DICOM PS3.17 2024b - Explanatory Information

AAAA.2.3 Acme

The first part of this example is shown above in Table AAAA.2-1.

Note

  1. The author of this protocol chose to use the code for the vertex of the head rather than the skull as the basis for the plane defining the extent of the scan and reconstructions.

  2. The Requested Series Description (0018,9937) is the same for both localizer acquisitions, however DICOM does not mandate series organization behavior so this does not guarantee that both localizers will be placed in the same series.

Table AAAA.2-3. AAPM Routine Brain Details - Acme

Attribute

Tag

Value

Model Specification Sequence

(0018,9912)

>Manufacturer

(0008,0070)

ACME

>Manufacturer's Model Name

(0008,1090)

Alpha

>Software Versions

(0018,1020)

V1.63\1.70

>Manufacturer

(0008,0070)

ACME

>Manufacturer's Model Name

(0008,1090)

Alpha Plus

>Software Versions

(0018,1020)

V1.63\1.70

Patient Specification Sequence

(0018,9911)

>See Table AAAA.2-3a “Patient Specification”

Acquisition Protocol Element Specification Sequence

(0018,991F)

>Protocol Element Number

(0018,9921)

1

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3b “First Acquisition Protocol Element Specification”

>Protocol Element Number

(0018,9921)

2

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3c “Second Acquisition Protocol Element Specification”

>Protocol Element Number

(0018,9921)

3

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3d “Third Acquisition Protocol Element Specification”

Reconstruction Protocol Element Specification Sequence

(0018,9933)

>Protocol Element Number

(0018,9921)

1

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3e “First Reconstruction Protocol Element Specification”

>Protocol Element Number

(0018,9921)

2

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3f “Second Reconstruction Protocol Element Specification”

Storage Protocol Element Specification Sequence

(0018,9935)

>Protocol Element Number

(0018,9921)

1

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3g “First Storage Protocol Element Specification”

>Protocol Element Number

(0018,9921)

2

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3h “Second Storage Protocol Element Specification”

>Protocol Element Number

(0018,9921)

3

>Parameters Specification Sequence

(0018,9913)

>>See Table AAAA.2-3i “Third Storage Protocol Element Specification”


The following tables reflect the semantic contents of constraint sequences but not the actual structure of the IOD. The centered rows in italics clarify the context of the constrained Attributes that follow by indicating which sequence in the performed module contains the constrained Attribute (as specified in the Selector Sequence Pointer).

Table AAAA.2-3a. Patient Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Patient's Age

(0010,1010)

1

absent

absent

GREATER_THAN

"16Y"


Table AAAA.2-3b. First Acquisition Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Acquistion Protocol Element Sequence (yym8, m9x1)

Protocol Element Name

(0018,9922)

1

(0018,9920)

1

EQUAL

Localizer: Lateral

Content Qualification

(0018,9004)

1

(0018,9934)

1

EQUAL

PRODUCT

Requested Series Description

(0018,9937)

1

(0018,9934)

1

EQUAL

"Localizers"

Acquisition Type

(0018,9302)

1

(0018,9920)

1

EQUAL

CONSTANT_ANGLE

Tube Angle

(0018,9303)

1

(0018,9920)

1

EQUAL

90

Acquisition Motion

(0018,9930)

1

(0018,9920)

1

EQUAL

FORWARD

>Acquisition Start Location Sequence (0018,9931)

Reference Location Label

(0018,9900)

1

(0018,9920), (0018,9931)

1\1

EQUAL

"Top of Head"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9920), (0018,9931)

1\1

EQUAL

(88986008, SCT, "Vertex of Head")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9920), (0018,9931)

1\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

>Acquisition End Location Sequence (0018,9932)

Reference Location Label

(0018,9900)

1

(0018,9920), (0018,9932)

1\1

EQUAL

"Bottom of 256mm Localizer"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9920), (0018,9932)

1\1

EQUAL

(88986008, SCT, "Vertex of Head")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9920), (0018,9932)

1\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

Offset Distance

(0018,9904)

1

(0018,9920), (0018,9932)

1\1

EQUAL

256

Offset Direction

(0018,9905)

1

(0018,9920), (0018,9932)

1\1

EQUAL

INFERIOR

>CT X-Ray Details Sequence (0018,9325) - First Beam

Beam Number

(300A,00C0)

