1
|
Performed Imaging Agent Administration
|
|
TID 11020
|
1.1
|
Language of Content Item and Descendants
|
English
|
TID 11020
TID 1204
|
1.1.1
|
Country of Language
|
United States
|
TID 1204
|
1.2
|
Observer Type
|
Person
|
TID 1002
CID 270
|
1.3
|
Person Observer Name
|
Doe^Jane
|
TID 1003
|
1.4
|
Observer Type
|
Device
|
TID 1002
Since Observer Type is Device
|
1.5
|
Device Observer UID
|
1.2.3.4.47110815.1
|
TID 1004
|
1.6
|
Device Observer Manufacturer
|
Injector Corporation
|
TID 1004
|
1.7
|
Device Observer Model Name
|
XYZ INJECTOR
|
TID 1004
|
1.8
|
Device Observer Serial Number
|
1234567890
|
TID 1004
|
1.9
|
Station AE Title
|
XYZINJAET
|
TID 1004
|
1.10
|
Procedure Study Instance UID
|
1.2.3.4.47110815.2
|
TID 1005
Defaults to Study Instance UID (0020,000D) of General Study Module
|
1.11
|
Accession Number
|
123456789
|
TID 1005
Defaults to (0008,0050)
|
1.12
|
Medication given
|
|
TID 8131
|
1.12.1
|
Route of administration
|
(47625008, SCT, "Intravenous route")
|
TID 8131
CID 11
|
1.12.2
|
Mixture
|
|
TID 8131
|
1.12.2.1
|
Drug administered
|
Prednisone
|
TID 8131 row 7
|
1.12.2..2
|
Medication Type
|
(130259, DCM, "Contrast Reaction Prophylactic Agent")
|
TID 8131
CID 76
|
1.12.2.3
|
Dosage
|
2 ml
|
TID 8131
CID 82 (Units)
|
1.12.2.4
|
Concentration
|
5 mg/ml
|
TID 8131
CID 82 (Units)
|
1.13
|
Patient Characteristics
|
|
TID 10024
|
1.13.1
|
Patient State
|
(414417004, SCT, "History of renal failure")
|
TID 10024
CID 64
|
1.13.2
|
Subject Age
|
25
|
TID 10024
CID 7456
UNITS=EV (a, UCUM, "year")
|
1.13.3
|
Subject Sex
|
(M, DCM, "Male")
|
TID 10024
CID 7455
|
1.13.4
|
Patient Height
|
175
|
TID 10024
UNITS=EV (cm, UCUM, "cm")
|
1.13.5
|
Patient Weight
|
65
|
TID 10024
UNITS=EV (kg, UCUM, "kg")
|
1.13.6
|
Body Mass Index
|
21.23
|
TID 10024
UNITS=EV (kg/m2, UCUM, "kg/m2")
|
1.13.6.1
|
Equation
|
(122265, DCM, "BMI=Wt/Ht^2")
|
TID 10024
|
1.13.7
|
Serum Creatinine
|
2.7
|
TID 10024
UNITS=DT (mg/dl, UCUM, "mg/dl")
|
1.13.8
|
Glomerular Filtration Rate
|
38
|
TID 10024
UNITS=DT (ml/min{1.73_m2}, UCUM, "ml/min/1.73m2")
|
1.13.8.1
|
Measurement Method
|
(113570, DCM, "Cockroft-Gault Formula estimation of GFR")
|
TID 10024
CID 10047
|
1.13.8.2
|
Equivalent meaning of concept name
|
(33914-3, LN, "Glomerular Filtration Rate (MDRD)")
|
TID 10024
CID 10046
|
1.14
|
Imaging Agent Information
|
|
TID 11002
|
1.14.1
|
Imaging Agent Identifier
|
INJECTOR_CONTRAST_AGENT
|
TID 11002
|
1.14.2
|
Imaging Agent Warmed
|
(373066001, SCT, "Yes")
|
TID 11002
CID 231
|
1.14.3
|
Imaging Agent Component Usage
|
|
TID 11002
|
1.14.3.1
|
Imaging Agent Component
|
|
TID 11004
|
1.14.3.1.1
|
Drug administered
|
(353903006, SCT, "Iopromide")
|
TID 11004
CID 12
|
1.14.3.1.2
|
Active Ingredient
|
(44588005, SCT, "Iodine")
|
TID 11004
CID 13
|
1.14.3.1.3
|
Concentration
|
370
|
TID 11004
UNITS = EV (mg/ml, "UCUM", "mg/ml")
|
1.14.3.1.4
|
Osmolality at 37C
|
770
|
TID 11004
UNITS=EV (mosm/kg, UCUM, "mosm/kg")
|
1.14.3.1.5
|
Viscosity at 37C
|
10
|
TID 11004
UNITS = EV (cP, "UCUM", "centi Poise")
|
1.14.3.1.6
|
Unit of Presentation
|
(68276009, SCT, "Bottle")
|
TID 11004
CID 68
|
1.14.3.1.7
|
Imaging Agent Volume Per Unit of Presentation
|
500
|
TID 11004
UNITS=EV (ml, UCUM, "ml")
|
