PS3.6

DICOM PS3.6 2026c - Data Dictionary

DICOM Standards Committee

A DICOM® publication


Table of Contents

Notice and Disclaimer
Foreword
1. Scope and Field of Application
2. Normative References
3. Definitions
4. Symbols and Abbreviations
5. Conventions
6. Registry of DICOM Data Elements
7. Registry of DICOM File Meta Elements
8. Registry of DICOM Directory Structuring Elements
9. Registry of DICOM Dynamic RTP Payload Elements
A. Registry of DICOM Unique Identifiers (UIDs) (Normative)
B. Well-Known Color Palettes (Normative)
B.1. Standard Color Palettes
B.1.1. Hot Iron Color Palette
B.1.1.1. Hot Iron Color Palette Description (Informative)
B.1.1.2. Hot Iron Color Palette Definition
B.1.2. PET Color Palette
B.1.2.1. PET Color Palette Description (Informative)
B.1.2.2. PET Color Palette Definition
B.1.3. Hot Metal Blue Color Palette
B.1.3.1. Hot Metal Blue Color Palette Description (Informative)
B.1.3.2. Hot Metal Blue Color Palette Definition
B.1.4. PET 20 Step Color Palette
B.1.4.1. PET 20 Step Color Palette Description (Informative)
B.1.4.2. PET 20 Step Color Palette Definition
B.1.5. Spring Color Palette
B.1.5.1. Spring Color Palette Description (Informative)
B.1.5.2. Spring Color Palette Definition
B.1.6. Summer Color Palette
B.1.6.1. Summer Color Palette Description (Informative)
B.1.6.2. Summer Color Palette Definition
B.1.7. Fall Color Palette
B.1.7.1. Fall Color Palette Description (Informative)
B.1.7.2. Fall Color Palette Definition
B.1.8. Winter Color Palette
B.1.8.1. Winter Color Palette Description (Informative)
B.1.8.2. Winter Color Palette Definition
B.2. Localized Standard Color Palette Description Values
B.2.1. French
B.2.2. German

List of Figures

B.1.1.1-1. Nuclear Medicine image with and without Hot Iron Palette applied.
B.1.2.1-1. PET image with PET Palette superimposed over grayscale CT image.
B.1.3.1-1. PET image with Hot Metal Blue Palette applied.
B.1.4.1-1. PET image with PET 20 Step Palette applied.
B.1.5.1-1. MR image with Spring LUT Palette applied.
B.1.6.1-1. MR image with Summer LUT Palette applied.
B.1.7.1-1. MR image with Fall LUT Palette applied.
B.1.8.1-1. MR image with Winter LUT Palette applied.

List of Tables

6-1. Registry of DICOM Data Elements
7-1. Registry of DICOM File Meta Elements
8-1. Registry of DICOM Directory Structuring Elements
9-1. Registry of DICOM Dynamic RTP Payload Elements
A-1. UID Values
A-2. Well-known Frames of Reference
A-3. Context Group UID Values
A-4. Template UID Values
B.1-1. Standard Color Palettes
B.1.1.2-1. Hot Iron Color Palette Descriptor
B.1.1.2-2. Hot Iron Color Palette Data
B.1.2.2-1. PET Color Palette Descriptor
B.1.2.2-2. PET Color Palette Data
B.1.3.2-1. Hot Metal Blue Color Palette Descriptor
B.1.3.2-2. Hot Metal Blue Color Palette Data
B.1.4.2-1. PET 20 Step Color Palette Descriptor
B.1.4.2-2. PET 20 Step Color Palette Data
B.1.5.2-1. Spring Color Palette Descriptor
B.1.5.2-2. Spring Segmented Color Palette Data
B.1.6.2-1. Summer Color Palette Descriptor
B.1.6.2-2. Summer Segmented Color Palette Data
B.1.7.2-1. Fall Color Palette Descriptor
B.1.7.2-2. Fall Segmented Color Palette Data
B.1.8.2-1. Winter Color Palette Descriptor
B.1.8.2-2. Winter Segmented Color Palette Data
B.2.1-1. French Standard Color Palette Description Values
B.2.2-1. German Standard Color Palette Description Values

Notice and Disclaimer

The information in this publication was considered technically sound by the consensus of persons engaged in the development and approval of the document at the time it was developed. Consensus does not necessarily mean that there is unanimous agreement among every person participating in the development of this document.

NEMA standards and guideline publications, of which the document contained herein is one, are developed through a voluntary consensus standards development process. This process brings together volunteers and/or seeks out the views of persons who have an interest in the topic covered by this publication. While NEMA administers the process and establishes rules to promote fairness in the development of consensus, it does not write the document and it does not independently test, evaluate, or verify the accuracy or completeness of any information or the soundness of any judgments contained in its standards and guideline publications.

NEMA disclaims liability for any personal injury, property, or other damages of any nature whatsoever, whether special, indirect, consequential, or compensatory, directly or indirectly resulting from the publication, use of, application, or reliance on this document. NEMA disclaims and makes no guaranty or warranty, expressed or implied, as to the accuracy or completeness of any information published herein, and disclaims and makes no warranty that the information in this document will fulfill any of your particular purposes or needs. NEMA does not undertake to guarantee the performance of any individual manufacturer or seller's products or services by virtue of this standard or guide.

In publishing and making this document available, NEMA is not undertaking to render professional or other services for or on behalf of any person or entity, nor is NEMA undertaking to perform any duty owed by any person or entity to someone else. Anyone using this document should rely on his or her own independent judgment or, as appropriate, seek the advice of a competent professional in determining the exercise of reasonable care in any given circumstances. Information and other standards on the topic covered by this publication may be available from other sources, which the user may wish to consult for additional views or information not covered by this publication.

NEMA has no power, nor does it undertake to police or enforce compliance with the contents of this document. NEMA does not certify, test, or inspect products, designs, or installations for safety or health purposes. Any certification or other statement of compliance with any health or safety-related information in this document shall not be attributable to NEMA and is solely the responsibility of the certifier or maker of the statement.

Foreword

This DICOM Standard was developed according to the procedures of the DICOM Standards Committee.

The DICOM Standard is structured as a multi-part document using the guidelines established in [ISO/IEC Directives, Part 2].

PS3.1 should be used as the base reference for the current parts of this Standard.

1 Scope and Field of Application

This Part of the DICOM Standard is PS 3.6 of a multi-part standard produced to facilitate the interchange of information between digital imaging computer systems in medical environments. This interchange will enhance diagnostic imaging and potentially other clinical applications. The multi-part DICOM Standard covers the protocols and data that shall be supplied to achieve this interchange of information.

This Part of the Standard contains the registry of all DICOM Data Elements and all DICOM Unique Identifiers that are defined within the DICOM Standard.

DICOM® is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications of medical information, all rights reserved.

HL7® and CDA® are the registered trademarks of Health Level Seven International, all rights reserved.

SNOMED®, SNOMED Clinical Terms®, SNOMED CT® are the registered trademarks of the International Health Terminology Standards Development Organisation (IHTSDO), all rights reserved.

LOINC® is the registered trademark of Regenstrief Institute, Inc, all rights reserved.

2 Normative References

The following standards contain provisions that, through references in this text, constitute provisions of this Standard. At the time of publication, the editions indicated were valid. All standards are subject to revision, and parties to agreements based on this Standard are encouraged to investigate the possibilities of applying the most recent editions of the standards indicated below.

[ACR-NEMA 300] ACR-NEMA. 1988. Digital Imaging and Communications.

[ASTM E2339-11] ASTM. 2011. Standard Practice for Digital Imaging and Communication in Nondestructive Evaluation (DICONDE).

[IEC 61217] IEC. 2011. Ed 2. Radiotherapy Equipment - Coordinates, Movements and Scales.

[ISO 8649] ISO. 1988. Information processing systems - Open Systems Interconnection - Service definition for the Association Control Service Element (ACSE).

[ISO 8822] ISO. 1988. Information processing systems - Open Systems Interconnection - Connection oriented presentation service definition.

[ISO/IEC Directives, Part 2] ISO/IEC. 2021. 9.0. Rules for the structure and drafting of International Standards. http://www.iso.org/sites/directives/current/part2/index.xhtml .

[NEMA IIC 1 v02] NEMA. 2012. Digital Imaging and Communications in Security (DICOS) Information Object Definitions (IODs).

3 Definitions

For the purposes of this Standard, the following definitions apply.

3.1 DICOM Introduction and Overview Definition

This Part of the Standard makes use of the following terms defined in PS3.1:

Attribute

See Attribute in PS3.1 .

Service-Object Pair Class (SOP Class)

See Service-Object Pair Class in PS3.1 .

3.2 DICOM Information Object Definition

This Part of the Standard makes use of the following term defined in PS3.3:

Attribute Tag

See Attribute Tag in PS3.3 .

3.3 DICOM Data Structures and Encoding Definitions

This Part of the Standard makes use of the following terms defined in PS3.5:

Data Element

See Data Element in PS3.5

Data Element Tag

See Data Element Tag in PS3.5

Element Number

See Element Number in PS3.5

Group Number

See Group Number in PS3.5

Repeating Group

See Repeating Group in PS3.5

Retired Data Element

See Retired Data Element in PS3.5

Standard Data Element

See Standard Data Element in PS3.5

Value Multiplicity (VM)

See Value Multiplicity in PS3.5

Value Representation (VR)

See Value Representation in PS3.5

3.4 DICOM Data Dictionary

The following definition is commonly used in this Standard:

Tag

A unique identifier for an element of information composed of an ordered pair of numbers (a Group Number followed by an Element Number), which is used to identify Attributes and corresponding Data Elements.

3.5 DICOM Service Class Specifications Definitions

This Part of the Standard makes use of the following terms defined in PS3.4:

Service-Object Pair Instance (SOP Instance)

See Service-Object Pair Instance in PS3.4

4 Symbols and Abbreviations

The following symbols and abbreviations are used in this Part of the Standard.

ACR

American College of Radiology

DICOM

Digital Imaging and Communications in Medicine

DICONDE

Digital Imaging and Communication in Nondestructive Evaluation

DICOS

Digital Imaging and Communication for Security

IOD

Information Object Definition

ISO

International Standards Organization

JIRA

Japan Medical Imaging and Radiological Systems Industries Association

NEMA

National Electrical Manufacturers Association

OSI

Open Systems Interconnection

RDSR

Radiation Dose Structured Report

TCP/IP

Transmission Control Protocol/Internet Protocol

UID

Unique Identifier

VM

Value Multiplicity

VR

Value Representation

5 Conventions

Word(s) are capitalized in this document to help the reader understand that these word(s) have been previously defined in Section 3 and are to be interpreted with that meaning.

A Data Element Tag is represented as (gggg,eeee), where gggg equates to the Group Number and eeee equates to the Element Number within that Group. Data Element Tags are represented in hexadecimal notation as specified for each named Data Element in this Standard.

Where an "x" is shown in a group or element number, e.g. (ggxx,eeee), "x" means any value from 0 through F inclusive. The resulting group number is still required to be even, as defined by PS3.5, since these are Standard Data Element Tags.

"RET" is used to indicate that the corresponding Data Element, SOP Class, or Transfer Syntax has been retired. Retired items are shown italicized. For retired items, the edition of the Standard in parentheses is the edition in which the item last appeared before it was retired. When the name of a retired Data Element has been reused, the retired element has the qualifier "(Retired)" added, or "(Trial)" in the cases in which the Data Element was used in a Draft For Trial Implementation but not standardized.

Note

The use of retired items is supported in this version of DICOM. However, new implementations are strongly encouraged to implement alternative Data Elements, SOP Classes or Transfer Syntaxes.

"Note n" is used to indicate that further information is provided at the end of the corresponding table. The "n" is the consecutive number of the note. This information is not inserted directly into the tables in order to preserve their simple structure, e.g., for automatic processing of the contents.

6 Registry of DICOM Data Elements

Note

For data elements that were present in ACR-NEMA 1.0 and 2.0 and that have been retired, the specifications of Value Representation and Value Multiplicity provided are recommendations for the purpose of interpreting their values in objects created in accordance with earlier editions of this Standard. These recommendations are suggested as most appropriate for a particular data element; however, there is no guarantee that historical objects will not violate some requirements or specified VR and/or VM.

Table 6-1. Registry of DICOM Data Elements

Tag

Name

Keyword

VR

VM

(0008,0001)

Length to End

Length​To​End

UL

1

RET

(0008,0005)

Specific Character Set

Specific​Character​Set

CS

1-n

(0008,0006)

Language Code Sequence

Language​Code​Sequence

SQ

1

(0008,0008)

Image Type

Image​Type

CS

2-n

(0008,0010)

Recognition Code

Recognition​Code

SH

1

RET

(0008,0012)

Instance Creation Date

Instance​Creation​Date

DA

1

(0008,0013)

Instance Creation Time

Instance​Creation​Time

TM

1

(0008,0014)

Instance Creator UID

Instance​Creator​UID

UI

1

(0008,0015)

Instance Coercion DateTime

Instance​Coercion​Date​Time

DT

1

(0008,0016)

SOP Class UID

SOP​Class​UID

UI

1

(0008,0017)

Acquisition UID

Acquisition​UID

UI

1

(0008,0018)

SOP Instance UID

SOP​Instance​UID

UI

1

(0008,0019)

Pyramid UID

Pyramid​UID

UI

1

(0008,001A)

Related General SOP Class UID

Related​General​SOP​Class​UID

UI

1-n

(0008,001B)

Original Specialized SOP Class UID

Original​Specialized​SOP​Class​UID

UI

1

(0008,001C)

Synthetic Data

Synthetic​Data

CS

1

(0008,001D)

Sensitive Content Code Sequence

Sensitive​Content​Code​Sequence

SQ

1

(0008,0020)

Study Date

Study​Date

DA

1

(0008,0021)

Series Date

Series​Date

DA

1

(0008,0022)

Acquisition Date

Acquisition​Date

DA

1

(0008,0023)

Content Date

Content​Date

DA

1

(0008,0024)

Overlay Date

Overlay​Date

DA

1

RET (2004)

(0008,0025)

Curve Date

Curve​Date

DA

1

RET (2004)

(0008,002A)

Acquisition DateTime

Acquisition​Date​Time

DT

1

(0008,0030)

Study Time

Study​Time

TM

1

(0008,0031)

Series Time

Series​Time

TM

1

(0008,0032)

Acquisition Time

Acquisition​Time

TM

1

(0008,0033)

Content Time

Content​Time

TM

1

(0008,0034)

Overlay Time

Overlay​Time

TM

1

RET (2004)

(0008,0035)

Curve Time

Curve​Time

TM

1

RET

(0008,0040)

Data Set Type

Data​Set​Type

US

1

RET (2004)

(0008,0041)

Data Set Subtype

Data​Set​Subtype

LO

1

RET (2004)

(0008,0042)

Nuclear Medicine Series Type

Nuclear​Medicine​Series​Type

CS

1

RET (2004)

(0008,0050)

Accession Number

Accession​Number

SH

1

(0008,0051)

Issuer of Accession Number Sequence

Issuer​Of​Accession​Number​Sequence

SQ

1

(0008,0052)

Query/Retrieve Level

Query​Retrieve​Level

CS

1

(0008,0053)

Query/Retrieve View

Query​Retrieve​View

CS

1

(0008,0054)

Retrieve AE Title

Retrieve​AE​Title

AE

1-n

(0008,0055)

Station AE Title

Station​AE​Title

AE

1

(0008,0056)

Instance Availability

Instance​Availability

CS

1

(0008,0058)

Failed SOP Instance UID List

Failed​SOP​Instance​UID​List

UI

1-n

(0008,0060)

Modality

Modality

CS

1

(0008,0061)

Modalities in Study

Modalities​In​Study

CS

1-n

(0008,0062)

SOP Classes in Study

SOP​Classes​In​Study

UI

1-n

(0008,0063)

Anatomic Regions in Study Code Sequence

Anatomic​Regions​In​StudyC​ode​Sequence

SQ

1

(0008,0064)

Conversion Type

Conversion​Type

CS

1

(0008,0068)

Presentation Intent Type

Presentation​Intent​Type

CS

1

(0008,0070)

Manufacturer

Manufacturer

LO

1

(0008,0080)

Institution Name

Institution​Name

LO

1

(0008,0081)

Institution Address

Institution​Address

ST

1

(0008,0082)

Institution Code Sequence

Institution​Code​Sequence

SQ

1

(0008,0090)

Referring Physician's Name

Referring​Physician​Name

PN

1

(0008,0092)

Referring Physician's Address

Referring​Physician​Address

ST

1

(0008,0094)

Referring Physician's Telephone Numbers

Referring​Physician​Telephone​Numbers

SH

1-n

(0008,0096)

Referring Physician Identification Sequence

Referring​Physician​Identification​Sequence

SQ

1

(0008,009C)

Consulting Physician's Name

Consulting​Physician​Name

PN

1-n

(0008,009D)

Consulting Physician Identification Sequence

Consulting​Physician​Identification​Sequence

SQ

1

(0008,0100)

Code Value

Code​Value

SH

1

(0008,0101)

Extended Code Value

Extended​Code​Value

LO

1

DICOS

(0008,0102)

Coding Scheme Designator

Coding​Scheme​Designator

SH

1

(0008,0103)

Coding Scheme Version

Coding​Scheme​Version

SH

1

(0008,0104)

Code Meaning

Code​Meaning

LO

1

(0008,0105)

Mapping Resource

Mapping​Resource

CS

1

(0008,0106)

Context Group Version

Context​Group​Version

DT

1

(0008,0107)

Context Group Local Version

Context​Group​Local​Version

DT

1

(0008,0108)

Extended Code Meaning

Extended​Code​Meaning

LT

1

DICOS

(0008,0109)

Coding Scheme Resources Sequence

Coding​Scheme​Resources​Sequence

SQ

1

(0008,010A)

Coding Scheme URL Type

Coding​Scheme​URL​Type

CS

1

(0008,010B)

Context Group Extension Flag

Context​Group​Extension​Flag

CS

1

(0008,010C)

Coding Scheme UID

Coding​Scheme​UID

UI

1

(0008,010D)

Context Group Extension Creator UID

Context​Group​Extension​Creator​UID

UI

1

(0008,010E)

Coding Scheme URL

Coding​Scheme​URL

UR

1

(0008,010F)

Context Identifier

Context​Identifier

CS

1

(0008,0110)

Coding Scheme Identification Sequence

Coding​Scheme​Identification​Sequence

SQ

1

(0008,0112)

Coding Scheme Registry

Coding​Scheme​Registry

LO

1

(0008,0114)

Coding Scheme External ID

Coding​Scheme​External​ID

ST

1

(0008,0115)

Coding Scheme Name

Coding​Scheme​Name

ST

1

(0008,0116)

Coding Scheme Responsible Organization

Coding​Scheme​Responsible​Organization

ST

1

(0008,0117)

Context UID

Context​UID

UI

1

(0008,0118)

Mapping Resource UID

Mapping​Resource​UID

UI

1

(0008,0119)

Long Code Value

Long​Code​Value

UC

1

(0008,0120)

URN Code Value

URN​Code​Value

UR

1

(0008,0121)

Equivalent Code Sequence

Equivalent​Code​Sequence

SQ

1

(0008,0122)

Mapping Resource Name

Mapping​Resource​Name

LO

1

(0008,0123)

Context Group Identification Sequence

Context​Group​Identification​Sequence

SQ

1

(0008,0124)

Mapping Resource Identification Sequence

Mapping​Resource​Identification​Sequence

SQ

1

(0008,0201)

Timezone Offset From UTC

Timezone​Offset​From​UTC

SH

1

(0008,0202)

RET (2020c)

(0008,0220)

Responsible Group Code Sequence

Responsible​Group​Code​Sequence

SQ

1

(0008,0221)

Equipment Modality

Equipment​Modality

CS

1

(0008,0222)

Manufacturer's Related Model Group

Manufacturer​Related​Model​Group

LO

1

(0008,0300)

Private Data Element Characteristics Sequence

Private​Data​Element​Characteristics​Sequence

SQ

1

(0008,0301)

Private Group Reference

Private​Group​Reference

US

1

(0008,0302)

Private Creator Reference

Private​Creator​Reference

LO

1

(0008,0303)

Block Identifying Information Status

Block​Identifying​Information​Status

CS

1

(0008,0304)

Nonidentifying Private Elements

Nonidentifying​Private​Elements

US

1-n

(0008,0306)

Identifying Private Elements

Identifying​Private​Elements

US

1-n

(0008,0305)

Deidentification Action Sequence

Deidentification​Action​Sequence

SQ

1

(0008,0307)

Deidentification Action

Deidentification​Action

CS

1

(0008,0308)

Private Data Element

Private​Data​Element

US

1

(0008,0309)

Private Data Element Value Multiplicity

Private​Data​Element​Value​Multiplicity

UL

1-3

(0008,030A)

Private Data Element Value Representation

Private​Data​Element​Value​Representation

CS

1

(0008,030B)

Private Data Element Number of Items

Private​Data​Element​Number​Of​Items

UL

1-2

(0008,030C)

Private Data Element Name

Private​Data​Element​Name

UC

1

(0008,030D)

Private Data Element Keyword

Private​Data​Element​Keyword

UC

1

(0008,030E)

Private Data Element Description

Private​Data​Element​Description

UT

1

(0008,030F)

Private Data Element Encoding

Private​Data​Element​Encoding

UT

1

(0008,0310)

Private Data Element Definition Sequence

Private​Data​Element​Definition​Sequence

SQ

1

(0008,0400)

Scope of Inventory Sequence

Scope​Of​Inventory​Sequence

SQ

1

(0008,0401)

Inventory Purpose

Inventory​Purpose

LT

1

(0008,0402)

Inventory Instance Description

Inventory​Instance​Description

LT

1

(0008,0403)

Inventory Level

Inventory​Level

CS

1

(0008,0404)

Item Inventory DateTime

Item​Inventory​Date​Time

DT

1

(0008,0405)

Removed from Operational Use

Removed​From​Operational​Use

CS

1

(0008,0406)

Reason for Removal Code Sequence

Reason​For​Removal​Code​Sequence

SQ

1

(0008,0407)

Stored Instance Base URI

Stored​Instance​Base​URI

UR

1

(0008,0408)

Folder Access URI

Folder​Access​URI

UR

1

(0008,0409)

File Access URI

File​Access​URI

UR

1

(0008,040A)

Container File Type

Container​File​Type

CS

1

(0008,040B)

Filename in Container

Filename​In​Container

UR

1

(0008,040C)

File Offset in Container

File​Offset​In​Container

UV

1

(0008,040D)

File Length in Container

File​Length​In​Container

UV

1

(0008,040E)

Stored Instance Transfer Syntax UID

Stored​Instance​Transfer​Syntax​UID

UI

1

(0008,040F)

Extended Matching Mechanisms

Extended​Matching​Mechanisms

CS

1-n

(0008,0410)

Range Matching Sequence

Range​Matching​Sequence

SQ

1

(0008,0411)

List of UID Matching Sequence

List​Of​UID​Matching​Sequence

SQ

1

(0008,0412)

Empty Value Matching Sequence

Empty​Value​Matching​Sequence

SQ

1

(0008,0413)

General Matching Sequence

General​Matching​Sequence

SQ

1

(0008,0414)

Requested Status Interval

Requested​Status​Interval

US

1

(0008,0415)

Retain Instances

Retain​Instances

CS

1

(0008,0416)

Expiration DateTime

Expiration​Date​Time

DT

1

(0008,0417)

Transaction Status

Transaction​Status

CS

1

(0008,0418)

Transaction Status Comment

Transaction​Status​Comment

LT

1

(0008,0419)

File Set Access Sequence

File​Set​Access​Sequence

SQ

1

(0008,041A)

File Access Sequence

File​Access​Sequence

SQ

1

(0008,041B)

Record Key

Record​Key

OB

1

(0008,041C)

Prior Record Key

Prior​Record​Key

OB

1

(0008,041D)

Metadata Sequence

Metadata​Sequence

SQ

1

(0008,041E)

Updated Metadata Sequence

Updated​Metadata​Sequence

SQ

1

(0008,041F)

Study Update DateTime

Study​Update​Date​Time

DT

1

(0008,0420)

Inventory Access End Points Sequence

Inventory​Access​End​Points​Sequence

SQ

1

(0008,0421)

Study Access End Points Sequence

Study​Access​End​Points​Sequence

SQ

1

(0008,0422)

Incorporated Inventory Instance Sequence

Incorporated​Inventory​Instance​Sequence

SQ

1

(0008,0423)

Inventoried Studies Sequence

Inventoried​Studies​Sequence

SQ

1

(0008,0424)

Inventoried Series Sequence

Inventoried​Series​Sequence

SQ

1

(0008,0425)

Inventoried Instances Sequence

Inventoried​Instances​Sequence

SQ

1

(0008,0426)

Inventory Completion Status

Inventory​Completion​Status

CS

1

(0008,0427)

Number of Study Records in Instance

Number​Of​Study​Records​In​Instance

UL

1

(0008,0428)

Total Number of Study Records

Total​Number​Of​Study​Records

UV

1

(0008,0429)

Maximum Number of Records

Maximum​Number​Of​Records

UV

1

(0008,1000)

Network ID

Network​ID

AE

1

RET

(0008,1010)

Station Name

Station​Name

SH

1

(0008,1030)

Study Description

Study​Description

LO

1

(0008,1032)

Procedure Code Sequence

Procedure​Code​Sequence

SQ

1

(0008,103E)

Series Description

Series​Description

LO

1

(0008,103F)

Series Description Code Sequence

Series​Description​Code​Sequence

SQ

1

(0008,1040)

Institutional Department Name

Institutional​Department​Name

LO

1

(0008,1041)

Institutional Department Type Code Sequence

Institutional​Department​Type​Code​Sequence

SQ

1

(0008,1048)

Physician(s) of Record

Physicians​Of​Record

PN

1-n

(0008,1049)

Physician(s) of Record Identification Sequence

Physicians​Of​Record​Identification​Sequence

SQ

1

(0008,1050)

Performing Physician's Name

Performing​Physician​Name

PN

1-n

(0008,1052)

Performing Physician Identification Sequence

Performing​Physician​Identification​Sequence

SQ

1

(0008,1060)

Name of Physician(s) Reading Study

Name​Of​Physicians​Reading​Study

PN

1-n

(0008,1062)

Physician(s) Reading Study Identification Sequence

Physicians​Reading​Study​Identification​Sequence

SQ

1

(0008,1070)

Operators' Name

Operators​Name

PN

1-n

(0008,1072)

Operator Identification Sequence

Operator​Identification​Sequence

SQ

1

(0008,1080)

Admitting Diagnoses Description

Admitting​Diagnoses​Description

LO

1-n

(0008,1084)

Admitting Diagnoses Code Sequence

Admitting​Diagnoses​Code​Sequence

SQ

1

(0008,1088)

Pyramid Description

Pyramid​Description

LO

1

(0008,1090)

Manufacturer's Model Name

Manufacturer​Model​Name

LO

1

(0008,1100)

Referenced Results Sequence

Referenced​Results​Sequence

SQ

1

RET (2004)

(0008,1110)

Referenced Study Sequence

Referenced​Study​Sequence

SQ

1

(0008,1111)

Referenced Performed Procedure Step Sequence

Referenced​Performed​Procedure​Step​Sequence

SQ

1

(0008,1112)

Referenced Instances by SOP Class Sequence

Referenced​Instances​By​SOP​Class​Sequence

SQ

1

(0008,1115)

Referenced Series Sequence

Referenced​Series​Sequence

SQ

1

(0008,1120)

Referenced Patient Sequence

Referenced​Patient​Sequence

SQ

1

(0008,1125)

Referenced Visit Sequence

Referenced​Visit​Sequence

SQ

1

(0008,1130)

Referenced Overlay Sequence

Referenced​Overlay​Sequence

SQ

1

RET (2004)

(0008,1134)

Referenced Stereometric Instance Sequence

Referenced​Stereometric​Instance​Sequence

SQ

1

(0008,113A)

Referenced Waveform Sequence

Referenced​Waveform​Sequence

SQ

1

(0008,1140)

Referenced Image Sequence

Referenced​Image​Sequence

SQ

1

(0008,1145)

Referenced Curve Sequence

Referenced​Curve​Sequence

SQ

1

RET (2004)

(0008,114A)

Referenced Instance Sequence

Referenced​Instance​Sequence

SQ

1

(0008,114B)

Referenced Real World Value Mapping Instance Sequence

Referenced​Real​World​Value​Mapping​Instance​Sequence

SQ

1

(0008,114C)

Referenced Segmentation Sequence

Referenced​Segmentation​Sequence

SQ

1

(0008,114D)

Referenced Surface Segmentation Sequence

Referenced​Surface​Segmentation​Sequence

SQ

1

(0008,1150)

Referenced SOP Class UID

Referenced​SOP​Class​UID

UI

1

(0008,1155)

Referenced SOP Instance UID

Referenced​SOP​Instance​UID

UI

1

(0008,1156)

Definition Source Sequence

Definition​Source​Sequence

SQ

1

(0008,115A)

SOP Classes Supported

SOP​Classes​Supported

UI

1-n

(0008,1160)

Referenced Frame Number

Referenced​Frame​Number

IS

1-n

(0008,1161)

Simple Frame List

Simple​Frame​List

UL

1-n

(0008,1162)

Calculated Frame List

Calculated​Frame​List

UL

3-3n

(0008,1163)

Time Range

Time​Range

FD

2

(0008,1164)

Frame Extraction Sequence

Frame​Extraction​Sequence

SQ

1

(0008,1167)

Multi-frame Source SOP Instance UID

Multi​Frame​Source​SOP​Instance​UID

UI

1

(0008,1190)

Retrieve URL

Retrieve​URL

UR

1

(0008,1195)

Transaction UID

Transaction​UID

UI

1

(0008,1196)

Warning Reason

Warning​Reason

US

1

(0008,1197)

Failure Reason

Failure​Reason

US

1

(0008,1198)

Failed SOP Sequence

Failed​SOP​Sequence

SQ

1

(0008,1199)

Referenced SOP Sequence

Referenced​SOP​Sequence

SQ

1

(0008,119A)

Other Failures Sequence

Other​Failures​Sequence

SQ

1

(0008,119B)

Failed Study Sequence

Failed​Study​Sequence

SQ

1

(0008,1200)

Studies Containing Other Referenced Instances Sequence

Studies​Containing​Other​Referenced​Instances​Sequence

SQ

1

(0008,1250)

Related Series Sequence

Related​Series​Sequence

SQ

1

(0008,1301)

Principal Diagnosis Code Sequence

Principal​Diagnosis​Code​Sequence

SQ

1

(0008,1302)

Primary Diagnosis Code Sequence

Primary​Diagnosis​Code​Sequence

SQ

1

(0008,1303)

Secondary Diagnoses Code Sequence

Secondary​Diagnoses​Code​Sequence

SQ

1

(0008,1304)

Histological Diagnoses Code Sequence

Histological​Diagnoses​Code​Sequence

SQ

1

(0008,2110)

Lossy Image Compression (Retired)

Lossy​Image​Compression​Retired

CS

1

RET (1993)

(0008,2111)

Derivation Description

Derivation​Description

ST

1

(0008,2112)

Source Image Sequence

Source​Image​Sequence

SQ

1

(0008,2120)

Stage Name

Stage​Name

SH

1

(0008,2122)

Stage Number

Stage​Number

IS

1

(0008,2124)

Number of Stages

Number​Of​Stages

IS

1

(0008,2127)

View Name

View​Name

SH

1

(0008,2128)

View Number

View​Number

IS

1

(0008,2129)

Number of Event Timers

Number​Of​Event​Timers

IS

1

(0008,212A)

Number of Views in Stage

Number​Of​Views​In​Stage

IS

1

(0008,2130)

Event Elapsed Time(s)

Event​Elapsed​Times

DS

1-n

(0008,2132)

Event Timer Name(s)

Event​Timer​Names

LO

1-n

(0008,2133)

Event Timer Sequence

Event​Timer​Sequence

SQ

1

(0008,2134)

Event Time Offset

Event​Time​Offset

FD

1

(0008,2135)

Event Code Sequence

Event​Code​Sequence

SQ

1

(0008,2142)

Start Trim

Start​Trim

IS

1

(0008,2143)

Stop Trim

Stop​Trim

IS

1

(0008,2144)

Recommended Display Frame Rate

Recommended​Display​Frame​Rate

IS

1

(0008,2200)

Transducer Position

Transducer​Position

CS

1

RET (1993)

(0008,2204)

Transducer Orientation

Transducer​Orientation

CS

1

RET (1993)

(0008,2208)

Anatomic Structure

Anatomic​Structure

CS

1

RET (1993)

(0008,2218)

Anatomic Region Sequence

Anatomic​Region​Sequence

SQ

1

(0008,2220)

Anatomic Region Modifier Sequence

Anatomic​Region​Modifier​Sequence

SQ

1

(0008,2228)

Primary Anatomic Structure Sequence

Primary​Anatomic​Structure​Sequence

SQ

1

(0008,2229)

Anatomic Structure, Space or Region Sequence

Anatomic​Structure​Space​Or​Region​Sequence

SQ

1

RET (2017c)

(0008,2230)

Primary Anatomic Structure Modifier Sequence

Primary​Anatomic​Structure​Modifier​Sequence

SQ

1

(0008,2240)

Transducer Position Sequence

Transducer​Position​Sequence

SQ

1

RET (2004)

(0008,2242)

Transducer Position Modifier Sequence

Transducer​Position​Modifier​Sequence

SQ

1

RET (2004)

(0008,2244)

Transducer Orientation Sequence

Transducer​Orientation​Sequence

SQ

1

RET (2004)

(0008,2246)

Transducer Orientation Modifier Sequence

Transducer​Orientation​Modifier​Sequence

SQ

1

RET (2004)

(0008,2251)

Anatomic Structure Space Or Region Code Sequence (Trial)

Anatomic​Structure​Space​Or​Region​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,2253)

Anatomic Portal Of Entrance Code Sequence (Trial)

Anatomic​Portal​Of​Entrance​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,2255)

Anatomic Approach Direction Code Sequence (Trial)

Anatomic​Approach​Direction​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,2256)

Anatomic Perspective Description (Trial)

Anatomic​Perspective​Description​Trial

ST

1

RET (2007)

(0008,2257)

Anatomic Perspective Code Sequence (Trial)

Anatomic​Perspective​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,2258)

Anatomic Location Of Examining Instrument Description (Trial)

Anatomic​Location​Of​Examining​Instrument​Description​Trial

ST

1

RET (2007)

(0008,2259)

Anatomic Location Of Examining Instrument Code Sequence (Trial)

Anatomic​Location​Of​Examining​Instrument​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,225A)

Anatomic Structure Space Or Region Modifier Code Sequence (Trial)

Anatomic​Structure​Space​Or​Region​Modifier​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,225C)

On Axis Background Anatomic Structure Code Sequence (Trial)

On​Axis​Background​Anatomic​Structure​Code​Sequence​Trial

SQ

1

RET (2007)

(0008,3001)

Alternate Representation Sequence

Alternate​Representation​Sequence

SQ

1

(0008,3002)

Available Transfer Syntax UID

Available​Transfer​Syntax​UID

UI

1-n

(0008,3010)

Irradiation Event UID

Irradiation​Event​UID

UI

1-n

(0008,3011)

Source Irradiation Event Sequence

Source​Irradiation​Event​Sequence

SQ

1

(0008,3012)

Radiopharmaceutical Administration Event UID

Radiopharmaceutical​Administration​Event​UID

UI

1

(0008,4000)

Identifying Comments

Identifying​Comments

LT

1

RET

(0008,9007)

Frame Type

Frame​Type

CS

4-5

(0008,9092)

Referenced Image Evidence Sequence

Referenced​Image​Evidence​Sequence

SQ

1

(0008,9121)

Referenced Raw Data Sequence

Referenced​Raw​Data​Sequence

SQ

1

(0008,9123)

Creator-Version UID

Creator​Version​UID

UI

1

(0008,9124)

Derivation Image Sequence

Derivation​Image​Sequence

SQ

1

(0008,9154)

Source Image Evidence Sequence

Source​Image​Evidence​Sequence

SQ

1

(0008,9205)

Pixel Presentation

Pixel​Presentation

CS

1

(0008,9206)

Volumetric Properties

Volumetric​Properties

CS

1

(0008,9207)

Volume Based Calculation Technique

Volume​Based​Calculation​Technique

CS

1

(0008,9208)

Complex Image Component

Complex​Image​Component

CS

1

(0008,9209)

Acquisition Contrast

Acquisition​Contrast

CS

1

(0008,9215)

Derivation Code Sequence

Derivation​Code​Sequence

SQ

1

(0008,9237)

Referenced Presentation State Sequence

Referenced​Presentation​State​Sequence

SQ

1

(0008,9410)

Referenced Other Plane Sequence

Referenced​Other​Plane​Sequence

SQ

1

(0008,9458)

Frame Display Sequence

Frame​Display​Sequence

SQ

1

(0008,9459)

Recommended Display Frame Rate in Float

Recommended​Display​Frame​Rate​In​Float

FL

1

(0008,9460)

Skip Frame Range Flag

Skip​Frame​Range​Flag

CS

1

(0010,0010)

Patient's Name

Patient​Name

PN

1

(0010,0011)

Person Names to Use Sequence

Person​Names​To​Use​Sequence

SQ

1

(0010,0012)

Name to Use

Name​To​Use

LT

1

(0010,0013)

Name to Use Comment

Name​To​Use​Comment

UT

1

(0010,0014)

Third Person Pronouns Sequence

Third​Person​Pronouns​Sequence

SQ

1

(0010,0015)

Pronoun Code Sequence

Pronoun​Code​Sequence

SQ

1

(0010,0016)

Pronoun Comment

Pronoun​Comment

UT

1

(0010,0020)

Patient ID

Patient​ID

LO

1

(0010,0021)

Issuer of Patient ID

Issuer​Of​Patient​ID

LO

1

(0010,0022)

Type of Patient ID

Type​Of​Patient​ID

CS

1

(0010,0024)

Issuer of Patient ID Qualifiers Sequence

Issuer​Of​Patient​ID​Qualifiers​Sequence

SQ

1

(0010,0026)

Source Patient Group Identification Sequence

Source​Patient​Group​Identification​Sequence

SQ

1

(0010,0027)

Group of Patients Identification Sequence

Group​Of​Patients​Identification​Sequence

SQ

1

(0010,0028)

Subject Relative Position in Image

Subject​Relative​Position​In​Image

US

3

(0010,0030)

Patient's Birth Date

Patient​Birth​Date

DA

1

(0010,0032)

Patient's Birth Time

Patient​Birth​Time

TM

1

(0010,0033)

Patient's Birth Date in Alternative Calendar

Patient​Birth​Date​In​Alternative​Calendar

LO

1

(0010,0034)

Patient's Death Date in Alternative Calendar

Patient​Death​Date​In​Alternative​Calendar

LO

1

(0010,0035)

Patient's Alternative Calendar

Patient​Alternative​Calendar

CS

1

(0010,0040)

Patient's Sex

Patient​Sex

CS

1

(0010,0041)

Gender Identity Sequence

Gender​Identity​Sequence

SQ

1

(0010,0042)

Sex Parameters for Clinical Use Category Comment

Sex​Parameters​For​Clinical​Use​Category​Comment

UT

1

(0010,0043)

Sex Parameters for Clinical Use Category Sequence

Sex​Parameters​For​Clinical​Use​Category​Sequence

SQ

1

(0010,0044)

