DICOM PS3.17 2026d - Explanatory Information

UUUU.5 Radiation Dose Structured Report Examples

This section contains examples of encoding of Radiation Dose Structured Report.

In these examples, the following convention is used to describe the Content Items and values:

Note that other alternative implementation patterns exist for these examples; the normative text in PS3.16 is constraining, these examples are not.

UUUU.5.1 Examples of Traditional RDSR

UUUU.5.1.1 Example of Traditional RDSR for XA

The following is an example RDSR for a routine XA procedure step combining diagnostic and interventional treatment. A single plane Interventional X-Ray acquisition system (Irradiating Device) performs an exam (one procedure step) with the following characteristics:

  • Patient position is Head First Supine.

  • The top of the patient head is located at 25 cm from the table head in the direction of the table foot.

  • One focused anti-scatter grid is fixed on the detector housing.

  • One pulsed fluoroscopy Irradiation Event lasting 10 seconds, with no XA image recorded.

  • One rotational acquisition (CBCT) Irradiation Event, rotating at 10 degrees per second and acquiring 30 frames per second over an arc of 200 degrees. An XA image is recorded.

  • A dose map image is created at the end of the procedure step.

  • The irradiation is administered by the performing physician Dr. Hugh Mann and by the operator John Doe.

  • The irradiation is authorized by the performing physician Dr. Hugh Mann.

Table UUUU.5.1.1-1 illustrates one possible encoding of this example. Note that other alternative implementation patterns exist.

Table UUUU.5.1.1-1. XA Traditional RDSR Example

Node

Code Meaning of Concept Name

Code or Example Value

TID

1

X-Ray Radiation Dose Report

TID 10001

1.1

Language of Content Item and Descendants

(en, RFC5646, "English")

TID 1204

1.2

Procedure reported

(113704, DCM, "Projection X-Ray")

TID 10001

1.2.1

Has Intent

(1279505009, SCT, "Combined Diagnostic and Therapeutic Intent")

TID 10001

1.3

Acquisition Device Type

(113957, DCM, "Fluoroscopy-Guided Projection Radiography System")

TID 10001

Start Observer Context

Observer #1: Irradiating device

1.4

Observer Type

(121007, DCM, "Device")

TID 1002

1.5

Device Observer UID

2.999.1

TID 1004

1.6

Device Observer Name

MyStationName

TID 1004

1.7

Device Observer Manufacturer

Manufacturer X

TID 1004

1.8

Device Observer Model Name

Model Y

TID 1004

1.9

Device Observer Serial Number

SerialNumber123

TID 1004

1.10

Device Role in Procedure

(113859, DCM, "Irradiating Device")

TID 1004

<End Observer Context>

1.11

Scope of Accumulation

(113016, DCM, "Performed Procedure Step")

TID 10001

1.12

Performed Procedure Step SOP Instance UID

2.999.2

TID 10001

1.13

X-Ray Detector Data Available

(373066001, SCT, "Yes")

TID 10001

1.14

X-Ray Source Data Available

(373066001, SCT, "Yes")

TID 10001

1.15

X-Ray Mechanical Data Available

(373066001, SCT, "Yes")

TID 10001

1.16

Accumulated X-Ray Dose Data

TID 10002

1.16.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10002

1.16.2

Fluoro Dose Area Product Total

0.00000350 Gy.m2

TID 10004

1.16.3

Fluoro Dose (RP) Total

0.00007700 Gy

TID 10004

1.16.4

Total Fluoro Time

10 s

TID 10004

1.16.5

Acquisition Dose Area Product Total

0.00003000 Gy.m2

TID 10004

1.16.6

Acquisition Dose (RP) Total

0.00081700 Gy

TID 10004

1.16.7

Total Acquisition Time

30 s

TID 10004

1.16.8

Dose Area Product Total

0.00003410 Gy.m2

TID 10007

1.16.9

Dose (RP) Total

0.00089400 Gy

TID 10007

1.16.10

Distance Source to Reference Point

570.00 mm

TID 10007

1.16.11

Total Number of Radiographic Frames

600 no units

TID 10007

1.16.12

Reference Point Definition

(113860, DCM, "15cm from Isocenter toward Source")

TID 10007

1.16.13

Equipment Landmark

(128751, DCM, "Center of Table Head")

TID 10002

1.16.13.1

Equipment Landmark X Position

50.00 mm

TID 10002

1.16.13.2

Equipment Landmark Z Position

400.00 mm

TID 10002

1.16.14

Patient Location Fiducial"

TID 10002

1.16.14.1

Reference Basis

(88986008, SCT, "Vertex of Head")

TID 400

1.16.14.2

Reference Geometry

(128120, DCM, "Plane through Superior Extent")

TID 400

1.16.14.3

Equipment Landmark to Patient Fiducial Z Distance

-250.00 mm

TID 10002

1.17

Irradiation Event X-Ray Data

TID 10003

1.17.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10003

1.17.2

Irradiation Event UID

2.999.3

TID 10003

1.17.3

DateTime Started

20231020125921.000

TID 10003

1.17.4

Irradiation Event Type

(44491008, SCT, "Fluoroscopy")

TID 10003

1.17.5

Acquisition Protocol

Cardiac and Vascular FLUORO

TID 10003

1.17.6

Patient Table Relationship

(102540008, SCT, "headfirst")

TID 10003

1.17.7

Patient Orientation

(102538003, SCT, "recumbent")

TID 10003

1.17.7.1

Patient Orientation Modifier

(40199007, SCT, "supine")

TID 10003

1.17.8

Target Region

(80891009, SCT, "Heart")

TID 10003

1.17.9

Dose Area Product

0.00000310 Gy.m2

TID 10003

1.17.10

Patient Equivalent Thickness

300.0 mm

TID 10003

1.17.11

Comment

Fluoro Loop

TID 10003

Start Person Participant within TID 10003: Irradiation Administering are the performing physician and the operator.