1

(0018,9920), (0018,9325)

1\1

EQUAL

1

KVP

(0018,0060)

1

(0018,9920), (0018,9325)

1\1

EQUAL

120

X-ray Tube Current in mA

(0018,9330)

1

(0018,9920), (0018,9325)

1\1

EQUAL

50


Table AAAA.2-3c. Second Acquisition Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Acquistion Protocol Element Sequence (yym8, m9x1)

Protocol Element Name

(0018,9922)

1

(0018,9920)

2

EQUAL

Localizer: AP

Content Qualification

(0018,9004)

1

(0018,9934)

2

EQUAL

PRODUCT

Requested Series Description

(0018,9937)

1

(0018,9934)

2

EQUAL

"Localizers"

Acquisition Type

(0018,9302)

1

(0018,9920)

2

EQUAL

CONSTANT_ANGLE

Tube Angle

(0018,9303)

1

(0018,9920)

2

EQUAL

0

Acquisition Motion

(0018,9930)

1

(0018,9920)

2

EQUAL

FORWARD

>Acquisition Start Location Sequence (0018,9931)

Reference Location Label

(0018,9900)

1

(0018,9920), (0018,9931)

2\1

EQUAL

"Top of Head"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9920), (0018,9931)

2\1

EQUAL

(88986008, SCT, "Vertex of Head")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9920), (0018,9931)

2\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

>Acquisition End Location Sequence (0018,9932)

Reference Location Label

(0018,9900)

1

(0018,9920), (0018,9932)

2\1

EQUAL

"Bottom of 256mm Localizer"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9920), (0018,9932)

2\1

EQUAL

(88986008, SCT, "Vertex of Head")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9920), (0018,9932)

2\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

Offset Distance

(0018,9904)

1

(0018,9920), (0018,9932)

2\1

EQUAL

256

Offset Direction

(0018,9905)

1

(0018,9920), (0018,9932)

2\1

EQUAL

INFERIOR

>CT X-Ray Details Sequence (0018,9325) - First Beam

Beam Number

(300A,00C0)

1

(0018,9920), (0018,9325)

2\1

EQUAL

1

KVP

(0018,0060)

1

(0018,9920), (0018,9325)

2\1

EQUAL

120

X-ray Tube Current in mA

(0018,9330)

1

(0018,9920), (0018,9325)

2\1

EQUAL

50


Table AAAA.2-3d. Third Acquisition Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Acquistion Protocol Element Sequence (yym8, m9x1)

Protocol Element Name

(0018,9922)

1

(0018,9920)

3

EQUAL

Helical

Content Qualification

(0018,9004)

1

(0018,9934)

3

EQUAL

PRODUCT

Requested Series Description

(0018,9937)

1

(0018,9934)

3

EQUAL

"Raw Data Brain"

Acquisition Type

(0018,9302)

1

(0018,9920)

3

EQUAL

SPIRAL

Revolution Time

(0018,9305)

1

(0018,9920)

3

EQUAL

0.75

Single Collimation Width

(0018,9306)

1

(0018,9920)

3

EQUAL

0.5

Total Collimation Width

(0018,9307)

1

(0018,9920)

3

EQUAL

16

Table Speed

(0018,9309)

1

(0018,9920)

3

EQUAL

14

Table Speed per Rotation

(0018,9310)

1

(0018,9920)

3

EQUAL

10.5

Spiral Pitch Factor

(0018,9311)

1

(0018,9920)

3

EQUAL

0.656

CTDIvol

(0018,9345)

1

(0018,9920)

3

EQUAL

55.7

CTDI Phantom Type Code Sequence

(0018,9346)

1

(0018,9920)

3

EQUAL

(113690, DCM, "IEC Head Dosimetry Phantom")

CTDIvol Notification Trigger

(0018,9942)

1

(0018,9920)

3

EQUAL

80

Acquisition Motion

(0018,9930)

1

(0018,9920)

3

EQUAL

FORWARD

>Acquisition Start Location Sequence (0018,9931)

Reference Location Label

(0018,9900)

1

(0018,9920), (0018,9931)

3\1

EQUAL

"C1 Lamina"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9920), (0018,9931)

3\1

EQUAL

(14806007, SCT, "C1 vertebra")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9920), (0018,9931)

3\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

>Acquisition End Location Sequence (0018,9932)

Reference Location Label

(0018,9900)