Gender Identity Code Sequence

Gender​Identity​Code​Sequence

SQ

1

(0010,0045)

Gender Identity Comment

Gender​Identity​Comment

UT

1

(0010,0046)

Sex Parameters for Clinical Use Category Code Sequence

Sex​Parameters​For​Clinical​Use​Category​Code​Sequence

SQ

1

(0010,0047)

Sex Parameters for Clinical Use Category Reference

Sex​Parameters​For​Clinical​Use​Category​Reference

UR

1-n

(0010,0050)

Patient's Insurance Plan Code Sequence

Patient​Insurance​Plan​Code​Sequence

SQ

1

(0010,0101)

Patient's Primary Language Code Sequence

Patient​Primary​Language​Code​Sequence

SQ

1

(0010,0102)

Patient's Primary Language Modifier Code Sequence

Patient​Primary​Language​Modifier​Code​Sequence

SQ

1

(0010,0200)

Quality Control Subject

Quality​Control​Subject

CS

1

(0010,0201)

Quality Control Subject Type Code Sequence

Quality​Control​Subject​Type​Code​Sequence

SQ

1

(0010,0212)

Strain Description

Strain​Description

UC

1

(0010,0213)

Strain Nomenclature

Strain​Nomenclature

LO

1

(0010,0214)

Strain Stock Number

Strain​Stock​Number

LO

1

(0010,0215)

Strain Source Registry Code Sequence

Strain​Source​Registry​Code​Sequence

SQ

1

(0010,0216)

Strain Stock Sequence

Strain​Stock​Sequence

SQ

1

(0010,0217)

Strain Source

Strain​Source

LO

1

(0010,0218)

Strain Additional Information

Strain​Additional​Information

UT

1

(0010,0219)

Strain Code Sequence

Strain​Code​Sequence

SQ

1

(0010,0221)

Genetic Modifications Sequence

Genetic​Modifications​​Sequence

SQ

1

(0010,0222)

Genetic Modifications Description

Genetic​Modifications​Description

UC

1

(0010,0223)

Genetic Modifications Nomenclature

Genetic​Modifications​Nomenclature

LO

1

(0010,0229)

Genetic Modifications Code Sequence

Genetic​Modifications​Code​Sequence

SQ

1

(0010,1000)

Other Patient IDs

Other​Patient​IDs

LO

1-n

RET (2017a)

(0010,1001)

Other Patient Names

Other​Patient​Names

PN

1-n

(0010,1002)

Other Patient IDs Sequence

Other​Patient​IDs​Sequence

SQ

1

(0010,1005)

Patient's Birth Name

Patient​Birth​Name

PN

1

(0010,1010)

Patient's Age

Patient​Age

AS

1

(0010,1020)

Patient's Size

Patient​Size

DS

1

(0010,1021)

Patient's Size Code Sequence

Patient​Size​Code​Sequence

SQ

1

(0010,1022)

Patient's Body Mass Index

Patient​Body​Mass​Index

DS

1

(0010,1023)

Measured AP Dimension

Measured​APDimension

DS

1

(0010,1024)

Measured Lateral Dimension

Measured​Lateral​Dimension

DS

1

(0010,1030)

Patient's Weight

Patient​Weight

DS

1

(0010,1040)

Patient's Address

Patient​Address

LO

1

(0010,1050)

Insurance Plan Identification

Insurance​Plan​Identification

LO

1-n

RET

(0010,1060)

Patient's Mother's Birth Name

Patient​Mother​Birth​Name

PN

1

(0010,1080)

Military Rank

Military​Rank

LO

1

(0010,1081)

Branch of Service

Branch​Of​Service

LO

1

(0010,1090)

Medical Record Locator

Medical​Record​Locator

LO

1

RET (2017a)

(0010,1100)

Referenced Patient Photo Sequence

Referenced​Patient​Photo​Sequence

SQ

1

(0010,2000)

Medical Alerts

Medical​Alerts

LO

1-n

(0010,2110)

Allergies

Allergies

LO

1-n

(0010,2150)

Country of Residence

Country​Of​Residence

LO

1

(0010,2152)

Region of Residence

Region​Of​Residence

LO

1

(0010,2154)

Patient's Telephone Numbers

Patient​Telephone​Numbers

SH

1-n

(0010,2155)

Patient's Telecom Information

Patient​Telecom​Information​

LT

1

(0010,2160)

Ethnic Group

Ethnic​Group

SH

1

RET (2025a)

(0010,2161)

Ethnic Group Code Sequence

Ethnic​Group​Code​Sequence

SQ

1

(0010,2162)

Ethnic Groups

Ethnic​Groups

UC

1-n

(0010,2180)

Occupation

Occupation

SH

1

(0010,21A0)

Smoking Status

Smoking​Status

CS

1

(0010,21B0)

Additional Patient History

Additional​Patient​History

LT

1

(0010,21C0)

Pregnancy Status

Pregnancy​Status

US

1

(0010,21D0)

Last Menstrual Date

Last​Menstrual​Date

DA

1

(0010,21F0)

Patient's Religious Preference

Patient​Religious​Preference

LO

1

(0010,2201)

Patient Species Description

Patient​Species​Description

LO

1

(0010,2202)

Patient Species Code Sequence

Patient​Species​Code​Sequence

SQ

1

(0010,2203)

Patient's Sex Neutered

Patient​Sex​Neutered

CS

1

(0010,2210)

Anatomical Orientation Type

Anatomical​Orientation​Type

CS

1

(0010,2292)

Patient Breed Description

Patient​Breed​Description

LO

1

(0010,2293)

Patient Breed Code Sequence

Patient​Breed​Code​Sequence

SQ

1

(0010,2294)

Breed Registration Sequence

Breed​Registration​Sequence

SQ

1

(0010,2295)

Breed Registration Number

Breed​Registration​Number

LO

1

(0010,2296)

Breed Registry Code Sequence

Breed​Registry​Code​Sequence

SQ

1

(0010,2297)

Responsible Person

Responsible​Person

PN

1

(0010,2298)

Responsible Person Role

Responsible​Person​Role

CS

1

(0010,2299)

Responsible Organization

Responsible​Organization

LO

1

(0010,4000)

Patient Comments

Patient​Comments

LT

1

(0010,9431)

Examined Body Thickness

Examined​Body​Thickness

FL

1

(0012,0010)

Clinical Trial Sponsor Name

Clinical​Trial​Sponsor​Name

LO

1

(0012,0020)

Clinical Trial Protocol ID

Clinical​Trial​Protocol​ID

LO

1

(0012,0021)

Clinical Trial Protocol Name

Clinical​Trial​Protocol​Name

LO

1

(0012,0022)

Issuer of Clinical Trial Protocol ID

Issuer​Of​Clinical​Tria​lProtocol​ID

LO

1

(0012,0023)

Other Clinical Trial Protocol IDs Sequence

Other​Clinical​Trial​Protocol​IDs​Sequence

SQ

1

(0012,0030)

Clinical Trial Site ID

Clinical​Trial​Site​ID

LO

1

(0012,0031)

Clinical Trial Site Name

Clinical​Trial​Site​Name

LO

1

(0012,0032)

Issuer of Clinical Trial Site ID

Issuer​Of​Clinical​Trial​Site​ID

LO

1

(0012,0040)

Clinical Trial Subject ID

Clinical​Trial​Subject​ID

LO

1

(0012,0041)

Issuer of Clinical Trial Subject ID

Issuer​Of​Clinical​Trial​Subject​ID

LO

1

(0012,0042)

Clinical Trial Subject Reading ID

Clinical​Trial​Subject​Reading​ID

LO

1

(0012,0043)

Issuer of Clinical Trial Subject Reading ID

Issuer​Of​Clinical​Trial​Subject​Reading​ID

LO

1

(0012,0050)

Clinical Trial Time Point ID

Clinical​Trial​Time​Point​ID

LO

1

(0012,0051)

Clinical Trial Time Point Description

Clinical​Trial​Time​Point​Description

ST

1

(0012,0052)

Longitudinal Temporal Offset from Event

Longitudinal​Temporal​Offset​From​Event

FD

1

(0012,0053)

Longitudinal Temporal Event Type

Longitudinal​Temporal​Event​Type

CS

1

(0012,0054)

Clinical Trial Time Point Type Code Sequence

Clinical​Trial​Time​Point​Type​Code​Sequence

SQ

1

(0012,0055)

Issuer of Clinical Trial Time Point ID

Issuer​Of​Clinical​Trial​Time​Point​ID

LO

1

(0012,0060)

Clinical Trial Coordinating Center Name

Clinical​Trial​Coordinating​Center​Name

LO

1

(0012,0062)

Patient Identity Removed

Patient​Identity​Removed

CS

1

(0012,0063)

De-identification Method

Deidentification​Method

LO

1-n

(0012,0064)

De-identification Method Code Sequence

Deidentification​Method​Code​Sequence

SQ

1

(0012,0071)

Clinical Trial Series ID

Clinical​Trial​Series​ID

LO

1

(0012,0072)

Clinical Trial Series Description

Clinical​Trial​Series​Description

LO

1

(0012,0073)

Issuer of Clinical Trial Series ID

Issuer​Of​Clinical​Trial​Series​ID

LO

1

(0012,0081)

Clinical Trial Protocol Ethics Committee Name

Clinical​Trial​Protocol​Ethics​Committee​Name

LO

1

(0012,0082)

Clinical Trial Protocol Ethics Committee Approval Number

Clinical​Trial​Protocol​Ethics​Committee​Approval​Number

LO

1

(0012,0083)

Consent for Clinical Trial Use Sequence

Consent​For​Clinical​Trial​Use​Sequence

SQ

1

(0012,0084)

Distribution Type

Distribution​Type

CS

1

(0012,0085)

Consent for Distribution Flag

Consent​For​Distribution​Flag

CS

1

(0012,0086)

Ethics Committee Approval Effectiveness Start Date

Ethics​Committee​Approval​Effectiveness​Start​Date

DA

1

(0012,0087)

Ethics Committee Approval Effectiveness End Date

Ethics​Committee​Approval​Effectiveness​End​Date

DA

1

(0014,0023)

CAD File Format

CAD​File​Format

ST

1

RET (2011)

(0014,0024)

Component Reference System

Component​Reference​System

ST

1

RET (2011)

(0014,0025)

Component Manufacturing Procedure

Component​Manufacturing​Procedure

ST

1

DICONDE

(0014,0028)

Component Manufacturer

Component​Manufacturer

ST

1

DICONDE

(0014,0030)

Material Thickness

Material​Thickness

DS

1-n

DICONDE

(0014,0032)

Material Pipe Diameter

Material​Pipe​Diameter

DS

1-n

DICONDE

(0014,0034)

Material Isolation Diameter

Material​Isolation​Diameter

DS

1-n

DICONDE

(0014,0042)

Material Grade

Material​Grade

ST

1

DICONDE

(0014,0044)

Material Properties Description

Material​Properties​Description

ST

1

DICONDE

(0014,0045)

Material Properties File Format (Retired)

Material​Properties​File​Format​Retired

ST

1

RET (2011)

(0014,0046)

Material Notes

Material​Notes

LT

1

DICONDE

(0014,0050)

Component Shape

Component​Shape

CS

1

DICONDE

(0014,0052)

Curvature Type

Curvature​Type

CS

1

DICONDE

(0014,0054)

Outer Diameter

Outer​Diameter

DS

1

DICONDE

(0014,0056)

Inner Diameter

Inner​Diameter

DS

1

DICONDE

(0014,0100)

Component Welder IDs

Component​Welder​IDs

LO

1-n

DICONDE

(0014,0101)

Secondary Approval Status

Secondary​Approval​Status

CS

1

DICONDE

(0014,0102)

Secondary Review Date

Secondary​Review​Date

DA

1

DICONDE

(0014,0103)

Secondary Review Time

Secondary​Review​Time

TM

1

DICONDE

(0014,0104)

Secondary Reviewer Name

Secondary​Reviewer​Name

PN

1

DICONDE

(0014,0105)

Repair ID

Repair​ID

ST

1

DICONDE

(0014,0106)

Multiple Component Approval Sequence

Multiple​Component​Approval​Sequence

SQ

1

DICONDE

(0014,0107)

Other Approval Status

Other​Approval​Status

CS

1-n

DICONDE

(0014,0108)

Other Secondary Approval Status

Other​Secondary​Approval​Status

CS

1-n

DICONDE

(0014,0200)

Data Element Label Sequence

Data​Element​Label​Sequence

SQ

1

DICONDE

(0014,0201)

Data Element Label Item Sequence

Data​Element​Label​Item​Sequence

SQ

1

DICONDE

(0014,0202)

Data Element

Data​Element

AT

1

DICONDE

(0014,0203)

Data Element Name

Data​Element​Name

LO

1

DICONDE

(0014,0204)

Data Element Description

Data​Element​Description

LO

1

DICONDE

(0014,0205)

Data Element Conditionality

Data​Element​Conditionality

CS

1

DICONDE

(0014,0206)

Data Element Minimum Characters

Data​Element​Minimum​Characters

IS

1

DICONDE

(0014,0207)

Data Element Maximum Characters

Data​Element​Maximum​Characters

IS

1

DICONDE

(0014,1010)

Actual Environmental Conditions

Actual​Environmental​Conditions

ST

1

DICONDE

(0014,1020)

Expiry Date

Expiry​Date

DA

1

DICONDE

(0014,1040)

Environmental Conditions

Environmental​Conditions

ST

1

DICONDE

(0014,2002)

Evaluator Sequence

Evaluator​Sequence

SQ

1

DICONDE

(0014,2004)

Evaluator Number

Evaluator​Number

IS

1

DICONDE

(0014,2006)

Evaluator Name

Evaluator​Name

PN

1

DICONDE

(0014,2008)

Evaluation Attempt

Evaluation​Attempt

IS

1

DICONDE

(0014,2012)

Indication Sequence

Indication​Sequence

SQ

1

DICONDE

(0014,2014)

Indication Number

Indication​Number

IS

1

DICONDE

(0014,2016)

Indication Label

Indication​Label

SH

1

DICONDE

(0014,2018)

Indication Description

Indication​Description

ST

1

DICONDE

(0014,201A)

Indication Type

Indication​Type

CS

1-n

DICONDE

(0014,201C)

Indication Disposition

Indication​Disposition

CS

1

DICONDE

(0014,201E)

Indication ROI Sequence

Indication​ROI​Sequence

SQ

1

DICONDE

(0014,2030)

Indication Physical Property Sequence

Indication​Physical​Property​Sequence

SQ

1

DICONDE

(0014,2032)

Property Label

Property​Label

SH

1

DICONDE

(0014,2202)

Coordinate System Number of Axes

Coordinate​System​Number​Of​Axes

IS

1

DICONDE

(0014,2204)

Coordinate System Axes Sequence

Coordinate​System​Axes​Sequence

SQ

1

DICONDE

(0014,2206)

Coordinate System Axis Description

Coordinate​System​Axis​Description

ST

1

DICONDE

(0014,2208)

Coordinate System Data Set Mapping

Coordinate​System​Data​Set​Mapping

CS

1

DICONDE

(0014,220A)

Coordinate System Axis Number

Coordinate​System​Axis​Number

IS

1

DICONDE

(0014,220C)

Coordinate System Axis Type

Coordinate​System​Axis​Type

CS

1

DICONDE

(0014,220E)

Coordinate System Axis Units

Coordinate​System​Axis​Units

CS

1

DICONDE

(0014,2210)

Coordinate System Axis Values

Coordinate​System​Axis​Values

OB

1

DICONDE

(0014,2220)

Coordinate System Transform Sequence

Coordinate​System​Transform​Sequence

SQ

1

DICONDE

(0014,2222)

Transform Description

Transform​Description

ST

1

DICONDE

(0014,2224)

Transform Number of Axes

Transform​Number​Of​Axes

IS

1

DICONDE

(0014,2226)

Transform Order of Axes

Transform​Order​Of​Axes

IS

1-n

DICONDE

(0014,2228)

Transformed Axis Units

Transformed​Axis​Units

CS

1

DICONDE

(0014,222A)

Coordinate System Transform Rotation and Scale Matrix

Coordinate​System​Transform​Rotation​And​Scale​Matrix

DS

1-n

DICONDE

(0014,222C)

Coordinate System Transform Translation Matrix

Coordinate​System​Transform​Translation​Matrix

DS

1-n

DICONDE

(0014,3011)

Internal Detector Frame Time

Internal​Detector​Frame​Time

DS

1

DICONDE

(0014,3012)

Number of Frames Integrated

Number​Of​Frames​Integrated

DS

1

DICONDE

(0014,3020)

Detector Temperature Sequence

Detector​Temperature​Sequence

SQ

1

DICONDE

(0014,3022)

Sensor Name

Sensor​Name

ST

1

DICONDE

(0014,3024)

Horizontal Offset of Sensor

Horizontal​Offset​Of​Sensor

DS

1

DICONDE

(0014,3026)

Vertical Offset of Sensor

Vertical​Offset​Of​Sensor

DS

1

DICONDE

(0014,3028)

Sensor Temperature

Sensor​Temperature

DS

1

DICONDE

(0014,3040)

Dark Current Sequence

Dark​Current​Sequence

SQ

1

DICONDE

(0014,3050)

Dark Current Counts

Dark​Current​Counts

OB or OW

1

DICONDE

(0014,3060)

Gain Correction Reference Sequence

Gain​Correction​Reference​Sequence

SQ

1

DICONDE

(0014,3070)

Air Counts

Air​Counts

OB or OW

1

DICONDE

(0014,3071)

KV Used in Gain Calibration

KV​Used​In​Gain​Calibration

DS

1

DICONDE

(0014,3072)

MA Used in Gain Calibration

MA​Used​In​Gain​Calibration

DS

1

DICONDE

(0014,3073)

Number of Frames Used for Integration

Number​Of​Frames​Used​For​Integration

DS

1

DICONDE

(0014,3074)

Filter Material Used in Gain Calibration

Filter​Material​Used​In​Gain​Calibration

LO

1

DICONDE

(0014,3075)

Filter Thickness Used in Gain Calibration

Filter​Thickness​Used​In​Gain​Calibration

DS

1

DICONDE

(0014,3076)

Date of Gain Calibration

Date​Of​Gain​Calibration

DA

1

DICONDE

(0014,3077)

Time of Gain Calibration

Time​Of​Gain​Calibration

TM

1

DICONDE

(0014,3080)

Bad Pixel Image

Bad​Pixel​Image

OB

1

DICONDE

(0014,3099)

Calibration Notes

Calibration​Notes

LT

1

DICONDE

(0014,3100)

Linearity Correction Technique

Linearity​Correction​Technique

LT

1

DICONDE

(0014,3101)

Beam Hardening Correction Technique

Beam​Hardening​Correction​Technique

LT

1

DICONDE

(0014,4002)

Pulser Equipment Sequence

Pulser​Equipment​Sequence

SQ

1

DICONDE

(0014,4004)

Pulser Type

Pulser​Type

CS

1

DICONDE

(0014,4006)

Pulser Notes

Pulser​Notes

LT

1

DICONDE

(0014,4008)

Receiver Equipment Sequence

Receiver​Equipment​Sequence

SQ

1

DICONDE

(0014,400A)

Amplifier Type

Amplifier​Type

CS

1

DICONDE

(0014,400C)

Receiver Notes

Receiver​Notes

LT

1

DICONDE

(0014,400E)

Pre-Amplifier Equipment Sequence

Pre​Amplifier​Equipment​Sequence

SQ

1

DICONDE

(0014,400F)

Pre-Amplifier Notes

Pre​Amplifier​Notes

LT

1

DICONDE

(0014,4010)

Transmit Transducer Sequence

Transmit​Transducer​Sequence

SQ

1

DICONDE

(0014,4011)

Receive Transducer Sequence

Receive​Transducer​Sequence

SQ

1

DICONDE

(0014,4012)

Number of Elements

Number​Of​Elements

US

1

DICONDE

(0014,4013)

Element Shape

Element​Shape

CS

1

DICONDE

(0014,4014)

Element Dimension A

Element​Dimension​A

DS

1

DICONDE

(0014,4015)

Element Dimension B

Element​Dimension​B

DS

1

DICONDE

(0014,4016)

Element Pitch A

Element​Pitch​A

DS

1

DICONDE

(0014,4017)

Measured Beam Dimension A

Measured​Beam​Dimension​A

DS

1

DICONDE

(0014,4018)

Measured Beam Dimension B

Measured​Beam​Dimension​B

DS

1

DICONDE

(0014,4019)

Location of Measured Beam Diameter

Location​Of​Measured​Beam​Diameter

DS

1

DICONDE

(0014,401A)

Nominal Frequency

Nominal​Frequency

DS

1

DICONDE

(0014,401B)

Measured Center Frequency

Measured​Center​Frequency

DS

1

DICONDE

(0014,401C)

Measured Bandwidth

Measured​Bandwidth

DS

1

DICONDE

(0014,401D)

Element Pitch B

Element​Pitch​B

DS

1

DICONDE

(0014,4020)

Pulser Settings Sequence

Pulser​Settings​Sequence

SQ

1

DICONDE

(0014,4022)

Pulse Width

Pulse​Width

DS

1

DICONDE

(0014,4024)

Excitation Frequency

Excitation​Frequency

DS

1

DICONDE

(0014,4026)

Modulation Type

Modulation​Type

CS

1

DICONDE

(0014,4028)

Damping

Damping

DS

1

DICONDE

(0014,4030)

Receiver Settings Sequence

Receiver​Settings​Sequence

SQ

1

DICONDE

(0014,4031)

Acquired Soundpath Length

Acquired​Soundpath​Length

DS

1

DICONDE

(0014,4032)

Acquisition Compression Type

Acquisition​Compression​Type

CS

1

DICONDE

(0014,4033)

Acquisition Sample Size

Acquisition​Sample​Size

IS

1

DICONDE

(0014,4034)

Rectifier Smoothing

Rectifier​Smoothing

DS

1

DICONDE

(0014,4035)

DAC Sequence

DAC​Sequence

SQ

1

DICONDE

(0014,4036)

DAC Type

DAC​Type

CS

1

DICONDE

(0014,4038)

DAC Gain Points

DAC​Gain​Points

DS

1-n

DICONDE

(0014,403A)

DAC Time Points

DAC​Time​Points

DS

1-n

DICONDE

(0014,403C)

DAC Amplitude

DAC​Amplitude

DS

1-n

DICONDE

(0014,4040)

Pre-Amplifier Settings Sequence

Pre​Amplifier​Settings​Sequence

SQ

1

DICONDE

(0014,4050)

Transmit Transducer Settings Sequence

Transmit​Transducer​Settings​Sequence

SQ

1

DICONDE

(0014,4051)

Receive Transducer Settings Sequence

Receive​Transducer​Settings​Sequence

SQ

1

DICONDE

(0014,4052)

Incident Angle

Incident​Angle

DS

1

DICONDE

(0014,4054)

Coupling Technique

Coupling​Technique

ST

1

DICONDE

(0014,4056)

Coupling Medium

Coupling​Medium

ST

1

DICONDE

(0014,4057)

Coupling Velocity

Coupling​Velocity

DS

1

DICONDE

(0014,4058)

Probe Center Location X

Probe​Center​Location​X

DS

1

DICONDE

(0014,4059)

Probe Center Location Z

Probe​Center​Location​Z

DS

1

DICONDE

(0014,405A)

Sound Path Length

Sound​Path​Length

DS

1

DICONDE

(0014,405C)

Delay Law Identifier

Delay​Law​Identifier

ST

1

DICONDE

(0014,4060)

Gate Settings Sequence

Gate​Settings​Sequence

SQ

1

DICONDE

(0014,4062)

Gate Threshold

Gate​Threshold

DS

1

DICONDE

(0014,4064)

Velocity of Sound

Velocity​Of​Sound

DS

1

DICONDE

(0014,4070)

Calibration Settings Sequence

Calibration​Settings​Sequence

SQ

1

DICONDE

(0014,4072)

Calibration Procedure

Calibration​Procedure

ST

1

DICONDE

(0014,4074)

Procedure Version

Procedure​Version

SH

1

DICONDE

(0014,4076)

Procedure Creation Date

Procedure​Creation​Date

DA

1

DICONDE

(0014,4078)

Procedure Expiration Date

Procedure​Expiration​Date

DA

1

DICONDE

(0014,407A)

Procedure Last Modified Date

Procedure​Last​Modified​Date

DA

1

DICONDE

(0014,407C)

Calibration Time

Calibration​Time

TM

1-n

DICONDE

(0014,407E)

Calibration Date

Calibration​Date

DA

1-n

DICONDE

(0014,4080)

Probe Drive Equipment Sequence

Probe​Drive​Equipment​Sequence

SQ

1

DICONDE

(0014,4081)

Drive Type

Drive​Type

CS

1

DICONDE

(0014,4082)

Probe Drive Notes

Probe​Drive​Notes

LT

1

DICONDE

(0014,4083)

Drive Probe Sequence

Drive​Probe​Sequence

SQ

1

DICONDE

(0014,4084)

Probe Inductance

Probe​Inductance

DS

1

DICONDE

(0014,4085)

Probe Resistance

Probe​Resistance

DS

1

DICONDE

(0014,4086)

Receive Probe Sequence

Receive​Probe​Sequence

SQ

1

DICONDE

(0014,4087)

Probe Drive Settings Sequence

Probe​Drive​Settings​Sequence

SQ

1

DICONDE

(0014,4088)

Bridge Resistors

Bridge​Resistors

DS

1

DICONDE

(0014,4089)

Probe Orientation Angle

Probe​Orientation​Angle

DS

1

DICONDE

(0014,408B)

User Selected Gain Y

User​Selected​Gain​Y

DS

1

DICONDE

(0014,408C)

User Selected Phase

User​Selected​Phase

DS

1

DICONDE

(0014,408D)

User Selected Offset X

User​Selected​Offset​X

DS

1

DICONDE

(0014,408E)

User Selected Offset Y

User​Selected​Offset​Y

DS

1

DICONDE

(0014,4091)

Channel Settings Sequence

Channel​Settings​Sequence

SQ

1

DICONDE

(0014,4092)

Channel Threshold

Channel​Threshold

DS

1

DICONDE

(0014,409A)

Scanner Settings Sequence

Scanner​Settings​Sequence

SQ

1

DICONDE

(0014,409B)

Scan Procedure

Scan​Procedure

ST

1

DICONDE

(0014,409C)

Translation Rate X

Translation​Rate​X

DS

1

DICONDE

(0014,409D)

Translation Rate Y

Translation​Rate​Y

DS

1

DICONDE

(0014,409F)

Channel Overlap

Channel​Overlap

DS

1

DICONDE

(0014,40A0)

Image Quality Indicator Type

Image​Quality​Indicator​Type

LO

1-n

DICONDE

(0014,40A1)

Image Quality Indicator Material

Image​Quality​Indicator​Material

LO

1-n

DICONDE

(0014,40A2)

Image Quality Indicator Size

Image​Quality​Indicator​Size

LO

1-n

DICONDE

(0014,4101)

Wave Dimensions Definition Sequence

Wave​Dimensions​Definition​Sequence

SQ

1

DICONDE

(0014,4102)

Wave Dimension Number

Wave​Dimension​Number

US

1

DICONDE

(0014,4103)

Wave Dimension Description

Wave​Dimension​Description

LO

1

DICONDE

(0014,4104)

Wave Dimension Unit

Wave​Dimension​Unit

US

1

DICONDE

(0014,4105)

Wave Dimension Value Type

Wave​Dimension​Value​Type

CS

1

DICONDE

(0014,4106)

Wave Dimension Values Sequence

Wave​Dimension​Values​Sequence

SQ

1-n

DICONDE

(0014,4107)

Referenced Wave Dimension

Referenced​Wave​Dimension

US

1

DICONDE

(0014,4108)

Integer Numeric Value

Integer​Numeric​Value

SL

1

DICONDE

(0014,4109)

Byte Numeric Value

Byte​Numeric​Value

OB

1

DICONDE

(0014,410A)

Short Numeric Value

Short​Numeric​Value

OW

1

DICONDE

(0014,410B)

Single Precision Floating Point Numeric Value

Single​Precision​Floating​Point​Numeric​Value

OF

1

DICONDE

(0014,410C)

Double Precision Floating Point Numeric Value

Double​Precision​Floating​Point​Numeric​Value

OD

1

DICONDE

(0014,5002)

LINAC Energy

LINAC​Energy

IS

1

DICONDE

(0014,5004)

LINAC Output

LINAC​Output

IS

1

DICONDE

(0014,5100)

Active Aperture

Active​Aperture

US

1

DICONDE

(0014,5101)

Total Aperture

Total​Aperture

DS

1

DICONDE

(0014,5102)

Aperture Elevation

Aperture​Elevation

DS

1

DICONDE

(0014,5103)

Main Lobe Angle

Main​Lobe​Angle

DS

1

DICONDE

(0014,5104)

Main Roof Angle

Main​Roof​Angle

DS

1

DICONDE

(0014,5105)

Connector Type

Connector​Type

CS

1

DICONDE

(0014,5106)

Wedge Model Number

Wedge​Model​Number

SH

1

DICONDE

(0014,5107)

Wedge Angle Float

Wedge​Angle​Float

DS

1

DICONDE

(0014,5108)

Wedge Roof Angle

Wedge​Roof​Angle

DS

1

DICONDE

(0014,5109)

Wedge Element 1 Position

Wedge​Element​1​Position

CS

1

DICONDE

(0014,510A)

Wedge Material Velocity

Wedge​Material​Velocity

DS

1

DICONDE

(0014,510B)

Wedge Material

Wedge​Material

SH

1

DICONDE

(0014,510C)

Wedge Offset Z

Wedge​Offset​Z

DS

1

DICONDE

(0014,510D)

Wedge Origin Offset X

Wedge​Origin​Offset​X

DS

1

DICONDE

(0014,510E)

Wedge Time Delay

Wedge​Time​Delay

DS

1

DICONDE

(0014,510F)

Wedge Name

Wedge​Name

SH

1

DICONDE

(0014,5110)

Wedge Manufacturer Name

Wedge​Manufacturer​Name

SH

1

DICONDE

(0014,5111)

Wedge Description

Wedge​Description

LO

1

DICONDE

(0014,5112)

Nominal Beam Angle

Nominal​Beam​Angle

DS

1

DICONDE

(0014,5113)

Wedge Offset X

Wedge​Offset​X

DS

1

DICONDE

(0014,5114)

Wedge Offset Y

Wedge​Offset​Y

DS

1

DICONDE

(0014,5115)

Wedge Total Length

Wedge​Total​Length

DS

1

DICONDE

(0014,5116)

Wedge In Contact Length

Wedge​In​Contact​Length

DS

1

DICONDE

(0014,5117)

Wedge Front Gap

Wedge​Front​Gap

DS

1

DICONDE

(0014,5118)

Wedge Total Height

Wedge​Total​Height

DS

1

DICONDE

(0014,5119)

Wedge Front Height

Wedge​Front​Height

DS

1

DICONDE

(0014,511A)

Wedge Rear Height

Wedge​Rear​Height

DS

1

DICONDE

(0014,511B)

Wedge Total Width

Wedge​Total​Width

DS

1

DICONDE

(0014,511C)

Wedge In Contact Width

Wedge​In​Contact​Width

DS

1

DICONDE

(0014,511D)

Wedge Chamfer Height

Wedge​Chamfer​Height

DS

1

DICONDE

(0014,511E)

Wedge Curve

Wedge​Curve

CS

1

DICONDE

(0014,511F)

Radius Along the Wedge

Radius​Along​Wedge

DS

1

DICONDE

(0014,6001)

Thermal Camera Settings Sequence

Thermal​Camera​Settings​Sequence

SQ

1

DICONDE

(0014,6002)

Acquisition Frame Rate

Acquisition​Frame​Rate

DS

1

DICONDE

(0014,6003)

Integration Time

Integration​Time

DS

1

DICONDE

(0014,6004)

Number of Calibration Frames

Number​Of​Calibration​Frames

DS

1

DICONDE

(0014,6005)

Number of Rows in Full Acquisition Image

Number​Of​Rows​In​Full​Acquisition​Image

DS

1

DICONDE

(0014,6006)

Number Of Columns in Full Acquisition Image

Number​Of​Columns​In​Full​Acquisition​Image

DS

1

DICONDE

(0014,6007)

Thermal Source Settings Sequence

Thermal​Source​Settings​Sequence

SQ

1

DICONDE

(0014,6008)

Source Horizontal Pitch

Source​Horizontal​Pitch

DS

1

DICONDE

(0014,6009)

Source Vertical Pitch

Source​Vertical​Pitch

DS

1

DICONDE

(0014,600A)

Source Horizontal Scan Speed

Source​Horizontal​Scan​Speed

DS

1

DICONDE

(0014,600B)

Thermal Source Modulation Frequency

Thermal​Source​Modulation​Frequency

DS

1

DICONDE

(0014,600C)

Induction Source Setting Sequence

Induction​Source​Setting​Sequence

SQ

1

DICONDE

(0014,600D)

Coil Frequency

Coil​Frequency

DS

1

DICONDE

(0014,600E)

Current Amplitude Across Coil

Current​Amplitude​Across​Coil

DS

1

DICONDE

(0014,600F)

Flash Source Setting Sequence

Flash​Source​Setting​Sequence

SQ

1

DICONDE

(0014,6010)

Flash Duration

Flash​Duration

DS

1

DICONDE

(0014,6011)

Flash Frame Number

Flash​Frame​Number

DS

1-n

DICONDE

(0014,6012)

Laser Source Setting Sequence

Laser​Source​Setting​Sequence

SQ

1

DICONDE

(0014,6013)

Horizontal Laser Spot Dimension

Horizontal​Laser​Spot​Dimension

DS

1

DICONDE

(0014,6014)

Vertical Laser Spot Dimension

Vertical​Laser​Spot​Dimension

DS

1

DICONDE

(0014,6015)

Laser Wavelength

Laser​Wavelength

DS

1

DICONDE

(0014,6016)

Laser Power

Laser​Power

DS

1

DICONDE

(0014,6017)

Forced Gas Setting Sequence

Forced​Gas​Setting​Sequence

SQ

1

DICONDE

(0014,6018)

Vibration Source Setting Sequence

Vibration​Source​Setting​Sequence

SQ

1

DICONDE

(0014,6019)

Vibration Excitation Frequency

Vibration​Excitation​Frequency

DS

1

DICONDE

(0014,601A)

Vibration Excitation Voltage

Vibration​Excitation​Voltage

DS

1

DICONDE

(0014,601B)

Thermography Data Capture Method

Thermography​Data​Capture​Method

CS

1

DICONDE

(0014,601C)

Thermal Technique

Thermal​Technique

CS

1

DICONDE

(0014,601D)

Thermal Camera Core Sequence

Thermal​Camera​Core​Sequence

SQ

1

DICONDE

(0014,601E)

Detector Wavelength Range

Detector​Wavelength​Range

CS

1

DICONDE

(0014,601F)

Thermal Camera Calibration Type

Thermal​Camera​Calibration​Type

CS

1

DICONDE

(0014,6020)

Acquisition Image Counter

Acquisition​Image​Counter

UV

1

DICONDE

(0014,6021)

Front Panel Temperature

Front​Panel​Temperature

DS

1

DICONDE

(0014,6022)

Air Gap Temperature

Air​Gap​Temperature

DS

1

DICONDE

(0014,6023)

Vertical Pixel Size

Vertical​Pixel​Size

DS

1

DICONDE

(0014,6024)

Horizontal Pixel Size

Horizontal​Pixel​Size

DS

1

DICONDE

(0014,6025)

Data Streaming Protocol

Data​Streaming​Protocol

ST

1-n

DICONDE

(0014,6026)

Lens Sequence

Lens​Sequence

SQ

1

DICONDE

(0014,6027)

Field of View

Field​Of​View

DS

1

DICONDE

(0014,6028)

Lens Filter Manufacturer

Lens​Filter​Manufacturer

LO

1

DICONDE

(0014,6029)

Cutoff Filter Type

Cutoff​Filter​Type

CS

1

DICONDE

(0014,602A)

Lens Filter Cut-Off Wavelength

Lens​Filter​Cut​Off​Wavelength

DS

1-n

DICONDE

(0014,602B)

Thermal Source Sequence

Thermal​Source​Sequence

SQ

1

DICONDE

(0014,602C)

Thermal Source Motion State

Thermal​Source​Motion​State

CS

1

DICONDE

(0014,602D)

Thermal Source Motion Type

Thermal​Source​Motion​Type

CS

1

DICONDE

(0014,602E)

Induction Heating Sequence

Induction​Heating​Sequence

SQ

1

DICONDE

(0014,602F)

Coil Configuration ID

Coil​Configuration​ID

ST

1

DICONDE

(0014,6030)

Number of Turns in Coil

Number​Of​Turns​In​Coil

DS

1

DICONDE

(0014,6031)

Shape of Individual Turn

Shape​Of​Individual​Turn

CS

1

DICONDE

(0014,6032)

Size of Individual Turn

Size​Of​Individual​Turn

DS

1-n

DICONDE

(0014,6033)

Distance Between Turns

Distance​Between​Turns

DS

1-n

DICONDE

(0014,6034)

Flash Heating Sequence

Flash​Heating​Sequence

SQ

1

DICONDE

(0014,6035)

Number of Lamps

Number​Of​Lamps

DS

1

DICONDE

(0014,6036)

Flash Synchronization Protocol

Flash​Synchronization​Protocol

ST

1

DICONDE

(0014,6037)

Flash Modification Status

Flash​Modification​Status

CS

1

DICONDE

(0014,6038)

Laser Heating Sequence

Laser​Heating​Sequence

SQ

1

DICONDE

(0014,6039)

Laser Manufacturer

Laser​Manufacturer

LO

1

DICONDE

(0014,603A)

Laser Model Number

Laser​Model​Number

LO

1

DICONDE

(0014,603B)

Laser Type Description

Laser​Type​Description

ST

1

DICONDE

(0014,603C)

Forced Gas Heating Sequence

Forced​Gas​Heating​Sequence

SQ

1

DICONDE

(0014,603D)

Gas Used for Heating/Cooling Part

Gas​Used​For​Heating​Cooling​Part

LO

1

DICONDE

(0014,603E)

Vibration/Sonic Heating Sequence

Vibration​Sonic​Heating​Sequence

SQ

1

DICONDE

(0014,603F)

Probe Manufacturer

Probe​Manufacturer

LO

1

DICONDE

(0014,6040)

Probe Model Number

Probe​Model​Number

LO

1

DICONDE

(0014,6041)

Aperture Size

Aperture​Size

DS

1

DICONDE

(0014,6042)

Probe Resonant Frequency

Probe​Resonant​Frequency

DS

1

DICONDE

(0014,6043)

Heat Source Description

Heat​Source​Description

UT

1

DICONDE

(0014,6044)

Surface Preparation with Optical Coating

Surface​Preparation​With​Optical​Coating

CS

1

DICONDE

(0014,6045)

Optical Coating Type

Optical​Coating​Type

ST

1

DICONDE

(0014,6046)

Thermal Conductivity of Exposed Surface

Thermal​Conductivity​Of​Exposed​Surface

DS

1

DICONDE

(0014,6047)

Material Density

Material​Density

DS

1

DICONDE

(0014,6048)

Specific Heat of Inspection Surface

Specific​Heat​Of​Inspection​Surface

DS

1

DICONDE

(0014,6049)

Emissivity of Inspection Surface

Emissivity​Of​Inspection​Surface

DS

1

DICONDE

(0014,604A)