1.17.12

Person Name

Mann^Hugh^^Dr

TID 1020

1.17.12.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.17.12.2

Person Role in Organization

(309343006, SCT, "Physician")

TID 1020

1.17.13

Person Name

Doe^John

TID 1020

1.17.13.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.17.13.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

Irradiation Event X-Ray Source Data TID 10003B

1.17.14

Dose (RP)

0.00007700 Gy

TID 10003B

1.17.15

Reference Point Definition

(113860, DCM, "15cm from Isocenter toward Source")

TID 10003B

1.17.16

Fluoro Mode

(113631, DCM, "Pulsed")

TID 10003B

1.17.17

Pulse Rate

15 pulse/s

TID 10003B

1.17.18

Number of Pulses

150 no units

TID 10003B

1.17.19

Pulse Width

2.01 ms

TID 10003B

1.17.20

Irradiation Duration

10.000 s

TID 10003B

1.17.21

KVP

71 kV

TID 10003B

1.17.22

X-Ray Tube Current

12 mA

TID 10003B

1.17.23

Average X-Ray Tube Current

12 mA

TID 10003B

1.17.24

Exposure Time

20 ms

TID 10003B

1.17.25

Exposure

250 uA.s

TID 10003B

1.17.26

Focal Spot Size

0.6 mm

TID 10003B

1.17.27

Anode Target Material

(26194003, SCT, "Tungsten")

TID 10003B

1.17.28

X-Ray Filters

TID 10003B

1.17.28.1

X-Ray Filter Type

(113653, DCM, "Flat filter")

TID 10003B

1.17.28.2

X-Ray Filter Material

(66925006, SCT, "Copper")

TID 10003B

1.17.28.3

X-Ray Filter Thickness Minimum

0.300000 mm

TID 10003B

1.17.28.4

X-Ray Filter Thickness Maximum

0.300000 mm

TID 10003B

1.17.29

X-Ray Filters

TID 10003B

1.17.29.1

X-Ray Filter Type

(113651, DCM, "Wedge filter")

TID 10003B

1.17.29.2

X-Ray Filter Material

(66925006, SCT, "Copper")

TID 10003B

1.17.29.3

X-Ray Filter Thickness Minimum

1.000000 mm

TID 10003B

1.17.29.4

X-Ray Filter Thickness Maximum

2.000000 mm

TID 10003B

1.17.30

Collimated Field Area

0.010781 m^2

TID 10003B

1.17.31

Collimated Field Height

129 mm

TID 10003B

1.17.32

Collimated Field Width

83 mm

TID 10003B

1.17.33

X-Ray Grid

(111641, DCM, "Fixed grid")

TID 10003B

1.17.34

X-Ray Grid

(111642, DCM, "Focused grid")

TID 10003B

Irradiation Event X-Ray Mechanical Data TID 10003C

1.17.35

Positioner Primary Angle

23.70 deg

TID 10003C

1.17.36

Positioner Secondary Angle

10.30 deg

TID 10003C

1.17.37

Table Head Tilt Angle

0.0 deg

TID 10003C

1.17.38

Table Horizontal Rotation Angle

0.0 deg

TID 10003C

1.17.39

Table Cradle Tilt Angle

0.0 deg

TID 10003C

Start CID 10008 “Dose Related Distance Measurement”

1.17.40

Distance Source to Isocenter

720 mm

TID 10003C

1.17.41

Distance Source to Reference Point

570.00 mm

TID 10003C

1.17.42

Distance Source to Detector

1195 mm

TID 10003C

1.17.43

Table Longitudinal Position

727.90 mm

TID 10003C

1.17.44

Table Lateral Position

50.90 mm

TID 10003C

1.17.45

Table Height Position

87.5 mm

TID 10003C

1.17.46

Table X Position to Isocenter

727.90 mm

TID 10003C

1.17.47

Table Y Position to Isocenter

87.5 mm

TID 10003C

1.17.48

Table Z Position to Isocenter

50.90 mm

TID 10003C

End CID 10008 “Dose Related Distance Measurement”

1.17.49

Positioner Isocenter Primary Angle

23.70 deg

TID 10003C

1.17.50

Positioner Isocenter Secondary Angle

10.30 deg

TID 10003C

1.17.51

Positioner Isocenter Detector Rotation Angle

0.0 deg

TID 10003C

1.18

Irradiation Event X-Ray Data

TID 10003

1.18.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10003

1.18.2

Irradiation Event UID

2.999.4

TID 10003

1.18.3

DateTime Started

20231020130412.000

TID 10003

1.18.4

Irradiation Event Type

(113613, DCM, "Rotational Acquisition")

TID 10003

1.18.5

Acquisition Protocol

Cardiac and Vascular FLUORO

TID 10003

1.18.6

Patient Table Relationship

(102540008, SCT, "headfirst")

TID 10003

1.18.7

Patient Orientation

(102538003, SCT, "recumbent")

TID 10003

1.18.7.1

Patient Orientation Modifier

(40199007, SCT, "supine")

TID 10003

1.18.8

Target Region

(80891009, SCT, "Heart")

TID 10003

1.18.9

Dose Area Product

0.00003100 Gy.m2

TID 10003

1.18.10

Patient Equivalent Thickness

300.0 mm

TID 10003

1.18.11

Comment

Rotational CBCT

TID 10003

Start Person Participant within TID 10003

Person #1: Performing physician

1.18.12

Person Name

Mann^Hugh^^Dr

TID 1020

1.18.12.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.18.12.2

Person Role in Organization

(309343006, SCT, "Physician")

TID 1020

Person #2: Operator

1.18.13

Person Name

Doe^John

TID 1020

1.18.13.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.18.13.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

Irradiation Event X-Ray Detector Data TID 10003A

1.18.14

Acquired Image

SOP Class UID: 1.2.840.10008.5.1.4.1.1.12.1.1

SOP Instance UID: 2.999.5

TID 10003A

Irradiation Event X-Ray Source Data TID 10003B

1.18.15

Dose (RP)

0.00081700 Gy

TID 10003B

1.18.16

Reference Point Definition

(113860, DCM, "15cm from Isocenter toward Source")

TID 10003B

1.18.17

Number of Pulses

600 no units

TID 10003B

1.18.18

Pulse Width

5.00 ms

TID 10003B

1.18.19

Irradiation Duration

20.000 s

TID 10003B

1.18.20

KVP

75 kV

TID 10003B

1.18.21

X-Ray Tube Current

20 mA

TID 10003B

1.18.22

Average X-Ray Tube Current

20 mA

TID 10003B

1.18.23

Exposure Time

3000 ms

TID 10003B

1.18.24

Exposure

940 uA.s

TID 10003B

1.18.25

Focal Spot Size

0.6 mm

TID 10003B

1.18.26

Anode Target Material

(26194003, SCT, "Tungsten")

TID 10003B

1.18.27

X-Ray Filters

TID 10003B

1.18.27.1

X-Ray Filter Type

(113653, DCM, "Flat filter")

TID 10003B

1.18.27.2

X-Ray Filter Material

(66925006, SCT, "Copper")