1.14.3.1.8
|
Medical Product Expiration Date
|
20190301
|
TID 11004
|
1.14.3.1.9
|
Manufacturer Name
|
ContrastMed Corp
|
TID 11004
|
1.14.3.1.10
|
Brand Name
|
ContrastStuff 370
|
TID 11004
|
1.14.3.1.11
|
Barcode Value
|
-07363935
|
TID 11004
PZN number
|
1.14.3.1.12
|
Lot Identifier
|
4B17010
|
TID 11004
|
1.14.3.2
|
Component Volume
|
97.84
|
TID 11002
UNITS=EV (ml, UCUM, "ml")
|
1.15
|
Imaging Agent Information
|
|
TID 11002
|
1.15.1
|
Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11002
|
1.15.2
|
Imaging Agent Warmed
|
(373066001, SCT, "Yes")
|
TID 11002
CID 231
|
1.15.3
|
Imaging Agent Component Usage
|
|
TID 11002
|
1.15.3.1
|
Imaging Agent Component
|
|
TID 11004
|
1.15.3.1.1
|
Drug administered
|
(262003004, SCT, "Saline")
|
TID 11004
CID 70
|
1.15.3.1.2
|
Unit of Presentation
|
(68276009, SCT, "Bottle")
|
TID 11004
CID 68
|
1.15.3.1.3
|
Imaging Agent Volume Per Unit of Presentation
|
500
|
TID 11004
UNITS=EV (ml, UCUM, "ml")
|
1.15.3.1.4
|
Manufacturer Name
|
Saline Water Corp
|
TID 11004
|
1.15.3.1.5
|
Brand Name
|
Isotonic Natriumchloride Solution
|
TID 11004
|
1.15.3.1.6
|
Barcode Value
|
-00854309
|
TID 11004
PZN number
|
1.15.3.1.7
|
Lot Identifier
|
13CQ4857
|
TID 11004
|
1.15.3.2
|
Component Volume
|
200
|
TID 11002
UNITS=EV (ml, UCUM, "ml")
|
1.16
|
Imaging Agent Information
|
|
TID 11002
|
1.16.1
|
Imaging Agent Identifier
|
ORAL_CONTRAST_AGENT
|
TID 11002
|
1.16.2
|
Imaging Agent Warmed
|
(373067005, SCT, "No")
|
TID 11002
CID 231
|
1.16.3
|
Imaging Agent Component Usage
|
|
TID 11002
|
1.16.3.1
|
Imaging Agent Component
|
|
TID 11004
|
1.16.3.1.1
|
Drug administered
|
(47192000, SCT, "Meglumine diatrizoate")
|
TID 11004
CID 12
|
1.16.3.1.2
|
Active Ingredient
|
(44588005, SCT, "Iodine")
|
TID 11004
CID 13
|
1.16.3.1.3
|
Concentration
|
370
|
TID 11004
UNITS = EV (mg/ml, "UCUM", "mg/ml")
|
1.16.3.1.4
|
Viscosity at 37C
|
8.9
|
TID 11004
UNITS = EV (cP, "UCUM", "centiPoise")
|
1.16.3.1.5
|
Unit of Presentation
|
(68276009, SCT, "Bottle")
|
TID 11004
CID 68
|
1.16.3.1.6
|
Imaging Agent Volume Per Unit of Presentation
|
100
|
TID 11004
UNITS=EV (ml, UCUM, "ml")
|
1.16.3.1.7
|
Manufacturer Name
|
ContrastMed Corp
|
TID 11004
|
1.16.3.1.8
|
Brand Name
|
OralContrastofin
|
TID 11004
|
1.16.3.1.9
|
Barcode Value
|
-00408497
|
TID 11004
PZN number
|
1.16.3.1.10
|
Lot Identifier
|
6X14325
|
TID 11004
|
1.16.3.2
|
Component Volume
|
24.4
|
TID 11002
UNITS=EV (ml, UCUM, "ml")
|
1.16.4
|
Imaging Agent Component Usage
|
|
TID 11002
|
1.16.4.1
|
Imaging Agent Component
|
|
TID 11002
TID 11004
|
1.16.4.1.1
|
Drug administered
|
(11713004, SCT, "Water")
|
TID 11004
CID 12
|
1.16.4.1.2
|
Unit of Presentation
|
(68276009, SCT, "Bottle")
|
TID 11004
CID 68
|
1.16.4.1.3
|
Imaging Agent Volume Per Unit of Presentation
|
1000
|
TID 11004
UNITS=EV (ml, UCUM, "ml")
|
1.16.4.1.4
|
Manufacturer Name
|
Fresh Water Corp
|
TID 11004
|
1.16.4.1.5
|
Brand Name
|
BestWaterEver
|
TID 11004
|
1.16.4.1.6
|
Barcode Value
|
-4801694
|
TID 11004
PZN number
|
1.16.4.2
|
Component Volume
|
975.6
|
TID 11002
UNITS=EV (ml, UCUM, "ml")
|
1.17
|
Summary
|
Administered 1000 ml of Oral Contrastofin via oral route and 88ml of ContrastStuff 370 via intravenous route in Left Arm Vein.
