>
                                                    <name code="121139" codeSystem="1.2.840.10008.2.16.4"
                                                        codeSystemName="DCM" displayName="Modality">
                                                    </name>
                                                    <value code="CR" codeSystem="1.2.840.10008.2.16.4"
                                                        codeSystemName="DCM" displayName="Computed Radiography">
                                                    </value>
                                                </qualifier>
                                            </code>
                                            <!--**************** SOP Instance UID *****************-->
                                            <!-- 2 References (chest PA and LAT) -->
                                            <entryRelationship typeCode="COMP">
                                                <observation classCode="DGIMG" moodCode="EVN">
                                                    <templateId root="2.16.840.1.113883.10.20.6.2.8"/>
                                                    <id root="1.2.840.113619.2.62.994044785528.20060823.200608232232322.3"/>
                                                    <code code="1.2.840.10008.5.1.4.1.1.1"codeSystem="1.2.840.10008.2.6.1"
                                                        codeSystemName="DCMUID"displayName="Computed Radiography Image Storage">
                                                    </code>
                                                    <text mediaType="application/dicom">
                                                        <reference value="http://www.example.org/wado?requestType=WADO
                                                            &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
                                                            &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823223142485051
                                                            &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232232322.3
                                                            &amp;contentType=application/dicom"/>
                                                        <!--reference to image 1 (PA) -->
                                                    </text>
                                                    <effectiveTime value="20060823223232"/>
                                                </observation>
                                            </entryRelationship>
                                            <entryRelationship typeCode="COMP">
                                                <observation classCode="DGIMG" moodCode="EVN">
                                                    <templateId root="2.16.840.1.113883.10.20.6.2.8"/>
                                                    <id root="1.2.840.113619.2.62.994044785528.20060823.200608232231422.3"/>
                                                    <code code="1.2.840.10008.5.1.4.1.1.1"codeSystem="1.2.840.10008.2.6.1"
                                                        codeSystemName="DCMUID"displayName="Computed Radiography Image Storage">
                                                    </code>
                                                    <text
                                                        mediaType="application/dicom">
                                                        <reference value="http://www.example.org/wado?requestType=WADO
                                                        &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
                                                        &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823223142485051
                                                        &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232231422.3
                                                        &amp;contentType=application/dicom"/>
                                                        <!--reference to image 2 (LAT) -->
                                                    </text>
                                                    <effectiveTime value="20060823223142"/>
                                                </observation>
                                            </entryRelationship>
                                        </act>
                                    </entryRelationship>
                                </act>
                            </entry>
                        </section>
                        <!--**************** End of DICOM Object Catalog Subsection *****************-->
                    </component>
                </section>
                <!--**************** End of Imaging Procedure Description Section *****************-->
            </component>
            <component>
                <!--**************** Findings Section  *****************-->
                <section>
                    <templateId root="2.16.840.1.113883.10.20.6.1.2"/>
                    <id root="1.2.840.10213.2.62.9940434234785528.114289545000804445"/>
                    <code code="59776-5" codeSystem="2.16.840.1.113883.6.1"codeSystemName="LOINC" displayName="Findings"/>
                    <title>Findings</title>
                    <text>
                        <paragraph>
                            <caption>Finding</caption>
                            <content ID="Fndng2">The cardiomediastinum is within normal limits. The trachea is midline.
                            The previously described opacity at the medial right lung base has cleared. There are no new
                            infiltrates. There is a new round density at the left hilus,superiorly (diameter about 45mm).