TID 10003B

1.18.27.3

X-Ray Filter Thickness Minimum

0.300000 mm

TID 10003B

1.18.27.4

X-Ray Filter Thickness Maximum

0.300000 mm

TID 10003B

1.18.28

Collimated Field Area

0.010781 m2

TID 10003B

1.18.29

Collimated Field Height

129 mm

TID 10003B

1.18.30

Collimated Field Width

83 mm

TID 10003B

1.18.31

X-Ray Grid

(111641, DCM, "Fixed grid")

TID 10003B

1.18.32

X-Ray Grid

(111642, DCM, "Focused grid")

TID 10003B

Irradiation Event X-Ray Mechanical Data TID 10003C

1.18.33

Positioner Primary Angle

-100.00 deg

TID 10003C

1.18.34

Positioner Secondary Angle

0.0 deg

TID 10003C

1.18.35

Positioner Primary End Angle

100.00 deg

TID 10003C

1.18.36

Positioner Secondary End Angle

0.0 deg

TID 10003C

1.18.37

Table Head Tilt Angle

0.0 deg

TID 10003C

1.18.38

Table Horizontal Rotation Angle

0.0 deg

TID 10003C

1.18.39

Table Cradle Tilt Angle

0.0 deg

TID 10003C

Start CID 10008 “Dose Related Distance Measurement”

1.18.40

Distance Source to Isocenter

720 mm

TID 10003C

1.18.41

Distance Source to Reference Point

570.00 mm

TID 10003C

1.18.42

Distance Source to Detector

1195 mm

TID 10003C

1.18.43

Table Longitudinal Position

727.90 mm

TID 10003C

1.18.44

Table Lateral Position

50.90 mm

TID 10003C

1.18.45

Table Height Position

87.5 mm

TID 10003C

1.18.46

Table X Position to Isocenter

727.90 mm

TID 10003C

1.18.47

Table Y Position to Isocenter

87.5 mm

TID 10003C

1.18.48

Table Z Position to Isocenter

50.90 mm

TID 10003C

End CID 10008 “Dose Related Distance Measurement”

1.18.49

Positioner Isocenter Primary Angle

-100.0 deg

TID 10003C

1.18.50

Positioner Isocenter Secondary Angle

0.0 deg

TID 10003C

1.18.51

Positioner Isocenter Detector Rotation Angle

0.0 deg

TID 10003C

1.18.52

Positioner Isocenter Primary End Angle

100.00 deg

TID 10003C

1.18.53

Positioner Isocenter Secondary End Angle

0.00 deg

TID 10003C

1.18.54

Positioner Isocenter Detector Rotation End Angle

0.0 deg

TID 10003C

1.19

Dose Image

SOP Class UID: 1.2.840.10008.5.1.4.1.1.7

SOP Instance UID: 2.999.6

TID 10001

Start Person Participant within TID 10001: Irradiation Authorizing is the performing physician.

1.20

Person Name

Mann^Hugh^^Dr

TID 1020

1.20.1

Person Role in Procedure

(113850, DCM, "Irradiation Authorizing")

TID 1020

1.20.2

Person ID

5321611

TID 1020

1.20.3

Person ID Issuer

Institution A

TID 1020

1.20.4

Person Role in Organization

(309343006, SCT, "Physician")

TID 1020

End Person Participant

1.21

Source of Dose Information

(113856, DCM, "Automated Data Collection")

TID 10001


UUUU.5.1.2 Example of Traditional RDSR for CT

The following is an example RDSR for a routine CT study. In this example, a CT scanner performs a CT localizer scan and a single CT acquisition.

In this example,

  • The irradiation is authorized by the referring physician who ordered the CT scan, Dr. Marcus Welby.

  • The irradiation is administered by the technologist, John Doe.

  • The dose check alert was configured for CTDIvol of 1 Gy, and the dose notification was configured for CTDIvol of 45 mGy. Neither the alert nor the notification was triggered during this study.

Table UUUU.5.1.2-1 illustrates one possible encoding of this example. Note that other alternative implementation patterns exist.

Table UUUU.5.1.2-1. CT Traditional RDSR

Node

Code Meaning of Concept Name

Code or Example Value

TID

1

X-Ray Radiation Dose Report

TID 10011

1.1

Language of Content Item and Descendants

(en, RFC5646, "English")

TID 1204

1.2

Procedure reported

(77477000, SCT, "Computed Tomography X-Ray")

TID 10011

1.2.1

Has Intent

(261004008, SCT, "Diagnostic Intent")

TID 10011

Start Observer Context

Observer #1: Irradiating Device

1.3

Observer Type

(121007, DCM, "Device")

TID 1002

1.4

Device Observer UID

2.999.1.2.3.4

TID 1004

1.5

Device Observer Name

CT1_HOSPITAL_A

TID 1004

1.6

Device Observer Manufacturer

Manufacturer X

TID 1004

1.7

Device Observer Model Name

Model Y

TID 1004

1.8

Device Observer Serial Number

123456789

TID 1004

1.9

Device Role in Procedure

(113859, DCM, "Irradiating Device")

TID 1004

1.10

Device Role in Procedure

(121097, DCM, "Recording")

TID 1004

End Observer Context

1.11

Start of X-Ray Irradiation

20230725120000.000

TID 10011

1.12

End of X-Ray Irradiation

20230725120300.000

TID 10011

1.13

Scope of Accumulation

(113014, DCM, "Study")

TID 10011

1.13.1

Study Instance UID

2.999.2.3.4.5

TID 10011

1.14

CT Accumulated Dose Data

TID 10012

1.14.1

Total Number of Irradiation Events

2 events

TID 10012

1.14.2

CT Dose Length Product Total

220 mGy.cm

TID 10012

Start first Irradiation Event

1.15

CT Acquisition

TID 10013

1.15.1

Acquisition Protocol

CT Abdomen W contrast IV

TID 10013

1.15.2

Target Region

(818981001, SCT, "Abdomen")

TID 10013

1.15.3

CT Acquisition Type

(113805, DCM, "Constant Angle Acquisition")

TID 10013

1.15.4

Irradiation Event UID

2.999.3.4.5.6

TID 10013

1.15.5

DateTime Started

20230725120000.000

TID 10013

1.15.6

CT Acquisition Parameters

TID 10013

1.15.6.1

Exposure Time

3.00 s

TID 10013

1.15.6.2

Scanning Length

250 mm

TID 10014

1.15.6.3

Top Z Location of Scanning Length

0.00 mm

TID 10014

1.15.6.4

Bottom Z Location of Scanning Length

-250.00 mm

TID 10014

1.15.6.5

Frame of Reference UID

2.999.4.5.6.7

TID 10014

1.15.6.6

Nominal Single Collimation Width

0.625 mm

TID 10013

1.15.6.7

Nominal Total Collimation Width

5.0 mm

TID 10013

1.15.6.8

Number of X-Ray Sources

1 X-Ray sources

TID 10013

1.15.6.9

CT X-Ray Source Parameters

TID 10013

1.15.6.9.1

Identification of the X-Ray Source

1

TID 10013

1.15.6.9.2

KVP

120.0 kV

TID 10013

1.15.6.9.3

Maximum X-Ray Tube Current

40 mA

TID 10013

1.15.6.9.4

X-Ray Tube Current

40 mA

TID 10013

1.15.7

Comment

Localizer

TID 10013

Start Person Participant within TID 10013

1.15.8

Person Name

Doe^John

TID 1020

1.15.8.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.15.8.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