Electromagnetic Classification of Inspection Surface

Electromagnetic​Classification​Of​Inspection​Surface

CS

1-n

DICONDE

(0014,604C)

Moving Window Size

Moving​Window​Size

DS

1

DICONDE

(0014,604D)

Moving Window Type

Moving​Window​Type

CS

1

DICONDE

(0014,604E)

Moving Window Weights

Moving​Window​Weights

DS

1-n

DICONDE

(0014,604F)

Moving Window Pitch

Moving​Window​Pitch

DS

1

DICONDE

(0014,6050)

Moving Window Padding Scheme

Moving​Window​Padding​Scheme

CS

1

DICONDE

(0014,6051)

Moving Window Padding Length

Moving​Window​Padding​Length

DS

1

DICONDE

(0014,6052)

Spatial Filtering Parameters Sequence

Spatial​Filtering​Parameters​Sequence

SQ

1

DICONDE

(0014,6053)

Spatial Filtering Scheme

Spatial​Filtering​Scheme

CS

1

DICONDE

(0014,6056)

Horizontal Moving Window Size

Horizontal​Moving​Window​Size

DS

1

DICONDE

(0014,6057)

Vertical Moving Window Size

Vertical​Moving​Window​Size

DS

1

DICONDE

(0014,6059)

Polynomial Fitting Sequence

Polynomial​Fitting​Sequence

SQ

1

DICONDE

(0014,605A)

Fitting Data Type

Fitting​Data​Type

CS

1-n

DICONDE

(0014,605B)

Operation on Time Axis Before Fitting

Operation​On​Time​Axis​Before​Fitting

CS

1

DICONDE

(0014,605C)

Operation on Pixel Intensity Before Fitting

Operation​On​Pixel​Intensity​Before​Fitting

CS

1

DICONDE

(0014,605D)

Order of Polynomial

Order​Of​Polynomial

DS

1

DICONDE

(0014,605E)

Independent Variable for Polynomial Fit

Independent​Variable​For​Polynomial​Fit

CS

1

DICONDE

(0014,605F)

Polynomial​Coefficients

Polynomial​Coefficients

DS

1-n

DICONDE

(0014,6060)

Thermography Pixel Data Unit

Thermography​Pixel​Data​Unit

CS

1

DICONDE

(0016,0001)

White Point

White​Point​

DS

1

(0016,0002)

Primary Chromaticities

Primary​Chromaticities

DS

3

(0016,0003)

Battery Level

Battery​Level

UT

1

(0016,0004)

Exposure Time in Seconds

Exposure​Time​In​Seconds

DS

1

(0016,0005)

F-Number

FNumber

DS

1

(0016,0006)

OECF Rows

OECF​Rows

IS

1

(0016,0007)

OECF Columns

OECF​Columns

IS

1

(0016,0008)

OECF Column Names

OECF​Column​Names

UC

1-n

(0016,0009)

OECF Values

OECF​Values

DS

1-n

(0016,000A)

Spatial Frequency Response Rows

Spatial​Frequency​Response​Rows

IS

1

(0016,000B)

Spatial Frequency Response Columns

Spatial​Frequency​Response​Columns

IS

1

(0016,000C)

Spatial Frequency Response Column Names

Spatial​Frequency​Response​Column​Names

UC

1-n

(0016,000D)

Spatial Frequency Response Values

Spatial​Frequency​Response​Values

DS

1-n

(0016,000E)

Color Filter Array Pattern Rows

Color​Filter​Array​Pattern​Rows

IS

1

(0016,000F)

Color Filter Array Pattern Columns

Color​Filter​Array​Pattern​Columns

IS

1

(0016,0010)

Color Filter Array Pattern Values

Color​Filter​Array​Pattern​Values

DS

1-n

(0016,0011)

Flash Firing Status

Flash​Firing​Status

US

1

(0016,0012)

Flash Return Status

Flash​Return​Status

US

1

(0016,0013)

Flash Mode

Flash​Mode

US

1

(0016,0014)

Flash Function Present

Flash​Function​Present

US

1

(0016,0015)

Flash Red Eye Mode

Flash​Red​Eye​Mode

US

1

(0016,0016)

Exposure Program

Exposure​Program

US

1

(0016,0017)

Spectral Sensitivity

Spectral​Sensitivity

UT

1

(0016,0018)

Photographic Sensitivity

Photographic​Sensitivity

IS

1

(0016,0019)

Self Timer Mode

Self​Timer​Mode

IS

1

(0016,001A)

Sensitivity Type

Sensitivity​Type

US

1

(0016,001B)

Standard Output Sensitivity

Standard​Output​Sensitivity

IS

1

(0016,001C)

Recommended Exposure Index

Recommended​Exposure​Index

IS

1

(0016,001D)

ISO Speed​

ISO​Speed​

IS

1

(0016,001E)

ISO Speed​​ Latitude yyy

ISO​Speed​Latitude​yyy

IS

1

(0016,001F)

ISO Speed​​ Latitude zzz

ISO​Speed​Latitude​zzz

IS

1

(0016,0020)

EXIF Version

EXIF​Version

UT

1

(0016,0021)

Shutter Speed Value

Shutter​Speed​Value

DS

1

(0016,0022)

Aperture Value

Aperture​Value

DS

1

(0016,0023)

Brightness Value

Brightness​Value

DS

1

(0016,0024)

Exposure Bias Value

Exposure​Bias​Value

DS

1

(0016,0025)

Max Aperture Value

Max​Aperture​Value

DS

1

(0016,0026)

Subject Distance

Subject​Distance

DS

1

(0016,0027)

Metering Mode

Metering​Mode

US

1

(0016,0028)

Light Source

Light​Source

US

1

(0016,0029)

Focal Length

Focal​Length

DS

1

(0016,002A)

Subject Area

Subject​Area

IS

2-4

(0016,002B)

Maker Note

Maker​Note

OB

1

(0016,0030)

Temperature

Temperature

DS

1

(0016,0031)

Humidity

Humidity

DS

1

(0016,0032)

Pressure

Pressure

DS

1

(0016,0033)

Water Depth

Water​Depth

DS

1

(0016,0034)

Acceleration

Acceleration

DS

1

(0016,0035)

Camera Elevation Angle

Camera​Elevation​Angle

DS

1

(0016,0036)

Flash Energy

Flash​Energy

DS

1-2

(0016,0037)

Subject Location

Subject​Location

IS

2

(0016,0038)

Photographic Exposure Index

Photographic​Exposure​Index

DS

1

(0016,0039)

Sensing Method

Sensing​Method

US

1

(0016,003A)

File Source

File​Source

US

1

(0016,003B)

Scene Type

Scene​Type

US

1

(0016,0041)

Custom Rendered

Custom​Rendered

US

1

(0016,0042)

Exposure Mode

Exposure​Mode

US

1

(0016,0043)

White Balance

White​Balance

US

1

(0016,0044)

Digital Zoom Ratio

Digital​Zoom​Ratio

DS

1

(0016,0045)

Focal Length In 35mm​ Film

Focal​Length​In​35mm​Film

IS

1

(0016,0046)

Scene Capture Type

Scene​Capture​Type

US

1

(0016,0047)

Gain Control

Gain​Control

US

1

(0016,0048)

Contrast

Contrast

US

1

(0016,0049)

Saturation

Saturation

US

1

(0016,004A)

Sharpness

Sharpness

US

1

(0016,004B)

Device Setting Description

Device​Setting​Description

OB

1

(0016,004C)

Subject Distance Range

Subject​Distance​Range

US

1

(0016,004D)

Camera Owner Name

Camera​Owner​Name

UT

1

(0016,004E)

Lens Specification

Lens​Specification

DS

4

(0016,004F)

Lens Make

Lens​Make

UT

1

(0016,0050)

Lens Model

Lens​Model

UT

1

(0016,0051)

Lens Serial Number

Lens​Serial​Number

UT

1

(0016,0061)

Interoperability Index

Interoperability​Index

CS

1

(0016,0062)

Interoperability Version

Interoperability​Version

OB

1

(0016,0070)

GPS Version ID

GPS​Version​ID

OB

1

(0016,0071)

GPS Latitude​ Ref

GPS​Latitude​Ref

CS

1

(0016,0072)

GPS Latitude​

GPS​Latitude​

DS

3

(0016,0073)

GPS Longitude Ref

GPS​Longitude​​Ref

CS

1

(0016,0074)

GPS Longitude

GPS​Longitude

DS

3

(0016,0075)

GPS Altitude​ Ref

GPS​Altitude​Ref

US

1

(0016,0076)

GPS Altitude​

GPS​Altitude​

DS

1

(0016,0077)

GPS Time​ Stamp

GPS​Time​Stamp

DT

1

(0016,0078)

GPS Satellites

GPS​Satellites

UT

1

(0016,0079)

GPS Status

GPS​Status

CS

1

(0016,007A)

GPS Measure ​Mode

GPS​Measure​​Mode

CS

1

(0016,007B)

GPS DOP

GPS​DOP

DS

1

(0016,007C)

GPS Speed​ Ref

GPS​Speed​Ref

CS

1

(0016,007D)

GPS Speed​

GPS​Speed​

DS

1

(0016,007E)

GPS Track ​Ref

GPS​Track​​Ref

CS

1

(0016,007F)

GPS Track

GPS​Track

DS

1

(0016,0080)

GPS Img​ Direction Ref

GPS​Img​Direction​Ref

CS

1

(0016,0081)

GPS Img​ Direction

GPS​Img​Direction

DS

1

(0016,0082)

GPS Map​ Datum

GPS​Map​​Datum

UT

1

(0016,0083)

GPS Dest ​Latitude Ref

GPS​Dest​Latitude​Ref

CS

1

(0016,0084)

GPS Dest​ Latitude

GPS​Dest​Latitude

DS

3

(0016,0085)

GPS Dest​ Longitude Ref

GPS​Dest​Longitude​Ref

CS

1

(0016,0086)

GPS Dest ​Longitude

GPS​Dest​​Longitude

DS

3

(0016,0087)

GPS Dest ​Bearing Ref

GPS​Dest​Bearing​Ref

CS

1

(0016,0088)

GPS Dest ​Bearing

GPS​Dest​Bearing

DS

1

(0016,0089)

GPS Dest ​Distance Ref

GPS​Dest​Distance​Ref

CS

1

(0016,008A)

GPS Dest​ Distance

GPS​Dest​Distance

DS

1

(0016,008B)

GPS Processing​ Method

GPS​Processing​Method

OB

1

(0016,008C)

GPS Area ​Information

GPS​Area​Information

OB

1

(0016,008D)

GPS Date​ Stamp

GPS​Date​Stamp

DT

1

(0016,008E)

GPS Differential

GPS​Differential

IS

1

(0016,1001)

Light Source Polarization

Light​Source​Polarization

CS

1

(0016,1002)

Emitter Color Temperature

Emitter​Color​Temperature

DS

1

(0016,1003)

Contact Method

Contact​Method

CS

1

(0016,1004)

Immersion Media

Immersion​Media

CS

1-n

(0016,1005)

Optical Magnification Factor

Optical​Magnification​Factor

DS

1

(0018,0010)

Contrast/Bolus Agent

Contrast​Bolus​Agent

LO

1

(0018,0012)

Contrast/Bolus Agent Sequence

Contrast​Bolus​Agent​Sequence

SQ

1

(0018,0013)

Contrast/Bolus T1 Relaxivity

Contrast​Bolus​​T1​Relaxivity

FL

1

(0018,0014)

Contrast/Bolus Administration Route Sequence

Contrast​Bolus​Administration​Route​Sequence

SQ

1

(0018,0015)

Body Part Examined

Body​Part​Examined

CS

1

(0018,0020)

Scanning Sequence

Scanning​Sequence

CS

1-n

(0018,0021)

Sequence Variant

Sequence​Variant

CS

1-n

(0018,0022)

Scan Options

Scan​Options

CS

1-n

(0018,0023)

MR Acquisition Type

MR​Acquisition​Type

CS

1

(0018,0024)

Sequence Name

Sequence​Name

SH

1

(0018,0025)

Angio Flag

Angio​Flag

CS

1

(0018,0026)

Intervention Drug Information Sequence

Intervention​Drug​Information​Sequence

SQ

1

(0018,0027)

Intervention Drug Stop Time

Intervention​Drug​Stop​Time

TM

1

(0018,0028)

Intervention Drug Dose

Intervention​Drug​Dose

DS

1

(0018,0029)

Intervention Drug Code Sequence

Intervention​Drug​Code​Sequence

SQ

1

(0018,002A)

Additional Drug Sequence

Additional​Drug​Sequence

SQ

1

(0018,0030)

Radionuclide

Radionuclide

LO

1-n

RET (1993)

(0018,0031)

Radiopharmaceutical

Radio​pharmaceutical

LO

1

(0018,0032)

Energy Window Centerline

Energy​Window​Centerline

DS

1

RET (1993)

(0018,0033)

Energy Window Total Width

Energy​Window​Total​Width

DS

1-n

RET (1993)

(0018,0034)

Intervention Drug Name

Intervention​Drug​Name

LO

1

(0018,0035)

Intervention Drug Start Time

Intervention​Drug​Start​Time

TM

1

(0018,0036)

Intervention Sequence

Intervention​Sequence

SQ

1

(0018,0037)

Therapy Type

Therapy​Type

CS

1

RET (2001)

(0018,0038)

Intervention Status

Intervention​Status

CS

1

(0018,0039)

Therapy Description

Therapy​Description

CS

1

RET (2003)

(0018,003A)

Intervention Description

Intervention​Description

ST

1

(0018,0040)

Cine Rate

Cine​Rate

IS

1

(0018,0042)

Initial Cine Run State

Initial​Cine​Run​State

CS

1

(0018,0050)

Slice Thickness

Slice​Thickness

DS

1

(0018,0060)

KVP

KVP

DS

1

(0018,0061)

DS

1

RET (2015c)

(0018,0070)

Counts Accumulated

Counts​Accumulated

IS

1

(0018,0071)

Acquisition Termination Condition

Acquisition​Termination​Condition

CS

1

(0018,0072)

Effective Duration

Effective​Duration

DS

1

(0018,0073)

Acquisition Start Condition

Acquisition​Start​Condition

CS

1

(0018,0074)

Acquisition Start Condition Data

Acquisition​Start​Condition​Data

IS

1

(0018,0075)

Acquisition Termination Condition Data

Acquisition​Termination​Condition​Data

IS

1

(0018,0080)

Repetition Time

Repetition​Time

DS

1

(0018,0081)

Echo Time

Echo​Time

DS

1

(0018,0082)

Inversion Time

Inversion​Time

DS

1

(0018,0083)

Number of Averages

Number​Of​Averages

DS

1

(0018,0084)

Imaging Frequency

Imaging​Frequency

DS

1

(0018,0085)

Imaged Nucleus

Imaged​Nucleus

SH

1

(0018,0086)

Echo Number(s)

Echo​Numbers

IS

1-n

(0018,0087)

Magnetic Field Strength

Magnetic​Field​Strength

DS

1

(0018,0088)

Spacing Between Slices

Spacing​Between​Slices

DS

1

(0018,0089)

Number of Phase Encoding Steps

Number​Of​Phase​Encoding​Steps

IS

1

(0018,0090)

Data Collection Diameter

Data​Collection​Diameter

DS

1

(0018,0091)

Echo Train Length

Echo​Train​Length

IS

1

(0018,0093)

Percent Sampling

Percent​Sampling

DS

1

(0018,0094)

Percent Phase Field of View

Percent​Phase​Field​Of​View

DS

1

(0018,0095)

Pixel Bandwidth

Pixel​Bandwidth

DS

1

(0018,1000)

Device Serial Number

Device​Serial​Number

LO

1

(0018,1002)

Device UID

Device​UID

UI

1

(0018,1003)

Device ID

Device​ID

LO

1

(0018,1004)

Plate ID

Plate​ID

LO

1

(0018,1005)

Generator ID

Generator​ID

LO

1

(0018,1006)

Grid ID

Grid​ID

LO

1

(0018,1007)

Cassette ID

Cassette​ID

LO

1

(0018,1008)

Gantry ID

Gantry​ID

LO

1

(0018,1009)

Unique Device Identifier

Unique​Device​Identifier

UT

1

(0018,100A)

UDI Sequence

UDI​Sequence

SQ

1

(0018,100B)

Manufacturer's Device Class UID

Manufacturer​Device​Class​UID

UI

1-n

(0018,1010)

Secondary Capture Device ID

Secondary​Capture​Device​ID

LO

1

(0018,1011)

Hardcopy Creation Device ID

Hardcopy​Creation​Device​ID

LO

1

RET (2007)

(0018,1012)

Date of Secondary Capture

Date​Of​Secondary​Capture

DA

1

(0018,1014)

Time of Secondary Capture

Time​Of​Secondary​Capture

TM

1

(0018,1016)

Secondary Capture Device Manufacturer

Secondary​Capture​Device​Manufacturer

LO

1

(0018,1017)

Hardcopy Device Manufacturer

Hardcopy​Device​Manufacturer

LO

1

RET (2007)

(0018,1018)

Secondary Capture Device Manufacturer's Model Name

Secondary​Capture​Device​Manufacturer​Model​Name

LO

1

(0018,1019)

Secondary Capture Device Software Versions

Secondary​Capture​Device​Software​Versions

LO

1-n

(0018,101A)

Hardcopy Device Software Version

Hardcopy​Device​Software​Version

LO

1-n

RET (2007)

(0018,101B)

Hardcopy Device Manufacturer's Model Name

Hardcopy​Device​Manufacturer​Model​Name

LO

1

RET (2007)

(0018,1020)

Software Versions

Software​Versions

LO

1-n

(0018,1022)

Video Image Format Acquired

Video​Image​Format​Acquired

SH

1

(0018,1023)

Digital Image Format Acquired

Digital​Image​Format​Acquired

LO

1

(0018,1030)

Protocol Name

Protocol​Name

LO

1

(0018,1040)

Contrast/Bolus Route

Contrast​Bolus​Route

LO

1

(0018,1041)

Contrast/Bolus Volume

Contrast​Bolus​Volume

DS

1

(0018,1042)

Contrast/Bolus Start Time

Contrast​Bolus​Start​Time

TM

1

(0018,1043)

Contrast/Bolus Stop Time

Contrast​Bolus​Stop​Time

TM

1

(0018,1044)

Contrast/Bolus Total Dose

Contrast​Bolus​Total​Dose

DS

1

(0018,1045)

Syringe Counts

Syringe​Counts

IS

1

(0018,1046)

Contrast Flow Rate

Contrast​Flow​Rate

DS

1-n

(0018,1047)

Contrast Flow Duration

Contrast​Flow​Duration

DS

1-n

(0018,1048)

Contrast/Bolus Ingredient

Contrast​Bolus​Ingredient

CS

1

(0018,1049)

Contrast/Bolus Ingredient Concentration

Contrast​Bolus​Ingredient​Concentration

DS

1

(0018,1050)

Spatial Resolution

Spatial​Resolution

DS

1

(0018,1060)

Trigger Time

Trigger​Time

DS

1

(0018,1061)

Trigger Source or Type

Trigger​Source​Or​Type

LO

1

(0018,1062)

Nominal Interval

Nominal​Interval

IS

1

(0018,1063)

Frame Time

Frame​Time

DS

1

(0018,1064)

Cardiac Framing Type

Cardiac​Framing​Type

LO

1

(0018,1065)

Frame Time Vector

Frame​Time​Vector

DS

1-n

(0018,1066)

Frame Delay

Frame​Delay

DS

1

(0018,1067)

Image Trigger Delay

Image​Trigger​Delay

DS

1

(0018,1068)

Multiplex Group Time Offset

Multiplex​Group​Time​Offset

DS

1

(0018,1069)

Trigger Time Offset

Trigger​Time​Offset

DS

1

(0018,106A)

Synchronization Trigger

Synchronization​Trigger

CS

1

(0018,106C)

Synchronization Channel

Synchronization​Channel

US

2

(0018,106E)

Trigger Sample Position

Trigger​Sample​Position

UL

1

(0018,1070)

Radiopharmaceutical Route

Radio​pharmaceutical​Route

LO

1

(0018,1071)

Radiopharmaceutical Volume

Radio​pharmaceutical​Volume

DS

1

(0018,1072)

Radiopharmaceutical Start Time

Radio​pharmaceutical​Start​Time

TM

1

(0018,1073)

Radiopharmaceutical Stop Time

Radio​pharmaceutical​Stop​Time

TM

1

(0018,1074)

Radionuclide Total Dose

Radionuclide​Total​Dose

DS

1

(0018,1075)

Radionuclide Half Life

Radionuclide​Half​Life

DS

1

(0018,1076)

Radionuclide Positron Fraction

Radionuclide​Positron​Fraction

DS

1

(0018,1077)

Radiopharmaceutical Specific Activity

Radio​pharmaceutical​Specific​Activity

DS

1

(0018,1078)

Radiopharmaceutical Start DateTime

Radio​pharmaceutical​Start​Date​Time

DT

1

(0018,1079)

Radiopharmaceutical Stop DateTime

Radio​pharmaceutical​Stop​Date​Time

DT

1

(0018,1080)

Beat Rejection Flag

Beat​Rejection​Flag

CS

1

(0018,1081)

Low R-R Value

Low​RR​Value

IS

1

(0018,1082)

High R-R Value

High​RR​Value

IS

1

(0018,1083)

Intervals Acquired

Intervals​Acquired

IS

1

(0018,1084)

Intervals Rejected

Intervals​Rejected

IS

1

(0018,1085)

PVC Rejection

PVC​Rejection

LO

1

(0018,1086)

Skip Beats

Skip​Beats

IS

1

(0018,1088)

Heart Rate

Heart​Rate

IS

1

(0018,1090)

Cardiac Number of Images

Cardiac​Number​Of​Images

IS

1

(0018,1094)

Trigger Window

Trigger​Window

IS

1

(0018,1100)

Reconstruction Diameter

Reconstruction​Diameter

DS

1

(0018,1110)

Distance Source to Detector

Distance​Source​To​Detector

DS

1

(0018,1111)

Distance Source to Patient

Distance​Source​To​Patient

DS

1

(0018,1114)

Estimated Radiographic Magnification Factor

Estimated​Radiographic​Magnification​Factor

DS

1

(0018,1120)

Gantry/Detector Tilt

Gantry​Detector​Tilt

DS

1

(0018,1121)

Gantry/Detector Slew

Gantry​Detector​Slew

DS

1

(0018,1130)

Table Height

Table​Height

DS

1

(0018,1131)

Table Traverse

Table​Traverse

DS

1

(0018,1134)

Table Motion

Table​Motion

CS

1

(0018,1135)

Table Vertical Increment

Table​Vertical​Increment

DS

1-n

(0018,1136)

Table Lateral Increment

Table​Lateral​Increment

DS

1-n

(0018,1137)

Table Longitudinal Increment

Table​Longitudinal​Increment

DS

1-n

(0018,1138)

Table Angle

Table​Angle

DS

1

(0018,113A)

Table Type

Table​Type

CS

1

(0018,1140)

Rotation Direction

Rotation​Direction

CS

1

(0018,1141)

Angular Position

Angular​Position

DS

1

RET (2007)

(0018,1142)

Radial Position

Radial​Position

DS

1-n

(0018,1143)

Scan Arc

Scan​Arc

DS

1

(0018,1144)

Angular Step

Angular​Step

DS

1

(0018,1145)

Center of Rotation Offset

Center​Of​Rotation​Offset

DS

1

(0018,1146)

Rotation Offset

Rotation​Offset

DS

1-n

RET (1993)

(0018,1147)

Field of View Shape

Field​Of​View​Shape

CS

1

(0018,1149)

Field of View Dimension(s)

Field​Of​View​Dimensions

IS

1-2

(0018,1150)

Exposure Time

Exposure​Time

IS

1

(0018,1151)

X-Ray Tube Current

X​Ray​Tube​Current

IS

1

(0018,1152)

Exposure

Exposure

IS

1

(0018,1153)

Exposure in µAs

Exposure​Inu​As

IS

1

(0018,1154)

Average Pulse Width

Average​Pulse​Width

DS

1

(0018,1155)

Radiation Setting

Radiation​Setting

CS

1

(0018,1156)

Rectification Type

Rectification​Type

CS

1

(0018,115A)

Radiation Mode

Radiation​Mode

CS

1

(0018,115E)

Image and Fluoroscopy Area Dose Product

Image​And​Fluoroscopy​Area​Dose​Product

DS

1

(0018,1160)

Filter Type

Filter​Type

SH

1

(0018,1161)

Type of Filters

Type​Of​Filters

LO

1-n

(0018,1162)

Intensifier Size

Intensifier​Size

DS

1

(0018,1164)

Imager Pixel Spacing

Imager​Pixel​Spacing

DS

2

(0018,1166)

Grid

Grid

CS

1-n

(0018,1170)

Generator Power

Generator​Power

IS

1

(0018,1180)

Collimator/grid Name

Collimator​Grid​Name

SH

1

(0018,1181)

Collimator Type

Collimator​Type

CS

1

(0018,1182)

Focal Distance

Focal​Distance

IS

1-2

(0018,1183)

X Focus Center

X​Focus​Center

DS

1-2

(0018,1184)

Y Focus Center

Y​Focus​Center

DS

1-2

(0018,1190)

Focal Spot(s)

Focal​Spots

DS

1-n

(0018,1191)

Anode Target Material

Anode​Target​Material

CS

1

(0018,11A0)

Body Part Thickness

Body​Part​Thickness

DS

1

(0018,11A2)

Compression Force

Compression​Force

DS

1

(0018,11A3)

Compression Pressure

Compression​​Pressure

DS

1

(0018,11A4)

Paddle Description

Paddle​Description

LO

1

(0018,11A5)

Compression Contact Area

Compression​​Contact​​Area

DS

1

(0018,11B0)

Acquisition Mode

Acquisition​Mode

LO

1

(0018,11B1)

Dose Mode Name

Dose​Mode​Name

LO

1

(0018,11B2)

Acquired Subtraction Mask Flag

Acquired​Subtraction​Mask​Flag

CS

1

(0018,11B3)

Fluoroscopy Persistence Flag

Fluoroscopy​Persistence​Flag

CS

1

(0018,11B4)

Fluoroscopy Last Image Hold Persistence Flag

Fluoroscopy​Last​Image​Hold​Persistence​Flag

CS

1

(0018,11B5)

Upper Limit Number Of Persistent Fluoroscopy Frames

Upper​Limit​Number​Of​Persistent​Fluoroscopy​Frames

IS

1

(0018,11B6)

Contrast/Bolus Auto Injection Trigger Flag

Contrast​Bolus​Auto​Injection​Trigger​Flag

CS

1

(0018,11B7)

Contrast/Bolus Injection Delay

Contrast​Bolus​Injection​Delay

FD

1

(0018,11B8)

XA Acquisition Phase Details Sequence

XA​Acquisition​Phase​Details​Sequence

SQ

1

(0018,11B9)

XA Acquisition Frame Rate

XA​Acquisition​Frame​Rate

FD

1

(0018,11BA)

XA Plane Details Sequence

XA​Plane​Details​Sequence

SQ

1

(0018,11BB)

Acquisition Field of View Label

Acquisition​Field​Of​View​Label

LO

1

(0018,11BC)

X-Ray Filter Details Sequence

XRay​Filter​Details​Sequence

SQ

1

(0018,11BD)

XA Acquisition Duration

XA​Acquisition​Duration

FD

1

(0018,11BE)

Reconstruction Pipeline Type

Reconstruction​Pipeline​Type

CS

1

(0018,11BF)

Image Filter Details Sequence

Image​Filter​Details​Sequence

SQ

1

(0018,11C0)

Applied Mask Subtraction Flag

Applied​Mask​Subtraction​Flag

CS

1

(0018,11C1)

Requested Series Description Code Sequence

Requested​Series​Description​Code​Sequence

SQ

1

(0018,1200)

Date of Last Calibration

Date​Of​Last​Calibration

DA

1-n

(0018,1201)

Time of Last Calibration

Time​Of​Last​Calibration

TM

1-n

(0018,1202)

DateTime of Last Calibration

Date​Time​Of​Last​Calibration

DT

1

(0018,1203)

Calibration DateTime

Calibration​Date​Time

DT

1

(0018,1204)

Date of Manufacture

Date​Of​Manufacture

DA

1

(0018,1205)

Date of Installation

Date​Of​Installation

DA

1

(0018,1210)

Convolution Kernel

Convolution​Kernel

SH

1-n

(0018,1240)

Upper/Lower Pixel Values

Upper​Lower​Pixel​Values

IS

1-n

RET

(0018,1242)

Actual Frame Duration

Actual​Frame​Duration

IS

1

(0018,1243)

Count Rate

Count​Rate

IS

1

(0018,1244)

Preferred Playback Sequencing

Preferred​Playback​Sequencing

US

1

(0018,1250)

Receive Coil Name

Receive​Coil​Name

SH

1

(0018,1251)

Transmit Coil Name

Transmit​Coil​Name

SH

1

(0018,1260)

Plate Type

Plate​Type

SH

1

(0018,1261)

Phosphor Type

Phosphor​Type

LO

1

(0018,1271)

Water Equivalent Diameter

Water​Equivalent​Diameter

FD

1

(0018,1272)

Water Equivalent Diameter Calculation Method Code Sequence

Water​Equivalent​Diameter​Calculation​Method​Code​Sequence

SQ

1

(0018,1300)

Scan Velocity

Scan​Velocity

DS

1

(0018,1301)

Whole Body Technique

Whole​Body​Technique

CS

1-n

(0018,1302)

Scan Length

Scan​Length

IS

1

(0018,1310)

Acquisition Matrix

Acquisition​Matrix

US

4

(0018,1312)

In-plane Phase Encoding Direction

In​Plane​Phase​Encoding​Direction

CS

1

(0018,1314)

Flip Angle

Flip​Angle

DS

1

(0018,1315)

Variable Flip Angle Flag

Variable​Flip​Angle​Flag

CS

1

(0018,1316)

SAR

SAR

DS

1

(0018,1318)

dB/dt

d​Bdt

DS

1

(0018,1320)

B1rms

B1rms

FL

1

(0018,1400)

Acquisition Device Processing Description

Acquisition​Device​Processing​Description

LO

1

(0018,1401)

Acquisition Device Processing Code

Acquisition​Device​Processing​Code

LO

1

(0018,1402)

Cassette Orientation

Cassette​Orientation

CS

1

(0018,1403)

Cassette Size

Cassette​Size

CS

1

(0018,1404)

Exposures on Plate

Exposures​On​Plate

US

1

(0018,1405)

Relative X-Ray Exposure

Relative​X​Ray​Exposure

IS

1

(0018,1411)

Exposure Index

Exposure​Index

DS

1

(0018,1412)

Target Exposure Index

Target​Exposure​Index

DS

1

(0018,1413)

Deviation Index

Deviation​Index

DS

1

(0018,1450)

Column Angulation

Column​Angulation

DS

1

(0018,1460)

Tomo Layer Height

Tomo​Layer​Height

DS

1

(0018,1470)

Tomo Angle

Tomo​Angle

DS

1

(0018,1480)

Tomo Time

Tomo​Time

DS

1

(0018,1490)

Tomo Type

Tomo​Type

CS

1

(0018,1491)

Tomo Class

Tomo​Class

CS

1

(0018,1495)

Number of Tomosynthesis Source Images

Number​Of​Tomosynthesis​Source​Images

IS

1

(0018,1500)

Positioner Motion

Positioner​Motion

CS

1

(0018,1508)

Positioner Type

Positioner​Type

CS

1

(0018,1510)

Positioner Primary Angle

Positioner​Primary​Angle

DS

1

(0018,1511)

Positioner Secondary Angle

Positioner​Secondary​Angle

DS

1

(0018,1520)

Positioner Primary Angle Increment

Positioner​Primary​Angle​Increment

DS

1-n

(0018,1521)

Positioner Secondary Angle Increment

Positioner​Secondary​Angle​Increment

DS

1-n

(0018,1530)

Detector Primary Angle

Detector​Primary​Angle

DS

1

(0018,1531)

Detector Secondary Angle

Detector​Secondary​Angle

DS

1

(0018,1600)

Shutter Shape

Shutter​Shape

CS

1-3

(0018,1602)

Shutter Left Vertical Edge

Shutter​Left​Vertical​Edge

IS

1

(0018,1604)

Shutter Right Vertical Edge

Shutter​Right​Vertical​Edge

IS

1

(0018,1606)

Shutter Upper Horizontal Edge

Shutter​Upper​Horizontal​Edge

IS

1

(0018,1608)

Shutter Lower Horizontal Edge

Shutter​Lower​Horizontal​Edge

IS

1

(0018,1610)

Center of Circular Shutter

Center​Of​Circular​Shutter

IS

2

(0018,1612)

Radius of Circular Shutter

Radius​Of​Circular​Shutter

IS

1

(0018,1620)

Vertices of the Polygonal Shutter

Vertices​Of​The​Polygonal​Shutter

IS

2-2n

(0018,1622)

Shutter Presentation Value

Shutter​Presentation​Value

US

1

(0018,1623)

Shutter Overlay Group

Shutter​Overlay​Group

US

1

(0018,1624)

Shutter Presentation Color CIELab Value

Shutter​Presentation​Color​CIE​Lab​Value

US

3

(0018,1630)

Outline Shape Type

Outline​Shape​Type

CS

1

(0018,1631)

Outline Left Vertical Edge

Outline​Left​Vertical​Edge

FD

1

(0018,1632)

Outline Right Vertical Edge

Outline​Right​Vertical​Edge

FD

1

(0018,1633)

Outline Upper Horizontal Edge

Outline​Upper​Horizontal​Edge

FD

1

(0018,1634)

Outline Lower Horizontal Edge

Outline​Lower​Horizontal​Edge

FD

1

(0018,1635)

Center of Circular Outline

Center​Of​Circular​Outline

FD

2

(0018,1636)

Diameter of Circular Outline

Diameter​Of​Circular​Outline

FD

1

(0018,1637)

Number of Polygonal Vertices

Number​Of​Polygonal​Vertices

UL

1

(0018,1638)

Vertices of the Polygonal Outline

Vertices​Of​The​Polygonal​Outline

OF

1

(0018,1700)

Collimator Shape

Collimator​Shape

CS

1-3

(0018,1702)

Collimator Left Vertical Edge

Collimator​Left​Vertical​Edge

IS

1

(0018,1704)

Collimator Right Vertical Edge

Collimator​Right​Vertical​Edge

IS

1

(0018,1706)

Collimator Upper Horizontal Edge

Collimator​Upper​Horizontal​Edge

IS

1

(0018,1708)

Collimator Lower Horizontal Edge

Collimator​Lower​Horizontal​Edge

IS

1

(0018,1710)

Center of Circular Collimator

Center​Of​Circular​Collimator

IS

2

(0018,1712)

Radius of Circular Collimator

Radius​Of​Circular​Collimator

IS

1

(0018,1720)

Vertices of the Polygonal Collimator

Vertices​Of​The​Polygonal​Collimator

IS

2-2n

(0018,1800)

Acquisition Time Synchronized

Acquisition​Time​Synchronized

CS

1

(0018,1801)

Time Source

Time​Source

SH

1

(0018,1802)

Time Distribution Protocol

Time​Distribution​Protocol

CS

1

(0018,1803)

NTP Source Address

NTP​Source​Address

LO

1

(0018,2001)

Page Number Vector

Page​Number​Vector

IS

1-n

(0018,2002)

Frame Label Vector

Frame​Label​Vector

SH

1-n

(0018,2003)

Frame Primary Angle Vector

Frame​Primary​Angle​Vector

DS

1-n

(0018,2004)

Frame Secondary Angle Vector

Frame​Secondary​Angle​Vector

DS

1-n

(0018,2005)

Slice Location Vector

Slice​Location​Vector

DS

1-n

(0018,2006)

Display Window Label Vector

Display​Window​Label​Vector

SH

1-n

(0018,2010)

Nominal Scanned Pixel Spacing

Nominal​Scanned​Pixel​Spacing

DS

2

(0018,2020)

Digitizing Device Transport Direction

Digitizing​Device​Transport​Direction

CS

1

(0018,2030)

Rotation of Scanned Film

Rotation​Of​Scanned​Film

DS

1

(0018,2041)

Biopsy Target Sequence

Biopsy​Target​Sequence

SQ

1

(0018,2042)

Target UID

Target​UID

UI

1

(0018,2043)

Localizing Cursor Position

Localizing​Cursor​Position

FL

2

(0018,2044)

Calculated Target Position

Calculated​Target​Position

FL

3

(0018,2045)

Target Label

Target​Label

SH

1

(0018,2046)

Displayed Z Value

Displayed​Z​Value

FL

1

(0018,3100)

IVUS Acquisition

IVUS​Acquisition

CS

1

(0018,3101)

IVUS Pullback Rate

IVUS​Pullback​Rate

DS

1

(0018,3102)

IVUS Gated Rate

IVUS​Gated​Rate

DS

1

(0018,3103)

IVUS Pullback Start Frame Number

IVUS​Pullback​Start​Frame​Number

IS

1

(0018,3104)

IVUS Pullback Stop Frame Number

IVUS​Pullback​Stop​Frame​Number

IS

1

(0018,3105)

Lesion Number

Lesion​Number

IS

1-n

(0018,4000)

Acquisition Comments

Acquisition​Comments

LT

1

RET

(0018,5000)

Output Power

Output​Power

SH

1-n

(0018,5010)

Transducer Data

Transducer​Data

LO

1-n

(0018,5011)

Transducer Identification Sequence

Transducer​Identification​Sequence

SQ

1

(0018,5012)

Focus Depth

Focus​Depth

DS

1

(0018,5020)

Processing Function

Processing​Function

LO

1

(0018,5021)

Postprocessing Function

Postprocessing​Function

LO

1

RET (2007)

(0018,5022)

Mechanical Index

Mechanical​Index

DS

1

(0018,5024)

Bone Thermal Index

Bone​Thermal​Index

DS

1

(0018,5026)

Cranial Thermal Index

Cranial​Thermal​Index

DS

1

(0018,5027)

Soft Tissue Thermal Index

Soft​Tissue​Thermal​Index

DS

1

(0018,5028)

Soft Tissue-focus Thermal Index

Soft​Tissue​Focus​Thermal​Index

DS

1

(0018,5029)

Soft Tissue-surface Thermal Index

Soft​Tissue​Surface​Thermal​Index

DS

1

(0018,5030)

Dynamic Range

Dynamic​Range

DS

1

RET

(0018,5040)

Total Gain

Total​Gain

DS

1

RET

(0018,5050)

Depth of Scan Field

Depth​Of​Scan​Field

IS

1

(0018,5100)

Patient Position

Patient​Position

CS

1

(0018,5101)

View Position

View​Position

CS

1

(0018,5104)

Projection Eponymous Name Code Sequence

Projection​Eponymous​Name​Code​Sequence

SQ

1

(0018,5210)

Image Transformation Matrix

Image​Transformation​Matrix

DS

6

RET (2003)

(0018,5212)

Image Translation Vector

Image​Translation​Vector

DS

3

RET (2003)

(0018,6000)

Sensitivity

Sensitivity

DS

1

(0018,6011)

Sequence of Ultrasound Regions

Sequence​Of​Ultrasound​Regions

SQ

1

(0018,6012)

Region Spatial Format

Region​Spatial​Format

US

1

(0018,6014)

Region Data Type

Region​Data​Type

US

1

(0018,6016)

Region Flags

Region​Flags

UL

1

(0018,6018)

