UUUU.4.9.2 Irradiation Event Dose Information Based on System Capability
Three flags in TID 10001 “Projection X-Ray Radiation Dose” indicate which technical data a system is capable of providing for each Irradiation Event:
Table UUUU.4.9.2-1. Required TIDs by System Capability Flags
|
TID 10001
|
Required TIDs
|
|
(113945, DCM, "X-Ray Detector Data Available") = (373066001, SCT, "Yes")
|
TID 10003A “Irradiation Event X-Ray Detector Data”
|
|
(113943, DCM, "X-Ray Source Data Available") = (373066001, SCT, "Yes")
|
TID 10003B “Irradiation Event X-Ray Source Data”
|
|
(113944, DCM, "X-Ray Mechanical Data Available") = (373066001, SCT, "Yes")
|
TID 10003C “Irradiation Event X-Ray Mechanical Data”
|
Table UUUU.4.9.2-2 shows examples of when the flags should be set to (373066001, SCT, "Yes") or (373067005, SCT, "No"), and recommended value for (113876, DCM, "Device Role in Procedure") in Procedure in TID 1004 “Device Observer Identifying Attributes” (refer to CID 7445 “Device Participating Role”):
Table UUUU.4.9.2-2. Examples of System Capability Flags
|
System Capability
|
X-Ray Detector Data Available
|
X-Ray Source Data Available
|
X-Ray Mechanical Data Available
|
Device Role in Procedure
|
|
Exposure index capability (detector) but no generator integration.
|
(373066001, SCT, "Yes")
|
(373067005, SCT, "No")
|
any
|
(113942, DCM, "X-Ray Reading Device")
|
|
Generator integration regardless other capabilities
|
any
|
(373066001, SCT, "Yes")
|
any
|
(113859, DCM, "Irradiating Device")
|
|
No detector nor generator integration
|
(373067005, SCT, "No")
|
(373067005, SCT, "No")
|
any
|
(121097, DCM, "Recording")
|