End first Irradiation Event

Start second Irradiation Event

1.16

CT Acquisition

TID 10013

1.16.1

Acquisition Protocol

CT Abdomen W contrast IV

TID 10013

1.16.2

Target Region

(818981001, SCT, "Abdomen")

TID 10013

1.16.3

CT Acquisition Type

(116152004, SCT, "Spiral Acquisition")

TID 10013

1.16.4

Irradiation Event UID

2.999.5.6.7.8

TID 10013

1.16.5

DateTime Started

20230725120258.000

TID 10013

1.16.6

CT Acquisition Parameters

TID 10013

1.16.6.1

Exposure Time

2.00 s

TID 10013

Start Scanning Length TID 10014

1.16.6.2

Scanning Length

220 mm

TID 10014

1.16.6.3

Length of Reconstructable Volume

200 mm

TID 10014

1.16.6.4

Exposed Range

260 mm

TID 10014

1.16.6.5

Top Z Location of Reconstructable Volume

-25.00 mm

TID 10014

1.16.6.6

Bottom Z Location of Reconstructable Volume

-225.00 mm

TID 10014

1.16.6.7

Top Z Location of Scanning Length

-15.00 mm

TID 10014

1.16.6.8

Bottom Z Location of Scanning Length

-235.00 mm

TID 10014

1.16.6.9

Frame of Reference UID

2.999.6.7.8.9

TID 10014

End Scanning Length

1.16.6.10

Nominal Single Collimation Width

0.625 mm

TID 10013

1.16.6.11

Nominal Total Collimation Width

40.0 mm

TID 10013

1.16.6.12

Pitch Factor

1.375

TID 10013

1.16.6.13

Number of X-Ray Sources

1 X-Ray sources

TID 10013

1.16.6.14

CT X-Ray Source Parameters

TID 10013

1.16.6.14.1

Identification of the X-Ray Source

1

TID 10013

1.16.6.14.2

KVP

120.0 kV

TID 10013

1.16.6.14.3

Maximum X-Ray Tube Current

500 mA

TID 10013

1.16.6.14.4

X-Ray Tube Current

394 mA

TID 10013

1.16.6.14.5

Exposure Time per Rotation

0.5 s

TID 10013

1.16.7

CT Dose

TID 10013

1.16.7.1

Mean CTDIvol

10.00 mGy

TID 10013

1.16.7.2

CTDIw Phantom Type

(113691, DCM, "IEC Body Dosimetry Phantom")

TID 10013

1.16.7.3

CTDIfreeair Calculation Factor

0.25 mGy/mA.s

TID 10013

1.16.7.4

Mean CTDIfreeair

49.25 mGy

TID 10013

1.16.7.5

DLP

220.00 mGy.cm

TID 10013

1.16.7.6

Size Specific Dose Estimate

12.30 mGy

TID 10013

1.16.7.6.1

Measurement Method

(113988, DCM, "Estimated from Water Equivalent Diameter")

TID 10013

1.16.7.6.2

Water Equivalent Diameter

300 mm

TID 10013

1.16.7.6.2.1

Measurement Method

(113984, DCM, "Water Equivalent Diameter From Localizer")

TID 10013

Start CT Dose Check Details TID 10015

1.16.7.4

Dose Check Alert Details

TID 10015

1.16.7.4.1

DLP Alert Value Configured

(373067005, SCT, "No")

TID 10015

1.16.7.4.2

CTDIvol Alert Value Configured

(373066001, SCT, "Yes")

TID 10015

1.16.7.4.3

CTDIvol Alert Value

1000.0 mGy

TID 10015

1.16.7.5

Dose Check Notification Details

TID 10015

1.16.7.5.1

DLP Notification Value Configured

(373067005, SCT, "No")

TID 10015

1.16.7.5.2

CTDIvol Notification Value Configured

(373066001, SCT, "Yes")

TID 10015

1.16.7.5.3

CTDIvol Notification Value

45.00 mGy

TID 10015

End CT Dose Check Details

Start Person Participant within TID 10013

1.16.8

Person Name

Doe^John

TID 1020

1.16.8.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.16.8.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

End Second Irradiation Event

1.17

Source of Dose Information

(113856, DCM, "Automated Data Collection")

TID 10011

Start Person Participant within TID 10011

1.18

Person Name

Welby^Marcus^^Dr

TID 1020

1.18.1

Person Role in Procedure

(113850, DCM, "Irradiation Authorizing")

TID 1020

1.18.2

Person Role in Organization

(309343006, SCT, "Physician")

TID 1020

End Person Participant


UUUU.5.1.2.1 Example of Traditional RDSR for CT with Notification Event

The following is an example RDSR for a routine CT study where the CTDIvol dose notification value is triggered (see [NEMA XR-25]). This example is largely identical to the example in Section UUUU.5.1.2 except for exceeding the notification value due to a change in scanning parameters.

In this example:

  • The dose check alert was configured for CTDIvol of 1 Gy, and the dose check notification was configured for CTDIvol of 20 mGy (lower than the previous example).

  • The X-Ray Tube Current and Exposure Time per Rotation are both increased from the previous example, resulting in a proportional increase in Mean CTDIvol, CTDIfreeair, DLP, and Size Specific Dose Estimate. As a result of the increase in reported dose metrics and the decrease in CTDIvol Notification Value, a notification event is triggered.

  • Exceeding a notification value allows the encoding of a Reason for Proceeding and a Person Participant for the RDSR encoding. This example includes a reason (patient with high body mass index (BMI) ) and the technologist initials (from the scanner console log in). Note that following the [NEMA XR-25] standard describing CT Dose Check capabilities, a reason need not necessarily be entered by the operator to proceed with a scan, meaning this row may be blank or absent in some instances.

  • The remaining content of the RDSR is identical for all other nodes to the example in Section UUUU.5.1.2.