Region Location Min X0

Region​Location​Min​X0

UL

1

(0018,601A)

Region Location Min Y0

Region​Location​Min​Y0

UL

1

(0018,601C)

Region Location Max X1

Region​Location​Max​X1

UL

1

(0018,601E)

Region Location Max Y1

Region​Location​Max​Y1

UL

1

(0018,6020)

Reference Pixel X0

Reference​Pixel​X0

SL

1

(0018,6022)

Reference Pixel Y0

Reference​Pixel​Y0

SL

1

(0018,6024)

Physical Units X Direction

Physical​Units​X​Direction

US

1

(0018,6026)

Physical Units Y Direction

Physical​Units​Y​Direction

US

1

(0018,6028)

Reference Pixel Physical Value X

Reference​Pixel​Physical​Value​X

FD

1

(0018,602A)

Reference Pixel Physical Value Y

Reference​Pixel​Physical​Value​Y

FD

1

(0018,602C)

Physical Delta X

Physical​Delta​X

FD

1

(0018,602E)

Physical Delta Y

Physical​Delta​Y

FD

1

(0018,6030)

Transducer Frequency

Transducer​Frequency

UL

1

(0018,6031)

Transducer Type

Transducer​Type

CS

1

(0018,6032)

Pulse Repetition Frequency

Pulse​Repetition​Frequency

UL

1

(0018,6034)

Doppler Correction Angle

Doppler​Correction​Angle

FD

1

(0018,6036)

Steering Angle

Steering​Angle

FD

1

(0018,6038)

Doppler Sample Volume X Position (Retired)

Doppler​Sample​Volume​X​Position​Retired

UL

1

RET (2001)

(0018,6039)

Doppler Sample Volume X Position

Doppler​Sample​Volume​X​Position

SL

1

(0018,603A)

Doppler Sample Volume Y Position (Retired)

Doppler​Sample​Volume​Y​Position​Retired

UL

1

RET (2001)

(0018,603B)

Doppler Sample Volume Y Position

Doppler​Sample​Volume​Y​Position

SL

1

(0018,603C)

TM-Line Position X0 (Retired)

TM​Line​Position​X0Retired

UL

1

RET (2001)

(0018,603D)

TM-Line Position X0

TM​Line​Position​X0

SL

1

(0018,603E)

TM-Line Position Y0 (Retired)

TM​Line​Position​Y0Retired

UL

1

RET (2001)

(0018,603F)

TM-Line Position Y0

TM​Line​Position​Y0

SL

1

(0018,6040)

TM-Line Position X1 (Retired)

TM​Line​Position​X1Retired

UL

1

RET (2001)

(0018,6041)

TM-Line Position X1

TM​Line​Position​X1

SL

1

(0018,6042)

TM-Line Position Y1 (Retired)

TM​Line​Position​Y1Retired

UL

1

RET (2001)

(0018,6043)

TM-Line Position Y1

TM​Line​Position​Y1

SL

1

(0018,6044)

Pixel Component Organization

Pixel​Component​Organization

US

1

(0018,6046)

Pixel Component Mask

Pixel​Component​Mask

UL

1

(0018,6048)

Pixel Component Range Start

Pixel​Component​Range​Start

UL

1

(0018,604A)

Pixel Component Range Stop

Pixel​Component​Range​Stop

UL

1

(0018,604C)

Pixel Component Physical Units

Pixel​Component​Physical​Units

US

1

(0018,604E)

Pixel Component Data Type

Pixel​Component​Data​Type

US

1

(0018,6050)

Number of Table Break Points

Number​Of​Table​Break​Points

UL

1

(0018,6052)

Table of X Break Points

Table​Of​X​Break​Points

UL

1-n

(0018,6054)

Table of Y Break Points

Table​Of​Y​Break​Points

FD

1-n

(0018,6056)

Number of Table Entries

Number​Of​Table​Entries

UL

1

(0018,6058)

Table of Pixel Values

Table​Of​Pixel​Values

UL

1-n

(0018,605A)

Table of Parameter Values

Table​Of​Parameter​Values

FL

1-n

(0018,6060)

R Wave Time Vector

R​Wave​Time​Vector

FL

1-n

(0018,6070)

Active Image Area Overlay Group

Active​Image​Area​Overlay​Group

US

1

(0018,7000)

Detector Conditions Nominal Flag

Detector​Conditions​Nominal​Flag

CS

1

(0018,7001)

Detector Temperature

Detector​Temperature

DS

1

(0018,7004)

Detector Type

Detector​Type

CS

1

(0018,7005)

Detector Configuration

Detector​Configuration

CS

1

(0018,7006)

Detector Description

Detector​Description

LT

1

(0018,7008)

Detector Mode

Detector​Mode

LT

1

(0018,700A)

Detector ID

Detector​ID

SH

1

(0018,700C)

Date of Last Detector Calibration

Date​Of​Last​Detector​Calibration

DA

1

(0018,700E)

Time of Last Detector Calibration

Time​Of​Last​Detector​Calibration

TM

1

(0018,7010)

Exposures on Detector Since Last Calibration

Exposures​On​Detector​Since​Last​Calibration

IS

1

(0018,7011)

Exposures on Detector Since Manufactured

Exposures​On​Detector​Since​Manufactured

IS

1

(0018,7012)

Detector Time Since Last Exposure

Detector​Time​Since​Last​Exposure

DS

1

(0018,7014)

Detector Active Time

Detector​Active​Time

DS

1

(0018,7016)

Detector Activation Offset From Exposure

Detector​Activation​Offset​From​Exposure

DS

1

(0018,701A)

Detector Binning

Detector​Binning

DS

2

(0018,7020)

Detector Element Physical Size

Detector​Element​Physical​Size

DS

2

(0018,7022)

Detector Element Spacing

Detector​Element​Spacing

DS

2

(0018,7024)

Detector Active Shape

Detector​Active​Shape

CS

1

(0018,7026)

Detector Active Dimension(s)

Detector​Active​Dimensions

DS

1-2

(0018,7028)

Detector Active Origin

Detector​Active​Origin

DS

2

(0018,702A)

Detector Manufacturer Name

Detector​Manufacturer​Name

LO

1

(0018,702B)

Detector Manufacturer's Model Name

Detector​Manufacturer​Model​Name

LO

1

(0018,7030)

Field of View Origin

Field​Of​View​Origin

DS

2

(0018,7032)

Field of View Rotation

Field​Of​View​Rotation

DS

1

(0018,7034)

Field of View Horizontal Flip

Field​Of​View​Horizontal​Flip

CS

1

(0018,7036)

Pixel Data Area Origin Relative To FOV

Pixel​Data​Area​Origin​Relative​To​FOV

FL

2

(0018,7038)

Pixel Data Area Rotation Angle Relative To FOV

Pixel​Data​Area​Rotation​Angle​Relative​To​FOV

FL

1

(0018,7040)

Grid Absorbing Material

Grid​Absorbing​Material

LT

1

(0018,7041)

Grid Spacing Material

Grid​Spacing​Material

LT

1

(0018,7042)

Grid Thickness

Grid​Thickness

DS

1

(0018,7044)

Grid Pitch

Grid​Pitch

DS

1

(0018,7046)

Grid Aspect Ratio

Grid​Aspect​Ratio

IS

2

(0018,7048)

Grid Period

Grid​Period

DS

1

(0018,704C)

Grid Focal Distance

Grid​Focal​Distance

DS

1

(0018,7050)

Filter Material

Filter​Material

CS

1-n

(0018,7052)

Filter Thickness Minimum

Filter​Thickness​Minimum

DS

1-n

(0018,7054)

Filter Thickness Maximum

Filter​Thickness​Maximum

DS

1-n

(0018,7056)

Filter Beam Path Length Minimum

Filter​Beam​Path​Length​Minimum

FL

1-n

(0018,7058)

Filter Beam Path Length Maximum

Filter​Beam​Path​Length​Maximum

FL

1-n

(0018,7060)

Exposure Control Mode

Exposure​Control​Mode

CS

1

(0018,7062)

Exposure Control Mode Description

Exposure​Control​Mode​Description

LT

1

(0018,7064)

Exposure Status

Exposure​Status

CS

1

(0018,7065)

Phototimer Setting

Phototimer​Setting

DS

1

(0018,8150)

Exposure Time in µS

Exposure​Time​InuS

DS

1

(0018,8151)

X-Ray Tube Current in µA

X​Ray​Tube​Current​InuA

DS

1

(0018,9004)

Content Qualification

Content​Qualification

CS

1

(0018,9005)

Pulse Sequence Name

Pulse​Sequence​Name

SH

1

(0018,9006)

MR Imaging Modifier Sequence

MR​Imaging​Modifier​Sequence

SQ

1

(0018,9008)

Echo Pulse Sequence

Echo​Pulse​Sequence

CS

1

(0018,9009)

Inversion Recovery

Inversion​Recovery

CS

1

(0018,9010)

Flow Compensation

Flow​Compensation

CS

1

(0018,9011)

Multiple Spin Echo

Multiple​Spin​Echo

CS

1

(0018,9012)

Multi-planar Excitation

Multi​Planar​Excitation

CS

1

(0018,9014)

Phase Contrast

Phase​Contrast

CS

1

(0018,9015)

Time of Flight Contrast

Time​Of​Flight​Contrast

CS

1

(0018,9016)

Spoiling

Spoiling

CS

1

(0018,9017)

Steady State Pulse Sequence

Steady​State​Pulse​Sequence

CS

1

(0018,9018)

Echo Planar Pulse Sequence

Echo​Planar​Pulse​Sequence

CS

1

(0018,9019)

Tag Angle First Axis

Tag​Angle​First​Axis

FD

1

(0018,9020)

Magnetization Transfer

Magnetization​Transfer

CS

1

(0018,9021)

T2 Preparation

T2Preparation

CS

1

(0018,9022)

Blood Signal Nulling

Blood​Signal​Nulling

CS

1

(0018,9024)

Saturation Recovery

Saturation​Recovery

CS

1

(0018,9025)

Spectrally Selected Suppression

Spectrally​Selected​Suppression

CS

1

(0018,9026)

Spectrally Selected Excitation

Spectrally​Selected​Excitation

CS

1

(0018,9027)

Spatial Pre-saturation

Spatial​Presaturation

CS

1

(0018,9028)

Tagging

Tagging

CS

1

(0018,9029)

Oversampling Phase

Oversampling​Phase

CS

1

(0018,9030)

Tag Spacing First Dimension

Tag​Spacing​First​Dimension

FD

1

(0018,9032)

Geometry of k-Space Traversal

Geometry​Of​K​Space​Traversal

CS

1

(0018,9033)

Segmented k-Space Traversal

Segmented​K​Space​Traversal

CS

1

(0018,9034)

Rectilinear Phase Encode Reordering

Rectilinear​Phase​Encode​Reordering

CS

1

(0018,9035)

Tag Thickness

Tag​Thickness

FD

1

(0018,9036)

Partial Fourier Direction

Partial​Fourier​Direction

CS

1

(0018,9037)

Cardiac Synchronization Technique

Cardiac​Synchronization​Technique

CS

1

(0018,9041)

Receive Coil Manufacturer Name

Receive​Coil​Manufacturer​Name

LO

1

(0018,9042)

MR Receive Coil Sequence

MR​Receive​Coil​Sequence

SQ

1

(0018,9043)

Receive Coil Type

Receive​Coil​Type

CS

1

(0018,9044)

Quadrature Receive Coil

Quadrature​Receive​Coil

CS

1

(0018,9045)

Multi-Coil Definition Sequence

Multi​Coil​Definition​Sequence

SQ

1

(0018,9046)

Multi-Coil Configuration

Multi​Coil​Configuration

LO

1

(0018,9047)

Multi-Coil Element Name

Multi​Coil​Element​Name

SH

1

(0018,9048)

Multi-Coil Element Used

Multi​Coil​Element​Used

CS

1

(0018,9049)

MR Transmit Coil Sequence

MR​Transmit​Coil​Sequence

SQ

1

(0018,9050)

Transmit Coil Manufacturer Name

Transmit​Coil​Manufacturer​Name

LO

1

(0018,9051)

Transmit Coil Type

Transmit​Coil​Type

CS

1

(0018,9052)

Spectral Width

Spectral​Width

FD

1-2

(0018,9053)

Chemical Shift Reference

Chemical​Shift​Reference

FD

1-2

(0018,9054)

Volume Localization Technique

Volume​Localization​Technique

CS

1

(0018,9058)

MR Acquisition Frequency Encoding Steps

MR​Acquisition​Frequency​Encoding​Steps

US

1

(0018,9059)

De-coupling

Decoupling

CS

1

(0018,9060)

De-coupled Nucleus

Decoupled​Nucleus

CS

1-2

(0018,9061)

De-coupling Frequency

Decoupling​Frequency

FD

1-2

(0018,9062)

De-coupling Method

Decoupling​Method

CS

1

(0018,9063)

De-coupling Chemical Shift Reference

Decoupling​Chemical​Shift​Reference

FD

1-2

(0018,9064)

k-space Filtering

K​Space​Filtering

CS

1

(0018,9065)

Time Domain Filtering

Time​Domain​Filtering

CS

1-2

(0018,9066)

Number of Zero Fills

Number​Of​Zero​Fills

US

1-2

(0018,9067)

Baseline Correction

Baseline​Correction

CS

1

(0018,9069)

Parallel Reduction Factor In-plane

Parallel​Reduction​Factor​In​Plane

FD

1

(0018,9070)

Cardiac R-R Interval Specified

Cardiac​RR​Interval​Specified

FD

1

(0018,9073)

Acquisition Duration

Acquisition​Duration

FD

1

(0018,9074)

Frame Acquisition DateTime

Frame​Acquisition​Date​Time

DT

1

(0018,9075)

Diffusion Directionality

Diffusion​Directionality

CS

1

(0018,9076)

Diffusion Gradient Direction Sequence

Diffusion​Gradient​Direction​Sequence

SQ

1

(0018,9077)

Parallel Acquisition

Parallel​Acquisition

CS

1

(0018,9078)

Parallel Acquisition Technique

Parallel​Acquisition​Technique

CS

1

(0018,9079)

Inversion Times

Inversion​Times

FD

1-n

(0018,9080)

Metabolite Map Description

Metabolite​Map​Description

ST

1

(0018,9081)

Partial Fourier

Partial​Fourier

CS

1

(0018,9082)

Effective Echo Time

Effective​Echo​Time

FD

1

(0018,9083)

Metabolite Map Code Sequence

Metabolite​Map​Code​Sequence

SQ

1

(0018,9084)

Chemical Shift Sequence

Chemical​Shift​Sequence

SQ

1

(0018,9085)

Cardiac Signal Source

Cardiac​Signal​Source

CS

1

(0018,9087)

Diffusion b-value

Diffusion​B​Value

FD

1

(0018,9089)

Diffusion Gradient Orientation

Diffusion​Gradient​Orientation

FD

3

(0018,9090)

Velocity Encoding Direction

Velocity​Encoding​Direction

FD

3

(0018,9091)

Velocity Encoding Minimum Value

Velocity​Encoding​Minimum​Value

FD

1

(0018,9092)

Velocity Encoding Acquisition Sequence

Velocity​Encoding​Acquisition​Sequence

SQ

1

(0018,9093)

Number of k-Space Trajectories

Number​Of​K​Space​Trajectories

US

1

(0018,9094)

Coverage of k-Space

Coverage​Of​K​Space

CS

1

(0018,9095)

Spectroscopy Acquisition Phase Rows

Spectroscopy​Acquisition​Phase​Rows

UL

1

(0018,9096)

Parallel Reduction Factor In-plane (Retired)

Parallel​Reduction​Factor​In​Plane​Retired

FD

1

RET (2007)

(0018,9098)

Transmitter Frequency

Transmitter​Frequency

FD

1-2

(0018,9100)

Resonant Nucleus

Resonant​Nucleus

CS

1-2

(0018,9101)

Frequency Correction

Frequency​Correction

CS

1

(0018,9103)

MR Spectroscopy FOV/Geometry Sequence

MR​Spectroscopy​FOV​Geometry​Sequence

SQ

1

(0018,9104)

Slab Thickness

Slab​Thickness

FD

1

(0018,9105)

Slab Orientation

Slab​Orientation

FD

3

(0018,9106)

Mid Slab Position

Mid​Slab​Position

FD

3

(0018,9107)

MR Spatial Saturation Sequence

MR​Spatial​Saturation​Sequence

SQ

1

(0018,9112)

MR Timing and Related Parameters Sequence

MR​Timing​And​Related​Parameters​Sequence

SQ

1

(0018,9114)

MR Echo Sequence

MR​Echo​Sequence

SQ

1

(0018,9115)

MR Modifier Sequence

MR​Modifier​Sequence

SQ

1

(0018,9117)

MR Diffusion Sequence

MR​Diffusion​Sequence

SQ

1

(0018,9118)

Cardiac Synchronization Sequence

Cardiac​Synchronization​Sequence

SQ

1

(0018,9119)

MR Averages Sequence

MR​Averages​Sequence

SQ

1

(0018,9125)

MR FOV/Geometry Sequence

MRFOV​Geometry​Sequence

SQ

1

(0018,9126)

Volume Localization Sequence

Volume​Localization​Sequence

SQ

1

(0018,9127)

Spectroscopy Acquisition Data Columns

Spectroscopy​Acquisition​Data​Columns

UL

1

(0018,9147)

Diffusion Anisotropy Type

Diffusion​Anisotropy​Type

CS

1

(0018,9151)

Frame Reference DateTime

Frame​Reference​Date​Time

DT

1

(0018,9152)

MR Metabolite Map Sequence

MR​Metabolite​Map​Sequence

SQ

1

(0018,9155)

Parallel Reduction Factor out-of-plane

Parallel​Reduction​Factor​Out​Of​Plane

FD

1

(0018,9159)

Spectroscopy Acquisition Out-of-plane Phase Steps

Spectroscopy​Acquisition​Out​Of​Plane​Phase​Steps

UL

1

(0018,9166)

Bulk Motion Status

Bulk​Motion​Status

CS

1

RET (2007)

(0018,9168)

Parallel Reduction Factor Second In-plane

Parallel​Reduction​Factor​Second​In​Plane

FD

1

(0018,9169)

Cardiac Beat Rejection Technique

Cardiac​Beat​Rejection​Technique

CS

1

(0018,9170)

Respiratory Motion Compensation Technique

Respiratory​Motion​Compensation​Technique

CS

1

(0018,9171)

Respiratory Signal Source

Respiratory​Signal​Source

CS

1

(0018,9172)

Bulk Motion Compensation Technique

Bulk​Motion​Compensation​Technique

CS

1

(0018,9173)

Bulk Motion Signal Source

Bulk​Motion​Signal​Source

CS

1

(0018,9174)

Applicable Safety Standard Agency

Applicable​Safety​Standard​Agency

CS

1

(0018,9175)

Applicable Safety Standard Description

Applicable​Safety​Standard​Description

LO

1

(0018,9176)

Operating Mode Sequence

Operating​Mode​Sequence

SQ

1

(0018,9177)

Operating Mode Type

Operating​Mode​Type

CS

1

(0018,9178)

Operating Mode

Operating​Mode

CS

1

(0018,9179)

Specific Absorption Rate Definition

Specific​Absorption​Rate​Definition

CS

1

(0018,9180)

Gradient Output Type

Gradient​Output​Type

CS

1

(0018,9181)

Specific Absorption Rate Value

Specific​Absorption​Rate​Value

FD

1

(0018,9182)

Gradient Output

Gradient​Output

FD

1

(0018,9183)

Flow Compensation Direction

Flow​Compensation​Direction

CS

1

(0018,9184)

Tagging Delay

Tagging​Delay

FD

1

(0018,9185)

Respiratory Motion Compensation Technique Description

Respiratory​Motion​Compensation​Technique​Description

ST

1

(0018,9186)

Respiratory Signal Source ID

Respiratory​Signal​Source​ID

SH

1

(0018,9195)

Chemical Shift Minimum Integration Limit in Hz

Chemical​Shift​Minimum​Integration​Limit​In​Hz

FD

1

RET (2003)

(0018,9196)

Chemical Shift Maximum Integration Limit in Hz

Chemical​Shift​Maximum​Integration​Limit​In​Hz

FD

1

RET (2003)

(0018,9197)

MR Velocity Encoding Sequence

MR​Velocity​Encoding​Sequence

SQ

1

(0018,9198)

First Order Phase Correction

First​Order​Phase​Correction

CS

1

(0018,9199)

Water Referenced Phase Correction

Water​Referenced​Phase​Correction

CS

1

(0018,9200)

MR Spectroscopy Acquisition Type

MR​Spectroscopy​Acquisition​Type

CS

1

(0018,9214)

Respiratory Cycle Position

Respiratory​Cycle​Position

CS

1

(0018,9217)

Velocity Encoding Maximum Value

Velocity​Encoding​Maximum​Value

FD

1

(0018,9218)

Tag Spacing Second Dimension

Tag​Spacing​Second​Dimension

FD

1

(0018,9219)

Tag Angle Second Axis

Tag​Angle​Second​Axis

SS

1

(0018,9220)

Frame Acquisition Duration

Frame​Acquisition​Duration

FD

1

(0018,9226)

MR Image Frame Type Sequence

MR​Image​Frame​Type​Sequence

SQ

1

(0018,9227)

MR Spectroscopy Frame Type Sequence

MR​Spectroscopy​Frame​Type​Sequence

SQ

1

(0018,9231)

MR Acquisition Phase Encoding Steps in-plane

MR​Acquisition​Phase​Encoding​Steps​In​Plane

US

1

(0018,9232)

MR Acquisition Phase Encoding Steps out-of-plane

MR​Acquisition​Phase​Encoding​Steps​Out​Of​Plane

US

1

(0018,9234)

Spectroscopy Acquisition Phase Columns

Spectroscopy​Acquisition​Phase​Columns

UL

1

(0018,9236)

Cardiac Cycle Position

Cardiac​Cycle​Position

CS

1

(0018,9239)

Specific Absorption Rate Sequence

Specific​Absorption​Rate​Sequence

SQ

1

(0018,9240)

RF Echo Train Length

RF​Echo​Train​Length

US

1

(0018,9241)

Gradient Echo Train Length

Gradient​Echo​Train​Length

US

1

(0018,9250)

Arterial Spin Labeling Contrast

Arterial​Spin​Labeling​Contrast

CS

1

(0018,9251)

MR Arterial Spin Labeling Sequence

MR​Arterial​Spin​Labeling​Sequence

SQ

1

(0018,9252)

ASL Technique Description

ASL​Technique​Description

LO

1

(0018,9253)

ASL Slab Number

ASL​Slab​Number

US

1

(0018,9254)

ASL Slab Thickness

ASL​Slab​Thickness

FD

1

(0018,9255)

ASL Slab Orientation

ASL​Slab​Orientation

FD

3

(0018,9256)

ASL Mid Slab Position

ASL​Mid​Slab​Position

FD

3

(0018,9257)

ASL Context

ASL​Context

CS

1

(0018,9258)

ASL Pulse Train Duration

ASL​Pulse​Train​Duration

UL

1

(0018,9259)

ASL Crusher Flag

ASL​Crusher​Flag

CS

1

(0018,925A)

ASL Crusher Flow Limit

ASL​Crusher​Flow​Limit

FD

1

(0018,925B)

ASL Crusher Description

ASL​Crusher​Description

LO

1

(0018,925C)

ASL Bolus Cut-off Flag

ASL​Bolus​Cutoff​Flag

CS

1

(0018,925D)

ASL Bolus Cut-off Timing Sequence

ASL​Bolus​Cutoff​Timing​Sequence

SQ

1

(0018,925E)

ASL Bolus Cut-off Technique

ASL​Bolus​Cutoff​Technique

LO

1

(0018,925F)

ASL Bolus Cut-off Delay Time

ASL​Bolus​Cutoff​Delay​Time

UL

1

(0018,9260)

ASL Slab Sequence

ASL​Slab​Sequence

SQ

1

(0018,9295)

Chemical Shift Minimum Integration Limit in ppm

Chemical​Shift​Minimum​Integration​Limit​Inppm

FD

1

(0018,9296)

Chemical Shift Maximum Integration Limit in ppm

Chemical​Shift​Maximum​Integration​Limit​Inppm

FD

1

(0018,9297)

Water Reference Acquisition

Water​Reference​Acquisition

CS

1

(0018,9298)

Echo Peak Position

Echo​Peak​Position

IS

1

(0018,9301)

CT Acquisition Type Sequence

CT​Acquisition​Type​Sequence

SQ

1

(0018,9302)

Acquisition Type

Acquisition​Type

CS

1

(0018,9303)

Tube Angle

Tube​Angle

FD

1

(0018,9304)

CT Acquisition Details Sequence

CT​Acquisition​Details​Sequence

SQ

1

(0018,9305)

Revolution Time

Revolution​Time

FD

1

(0018,9306)

Single Collimation Width

Single​Collimation​Width

FD

1

(0018,9307)

Total Collimation Width

Total​Collimation​Width

FD

1

(0018,9308)

CT Table Dynamics Sequence

CT​Table​Dynamics​Sequence

SQ

1

(0018,9309)

Table Speed

Table​Speed

FD

1

(0018,9310)

Table Feed per Rotation

Table​Feed​Per​Rotation

FD

1

(0018,9311)

Spiral Pitch Factor

Spiral​Pitch​Factor

FD

1

(0018,9312)

CT Geometry Sequence

CT​Geometry​Sequence

SQ

1

(0018,9313)

Data Collection Center (Patient)

Data​Collection​Center​Patient

FD

3

(0018,9314)

CT Reconstruction Sequence

CT​Reconstruction​Sequence

SQ

1

(0018,9315)

Reconstruction Algorithm

Reconstruction​Algorithm

CS

1

(0018,9316)

Convolution Kernel Group

Convolution​Kernel​Group

CS

1

(0018,9317)

Reconstruction Field of View

Reconstruction​Field​Of​View

FD

2

(0018,9318)

Reconstruction Target Center (Patient)

Reconstruction​Target​Center​Patient

FD

3

(0018,9319)

Reconstruction Angle

Reconstruction​Angle

FD

1

(0018,9320)

Image Filter

Image​Filter

SH

1

(0018,9321)

CT Exposure Sequence

CT​Exposure​Sequence

SQ

1

(0018,9322)

Reconstruction Pixel Spacing

Reconstruction​Pixel​Spacing

FD

2

(0018,9323)

Exposure Modulation Type

Exposure​Modulation​Type

CS

1-n

(0018,9324)

Estimated Dose Saving

Estimated​Dose​Saving

FD

1

RET (2020a)

(0018,9325)

CT X-Ray Details Sequence

CT​X​Ray​Details​Sequence

SQ

1

(0018,9326)

CT Position Sequence

CT​Position​Sequence

SQ

1

(0018,9327)

Table Position

Table​Position

FD

1

(0018,9328)

Exposure Time in ms

Exposure​Time​Inms

FD

1

(0018,9329)

CT Image Frame Type Sequence

CT​Image​Frame​Type​Sequence

SQ

1

(0018,9330)

X-Ray Tube Current in mA

X​Ray​Tube​Current​InmA

FD

1

(0018,9332)

Exposure in mAs

Exposure​Inm​As

FD

1

(0018,9333)

Constant Volume Flag

Constant​Volume​Flag

CS

1

(0018,9334)

Fluoroscopy Flag

Fluoroscopy​Flag

CS

1

(0018,9335)

Distance Source to Data Collection Center

Distance​Source​To​Data​Collection​Center

FD

1

(0018,9337)

Contrast/Bolus Agent Number

Contrast​Bolus​Agent​Number

US

1

(0018,9338)

Contrast/Bolus Ingredient Code Sequence

Contrast​Bolus​Ingredient​Code​Sequence

SQ

1

(0018,9340)

Contrast Administration Profile Sequence

Contrast​Administration​Profile​Sequence

SQ

1

(0018,9341)

Contrast/Bolus Usage Sequence

Contrast​Bolus​Usage​Sequence

SQ

1

(0018,9342)

Contrast/Bolus Agent Administered

Contrast​Bolus​Agent​Administered

CS

1

(0018,9343)

Contrast/Bolus Agent Detected

Contrast​Bolus​Agent​Detected

CS

1

(0018,9344)

Contrast/Bolus Agent Phase

Contrast​Bolus​Agent​Phase

CS

1

(0018,9345)

CTDIvol

CTD​Ivol

FD

1

(0018,9346)

CTDI Phantom Type Code Sequence

CTDI​Phantom​Type​Code​Sequence

SQ

1

(0018,9351)

Calcium Scoring Mass Factor Patient

Calcium​Scoring​Mass​Factor​Patient

FL

1

(0018,9352)

Calcium Scoring Mass Factor Device

Calcium​Scoring​Mass​Factor​Device

FL

3

(0018,9353)

Energy Weighting Factor

Energy​Weighting​Factor

FL

1

(0018,9360)

CT Additional X-Ray Source Sequence

CT​Additional​X​Ray​Source​Sequence

SQ

1

(0018,9361)

Multi-energy CT Acquisition

Multienergy​CT​Acquisition

CS

1

(0018,9362)

Multi-energy CT Acquisition Sequence

Multienergy​CT​Acquisition​Sequence

SQ

1

(0018,9363)

Multi-energy CT Processing Sequence

Multienergy​CT​Processing​Sequence

SQ

1

(0018,9364)

Multi-energy CT Characteristics Sequence

Multienergy​CT​Characteristics​Sequence

SQ

1

(0018,9365)

Multi-energy CT X-Ray Source Sequence

Multienergy​CT​XRay​Source​Sequence

SQ

1

(0018,9366)

X-Ray Source Index

XRay​Source​Index

US

1

(0018,9367)

X-Ray Source ID

XRay​Source​ID

UC

1

(0018,9368)

Multi-energy Source Technique

Multienergy​Source​Technique

CS

1

(0018,9369)

Source Start DateTime

Source​Start​Date​Time

DT

1

(0018,936A)

Source End DateTime

Source​End​Date​Time

DT

1

(0018,936B)

Switching Phase Number

Switching​Phase​Number

US

1

(0018,936C)

Switching Phase Nominal Duration

Switching​Phase​Nominal​Duration

DS

1

(0018,936D)

Switching Phase Transition Duration

Switching​Phase​Transition​Duration

DS

1

(0018,936E)

Effective Bin Energy

Effective​Bin​Energy

DS

1

(0018,936F)

Multi-energy CT X-Ray Detector Sequence

Multienergy​CT​XRay​Detector​Sequence

SQ

1

(0018,9370)

X-Ray Detector Index

XRay​Detector​Index

US

1

(0018,9371)

X-Ray Detector ID

XRay​Detector​ID

UC

1

(0018,9372)

Multi-energy Detector Type

Multienergy​Detector​Type

CS

1

(0018,9373)

X-Ray Detector Label

XRay​Detector​Label

ST

1

(0018,9374)

Nominal Max Energy

Nominal​Max​Energy

DS

1

(0018,9375)

Nominal Min Energy

Nominal​Min​Energy

DS

1

(0018,9376)

Referenced X-Ray Detector Index

Referenced​XRay​Detector​Index

US

1-n

(0018,9377)

Referenced X-Ray Source Index

Referenced​XRay​Source​Index

US

1-n

(0018,9378)

Referenced Path Index

Referenced​Path​Index

US

1-n

(0018,9379)

Multi-energy CT Path Sequence

Multienergy​CT​Path​Sequence

SQ

1

(0018,937A)

Multi-energy CT Path Index

Multienergy​CT​Path​Index

US

1

(0018,937B)

Multi-energy Acquisition Description

Multienergy​Acquisition​Description

UT

1

(0018,937C)

Monoenergetic Energy Equivalent

Monoenergetic​Energy​Equivalent

FD

1

(0018,937D)

Material Code Sequence

Material​Code​Sequence

SQ

1

(0018,937E)

Decomposition Method

Decomposition​Method

CS

1

(0018,937F)

Decomposition Description

Decomposition​Description

UT

1

(0018,9380)

Decomposition Algorithm Identification Sequence

Decomposition​Algorithm​Identification​Sequence

SQ

1

(0018,9381)

Decomposition Material Sequence

Decomposition​Material​Sequence

SQ

1

(0018,9382)

Material Attenuation Sequence

Material​Attenuation​Sequence

SQ

1

(0018,9383)

Photon Energy

Photon​Energy

DS

1

(0018,9384)

X-Ray Mass Attenuation Coefficient

XRay​Mass​Attenuation​Coefficient

DS

1

(0018,9390)

Metal Artifact Reduction Sequence

Metal​Artifact​Reduction​Sequence

SQ

1

(0018,9391)

Metal Artifact Reduction Applied

Metal​Artifact​Reduction​Applied

CS

1

(0018,9392)

Metal Artifact Reduction Algorithm Identification Sequence

Metal​Artifact​Reduction​Algorithm​Identification​Sequence

SQ

1

(0018,9401)

Projection Pixel Calibration Sequence

Projection​Pixel​Calibration​Sequence

SQ

1

(0018,9402)

Distance Source to Isocenter

Distance​Source​To​Isocenter

FL

1

(0018,9403)

Distance Object to Table Top

Distance​Object​To​Table​Top

FL

1

(0018,9404)

Object Pixel Spacing in Center of Beam

Object​Pixel​Spacing​In​Center​Of​Beam

FL

2

(0018,9405)

Positioner Position Sequence

Positioner​Position​Sequence

SQ

1

(0018,9406)

Table Position Sequence

Table​Position​Sequence

SQ

1

(0018,9407)

Collimator Shape Sequence

Collimator​Shape​Sequence

SQ

1

(0018,9410)

Planes in Acquisition

Planes​In​Acquisition

CS

1

(0018,9412)

XA/XRF Frame Characteristics Sequence

XAXRF​Frame​Characteristics​Sequence

SQ

1

(0018,9417)

Frame Acquisition Sequence

Frame​Acquisition​Sequence

SQ

1

(0018,9420)

X-Ray Receptor Type

X​Ray​Receptor​Type

CS

1

(0018,9423)

Acquisition Protocol Name

Acquisition​Protocol​Name

LO

1

(0018,9424)

Acquisition Protocol Description

Acquisition​Protocol​Description

LT

1

(0018,9425)

Contrast/Bolus Ingredient Opaque

Contrast​Bolus​Ingredient​Opaque

CS

1

(0018,9426)

Distance Receptor Plane to Detector Housing

Distance​Receptor​Plane​To​Detector​Housing

FL

1

(0018,9427)

Intensifier Active Shape

Intensifier​Active​Shape

CS

1

(0018,9428)

Intensifier Active Dimension(s)

Intensifier​Active​Dimensions

FL

1-2

(0018,9429)

Physical Detector Size

Physical​Detector​Size

FL

2

(0018,9430)

Position of Isocenter Projection

Position​Of​Isocenter​Projection

FL

2

(0018,9432)

Field of View Sequence

Field​Of​View​Sequence

SQ

1

(0018,9433)

Field of View Description

Field​Of​View​Description

LO

1

(0018,9434)

Exposure Control Sensing Regions Sequence

Exposure​Control​Sensing​Regions​Sequence

SQ

1

(0018,9435)

Exposure Control Sensing Region Shape

Exposure​Control​Sensing​Region​Shape

CS

1

(0018,9436)

Exposure Control Sensing Region Left Vertical Edge

Exposure​Control​Sensing​Region​Left​Vertical​Edge

SS

1

(0018,9437)

Exposure Control Sensing Region Right Vertical Edge

Exposure​Control​Sensing​Region​Right​Vertical​Edge

SS

1

(0018,9438)

Exposure Control Sensing Region Upper Horizontal Edge

Exposure​Control​Sensing​Region​Upper​Horizontal​Edge

SS

1

(0018,9439)

Exposure Control Sensing Region Lower Horizontal Edge

Exposure​Control​Sensing​Region​Lower​Horizontal​Edge

SS

1

(0018,9440)

Center of Circular Exposure Control Sensing Region

Center​Of​Circular​Exposure​Control​Sensing​Region

SS

2

(0018,9441)

Radius of Circular Exposure Control Sensing Region

Radius​Of​Circular​Exposure​Control​Sensing​Region

US

1

(0018,9442)

Vertices of the Polygonal Exposure Control Sensing Region

Vertices​Of​The​Polygonal​Exposure​Control​Sensing​Region

SS

2-n

(0018,9445)

RET (2004) - See Note 3

(0018,9447)

Column Angulation (Patient)

Column​Angulation​Patient

FL

1

(0018,9449)

Beam Angle

Beam​Angle

FL

1

(0018,9451)

Frame Detector Parameters Sequence

Frame​Detector​Parameters​Sequence

SQ

1

(0018,9452)

Calculated Anatomy Thickness

Calculated​Anatomy​Thickness

FL

1

(0018,9455)

Calibration Sequence

Calibration​Sequence

SQ

1

(0018,9456)

Object Thickness Sequence

Object​Thickness​Sequence

SQ

1

(0018,9457)

Plane Identification

Plane​Identification

CS

1

(0018,9461)

Field of View Dimension(s) in Float

Field​Of​View​Dimensions​In​Float

FL

1-2

(0018,9462)

Isocenter Reference System Sequence

Isocenter​Reference​System​Sequence

SQ

1

(0018,9463)

Positioner Isocenter Primary Angle

Positioner​Isocenter​Primary​Angle

FL

1

(0018,9464)

Positioner Isocenter Secondary Angle

Positioner​Isocenter​Secondary​Angle

FL

1

(0018,9465)

Positioner Isocenter Detector Rotation Angle

Positioner​Isocenter​Detector​Rotation​Angle

FL

1

(0018,9466)

Table X Position to Isocenter

Table​X​Position​To​Isocenter

FL

1

(0018,9467)

Table Y Position to Isocenter

Table​Y​Position​To​Isocenter

FL

1

(0018,9468)

Table Z Position to Isocenter

Table​Z​Position​To​Isocenter

FL

1

(0018,9469)

Table Horizontal Rotation Angle

Table​Horizontal​Rotation​Angle

FL

1

(0018,9470)

Table Head Tilt Angle

Table​Head​Tilt​Angle

FL

1

(0018,9471)

Table Cradle Tilt Angle

Table​Cradle​Tilt​Angle

FL

1

(0018,9472)

Frame Display Shutter Sequence

Frame​Display​Shutter​Sequence

SQ

1

(0018,9473)

Acquired Image Area Dose Product

Acquired​Image​Area​Dose​Product

FL

1

(0018,9474)

C-arm Positioner Tabletop Relationship

C​Arm​Positioner​Tabletop​Relationship

CS

1

(0018,9476)

X-Ray Geometry Sequence

X​Ray​Geometry​Sequence

SQ

1

(0018,9477)

Irradiation Event Identification Sequence

Irradiation​Event​Identification​Sequence

SQ

1

(0018,9504)

X-Ray 3D Frame Type Sequence

X​Ray3D​Frame​Type​Sequence

SQ

1

(0018,9506)

Contributing Sources Sequence

Contributing​Sources​Sequence

SQ

1

(0018,9507)

X-Ray 3D Acquisition Sequence

X​Ray3D​Acquisition​Sequence

SQ

1

(0018,9508)

Primary Positioner Scan Arc

Primary​Positioner​Scan​Arc

FL

1

(0018,9509)

Secondary Positioner Scan Arc

Secondary​Positioner​Scan​Arc

FL

1

(0018,9510)

Primary Positioner Scan Start Angle

Primary​Positioner​Scan​Start​Angle

FL

1

(0018,9511)

