DICOM PS3.18 2026d - Web Services

K Rendered Volume Response Module

The Rendered Volume Response Module provides the user agent with a representation of the parameters applied by the origin server to generate the volumetric rendering.

The user agent may use this information to:

K.1 Response Message Body

Table K.1-1 defines the Attributes that are returned in a Rendered MPR Volume Resource or a Rendered 3D Volume Resource response message body.

Note

  1. These represent Query Parameters that may be specified by the user agent in Rendered MPR Volume Resources or Rendered 3D Volume Resources. See Section 8.3.5.3.

  2. Anatomic orientation parameters (see Section 8.3.5.3.4) are converted to camera orientation parameters to facilitate fine grain adjustments in a subsequent request.

Table K.1-1. Rendered Volume Response Module Attributes

Attribute Name

Tag

Type

Attribute Description

Reformatting Operation Type

(0072,0510)

1

Reformatting operation to be applied to the Image Set.

Rendering Method

(0070,120D)

1

Specifies the display algorithm to be applied to the Volume Data.

Viewpoint Position

(0070,1603)

1

Position of the viewpoint in volume space.

Viewpoint LookAt Point

(0070,1604)

1

Point the viewpoint is looking at.

Viewpoint Up Direction

(0070,1605)

1

Vertical orientation of the view.

MPR Slab Thickness

(0070,1503)

1C

Required if Reformatting Operation Type (0072,0510) has a value of MPR and there is a specified thickness.

VOI LUT Function

(0028,1056)

1C

Required if Rendering Method (0070,120D) is not VOLUME_RENDERED.

Window Width

(0028,1051)

1C

Required if Rendering Method (0070,120D) is not VOLUME_RENDERED.

Window Center

(0028,1050)

1C

Required if Rendering Method (0070,120D) is not VOLUME_RENDERED.

Swivel Range

(0070,1A06)

1C

Required for SWIVEL animations.

Animation Step Size

(0070,1A05)

1C

Required for SWIVEL or CROSSCURVE animations.

Recommended Animation Rate

(0070,1A03)

1C

Required for video Media Types.


DICOM PS3.18 2026d - Web Services