DICOM PS3.17 2021e - Explanatory Information

II.3 Device Module

Table II-3. Device Module Attribute Mapping

Device Module Attribute Name

Tag

Product Characteristics Module Source

Device Sequence

(0050,0010)

--

>Include Table 8.8-1 “Code Sequence Macro Attributes” in PS3.3

Product Type Code Sequence (0044,0007) > 'Code Sequence Macro'

>Device Length

(0050,0014)

Product Parameter Sequence (0044,0013) > Numeric Value (0040,A30A), where:

  • Product Parameter Sequence > Concept Name Code Sequence (0040,A043) value is (410668003, SCT, "Length")

  • Product Parameter Sequence > Measurement Units Code Sequence (0040,08EA) is (mm, UCUM, "mm")

>Device Diameter

(0050,0016)

Product Parameter Sequence (0044,0013) > Numeric Value (0040,A30A), where:

>Device Diameter Units

(0050,0017)

Product Parameter Sequence (0044,0013) > Measurement Units Code Sequence (0040,08EA) > Code Meaning (0008,0104), where:

Note

Device Diameter Units is a CS VR (16 characters max, upper case), so a conversion from the LO VR is required.

>Device Volume

(0050,0018)

Product Parameter Sequence (0044,0013) > Numeric Value (0040,A30A), where:

  • Product Parameter Sequence > Concept Name Code Sequence (0040,A043) value is (118565006, SCT, "Volume")

  • Product Parameter Sequence > Measurement Units Code Sequence (0040,08EA) is (ml, UCUM, "ml")

>Inter-Marker Distance

(0050,0019)

Product Parameter Sequence (0044,0013) > Numeric Value (0040,A30A), where:

  • Product Parameter Sequence > Concept Name Code Sequence (0040,A043) value is (121208, DCM, "Inter-Marker Distance")

  • Product Parameter Sequence > Measurement Units Code Sequence (0040,08EA) is (mm, UCUM, "mm")

>Device Description

(0050,0020)

Product Name (0044,0008) and/or Product Description (0044,0009)


DICOM PS3.17 2021e - Explanatory Information