Table UUUU.5.1.2.1-1 illustrates one possible encoding of this example. Note that other alternative implementation patterns exist.

Table UUUU.5.1.2.1-1. CT Traditional RDSR with Notification Event

Node

Code Meaning of Concept Name

Code or Example Value

TID

1

X-Ray Radiation Dose Report

TID 10011

<Identical Content to Nodes 1.1-1.16.6.13 from Table UUUU.5.1.2-1>

1.16.6.14

CT X-Ray Source Parameters

TID 10013

1.16.6.14.1

Identification of the X-Ray Source

1

TID 10013

1.16.6.14.2

KVP

120.0 kV

TID 10013

1.16.6.14.3

Maximum X-Ray Tube Current

500 mA

TID 10013

1.16.6.14.4

X-Ray Tube Current

400 mA

TID 10013

1.16.6.14.5

Exposure Time per Rotation

1.0 s

TID 10013

1.16.7

CT Dose

TID 10013

1.16.7.1

Mean CTDIvol

20.30 mGy

TID 10013

1.16.7.2

CTDIw Phantom Type

(113691, DCM, "IEC Body Dosimetry Phantom")

TID 10013

1.16.7.3

CTDIfreeair Calculation Factor

0.25 mGy/mA.s

TID 10013

1.16.7.4

Mean CTDIfreeair

100.0 mGy

TID 10013

1.16.7.5

DLP

446.70 mGy.cm

TID 10013

1.16.7.6

Size Specific Dose Estimate

24.97 mGy

TID 10013

1.16.7.6.1

Measurement Method

(113988, DCM, "Estimated from Water Equivalent Diameter")

TID 10013

1.16.7.6.2

Water Equivalent Diameter

300 mm

TID 10013

1.16.7.6.2.1

Measurement Method

(113984, DCM, "Water Equivalent Diameter From Localizer")

TID 10013

Start CT Dose Check Details TID 10015

1.16.7.4

Dose Check Alert Details

TID 10015

1.16.7.4.1

DLP Alert Value Configured

(373067005, SCT, "No")

TID 10015

1.16.7.4.2

CTDIvol Alert Value Configured

(373066001, SCT, "Yes")

TID 10015

1.16.7.4.3

CTDIvol Alert Value

1000.0 mGy

TID 10015

1.16.7.5

Dose Check Notification Details

TID 10015

1.16.7.5.1

DLP Notification Value Configured

(373067005, SCT, "No")

TID 10015

1.16.7.5.2

CTDIvol Notification Value Configured

(373066001, SCT, "Yes")

TID 10015

1.16.7.5.3

CTDIvol Notification Value

20.00 mGy

TID 10015

1.16.7.5.4

CTDIvol Forward Estimate

20.30 mGy

TID 10015

1.16.7.5.5

Reason for Proceeding

High BMI patient

TID 10015

Start Person Participant within TID 10015

1.18

Person Name

NBB

TID 1020

1.18.1

Person Role in Procedure

(113850, DCM, "Irradiation Authorizing")

TID 1020

End Person Participant

End CT Dose Check Details

<Identical Content to Nodes 1.16.8-1.18.2 from Table UUUU.5.1.2-1>


UUUU.5.1.3 Example of Traditional RDSR for DX

The following is an example RDSR for a routine radiography study. In this example, an X-Ray system performs the acquisitions with the following characteristics:

  • This is an integrated system, therefore the conditions of TIDs (i.e., X-Ray Detector Data Available, X-Ray Source Data Available, X-Ray Mechanical Data Available) are satisfied.

  • Two acquisitions are performed: one lateral and one Anterior-Posterior (AP) views of the lumbar spine, with the patient standing.

  • The dose is estimated at the detector plane (dose reference point definition).

  • The irradiation is administered by the operator John Doe.

  • The irradiation is authorized by the physician Dr. Hugh Mann.

Table UUUU.5.1.3-1 illustrates one possible encoding of this example. Note that other alternative implementation patterns exist.

Table UUUU.5.1.3-1. DX Traditional RDSR of a Fully Integrated System

Node

Code Meaning of Concept Name

Code or Example Value

TID

1

X-Ray Radiation Dose Report

Section TID 10001

1.1

Language of Content Item and Descendants

(en, RFC5646, "English")

TID 1204

1.2

Procedure reported

(113704, DCM, "Projection X-Ray")

TID 10001

1.2.1

Has Intent

(261004008, SCT, "Diagnostic Intent")

TID 10001

1.3

Acquisition Device Type

(113958, DCM, "Integrated Projection Radiography System")

TID 10001

Start Observer Context

Observer #1: Irradiating device

1.4

Observer Type

(121007, DCM, "Device")

TID 1002

1.5

Device Observer UID

2.999.1

TID 1004

1.6

Device Observer Name

MyStationName

TID 1004

1.7

Device Observer Manufacturer

Manufacturer X

TID 1004

1.8

Device Observer Model Name

Model Y

TID 1004

1.9

Device Observer Serial Number

SerialNumber123

TID 1004

1.10

Device Role in Procedure

(113859, DCM, "Irradiating Device")

TID 1004

End Observer Context

1.11

Scope of Accumulation

(113014, DCM, "Study")

TID 10001

1.11.1

Study Instance UID

2.999.2

TID 10001

1.12

X-Ray Detector Data Available

(373066001, SCT, "Yes")

TID 10001

1.13

X-Ray Source Data Available

(373066001, SCT, "Yes")

TID 10001

1.14

X-Ray Mechanical Data Available

(373066001, SCT, "Yes")

TID 10001

1.15

Accumulated X-Ray Dose Data

TID 10002

1.15.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10002

1.15.2

Dose Area Product Total

0.000178600000 Gy.m2

TID 10007

1.15.3

Dose (RP) Total

0.00211933 Gy

TID 10007

1.15.4

Distance Source to Reference Point

1150 mm

TID 10007

1.15.5

Total Number of Radiographic Frames

2 no units

TID 10007

1.15.6

Reference Point Definition

(113941, DCM, "In Detector Plane")

TID 10007

Start Irradiation Event X-Ray Data of LATERAL acquisition

1.16

Irradiation Event X-Ray Data

TID 10003

1.16.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10003

1.16.2

Irradiation Event UID

2.999.3

TID 10003

1.16.3

DateTime Started

20241016213726.000

TID 10003

1.16.4

Irradiation Event Type

(113611, DCM, "Stationary Acquisition")

TID 10003

1.16.5

Acquisition Protocol

Wallstand L-Spine Lateral

TID 10003

1.16.6

Image View

(399198007, SCT, "right lateral")

TID 10003

1.16.7

Patient Orientation

(C86043, NCIt, "erect")

TID 10003

1.16.7.1

Patient Orientation Modifier

(10904000, SCT, "standing")

TID 10003

1.16.8

Target Region

(122496007, SCT, "Lumbar spine")

TID 10003

1.16.9

Dose Area Product

0.0001156000 Gy.m2

TID 10003

Start Person Participant within TID 10003: Irradiation Administering is the operator.

