Volumetric Presentation State Input Sequence
|
(0070,1201)
|
Four items are this sequence.
Most Attributes of all 4 items are set to exactly the same values, except Crop (0070,1204) is set to "YES" for the last three items, and the segmentations for the different lung nodule classifications are referenced via the corresponding Cropping Specification Index (0070,1205).
|
>ITEM 1
|
>Presentation Input Type
|
(0070,1202)
|
Set to "VOLUME".
|
>Referenced Image Sequence
|
(0008,1140)
|
Set reference(s) to the image(s) that make up the input volume.
|
>Crop
|
(0070,1204)
|
Set to "NO".
|
>ITEM 2
|
>Presentation Input Type
|
(0070,1202)
|
Set to "VOLUME".
|
>Referenced Image Sequence
|
(0008,1140)
|
Set reference(s) to the image(s) that make up the input volume.
|
>Crop
|
(0070,1204)
|
Set to "YES".
|
>Cropping Specification Index
|
(0070,1205)
|
Set to "1" to identify the segmentation cropping of the large nodules.
|
>ITEM 3
|
>Presentation Input Type
|
(0070,1202)
|
Set to "VOLUME".
|
>Referenced Image Sequence
|
(0008,1140)
|
Set reference(s) to the image(s) that make up the input volume.
|
>Crop
|
(0070,1204)
|
Set to "YES".
|
>Cropping Specification Index
|
(0070,1205)
|
Set to "2" to identify the segmentation cropping of the medium nodules.
|
>ITEM 4
|
>Presentation Input Type
|
(0070,1202)
|
Set to "VOLUME".
|
>Referenced Image Sequence
|
(0008,1140)
|
Set reference(s) to the image(s) that make up the input volume.
|
>Crop
|
(0070,1204)
|
Set to "YES".
|
>Cropping Specification Index
|
(0070,1205)
|
Set to "3" to identify the segmentation cropping of the small nodules.
|