|
TID 11020
|
1.18
|
Imaging Agent Administration Consumable
|
|
TID 11005
|
1.18.1
|
Imaging Agent Administration Consumable Type
|
(467354001, SCT, "Contrast medium injection system manifold kit")
|
TID 11005
CID 69
|
1.18.2
|
Quantity of Material
|
1
|
TID 11005
|
1.18.2.1
|
Consumable is New
|
(373066001, SCT, "Yes")
|
TID 11005
CID 231
|
1.18.3
|
Billing Code
|
317627C
|
TID 11005
|
1.18.4
|
Medical Product Expiration Date
|
20221031
|
TID 11005
|
1.18.5
|
Manufacturer Name
|
Injector Corp
|
TID 11005
|
1.18.6
|
Barcode Value
|
(01) 14250299676272(19)13111501(17)181000
|
TID 11005
|
1.18.7
|
Lot Identifier
|
13111501
|
TID 11005
|
1.19
|
Imaging Agent Administration Consumable
|
|
TID 11005
|
1.19.1
|
Imaging Agent Administration Consumable Type
|
(79068005, SCT, "Needle")
|
TID 11005
CID 69
|
1.19.2
|
Quantity of Material
|
1
|
TID 11005
|
1.19.2.1
|
Consumable is New
|
(373066001, SCT, "Yes")
|
TID 11005
CID 231
|
1.19.3
|
Billing Code
|
206342
|
TID 11005
|
1.19.4
|
Medical Product Expiration Date
|
20181130
|
TID 11005
|
1.19.5
|
Manufacturer Name
|
Dr. Poke Inc.
|
TID 11005
|
1.19.6
|
Brand Name
|
Sterile Standard, Green
|
TID 11005
|
1.20
|
Imaging Agent Administration Consumable
|
|
TID 11005
|
1.20.1
|
Imaging Agent Administration Consumable Type
|
(68276009, SCT, "Bottle")
|
TID 11005
CID 69
|
1.20.2
|
Quantity of Material
|
1
|
TID 11005
|
1.20.2.1
|
Consumable is New
|
(373066001, SCT, "Yes")
|
TID 11005
CID 231
|
1.20.3
|
Billing Code
|
47110815
|
TID 11005
|
1.20.4
|
Medical Product Expiration Date
|
20191001
|
TID 11005
|
1.21
|
Imaging Agent Administration Steps
|
|
TID 11006
|
1.21.1
|
Imaging Agent Administration Steps Name
|
Abdomen intestinal and vessel contrast processing
|
TID 11006
|
1.21.2
|
Imaging Agent Administration Steps Description
|
This contrast processing is given by an oral administration of first 2 hours in advance of the procedure. I.v. administration is done with a pre-inject to determine scan delay time. Patent test injection applies as a default procedure.
|
TID 11006
|
1.21.3
|
Imaging Agent Administration Step
|
|
TID 11007
|
1.21.3.1.
