DICOM PS3.17 2026c - Explanatory Information
XXX.3.10.3.4 Render Display Module Recommendations

Table XXX.3.10.3.4-1. Render Display Module Recommendations

Attribute Name

Tag

Comment

Pixel Presentation

(0008,9205)

Set to "TRUE_COLOR"

Volume Stream Sequence

(0070,1A08)

Item #1 of Volume Stream Sequence

>Volumetric Presentation Input Set UID

(0070,1209)

Set to UID1, referencing the first item in the Volumetric Presentation Input Set Sequence (0070,120A)

>Presentation State Classification Component Sequence

(0070,1801)

>Item #1 of Presentation State Classification Component Sequence

>>Component Type

(0070,1802)

Set to ONE_TO_RGBA

>>Component Input Sequence

(0070,1803)

>>Item #1 of Component Input Sequence

>>>Volumetric Presentation Input Index

(0070,1804)

Set to 1 referencing the first item in Volumetric Presentation State Input Sequence (0070,1201)

>>RGB LUT Transfer Function

(0028,140F)

Set to TABLE

>>Alpha LUT Transfer Function

(0028,1410)

Set to TABLE

>>RGB Palette Color Lookup Table Descriptors and Data

>>Alpha Palette Color Lookup Table Descriptor and Data

Presentation State Compositor Component Sequence

(0070,1805)

empty

ICC Profile

(0028,2000)

Value corresponding to IEC 61966-2-1:1999

Color Space

(0028,2002)

Set to SRGB


DICOM PS3.17 2026c - Explanatory Information