1

(0018,9920), (0018,9932)

3\1

EQUAL

"Top of Head"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9920), (0018,9932)

3\1

EQUAL

(88986008, SCT, "Vertex of Head")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9920), (0018,9932)

3\1

EQUAL

(128120, DCM, "Plane through Superior Extent")

>CT X-Ray Details Sequence (0018,9325) - First Beam

Beam Number

(300A,00C0)

1

(0018,9920), (0018,9325)

3\1

EQUAL

1

KVP

(0018,0060)

1

(0018,9920), (0018,9325)

3\1

EQUAL

120

X-ray Tube Current in mA

(0018,9330)

1

(0018,9920), (0018,9325)

3\1

EQUAL

220

Exposure Modulation Type

(0018,9323)

1

(0018,9920), (0018,9325)

3\1

EQUAL

NONE

Data Collection Diameter

(0018,0090)

1

(0018,9920), (0018,9325)

3\1

EQUAL

240


Table AAAA.2-3e. First Reconstruction Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Reconstruction Protocol Element Sequence (yym8, m9x1)

Protocol Element Name

(0018,9922)

1

(0018,9934)

1

EQUAL

"Transverse"

Requested Series Description

(0018,9937)

1

(0018,9934)

1

EQUAL

"Transverse without Contrast"

Source Acquisition Protocol Element Number

(0018,9938)

1

(0018,9934)

1

EQUAL

3

Source Acquisition Beam Number

(0018,9939)

1

(0018,9934)

1

EQUAL

1

Convolution Kernel Group

(0018,9316)

1

(0018,9934)

1

EQUAL

"BRAIN"

Reconstruction Diameter

(0018,1100)

1

(0018,9934)

1

EQUAL

240

Rows

(0028,0010)

1

(0018,9934)

1

EQUAL

512

Columns

(0028,0011)

1

(0018,9934)

1

EQUAL

512

Slice Thickness

(0018,0050)

1

(0018,9934)

1

EQUAL

5

Spacing Between Slices

(0018,0088)

1

(0018,9934)

1

EQUAL

5

>Reconstruction Algorithm Sequence (0018,993D)

Algorithm Name

(0066,0036)

1

(0018,9934) (0018,993D)

1/1

EQUAL

"Opti-Brain"

Algorithm Name Code Sequence

(0066,0030)

1

(0018,9934) (0018,993D)

1/1

EQUAL

(B506, 99ACME, "OptiBrain")

>Reconstruction Start Location Sequence (0018,993B)

Reference Location Label

(0018,9900)

1

(0018,9934), (0018,993B)

1\1

EQUAL

"Bottom of Scan"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9934), (0018,993B)

1\1

EQUAL

(128160, DCM, "Acquired Volume")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9934), (0018,993B)

1\1

EQUAL

(128121, DCM, "Plane through Inferior Extent")

>Reconstruction End Location Sequence (0018,993C)

Reference Location Label

(0018,9900)

1

(0018,9934), (0018,993C)

1\1

EQUAL

"Top of Scan"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9934), (0018,993C)

1\1

EQUAL

(128160, DCM, "Acquired Volume")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9934), (0018,993C)

1\1

EQUAL

(128120, DCM, "Plane through Superior Extent")


Table AAAA.2-3f. Second Reconstruction Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Reconstruction Protocol Element Sequence (yym8, m9x1)

Protocol Element Name

(0018,9922)

1

(0018,9934)

2

EQUAL

"Volume"

Protocol Element Purpose

(0018,9924)

1

(0018,9934)

2

EQUAL

"For volume processing"

Protocol Element Characteristics Summary

(0018,9923)

1

(0018,9934)

2

EQUAL

"Thin slices"

Requested Series Description

(0018,9937)

1

(0018,9934)

1

EQUAL

"Brain Volume"

Source Acquisition Protocol Element Number

(0018,9938)

1

(0018,9934)

2

EQUAL

3

Source Acquisition Beam Number

(0018,9939)

1

(0018,9934)

2

EQUAL

1

Convolution Kernel Group

(0018,9316)

1

(0018,9934)

2

EQUAL

"BRAIN"

Reconstruction Diameter

(0018,1100)

1

(0018,9934)

2

EQUAL

240

Rows

(0028,0010)

1

(0018,9934)

2

EQUAL

512

Columns

(0028,0011)