|
Imaging Agent Administration Step Identifier
|
ORAL_STEP_1
|
TID 11007
|
1.21.3.2
|
Imaging Agent Administration Performed Step UID
|
1.2.3.4.47110815.3
|
TID 11007
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.3.3
|
Administration Mode
|
(130174, DCM, "Manual Administration")
|
TID 11007
CID 63
|
1.21.3.4
|
Person Role in Organization
|
(121025, DCM, "Patient")
|
TID 11007
CID 7450
Since "Administration Mode" is "Manual Administration", condition holds (self-administration)
|
1.21.3.5
|
Administration Step Type
|
(130249, DCM, "Diagnostic Administration")
|
TID 11007
CID 72
|
1.21.3.6
|
Scan Delay
|
7200
|
TID 11007
UNITS = EV (s, UCUM, "s")
|
1.21.3.7
|
Route of Administration
|
(26643006, SCT, "Oral route")
|
TID 11007
CID 11
|
1.21.3.8
|
Imaging Agent Administration Phase
|
|
TID 11008
|
1.21.3.8.1
|
Imaging Agent Administration Phase Identifier
|
ORAL_PHASE
|
TID 11008
|
1.21.3.8.2
|
Imaging Agent Administration Performed Phase UID
|
1.2.3.4.47110815.4
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.3.8.3
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.3.8.3.1
|
Referenced Imaging Agent Identifier
|
ORAL_CONTRAST_AGENT
|
TID 11003
Value of 1.16.1
|
1.21.3.8.3.2
|
Volume Administered
|
1000
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.3.8.4
|
1.21.3.8.3.3
|
DateTime Started
|
20181012101531
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.3.8.3.4
|
Duration
|
2700
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.3.8.4
|
Total Phase Volume Administered
|
1000
|
TID 11008
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.3.8.3.2
|
1.21.3.8.5
|
DateTime Started
|
20181012101531
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.3.8.6
|
Duration
|
2700
|
TID 11008
UNITS = EV (s, UCUM, "s")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4
|
Imaging Agent Administration Step
|
|
TID 11007
|
1.21.4.1
|
Imaging Agent Administration Step Identifier
|
EXTRAVASATION_TEST_STEP_2
|
TID 11007
|
1.21.4.2
|
Imaging Agent Administration Performed Step UID
|
1.2.3.4.47110815.5
|
TID 11007
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4.3
|
Administration Mode
|
(130173, DCM, "Automated Administration")
|
TID 11007
CID 63
|
1.21.4.4
|
Administration Step Type
|
(130247, DCM, "Patency Test Injection")
|
TID 11007
CID 72
|
1.21.4.5
|
Route of Administration
|
(47625008, SCT, "Intravenous route")
|
TID 11007
CID 11
|
1.21.4.5.1
|
Site of
|
(261459001, SCT, "Via arm vein")
|
TID 11007
CID 3746
Since "Route of Administration" is "Intravenous route"
|
1.21.4.5.1.1
|
Laterality
|
(7771000, SCT, "Left")
|
TID 11007
CID 244
Since "Site of" is "Via arm vein"
|
1.21.4.6
|
Imaging Agent Administration Phase
|
|
TID 11008
|
1.21.4.6.1
|
Imaging Agent Administration Phase Identifier
|
EXTRAVASATION_TEST_PHASE
|
TID 11008
|
1.21.4.6.2
|
Imaging Agent Administration Performed Phase UID
|
1.2.3.4.47110815.6
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4.6.3
|
Imaging Agent Administration Phase Type
|
(130171, DCM, "Automatic with Manual Inject Phase")
|
TID 11008
CID 62
Since 1.21.4.3 (Administration Mode) is "Automated Administration"
|
1.21.4.6.4
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.4.6.4.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11003
Value of 1.15.1
|
1.21.4.6.4.2
|
Volume Administered
|
30
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.4.6.5 and 1.21.4.9.1
|
1.21.4.6.4.3
|
Starting Flow Rate of administration
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
|
1.21.4.6.4.4
|
Peak Flow Rate in Phase Activity
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
Since 1.21.4.3 (Administration Mode) is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4.6.4.5
|
Peak Pressure in Phase Activity
|
2.5
|
TID 11003
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.4.3 (Administration Mode) is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4.6.4.6
|
Initial Volume of Imaging Agent in Container
|
197
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4.6.4.7
|
Residual Volume of Imaging Agent in Container
|
167
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.4.6.4.8
|
DateTime Started
|
20181012121537
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.4.6.4.9
|
Duration
|
10
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.4.6.5