1.16.10

Person Name

Doe^John

TID 1020

1.16.10.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.16.10.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

1.16.11

Exposure Index

227.00 no units

TID 10003A

1.16.12

Target Exposure Index

250.00 no units

TID 10003A

1.16.13

Deviation Index

-0.41914 no units

TID 10003A

1.16.14

Acquired Image

2.999.4

TID 10003A

1.16.15

Dose (RP)

0.00123015 Gy

TID 10003B

1.16.16

Reference Point Definition

(113941, DCM, "In Detector Plane")

TID 10003B

1.16.17

Number of Pulses

1 no units

TID 10003B

1.16.18

KVP

89.800000 kV

TID 10003B

1.16.19

X-Ray Tube Current

724.00 mA

TID 10003B

1.16.20

Exposure Time

32.00 ms

TID 10003B

1.16.21

Exposure

23160.00 uA.s

TID 10003B

1.16.22

Focal Spot Size

1.6 mm

TID 10003B

1.16.23

X-Ray Filters

TID 10003B

1.16.23.1

X-Ray Filter Type

(111609, DCM, "No filter")

TID 10003B

1.16.24

Collimated Field Area

0.0886542 m2

TID 10003B

1.16.25

Collimated Field Height

425.2 mm

TID 10003B

1.16.26

Collimated Field Width

208.5 mm

TID 10003B

1.16.27

X-Ray Grid

(111642, DCM, "Focused grid")

TID 10003B

1.16.28

Distance Source to Detector

1150.00 mm

TID 10003C

Start Irradiation Event X-Ray Data of ANTERIOR-POSTERIOR acquisition

1.17

Irradiation Event X-Ray Data

TID 10003

1.17.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10003

1.17.2

Irradiation Event UID

2.999.5

TID 10003

1.17.3

DateTime Started

20241016214647.000

TID 10003

1.17.4

Irradiation Event Type

(113611, DCM, "Stationary Acquisition")

TID 10003

1.17.5

Acquisition Protocol

Wallstand L-Spine AP

TID 10003

1.17.6

Image View

(399348003, SCT, "antero-posterior")

TID 10003

1.17.7

Patient Orientation

(C86043, NCIt, "erect")

TID 10003

1.17.7.1

Patient Orientation Modifier

(10904000, SCT, "standing")

TID 10003

1.17.8

Target Region

(122496007, SCT, "Lumbar spine")

TID 10003

1.17.9

Dose Area Product

0.000063000 Gy.m2

TID 10003

Start Person Participant within TID 10003: Irradiation Administering is the operator.

1.16.10

Person Name

Doe^John

TID 1020

1.16.10.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.16.10.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

1.17.11

Exposure Index

243.00 no units

TID 10003A

1.17.12

Target Exposure Index

250.00 no units

TID 10003A

1.17.13

Deviation Index

-0.123 no units

TID 10003A

1.17.14

Acquired Image

2.999.6

TID 10003A

1.17.15

Dose (RP)

0.00088918 Gy

TID 10003B

1.17.16

Reference Point Definition

(113941, DCM, "In Detector Plane")

TID 10003B

1.17.17

Number of Pulses

1 no units

TID 10003B

1.17.18

KVP

80.90 kV

TID 10003B

1.17.19

X-Ray Tube Current

780.00 mA

TID 10003B

1.17.20

Exposure Time

26.00 ms

TID 10003B

1.17.21

Exposure

20360 uA.s

TID 10003B

1.17.22

Focal Spot Size

1.6 mm

TID 10003B

1.17.23

X-Ray Filters

TID 10003B

1.17.23.1

X-Ray Filter Type

(111609, DCM, "No filter")

TID 10003B

1.17.24

Collimated Field Area

0.067497 m2

TID 10003B

1.17.25

Collimated Field Height

372.500 mm

TID 10003B

1.17.26

Collimated Field Width

181.200 mm

TID 10003B

1.17.27

X-Ray Grid

(111642, DCM, "Focused grid")

TID 10003B

1.17.28

Distance Source to Detector

1150.00 mm

TID 10003C

End Irradiation Event X-Ray Data

Start Person Participant within TID 10001: Irradiation Authorizing is the physician.

1.18

Person Name

Mann^Hugh^^Dr

TID 1020

1.18.1

Person Role in Procedure

(113850, DCM, "Irradiation Authorizing")

TID 1020

1.18.2

Person ID

5321611

TID 1020

1.18.3

Person ID Issuer

Institution A

TID 1020

1.18.4

Person Role in Organization

(309343006, SCT, "Physician")

TID 1020

End Person Participant

1.19

Source of Dose Information

(113856, DCM, "Automated Data Collection")

TID 10001


UUUU.5.1.3.1 Example of Traditional RDSR for DX Non-Integrated Equipment

The following is an example of the previous routine radiography study, but performed on a less integrated equipment. In this example, a system performs the acquisitions with the following characteristics:

  • The system is a radiography system with mobile digital detector, which has Exposure Index capability but there is no X-Ray generator integration. Only X-Ray Detector Data is available. Other data such as X-Ray Source Data and X-Ray Mechanical Data are not available.

Table UUUU.5.1.3-2 includes values that are different vs. the fully integrated example.

Table UUUU.5.1.3-2. DX Traditional RDSR with only X-Ray Detector Data Available

Node

Code Meaning of Concept Name

Code or Example Value

TID

X-Ray Radiation Dose Report

Section TID 10001

…

TID 10001

Procedure reported

(113704, DCM, "Projection X-Ray")

TID 10001

Acquisition Device Type

(113959, DCM, "Cassette-based Projection Radiography System")

TID 10001

Start Observer Context

Observer #1: Reading Device

Observer Type

(121007, DCM, "Device")

TID 1002

…

TID 1004

Device Role in Procedure

(113942, DCM, "X-Ray Reading Device")

TID 1004

End Observer Context

…

TID 10001

X-Ray Detector Data Available

(373066001, SCT, "Yes")

TID 10001

X-Ray Source Data Available

(373067005, SCT, "No")

TID 10001

X-Ray Mechanical Data Available

(373067005, SCT, "No")

TID 10001

Include TID 10002 “Accumulated X-Ray Dose”