1

(0018,9934)

2

EQUAL

512

Slice Thickness

(0018,0050)

1

(0018,9934)

2

EQUAL

0.5

Spacing Between Slices

(0018,0088)

1

(0018,9934)

2

EQUAL

0.25

>Reconstruction Algorithm Sequence (0018,993D)

Algorithm Name

(0066,0036)

1

(0018,9934) (0018,993D)

2/1

EQUAL

"Opti-Brain"

Algorithm Name Code Sequence

(0066,0030)

1

(0018,9934) (0018,993D)

2/1

EQUAL

(B506, 99ACME, "OptiBrain")

>Reconstruction Start Location Sequence (0018,993B)

Reference Location Label

(0018,9900)

1

(0018,9934), (0018,993B)

2\1

EQUAL

"Bottom of Scan"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9934), (0018,993B)

2\1

EQUAL

(128160, DCM, "Acquired Volume")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9934), (0018,993B)

2\1

EQUAL

(128121, DCM, "Plane through Inferior Extent")

>Reconstruction End Location Sequence (0018,993C)

Reference Location Label

(0018,9900)

1

(0018,9934), (0018,993C)

2\1

EQUAL

"Top of Scan"

Reference Basis Code Sequence

(0018,9902)

1

(0018,9934), (0018,993C)

2\1

EQUAL

(128160, DCM, "Acquired Volume")

Reference Geometry Code Sequence

(0018,9903)

1

(0018,9934), (0018,993C)

2\1

EQUAL

(128120, DCM, "Plane through Superior Extent")


Table AAAA.2-3g. First Storage Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Storage Protocol Element Sequence (0018,9936)

Protocol Element Name

(0018,9922)

1

(0018,9936)

1

EQUAL

"To PACS"

Protocol Element Purpose

(0018,9924)

1

(0018,9936)

1

EQUAL

"For reading"

Protocol Element Characteristics Summary

(0018,9923)

1

(0018,9936)

1

EQUAL

"Localizers, transverse and thin images"

Source Acquisition Protocol Element Number

(0018,9938)

1

(0018,9936)

1

EQUAL

1\2

Source Reconstruction Protocol Element Number

(0018,993A)

1

(0018,9936)

1

EQUAL

1\2

>Output Information Sequence (0040,4033)

>>DICOM Retrieval Sequence (0040,4071)

Destination AE

(2100,0140)

1

(0018,9936), (0040,4033), (0040,4071)

1\1\1

EQUAL

"MHPACS"


Table AAAA.2-3h. Second Storage Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Storage Protocol Element Sequence (0018,9936)

Protocol Element Name

(0018,9922)

1

(0018,9936)

2

EQUAL

"To 3D"

Protocol Element Purpose

(0018,9924)

1

(0018,9936)

2

EQUAL

"For 3D Processing"

Protocol Element Characteristics Summary

(0018,9923)

1

(0018,9936)

2

EQUAL

"Thin images"

Source Reconstruction Protocol Element Number

(0018,993A)

1

(0018,9936)

2

EQUAL

2

>Output Information Sequence (0040,4033)

>>DICOM Retrieval Sequence (0040,4071)

DestinationAE

(2100,0140)

1

(0018,9936), (0040,4033), (0040,4071)

2\1\1

EQUAL

"3DWS"


Table AAAA.2-3i. Third Storage Protocol Element Specification

Attribute

Selector Attribute

Selector Value Number

Selector Sequence Pointer

Selector Sequence Pointer Items

Constraint Type

Constraint Value

Storage Protocol Element Sequence (0018,9936)

Protocol Element Name

(0018,9922)

1

(0018,9936)

3

EQUAL

"Raw Data Archive"

Protocol Element Purpose

(0018,9924)

1

(0018,9936)

3

EQUAL

"For later recons"

Protocol Element Characteristics Summary

(0018,9923)

1

(0018,9936)

3

EQUAL

"Raw acq data"

Source Acquisition Protocol Element Number

(0018,9938)

1

(0018,9936)

3

EQUAL

3

>Output Information Sequence (0040,4033)

>>DICOM Retrieval Sequence (0040,4071)

DestinationAE

(2100,0140)

1

(0018,9936), (0040,4033), (0040,4071)

3\1\1

EQUAL

"RAWCACHE"


DICOM PS3.17 2024b - Explanatory Information