Secondary Positioner Scan Start Angle

Secondary​Positioner​Scan​Start​Angle

FL

1

(0018,9514)

Primary Positioner Increment

Primary​Positioner​Increment

FL

1

(0018,9515)

Secondary Positioner Increment

Secondary​Positioner​Increment

FL

1

(0018,9516)

Start Acquisition DateTime

Start​Acquisition​Date​Time

DT

1

(0018,9517)

End Acquisition DateTime

End​Acquisition​Date​Time

DT

1

(0018,9518)

Primary Positioner Increment Sign

Primary​Positioner​Increment​Sign

SS

1

(0018,9519)

Secondary Positioner Increment Sign

Secondary​Positioner​Increment​Sign

SS

1

(0018,9524)

Application Name

Application​Name

LO

1

(0018,9525)

Application Version

Application​Version

LO

1

(0018,9526)

Application Manufacturer

Application​Manufacturer

LO

1

(0018,9527)

Algorithm Type

Algorithm​Type

CS

1

(0018,9528)

Algorithm Description

Algorithm​Description

LO

1

(0018,9530)

X-Ray 3D Reconstruction Sequence

X​Ray3D​Reconstruction​Sequence

SQ

1

(0018,9531)

Reconstruction Description

Reconstruction​Description

LO

1

(0018,9538)

Per Projection Acquisition Sequence

Per​Projection​Acquisition​Sequence

SQ

1

(0018,9541)

Detector Position Sequence

Detector​Position​Sequence

SQ

1

(0018,9542)

X-Ray Acquisition Dose Sequence

XRay​Acquisition​Dose​Sequence

SQ

1

(0018,9543)

X-Ray Source Isocenter Primary Angle

XRay​Source​Isocenter​Primary​Angle

FD

1

(0018,9544)

X-Ray Source Isocenter Secondary Angle

XRay​Source​Isocenter​Secondary​Angle

FD

1

(0018,9545)

Breast Support Isocenter Primary Angle

Breast​Support​Isocenter​Primary​Angle

FD

1

(0018,9546)

Breast Support Isocenter Secondary Angle

Breast​Support​Isocenter​Secondary​Angle

FD

1

(0018,9547)

Breast Support X Position to Isocenter

Breast​Support​XPosition​To​Isocenter

FD

1

(0018,9548)

Breast Support Y Position to Isocenter

Breast​Support​YPosition​To​Isocenter

FD

1

(0018,9549)

Breast Support Z Position to Isocenter

Breast​Support​ZPosition​To​Isocenter

FD

1

(0018,9550)

Detector Isocenter Primary Angle

Detector​Isocenter​Primary​Angle

FD

1

(0018,9551)

Detector Isocenter Secondary Angle

Detector​Isocenter​Secondary​Angle

FD

1

(0018,9552)

Detector X Position to Isocenter

Detector​XPosition​To​Isocenter

FD

1

(0018,9553)

Detector Y Position to Isocenter

Detector​YPosition​To​Isocenter

FD

1

(0018,9554)

Detector Z Position to Isocenter

Detector​ZPosition​To​Isocenter

FD

1

(0018,9555)

X-Ray Grid Sequence

XRay​Grid​Sequence

SQ

1

(0018,9556)

X-Ray Filter Sequence

XRay​Filter​Sequence

SQ

1

(0018,9557)

Detector Active Area TLHC Position

Detector​Active​Area​TLHCPosition

FD

3

(0018,9558)

Detector Active Area Orientation

Detector​Active​Area​Orientation

FD

6

(0018,9559)

Positioner Primary Angle Direction

Positioner​Primary​Angle​Direction

CS

1

(0018,9601)

Diffusion b-matrix Sequence

Diffusion​B​Matrix​Sequence

SQ

1

(0018,9602)

Diffusion b-value XX

Diffusion​B​Value​XX

FD

1

(0018,9603)

Diffusion b-value XY

Diffusion​B​Value​XY

FD

1

(0018,9604)

Diffusion b-value XZ

Diffusion​B​Value​XZ

FD

1

(0018,9605)

Diffusion b-value YY

Diffusion​B​Value​YY

FD

1

(0018,9606)

Diffusion b-value YZ

Diffusion​B​Value​YZ

FD

1

(0018,9607)

Diffusion b-value ZZ

Diffusion​B​Value​ZZ

FD

1

(0018,9621)

Functional MR Sequence

Functional​MR​Sequence

SQ

1

(0018,9622)

Functional Settling Phase Frames Present

Functional​Settling​Phase​Frames​Present

CS

1

(0018,9623)

Functional Sync Pulse

Functional​Sync​Pulse

DT

1

(0018,9624)

Settling Phase Frame

Settling​Phase​Frame

CS

1

(0018,9701)

Decay Correction DateTime

Decay​Correction​Date​Time

DT

1

(0018,9715)

Start Density Threshold

Start​Density​Threshold

FD

1

(0018,9716)

Start Relative Density Difference Threshold

Start​Relative​Density​Difference​Threshold

FD

1

(0018,9717)

Start Cardiac Trigger Count Threshold

Start​Cardiac​Trigger​Count​Threshold

FD

1

(0018,9718)

Start Respiratory Trigger Count Threshold

Start​Respiratory​Trigger​Count​Threshold

FD

1

(0018,9719)

Termination Counts Threshold

Termination​Counts​Threshold

FD

1

(0018,9720)

Termination Density Threshold

Termination​Density​Threshold

FD

1

(0018,9721)

Termination Relative Density Threshold

Termination​Relative​Density​Threshold

FD

1

(0018,9722)

Termination Time Threshold

Termination​Time​Threshold

FD

1

(0018,9723)

Termination Cardiac Trigger Count Threshold

Termination​Cardiac​Trigger​Count​Threshold

FD

1

(0018,9724)

Termination Respiratory Trigger Count Threshold

Termination​Respiratory​Trigger​Count​Threshold

FD

1

(0018,9725)

Detector Geometry

Detector​Geometry

CS

1

(0018,9726)

Transverse Detector Separation

Transverse​Detector​Separation

FD

1

(0018,9727)

Axial Detector Dimension

Axial​Detector​Dimension

FD

1

(0018,9729)

Radiopharmaceutical Agent Number

Radio​pharmaceutical​Agent​Number

US

1

(0018,9732)

PET Frame Acquisition Sequence

PET​Frame​Acquisition​Sequence

SQ

1

(0018,9733)

PET Detector Motion Details Sequence

PET​Detector​Motion​Details​Sequence

SQ

1

(0018,9734)

PET Table Dynamics Sequence

PET​Table​Dynamics​Sequence

SQ

1

(0018,9735)

PET Position Sequence

PET​Position​Sequence

SQ

1

(0018,9736)

PET Frame Correction Factors Sequence

PET​Frame​Correction​Factors​Sequence

SQ

1

(0018,9737)

Radiopharmaceutical Usage Sequence

Radio​pharmaceutical​Usage​Sequence

SQ

1

(0018,9738)

Attenuation Correction Source

Attenuation​Correction​Source

CS

1

(0018,9739)

Number of Iterations

Number​Of​Iterations

US

1

(0018,9740)

Number of Subsets

Number​Of​Subsets

US

1

(0018,9749)

PET Reconstruction Sequence

PET​Reconstruction​Sequence

SQ

1

(0018,9751)

PET Frame Type Sequence

PET​Frame​Type​Sequence

SQ

1

(0018,9755)

Time of Flight Information Used

Time​Of​Flight​Information​Used

CS

1

(0018,9756)

Reconstruction Type

Reconstruction​Type

CS

1

(0018,9758)

Decay Corrected

Decay​Corrected

CS

1

(0018,9759)

Attenuation Corrected

Attenuation​Corrected

CS

1

(0018,9760)

Scatter Corrected

Scatter​Corrected

CS

1

(0018,9761)

Dead Time Corrected

Dead​Time​Corrected

CS

1

(0018,9762)

Gantry Motion Corrected

Gantry​Motion​Corrected

CS

1

(0018,9763)

Patient Motion Corrected

Patient​Motion​Corrected

CS

1

(0018,9764)

Count Loss Normalization Corrected

Count​Loss​Normalization​Corrected

CS

1

(0018,9765)

Randoms Corrected

Randoms​Corrected

CS

1

(0018,9766)

Non-uniform Radial Sampling Corrected

Non​Uniform​Radial​Sampling​Corrected

CS

1

(0018,9767)

Sensitivity Calibrated

Sensitivity​Calibrated

CS

1

(0018,9768)

Detector Normalization Correction

Detector​Normalization​Correction

CS

1

(0018,9769)

Iterative Reconstruction Method

Iterative​Reconstruction​Method

CS

1

(0018,9770)

Attenuation Correction Temporal Relationship

Attenuation​Correction​Temporal​Relationship

CS

1

(0018,9771)

Patient Physiological State Sequence

Patient​Physiological​State​Sequence

SQ

1

(0018,9772)

Patient Physiological State Code Sequence

Patient​Physiological​State​Code​Sequence

SQ

1

(0018,9801)

Depth(s) of Focus

Depths​Of​Focus

FD

1-n

(0018,9803)

Excluded Intervals Sequence

Excluded​Intervals​Sequence

SQ

1

(0018,9804)

Exclusion Start DateTime

Exclusion​Start​Date​Time

DT

1

(0018,9805)

Exclusion Duration

Exclusion​Duration

FD

1

(0018,9806)

US Image Description Sequence

US​Image​Description​Sequence

SQ

1

(0018,9807)

Image Data Type Sequence

Image​Data​Type​Sequence

SQ

1

(0018,9808)

Data Type

Data​Type

CS

1

(0018,9809)

Transducer Scan Pattern Code Sequence

Transducer​Scan​Pattern​Code​Sequence

SQ

1

(0018,980B)

Aliased Data Type

Aliased​Data​Type

CS

1

(0018,980C)

Position Measuring Device Used

Position​Measuring​Device​Used

CS

1

(0018,980D)

Transducer Geometry Code Sequence

Transducer​Geometry​Code​Sequence

SQ

1

(0018,980E)

Transducer Beam Steering Code Sequence

Transducer​Beam​Steering​Code​Sequence

SQ

1

(0018,980F)

Transducer Application Code Sequence

Transducer​Application​Code​Sequence

SQ

1

(0018,9810)

Zero Velocity Pixel Value

Zero​Velocity​Pixel​Value

US or SS

1

(0018,9821)

Photoacoustic Excitation Characteristics Sequence

Photoacoustic​Excitation​Characteristics​Sequence

SQ

1

(0018,9822)

Excitation Spectral Width

Excitation​Spectral​Width

FD

1

(0018,9823)

Excitation Energy

Excitation​Energy

FD

1

(0018,9824)

Excitation Pulse Duration

Excitation​Pulse​Duration

FD

1

(0018,9825)

Excitation Wavelength Sequence

Excitation​Wavelength​Sequence

SQ

1

(0018,9826)

Excitation Wavelength

Excitation​Wavelength

FD

1

(0018,9828)

Illumination Translation Flag

Illumination​Translation​Flag

CS

1

(0018,9829)

Acoustic Coupling Medium Flag

Acoustic​Coupling​Medium​Flag

CS

1

(0018,982A)

Acoustic Coupling Medium Code Sequence

Acoustic​Coupling​Medium​Code​Sequence

SQ

1

(0018,982B)

Acoustic Coupling Medium Temperature

Acoustic​Coupling​Medium​Temperature

FD

1

(0018,982C)

Transducer Response Sequence

Transducer​Response​Sequence

SQ

1

(0018,982D)

Center Frequency

Center​Frequency

FD

1

(0018,982E)

Fractional Bandwidth

Fractional​Bandwidth

FD

1

(0018,982F)

Lower Cutoff Frequency

Lower​Cutoff​Frequency

FD

1

(0018,9830)

Upper Cutoff Frequency

Upper​Cutoff​Frequency

FD

1

(0018,9831)

Transducer Technology Sequence

Transducer​Technology​Sequence

SQ

1

(0018,9832)

Sound Speed Correction Mechanism Code Sequence

Sound​Speed​Correction​Mechanism​Code​Sequence

SQ

1

(0018,9833)

Object Sound Speed

Object​Sound​Speed

FD

1

(0018,9834)

Acoustic Coupling Medium Sound Speed

Acoustic​Coupling​Medium​Sound​Speed

FD

1

(0018,9835)

Photoacoustic Image Frame Type Sequence

Photoacoustic​Image​Frame​Type​Sequence

SQ

1

(0018,9836)

Image Data Type Code Sequence

Image​Data​Type​Code​Sequence

SQ

1

(0018,9900)

Reference Location Label

Reference​Location​L​abel

LO

1

(0018,9901)

Reference Location Description

Reference​Location​Description

UT

1

(0018,9902)

Reference Basis Code Sequence

Reference​Basis​Code​Sequence

SQ

1

(0018,9903)

Reference Geometry Code Sequence

Reference​Geometry​Code​Sequence

SQ

1

(0018,9904)

Offset Distance

Offset​Distance

DS

1

(0018,9905)

Offset Direction

Offset​Direction

CS

1

(0018,9906)

Potential Scheduled Protocol Code Sequence

Potential​Scheduled​Protocol​Code​Sequence

SQ

1

(0018,9907)

Potential Requested Procedure Code Sequence

Potential​Requested​Procedure​Code​Sequence

SQ

1

(0018,9908)

Potential Reasons for Procedure

Potential​Reasons​For​Procedure

UC

1-n

(0018,9909)

Potential Reasons for Procedure Code Sequence

Potential​Reasons​For​Procedure​Code​Sequence

SQ

1

(0018,990A)

Potential Diagnostic Tasks

Potential​Diagnostic​Tasks

UC

1-n

(0018,990B)

Contraindications Code Sequence

Contraindications​Code​Sequence

SQ

1

(0018,990C)

Referenced Defined Protocol Sequence

Referenced​Defined​Protocol​Sequence

SQ

1

(0018,990D)

Referenced Performed Protocol Sequence

Referenced​Performed​Protocol​Sequence

SQ

1

(0018,990E)

Predecessor Protocol Sequence

Predecessor​Protocol​Sequence

SQ

1

(0018,990F)

Protocol Planning Information

Protocol​Planning​Information

UT

1

(0018,9910)

Protocol Design Rationale

Protocol​Design​Rationale

UT

1

(0018,9911)

Patient Specification Sequence

Patient​Specification​Sequence

SQ

1

(0018,9912)

Model Specification Sequence

Model​Specification​Sequence

SQ

1

(0018,9913)

Parameters Specification Sequence

Parameters​Specification​Sequence

SQ

1

(0018,9914)

Instruction Sequence

Instruction​Sequence

SQ

1

(0018,9915)

Instruction Index

Instruction​Index

US

1

(0018,9916)

Instruction Text

Instruction​Text

LO

1

(0018,9917)

Instruction Description

Instruction​Description

UT

1

(0018,9918)

Instruction Performed Flag

Instruction​Performed​Flag

CS

1

(0018,9919)

Instruction Performed DateTime

Instruction​Performed​Date​Time

DT

1

(0018,991A)

Instruction Performance Comment

Instruction​Performance​Comment

UT

1

(0018,991B)

Patient Positioning Instruction Sequence

Patient​Positioning​Instruction​Sequence

SQ

1

(0018,991C)

Positioning Method Code Sequence

Positioning​Method​Code​Sequence

SQ

1

(0018,991D)

Positioning Landmark Sequence

Positioning​Landmark​Sequence

SQ

1

(0018,991E)

Target Frame of Reference UID

Target​Frame​Of​Reference​UID

UI

1

(0018,991F)

Acquisition Protocol Element Specification Sequence

Acquisition​Protocol​Element​Specification​Sequence

SQ

1

(0018,9920)

Acquisition Protocol Element Sequence

Acquisition​Protocol​Element​Sequence

SQ

1

(0018,9921)

Protocol Element Number

Protocol​Element​Number

US

1

(0018,9922)

Protocol Element Name

Protocol​Element​Name

LO

1

(0018,9923)

Protocol Element Characteristics Summary

Protocol​Element​Characteristics​Summary

UT

1

(0018,9924)

Protocol Element Purpose

Protocol​Element​Purpose

UT

1

(0018,9930)

Acquisition Motion

Acquisition​Motion

CS

1

(0018,9931)

Acquisition Start Location Sequence

Acquisition​Start​Location​Sequence

SQ

1

(0018,9932)

Acquisition End Location Sequence

Acquisition​End​Location​Sequence

SQ

1

(0018,9933)

Reconstruction Protocol Element Specification Sequence

Reconstruction​Protocol​Element​Specification​Sequence

SQ

1

(0018,9934)

Reconstruction Protocol Element Sequence

Reconstruction​Protocol​Element​Sequence

SQ

1

(0018,9935)

Storage Protocol Element Specification Sequence

Storage​Protocol​Element​Specification​Sequence

SQ

1

(0018,9936)

Storage Protocol Element Sequence

Storage​Protocol​Element​Sequence

SQ

1

(0018,9937)

Requested Series Description

Requested​Series​Description

LO

1

(0018,9938)

Source Acquisition Protocol Element Number

Source​Acquisition​Protocol​Element​Number

US

1-n

(0018,9939)

Source Acquisition Beam Number

Source​Acquisition​Beam​Number

US

1-n

(0018,993A)

Source Reconstruction Protocol Element Number

Source​Reconstruction​Protocol​Element​Number

US

1-n

(0018,993B)

Reconstruction Start Location Sequence

Reconstruction​Start​Location​Sequence

SQ

1

(0018,993C)

Reconstruction End Location Sequence

Reconstruction​End​Location​Sequence

SQ

1

(0018,993D)

Reconstruction Algorithm Sequence

Reconstruction​Algorithm​Sequence

SQ

1

(0018,993E)

Reconstruction Target Center Location Sequence

Reconstruction​Target​Center​Location​Sequence

SQ

1

(0018,9941)

Image Filter Description

Image​Filter​Description

UT

1

(0018,9942)

CTDIvol Notification Trigger

CTDIvol​Notification​Trigger

FD

1

(0018,9943)

DLP Notification Trigger

DLP​Notification​Trigger

FD

1

(0018,9944)

Auto KVP Selection Type

Auto​KVP​Selection​Type

CS

1

(0018,9945)

Auto KVP Upper Bound

Auto​KVP​Upper​Bound

FD

1

(0018,9946)

Auto KVP Lower Bound

Auto​KVP​Lower​Bound

FD

1

(0018,9947)

Protocol Defined Patient Position

Protocol​Defined​Patient​Position

CS

1

(0018,A001)

Contributing Equipment Sequence

Contributing​Equipment​Sequence

SQ

1

(0018,A002)

Contribution DateTime

Contribution​Date​Time

DT

1

(0018,A003)

Contribution Description

Contribution​Description

ST

1

(0020,000D)

Study Instance UID

Study​Instance​UID

UI

1

(0020,000E)

Series Instance UID

Series​Instance​UID

UI

1

(0020,0010)

Study ID

Study​ID

SH

1

(0020,0011)

Series Number

Series​Number

IS

1

(0020,0012)

Acquisition Number

Acquisition​Number

IS

1

(0020,0013)

Instance Number

Instance​Number

IS

1

(0020,0014)

Isotope Number

Isotope​Number

IS

1

RET (1993)

(0020,0015)

Phase Number

Phase​Number

IS

1

RET (1993)

(0020,0016)

Interval Number

Interval​Number

IS

1

RET (1993)

(0020,0017)

Time Slot Number

Time​Slot​Number

IS

1

RET (1993)

(0020,0018)

Angle Number

Angle​Number

IS

1

RET (1993)

(0020,0019)

Item Number

Item​Number

IS

1

(0020,0020)

Patient Orientation

Patient​Orientation

CS

2

(0020,0022)

Overlay Number

Overlay​Number

IS

1

RET (2004)

(0020,0024)

Curve Number

Curve​Number

IS

1

RET (2004)

(0020,0026)

LUT Number

LUT​Number

IS

1

RET (2004)

(0020,0027)

Pyramid Label

PyramidLabel

LO

1

(0020,0030)

Image Position

Image​Position

DS

3

RET

(0020,0032)

Image Position (Patient)

Image​Position​Patient

DS

3

(0020,0035)

Image Orientation

Image​Orientation

DS

6

RET

(0020,0037)

Image Orientation (Patient)

Image​Orientation​Patient

DS

6

(0020,0050)

Location

Location

DS

1

RET

(0020,0052)

Frame of Reference UID

Frame​Of​Reference​UID

UI

1

(0020,0060)

Laterality

Laterality

CS

1

(0020,0062)

Image Laterality

Image​Laterality

CS

1

(0020,0070)

Image Geometry Type

Image​Geometry​Type

LO

1

RET

(0020,0080)

Masking Image

Masking​Image

CS

1-n

RET

(0020,00AA)

Report Number

Report​Number

IS

1

RET (2009)

(0020,0100)

Temporal Position Identifier

Temporal​Position​Identifier

IS

1

(0020,0105)

Number of Temporal Positions

Number​Of​Temporal​Positions

IS

1

(0020,0110)

Temporal Resolution

Temporal​Resolution

DS

1

(0020,0200)

Synchronization Frame of Reference UID

Synchronization​Frame​Of​Reference​UID

UI

1

(0020,0242)

SOP Instance UID of Concatenation Source

SOP​Instance​UID​Of​Concatenation​Source

UI

1

(0020,1000)

Series in Study

Series​In​Study

IS

1

RET (2004)

(0020,1001)

Acquisitions in Series

Acquisitions​In​Series

IS

1

RET

(0020,1002)

Images in Acquisition

Images​In​Acquisition

IS

1

(0020,1003)

Images in Series

Images​In​Series

IS

1

RET (2006)

(0020,1004)

Acquisitions in Study

Acquisitions​In​Study

IS

1

RET (2004)

(0020,1005)

Images in Study

Images​In​Study

IS

1

RET (2006)

(0020,1020)

Reference

Reference

LO

1-n

RET

(0020,103F)

Target Position Reference Indicator

Target​Position​Reference​Indicator

LO

1

(0020,1040)

Position Reference Indicator

Position​Reference​Indicator

LO

1

(0020,1041)

Slice Location

Slice​Location

DS

1

(0020,1070)

Other Study Numbers

Other​Study​Numbers

IS

1-n

RET (2004)

(0020,1200)

Number of Patient Related Studies

Number​Of​Patient​Related​Studies

IS

1

(0020,1202)

Number of Patient Related Series

Number​Of​Patient​Related​Series

IS

1

(0020,1204)

Number of Patient Related Instances

Number​Of​Patient​Related​Instances

IS

1

(0020,1206)

Number of Study Related Series

Number​Of​Study​Related​Series

IS

1

(0020,1208)

Number of Study Related Instances

Number​Of​Study​Related​Instances

IS

1

(0020,1209)

Number of Series Related Instances

Number​Of​Series​Related​Instances

IS

1

(0020,31xx)

Source Image IDs

Source​Image​IDs

CS

1-n

RET

(0020,3401)

Modifying Device ID

Modifying​Device​ID

CS

1

RET

(0020,3402)

Modified Image ID

Modified​Image​ID

CS

1

RET

(0020,3403)

Modified Image Date

Modified​Image​Date

DA

1

RET

(0020,3404)

Modifying Device Manufacturer

Modifying​Device​Manufacturer

LO

1

RET

(0020,3405)

Modified Image Time

Modified​Image​Time

TM

1

RET

(0020,3406)

Modified Image Description

Modified​Image​Description

LO

1

RET

(0020,4000)

Image Comments

Image​Comments

LT

1

(0020,5000)

Original Image Identification

Original​Image​Identification

AT

1-n

RET

(0020,5002)

Original Image Identification Nomenclature

Original​Image​Identification​Nomenclature

LO

1-n

RET

(0020,9056)

Stack ID

Stack​ID

SH

1

(0020,9057)

In-Stack Position Number

In​Stack​Position​Number

UL

1

(0020,9071)

Frame Anatomy Sequence

Frame​Anatomy​Sequence

SQ

1

(0020,9072)

Frame Laterality

Frame​Laterality

CS

1

(0020,9111)

Frame Content Sequence

Frame​Content​Sequence

SQ

1

(0020,9113)

Plane Position Sequence

Plane​Position​Sequence

SQ

1

(0020,9116)

Plane Orientation Sequence

Plane​Orientation​Sequence

SQ

1

(0020,9128)

Temporal Position Index

Temporal​Position​Index

UL

1

(0020,9153)

Nominal Cardiac Trigger Delay Time

Nominal​Cardiac​Trigger​Delay​Time

FD

1

(0020,9154)

Nominal Cardiac Trigger Time Prior To R-Peak

Nominal​Cardiac​Trigger​Time​Prior​To​R​Peak

FL

1

(0020,9155)

Actual Cardiac Trigger Time Prior To R-Peak

Actual​Cardiac​Trigger​Time​Prior​To​R​Peak

FL

1

(0020,9156)

Frame Acquisition Number

Frame​Acquisition​Number

US

1

(0020,9157)

Dimension Index Values

Dimension​Index​Values

UL

1-n

(0020,9158)

Frame Comments

Frame​Comments

LT

1

(0020,9161)

Concatenation UID

Concatenation​UID

UI

1

(0020,9162)

In-concatenation Number

In​Concatenation​Number

US

1

(0020,9163)

In-concatenation Total Number

In​Concatenation​Total​Number

US

1

(0020,9164)

Dimension Organization UID

Dimension​Organization​UID

UI

1

(0020,9165)

Dimension Index Pointer

Dimension​Index​Pointer

AT

1

(0020,9167)

Functional Group Pointer

Functional​Group​Pointer

AT

1

(0020,9170)

Unassigned Shared Converted Attributes Sequence

Unassigned​Shared​Converted​Attributes​Sequence

SQ

1

(0020,9171)

Unassigned Per-Frame Converted Attributes Sequence

Unassigned​Per​Frame​Converted​Attributes​Sequence

SQ

1

(0020,9172)

Conversion Source Attributes Sequence

Conversion​Source​Attributes​Sequence

SQ

1

(0020,9213)

Dimension Index Private Creator

Dimension​Index​Private​Creator

LO

1

(0020,9221)

Dimension Organization Sequence

Dimension​Organization​Sequence

SQ

1

(0020,9222)

Dimension Index Sequence

Dimension​Index​Sequence

SQ

1

(0020,9228)

Concatenation Frame Offset Number

Concatenation​Frame​Offset​Number

UL

1

(0020,9238)

Functional Group Private Creator

Functional​Group​Private​Creator

LO

1

(0020,9241)

Nominal Percentage of Cardiac Phase

Nominal​Percentage​Of​Cardiac​Phase

FL

1

(0020,9245)

Nominal Percentage of Respiratory Phase

Nominal​Percentage​Of​Respiratory​Phase

FL

1

(0020,9246)

Starting Respiratory Amplitude

Starting​Respiratory​Amplitude

FL

1

(0020,9247)

Starting Respiratory Phase

Starting​Respiratory​Phase

CS

1

(0020,9248)

Ending Respiratory Amplitude

Ending​Respiratory​Amplitude

FL

1

(0020,9249)

Ending Respiratory Phase

Ending​Respiratory​Phase

CS

1

(0020,9250)

Respiratory Trigger Type

Respiratory​Trigger​Type

CS

1

(0020,9251)

R-R Interval Time Nominal

RR​Interval​Time​Nominal

FD

1

(0020,9252)

Actual Cardiac Trigger Delay Time

Actual​Cardiac​Trigger​Delay​Time

FD

1

(0020,9253)

Respiratory Synchronization Sequence

Respiratory​Synchronization​Sequence

SQ

1

(0020,9254)

Respiratory Interval Time

Respiratory​Interval​Time

FD

1

(0020,9255)

Nominal Respiratory Trigger Delay Time

Nominal​Respiratory​Trigger​Delay​Time

FD

1

(0020,9256)

Respiratory Trigger Delay Threshold

Respiratory​Trigger​Delay​Threshold

FD

1

(0020,9257)

Actual Respiratory Trigger Delay Time

Actual​Respiratory​Trigger​Delay​Time

FD

1

(0020,9301)

Image Position (Volume)

Image​Position​Volume

FD

3

(0020,9302)

Image Orientation (Volume)

Image​Orientation​Volume

FD

6

(0020,9307)

Ultrasound Acquisition Geometry

Ultrasound​Acquisition​Geometry

CS

1

(0020,9308)

Apex Position

Apex​Position

FD

3

(0020,9309)

Volume to Transducer Mapping Matrix

Volume​To​Transducer​Mapping​Matrix

FD

16

(0020,930A)

Volume to Table Mapping Matrix

Volume​To​Table​Mapping​Matrix

FD

16

(0020,930B)

Volume to Transducer Relationship

Volume​To​Transducer​Relationship

CS

1

(0020,930C)

Patient Frame of Reference Source

Patient​Frame​Of​Reference​Source

CS

1

(0020,930D)

Temporal Position Time Offset

Temporal​Position​Time​Offset

FD

1

(0020,930E)

Plane Position (Volume) Sequence

Plane​Position​Volume​Sequence

SQ

1

(0020,930F)

Plane Orientation (Volume) Sequence

Plane​Orientation​Volume​Sequence

SQ

1

(0020,9310)

Temporal Position Sequence

Temporal​Position​Sequence

SQ

1

(0020,9311)

Dimension Organization Type

Dimension​Organization​Type

CS

1

(0020,9312)

Volume Frame of Reference UID

Volume​Frame​Of​Reference​UID

UI

1

(0020,9313)

Table Frame of Reference UID

Table​Frame​Of​Reference​UID

UI

1

(0020,9421)

Dimension Description Label

Dimension​Description​Label

LO

1

(0020,9450)

Patient Orientation in Frame Sequence

Patient​Orientation​In​Frame​Sequence

SQ

1

(0020,9453)

Frame Label

Frame​Label

LO

1

(0020,9518)

Acquisition Index

Acquisition​Index

US

1-n

(0020,9529)

Contributing SOP Instances Reference Sequence

Contributing​SOP​Instances​Reference​Sequence

SQ

1

(0020,9536)

Reconstruction Index

Reconstruction​Index

US

1

(0022,0001)

Light Path Filter Pass-Through Wavelength

Light​Path​Filter​Pass​Through​Wavelength

US

1

(0022,0002)

Light Path Filter Pass Band

Light​Path​Filter​Pass​Band

US

2

(0022,0003)

Image Path Filter Pass-Through Wavelength

Image​Path​Filter​Pass​Through​Wavelength

US

1

(0022,0004)

Image Path Filter Pass Band

Image​Path​Filter​Pass​Band

US

2

(0022,0005)

Patient Eye Movement Commanded

Patient​Eye​Movement​Commanded

CS

1

(0022,0006)

Patient Eye Movement Command Code Sequence

Patient​Eye​Movement​Command​Code​Sequence

SQ

1

(0022,0007)

Spherical Lens Power

Spherical​Lens​Power

FL

1

(0022,0008)

Cylinder Lens Power

Cylinder​Lens​Power

FL

1

(0022,0009)

Cylinder Axis

Cylinder​Axis

FL

1

(0022,000A)

Emmetropic Magnification

Emmetropic​Magnification

FL

1

(0022,000B)

Intra Ocular Pressure

Intra​Ocular​Pressure

FL

1

(0022,000C)

Horizontal Field of View

Horizontal​Field​Of​View

FL

1

(0022,000D)

Pupil Dilated

Pupil​Dilated

CS

1

(0022,000E)

Degree of Dilation

Degree​Of​Dilation

FL

1

(0022,000F)

Vertex Distance

Vertex​Distance

FD

1

(0022,0010)

Stereo Baseline Angle

Stereo​Baseline​Angle

FL

1

(0022,0011)

Stereo Baseline Displacement

Stereo​Baseline​Displacement

FL

1

(0022,0012)

Stereo Horizontal Pixel Offset

Stereo​Horizontal​Pixel​Offset

FL

1

(0022,0013)

Stereo Vertical Pixel Offset

Stereo​Vertical​Pixel​Offset

FL

1

(0022,0014)

Stereo Rotation

Stereo​Rotation

FL

1

(0022,0015)

Acquisition Device Type Code Sequence

Acquisition​Device​Type​Code​Sequence

SQ

1

(0022,0016)

Illumination Type Code Sequence

Illumination​Type​Code​Sequence

SQ

1

(0022,0017)

Light Path Filter Type Stack Code Sequence

Light​Path​Filter​Type​Stack​Code​Sequence

SQ

1

(0022,0018)

Image Path Filter Type Stack Code Sequence

Image​Path​Filter​Type​Stack​Code​Sequence

SQ

1

(0022,0019)

Lenses Code Sequence

Lenses​Code​Sequence

SQ

1

(0022,001A)

Channel Description Code Sequence

Channel​Description​Code​Sequence

SQ

1

(0022,001B)

Refractive State Sequence

Refractive​State​Sequence

SQ

1

(0022,001C)

Mydriatic Agent Code Sequence

Mydriatic​Agent​Code​Sequence

SQ

1

(0022,001D)

Relative Image Position Code Sequence

Relative​Image​Position​Code​Sequence

SQ

1

(0022,001E)

Camera Angle of View

Camera​Angle​Of​View

FL

1

(0022,0020)

Stereo Pairs Sequence

Stereo​Pairs​Sequence

SQ

1

(0022,0021)

Left Image Sequence

Left​Image​Sequence

SQ

1

(0022,0022)

Right Image Sequence

Right​Image​Sequence

SQ

1

(0022,0028)

Stereo Pairs Present

Stereo​Pairs​Present

CS

1

(0022,0030)

Axial Length of the Eye

Axial​Length​Of​The​Eye

FL

1

(0022,0031)

Ophthalmic Frame Location Sequence

Ophthalmic​Frame​Location​Sequence

SQ

1

(0022,0032)

Reference Coordinates

Reference​Coordinates

FL

2-2n

(0022,0035)

Depth Spatial Resolution

Depth​Spatial​Resolution

FL

1

(0022,0036)

Maximum Depth Distortion

Maximum​Depth​Distortion

FL

1

(0022,0037)

Along-scan Spatial Resolution

Along​Scan​Spatial​Resolution

FL

1

(0022,0038)

Maximum Along-scan Distortion

Maximum​Along​Scan​Distortion

FL

1

(0022,0039)

Ophthalmic Image Orientation

Ophthalmic​Image​Orientation

CS

1

(0022,0041)

Depth of Transverse Image

Depth​Of​Transverse​Image

FL

1

(0022,0042)

Mydriatic Agent Concentration Units Sequence

Mydriatic​Agent​Concentration​Units​Sequence

SQ

1

(0022,0048)

Across-scan Spatial Resolution

Across​Scan​Spatial​Resolution

FL

1

(0022,0049)

Maximum Across-scan Distortion

Maximum​Across​Scan​Distortion

FL

1

(0022,004E)

Mydriatic Agent Concentration

Mydriatic​Agent​Concentration

DS

1

(0022,0055)

Illumination Wave Length

Illumination​Wave​Length

FL

1

(0022,0056)

Illumination Power

Illumination​Power

FL

1

(0022,0057)

Illumination Bandwidth

Illumination​Bandwidth

FL

1

(0022,0058)

Mydriatic Agent Sequence

Mydriatic​Agent​Sequence

SQ

1

(0022,1007)

Ophthalmic Axial Measurements Right Eye Sequence

Ophthalmic​Axial​Measurements​Right​Eye​Sequence

SQ

1

(0022,1008)

Ophthalmic Axial Measurements Left Eye Sequence

Ophthalmic​Axial​Measurements​Left​Eye​Sequence

SQ

1

(0022,1009)

Ophthalmic Axial Measurements Device Type

Ophthalmic​Axial​Measurements​Device​Type

CS

1

(0022,1010)

Ophthalmic Axial Length Measurements Type

Ophthalmic​Axial​Length​Measurements​Type

CS

1

(0022,1012)

Ophthalmic Axial Length Sequence

Ophthalmic​Axial​Length​Sequence

SQ

1

(0022,1019)

Ophthalmic Axial Length

Ophthalmic​Axial​Length

FL

1

(0022,1024)

Lens Status Code Sequence

Lens​Status​Code​Sequence

SQ

1

(0022,1025)

Vitreous Status Code Sequence

Vitreous​Status​Code​Sequence

SQ

1

(0022,1028)

IOL Formula Code Sequence

IOL​Formula​Code​Sequence

SQ

1

(0022,1029)

IOL Formula Detail

IOL​Formula​Detail

LO

1

(0022,1033)

Keratometer Index

Keratometer​Index

FL

1

(0022,1035)

Source of Ophthalmic Axial Length Code Sequence

Source​Of​Ophthalmic​Axial​Length​Code​Sequence

SQ

1

(0022,1036)

Source of Corneal Size Data Code Sequence

Source​Of​Corneal​Size​Data​Code​Sequence

SQ

1

(0022,1037)

Target Refraction

Target​Refraction

FL

1

(0022,1039)

Refractive Procedure Occurred

Refractive​Procedure​Occurred

CS

1

(0022,1040)

Refractive Surgery Type Code Sequence

Refractive​Surgery​Type​Code​Sequence

SQ

1

(0022,1044)

Ophthalmic Ultrasound Method Code Sequence

Ophthalmic​Ultrasound​Method​Code​Sequence

SQ

1

(0022,1045)

Surgically Induced Astigmatism Sequence

Surgically​Induced​Astigmatism​Sequence

SQ

1

(0022,1046)

Type of Optical Correction

Type​Of​Optical​Correction

CS

1

(0022,1047)

Toric IOL Power Sequence

Toric​IOL​Power​Sequence

SQ

1

(0022,1048)

Predicted Toric Error Sequence

PredictedT​oric​Error​Sequence

SQ

1

(0022,1049)

Pre-Selected for Implantation

Pre​Selected​For​Implantation

CS

1

(0022,104A)

Toric IOL Power for Exact Emmetropia Sequence

Toric​IOL​Power​For​Exact​Emmetropia​Sequence

SQ

1

(0022,104B)

Toric IOL Power for Exact Target Refraction Sequence

Toric​IOL​Power​For​Exact​Target​Refraction​Sequence

SQ

1

(0022,1050)

Ophthalmic Axial Length Measurements Sequence

Ophthalmic​Axial​Length​Measurements​Sequence

SQ

1

(0022,1053)

IOL Power

IOL​Power

FL

1

(0022,1054)

Predicted Refractive Error

Predicted​Refractive​Error

FL

1

(0022,1059)

Ophthalmic Axial Length Velocity

Ophthalmic​Axial​Length​Velocity

FL

1

(0022,1065)

Lens Status Description

Lens​Status​Description

LO

1

(0022,1066)

Vitreous Status Description

Vitreous​Status​Description

LO

1

(0022,1090)

IOL Power Sequence

IOL​Power​Sequence

SQ

1

(0022,1092)

Lens Constant Sequence

Lens​Constant​Sequence

SQ

1

(0022,1093)

IOL Manufacturer

IOL​Manufacturer

LO

1

(0022,1094)

Lens Constant Description

Lens​Constant​Description

LO

1

RET (2011)

(0022,1095)

Implant Name

Implant​Name

LO

1

(0022,1096)

Keratometry Measurement Type Code Sequence

Keratometry​Measurement​Type​Code​Sequence

SQ

1

(0022,1097)

Implant Part Number

Implant​Part​Number

LO

1

(0022,1100)

Referenced Ophthalmic Axial Measurements Sequence

Referenced​Ophthalmic​Axial​Measurements​Sequence

SQ

1

(0022,1101)

Ophthalmic Axial Length Measurements Segment Name Code Sequence

Ophthalmic​Axial​Length​Measurements​Segment​Name​Code​Sequence

SQ

1

(0022,1103)