|
Total Phase Volume Administered
|
30
|
TID 11008
UNITS = EV (ml, UCUM, "ml")
In this case the same value as 1.21.4.6.4.2 and 1.21.4.9.1
|
1.21.4.6.6
|
DateTime Started
|
20181012121537
|
TID 11008
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration"
|
1.21.4.6.7
|
Duration
|
10
|
TID 11008
UNITS = EV (s, UCUM, "s")
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration")
|
1.21.4.7
|
Number of Injector Heads
|
2
|
TID 11007
|
1.21.4.8
|
Programmable Device
|
(373066001, SCT, "Yes")
|
TID 11007
CID 231
|
1.21.4.9
|
Manually triggered injection information
|
|
TID 11007
Since 1.21.4.3 (Administration Mode) is "Automated Administration" and root Concept Name Code Sequence is "Performed Imaging Agent Administration"
|
1.21.4.9.1
|
Total Step Volume Administered
|
30
|
TID 11007
UNITS = EV (ml, UCUM, "ml")
In this case the same value as 1.21.4.6.4.2 and 1.21.4.6.5
|
1.21.4.9.2
|
Total number of manually triggered injections
|
1
|
TID 11007
|
1.21.5
|
Imaging Agent Administration Step
|
|
TID 11007
|
1.21.5.1
|
Imaging Agent Administration Step Identifier
|
DELAY_ESTIMATE_STEP_3
|
TID 11007
|
1.21.5.2
|
Imaging Agent Administration Performed Step UID
|
1.2.3.4.47110815.7
|
TID 11007
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.3
|
Administration Mode
|
(130173, DCM, "Automated Administration")
|
TID 11007
CID 63
|
1.21.5.4
|
Administration Step Type
|
(130248, DCM, "Transit Time Test Injection")
|
TID 11007
CID 72
|
1.21.5.5
|
Pressure Limit
|
15
|
TID 11007
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.5.3 is "Automated Administration"
|
1.21.5.6
|
Route of Administration
|
(47625008, SCT, "Intravenous route")
|
TID 11007
CID 11
|
1.21.5.6.1
|
Site of
|
(261459001, SCT, "Via arm vein")
|
TID 11007
CID 3746
Since "Route of Administration" is "Intravenous route"
|
1.21.5.6.1.1
|
Laterality
|
(7771000, SCT, "Left")
|
TID 11007
CID 244
Since "Site of" is "Via arm vein"
|
1.21.5.7
|
Imaging Agent Administration Phase
|
|
TID 11008
|
1.21.5.7.1
|
Imaging Agent Administration Phase Identifier
|
DELAY_ESTIMATE_PHASE_1
|
TID 11008
|
1.21.5.7.2
|
Imaging Agent Administration Performed Phase UID
|
1.2.3.4.47110815.8
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.7.3
|
Imaging Agent Administration Phase Type
|
(130168, DCM, "Automatic Administration Phase")
|
TID 11008
CID 62
Since 1.21.5.3 is "Automated Administration"
|
1.21.5.7.4
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.5.7.4.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_CONTRAST_AGENT
|
TID 11003
Value of 1.14.1
|
1.21.5.7.4.2
|
Volume Administered
|
10
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.5.7.5
|
1.21.5.7.4.3
|
Starting Flow Rate of administration
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
|
1.21.5.7.4.4
|
Peak Flow Rate in Phase Activity
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
Since 1.21.5.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.7.4.5
|
Peak Pressure in Phase Activity
|
2
|
TID 11003
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.5.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.7.4.6
|
Initial Volume of Imaging Agent in Container
|
195
|
TID 11003
UNITS = EV (ml, UCUM,
"ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.7.4.7
|
Residual Volume of Imaging Agent in Container
|
185
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.7.4.8
|
DateTime Started
|
20181012121637
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.5.7.4.9
|
Duration
|
3.3
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.5.7.5
|
Total Phase Volume Administered
|
10
|
TID 11008
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.5.7.4.2
|
1.21.5.7.6
|
DateTime Started
|
20181012121637
|
TID 11008
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration"
|
1.21.5.7.7
|
Duration
|
3.3
|
TID 11008
UNITS = EV (s, UCUM, "s")
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration")
|
1.21.5.8
|
Imaging Agent Administration Phase
|
|
TID 11008
|
1.21.5.8.1
|
Imaging Agent Administration Phase Identifier
|
DELAY_ESTIMATE_PHASE_2
|
TID 11008
|
1.21.5.8.2
|
Imaging Agent Administration Performed Phase UID
|
1.2.3.4.47110815.9
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.8.3
|
Imaging Agent Administration Phase Type
|
