DICOM PS3.17 2026c - Explanatory Information

XXX.3.4.3 Encoding Details

XXX.3.4.3.1 Presentation Animation Module Recommendations

Table XXX.3.4-1. Presentation Animation Module Recommendations

Attribute Name

Tag

Comment

Presentation Animation Style

(0070,1A01)

Set to "CROSSCURVE"

Recommended Animation Rate

(0070,1A03)

Set to provide a hint to the display application to automatically move the Planar MPR view along the curve through the volume, or omitted to leave the animation method to the discretion of the display application.

Animation Curve Sequence

(0070,1A04)

Set to the start point and end point of the straight-line curve.

Animation Step Size

(0070,1A05)

Set to (line length / (number of slices - 1)) as the distance between MPR views along the straight-line curve.


DICOM PS3.17 2026c - Explanatory Information