The recommendations in this section follows the description in Section C.11.27 “MPR Volumetric Presentation State Display Module” in PS3.3.
Table XXX.3.8-3. Volumetric Presentation State Display Module Recommendations
|
Contains two items, one for classifying the CT data and one for classifying the PET data. |
||
|
Contains one item in this sequence since the component has only one input. |
||
|
Set to "1" for the anatomy view - this is the CT input set. For reference, see Table XXX.3.8-1, Item 1, Volumetric Presentation State Input Sequence (0070,1201) > Input Number (0070,1207) = “1”. |
||
|
Contains one item in this sequence since the component has only one input. |
||
|
Set to "2" for the functional view - this is the PET input set. For reference, see Table XXX.3.8-1, Item 2, Volumetric Presentation State Input Sequence (0070,1201) > Input Number (0070,1207) = “2”. |
||
|
Set to "TABLE" to be able to map functional data to a color range. |
||
|
Set to "TABLE" to be able to set a transparency for the segmentation. |
||
|
Contains the two Weighting LUTs from Section XXX.5.2 to create the "Partially Transparent A over B" composting of two RGBA inputs. |
||
|
Set to an ICC Profile describing the transformation of the resulting RGB image into PCS-Values. |
||