DICOM PS3.17 2025e - Explanatory Information

XXX.4 Uses of Presentation View Description in the Identification Module

XXX.4.1 Hanging Protocols

A Hanging Protocol Instance could select a set of orthogonal MPRs by use of the Image Sets Sequence (0072,0020).

Table XXX.4.1-1. Hanging Protocol Image Set Sequence Recommendations

Attribute Name

Tag

Comment

Image Sets Sequence

(0072,0020)

>Image Set Selector Sequence

(0072,0022)

>>Image Set Selector Usage Flag

(0072,0024)

Set to "MATCH"

>>Selector Attribute

(0072,0026)

Set to "0008,0016", the SOP Class UID Attribute Tag

>>Selector Attribute VR

(0072,0050)

Set to "UI", the SOP Class UID Attribute VR

>>Selector UI Value[i]

(0072,XXXX)

Set to "1.2.840.10008.​5.​1.​4.​1.​1.​11.​6", SOP Class UID of the Planar MPR VPS SOP Class

>>Image Set Selector Usage Flag

(0072,0024)

Set to "MATCH"

>>Selector Attribute

(0072,0026)

Set to "0054,0220", the View Code Sequence Attribute Tag

>>Selector Attribute VR

(0072,0050)

Set to "SQ", the View Code Sequence Attribute VR

>>Selector Code Value

(0072,0080)

Set to (81654009, SCT, "Coronal")

>>Image Set Number

(0072,0032)

Set to "1" for Image Set 1

>>Image Set Selector Usage Flag

(0072,0024)

Set to "MATCH"

>>Selector Attribute

(0072,0026)

Set to "0008,0016", the SOP Class UID Attribute Tag

>>Selector Attribute VR

(0072,0050)

Set to "UI", the SOP Class UID Attribute VR

>>Selector UI Value

(0072,XXXX)

Set to "1.2.840.10008.​5.​1.​4.​1.​1.​11.​6", SOP Class UID of the Planar MPR VPS SOP Class

>>Image Set Selector Usage Flag

(0072,0024)

Set to "MATCH"

>>Selector Attribute

(0072,0026)

Set to "0054,0220", the View Code Sequence Attribute Tag

>>Selector Attribute VR

(0072,0050)

Set to "SQ", the View Code Sequence Attribute VR

>>Selector Code Value

(0072,0080)

Set to (30730003, SCT, "Sagittal")

>>Image Set Number

(0072,0032)

Set to "2" for Image Set 2

>>Image Set Selector Usage Flag

(0072,0024)

Set to "MATCH"

>>Selector Attribute

(0072,0026)

Set to "0008,0016", the SOP Class UID Attribute Tag

>>Selector Attribute VR

(0072,0050)

Set to "UI", the SOP Class UID Attribute VR

>>Selector UI Value

(0072,XXXX)

Set to "1.2.840.10008.​5.​1.​4.​1.​1.​11.​6", SOP Class UID of the Planar MPR VPS SOP Class

>>Image Set Selector Usage Flag

(0072,0024)

Set to "MATCH"

>>Selector Attribute

(0072,0026)

Set to "0054,0220", the View Code Sequence Attribute Tag

>>Selector Attribute VR

(0072,0050)

Set to "SQ", the View Code Sequence Attribute VR

>>Selector Code Value

(0072,0080)

Set to (62824007, SCT, "Transverse")

>>Image Set Number

(0072,0032)

Set to "3" for Image Set 3


The display application would look for three Planar MPR Volumetric Presentation States - one Coronal, one Sagittal and one Transverse - and associate them with Image Sets in the view.

DICOM PS3.17 2025e - Explanatory Information