Refractive Error Before Refractive Surgery Code Sequence

Refractive​Error​Before​Refractive​Surgery​Code​Sequence

SQ

1

(0022,1121)

IOL Power For Exact Emmetropia

IOL​Power​For​Exact​Emmetropia

FL

1

(0022,1122)

IOL Power For Exact Target Refraction

IOL​Power​For​Exact​Target​Refraction

FL

1

(0022,1125)

Anterior Chamber Depth Definition Code Sequence

Anterior​Chamber​Depth​Definition​Code​Sequence

SQ

1

(0022,1127)

Lens Thickness Sequence

Lens​Thickness​Sequence

SQ

1

(0022,1128)

Anterior Chamber Depth Sequence

Anterior​Chamber​Depth​Sequence

SQ

1

(0022,112A)

Calculation Comment Sequence

Calculation​Comment​Sequence

SQ

1

(0022,112B)

Calculation Comment Type

Calculation​Comment​Type

CS

1

(0022,112C)

Calculation Comment

Calculation​Comment

LT

1

(0022,1130)

Lens Thickness

Lens​Thickness

FL

1

(0022,1131)

Anterior Chamber Depth

Anterior​Chamber​Depth

FL

1

(0022,1132)

Source of Lens Thickness Data Code Sequence

Source​Of​Lens​Thickness​Data​Code​Sequence

SQ

1

(0022,1133)

Source of Anterior Chamber Depth Data Code Sequence

Source​Of​Anterior​Chamber​Depth​Data​Code​Sequence

SQ

1

(0022,1134)

Source of Refractive Measurements Sequence

Source​Of​Refractive​Measurements​Sequence

SQ

1

(0022,1135)

Source of Refractive Measurements Code Sequence

Source​Of​Refractive​Measurements​Code​Sequence

SQ

1

(0022,1140)

Ophthalmic Axial Length Measurement Modified

Ophthalmic​Axial​Length​Measurement​Modified

CS

1

(0022,1150)

Ophthalmic Axial Length Data Source Code Sequence

Ophthalmic​Axial​Length​Data​Source​Code​Sequence

SQ

1

(0022,1153)

Ophthalmic Axial Length Acquisition Method Code Sequence

Ophthalmic​Axial​Length​Acquisition​Method​Code​Sequence

SQ

1

RET (2011)

(0022,1155)

Signal to Noise Ratio

Signal​To​Noise​Ratio

FL

1

(0022,1159)

Ophthalmic Axial Length Data Source Description

Ophthalmic​Axial​Length​Data​Source​Description

LO

1

(0022,1210)

Ophthalmic Axial Length Measurements Total Length Sequence

Ophthalmic​Axial​Length​Measurements​Total​Length​Sequence

SQ

1

(0022,1211)

Ophthalmic Axial Length Measurements Segmental Length Sequence

Ophthalmic​Axial​Length​Measurements​Segmental​Length​Sequence

SQ

1

(0022,1212)

Ophthalmic Axial Length Measurements Length Summation Sequence

Ophthalmic​Axial​Length​Measurements​Length​Summation​Sequence

SQ

1

(0022,1220)

Ultrasound Ophthalmic Axial Length Measurements Sequence

Ultrasound​Ophthalmic​Axial​Length​Measurements​Sequence

SQ

1

(0022,1225)

Optical Ophthalmic Axial Length Measurements Sequence

Optical​Ophthalmic​Axial​Length​Measurements​Sequence

SQ

1

(0022,1230)

Ultrasound Selected Ophthalmic Axial Length Sequence

Ultrasound​Selected​Ophthalmic​Axial​Length​Sequence

SQ

1

(0022,1250)

Ophthalmic Axial Length Selection Method Code Sequence

Ophthalmic​Axial​Length​Selection​Method​Code​Sequence

SQ

1

(0022,1255)

Optical Selected Ophthalmic Axial Length Sequence

Optical​Selected​Ophthalmic​Axial​Length​Sequence

SQ

1

(0022,1257)

Selected Segmental Ophthalmic Axial Length Sequence

Selected​Segmental​Ophthalmic​Axial​Length​Sequence

SQ

1

(0022,1260)

Selected Total Ophthalmic Axial Length Sequence

Selected​Total​Ophthalmic​Axial​Length​Sequence

SQ

1

(0022,1262)

Ophthalmic Axial Length Quality Metric Sequence

Ophthalmic​Axial​Length​Quality​Metric​Sequence

SQ

1

(0022,1265)

Ophthalmic Axial Length Quality Metric Type Code Sequence

Ophthalmic​Axial​Length​Quality​Metric​Type​Code​Sequence

SQ

1

RET

(0022,1273)

Ophthalmic Axial Length Quality Metric Type Description

Ophthalmic​Axial​Length​Quality​Metric​Type​Description

LO

1

RET (2011)

(0022,1300)

Intraocular Lens Calculations Right Eye Sequence

Intraocular​Lens​Calculations​Right​Eye​Sequence

SQ

1

(0022,1310)

Intraocular Lens Calculations Left Eye Sequence

Intraocular​Lens​Calculations​Left​Eye​Sequence

SQ

1

(0022,1330)

Referenced Ophthalmic Axial Length Measurement QC Image Sequence

Referenced​Ophthalmic​Axial​Length​Measurement​QC​Image​Sequence

SQ

1

(0022,1415)

Ophthalmic Mapping Device Type

Ophthalmic​Mapping​Device​Type

CS

1

(0022,1420)

Acquisition Method Code Sequence

Acquisition​Method​Code​Sequence

SQ

1

(0022,1423)

Acquisition Method Algorithm Sequence

Acquisition​Method​Algorithm​Sequence

SQ

1

(0022,1436)

Ophthalmic Thickness Map Type Code Sequence

Ophthalmic​Thickness​Map​Type​Code​Sequence

SQ

1

(0022,1443)

Ophthalmic Thickness Mapping Normals Sequence

Ophthalmic​Thickness​Mapping​Normals​Sequence

SQ

1

(0022,1445)

Retinal Thickness Definition Code Sequence

Retinal​Thickness​Definition​Code​Sequence

SQ

1

(0022,1450)

Pixel Value Mapping to Coded Concept Sequence

Pixel​Value​Mapping​To​Coded​Concept​Sequence

SQ

1

(0022,1452)

Mapped Pixel Value

Mapped​Pixel​Value

US or SS

1

(0022,1454)

Pixel Value Mapping Explanation

Pixel​Value​Mapping​Explanation

LO

1

(0022,1458)

Ophthalmic Thickness Map Quality Threshold Sequence

Ophthalmic​Thickness​Map​Quality​Threshold​Sequence

SQ

1

(0022,1460)

Ophthalmic Thickness Map Threshold Quality Rating

Ophthalmic​Thickness​Map​Threshold​Quality​Rating

FL

1

(0022,1463)

Anatomic Structure Reference Point

Anatomic​Structure​Reference​Point

FL

2

(0022,1465)

Registration to Localizer Sequence

Registration​To​Localizer​Sequence

SQ

1

(0022,1466)

Registered Localizer Units

Registered​Localizer​Units

CS

1

(0022,1467)

Registered Localizer Top Left Hand Corner

Registered​Localizer​Top​Left​Hand​Corner

FL

2

(0022,1468)

Registered Localizer Bottom Right Hand Corner

Registered​Localizer​Bottom​Right​Hand​Corner

FL

2

(0022,1470)

Ophthalmic Thickness Map Quality Rating Sequence

Ophthalmic​Thickness​Map​Quality​Rating​Sequence

SQ

1

(0022,1472)

Relevant OPT Attributes Sequence

Relevant​OPT​Attributes​Sequence

SQ

1

(0022,1512)

Transformation Method Code Sequence

Transformation​Method​Code​Sequence

SQ

1

(0022,1513)

Transformation Algorithm Sequence

Transformation​Algorithm​Sequence

SQ

1

(0022,1515)

Ophthalmic Axial Length Method

Ophthalmic​Axial​Length​Method

CS

1

(0022,1517)

Ophthalmic FOV

Ophthalmic​FOV

FL

1

(0022,1518)

Two Dimensional to Three Dimensional Map Sequence

Two​Dimensional​To​Three​Dimensional​Map​Sequence

SQ

1

(0022,1525)

Wide Field Ophthalmic Photography Quality Rating Sequence

Wide​Field​Ophthalmic​Photography​Quality​Rating​Sequence

SQ

1

(0022,1526)

Wide Field Ophthalmic Photography Quality Threshold Sequence

Wide​Field​Ophthalmic​Photography​Quality​Threshold​Sequence

SQ

1

(0022,1527)

Wide Field Ophthalmic Photography Threshold Quality Rating

Wide​Field​Ophthalmic​Photography​Threshold​Quality​Rating

FL

1

(0022,1528)

X Coordinates Center Pixel View Angle

X​Coordinates​Center​Pixel​View​Angle

FL

1

(0022,1529)

Y Coordinates Center Pixel View Angle

Y​Coordinates​Center​Pixel​View​Angle

FL

1

(0022,1530)

Number of Map Points

Number​Of​Map​Points

UL

1

(0022,1531)

Two Dimensional to Three Dimensional Map Data

Two​Dimensional​To​Three​Dimensional​Map​Data

OF

1

(0022,1612)

Derivation Algorithm Sequence

Derivation​Algorithm​Sequence

SQ

1

(0022,1615)

Ophthalmic Image Type Code Sequence

Ophthalmic​Image​Type​Code​Sequence

SQ

1

(0022,1616)

Ophthalmic Image Type Description

Ophthalmic​Image​Type​Description

LO

1

(0022,1618)

Scan Pattern Type Code Sequence

Scan​Pattern​Type​Code​Sequence

SQ

1

(0022,1620)

Referenced Surface Mesh Identification Sequence

Referenced​Surface​Mesh​Identification​Sequence

SQ

1

(0022,1622)

Ophthalmic Volumetric Properties Flag

Ophthalmic​Volumetric​Properties​Flag

CS

1

(0022,1623)

Ophthalmic Anatomic Reference Point Frame Coordinate

Ophthalmic​Anatomic​Reference​Point​Frame​Coordinate

FL

1

(0022,1624)

Ophthalmic Anatomic Reference Point X-Coordinate

Ophthalmic​Anatomic​Reference​Point​XCoordinate

FL

1

(0022,1626)

Ophthalmic Anatomic Reference Point Y-Coordinate

Ophthalmic​Anatomic​Reference​Point​YCoordinate

FL

1

(0022,1627)

Ophthalmic En Face Volume Descriptor Sequence

Ophthalmic​En​Face​Volume​Descriptor​Sequence

SQ

1

(0022,1628)

Ophthalmic En Face Image Quality Rating Sequence

Ophthalmic​En​Face​Image​Quality​Rating​Sequence

SQ

1

(0022,1629)

Ophthalmic En Face Volume Descriptor Scope

Ophthalmic​En​Face​Volume​Descriptor​Scope

CS

1

(0022,1630)

Quality Threshold

Quality​Threshold

DS

1

(0022,1632)

Ophthalmic Anatomic Reference Point Sequence

Ophthalmic​Anatomic​Reference​Point​Sequence

SQ

1

(0022,1633)

Ophthalmic Anatomic Reference Point Localization Type

Ophthalmic​Anatomic​Reference​Point​Localization​Type

CS

1

(0022,1634)

Primary Anatomic Structure Item Index

Primary​Anatomic​Structure​Item​Index

IS

1

(0022,1640)

OCT B-scan Analysis Acquisition Parameters Sequence

OCTBscan​Analysis​Acquisition​Parameters​Sequence

SQ

1

(0022,1642)

Number of B-scans Per Frame

Number​Of​Bscans​Per​Frame

UL

1

(0022,1643)

B-scan Slab Thickness

Bscan​Slab​Thickness

FL

1

(0022,1644)

Distance Between B-scan Slabs

Distance​Between​Bscan​Slabs

FL

1

(0022,1645)

B-scan Cycle Time

Bscan​Cycle​Time

FL

1

(0022,1646)

B-scan Cycle Time Vector

Bscan​Cycle​Time​Vector

FL

1-n

(0022,1649)

A-scan Rate

Ascan​Rate

FL

1

(0022,1650)

B-scan Rate

Bscan​Rate

FL

1

(0022,1658)

Surface Mesh Z-Pixel Offset

Surface​Mesh​ZPixel​Offset

UL

1

(0024,0010)

Visual Field Horizontal Extent

Visual​Field​Horizontal​Extent

FL

1

(0024,0011)

Visual Field Vertical Extent

Visual​Field​Vertical​Extent

FL

1

(0024,0012)

Visual Field Shape

Visual​Field​Shape

CS

1

(0024,0016)

Screening Test Mode Code Sequence

Screening​Test​Mode​Code​Sequence

SQ

1

(0024,0018)

Maximum Stimulus Luminance

Maximum​Stimulus​Luminance

FL

1

(0024,0020)

Background Luminance

Background​Luminance

FL

1

(0024,0021)

Stimulus Color Code Sequence

Stimulus​Color​Code​Sequence

SQ

1

(0024,0024)

Background Illumination Color Code Sequence

Background​Illumination​Color​Code​Sequence

SQ

1

(0024,0025)

Stimulus Area

Stimulus​Area

FL

1

(0024,0028)

Stimulus Presentation Time

Stimulus​Presentation​Time

FL

1

(0024,0032)

Fixation Sequence

Fixation​Sequence

SQ

1

(0024,0033)

Fixation Monitoring Code Sequence

Fixation​Monitoring​Code​Sequence

SQ

1

(0024,0034)

Visual Field Catch Trial Sequence

Visual​Field​Catch​Trial​Sequence

SQ

1

(0024,0035)

Fixation Checked Quantity

Fixation​Checked​Quantity

US

1

(0024,0036)

Patient Not Properly Fixated Quantity

Patient​Not​Properly​Fixated​Quantity

US

1

(0024,0037)

Presented Visual Stimuli Data Flag

Presented​Visual​Stimuli​Data​Flag

CS

1

(0024,0038)

Number of Visual Stimuli

Number​Of​Visual​Stimuli

US

1

(0024,0039)

Excessive Fixation Losses Data Flag

Excessive​Fixation​Losses​Data​Flag

CS

1

(0024,0040)

Excessive Fixation Losses

Excessive​Fixation​Losses

CS

1

(0024,0042)

Stimuli Retesting Quantity

Stimuli​Retesting​Quantity

US

1

(0024,0044)

Comments on Patient's Performance of Visual Field

Comments​On​Patient​Performance​Of​Visual​Field

LT

1

(0024,0045)

False Negatives Estimate Flag

False​Negatives​Estimate​Flag

CS

1

(0024,0046)

False Negatives Estimate

False​Negatives​Estimate

FL

1

(0024,0048)

Negative Catch Trials Quantity

Negative​Catch​Trials​Quantity

US

1

(0024,0050)

False Negatives Quantity

False​Negatives​Quantity

US

1

(0024,0051)

Excessive False Negatives Data Flag

Excessive​False​Negatives​Data​Flag

CS

1

(0024,0052)

Excessive False Negatives

Excessive​False​Negatives

CS

1

(0024,0053)

False Positives Estimate Flag

False​Positives​Estimate​Flag

CS

1

(0024,0054)

False Positives Estimate

False​Positives​Estimate

FL

1

(0024,0055)

Catch Trials Data Flag

Catch​Trials​Data​Flag

CS

1

(0024,0056)

Positive Catch Trials Quantity

Positive​Catch​Trials​Quantity

US

1

(0024,0057)

Test Point Normals Data Flag

Test​Point​Normals​Data​Flag

CS

1

(0024,0058)

Test Point Normals Sequence

Test​Point​Normals​Sequence

SQ

1

(0024,0059)

Global Deviation Probability Normals Flag

Global​Deviation​Probability​Normals​Flag

CS

1

(0024,0060)

False Positives Quantity

False​Positives​Quantity

US

1

(0024,0061)

Excessive False Positives Data Flag

Excessive​False​Positives​Data​Flag

CS

1

(0024,0062)

Excessive False Positives

Excessive​False​Positives

CS

1

(0024,0063)

Visual Field Test Normals Flag

Visual​Field​Test​Normals​Flag

CS

1

(0024,0064)

Results Normals Sequence

Results​Normals​Sequence

SQ

1

(0024,0065)

Age Corrected Sensitivity Deviation Algorithm Sequence

Age​Corrected​Sensitivity​Deviation​Algorithm​Sequence

SQ

1

(0024,0066)

Global Deviation From Normal

Global​Deviation​From​Normal

FL

1

(0024,0067)

Generalized Defect Sensitivity Deviation Algorithm Sequence

Generalized​Defect​Sensitivity​Deviation​Algorithm​Sequence

SQ

1

(0024,0068)

Localized Deviation From Normal

Localized​Deviation​From​Normal

FL

1

(0024,0069)

Patient Reliability Indicator

Patient​Reliability​Indicator

LO

1

(0024,0070)

Visual Field Mean Sensitivity

Visual​Field​Mean​Sensitivity

FL

1

(0024,0071)

Global Deviation Probability

Global​Deviation​Probability

FL

1

(0024,0072)

Local Deviation Probability Normals Flag

Local​Deviation​Probability​Normals​Flag

CS

1

(0024,0073)

Localized Deviation Probability

Localized​Deviation​Probability

FL

1

(0024,0074)

Short Term Fluctuation Calculated

Short​Term​Fluctuation​Calculated

CS

1

(0024,0075)

Short Term Fluctuation

Short​Term​Fluctuation

FL

1

(0024,0076)

Short Term Fluctuation Probability Calculated

Short​Term​Fluctuation​Probability​Calculated

CS

1

(0024,0077)

Short Term Fluctuation Probability

Short​Term​Fluctuation​Probability

FL

1

(0024,0078)

Corrected Localized Deviation From Normal Calculated

Corrected​Localized​Deviation​From​Normal​Calculated

CS

1

(0024,0079)

Corrected Localized Deviation From Normal

Corrected​Localized​Deviation​From​Normal

FL

1

(0024,0080)

Corrected Localized Deviation From Normal Probability Calculated

Corrected​Localized​Deviation​From​Normal​Probability​Calculated

CS

1

(0024,0081)

Corrected Localized Deviation From Normal Probability

Corrected​Localized​Deviation​From​Normal​Probability

FL

1

(0024,0083)

Global Deviation Probability Sequence

Global​Deviation​Probability​Sequence

SQ

1

(0024,0085)

Localized Deviation Probability Sequence

Localized​Deviation​Probability​Sequence

SQ

1

(0024,0086)

Foveal Sensitivity Measured

Foveal​Sensitivity​Measured

CS

1

(0024,0087)

Foveal Sensitivity

Foveal​Sensitivity

FL

1

(0024,0088)

Visual Field Test Duration

Visual​Field​Test​Duration

FL

1

(0024,0089)

Visual Field Test Point Sequence

Visual​Field​Test​Point​Sequence

SQ

1

(0024,0090)

Visual Field Test Point X-Coordinate

Visual​Field​Test​Point​X​Coordinate

FL

1

(0024,0091)

Visual Field Test Point Y-Coordinate

Visual​Field​Test​Point​Y​Coordinate

FL

1

(0024,0092)

Age Corrected Sensitivity Deviation Value

Age​Corrected​Sensitivity​Deviation​Value

FL

1

(0024,0093)

Stimulus Results

Stimulus​Results

CS

1

(0024,0094)

Sensitivity Value

Sensitivity​Value

FL

1

(0024,0095)

Retest Stimulus Seen

Retest​Stimulus​Seen

CS

1

(0024,0096)

Retest Sensitivity Value

Retest​Sensitivity​Value

FL

1

(0024,0097)

Visual Field Test Point Normals Sequence

Visual​Field​Test​Point​Normals​Sequence

SQ

1

(0024,0098)

Quantified Defect

Quantified​Defect

FL

1

(0024,0100)

Age Corrected Sensitivity Deviation Probability Value

Age​Corrected​Sensitivity​Deviation​Probability​Value

FL

1

(0024,0102)

Generalized Defect Corrected Sensitivity Deviation Flag

Generalized​Defect​Corrected​Sensitivity​Deviation​Flag

CS

1

(0024,0103)

Generalized Defect Corrected Sensitivity Deviation Value

Generalized​Defect​Corrected​Sensitivity​Deviation​Value

FL

1

(0024,0104)

Generalized Defect Corrected Sensitivity Deviation Probability Value

Generalized​Defect​Corrected​Sensitivity​Deviation​Probability​Value

FL

1

(0024,0105)

Minimum Sensitivity Value

Minimum​Sensitivity​Value

FL

1

(0024,0106)

Blind Spot Localized

Blind​Spot​Localized

CS

1

(0024,0107)

Blind Spot X-Coordinate

Blind​Spot​X​Coordinate

FL

1

(0024,0108)

Blind Spot Y-Coordinate

Blind​Spot​Y​Coordinate

FL

1

(0024,0110)

Visual Acuity Measurement Sequence

Visual​Acuity​Measurement​Sequence

SQ

1

(0024,0112)

Refractive Parameters Used on Patient Sequence

Refractive​Parameters​Used​On​Patient​Sequence

SQ

1

(0024,0113)

Measurement Laterality

Measurement​Laterality

CS

1

(0024,0114)

Ophthalmic Patient Clinical Information Left Eye Sequence

Ophthalmic​Patient​Clinical​Information​Left​Eye​Sequence

SQ

1

(0024,0115)

Ophthalmic Patient Clinical Information Right Eye Sequence

Ophthalmic​Patient​Clinical​Information​Right​Eye​Sequence

SQ

1

(0024,0117)

Foveal Point Normative Data Flag

Foveal​Point​Normative​Data​Flag

CS

1

(0024,0118)

Foveal Point Probability Value

Foveal​Point​Probability​Value

FL

1

(0024,0120)

Screening Baseline Measured

Screening​Baseline​Measured

CS

1

(0024,0122)

Screening Baseline Measured Sequence

Screening​Baseline​Measured​Sequence

SQ

1

(0024,0124)

Screening Baseline Type

Screening​Baseline​Type

CS

1

(0024,0126)

Screening Baseline Value

Screening​Baseline​Value

FL

1

(0024,0202)

Algorithm Source

Algorithm​Source

LO

1

(0024,0306)

Data Set Name

Data​Set​Name

LO

1

(0024,0307)

Data Set Version

Data​Set​Version

LO

1

(0024,0308)

Data Set Source

Data​Set​Source

LO

1

(0024,0309)

Data Set Description

Data​Set​Description

LO

1

(0024,0317)

Visual Field Test Reliability Global Index Sequence

Visual​Field​Test​Reliability​Global​Index​Sequence

SQ

1

(0024,0320)

Visual Field Global Results Index Sequence

Visual​Field​Global​Results​Index​Sequence

SQ

1

(0024,0325)

Data Observation Sequence

Data​Observation​Sequence

SQ

1

(0024,0338)

Index Normals Flag

Index​Normals​Flag

CS

1

(0024,0341)

Index Probability

Index​Probability

FL

1

(0024,0344)

Index Probability Sequence

Index​Probability​Sequence

SQ

1

(0028,0002)

Samples per Pixel

Samples​Per​Pixel

US

1

(0028,0003)

Samples per Pixel Used

Samples​Per​Pixel​Used

US

1

(0028,0004)

Photometric Interpretation

Photometric​Interpretation

CS

1

(0028,0005)

Image Dimensions

Image​Dimensions

US

1

RET

(0028,0006)

Planar Configuration

Planar​Configuration

US

1

(0028,0008)

Number of Frames

Number​Of​Frames

IS

1

(0028,0009)

Frame Increment Pointer

Frame​Increment​Pointer

AT

1-n

(0028,000A)

Frame Dimension Pointer

Frame​Dimension​Pointer

AT

1-n

(0028,0010)

Rows

Rows

US

1

(0028,0011)

Columns

Columns

US

1

(0028,0012)

Planes

Planes

US

1

RET (2007)

(0028,0014)

Ultrasound Color Data Present

Ultrasound​Color​Data​Present

US

1

(0028,0020)

RET (2007) - See Note 3

(0028,0030)

Pixel Spacing

Pixel​Spacing

DS

2

(0028,0031)

Zoom Factor

Zoom​Factor

DS

2

(0028,0032)

Zoom Center

Zoom​Center

DS

2

(0028,0034)

Pixel Aspect Ratio

Pixel​Aspect​Ratio

IS

2

(0028,0040)

Image Format

Image​Format

CS

1

RET

(0028,0050)

Manipulated Image

Manipulated​Image

LO

1-n

RET

(0028,0051)

Corrected Image

Corrected​Image

CS

1-n

(0028,005F)

Compression Recognition Code

Compression​Recognition​Code

LO

1

RET (2007)

(0028,0060)

Compression Code

Compression​Code

CS

1

RET

(0028,0061)

Compression Originator

Compression​Originator

SH

1

RET (2007)

(0028,0062)

Compression Label

Compression​Label

LO

1

RET (2007)

(0028,0063)

Compression Description

Compression​Description

SH

1

RET (2007)

(0028,0065)

Compression Sequence

Compression​Sequence

CS

1-n

RET (2007)

(0028,0066)

Compression Step Pointers

Compression​Step​Pointers

AT

1-n

RET (2007)

(0028,0068)

Repeat Interval

Repeat​Interval

US

1

RET (2007)

(0028,0069)

Bits Grouped

Bits​Grouped

US

1

RET (2007)

(0028,0070)

Perimeter Table

Perimeter​Table

US

1-n

RET (2007)

(0028,0071)

Perimeter Value

Perimeter​Value

US or SS

1

RET (2007)

(0028,0080)

Predictor Rows

Predictor​Rows

US

1

RET (2007)

(0028,0081)

Predictor Columns

Predictor​Columns

US

1

RET (2007)

(0028,0082)

Predictor Constants

Predictor​Constants

US

1-n

RET (2007)

(0028,0090)

Blocked Pixels

Blocked​Pixels

CS

1

RET (2007)

(0028,0091)

Block Rows

Block​Rows

US

1

RET (2007)

(0028,0092)

Block Columns

Block​Columns

US

1

RET (2007)

(0028,0093)

Row Overlap

Row​Overlap

US

1

RET (2007)

(0028,0094)

Column Overlap

Column​Overlap

US

1

RET (2007)

(0028,0100)

Bits Allocated

Bits​Allocated

US

1

(0028,0101)

Bits Stored

Bits​Stored

US

1

(0028,0102)

High Bit

High​Bit

US

1

(0028,0103)

Pixel Representation

Pixel​Representation

US

1

(0028,0104)

Smallest Valid Pixel Value

Smallest​Valid​Pixel​Value

US or SS

1

RET

(0028,0105)

Largest Valid Pixel Value

Largest​Valid​Pixel​Value

US or SS

1

RET

(0028,0106)

Smallest Image Pixel Value

Smallest​Image​Pixel​Value

US or SS

1

(0028,0107)

Largest Image Pixel Value

Largest​Image​Pixel​Value

US or SS

1

(0028,0108)

Smallest Pixel Value in Series

Smallest​Pixel​Value​In​Series

US or SS

1

(0028,0109)

Largest Pixel Value in Series

Largest​Pixel​Value​In​Series

US or SS

1

(0028,0110)

Smallest Image Pixel Value in Plane

Smallest​Image​Pixel​Value​In​Plane

US or SS

1

RET (2007)

(0028,0111)

Largest Image Pixel Value in Plane

Largest​Image​Pixel​Value​In​Plane

US or SS

1

RET (2007)

(0028,0120)

Pixel Padding Value

Pixel​Padding​Value

US or SS

1

(0028,0121)

Pixel Padding Range Limit

Pixel​Padding​Range​Limit

US or SS

1

(0028,0122)

Float Pixel Padding Value

Float​Pixel​Padding​Value

FL

1

(0028,0123)

Double Float Pixel Padding Value

Double​Float​Pixel​Padding​Value

FD

1

(0028,0124)

Float Pixel Padding Range Limit

Float​Pixel​Padding​Range​Limit

FL

1

(0028,0125)

Double Float Pixel Padding Range Limit

Double​Float​Pixel​Padding​Range​Limit

FD

1

(0028,0200)

Image Location

Image​Location

US

1

RET

(0028,0300)

Quality Control Image

Quality​Control​Image

CS

1

(0028,0301)

Burned In Annotation

Burned​In​Annotation

CS

1

(0028,0302)

Recognizable Visual Features

Recognizable​Visual​Features

CS

1

(0028,0303)

Longitudinal Temporal Information Modified

Longitudinal​Temporal​Information​Modified

CS

1

(0028,0304)

Referenced Color Palette Instance UID

Referenced​Color​Palette​Instance​UID

UI

1

(0028,0400)

Transform Label

Transform​Label

LO

1

RET (2007)

(0028,0401)

Transform Version Number

Transform​Version​Number

LO

1

RET (2007)

(0028,0402)

Number of Transform Steps

Number​Of​Transform​Steps

US

1

RET (2007)

(0028,0403)

Sequence of Compressed Data

Sequence​Of​Compressed​Data

LO

1-n

RET (2007)

(0028,0404)

Details of Coefficients

Details​Of​Coefficients

AT

1-n

RET (2007)

(0028,04x0)

Rows For Nth Order Coefficients

Rows​For​Nth​Order​Coefficients

US

1

RET (2007)

(0028,04x1)

Columns For Nth Order Coefficients

Columns​For​Nth​Order​Coefficients

US

1

RET (2007)

(0028,04x2)

Coefficient Coding

Coefficient​Coding

LO

1-n

RET (2007)

(0028,04x3)

Coefficient Coding Pointers

Coefficient​Coding​Pointers

AT

1-n

RET (2007)

(0028,0700)

DCT Label

DCT​Label

LO

1

RET (2007)

(0028,0701)

Data Block Description

Data​Block​Description

CS

1-n

RET (2007)

(0028,0702)

Data Block

Data​Block

AT

1-n

RET (2007)

(0028,0710)

Normalization Factor Format

Normalization​Factor​Format

US

1

RET (2007)

(0028,0720)

Zonal Map Number Format

Zonal​Map​Number​Format

US

1

RET (2007)

(0028,0721)

Zonal Map Location

Zonal​Map​Location

AT

1-n

RET (2007)

(0028,0722)

Zonal Map Format

Zonal​Map​Format

US

1

RET (2007)

(0028,0730)

Adaptive Map Format

Adaptive​Map​Format

US

1

RET (2007)

(0028,0740)

Code Number Format

Code​Number​Format

US

1

RET (2007)

(0028,08x0)

Code Label

Code​Label

CS

1-n

RET (2007)

(0028,08x2)

Number of Tables

Number​Of​Tables

US

1

RET (2007)

(0028,08x3)

Code Table Location

Code​Table​Location

AT

1-n

RET (2007)

(0028,08x4)

Bits For Code Word

Bits​For​Code​Word

US

1

RET (2007)

(0028,08x8)

Image Data Location

Image​Data​Location

AT

1-n

RET (2007)

(0028,0A02)

Pixel Spacing Calibration Type

Pixel​Spacing​Calibration​Type

CS

1

(0028,0A04)

Pixel Spacing Calibration Description

Pixel​Spacing​Calibration​Description

LO

1

(0028,1040)

Pixel Intensity Relationship

Pixel​Intensity​Relationship

CS

1

(0028,1041)

Pixel Intensity Relationship Sign

Pixel​Intensity​Relationship​Sign

SS

1

(0028,1050)

Window Center

Window​Center

DS

1-n

(0028,1051)

Window Width

Window​Width

DS

1-n

(0028,1052)

Rescale Intercept

Rescale​Intercept

DS

1

(0028,1053)

Rescale Slope

Rescale​Slope

DS

1

(0028,1054)

Rescale Type

Rescale​Type

LO

1

(0028,1055)

Window Center & Width Explanation

Window​Center​Width​Explanation

LO

1-n

(0028,1056)

VOI LUT Function

VOILUT​Function

CS

1

(0028,1080)

Gray Scale

Gray​Scale

CS

1

RET

(0028,1090)

Recommended Viewing Mode

Recommended​Viewing​Mode

CS

1

(0028,1100)

Gray Lookup Table Descriptor

Gray​Lookup​Table​Descriptor

US or SS

3

RET

(0028,1101)

Red Palette Color Lookup Table Descriptor

Red​Palette​Color​Lookup​Table​Descriptor

US or SS

3

(0028,1102)

Green Palette Color Lookup Table Descriptor

Green​Palette​Color​Lookup​Table​Descriptor

US or SS

3

(0028,1103)

Blue Palette Color Lookup Table Descriptor

Blue​Palette​Color​Lookup​Table​Descriptor

US or SS

3

(0028,1104)

Alpha Palette Color Lookup Table Descriptor

Alpha​Palette​Color​Lookup​Table​Descriptor

US

3

(0028,1111)

Large Red Palette Color Lookup Table Descriptor

Large​Red​Palette​Color​Lookup​Table​Descriptor

US or SS

4

RET (2007)

(0028,1112)

Large Green Palette Color Lookup Table Descriptor

Large​Green​Palette​Color​Lookup​Table​Descriptor

US or SS

4

RET (2007)

(0028,1113)

Large Blue Palette Color Lookup Table Descriptor

Large​Blue​Palette​Color​Lookup​Table​Descriptor

US or SS

4

RET (2007)

(0028,1199)

Palette Color Lookup Table UID

Palette​Color​Lookup​Table​UID

UI

1

(0028,1200)

Gray Lookup Table Data

Gray​Lookup​Table​Data

US or SS or OW

1-n or 1

RET

(0028,1201)

Red Palette Color Lookup Table Data

Red​Palette​Color​Lookup​Table​Data

OW

1

(0028,1202)

Green Palette Color Lookup Table Data

Green​Palette​Color​Lookup​Table​Data

OW

1

(0028,1203)

Blue Palette Color Lookup Table Data

Blue​Palette​Color​Lookup​Table​Data

OW

1

(0028,1204)

Alpha Palette Color Lookup Table Data

Alpha​Palette​Color​Lookup​Table​Data

OW

1

(0028,1211)

Large Red Palette Color Lookup Table Data

Large​Red​Palette​Color​Lookup​Table​Data

OW

1

RET (2007)

(0028,1212)

Large Green Palette Color Lookup Table Data

Large​Green​Palette​Color​Lookup​Table​Data

OW

1

RET (2007)

(0028,1213)

Large Blue Palette Color Lookup Table Data

Large​Blue​Palette​Color​Lookup​Table​Data

OW

1

RET (2007)

(0028,1214)

Large Palette Color Lookup Table UID

Large​Palette​Color​Lookup​Table​UID

UI

1

RET (2007)

(0028,1221)

Segmented Red Palette Color Lookup Table Data

Segmented​Red​Palette​Color​Lookup​Table​Data

OW

1

(0028,1222)

Segmented Green Palette Color Lookup Table Data

Segmented​Green​Palette​Color​Lookup​Table​Data

OW

1

(0028,1223)

Segmented Blue Palette Color Lookup Table Data

Segmented​Blue​Palette​Color​Lookup​Table​Data

OW

1

(0028,1224)

Segmented Alpha Palette Color Lookup Table Data

Segmented​Alpha​Palette​Color​Lookup​Table​Data

OW

1

(0028,1230)

Stored Value Color Range Sequence

Stored​Value​Color​Range​Sequence

SQ

1

(0028,1231)

Minimum Stored Value Mapped

Minimum​Stored​Value​Mapped

FD

1

(0028,1232)

Maximum Stored Value Mapped

Maximum​Stored​Value​Mapped

FD

1

(0028,1300)

Breast Implant Present

Breast​Implant​Present

CS

1

(0028,1350)

Partial View

Partial​View

CS

1

(0028,1351)

Partial View Description

Partial​View​Description

ST

1

(0028,1352)

Partial View Code Sequence

Partial​View​Code​Sequence

SQ

1

(0028,135A)

Spatial Locations Preserved

Spatial​Locations​Preserved

CS

1

(0028,1401)

Data Frame Assignment Sequence

Data​Frame​Assignment​Sequence

SQ

1

(0028,1402)

Data Path Assignment

Data​Path​Assignment

CS

1

(0028,1403)

Bits Mapped to Color Lookup Table

Bits​Mapped​To​Color​Lookup​Table

US

1

(0028,1404)

Blending LUT 1 Sequence

Blending​LUT1Sequence

SQ

1

(0028,1405)

Blending LUT 1 Transfer Function

Blending​LUT1Transfer​Function

CS

1

(0028,1406)

Blending Weight Constant

Blending​Weight​Constant

FD

1

(0028,1407)

Blending Lookup Table Descriptor

Blending​Lookup​Table​Descriptor

US

3

(0028,1408)

Blending Lookup Table Data

Blending​Lookup​Table​Data

OW

1

(0028,140B)

Enhanced Palette Color Lookup Table Sequence

Enhanced​Palette​Color​Lookup​Table​Sequence

SQ

1

(0028,140C)

Blending LUT 2 Sequence

Blending​LUT2Sequence

SQ

1

(0028,140D)

Blending LUT 2 Transfer Function

Blending​LUT2Transfer​Function

CS

1

(0028,140E)

Data Path ID

Data​Path​ID

CS

1

(0028,140F)

RGB LUT Transfer Function

RGBLUT​Transfer​Function

CS

1

(0028,1410)

Alpha LUT Transfer Function

Alpha​LUT​Transfer​Function

CS

1

(0028,2000)

ICC Profile

ICC​Profile

OB

1

(0028,2002)

Color Space

Color​Space

CS

1

(0028,2110)

Lossy Image Compression

Lossy​Image​Compression

CS

1

(0028,2112)

Lossy Image Compression Ratio

Lossy​Image​Compression​Ratio

DS

1-n

(0028,2114)

Lossy Image Compression Method

Lossy​Image​Compression​Method

CS

1-n

(0028,3000)

Modality LUT Sequence

Modality​LUT​Sequence

SQ

1

(0028,3001)

Variable Modality LUT Sequence

Variable​Modality​LUT​Sequence

SQ

1

(0028,3002)

LUT Descriptor

LUT​Descriptor

US or SS

3

(0028,3003)

LUT Explanation

LUT​Explanation

LO

1

(0028,3004)

Modality LUT Type

Modality​LUT​Type

LO

1

(0028,3006)

LUT Data

LUT​Data

US or OW

1-n or 1

(0028,3010)

VOI LUT Sequence

VOILUT​Sequence

SQ

1

(0028,3110)

Softcopy VOI LUT Sequence

Softcopy​VOILUT​Sequence

SQ

1

(0028,4000)

Image Presentation Comments

Image​Presentation​Comments

LT

1

RET

(0028,5000)

Bi-Plane Acquisition Sequence

Bi​Plane​Acquisition​Sequence

SQ

1

RET (2007)

(0028,6010)

Representative Frame Number

Representative​Frame​Number

US

1

(0028,6020)

Frame Numbers of Interest (FOI)

Frame​Numbers​Of​Interest

US

1-n

(0028,6022)

Frame of Interest Description

Frame​Of​Interest​Description

LO

1-n

(0028,6023)

Frame of Interest Type

Frame​Of​Interest​Type

CS

1-n

(0028,6030)

Mask Pointer(s)

Mask​Pointers

US

1-n

RET (2001)

(0028,6040)

R Wave Pointer

R​Wave​Pointer

US

1-n

(0028,6100)

Mask Subtraction Sequence

Mask​Subtraction​Sequence

SQ

1

(0028,6101)

Mask Operation

Mask​Operation

CS

1

(0028,6102)

Applicable Frame Range

Applicable​Frame​Range

US

2-2n

(0028,6110)

Mask Frame Numbers

Mask​Frame​Numbers

US

1-n

(0028,6112)

