| DICOM PS3.18 2025e - Web Services |
|---|
The "swivelrange" parameter specifies the angular range over which a rendered volume rotates around the swivel axis, which is defined as the axis parallel to the "viewpointup" intersecting the "viewpointlookat". The rendered volume rotates back and forth.
The syntax of this parameter is:
%s"swivelrange =" sr
This corresponds to the differently capitalized SWIVEL value of Presentation Animation Style (0070,1A01) and Swivel Range (0070,1A06). See Section C.11.29 in PS3.3 and Section FF.2.4.2 in PS3.4.
The origin server shall create an animation with a number of frames equal toSwivel Range divided by the "animationstepsize".
If the "swivelrange" parameter is present and the "animationrate" parameter is not present, the origin server shall determine the animation rate.
If the Target Resource is a Volumetric Presentation State, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.
| DICOM PS3.18 2025e - Web Services |
|---|