(130168, DCM, "Automatic Administration Phase")
|
TID 11008
CID 62
Since 1.21.5.3 is "Automated Administration"
|
1.21.5.8.4
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.5.8.4.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11003
Value of 1.15.1
|
1.21.5.8.4.2
|
Volume Administered
|
30
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.5.9.5
|
1.21.5.8.4.3
|
Starting Flow Rate of administration
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
|
1.21.5.8.4.4
|
Peak Flow Rate in Phase Activity
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
Since 1.21.5.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.8.4.5
|
Peak Pressure in Phase Activity
|
5
|
TID 11003
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.5.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.8.4.6
|
Initial Volume of Imaging Agent in Container
|
166
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.8.4.7
|
Residual Volume of Imaging Agent in Container
|
136
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.5.8.4.8
|
DateTime Started
|
20181012121640.3
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.5.8.4.9
|
Duration
|
10
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.5.8.5
|
Total Phase Volume Administered
|
30
|
TID 11008
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.5.9.4.2
|
1.21.5.8.6
|
DateTime Started
|
20181012121640.3
|
TID 11008
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration"
|
1.21.5.8.7
|
Duration
|
10
|
TID 11008
UNITS = EV (s, UCUM, "s")
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration")
|
1.21.5.9
|
Imaging Agent Administration Graph
|
|
TID 11023
|
1.21.5.9.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_CONTRAST_AGENT
|
TID 11023
|
1.21.5.9.2
|
Flow Rate vs time
|
|
TID 3990
Concept name is parameter $MeasurmentGraph
|
1.21.5.9.2.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
TID 3990
Parameter $X-Concept
|
1.21.5.9.2.2
|
Y-Concept
|
(122094, DCM, "Rate of administration")
|
TID 3990
Parameter $Y-Concept
|
1.21.5.9.2.3
|
Flow Rate vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.10
|
TID 3990
|
1.21.5.9.3
|
Pressure vs time
|
|
TID 3990
Concept name is parameter $MeasurementGraph of TID 3990
|
1.21.5.9.3.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
Parameter $X-Concept of TID 3990
|
1.21.5.9.3.2
|
Y-Concept
|
(279046003, SCT, "Pressure")
|
Parameter $Y-Concept of TID 3990
|
1.21.5.9.3.3
|
Pressure vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.10
|
TID 3990
All graphs are in the same image in this example.
|
1.21.5.10
|
Imaging Agent Administration Graph
|
|
TID 11023
|
1.21.5.10.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11023
|
1.21.5.10.2
|
Flow Rate vs time
|
|
TID 3990
Concept name is parameter $MeasurementGraph of TID 3990
|
1.21.5.10.2.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
TID 3990
Parameter $X-Concept
|
1.21.5.10.2.2
|
Y-Concept
|
(122094, DCM, "Rate of administration")
|
TID 3990
Parameter $Y-Concept
|
1.21.5.10.2.3
|
Flow Rate vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.10
|
TID 3990
|
1.21.5.10.3
|
Pressure vs time
|
|
TID 3990
Concept name is parameter $MeasurementGraph of TID 3990
|
1.21.5.10.3.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
Parameter $X-Concept of TID 3990
|
1.21.5.10.3.2
|
Y-Concept
|
(279046003, SCT, "Pressure")
|
Parameter $Y-Concept of TID 3990
|
1.21.5.10.3.3
|
Pressure vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.10
|
TID 3990
|
1.21.5.11
|
Number of Injector Heads
|
2
|
TID 11007
|
1.21.5.12
|
Programmable Device
|
(373066001, SCT, "Yes")
|
TID 11007
CID 231
|
1.21.6
|
Imaging Agent Administration Step
|
|
TID 11007
|
1.21.6.1
|
Imaging Agent Administration Step Identifier
|
DIAGNOSTIC_STEP_4
|
TID 11007
|
1.21.6.2
|
Imaging Agent Administration Performed Step UID
|
1.2.3.4.47110815.10
|
TID 11007
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.3
|
Administration Mode
|
(130173, DCM, "Automated Administration")
|
TID 11007
CID 63
|
1.21.6.4
|
Administration Step Type
|
(130249, DCM, "Diagnostic Administration")
|
TID 11007
CID 72
|
1.21.6.5
|
Scan Delay
|
12
|
TID 11007
UNITS = EV (s, UCUM, "s")
|
1.21.6.6
|
Pressure Limit
|
15
|
TID 11007
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.6.3 is "Automated Administration"