…

TID 10002

Include TID 10006 “Accumulated Cassette-based Projection Radiography Dose”

Detector Type

(113948, DCM, "Direct Detector")

TID 10006

Total Number of Radiographic Frames

2

TID 10006

…

TID 10006

Include TID 10003 “Irradiation Event X-Ray Data” for irradiation #1

…

TID 10003

Include TID 10003A “Irradiation Event X-Ray Detector Data” for irradiation #1

(TID 10003B and TID 10003C are not allowed)

Exposure Index

227.00 no units

TID 10003A

Target Exposure Index

250.00 no units

TID 10003A

Deviation Index

-0.41914 no units

TID 10003A

…

TID 10003A

Include TID 10003 “Irradiation Event X-Ray Data” for irradiation #2

…

TID 10003

Include TID 10003A “Irradiation Event X-Ray Detector Data” for irradiation #2

(TID 10003B and TID 10003C are not allowed)

Exposure Index

243.00 no units

TID 10003A

Target Exposure Index

250.00 no units

TID 10003A

Deviation Index

-0.123 no units

TID 10003A

…

TID 10003A


The following is an example of the previous routine radiography study, but performed on a less integrated equipment. In this example, a system performs the acquisitions with the following characteristics:

  • The system is a mobile X-Ray source which has generator integration. Only X-Ray Source Data is available. Other data such as X-Ray Detector Data and X-Ray Mechanical Data are not available.

Table UUUU.5.1.3-3 includes values that are different vs. the fully integrated example.

Table UUUU.5.1.3-3. DX Traditional RDSR with only X-Ray Source Data Available

Node

Code Meaning of Concept Name

Code or Example Value

TID

X-Ray Radiation Dose Report

Section TID 10001

…

TID 10001

Procedure reported

(113704, DCM, "Projection X-Ray")

TID 10001

Acquisition Device Type

(113959, DCM, "Cassette-based Projection Radiography System")

TID 10001

Start Observer Context

Observer #1: Irradiating Device

Observer Type

(121007, DCM, "Device")

TID 1002

…

TID 1004

Device Role in Procedure

(113859, DCM, "Irradiating Device")

TID 1004

End Observer Context

X-Ray Detector Data Available

(373067005, SCT, "No")

TID 10001

X-Ray Source Data Available

(373066001, SCT, "Yes")

TID 10001

X-Ray Mechanical Data Available

(373067005, SCT, "No")

TID 10001

Include TID 10002 “Accumulated X-Ray Dose”

…

TID 10002

Include TID 10006 “Accumulated Cassette-based Projection Radiography Dose”

Dose Area Product Total

0.000178600000 Gy.m2

TID 10006

Include TID 10003 “Irradiation Event X-Ray Data” for irradiation #1

Dose Area Product

0.0001156000 Gy.m2

TID 10003

…

TID 10003

Include TID 10003B “Irradiation Event X-Ray Source Data” for irradiation #1

(TID 10003A and TID 10003C are not allowed)

Dose (RP)

0.00123015 Gy

TID 10003B

Reference Point Definition

(113965, DCM, "100cm from X-Ray Source")

TID 10003B

Number of Pulses

1 no units

TID 10003B

…

TID 10003B

Include TID 10003 “Irradiation Event X-Ray Data” for irradiation #2

Dose Area Product

0.000063000 Gy.m2

TID 10003

…

TID 10003

Include TID 10003B “Irradiation Event X-Ray Source Data” for irradiation #2

(TID 10003A and TID 10003C are not allowed)

Dose (RP)

0.00088918 Gy

TID 10003B

Reference Point Definition

(113965, DCM, "100cm from X-Ray Source")

TID 10003B

Number of Pulses

1 no units

TID 10003B

…

TID 10003B


The following is an example of the previous routine radiography study, but performed on a less integrated equipment. In this example, a system performs the acquisitions with the following characteristics:

  • The system has no detector and no generator integration. Only X-Ray Mechanical Data is Available. Other data such as X-Ray Detector Data and X-Ray Source Data are not available.

  • The Dose Area Product (DAP) is measured with a DAP meter.

Table UUUU.5.1.3-4 includes values that are different vs. the fully integrated example.

Table UUUU.5.1.3-4. DX Traditional RDSR with only X-Ray Mechanical Data Available

Node

Code Meaning of Concept Name

Code or Example Value

TID

X-Ray Radiation Dose Report

Section TID 10001

…

Procedure reported

(113704, DCM, "Projection X-Ray")

TID 10001

Acquisition Device Type

(113959, DCM, "Cassette-based Projection Radiography System")

TID 10001

Start Observer Context

Observer #1: Recording Device

Observer Type

(121007, DCM, "Device")

TID 1002

…

Device Role in Procedure

(121097, DCM, "Recording")

TID 1004

End Observer Context

X-Ray Detector Data Available

(373067005, SCT, "No")

TID 10001

X-Ray Source Data Available

(373067005, SCT, "No")

TID 10001

X-Ray Mechanical Data Available

(373066001, SCT, "Yes")

TID 10001

Include TID 10002 “Accumulated X-Ray Dose”

…

TID 10002

Include TID 10006 “Accumulated Cassette-based Projection Radiography Dose”

Dose Area Product Total

0.000178600000 Gy.m2

TID 10006

Include TID 10003 “Irradiation Event X-Ray Data” for irradiation #1

Dose Area Product

0.0001156000 Gy.m2

TID 10003

…

TID 10003

Include TID 10003C “Irradiation Event X-Ray Mechanical Data” for irradiation #1

(TID 10003A and TID 10003B are not allowed)

Distance Source to Detector

1150.00 mm

TID 10003C

Include TID 10003 “Irradiation Event X-Ray Data” for irradiation #2

Dose Area Product

0.000063000 Gy.m2

TID 10003

…

TID 10003

Include TID 10003C “Irradiation Event X-Ray Mechanical Data” for irradiation #2

(TID 10003A and TID 10003B are not allowed)

Distance Source to Detector

1150.00 mm

TID 10003C


UUUU.5.1.4 Example of Traditional RDSR for MG

The following is an example RDSR for a routine mammography imaging procedure. In this example, a typical four-view screening exam is performed with the following characteristics:

  • This is an integrated system: X-Ray Detector / Source / Mechanical data is available.

  • One focused anti-scatter grid is fixed on the detector housing.

  • Four acquisitions are performed with CC and MLO views for both the left and right breasts.

  • The irradiation is administered by the operator John Doe.

  • The irradiation is authorized by the physician Dr. Hugh Mann.

Table UUUU.5.1.4-1 illustrates one possible encoding of this example. Note that other alternative implementation patterns exist.