Contrast Frame Averaging

Contrast​Frame​Averaging

US

1

(0028,6114)

Mask Sub-pixel Shift

Mask​Sub​Pixel​Shift

FL

2

(0028,6120)

TID Offset

TID​Offset

SS

1

(0028,6190)

Mask Operation Explanation

Mask​Operation​Explanation

ST

1

(0028,7000)

Equipment Administrator Sequence

Equipment​Administrator​Sequence

SQ

1

(0028,7001)

Number of Display Subsystems

Number​Of​Display​Subsystems

US

1

(0028,7002)

Current Configuration ID

Current​Configuration​ID

US

1

(0028,7003)

Display Subsystem ID

Display​Subsystem​ID

US

1

(0028,7004)

Display Subsystem Name

Display​Subsystem​Name

SH

1

(0028,7005)

Display Subsystem Description

Display​Subsystem​Description

LO

1

(0028,7006)

System Status

System​Status

CS

1

(0028,7007)

System Status Comment

System​Status​Comment

LO

1

(0028,7008)

Target Luminance Characteristics Sequence

Target​Luminance​Characteristics​Sequence

SQ

1

(0028,7009)

Luminance Characteristics ID

Luminance​Characteristics​ID

US

1

(0028,700A)

Display Subsystem Configuration Sequence

Display​Subsystem​Configuration​Sequence

SQ

1

(0028,700B)

Configuration ID

Configuration​ID

US

1

(0028,700C)

Configuration Name

Configuration​Name

SH

1

(0028,700D)

Configuration Description

Configuration​Description

LO

1

(0028,700E)

Referenced Target Luminance Characteristics ID

Referenced​Target​Luminance​Characteristics​ID

US

1

(0028,700F)

QA Results Sequence

QAResults​Sequence

SQ

1

(0028,7010)

Display Subsystem QA Results Sequence

Display​Subsystem​QAResults​Sequence

SQ

1

(0028,7011)

Configuration QA Results Sequence

Configuration​QAResults​Sequence

SQ

1

(0028,7012)

Measurement Equipment Sequence

Measurement​Equipment​Sequence

SQ

1

(0028,7013)

Measurement Functions

Measurement​Functions

CS

1-n

(0028,7014)

Measurement Equipment Type

Measurement​Equipment​Type

CS

1

(0028,7015)

Visual Evaluation Result Sequence

Visual​Evaluation​Result​Sequence

SQ

1

(0028,7016)

Display Calibration Result Sequence

Display​Calibration​Result​Sequence

SQ

1

(0028,7017)

DDL Value

DDLValue

US

1

(0028,7018)

CIExy White Point

CIExy​White​Point

FL

2

(0028,7019)

Display Function Type

Display​Function​Type

CS

1

(0028,701A)

Gamma Value

Gamma​Value

FL

1

(0028,701B)

Number of Luminance Points

Number​Of​Luminance​Points

US

1

(0028,701C)

Luminance Response Sequence

Luminance​Response​Sequence

SQ

1

(0028,701D)

Target Minimum Luminance

Target​Minimum​Luminance

FL

1

(0028,701E)

Target Maximum Luminance

Target​Maximum​Luminance

FL

1

(0028,701F)

Luminance Value

Luminance​Value

FL

1

(0028,7020)

Luminance Response Description

Luminance​Response​Description

LO

1

(0028,7021)

White Point Flag

White​Point​Flag

CS

1

(0028,7022)

Display Device Type Code Sequence

Display​Device​Type​Code​Sequence

SQ

1

(0028,7023)

Display Subsystem Sequence

Display​Subsystem​Sequence

SQ

1

(0028,7024)

Luminance Result Sequence

Luminance​Result​Sequence

SQ

1

(0028,7025)

Ambient Light Value Source

Ambient​Light​Value​Source

CS

1

(0028,7026)

Measured Characteristics

Measured​Characteristics

CS

1-n

(0028,7027)

Luminance Uniformity Result Sequence

Luminance​Uniformity​Result​Sequence

SQ

1

(0028,7028)

Visual Evaluation Test Sequence

Visual​Evaluation​Test​Sequence

SQ

1

(0028,7029)

Test Result

Test​Result

CS

1

(0028,702A)

Test Result Comment

Test​Result​Comment

LO

1

(0028,702B)

Test Image Validation

Test​Image​Validation

CS

1

(0028,702C)

Test Pattern Code Sequence

Test​Pattern​Code​Sequence

SQ

1

(0028,702D)

Measurement Pattern Code Sequence

Measurement​Pattern​Code​Sequence

SQ

1

(0028,702E)

Visual Evaluation Method Code Sequence

Visual​Evaluation​Method​Code​Sequence

SQ

1

(0028,7FE0)

Pixel Data Provider URL

Pixel​Data​Provider​URL

UR

1

(0028,9001)

Data Point Rows

Data​Point​Rows

UL

1

(0028,9002)

Data Point Columns

Data​Point​Columns

UL

1

(0028,9003)

Signal Domain Columns

Signal​Domain​Columns

CS

1

(0028,9099)

Largest Monochrome Pixel Value

Largest​Monochrome​Pixel​Value

US

1

RET (2003)

(0028,9108)

Data Representation

Data​Representation

CS

1

(0028,9110)

Pixel Measures Sequence

Pixel​Measures​Sequence

SQ

1

(0028,9132)

Frame VOI LUT Sequence

Frame​VOILUT​Sequence

SQ

1

(0028,9145)

Pixel Value Transformation Sequence

Pixel​Value​Transformation​Sequence

SQ

1

(0028,9235)

Signal Domain Rows

Signal​Domain​Rows

CS

1

(0028,9411)

Display Filter Percentage

Display​Filter​Percentage

FL

1

(0028,9415)

Frame Pixel Shift Sequence

Frame​Pixel​Shift​Sequence

SQ

1

(0028,9416)

Subtraction Item ID

Subtraction​Item​ID

US

1

(0028,9422)

Pixel Intensity Relationship LUT Sequence

Pixel​Intensity​Relationship​LUT​Sequence

SQ

1

(0028,9443)

Frame Pixel Data Properties Sequence

Frame​Pixel​Data​Properties​Sequence

SQ

1

(0028,9444)

Geometrical Properties

Geometrical​Properties

CS

1

(0028,9445)

Geometric Maximum Distortion

Geometric​Maximum​Distortion

FL

1

(0028,9446)

Image Processing Applied

Image​Processing​Applied

CS

1-n

(0028,9454)

Mask Selection Mode

Mask​Selection​Mode

CS

1

(0028,9474)

LUT Function

LUT​Function

CS

1

(0028,9478)

Mask Visibility Percentage

Mask​Visibility​Percentage

FL

1

(0028,9501)

Pixel Shift Sequence

Pixel​Shift​Sequence

SQ

1

(0028,9502)

Region Pixel Shift Sequence

Region​Pixel​Shift​Sequence

SQ

1

(0028,9503)

Vertices of the Region

Vertices​Of​The​Region

SS

2-2n

(0028,9505)

Multi-frame Presentation Sequence

Multi​Frame​Presentation​Sequence

SQ

1

(0028,9506)

Pixel Shift Frame Range

Pixel​Shift​Frame​Range

US

2-2n

(0028,9507)

LUT Frame Range

LUT​Frame​Range

US

2-2n

(0028,9520)

Image to Equipment Mapping Matrix

Image​To​Equipment​Mapping​Matrix

DS

16

(0028,9537)

Equipment Coordinate System Identification

Equipment​Coordinate​System​Identification

CS

1

(0032,000A)

Study Status ID

Study​Status​ID

CS

1

RET (2004)

(0032,000C)

Study Priority ID

Study​Priority​ID

CS

1

RET (2004)

(0032,0012)

Study ID Issuer

Study​ID​Issuer

LO

1

RET (2004)

(0032,0032)

Study Verified Date

Study​Verified​Date

DA

1

RET (2004)

(0032,0033)

Study Verified Time

Study​Verified​Time

TM

1

RET (2004)

(0032,0034)

Study Read Date

Study​Read​Date

DA

1

RET (2004)

(0032,0035)

Study Read Time

Study​Read​Time

TM

1

RET (2004)

(0032,1000)

Scheduled Study Start Date

Scheduled​Study​Start​Date

DA

1

RET (2004)

(0032,1001)

Scheduled Study Start Time

Scheduled​Study​Start​Time

TM

1

RET (2004)

(0032,1010)

Scheduled Study Stop Date

Scheduled​Study​Stop​Date

DA

1

RET (2004)

(0032,1011)

Scheduled Study Stop Time

Scheduled​Study​Stop​Time

TM

1

RET (2004)

(0032,1020)

Scheduled Study Location

Scheduled​Study​Location

LO

1

RET (2004)

(0032,1021)

Scheduled Study Location AE Title

Scheduled​Study​Location​AE​Title

AE

1-n

RET (2004)

(0032,1030)

Reason for Study

Reason​For​Study

LO

1

RET (2004)

(0032,1031)

Requesting Physician Identification Sequence

Requesting​Physician​Identification​Sequence

SQ

1

(0032,1032)

Requesting Physician

Requesting​Physician

PN

1

(0032,1033)

Requesting Service

Requesting​Service

LO

1

(0032,1034)

Requesting Service Code Sequence

Requesting​Service​Code​Sequence

SQ

1

(0032,1040)

Study Arrival Date

Study​Arrival​Date

DA

1

RET (2004)

(0032,1041)

Study Arrival Time

Study​Arrival​Time

TM

1

RET (2004)

(0032,1050)

Study Completion Date

Study​Completion​Date

DA

1

RET (2004)

(0032,1051)

Study Completion Time

Study​Completion​Time

TM

1

RET (2004)

(0032,1055)

Study Component Status ID

Study​Component​Status​ID

CS

1

RET (2004)

(0032,1060)

Requested Procedure Description

Requested​Procedure​Description

LO

1

(0032,1064)

Requested Procedure Code Sequence

Requested​Procedure​Code​Sequence

SQ

1

(0032,1065)

Requested Laterality Code Sequence

Requested​Laterality​Code​Sequence

SQ

1

(0032,1066)

Reason for Visit

Reason​For​Visit

UT

1

(0032,1067)

Reason for Visit Code Sequence

Reason​For​Visit​Code​Sequence

SQ

1

(0032,1070)

Requested Contrast Agent

Requested​Contrast​Agent

LO

1

(0032,4000)

Study Comments

Study​Comments

LT

1

RET (2007)

(0034,0001)

Flow Identifier Sequence

Flow​Identifier​Sequence

SQ

1

(0034,0002)

Flow Identifier

Flow​Identifier

OB

1

(0034,0003)

Flow Transfer Syntax UID

Flow​Transfer​Syntax​UID

UI

1

(0034,0004)

Flow RTP Sampling Rate

Flow​RTP​Sampling​Rate

UL

1

(0034,0005)

Source Identifier

Source​Identifier

OB

1

(0034,0007)

Frame Origin Timestamp

Frame​Origin​Timestamp

OB

1

(0034,0008)

Includes Imaging Subject

Includes​Imaging​Subject

CS

1

(0034,0009)

Frame Usefulness Group Sequence

Frame​Usefulness​Group​Sequence

SQ

1

(0034,000A)

Real-Time Bulk Data Flow Sequence

Real​Time​Bulk​Data​Flow​Sequence

SQ

1

(0034,000B)

Camera Position Group Sequence

Camera​Position​Group​Sequence

SQ

1

(0034,000C)

Includes Information

Includes​Information

CS

1

(0034,000D)

Time of Frame Group Sequence

Time​Of​Frame​Group​Sequence

SQ

1

(0038,0004)

Referenced Patient Alias Sequence

Referenced​Patient​Alias​Sequence

SQ

1

RET (2023d)

(0038,0008)

Visit Status ID

Visit​Status​ID

CS

1

(0038,0010)

Admission ID

Admission​ID

LO

1

(0038,0011)

Issuer of Admission ID

Issuer​Of​Admission​ID

LO

1

RET (2008)

(0038,0014)

Issuer of Admission ID Sequence

Issuer​Of​Admission​ID​Sequence

SQ

1

(0038,0016)

Route of Admissions

Route​Of​Admissions

LO

1

(0038,001A)

Scheduled Admission Date

Scheduled​Admission​Date

DA

1

RET (2004)

(0038,001B)

Scheduled Admission Time

Scheduled​Admission​Time

TM

1

RET (2004)

(0038,001C)

Scheduled Discharge Date

Scheduled​Discharge​Date

DA

1

RET (2004)

(0038,001D)

Scheduled Discharge Time

Scheduled​Discharge​Time

TM

1

RET (2004)

(0038,001E)

Scheduled Patient Institution Residence

Scheduled​Patient​Institution​Residence

LO

1

RET (2004)

(0038,0020)

Admitting Date

Admitting​Date

DA

1

(0038,0021)

Admitting Time

Admitting​Time

TM

1

(0038,0030)

Discharge Date

Discharge​Date

DA

1

RET (2004)

(0038,0032)

Discharge Time

Discharge​Time

TM

1

RET (2004)

(0038,0040)

Discharge Diagnosis Description

Discharge​Diagnosis​Description

LO

1

RET (2004)

(0038,0044)

Discharge Diagnosis Code Sequence

Discharge​Diagnosis​Code​Sequence

SQ

1

RET (2004)

(0038,0050)

Special Needs

Special​Needs

LO

1

(0038,0060)

Service Episode ID

Service​Episode​ID

LO

1

(0038,0061)

Issuer of Service Episode ID

Issuer​Of​Service​Episode​ID

LO

1

RET (2008)

(0038,0062)

Service Episode Description

Service​Episode​Description

LO

1

(0038,0064)

Issuer of Service Episode ID Sequence

Issuer​Of​Service​Episode​ID​Sequence

SQ

1

(0038,0100)

Pertinent Documents Sequence

Pertinent​Documents​Sequence

SQ

1

(0038,0101)

Pertinent Resources Sequence

Pertinent​Resources​Sequence

SQ

1

(0038,0102)

Resource Description

Resource​Description

LO

1

(0038,0300)

Current Patient Location

Current​Patient​Location

LO

1

(0038,0400)

Patient's Institution Residence

Patient​Institution​Residence

LO

1

(0038,0500)

Patient State

Patient​State

LO

1

(0038,0502)

Patient Clinical Trial Participation Sequence

Patient​Clinical​Trial​Participation​Sequence

SQ

1

(0038,4000)

Visit Comments

Visit​Comments

LT

1

(003A,0004)

Waveform Originality

Waveform​Originality

CS

1

(003A,0005)

Number of Waveform Channels

Number​Of​Waveform​Channels

US

1

(003A,0010)

Number of Waveform Samples

Number​Of​Waveform​Samples

UL

1

(003A,001A)

Sampling Frequency

Sampling​Frequency

DS

1

(003A,0020)

Multiplex Group Label

Multiplex​Group​Label

SH

1

(003A,0200)

Channel Definition Sequence

Channel​Definition​Sequence

SQ

1

(003A,0202)

Waveform Channel Number

Waveform​Channel​Number

IS

1

(003A,0203)

Channel Label

Channel​Label

SH

1

(003A,0205)

Channel Status

Channel​Status

CS

1-n

(003A,0208)

Channel Source Sequence

Channel​Source​Sequence

SQ

1

(003A,0209)

Channel Source Modifiers Sequence

Channel​Source​Modifiers​Sequence

SQ

1

(003A,020A)

Source Waveform Sequence

Source​Waveform​Sequence

SQ

1

(003A,020C)

Channel Derivation Description

Channel​Derivation​Description

LO

1

(003A,0210)

Channel Sensitivity

Channel​Sensitivity

DS

1

(003A,0211)

Channel Sensitivity Units Sequence

Channel​Sensitivity​Units​Sequence

SQ

1

(003A,0212)

Channel Sensitivity Correction Factor

Channel​Sensitivity​Correction​Factor

DS

1

(003A,0213)

Channel Baseline

Channel​Baseline

DS

1

(003A,0214)

Channel Time Skew

Channel​Time​Skew

DS

1

(003A,0215)

Channel Sample Skew

Channel​Sample​Skew

DS

1

(003A,0218)

Channel Offset

Channel​Offset

DS

1

(003A,021A)

Waveform Bits Stored

Waveform​Bits​Stored

US

1

(003A,0220)

Filter Low Frequency

Filter​Low​Frequency

DS

1

(003A,0221)

Filter High Frequency

Filter​High​Frequency

DS

1

(003A,0222)

Notch Filter Frequency

Notch​Filter​Frequency

DS

1

(003A,0223)

Notch Filter Bandwidth

Notch​Filter​Bandwidth

DS

1

(003A,0230)

Waveform Data Display Scale

Waveform​Data​Display​Scale

FL

1

(003A,0231)

Waveform Display Background CIELab Value

Waveform​Display​Background​CIE​Lab​Value

US

3

(003A,0240)

Waveform Presentation Group Sequence

Waveform​Presentation​Group​Sequence

SQ

1

(003A,0241)

Presentation Group Number

Presentation​Group​Number

US

1

(003A,0242)

Channel Display Sequence

Channel​Display​Sequence

SQ

1

(003A,0244)

Channel Recommended Display CIELab Value

Channel​Recommended​Display​CIE​Lab​Value

US

3

(003A,0245)

Channel Position

Channel​Position

FL

1

(003A,0246)

Display Shading Flag

Display​Shading​Flag

CS

1

(003A,0247)

Fractional Channel Display Scale

Fractional​Channel​Display​Scale

FL

1

(003A,0248)

Absolute Channel Display Scale

Absolute​Channel​Display​Scale

FL

1

(003A,0300)

Multiplexed Audio Channels Description Code Sequence

Multiplexed​Audio​Channels​Description​Code​Sequence

SQ

1

(003A,0301)

Channel Identification Code

Channel​Identification​Code

IS

1

(003A,0302)

Channel Mode

Channel​Mode

CS

1

(003A,0310)

Multiplex Group UID

Multiplex​Group​UID

UI

1

(003A,0311)

Powerline Frequency

Powerline​Frequency

DS

1

(003A,0312)

Channel Impedance Sequence

Channel​Impedance​Sequence

SQ

1

(003A,0313)

Impedance Value

Impedance​Value

DS

1

(003A,0314)

Impedance Measurement DateTime

Impedance​Measurement​DateTime

DT

1

(003A,0315)

Impedance Measurement Frequency

Impedance​Measurement​Frequency

DS

1

(003A,0316)

Impedance Measurement Current Type

Impedance​Measurement​Current​Type

CS

1

(003A,0317)

Waveform Amplifier Type

Waveform​Amplifier​Type

CS

1

(003A,0318)

Filter Low Frequency Characteristics Sequence

Filter​Low​Frequency​Characteristics​Sequence

SQ

1

(003A,0319)

Filter High Frequency Characteristics Sequence

Filter​High​Frequency​Characteristics​Sequence

SQ

1

(003A,0320)

Summarized Filter Lookup Table Sequence

Summarized​Filter​Lookup​Table​Sequence

SQ

1

(003A,0321)

Notch Filter Characteristics Sequence

Notch​Filter​Characteristics​Sequence

SQ

1

(003A,0322)

Waveform Filter Type

Waveform​Filter​Type

CS

1

(003A,0323)

Analog Filter Characteristics Sequence

Analog​Filter​Characteristics​Sequence

SQ

1

(003A,0324)

Analog Filter Roll Off

Analog​Filter​Roll​Off

DS

1

(003A,0325)

Analog Filter Type Code Sequence

Analog​Filter​Type​Code​Sequence

SQ

1

(003A,0326)

Digital Filter Characteristics Sequence

Digital​Filter​Characteristics​Sequence

SQ

1

(003A,0327)

Digital Filter Order

Digital​Filter​Order

IS

1

(003A,0328)

Digital Filter Type Code Sequence

Digital​Filter​Type​Code​Sequence

SQ

1

(003A,0329)

Waveform Filter Description

Waveform​Filter​Description

ST

1

(003A,032A)

Filter Lookup Table Sequence

Filter​Lookup​Table​Sequence

SQ

1

(003A,032B)

Filter Lookup Table Description

Filter​Lookup​Table​Description

ST

1

(003A,032C)

Frequency Encoding Code Sequence

Frequency​Encoding​Code​Sequence

SQ

1

(003A,032D)

Magnitude Encoding Code Sequence

Magnitude​Encoding​Code​Sequence

SQ

1

(003A,032E)

Filter Lookup Table Data

Filter​Lookup​Table​Data

OD

1

(0040,0001)

Scheduled Station AE Title

Scheduled​Station​AE​Title

AE

1-n

(0040,0002)

Scheduled Procedure Step Start Date

Scheduled​Procedure​Step​Start​Date

DA

1

(0040,0003)

Scheduled Procedure Step Start Time

Scheduled​Procedure​Step​Start​Time

TM

1

(0040,0004)

Scheduled Procedure Step End Date

Scheduled​Procedure​Step​End​Date

DA

1

(0040,0005)

Scheduled Procedure Step End Time

Scheduled​Procedure​Step​End​Time

TM

1

(0040,0006)

Scheduled Performing Physician's Name

Scheduled​Performing​Physician​Name

PN

1

(0040,0007)

Scheduled Procedure Step Description

Scheduled​Procedure​Step​Description

LO

1

(0040,0008)

Scheduled Protocol Code Sequence

Scheduled​Protocol​Code​Sequence

SQ

1

(0040,0009)

Scheduled Procedure Step ID

Scheduled​Procedure​Step​ID

SH

1

(0040,000A)

Stage Code Sequence

Stage​Code​Sequence

SQ

1

(0040,000B)

Scheduled Performing Physician Identification Sequence

Scheduled​Performing​Physician​Identification​Sequence

SQ

1

(0040,0010)

Scheduled Station Name

Scheduled​Station​Name

SH

1-n

(0040,0011)

Scheduled Procedure Step Location

Scheduled​Procedure​Step​Location

SH

1

(0040,0012)

Pre-Medication

Pre​Medication

LO

1

(0040,0020)

Scheduled Procedure Step Status

Scheduled​Procedure​Step​Status

CS

1

(0040,0026)

Order Placer Identifier Sequence

Order​Placer​Identifier​Sequence

SQ

1

(0040,0027)

Order Filler Identifier Sequence

Order​Filler​Identifier​Sequence

SQ

1

(0040,0031)

Local Namespace Entity ID

Local​Namespace​Entity​ID

UT

1

(0040,0032)

Universal Entity ID

Universal​Entity​ID

UT

1

(0040,0033)

Universal Entity ID Type

Universal​Entity​ID​Type

CS

1

(0040,0035)

Identifier Type Code

Identifier​Type​Code

CS

1

(0040,0036)

Assigning Facility Sequence

Assigning​Facility​Sequence

SQ

1

(0040,0039)

Assigning Jurisdiction Code Sequence

Assigning​Jurisdiction​Code​Sequence

SQ

1

(0040,003A)

Assigning Agency or Department Code Sequence

Assigning​Agency​Or​Department​Code​Sequence

SQ

1

(0040,0100)

Scheduled Procedure Step Sequence

Scheduled​Procedure​Step​Sequence

SQ

1

(0040,0220)

Referenced Non-Image Composite SOP Instance Sequence

Referenced​Non​Image​Composite​SOP​Instance​Sequence

SQ

1

(0040,0241)

Performed Station AE Title

Performed​Station​AE​Title

AE

1

(0040,0242)

Performed Station Name

Performed​Station​Name

SH

1

(0040,0243)

Performed Location

Performed​Location

SH

1

(0040,0244)

Performed Procedure Step Start Date

Performed​Procedure​Step​Start​Date

DA

1

(0040,0245)

Performed Procedure Step Start Time

Performed​Procedure​Step​Start​Time

TM

1

(0040,0250)

Performed Procedure Step End Date

Performed​Procedure​Step​End​Date

DA

1

(0040,0251)

Performed Procedure Step End Time

Performed​Procedure​Step​End​Time

TM

1

(0040,0252)

Performed Procedure Step Status

Performed​Procedure​Step​Status

CS

1

(0040,0253)

Performed Procedure Step ID

Performed​Procedure​Step​ID

SH

1

(0040,0254)

Performed Procedure Step Description

Performed​Procedure​Step​Description

LO

1

(0040,0255)

Performed Procedure Type Description

Performed​Procedure​Type​Description

LO

1

(0040,0260)

Performed Protocol Code Sequence

Performed​Protocol​Code​Sequence

SQ

1

(0040,0261)

Performed Protocol Type

Performed​Protocol​Type

CS

1

(0040,0270)

Scheduled Step Attributes Sequence

Scheduled​Step​Attributes​Sequence

SQ

1

(0040,0275)

Request Attributes Sequence

Request​Attributes​Sequence

SQ

1

(0040,0280)

Comments on the Performed Procedure Step

Comments​On​The​Performed​Procedure​Step

ST

1

(0040,0281)

Performed Procedure Step Discontinuation Reason Code Sequence

Performed​Procedure​Step​Discontinuation​Reason​Code​Sequence

SQ

1

(0040,0293)

Quantity Sequence

Quantity​Sequence

SQ

1

(0040,0294)

Quantity

Quantity

DS

1

(0040,0295)

Measuring Units Sequence

Measuring​Units​Sequence

SQ

1

(0040,0296)

Billing Item Sequence

Billing​Item​Sequence

SQ

1

(0040,0300)

Total Time of Fluoroscopy

Total​Time​Of​Fluoroscopy

US

1

RET (2017c)

(0040,0301)

Total Number of Exposures

Total​Number​Of​Exposures

US

1

RET (2017c)

(0040,0302)

Entrance Dose

Entrance​Dose

US

1

(0040,0303)

Exposed Area

Exposed​Area

US

1-2

(0040,0306)

Distance Source to Entrance

Distance​Source​To​Entrance

DS

1

(0040,0307)

Distance Source to Support

Distance​Source​To​Support

DS

1

RET (2001)

(0040,030E)

Exposure Dose Sequence

Exposure​Dose​Sequence

SQ

1

RET (2017c)

(0040,0310)

Comments on Radiation Dose

Comments​On​Radiation​Dose

ST

1

(0040,0312)

X-Ray Output

X​Ray​Output

DS

1

(0040,0314)

Half Value Layer

Half​Value​Layer

DS

1

(0040,0316)

Organ Dose

Organ​Dose

DS

1

(0040,0318)

Organ Exposed

Organ​Exposed

CS

1

(0040,0320)

Billing Procedure Step Sequence

Billing​Procedure​Step​Sequence

SQ

1

(0040,0321)

Film Consumption Sequence

Film​Consumption​Sequence

SQ

1

(0040,0324)

Billing Supplies and Devices Sequence

Billing​Supplies​And​Devices​Sequence

SQ

1

(0040,0330)

Referenced Procedure Step Sequence

Referenced​Procedure​Step​Sequence

SQ

1

RET (2001)

(0040,0340)

Performed Series Sequence

Performed​Series​Sequence

SQ

1

(0040,0400)

Comments on the Scheduled Procedure Step

Comments​On​The​Scheduled​Procedure​Step

LT

1

(0040,0440)

Protocol Context Sequence

Protocol​Context​Sequence

SQ

1

(0040,0441)

Content Item Modifier Sequence

Content​Item​Modifier​Sequence

SQ

1

(0040,0500)

Scheduled Specimen Sequence

Scheduled​Specimen​Sequence

SQ

1

(0040,050A)

Specimen Accession Number

Specimen​Accession​Number

LO

1

RET (2008)

(0040,0512)

Container Identifier

Container​Identifier

LO

1

(0040,0513)

Issuer of the Container Identifier Sequence

Issuer​Of​The​Container​Identifier​Sequence

SQ

1

(0040,0515)

Alternate Container Identifier Sequence

Alternate​Container​Identifier​Sequence

SQ

1

(0040,0518)

Container Type Code Sequence

Container​Type​Code​Sequence

SQ

1

(0040,051A)

Container Description

Container​Description

LO

1

(0040,0520)

Container Component Sequence

Container​Component​Sequence

SQ

1

(0040,0550)

Specimen Sequence

Specimen​Sequence

SQ

1

RET (2008)

(0040,0551)

Specimen Identifier

Specimen​Identifier

LO

1

(0040,0552)

Specimen Description Sequence (Trial)

Specimen​Description​Sequence​Trial

SQ

1

RET (2007)

(0040,0553)

Specimen Description (Trial)

Specimen​Description​Trial

ST

1

RET (2007)

(0040,0554)

Specimen UID

Specimen​UID

UI

1

(0040,0555)

Acquisition Context Sequence

Acquisition​Context​Sequence

SQ

1

(0040,0556)

Acquisition Context Description

Acquisition​Context​Description

ST

1

(0040,059A)

Specimen Type Code Sequence

Specimen​Type​Code​Sequence

SQ

1

(0040,0560)

Specimen Description Sequence

Specimen​Description​Sequence

SQ

1

(0040,0562)

Issuer of the Specimen Identifier Sequence

Issuer​Of​The​Specimen​Identifier​Sequence

SQ

1

(0040,0600)

Specimen Short Description

Specimen​Short​Description

LO

1

(0040,0602)

Specimen Detailed Description

Specimen​Detailed​Description

UT

1

(0040,0610)

Specimen Preparation Sequence

Specimen​Preparation​Sequence

SQ

1

(0040,0612)

Specimen Preparation Step Content Item Sequence

Specimen​Preparation​Step​Content​Item​Sequence

SQ

1

(0040,0620)

Specimen Localization Content Item Sequence

Specimen​Localization​Content​Item​Sequence

SQ

1

(0040,06FA)

Slide Identifier

Slide​Identifier

LO

1

RET (2008)

(0040,0710)

Whole Slide Microscopy Image Frame Type Sequence

Whole​Slide​Microscopy​Image​Frame​Type​Sequence

SQ

1

(0040,071A)

Image Center Point Coordinates Sequence

Image​Center​Point​Coordinates​Sequence

SQ

1

(0040,072A)

X Offset in Slide Coordinate System

X​Offset​In​Slide​Coordinate​System

DS

1

(0040,073A)

Y Offset in Slide Coordinate System

Y​Offset​In​Slide​Coordinate​System

DS

1

(0040,074A)

Z Offset in Slide Coordinate System

Z​Offset​In​Slide​Coordinate​System

DS

1

(0040,08D8)

Pixel Spacing Sequence

Pixel​Spacing​Sequence

SQ

1

RET (2009)

(0040,08DA)

Coordinate System Axis Code Sequence

Coordinate​System​Axis​Code​Sequence

SQ

1

RET (2009)

(0040,08EA)

Measurement Units Code Sequence

Measurement​Units​Code​Sequence

SQ

1

(0040,09F8)

Vital Stain Code Sequence (Trial)

Vital​Stain​Code​Sequence​Trial

SQ

1

RET (2007)

(0040,1001)

Requested Procedure ID

Requested​Procedure​ID

SH

1

(0040,1002)

Reason for the Requested Procedure

Reason​For​The​Requested​Procedure

LO

1

(0040,1003)

Requested Procedure Priority

Requested​Procedure​Priority

SH

1

(0040,1004)

Patient Transport Arrangements

Patient​Transport​Arrangements

LO

1

(0040,1005)

Requested Procedure Location

Requested​Procedure​Location

LO

1

(0040,1006)

Placer Order Number / Procedure

Placer​Order​Number​Procedure

SH

1

RET (1998)

(0040,1007)

Filler Order Number / Procedure

Filler​Order​Number​Procedure

SH

1

RET (1998)

(0040,1008)

Confidentiality Code

Confidentiality​Code

LO

1

(0040,1009)

Reporting Priority

Reporting​Priority

SH

1

(0040,100A)

Reason for Requested Procedure Code Sequence

Reason​For​Requested​Procedure​Code​Sequence

SQ

1

(0040,1010)

Names of Intended Recipients of Results

Names​Of​Intended​Recipients​Of​Results

PN

1-n

(0040,1011)

Intended Recipients of Results Identification Sequence

Intended​Recipients​Of​Results​Identification​Sequence

SQ

1

(0040,1012)

Reason For Performed Procedure Code Sequence

Reason​For​Performed​Procedure​Code​Sequence

SQ

1

(0040,1060)

Requested Procedure Description (Trial)

Requested​Procedure​Description​Trial

LO

1

RET (2009)

(0040,1101)

Person Identification Code Sequence

Person​Identification​Code​Sequence

SQ

1

(0040,1102)

Person's Address

Person​Address

ST

1

(0040,1103)

Person's Telephone Numbers

Person​Telephone​Numbers

LO

1-n

(0040,1104)

Person's Telecom Information

Person​Telecom​Information

LT

1

(0040,1400)

Requested Procedure Comments

Requested​Procedure​Comments

LT

1

(0040,2001)

Reason for the Imaging Service Request

Reason​For​The​Imaging​Service​Request

LO

1

RET (2003)

(0040,2004)

Issue Date of Imaging Service Request

Issue​Date​Of​Imaging​Service​Request

DA

1

(0040,2005)

Issue Time of Imaging Service Request

Issue​Time​Of​Imaging​Service​Request

TM

1

(0040,2006)

Placer Order Number / Imaging Service Request (Retired)

Placer​Order​Number​Imaging​Service​Request​Retired

SH

1

RET (1998)

(0040,2007)

Filler Order Number / Imaging Service Request (Retired)

Filler​Order​Number​Imaging​Service​Request​Retired

SH

1

RET (1998)

(0040,2008)

Order Entered By

Order​Entered​By

PN

1

(0040,2009)

Order Enterer's Location

Order​Enterer​Location

SH

1

(0040,2010)

Order Callback Phone Number

Order​Callback​Phone​Number

SH

1

(0040,2011)

Order Callback Telecom Information

Order​Callback​Telecom​Information

LT

1

(0040,2016)

Placer Order Number / Imaging Service Request

Placer​Order​Number​Imaging​Service​Request

LO

1

(0040,2017)

Filler Order Number / Imaging Service Request

Filler​Order​Number​Imaging​Service​Request

LO

1

(0040,2400)

Imaging Service Request Comments

Imaging​Service​Request​Comments

LT

1

(0040,3001)

Confidentiality Constraint on Patient Data Description

Confidentiality​Constraint​On​Patient​Data​Description

LO

1

(0040,4001)

General Purpose Scheduled Procedure Step Status

General​Purpose​Scheduled​Procedure​Step​Status

CS

1

RET (2011)

(0040,4002)

General Purpose Performed Procedure Step Status

General​Purpose​Performed​Procedure​Step​Status

CS

1

RET (2011)

(0040,4003)

General Purpose Scheduled Procedure Step Priority

General​Purpose​Scheduled​Procedure​Step​Priority

CS

1

RET (2011)

(0040,4004)

Scheduled Processing Applications Code Sequence

Scheduled​Processing​Applications​Code​Sequence

SQ

1

RET (2011)

(0040,4005)

Scheduled Procedure Step Start DateTime

Scheduled​Procedure​Step​Start​Date​Time

DT

1

(0040,4006)

Multiple Copies Flag

Multiple​Copies​Flag

CS

1

RET (2011)

(0040,4007)

Performed Processing Applications Code Sequence

Performed​Processing​Applications​Code​Sequence

SQ

1

RET (2017e)

(0040,4008)

Scheduled Procedure Step Expiration DateTime

Scheduled​Procedure​Step​Expiration​Date​Time

DT

1

(0040,4009)

Human Performer Code Sequence

Human​Performer​Code​Sequence

SQ

1

(0040,4010)

Scheduled Procedure Step Modification DateTime

Scheduled​Procedure​Step​Modification​Date​Time

DT

1

(0040,4011)

Expected Completion DateTime

Expected​Completion​Date​Time

DT

1

(0040,4015)

Resulting General Purpose Performed Procedure Steps Sequence

Resulting​General​Purpose​Performed​Procedure​Steps​Sequence

SQ

1

RET (2011)

(0040,4016)

Referenced General Purpose Scheduled Procedure Step Sequence

Referenced​General​Purpose​Scheduled​Procedure​Step​Sequence

SQ

1

RET (2011)

(0040,4018)

Scheduled Workitem Code Sequence

Scheduled​Workitem​Code​Sequence

SQ

1

(0040,4019)

Performed Workitem Code Sequence

Performed​Workitem​Code​Sequence

SQ

1

(0040,4020)

Input Availability Flag

Input​Availability​Flag

CS

1

RET (2017b)

(0040,4021)

Input Information Sequence

Input​Information​Sequence

SQ

1

(0040,4022)

Relevant Information Sequence

Relevant​Information​Sequence

SQ

1

RET (2011)

(0040,4023)

Referenced General Purpose Scheduled Procedure Step Transaction UID

Referenced​General​Purpose​Scheduled​Procedure​Step​Transaction​UID

UI

1

RET (2011)

(0040,4025)

Scheduled Station Name Code Sequence

Scheduled​Station​Name​Code​Sequence

SQ

1

(0040,4026)

Scheduled Station Class Code Sequence

Scheduled​Station​Class​Code​Sequence

SQ

1

(0040,4027)

Scheduled Station Geographic Location Code Sequence

Scheduled​Station​Geographic​Location​Code​Sequence

SQ

1

(0040,4028)

Performed Station Name Code Sequence

Performed​Station​Name​Code​Sequence

SQ

1

(0040,4029)

Performed Station Class Code Sequence

Performed​Station​Class​Code​Sequence

SQ

1

(0040,4030)

Performed Station Geographic Location Code Sequence

Performed​Station​Geographic​Location​Code​Sequence

SQ

1

(0040,4031)

Requested Subsequent Workitem Code Sequence

Requested​Subsequent​Workitem​Code​Sequence

SQ

1

RET (2011)

(0040,4032)

Non-DICOM Output Code Sequence

Non​DICOM​Output​Code​Sequence

SQ

1

RET (2011)

(0040,4033)

Output Information Sequence

Output​Information​Sequence

SQ

1

(0040,4034)

Scheduled Human Performers Sequence

Scheduled​Human​Performers​Sequence

SQ

1

(0040,4035)

Actual Human Performers Sequence

Actual​Human​Performers​Sequence

SQ

1

(0040,4036)

Human Performer's Organization

Human​Performer​Organization

LO

1

(0040,4037)

Human Performer's Name

Human​Performer​Name

PN

1

(0040,4040)

Raw Data Handling

Raw​Data​Handling

CS

1

(0040,4041)

Input Readiness State

Input​Readiness​State

CS

1

(0040,4050)

Performed Procedure Step Start DateTime

Performed​Procedure​Step​Start​Date​Time

DT

1

(0040,4051)

Performed Procedure Step End DateTime

Performed​Procedure​Step​End​Date​Time

DT

1

(0040,4052)

Procedure Step Cancellation DateTime

Procedure​Step​Cancellation​Date​Time

DT

1

(0040,4070)

Output Destination Sequence

Output​Destination​Sequence

SQ

1

(0040,4071)

DICOM Storage Sequence

DICOM​Storage​Sequence

SQ

1

(0040,4072)

STOW-RS Storage Sequence

STOWRS​Storage​Sequence

SQ

1

(0040,4073)

Storage URL

Storage​URL

UR

1

(0040,4074)

XDS Storage Sequence

XDS​Storage​Sequence

SQ

1

(0040,8302)

Entrance Dose in mGy

Entrance​Dose​In​mGy

DS

1

(0040,8303)

Entrance Dose Derivation

Entrance​Dose​Derivation

CS

1

(0040,9092)

Parametric Map Frame Type Sequence

Parametric​​Map​​Frame​Type​Sequence

SQ

1

(0040,9094)

Referenced Image Real World Value Mapping Sequence

Referenced​Image​Real​World​Value​Mapping​Sequence

SQ

1

(0040,9096)

