DICOM PS3.17 2024d - Explanatory Information
XXX.3.11.3.4 Render Display Module Recommendations

Table XXX.3.11.3.4-1. Render Display Module Recommendations

Attribute Name

Tag

Comment

Pixel Presentation

(0008, 9205)

Set to 'TRUE_COLOR'

Volume Stream Sequence

(0070,1A08)

One item in this sequence

Item #1 in Volume Stream Sequence

>Volumetric Presentation Input Set UID

(0070,1209)

UID1

>Presentation State

Classification Component

Sequence

(0070,1801)

Include three items, one for each classification component

>Item #1 in Presentation State Classification Component Sequence

>>Component Type

(0070, 1802)

Set to 'ONE_TO_RGBA'

>>Component Input Sequence

(0070, 1803)

Set only one item in this sequence, since the

component has only one input.

>>Item #1 in Component Input Sequence

>>>Volumetric Presentation Input Index

(0070,1804)

Set to '1' for the bones.

>>RGB LUT Transfer Function

(0028,140F)

Set to 'TABLE'.

>>Alpha LUT Transfer Function

(0028,1410)

Set to 'TABLE'.

>> RGB Palette Color Lookup Table Descriptors and Data

>> Alpha Palette Color Lookup Table Descriptor and Data

>Item #2 in Presentation State Classification Component Sequence

>>Component Type

(0070, 1802)

Set to 'ONE_TO_RGBA'

>>Component Input Sequence

(0070, 1803)

Set only one item in this sequence, since the

component has only one input.

>>Item #1 in Component Input Sequence

>>>Volumetric Presentation Input Index

(0070,1804)

Set to '2' for the segmented vessels and kidneys

>>RGB LUT Transfer Function

(0028,140F)

Set to 'TABLE'

>>Alpha LUT Transfer Function

(0028,1410)

Set to 'TABLE'.

>>RGB Palette Color Lookup Table Descriptors and Data

>>Alpha Palette Color Lookup Table Descriptor and Data

>Item #3 in Presentation State Classification Component Sequence

>>Component Type

(0070, 1802)

Set to 'ONE_TO_RGBA'

>>Component Input Sequence

(0070, 1803)

Set only one item in this sequence, since the

component has only one input.

>>Item #1 in Component Input Sequence

>>Volumetric Presentation Input Index

(0070,1804)

Set to '3' for the segmented tumor.

>RGB LUT Transfer Function

(0028,140F)

Set to 'EQUAL_RGB'

>Alpha LUT Transfer Function

(0028,1410)

Set to "NONE".

Presentation State Compositor Component Sequence

(0070,1805)

empty

ICC Profile

(0028,2000)

Value corresponding to IEC 61966-2-1:1999

Color Space

(0028,2002)

Set to SRGB


DICOM PS3.17 2024d - Explanatory Information