UUUU.4.2.2 Enhanced RDSR

TID 10040 requires the Content Item (121058, DCM, "Procedure reported") to be present, with possible values defined in the CID 10005 “X-Ray Radiation Dose Procedure Type Reported”. This provides an indication of the modality and type of equipment used, similarly to the information in Traditional RDSR. Note that in Enhanced RDSR there's no condition on other Content Items based on the value of the Procedure Reported.

CID 10005 “X-Ray Radiation Dose Procedure Type Reported” specifies a limited set of general procedures for the purpose of sorting/processing RDSRs based on their radiation dose reporting characteristics. This list is extensible and extension codes should follow a recognized standard for coded terminologies. There are other Context Groups used in PS3.16 for "Procedure Reported" on other SR Templates, which include more detailed clinical procedures.

Enhanced RDSR does not include (122142, DCM, "Acquisition Device Type") because CID 10005 is used in (121058, DCM, "Procedure reported") and contains equivalent granularity.

Table UUUU.4.2.2-1 provides the recommended values of the Content Item (121058, DCM, "Procedure reported") for the Enhanced RDSR.

Table UUUU.4.2.2-1. Recommended Values for Procedure Reported