Real World Value Mapping Sequence

Real​World​Value​Mapping​Sequence

SQ

1

(0040,9098)

Pixel Value Mapping Code Sequence

Pixel​Value​Mapping​Code​Sequence

SQ

1

(0040,9210)

LUT Label

LUT​Label

SH

1

(0040,9211)

Real World Value Last Value Mapped

Real​World​Value​Last​Value​Mapped

US or SS

1

(0040,9212)

Real World Value LUT Data

Real​World​Value​LUT​Data

FD

1-n

(0040,9213)

Double Float Real World Value Last Value Mapped

Double​Float​Real​World​Value​Last​Value​Mapped

FD

1

(0040,9214)

Double Float Real World Value First Value Mapped

Double​Float​Real​World​Value​First​Value​Mapped

FD

1

(0040,9216)

Real World Value First Value Mapped

Real​World​Value​First​Value​Mapped

US or SS

1

(0040,9220)

Quantity Definition Sequence

Quantity​​Definition​​Sequence

SQ

1

(0040,9224)

Real World Value Intercept

Real​World​Value​Intercept

FD

1

(0040,9225)

Real World Value Slope

Real​World​Value​Slope

FD

1

(0040,A007)

Findings Flag (Trial)

Findings​Flag​Trial

CS

1

RET (2009)

(0040,A010)

Relationship Type

Relationship​Type

CS

1

(0040,A020)

Findings Sequence (Trial)

Findings​Sequence​Trial

SQ

1

RET (2009)

(0040,A021)

Findings Group UID (Trial)

Findings​Group​UID​Trial

UI

1

RET (2009)

(0040,A022)

Referenced Findings Group UID (Trial)

Referenced​Findings​Group​UID​Trial

UI

1

RET (2009)

(0040,A023)

Findings Group Recording Date (Trial)

Findings​Group​Recording​Date​Trial

DA

1

RET (2009)

(0040,A024)

Findings Group Recording Time (Trial)

Findings​Group​Recording​Time​Trial

TM

1

RET (2009)

(0040,A026)

Findings Source Category Code Sequence (Trial)

Findings​Source​Category​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A027)

Verifying Organization

Verifying​Organization

LO

1

(0040,A028)

Documenting Organization Identifier Code Sequence (Trial)

Documenting​Organization​Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A030)

Verification DateTime

Verification​Date​Time

DT

1

(0040,A032)

Observation DateTime

Observation​Date​Time

DT

1

(0040,A033)

Observation Start DateTime

Observation​Start​Date​Time

DT

1

(0040,A034)

Effective Start DateTime

Effective​Start​Date​Time

DT

1

(0040,A035)

Effective Stop DateTime

Effective​Stop​Date​Time

DT

1

(0040,A040)

Value Type

Value​Type

CS

1

(0040,A043)

Concept Name Code Sequence

Concept​Name​Code​Sequence

SQ

1

(0040,A047)

Measurement Precision Description (Trial)

Measurement​Precision​Description​Trial

LO

1

RET (2009)

(0040,A050)

Continuity Of Content

Continuity​Of​Content

CS

1

(0040,A057)

Urgency or Priority Alerts (Trial)

Urgency​Or​Priority​Alerts​Trial

CS

1-n

RET (2009)

(0040,A060)

Sequencing Indicator (Trial)

Sequencing​Indicator​Trial

LO

1

RET (2009)

(0040,A066)

Document Identifier Code Sequence (Trial)

Document​Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A067)

Document Author (Trial)

Document​Author​Trial

PN

1

RET (2009)

(0040,A068)

Document Author Identifier Code Sequence (Trial)

Document​Author​Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A070)

Identifier Code Sequence (Trial)

Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A073)

Verifying Observer Sequence

Verifying​Observer​Sequence

SQ

1

(0040,A074)

Object Binary Identifier (Trial)

Object​Binary​Identifier​Trial

OB

1

RET (2009)

(0040,A075)

Verifying Observer Name

Verifying​Observer​Name

PN

1

(0040,A076)

Documenting Observer Identifier Code Sequence (Trial)

Documenting​Observer​Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A078)

Author Observer Sequence

Author​Observer​Sequence

SQ

1

(0040,A07A)

Participant Sequence

Participant​Sequence

SQ

1

(0040,A07C)

Custodial Organization Sequence

Custodial​Organization​Sequence

SQ

1

(0040,A080)

Participation Type

Participation​Type

CS

1

(0040,A082)

Participation DateTime

Participation​Date​Time

DT

1

(0040,A084)

Observer Type

Observer​Type

CS

1

(0040,A085)

Procedure Identifier Code Sequence (Trial)

Procedure​Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A088)

Verifying Observer Identification Code Sequence

Verifying​Observer​Identification​Code​Sequence

SQ

1

(0040,A089)

Object Directory Binary Identifier (Trial)

Object​Directory​Binary​Identifier​Trial

OB

1

RET (2009)

(0040,A090)

Equivalent CDA Document Sequence

Equivalent​CDA​Document​Sequence

SQ

1

RET (2006)

(0040,A0B0)

Referenced Waveform Channels

Referenced​Waveform​Channels

US

2-2n

(0040,A110)

Date of Document or Verbal Transaction (Trial)

Date​Of​Document​Or​Verbal​Transaction​Trial

DA

1

RET (2009)

(0040,A112)

Time of Document Creation or Verbal Transaction (Trial)

Time​Of​Document​Creation​Or​Verbal​Transaction​Trial

TM

1

RET (2009)

(0040,A120)

DateTime

Date​Time

DT

1

(0040,A121)

Date

Date

DA

1

(0040,A122)

Time

Time

TM

1

(0040,A123)

Person Name

Person​Name

PN

1

(0040,A124)

UID

UID

UI

1

(0040,A125)

Report Status ID (Trial)

Report​Status​ID​Trial

CS

2

RET (2009)

(0040,A130)

Temporal Range Type

Temporal​Range​Type

CS

1

(0040,A132)

Referenced Sample Positions

Referenced​Sample​Positions

UL

1-n

(0040,A136)

Referenced Frame Numbers

Referenced​Frame​Numbers

US

1-n

RET (2019)

(0040,A138)

Referenced Time Offsets

Referenced​Time​Offsets

DS

1-n

(0040,A13A)

Referenced DateTime

Referenced​Date​Time

DT

1-n

(0040,A160)

Text Value

Text​Value

UT

1

(0040,A161)

Floating Point Value

Floating​Point​Value

FD

1-n

(0040,A162)

Rational Numerator Value

Rational​Numerator​Value

SL

1-n

(0040,A163)

Rational Denominator Value

Rational​Denominator​Value

UL

1-n

(0040,A167)

Observation Category Code Sequence (Trial)

Observation​Category​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A168)

Concept Code Sequence

Concept​Code​Sequence

SQ

1

(0040,A16A)

Bibliographic Citation (Trial)

Bibliographic​Citation​Trial

ST

1

RET (2009)

(0040,A170)

Purpose of Reference Code Sequence

Purpose​Of​Reference​Code​Sequence

SQ

1

See Note 1

(0040,A171)

Observation UID

Observation​UID

UI

1

(0040,A172)

Referenced Observation UID (Trial)

Referenced​Observation​UID​Trial

UI

1

RET (2009)

(0040,A173)

Referenced Observation Class (Trial)

Referenced​Observation​Class​Trial

CS

1

RET (2009)

(0040,A174)

Referenced Object Observation Class (Trial)

Referenced​Object​Observation​Class​Trial

CS

1

RET (2009)

(0040,A180)

Annotation Group Number

Annotation​Group​Number

US

1

(0040,A192)

Observation Date (Trial)

Observation​Date​Trial

DA

1

RET (2009)

(0040,A193)

Observation Time (Trial)

Observation​Time​Trial

TM

1

RET (2009)

(0040,A194)

Measurement Automation (Trial)

Measurement​Automation​Trial

CS

1

RET (2009)

(0040,A195)

Modifier Code Sequence

Modifier​Code​Sequence

SQ

1

(0040,A224)

Identification Description (Trial)

Identification​Description​Trial

ST

1

RET (2009)

(0040,A290)

Coordinates Set Geometric Type (Trial)

Coordinates​Set​Geometric​Type​Trial

CS

1

RET (2009)

(0040,A296)

Algorithm Code Sequence (Trial)

Algorithm​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A297)

Algorithm Description (Trial)

Algorithm​Description​Trial

ST

1

RET (2009)

(0040,A29A)

Pixel Coordinates Set (Trial)

Pixel​Coordinates​Set​Trial

SL

2-2n

RET (2009)

(0040,A300)

Measured Value Sequence

Measured​Value​Sequence

SQ

1

(0040,A301)

Numeric Value Qualifier Code Sequence

Numeric​Value​Qualifier​Code​Sequence

SQ

1

(0040,A307)

Current Observer (Trial)

Current​Observer​Trial

PN

1

RET (2009)

(0040,A30A)

Numeric Value

Numeric​Value

DS

1-n

(0040,A313)

Referenced Accession Sequence (Trial)

Referenced​Accession​Sequence​Trial

SQ

1

RET (2009)

(0040,A33A)

Report Status Comment (Trial)

Report​Status​Comment​Trial

ST

1

RET (2009)

(0040,A340)

Procedure Context Sequence (Trial)

Procedure​Context​Sequence​Trial

SQ

1

RET (2009)

(0040,A352)

Verbal Source (Trial)

Verbal​Source​Trial

PN

1

RET (2009)

(0040,A353)

Address (Trial)

Address​Trial

ST

1

RET (2009)

(0040,A354)

Telephone Number (Trial)

Telephone​Number​Trial

LO

1

RET (2009)

(0040,A358)

Verbal Source Identifier Code Sequence (Trial)

Verbal​Source​Identifier​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A360)

Predecessor Documents Sequence

Predecessor​Documents​Sequence

SQ

1

(0040,A370)

Referenced Request Sequence

Referenced​Request​Sequence

SQ

1

(0040,A372)

Performed Procedure Code Sequence

Performed​Procedure​Code​Sequence

SQ

1

(0040,A375)

Current Requested Procedure Evidence Sequence

Current​Requested​Procedure​Evidence​Sequence

SQ

1

(0040,A380)

Report Detail Sequence (Trial)

Report​Detail​Sequence​Trial

SQ

1

RET (2009)

(0040,A385)

Pertinent Other Evidence Sequence

Pertinent​Other​Evidence​Sequence

SQ

1

(0040,A390)

HL7 Structured Document Reference Sequence

HL7Structured​Document​Reference​Sequence

SQ

1

(0040,A402)

Observation Subject UID (Trial)

Observation​Subject​UID​Trial

UI

1

RET (2009)

(0040,A403)

Observation Subject Class (Trial)

Observation​Subject​Class​Trial

CS

1

RET (2009)

(0040,A404)

Observation Subject Type Code Sequence (Trial)

Observation​Subject​Type​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A491)

Completion Flag

Completion​Flag

CS

1

(0040,A492)

Completion Flag Description

Completion​Flag​Description

LO

1

(0040,A493)

Verification Flag

Verification​Flag

CS

1

(0040,A494)

Archive Requested

Archive​Requested

CS

1

(0040,A496)

Preliminary Flag

Preliminary​Flag

CS

1

(0040,A504)

Content Template Sequence

Content​Template​Sequence

SQ

1

(0040,A525)

Identical Documents Sequence

Identical​Documents​Sequence

SQ

1

(0040,A600)

Observation Subject Context Flag (Trial)

Observation​Subject​Context​Flag​Trial

CS

1

RET (2009)

(0040,A601)

Observer Context Flag (Trial)

Observer​Context​Flag​Trial

CS

1

RET (2009)

(0040,A603)

Procedure Context Flag (Trial)

Procedure​Context​Flag​Trial

CS

1

RET (2009)

(0040,A730)

Content Sequence

Content​Sequence

SQ

1

(0040,A731)

Relationship Sequence (Trial)

Relationship​Sequence​Trial

SQ

1

RET (2009)

(0040,A732)

Relationship Type Code Sequence (Trial)

Relationship​Type​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A744)

Language Code Sequence (Trial)

Language​Code​Sequence​Trial

SQ

1

RET (2009)

(0040,A801)

Tabulated Values Sequence

Tabulated​Values​Sequence

SQ

1

(0040,A802)

Number of Table Rows

Number​Of​Table​Rows

UL

1

(0040,A803)

Number of Table Columns

Number​Of​Table​Columns

UL

1

(0040,A804)

Table Row Number

Table​Row​Number

UL

1

(0040,A805)

Table Column Number

Table​Column​Number

UL

1

(0040,A806)

Table Row Definition Sequence

Table​Row​Definition​Sequence

SQ

1

(0040,A807)

Table Column Definition Sequence

Table​Column​Definition​Sequence

SQ

1

(0040,A808)

Cell Values Sequence

Cell​Values​Sequence

SQ

1

(0040,A992)

Uniform Resource Locator (Trial)

Uniform​Resource​Locator​Trial

ST

1

RET (2009)

(0040,B020)

Waveform Annotation Sequence

Waveform​Annotation​Sequence

SQ

1

(0040,B030)

Structured Waveform Annotation Sequence

Structured​Waveform​Annotation​Sequence

SQ

1

(0040,B031)

Waveform Annotation Display Selection Sequence

Waveform​Annotation​Display​Selection​Sequence

SQ

1

(0040,B032)

Referenced Montage Index

Referenced​Montage​Index

US

1

(0040,B033)

Waveform Textual Annotation Sequence

Waveform​Textual​Annotation​Sequence

SQ

1

(0040,B034)

Annotation DateTime

Annotation​Date​Time

DT

1

(0040,B035)

Displayed Waveform Segment Sequence

Displayed​Waveform​Segment​Sequence

SQ

1

(0040,B036)

Segment Definition DateTime

Segment​Definition​Date​Time

DT

1

(0040,B037)

Montage Activation Sequence

Montage​Activation​Sequence

SQ

1

(0040,B038)

Montage Activation Time Offset

Montage​Activation​Time​Offset

DS

1

(0040,B039)

Waveform Montage Sequence

Waveform​Montage​Sequence

SQ

1

(0040,B03A)

Referenced Montage Channel Number

Referenced​Montage​Channel​Number

IS

1

(0040,B03B)

Montage Name

Montage​Name

LT

1

(0040,B03C)

Montage Channel Sequence

Montage​Channel​Sequence

SQ

1

(0040,B03D)

Montage Index

Montage​Index

US

1

(0040,B03E)

Montage Channel Number

Montage​Channel​Number

IS

1

(0040,B03F)

Montage Channel Label

Montage​Channel​Label

LO

1

(0040,B040)

Montage Channel Source Code Sequence

Montage​Channel​Source​Code​Sequence

SQ

1

(0040,B041)

Contributing Channel Sources Sequence

Contributing​Channel​Sources​Sequence

SQ

1

(0040,B042)

Channel Weight

Channel​Weight

FL

1

(0040,DB00)

Template Identifier

Template​Identifier

CS

1

(0040,DB06)

Template Version

Template​Version

DT

1

RET (2001)

(0040,DB07)

Template Local Version

Template​Local​Version

DT

1

RET (2001)

(0040,DB0B)

Template Extension Flag

Template​Extension​Flag

CS

1

RET (2001)

(0040,DB0C)

Template Extension Organization UID

Template​Extension​Organization​UID

UI

1

RET (2001)

(0040,DB0D)

Template Extension Creator UID

Template​Extension​Creator​UID

UI

1

RET (2001)

(0040,DB73)

Referenced Content Item Identifier

Referenced​Content​Item​Identifier

UL

1-n

(0040,E001)

HL7 Instance Identifier

HL7Instance​Identifier

ST

1

(0040,E004)

HL7 Document Effective Time

HL7Document​Effective​Time

DT

1

(0040,E006)

HL7 Document Type Code Sequence

HL7Document​Type​Code​Sequence

SQ

1

(0040,E008)

Document Class Code Sequence

Document​Class​Code​Sequence

SQ

1

(0040,E010)

Retrieve URI

Retrieve​URI

UR

1

(0040,E011)

Retrieve Location UID

Retrieve​Location​UID

UI

1

(0040,E012)

Display URI

Display​URI

UR

1

(0040,E020)

Type of Instances

Type​Of​Instances

CS

1

(0040,E021)

DICOM Retrieval Sequence

DICOM​Retrieval​Sequence

SQ

1

(0040,E022)

DICOM Media Retrieval Sequence

DICOM​Media​Retrieval​Sequence

SQ

1

(0040,E023)

WADO Retrieval Sequence

WADO​Retrieval​Sequence

SQ

1

(0040,E024)

XDS Retrieval Sequence

XDS​Retrieval​Sequence

SQ

1

(0040,E025)

WADO-RS Retrieval Sequence

WADORS​Retrieval​Sequence

SQ

1

(0040,E030)

Repository Unique ID

Repository​Unique​ID

UI

1

(0040,E031)

Home Community ID

Home​Community​ID

UI

1

(0042,0010)

Document Title

Document​Title

ST

1

(0042,0011)

Encapsulated Document

Encapsulated​Document

OB

1

(0042,0012)

MIME Type of Encapsulated Document

MIME​Type​Of​Encapsulated​Document

LO

1

(0042,0013)

Source Instance Sequence

Source​Instance​Sequence

SQ

1

(0042,0014)

List of MIME Types

List​Of​MIME​Types

LO

1-n

(0042,0015)

Encapsulated Document Length

Encapsulated​Document​Length

UL

1

(0044,0001)

Product Package Identifier

Product​Package​Identifier

ST

1

(0044,0002)

Substance Administration Approval

Substance​Administration​Approval

CS

1

(0044,0003)

Approval Status Further Description

Approval​Status​Further​Description

LT

1

(0044,0004)

Approval Status DateTime

Approval​Status​Date​Time

DT

1

(0044,0007)

Product Type Code Sequence

Product​Type​Code​Sequence

SQ

1

(0044,0008)

Product Name

Product​Name

LO

1-n

(0044,0009)

Product Description

Product​Description

LT

1

(0044,000A)

Product Lot Identifier

Product​Lot​Identifier

LO

1

(0044,000B)

Product Expiration DateTime

Product​Expiration​Date​Time

DT

1

(0044,0010)

Substance Administration DateTime

Substance​Administration​Date​Time

DT

1

(0044,0011)

Substance Administration Notes

Substance​Administration​Notes

LO

1

(0044,0012)

Substance Administration Device ID

Substance​Administration​Device​ID

LO

1

(0044,0013)

Product Parameter Sequence

Product​Parameter​Sequence

SQ

1

(0044,0019)

Substance Administration Parameter Sequence

Substance​Administration​Parameter​Sequence

SQ

1

(0044,0100)

Approval Sequence

Approval​Sequence

SQ

1

(0044,0101)

Assertion Code Sequence

Assertion​Code​Sequence

SQ

1

(0044,0102)

Assertion UID

Assertion​UID

UI

1

(0044,0103)

Asserter Identification Sequence

Asserter​Identification​Sequence

SQ

1

(0044,0104)

Assertion DateTime

Assertion​DateTime

DT

1

(0044,0105)

Assertion Expiration DateTime

Assertion​Expiration​DateTime

DT

1

(0044,0106)

Assertion Comments

Assertion​Comments

UT

1

(0044,0107)

Related Assertion Sequence

Related​Assertion​Sequence

SQ

1

(0044,0108)

Referenced Assertion UID

Referenced​Assertion​UID

UI

1

(0044,0109)

Approval Subject Sequence

Approval​Subject​Sequence

SQ

1

(0044,010A)

Organizational Role Code Sequence

Organizational​Role​Code​Sequence

SQ

1

(0044,0110)

RT Assertions Sequence

RT​Assertions​Sequence

SQ

1

(0046,0012)

Lens Description

Lens​Description

LO

1

(0046,0014)

Right Lens Sequence

Right​Lens​Sequence

SQ

1

(0046,0015)

Left Lens Sequence

Left​Lens​Sequence

SQ

1

(0046,0016)

Unspecified Laterality Lens Sequence

Unspecified​Laterality​Lens​Sequence

SQ

1

(0046,0018)

Cylinder Sequence

Cylinder​Sequence

SQ

1

(0046,0028)

Prism Sequence

Prism​Sequence

SQ

1

(0046,0030)

Horizontal Prism Power

Horizontal​Prism​Power

FD

1

(0046,0032)

Horizontal Prism Base

Horizontal​Prism​Base

CS

1

(0046,0034)

Vertical Prism Power

Vertical​Prism​Power

FD

1

(0046,0036)

Vertical Prism Base

Vertical​Prism​Base

CS

1

(0046,0038)

Lens Segment Type

Lens​Segment​Type

CS

1

(0046,0040)

Optical Transmittance

Optical​Transmittance

FD

1

(0046,0042)

Channel Width

Channel​Width

FD

1

(0046,0044)

Pupil Size

Pupil​Size

FD

1

(0046,0046)

Corneal Size

Corneal​Size

FD

1

(0046,0047)

Corneal Size Sequence

Corneal​Size​Sequence

SQ

1

(0046,0050)

Autorefraction Right Eye Sequence

Autorefraction​Right​Eye​Sequence

SQ

1

(0046,0052)

Autorefraction Left Eye Sequence

Autorefraction​Left​Eye​Sequence

SQ

1

(0046,0060)

Distance Pupillary Distance

Distance​Pupillary​Distance

FD

1

(0046,0062)

Near Pupillary Distance

Near​Pupillary​Distance

FD

1

(0046,0063)

Intermediate Pupillary Distance

Intermediate​Pupillary​Distance

FD

1

(0046,0064)

Other Pupillary Distance

Other​Pupillary​Distance

FD

1

(0046,0070)

Keratometry Right Eye Sequence

Keratometry​Right​Eye​Sequence

SQ

1

(0046,0071)

Keratometry Left Eye Sequence

Keratometry​Left​Eye​Sequence

SQ

1

(0046,0074)

Steep Keratometric Axis Sequence

Steep​Keratometric​Axis​Sequence

SQ

1

(0046,0075)

Radius of Curvature

Radius​Of​Curvature

FD

1

(0046,0076)

Keratometric Power

Keratometric​Power

FD

1

(0046,0077)

Keratometric Axis

Keratometric​Axis

FD

1

(0046,0080)

Flat Keratometric Axis Sequence

Flat​Keratometric​Axis​Sequence

SQ

1

(0046,0092)

Background Color

Background​Color

CS

1

(0046,0094)

Optotype

Optotype

CS

1

(0046,0095)

Optotype Presentation

Optotype​Presentation

CS

1

(0046,0097)

Subjective Refraction Right Eye Sequence

Subjective​Refraction​Right​Eye​Sequence

SQ

1

(0046,0098)

Subjective Refraction Left Eye Sequence

Subjective​Refraction​Left​Eye​Sequence

SQ

1

(0046,0100)

Add Near Sequence

Add​Near​Sequence

SQ

1

(0046,0101)

Add Intermediate Sequence

Add​Intermediate​Sequence

SQ

1

(0046,0102)

Add Other Sequence

Add​Other​Sequence

SQ

1

(0046,0104)

Add Power

Add​Power

FD

1

(0046,0106)

Viewing Distance

Viewing​Distance

FD

1

(0046,0110)

Cornea Measurements Sequence

Cornea​Measurements​Sequence

SQ

1

(0046,0111)

Source of Cornea Measurement Data Code Sequence

Sourc​eOf​Cornea​Measurement​Data​Code​Sequence

SQ

1

(0046,0112)

Steep Corneal Axis Sequence

Steep​Corneal​Axis​Sequence

SQ

1

(0046,0113)

Flat Corneal Axis Sequence

Flat​Corneal​Axis​Sequence

SQ

1

(0046,0114)

Corneal Power

Corneal​Power

FD

1

(0046,0115)

Corneal Axis

Corneal​Axis

FD

1

(0046,0116)

Cornea Measurement Method Code Sequence

Cornea​Measurement​Method​CodeS​equence

SQ

1

(0046,0117)

Refractive Index of Cornea

Refractive​Index​Of​Cornea

FL

1

(0046,0118)

Refractive Index of Aqueous Humor

Refractive​Index​Of​Aqueous​Humor

FL

1

(0046,0121)

Visual Acuity Type Code Sequence

Visual​Acuity​Type​Code​Sequence

SQ

1

(0046,0122)

Visual Acuity Right Eye Sequence

Visual​Acuity​Right​Eye​Sequence

SQ

1

(0046,0123)

Visual Acuity Left Eye Sequence

Visual​Acuity​Left​Eye​Sequence

SQ

1

(0046,0124)

Visual Acuity Both Eyes Open Sequence

Visual​Acuity​Both​Eyes​Open​Sequence

SQ

1

(0046,0125)

Viewing Distance Type

Viewing​Distance​Type

CS

1

(0046,0135)

Visual Acuity Modifiers

Visual​Acuity​Modifiers

SS

2

(0046,0137)

Decimal Visual Acuity

Decimal​Visual​Acuity

FD

1

(0046,0139)

Optotype Detailed Definition

Optotype​Detailed​Definition

LO

1

(0046,0145)

Referenced Refractive Measurements Sequence

Referenced​Refractive​Measurements​Sequence

SQ

1

(0046,0146)

Sphere Power

Sphere​Power

FD

1

(0046,0147)

Cylinder Power

Cylinder​Power

FD

1

(0046,0201)

Corneal Topography Surface

Corneal​Topography​Surface

CS

1

(0046,0202)

Corneal Vertex Location

Corneal​Vertex​Location

FL

2

(0046,0203)

Pupil Centroid X-Coordinate

Pupil​Centroid​X​Coordinate

FL

1

(0046,0204)

Pupil Centroid Y-Coordinate

Pupil​Centroid​Y​Coordinate

FL

1

(0046,0205)

Equivalent Pupil Radius

Equivalent​Pupil​Radius

FL

1

(0046,0207)

Corneal Topography Map Type Code Sequence

Corneal​Topography​Map​Type​Code​Sequence

SQ

1

(0046,0208)

Vertices of the Outline of Pupil

Vertices​Of​The​Outline​Of​Pupil

IS

2-2n

(0046,0210)

Corneal Topography Mapping Normals Sequence

Corneal​Topography​Mapping​Normals​Sequence

SQ

1

(0046,0211)

Maximum Corneal Curvature Sequence

Maximum​Corneal​Curvature​Sequence

SQ

1

(0046,0212)

Maximum Corneal Curvature

Maximum​Corneal​Curvature

FL

1

(0046,0213)

Maximum Corneal Curvature Location

Maximum​Corneal​Curvature​Location

FL

2

(0046,0215)

Minimum Keratometric Sequence

Minimum​Keratometric​Sequence

SQ

1

(0046,0218)

Simulated Keratometric Cylinder Sequence

Simulated​Keratometric​Cylinder​Sequence

SQ

1

(0046,0220)

Average Corneal Power

Average​Corneal​Power

FL

1

(0046,0224)

Corneal I-S Value

Corneal​IS​Value

FL

1

(0046,0227)

Analyzed Area

Analyzed​Area

FL

1

(0046,0230)

Surface Regularity Index

Surface​Regularity​Index

FL

1

(0046,0232)

Surface Asymmetry Index

Surface​Asymmetry​Index

FL

1

(0046,0234)

Corneal Eccentricity Index

Corneal​Eccentricity​Index

FL

1

(0046,0236)

Keratoconus Prediction Index

Keratoconus​Prediction​Index

FL

1

(0046,0238)

Decimal Potential Visual Acuity

Decimal​Potential​Visual​Acuity

FL

1

(0046,0242)

Corneal Topography Map Quality Evaluation

Corneal​Topography​Map​Quality​Evaluation

CS

1

(0046,0244)

Source Image Corneal Processed Data Sequence

Source​Image​Corneal​Processed​Data​Sequence

SQ

1

(0046,0247)

Corneal Point Location

Corneal​Point​Location

FL

3

(0046,0248)

Corneal Point Estimated

Corneal​Point​Estimated

CS

1

(0046,0249)

Axial Power

Axial​Power

FL

1

(0046,0250)

Tangential Power

Tangential​Power

FL

1

(0046,0251)

Refractive Power

Refractive​Power

FL

1

(0046,0252)

Relative Elevation

Relative​Elevation

FL

1

(0046,0253)

Corneal Wavefront

Corneal​Wavefront

FL

1

(0048,0001)

Imaged Volume Width

Imaged​Volume​Width

FL

1

(0048,0002)

Imaged Volume Height

Imaged​Volume​Height

FL

1

(0048,0003)

Imaged Volume Depth

Imaged​Volume​Depth

FL

1

(0048,0006)

Total Pixel Matrix Columns

Total​Pixel​Matrix​Columns

UL

1

(0048,0007)

Total Pixel Matrix Rows

Total​Pixel​Matrix​Rows

UL

1

(0048,0008)

Total Pixel Matrix Origin Sequence

Total​Pixel​Matrix​Origin​Sequence

SQ

1

(0048,0010)

Specimen Label in Image

Specimen​Label​In​Image

CS

1

(0048,0011)

Focus Method

Focus​Method

CS

1

(0048,0012)

Extended Depth of Field

Extended​Depth​Of​Field

CS

1

(0048,0013)

Number of Focal Planes

Number​Of​Focal​Planes

US

1

(0048,0014)

Distance Between Focal Planes

Distance​Between​Focal​Planes

FL

1

(0048,0015)

Recommended Absent Pixel CIELab Value

Recommended​Absent​Pixel​CIE​Lab​Value

US

3

(0048,0100)

Illuminator Type Code Sequence

Illuminator​Type​Code​Sequence

SQ

1

(0048,0102)

Image Orientation (Slide)

Image​Orientation​Slide

DS

6

(0048,0105)

Optical Path Sequence

Optical​Path​Sequence

SQ

1

(0048,0106)

Optical Path Identifier

Optical​Path​Identifier

SH

1

(0048,0107)

Optical Path Description

Optical​Path​Description

ST

1

(0048,0108)

Illumination Color Code Sequence

Illumination​Color​Code​Sequence

SQ

1

(0048,0110)

Specimen Reference Sequence

Specimen​Reference​Sequence

SQ

1

(0048,0111)

Condenser Lens Power

Condenser​Lens​Power

DS

1

(0048,0112)

Objective Lens Power

Objective​Lens​Power

DS

1

(0048,0113)

Objective Lens Numerical Aperture

Objective​Lens​Numerical​Aperture

DS

1

(0048,0114)

Confocal Mode

Confocal​Mode

CS

1

(0048,0115)

Tissue Location

Tissue​Location

CS

1

(0048,0116)

Confocal Microscopy Image Frame Type Sequence

Confocal​Microscopy​Image​Frame​Type​Sequence

SQ

1

(0048,0117)

Image Acquisition Depth

Image​Acquisition​Depth

FD

1

(0048,0120)

Palette Color Lookup Table Sequence

Palette​Color​Lookup​Table​Sequence

SQ

1

(0048,0200)

Referenced Image Navigation Sequence

Referenced​Image​Navigation​Sequence

SQ

1

RET (2021c)

(0048,0201)

Top Left Hand Corner of Localizer Area

Top​Left​Hand​Corner​Of​Localizer​Area

US

2

RET (2021c)

(0048,0202)

Bottom Right Hand Corner of Localizer Area

Bottom​Right​Hand​Corner​Of​Localizer​Area

US

2

RET (2021c)

(0048,0207)

Optical Path Identification Sequence

Optical​Path​Identification​Sequence

SQ

1

(0048,021A)

Plane Position (Slide) Sequence

Plane​Position​Slide​Sequence

SQ

1

(0048,021E)

Column Position In Total Image Pixel Matrix

Column​Position​In​Total​Image​Pixel​Matrix

SL

1

(0048,021F)

Row Position In Total Image Pixel Matrix

Row​Position​In​Total​Image​Pixel​Matrix

SL

1

(0048,0301)

Pixel Origin Interpretation

Pixel​Origin​Interpretation

CS

1

(0048,0302)

Number of Optical Paths

Number​Of​Optical​Paths

UL

1

(0048,0303)

Total Pixel Matrix Focal Planes

Total​Pixel​Matrix​Focal​Planes

UL

1

(0048,0304)

Tiles Overlap

Tiles​Overlap

CS

1

(0050,0004)

Calibration Image

Calibration​Image

CS

1

(0050,0010)

Device Sequence

Device​Sequence

SQ

1

(0050,0012)

Container Component Type Code Sequence

Container​Component​Type​Code​Sequence

SQ

1

(0050,0013)

Container Component Thickness

Container​Component​Thickness

FD

1

(0050,0014)

Device Length

Device​Length

DS

1

(0050,0015)

Container Component Width

Container​Component​Width

FD

1

(0050,0016)

Device Diameter

Device​Diameter

DS

1

(0050,0017)

Device Diameter Units

Device​Diameter​Units

CS

1

(0050,0018)

Device Volume

Device​Volume

DS

1

(0050,0019)

Inter-Marker Distance

Inter​Marker​Distance

DS

1

(0050,001A)

Container Component Material

Container​Component​Material

CS

1

(0050,001B)

Container Component ID

Container​Component​ID

LO

1

(0050,001C)

Container Component Length

Container​Component​Length

FD

1

(0050,001D)

Container Component Diameter

Container​Component​Diameter

FD

1

(0050,001E)

Container Component Description

Container​Component​Description

LO

1

(0050,0020)

Device Description

Device​Description

LO

1

(0050,0021)

Long Device Description

Long​Device​Description

ST

1

(0052,0001)

Contrast/Bolus Ingredient Percent by Volume

Contrast​Bolus​Ingredient​Percent​By​Volume

FL

1

(0052,0002)

OCT Focal Distance

OCT​Focal​Distance

FD

1

(0052,0003)

Beam Spot Size

Beam​Spot​Size

FD

1

(0052,0004)

Effective Refractive Index

Effective​Refractive​Index

FD

1

(0052,0006)

OCT Acquisition Domain

OCT​Acquisition​Domain

CS

1

(0052,0007)

OCT Optical Center Wavelength

OCT​Optical​Center​Wavelength

FD

1

(0052,0008)

Axial Resolution

Axial​Resolution

FD

1

(0052,0009)

Ranging Depth

Ranging​Depth

FD

1

(0052,0011)

A-line Rate

A​Line​Rate

FD

1

(0052,0012)

A-lines Per Frame

A​Lines​Per​Frame

US

1

(0052,0013)

Catheter Rotational Rate

Catheter​Rotational​Rate

FD

1

(0052,0014)

A-line Pixel Spacing

A​Line​Pixel​Spacing

FD

1

(0052,0016)

Mode of Percutaneous Access Sequence

Mode​Of​Percutaneous​Access​Sequence

SQ

1

(0052,0025)

Intravascular OCT Frame Type Sequence

Intravascular​OCT​Frame​Type​Sequence

SQ

1

(0052,0026)

OCT Z Offset Applied

OCTZ​Offset​Applied

CS

1

(0052,0027)

Intravascular Frame Content Sequence

Intravascular​Frame​Content​Sequence

SQ

1

(0052,0028)

Intravascular Longitudinal Distance

Intravascular​Longitudinal​Distance

FD

1

(0052,0029)

Intravascular OCT Frame Content Sequence

Intravascular​OCT​Frame​Content​Sequence

SQ

1

(0052,0030)

OCT Z Offset Correction

OCTZ​Offset​Correction

SS

1

(0052,0031)

Catheter Direction of Rotation

Catheter​Direction​Of​Rotation

CS

1

(0052,0033)

Seam Line Location

Seam​Line​Location

FD

1

(0052,0034)

First A-line Location

First​A​Line​Location

FD

1

(0052,0036)

Seam Line Index

Seam​Line​Index

US

1

(0052,0038)

Number of Padded A-lines

Number​Of​Padded​ALines

US

1

(0052,0039)

Interpolation Type

Interpolation​Type

CS

1

(0052,003A)

Refractive Index Applied

Refractive​Index​Applied

CS

1

(0054,0010)

Energy Window Vector

Energy​Window​Vector

US

1-n

(0054,0011)

Number of Energy Windows

Number​Of​Energy​Windows

US

1

(0054,0012)

Energy Window Information Sequence

Energy​Window​Information​Sequence

SQ

1

(0054,0013)

Energy Window Range Sequence

Energy​Window​Range​Sequence

SQ

1

(0054,0014)

Energy Window Lower Limit

Energy​Window​Lower​Limit

DS

1

(0054,0015)

Energy Window Upper Limit

Energy​Window​Upper​Limit

DS

1

(0054,0016)

Radiopharmaceutical Information Sequence

Radio​pharmaceutical​Information​Sequence

SQ

1

(0054,0017)

Residual Syringe Counts

Residual​Syringe​Counts

IS

1

(0054,0018)

Energy Window Name

Energy​Window​Name

SH

1

(0054,0020)

Detector Vector

Detector​Vector

US

1-n

(0054,0021)

Number of Detectors

Number​Of​Detectors

US

1

(0054,0022)

Detector Information Sequence

Detector​Information​Sequence

SQ

1

(0054,0030)

Phase Vector

Phase​Vector

US

1-n

(0054,0031)

Number of Phases

Number​Of​Phases

US

1

(0054,0032)

Phase Information Sequence

Phase​Information​Sequence

SQ

1

(0054,0033)

Number of Frames in Phase

Number​Of​Frames​In​Phase

US

1

(0054,0036)

Phase Delay

Phase​Delay

IS

1

(0054,0038)

Pause Between Frames

Pause​Between​Frames

IS

1

(0054,0039)

Phase Description

Phase​Description

CS

1

(0054,0050)

Rotation Vector

Rotation​Vector

US

1-n

(0054,0051)

Number of Rotations

Number​Of​Rotations

US

1

(0054,0052)

Rotation Information Sequence

Rotation​Information​Sequence

SQ

1

(0054,0053)

Number of Frames in Rotation

Number​Of​Frames​In​Rotation

US

1

(0054,0060)

R-R Interval Vector

RR​Interval​Vector

US

1-n

(0054,0061)

Number of R-R Intervals

Number​Of​RR​Intervals

US

1

(0054,0062)

Gated Information Sequence

Gated​Information​Sequence

SQ

1

(0054,0063)

Data Information Sequence

Data​Information​Sequence

SQ

1

(0054,0070)

Time Slot Vector

Time​Slot​Vector

US

1-n

(0054,0071)

Number of Time Slots

Number​Of​Time​Slots

US

1

(0054,0072)

Time Slot Information Sequence

Time​Slot​Information​Sequence

SQ

1

(0054,0073)

Time Slot Time

Time​Slot​Time

DS

1

(0054,0080)

Slice Vector

Slice​Vector

US

1-n

(0054,0081)

Number of Slices

Number​Of​Slices

US

1

(0054,0090)

Angular View Vector

Angular​View​Vector

US

1-n

(0054,0100)

Time Slice Vector

Time​Slice​Vector

US

1-n

(0054,0101)

Number of Time Slices

Number​Of​Time​Slices

US

1

(0054,0200)

Start Angle

Start​Angle

DS

1

(0054,0202)

Type of Detector Motion

Type​Of​Detector​Motion

CS

1

(0054,0210)

Trigger Vector

Trigger​Vector

IS

1-n

(0054,0211)

Number of Triggers in Phase

Number​Of​Triggers​In​Phase

US

1

(0054,0220)

View Code Sequence

View​Code​Sequence

SQ

1

(0054,0222)

View Modifier Code Sequence

View​Modifier​Code​Sequence

SQ

1

(0054,0300)

Radionuclide Code Sequence

Radionuclide​Code​Sequence

SQ

1

(0054,0302)

Administration Route Code Sequence

Administration​Route​Code​Sequence

SQ