| DICOM PS3.18 2025e - Web Services |
|---|
The "mprslab" parameter specifies the thickness of the MPR plane. This parameter results in an orthographic rendering with a defined thickness using the method defined by "renderingmethod". See Section C.11.26.1.1 in PS3.3 for more information.
The syntax of this parameter for a Rendered MPR Volume is:
%s"mprslab =" st
This corresponds to the MPR Slab Thickness (0070,1503) attribute. See Section C.11.26 in PS3.3.
The slab thickness of the returned media might not match the requested thickness due to the voxel size of the Target Resource.
If "renderingmethod" is not present, the origin server may apply a default rendering method, based on the resource and/or slab thickness, or alternatively, return 400 (Bad Request) and may include an appropriate Status Report.
If the Target Resource is a Volumetric Presentation State, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.
| DICOM PS3.18 2025e - Web Services |
|---|