|
1.21.6.7
|
Route of Administration
|
(47625008, SCT, "Intravenous route")
|
TID 11007
CID 11
|
1.21.6.7.1
|
Site of
|
(261459001, SCT, "Via arm vein")
|
TID 11007
CID 3746
Since "Route of Administration" is "Intravenous route"
|
1.21.6.7.1.1
|
Laterality
|
(7771000, SCT, "Left")
|
TID 11007
CID 244
Since "Site of" is "Via arm vein"
|
1.21.6.8
|
Imaging Agent Administration Phase
|
|
TID 11008
|
1.21.6.8.1
|
Imaging Agent Administration Phase Identifier
|
DIAGNOSTIC_INJECTION_PHASE_1
|
TID 11008
|
1.21.6.8.2
|
Imaging Agent Administration Performed Phase UID
|
1.2.3.4.47110815.11
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.3
|
Imaging Agent Administration Phase Type
|
(130168, DCM, "Automatic Administration Phase")
|
TID 11008
CID 62
Since 1.21.6.3 is "Automated Administration"
|
1.21.6.8.4
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.6.8.4.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_CONTRAST_AGENT
|
TID 11003
Value of 1.14.1
|
1.21.6.8.4.2
|
Volume Administered
|
88
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
See 1.21.6.8.6 (Phase Volume) also
|
1.21.6.8.4.3
|
Starting Flow Rate of administration
|
1.5
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
|
1.21.6.8.4.4
|
Peak Flow Rate in Phase Activity
|
1.5
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
Since 1.21.6.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.4.5
|
Peak Pressure in Phase Activity
|
5
|
TID 11003
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.6.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.4.6
|
Initial Volume of Imaging Agent in Container
|
185
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.4.7
|
Residual Volume of Imaging Agent in Container
|
97
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.4.8
|
DateTime Started
|
20181012121900
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.6.8.4.9
|
Duration
|
58.6
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.6.8.5
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.6.8.5.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11003
Value of 1.15.1
|
1.21.6.8.5.2
|
Volume Administered
|
88
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
See 1.21.6.8.6 (Phase Volume) also
|
1.21.6.8.5.3
|
Starting Flow Rate of administration
|
1.5
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
|
1.21.6.8.5.4
|
Peak Flow Rate in Phase Activity
|
1.5
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
Since 1.21.6.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.5.5
|
Peak Pressure in Phase Activity
|
5
|
TID 11003
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.6.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.5.6
|
Initial Volume of Imaging Agent in Container
|
134
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Value results from 136ml - 2ml KVO within 1 min 10 sec until now
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.5.7
|
Residual Volume of Imaging Agent in Container
|
46
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.8.5.8
|
DateTime Started
|
20181012121900
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.6.8.5.9
|
Duration
|
58.6
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.6.8.6
|
Total Phase Volume Administered
|
176
|
TID 11008
UNITS = EV (ml, UCUM, "ml")
Sum of 1.21.6.8.4.2 and 1.21.6.8.5.2
|
1.21.6.8.7
|
DateTime Started
|
20181012121900
|
TID 11008
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration"
|
1.21.6.8.8
|
Duration
|
58.56
|
TID 11008
UNITS = EV (s, UCUM, "s")
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration")
|
1.21.6.9
|
Imaging Agent Administration Phase
|
|
TID 11008
|
1.21.6.9.1
|
Imaging Agent Administration Phase Identifier
|
DIAGNOSTIC_INJECTION_PHASE_2
|
TID 11008
|
1.21.6.9.2
|
Imaging Agent Administration Performed Phase UID
|
1.2.3.4.47110815.12
|
TID 11008
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.9.3
|
Imaging Agent Administration Phase Type
|
(130168, DCM, "Automatic Administration Phase")
|
TID 11008
CID 62
Since 1.21.6.3 is "Automated Administration"
|
1.21.6.9.4
|
Imaging Agent Administration Activity
|
|
TID 11003
|
1.21.6.9.4.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11003
Value of 1.15.1
|
1.21.6.9.4.2
|
Volume Administered
|
30