                            A CT scan is recommended for further evaluation. The pleural spaces are clear. The visualized
                            musculoskeletal structures and the upper abdomen are stable and unremarkable.</content>
                        </paragraph>
                        <paragraph>
                            <caption>Diameter</caption>
                            <content ID="Diam2">45mm</content>
                        </paragraph>
                        <paragraph>
                            <caption>Source of Measurement</caption>
                            <content ID="SrceOfMeas2">
                                <linkHtml
                                    href="http://www.example.org/wado?requestType=WADO
                                        &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
                                        &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823223142485051
                                        &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232232322.3
                                        &amp;contentType=application/dicom">Chest_PA</linkHtml>
                            </content>
                        </paragraph>
                    </text>
                    <entry>
                        <observation classCode="OBS" moodCode="EVN">
                            <!-- Text Observation -->
                            <templateId root="2.16.840.1.113883.10.20.6.2.12"/>
                            <code code="121071" codeSystem="1.2.840.10008.2.16.4"codeSystemName="DCM" displayName="Finding"/>
                            <value xsi:type="ED">
                                <reference value="#Fndng2"/>
                            </value>
                            <!-- inferred from measurement -->
                            <entryRelationship typeCode="SPRT">
                                <observation classCode="OBS" moodCode="EVN">
                                    <templateId root="2.16.840.1.113883.10.20.6.2.14"/>
                                    <code code="246120007" codeSystem="2.16.840.1.113883.6.96"codeSystemName="SNOMED"
                                        displayName="Nodule size">
                                        <originalText>
                                            <reference value="#Diam2"/>
                                        </originalText>
                                    </code>
                                    <!-- no DICOM attribute <statusCode code="completed"/> -->
                                    <effectiveTime value="20060823223912"/>
                                    <value xsi:type="PQ" value="45" unit="mm"/>
                                    <!-- inferred from image -->
                                    <entryRelationship typeCode="SUBJ">
                                        <observation classCode="DGIMG" moodCode="EVN">
                                            <templateId root="2.16.840.1.113883.10.20.6.2.8"/>
                                            <!-- (0008,1155) Referenced SOP Instance UID-->
                                            <id root="1.2.840.113619.2.62.994044785528.20060823.200608232232322.3"/>
                                            <!-- (0008,1150) Referenced SOP Class UID -->
                                            <code code="1.2.840.10008.5.1.4.1.1.1"codeSystem="1.2.840.10008.2.6.1"
                                                codeSystemName="DCMUID"displayName="Computed Radiography Image Storage">
                                            </code>
                                            <text mediaType="application/dicom">
                                                <!--reference to CR DICOM image (PA view) -->
                                                <reference
                                                value="http://www.example.org/wado?requestType=WADO
                                                    &amp;studyUID=1.2.840.113619.2.62.994044785528.114289542805
                                                    &amp;seriesUID=1.2.840.113619.2.62.994044785528.20060823223142485051
                                                    &amp;objectUID=1.2.840.113619.2.62.994044785528.20060823.200608232232322.3
                                                    &amp;contentType=application/dicom"/>
                                            </text>
                                            <effectiveTime value="20060823223232"/>
                                            <!-- Purpose of Reference -->
                                            <entryRelationship typeCode="RSON">
                                                <observation classCode="OBS" moodCode="EVN">
                                                    <templateId root="2.16.840.1.113883.10.20.6.2.9"/>
                                                    <code code="ASSERTION"codeSystem="2.16.840.1.113883.5.4"/>
                                                    <value xsi:type="CD" code="121112"codeSystem="1.2.840.10008.2.16.4"
                                                            codeSystemName="DCM"displayName="Source of Measurement">
                                                        <originalText>
                                                            <reference value="#SrceOfMeas2"
                                                            />
                                                        </originalText>
                                                    </value>
                                                </observation>
                                            </entryRelationship>
                                        </observation>
                                    </entryRelationship>
                                </observation>
                            </entryRelationship>
                        </observation>
                    </entry>
                </section>
                <!--**************** End of Findings Section *****************-->
            </component>
            <component>
                <!--**************** Impressions Section *****************-->
                <section>
                    <templateId root="1.2.840.10008.9.5"/>
                    <id root="1.2.840.10213.2.62.9940434234785528.114289545345927752"/>
                    <code code="19005-8" codeSystem="2.16.840.1.113883.6.1"codeSystemName="LOINC" displayName="Impressions"/>
                    <title>Impressions</title>
                    <text>
                        <paragraph>
                            <caption>Impression</caption>
                            <content ID="Fndng3">No acute cardiopulmonary process. Round density in left superior hilus, further
                            evaluation with CT is recommended as underlying malignancy is not excluded.</content>
                        </paragraph>
                    </text>
                    <entry>
                        <!-- Impression report element (TEXT) -->
                        <observation classCode="OBS" moodCode="EVN">
                            <!-- Text Observation -->
                            <templateId root="2.16.840.1.113883.10.20.6.2.12"/>
                            <code code="121073" codeSystem="1.2.840.10008.2.16.4"codeSystemName="DCM" displayName="Impression"/>
                            <value xsi:type="ED">
                                <reference value="#Fndng3"/>
                            </value>
                        </observation>
                    </entry>
                </section>
                <!--**************** End of Impressions
                 Section *****************-->
            </component>
        </structuredBody>
    </component>
</ClinicalDocument>]]></programlisting>
        
      </section>
    </section>
  </chapter>
</book>