DICOM PS3.18 2025e - Web Services

14.3 Transactions Overview

The Modality Scheduled Procedure Step Service consists of the Transactions listed in Table 14.3-1.

Table 14.3-1. Modality Scheduled Procedure Step Service Transactions

Transaction Name

Method

Payload

Description

Request

Success Response

Search

GET

none

Data Set according to Table K.6-1 in PS3.4

Searches for Modality Scheduled Procedure Steps


Table 14.3-2 lists the Modality Scheduled Procedure Step Service Transactions and their corresponding DIMSE Operations used in MWL.

Table 14.3-2. Mapping of Modality Scheduled Procedure Step Service Transactions and DIMSE Operations

Transaction

Operation

Reference

DIMSE Service

Search

Query Worklist

Section K.4 in PS3.4

C-FIND


Note

As in DIMSE, the Transactions do not provide a complete CRUDL interface for the respective resource. For instance, it is not possible to create Modality Scheduled Procedure Steps using DICOM, neither with DIMSE, nor with DICOMweb. What DICOM does provide is access to scheduled procedure steps at the level required for modalities.

DICOM PS3.18 2025e - Web Services