DICOM PS3.18 2026c - Web Services

H Capabilities Description

A Capabilities Description is a WADL Document. See [WADL].

The Capabilities Description resource follows directly and unambiguously from the Representational State Transfer (REST) Web Service resources defined in Section 10, Section 11, Section 12, Section 13, Section 14, and Section 15.

The WADL document shall contain one top-level "application" element.

The "application" element shall contain one "resources" element whose "base" attribute value is the base URL for the service. This may be a combination of protocol (either http or https), host, port, and application.

Additionally, the WADL content shall include a "resource" element for the Target Resource specified in the request describing all methods and child resources for the specified resource and each of its children.

The full resource tree and the methods defined for each resource are described in Table H-1.

Note

The Retrieve Capabilities Transaction is excluded from Table H-1 because that Transaction is used to retrieve this document and WADL is not self-describing.

Table H-1. Resources and Methods

Service

Resource

Transactions

Reference

Studies (see Section 10.1.1)

studies

Search for Studies

Store Instances

Section 10.6

Section 10.5

{StudyInstance}

Retrieve Study

Store Study Instances

Section 10.4

Section 10.5

metadata

Retrieve Study Metadata

Section 10.4

rendered

Retrieve Rendered Study

Section 10.4

renderedmpr

Retrieve Rendered MPR Volume Study

Section 10.4

rendered3d

Retrieve Rendered 3D Volume Study

Section 10.4

thumbnail

Retrieve Study Thumbnail

Section 10.4

bulkdata

Retrieve Study Bulk Data

Section 10.4

pixeldata

Retrieve Study Pixel Data

Section 10.4

series

Search for Study Series

Section 10.6

{SeriesInstance}

Retrieve Series

Section 10.4

metadata

Retrieve Series Metadata

Section 10.4

rendered

Retrieve Rendered Series

Section 10.4

renderedmpr

Retrieve Rendered MPR Volume Series

Section 10.4

rendered3d

Retrieve Rendered 3D Volume Series

Section 10.4

thumbnail

Retrieve Series Thumbnail

Section 10.4

bulkdata

Retrieve Series Bulk Data

Section 10.4

pixeldata

Retrieve Series Pixel Data

Section 10.4

instances

Search for Study Series Instances

Section 10.4

{SOPInstance}

Retrieve Instance

Section 10.4

metadata

Retrieve Instance Metadata

Section 10.4

rendered

Retrieve Rendered Instance

Section 10.4

renderedmpr

Retrieve Rendered MPR Volume Instance

Section 10.4

rendered3d

Retrieve Rendered 3D Volume Instance

Section 10.4

thumbnail

Retrieve Instance Thumbnail

Section 10.4

bulkdata

Retrieve Instance Bulk Data

Section 10.4

pixeldata

Retrieve Instance Pixel Data

Section 10.4

frames

N/A

N/A

{frames}

Retrieve Frames

Section 10.4

rendered

Retrieve Rendered Frames

Section 10.4

renderedmpr

Retrieve Rendered MPR Volume Frames

Section 10.4

rendered3d

Retrieve Rendered 3D Volume Frames

Section 10.4

thumbnail

Retrieve Frame Thumbnail

Section 10.4

pixeldata

Retrieve Frame Pixel Data

Section 10.4

instances

Search for Study Instances

Section 10.6

series

Search for Series

Section 10.6

{SeriesInstance}

N/A

N/A

{instances}

Search for Instances

Section 10.6

instances

Search for Instances

Section 10.6

{BulkDataReference}

Retrieve Bulk Data

Section 10.4

Worklist (see Section 11.1.1)

workitems

Search for Workitem

Create Workitem

Section 11.9

Section 11.4

{Workitem}

Retrieve Workitem

Update Workitem

Section 11.4

Section 11.6

state

Change Workitem State

Section 11.7

cancelrequest

Request Workitem Cancellation

Section 11.8

subscribers

N/A

N/A

{subscriber}

Subscribe

Unsubscribe

Section 11.10

Section 11.11

1.2.840.10008.5.1.4.34.5

N/A

N/A

subscribers

N/A

N/A

{subscriber}

Subscribe

Unsubscribe

Section 11.10

Section 11.11

suspend

Unsubscribe

Section 11.11

1.2.840.10008.5.1.4.34.5.1

N/A

N/A

subscribers

N/A

N/A

{subscriber}

Subscribe

Unsubscribe

Section 11.10

Section 11.11

suspend

Suspend Worklist Subscription

Section 11.11

Non-Patient Instance (see Section 12.1.1)

color-palettes

N/A

N/A

{uid}

Retrieve

Store

Search

Section 12.4

Section 12.5

Section 12.6

defined-procedure-protocol

N/A

N/A

{uid}

Retrieve

Store

Search

Section 12.4

Section 12.5

Section 12.6

hanging-protocol

N/A

N/A

{uid}

Retrieve

Store

Search

Section 12.4

Section 12.5

Section 12.6

implant-templates

N/A

N/A

{uid}

Retrieve

Store

Search

Section 12.4

Section 12.5

Section 12.6

inventories

N/A

N/A

{uid}

Retrieve

Store

Search

Section 12.4

Section 12.5

Section 12.6

protocol-approvals

N/A

N/A

{uid}

Retrieve

Store

Search

Section 12.4

Section 12.5

Section 12.6

Storage Commitment requests (see Section 13.1.1)

commitment-requests

Commit

Section 13.4

Check Commit Result

Section 13.5

Modality Scheduled Procedure Step Service (see Section 14.1.1)

modality-scheduled-procedure-steps

Search

Section 14.4

Modality Performed Procedure Step Service (see Section 15.1.1)

modality-performed-procedure-steps

Create

Section 15.4

Update

Section 15.5

Retrieve

Section 15.6


DICOM PS3.18 2026c - Web Services