| DICOM PS3.18 2026c - Web Services |
|---|
The Modality Scheduled Procedure Step Service enables a user agent to search for Scheduled Procedure Steps, and entities related to these steps, intended to be performed on an imaging modality. It corresponds to the DIMSE Modality Worklist (MWL) service as defined in Annex K of PS3.4 and has the same semantics.
The Modality Scheduled Procedure Step Service provides access to a collection of Modality Scheduled Procedure Steps, defined as the resource given in Table 14.1.1-1.
The origin server shall support Query Parameters as required in Table 14.1.2-1.
The user agent shall supply in the request Query Parameters as required in Table 14.1.2-1.
The origin server shall support the media types specified as Default or Required in Table 14.1.3-1.
| DICOM PS3.18 2026c - Web Services |
|---|