A success response payload shall contain the search results in an Acceptable Media Type. See Section 8.7.5. If (results = 0) then there are no matches, and a 200 (OK) response shall be returned with an empty payload. The empty payload shall be formatted as specified in Section 8.3.4.4.1. See Section 8.6.
Each search result in the payload shall comply with all requirements in the Matching Key Type and Return Key Type columns of Table CC.2.5-3 “UPS SOP Class N-CREATE/N-SET/N-GET/C-FIND Attributes” in PS3.4.
A failure response payload may contain a Status Report describing any failures, warnings, or other useful information.