DICOM PS3.18 2025b - Web Services |
---|
The "animationrate" parameter specifies the rate at which an animated 3D or MPR Volumetric Presentation is displayed.
The syntax of this parameter is:
%s"animationrate=" rt
This corresponds to Recommended Animation Rate (0070,1A03) in Section C.11.29 in PS3.3 and Section FF.2.4.2 in PS3.4.
Playback of the returned media on a client may or may not achieve the requested animation rate.
If "animationrate" is notpresent, and other animation parameters are present (e.g., "swivelrange", "animationstepsize", or "volumetriccurvepoint"), the origin server may apply a default animation rate, or alternatively, return 400 (Bad Request) and may include an appropriate Status Report.
If the Target Resource is a Volumetric Presentation State, the origin server shall return a 400 (Bad Request) and may include an appropriate Status Report.
DICOM PS3.18 2025b - Web Services |
---|