DICOM PS3.17 2024a - Explanatory Information

SSS Use of Image Libraries in SR Documents (Informative)

This Annex contains examples of the use of Image Library templates within SR Documents.

SSS.1 Image Library for PET-CT Example

This PET-CT example dillustrates an Image Library in which Attributes of images for two modalities are described, with common Attributes factored out of the individual image references.

Note

  1. Only the Attributes of relevance to SUV and spatial measurements are included, not a complete description of all aspects of acquisition.

  2. Only two images for each modality are described, rather than all slices acquired, since it is usually only necessary to describe images that are referenced elsewhere in the SR content tree, e.g., on which a region of interest is specified from which measurements are made.

Table SSS.1-1. Image Library for PET-CT Example

Node

Code Meaning of Concept Name

Code Meaning or Example Value

TID

1.n

Image Library

TID 1600

1.n.1

Image Library Group

TID 1600

1.n.1.1

Modality

PET

TID 1602

1.n.1.2

Target Region

Whole Body

TID 1602

1.n.1.3

Study Date

20030417

TID 1602

1.n.1.4

Acquisition Date

20030417

TID 1602

1.n.1.5

Acquisition Time

094513

TID 1602

1.n.1.6

Frame of Reference UID

1.2.3.xyz

TID 1602

1.n.1.7

Pixel Data Rows

128

TID 1602

1.n.1.8

Pixel Data Columns

128

TID 1602

1.n.1.9

Horizontal Pixel Spacing

4.0 mm

TID 1604

1.n.1.10

Vertical Pixel Spacing

4.0 mm

TID 1604

1.n.1.11

Spacing Between Slices

4.0 mm

TID 1604

1.n.1.12

Slice Thickness

4.0 mm

TID 1604

1.n.1.13

Image Orientation (Patient) Row X

1

TID 1604

1.n.1.14

Image Orientation (Patient) Row Y

0

TID 1604

1.n.1.15

Image Orientation (Patient) Row Z

0

TID 1604

1.n.1.16

Image Orientation (Patient) Column X

0

TID 1604

1.n.1.17

Image Orientation (Patient) Column Y

1

TID 1604

1.n.1.18

Image Orientation (Patient) Column Z

0

TID 1604

1.n.1.19

Radionuclide

^18^Fluorine

TID 1607

1.n.1.20

Radiopharmaceutical agent

Fluorodeoxyglucose F^18^

TID 1607

1.n.1.21

Radiopharmaceutical Start DateTime

20030417084513

TID 1607

1.n.1.22

Radionuclide Total Dose

277000000 Bq

TID 1607

1.n.1.23

PET Radionuclide Incubation Time

60 min

TID 1607

1.n.1.24

Glucose

5.5 mmol/l

TID 1607

1.n.1.24.1

Glucose Measurement Date

20030417

TID 1607

1.n.1.24.2

Glucose Measurement Time

083043

TID 1607

1.n.1.25

IMAGE - PET image #1

TID 1601

1.n.1.25.1

Image Position (Patient) X

-288.0

TID 1604

1.n.1.25.2

Image Position (Patient) Y

288.0

TID 1604

1.n.1.25.3

Image Position (Patient) Z

136.0

TID 1604

1.n.1.26

IMAGE - PET image #2

TID 1601

1.n.1.26.1

Image Position (Patient) X

-288.0

TID 1604

1.n.1.26.2

Image Position (Patient) Y

288.0

TID 1604

1.n.1.26.3

Image Position (Patient) Z

140.0

TID 1604

1.n.2

Image Library Group

TID 1600

1.n.2.1

Modality

CT

TID 1602

1.n.2.2

Target Region

Whole Body

TID 1602

1.n.2.3

Study Date

20030417

TID 1602

1.n.2.4

Frame of Reference UID

1.2.3.xyz

TID 1602

1.n.2.5

Pixel Data Rows

512

TID 1602

1.n.2.6

Pixel Data Columns

512

TID 1602

1.n.2.7

Horizontal Pixel Spacing

1.171875 mm

TID 1604

1.n.2.8

Vertical Pixel Spacing

1.171875 mm

TID 1604

1.n.2.9

Spacing Between Slices

4 mm

TID 1604

1.n.2.10

Slice Thickness

4 mm

TID 1604

1.n.2.11

Image Orientation (Patient) Row X

1

TID 1604

1.n.2.12

Image Orientation (Patient) Row Y

0

TID 1604

1.n.2.13

Image Orientation (Patient) Row Z

0

TID 1604

1.n.2.14

Image Orientation (Patient) Column X

0

TID 1604

1.n.2.15

Image Orientation (Patient) Column Y

1

TID 1604

1.n.2.16

Image Orientation (Patient) Column Z

0

TID 1604

1.n.2.17

CTAcquisition Type

Spiral Acquisition

TID 1605

1.n.2.18

Reconstruction Algorithm

Filtered Back Projection

TID 1605

1.n.2.19

IMAGE - CT image #1

TID 1601

1.n.2.19.1

Image Position (Patient) X

-288.0

TID 1604

1.n.2.19.2

Image Position (Patient) Y

288.0

TID 1604

1.n.2.19.3

Image Position (Patient) Z

136.0

TID 1604

1.n.2.20

IMAGE - CT image #2

TID 1601

1.n.2.20.1

Image Position (Patient) X

-288.0

TID 1604

1.n.2.20.2

Image Position (Patient) Y

288.0

TID 1604

1.n.2.20.3

Image Position (Patient) Z

140.0

TID 1604


DICOM PS3.17 2024a - Explanatory Information