Table UUUU.5.1.4-1. MG Traditional RDSR

Node

Code Meaning of Concept Name

Code or Example Value

TID

1

X-Ray Radiation Dose Report

TID 10001

1.1

Language of Content Item and Descendants

(en, RFC5646, "English")

TID 1204

1.2

Procedure reported

(71651007, SCT, "Mammography")

TID 10001

1.2.1

Has Intent

(360156006, SCT, "Screening Intent")

TID 10001

Start Observer Context

Observer #1: Irradiating device

1.3

Observer Type

(121007, DCM, "Device")

TID 1002

1.4

Device Observer UID

2.999.1

TID 1004

1.5

Device Observer Name

MyStationName

TID 1004

1.6

Device Observer Manufacturer

Manufacturer X

TID 1004

1.7

Device Observer Model Name

Model Y

TID 1004

1.8

Device Observer Serial Number

SerialNumber123

TID 1004

1.9

Device Role in Procedure

(113859, DCM, "Irradiating Device")

TID 1004

End Observer Context

1.10

Scope of Accumulation

(113014, DCM, "Study")

TID 10001

1.11

Study Instance UID

2.999.2

TID 10001

1.12

X-Ray Detector Data Available

(373066001, SCT, "Yes")

TID 10001

1.13

X-Ray Source Data Available

(373066001, SCT, "Yes")

TID 10001

1.14

X-Ray Mechanical Data Available

(373066001, SCT, "Yes")

TID 10001

1.15

Accumulated X-Ray Dose Data

TID 10002

1.15.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10002

1.15.2

Accumulated Average Glandular Dose

2.50 mGy

TID 10005

1.15.2.1

Laterality

(80248007, SCT, "Left Breast")

TID 10005

1.15.3

Accumulated Average Glandular Dose

2.80 mGy

TID 10005

1.15.3.1

Laterality

(73056007, SCT, "Right Breast")

TID 10005

1.16

Irradiation Event X-Ray Data

TID 10003

1.16.1

Acquisition Plane

(113622, DCM, "Single Plane")

TID 10003

1.16.2

Irradiation Event UID

2.999.3

TID 10003

1.16.3

DateTime Started

20240418123000.000

TID 10003

1.16.4

Irradiation Event Type

(113611, DCM, "Stationary Acquisition")

TID 10003

1.16.5

Image View

(SCT, 399162004, "cranio-caudal")

TID 10003

1.16.6

Target Region

(76752008, SCT, "Breast")

TID 10003

1.16.6.1

Laterality

(7771000, SCT, "Left")

TID 10003

1.16.7

Half Value Layer

0.50 mm

TID 10003

1.16.8

Patient Equivalent Thickness

44.0 mm

TID 10003

1.16.9

Entrance Exposure at RP

4.10 mGy

TID 10003

1.16.10

Reference Point Definition

(113865, DCM, "4.2cm above Breast Support Surface")

TID 10003

Start Person Participant within TID 10003: Irradiation Administering is the operator.

1.16.11

Person Name

Doe^John

TID 1020

1.16.11.1

Person Role in Procedure

(113851, DCM, "Irradiation Administering")

TID 1020

1.16.11.2

Person Role in Organization

(159016003, SCT, "Radiologic Technologist")

TID 1020

End Person Participant

Irradiation Event X-Ray Detector Data TID 10003A

1.16.12

Acquired Image

SOP Class UID: 1.2.840.10008.5.1.4.1.1.1.2

SOP Instance UID: 2.999.6

TID 10003A

Irradiation Event X-Ray Source Data TID 10003B

1.16.13

Average Glandular Dose

1.20 mGy

TID 10003B

1.16.14

KVP

28.00 kV

TID 10003B

1.16.15

X-Ray Tube Current

160.00 mA

TID 10003B

1.16.16

Exposure Time

625.00 ms

TID 10003B

1.16.17

Exposure

100000 uA.s

TID 10003B

1.16.18

Focal Spot Size

0.3 mm

TID 10003B

1.16.19

Anode Target Material

(26194003, SCT, "Tungsten")

TID 10003B

1.16.20

X-Ray Filters

TID 10003B

1.16.20.1

X-Ray Filter Type

(113653, DCM, "Flat filter")

TID 10003B

1.16.20.2

X-Ray Filter Material

(59801003, SCT, "Rhodium")

TID 10003B

1.16.20.3

X-Ray Filter Thickness Minimum

0.050 mm

TID 10003B

1.16.20.4

X-Ray Filter Thickness Maximum

0.050 mm

TID 10003B

1.16.21

Collimated Field Area

0.0696 m2

TID 10003B

1.16.22

Collimated Field Height

240 mm

TID 10003B

1.16.23

Collimated Field Width

290 mm

TID 10003B

1.16.24

X-Ray Grid

(111642, DCM, "Focused grid")

TID 10003B

1.16.25

X-Ray Grid

(111642, DCM, "Reciprocating grid")

TID 10003B

Irradiation Event X-Ray Mechanical Data TID 10003C

1.16.26

Positioner Primary Angle

0.00 deg

TID 10003C

1.16.27

Compression Thickness

44.00 mm

TID 10003C

1.16.28

Compression Force

70 Newton

TID 10003C

Start CID 10008 “Dose Related Distance Measurement”

1.16.29

Distance Source to Reference Point

633.00 mm

TID 10003C

1.16.30

Distance Source to Detector

700.00 mm

TID 10003C

End CID 10008 “Dose Related Distance Measurement”

1.17

Irradiation Event X-Ray Data

TID 10003

Irradiation Event X-Ray Data TID 10003 for LMLO Acquisition not shown for brevity

1.18

Irradiation Event X-Ray Data

TID 10003

Irradiation Event X-Ray Data TID 10003 for RCC Acquisition not shown for brevity

1.19

Irradiation Event X-Ray Data

TID 10003

Irradiation Event X-Ray Data TID 10003 for RMLO Acquisition not shown for brevity

Start Person Participant within TID 10001: Irradiation Authorizing is the physician.

1.20

Person Name

Mann^Hugh^^Dr

TID 1020

1.20.1

Person Role in Procedure

(113850, DCM, "Irradiation Authorizing")

TID 1020

1.20.2

Person ID

5321611

TID 1020

1.20.3

Person ID Issuer

Institution A

TID 1020

1.20.4

Person Role in Organization

(309343006, SCT, "Physician")

TID 1020

End Person Participant

1.21

Source of Dose Information

(113856, DCM, "Automated Data Collection")

TID 10001


DICOM PS3.17 2026d - Explanatory Information