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.6.9.5
|
1.21.6.9.4.3
|
Starting Flow Rate of administration
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
|
1.21.6.9.4.4
|
Peak Flow Rate in Phase Activity
|
3
|
TID 11003
UNITS = EV (ml/s, UCUM "ml/s")
Since 1.21.6.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.9.4.5
|
Peak Pressure in Phase Activity
|
5
|
TID 11003
UNITS = EV (kPa, UCUM "kPa")
Since 1.21.6.3 is "Automated Administration" and "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.9.4.6
|
Initial Volume of Imaging Agent in Container
|
46
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.9.4.7
|
Residual Volume of Imaging Agent in Container
|
16
|
TID 11003
UNITS = EV (ml, UCUM, "ml")
Since "Root Concept Name Code Sequence" is "Performed Imaging Agent Administration"
|
1.21.6.9.4.8
|
DateTime Started
|
20181012121958.56
|
TID 11003
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.6.9.4.9
|
Duration
|
10
|
TID 11003
UNITS = EV (s, UCUM, "s")
Since root Concept Name is "Performed Imaging Agent Administration"
|
1.21.6.9.5
|
Total Phase Volume Administered
|
30
|
TID 11008
UNITS = EV (ml, UCUM, "ml")
Same value as 1.21.6.9.4.2
|
1.21.6.9.6
|
DateTime Started
|
20181012121958.56
|
TID 11008
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration"
|
1.21.6.9.7
|
Duration
|
10
|
TID 11008
UNITS = EV (s, UCUM, "s")
Since root Concept Name Code Sequence is "Performed Imaging Agent Administration")
|
1.21.6.10
|
Imaging Agent Administration Graph
|
|
TID 11023
|
1.21.6.10.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_CONTRAST_AGENT
|
TID 11023
|
1.21.6.10.2
|
Flow Rate vs time
|
|
TID 3990
Concept name is parameter $MeasurementGraph
|
1.21.6.10.2.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
TID 3990
Parameter $X-Concept
|
1.21.6.10.2.2
|
Y-Concept
|
(122094, DCM, "Rate of administration")
|
TID 3990
Parameter $Y-Concept
|
1.21.6.10.2.3
|
Flow Rate vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.11
|
TID 3990
|
1.21.6.10.3
|
Pressure vs time
|
|
TID 3990
Named by parameter $MeasurementGraph
|
1.21.6.10.3.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
TID 3990
Parameter $X-Concept
|
1.21.6.10.3.2
|
Y-Concept
|
(279046003, SCT, "Pressure")
|
TID 3990
Parameter $Y-Concept
|
1.21.6.10.3.3
|
Pressure vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.11
|
TID 3990
|
1.21.6.11
|
Imaging Agent Administration Graph
|
|
TID 11023
|
1.21.6.11.1
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11023
|
1.21.6.11.2
|
Flow Rate vs time
|
|
TID 3990
Concept name is parameter $MeasurementGraph
|
1.21.6.11.2.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
TID 3990
Parameter
|
1.21.6.11.2.2
|
Y-Concept
|
(122094, DCM, "Rate of administration")
|
TID 3990
Parameter $Y-Concept
|
1.21.6.11.2.3
|
Flow Rate vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.11
|
TID 3990
|
1.21.6.11.3
|
Pressure vs time
|
|
TID 3990
Named by parameter $MeasurementGraph
|
1.21.6.11.3.1
|
X-Concept
|
(130194, DCM, "Time after the start of injection")
|
TID 3990
Parameter $X-Concept
|
1.21.6.11.3.2
|
Y-Concept
|
(279046003, SCT, "Pressure")
|
TID 3990
Parameter $Y-Concept
|
1.21.6.11.3.3
|
Pressure vs time
|
IMAGE = 1.2.3.4.5.6.7.8.9.11
|
TID 3990
|
1.21.6.12
|
Number of Injector Heads
|
2
|
TID 11007
|
1.21.6.13
|
Programmable Device
|
(373066001, SCT, "Yes")
|
TID 11007
CID 231
|
1.22
|
Planned Imaging Agent Administration SOP Instance
|
1.2.3.4.47110815.13
|
TID 11020
SInce this administration was based on a Imaging Administration Plan
|
1.23
|
Imaging Agent Administration Completion Status
|
(255594003, SCT, "Complete")
|
TID 11020
CID 67
|
1.24
|
Imaging Agent Administration Adverse Events
|
|
TID 11021
|
1.24.1
|
Administration discontinued
|
(373067005, SCT, "No")
|
TID 11021
CID 231
|
1.24.2
|
Adverse Event
|
(415690000, SCT, "Sweating")
|
TID 11021
CID 60
|
1.24.2.1
|
Severity
|
(255604002, SCT, "Mild")
|
TID 11021
CID 3716
|
1.24.2.2
|
Relative Time
|
(307153007, SCT, "Before Procedure")
|
TID 11021
CID 61
|
1.24.2.3
|
Adverse Event Detection DateTime
|
20181012121500
|
TID 11021
|
1.24.2.4
|
Referenced Imaging Agent Administration Step UID
|
1.2.3.4.47110815.5
|
TID 11021
Same value as 1.21.4.2
|
1.24.2.5
|
Referenced Imaging Agent Administration Phase UID
|
1.2.3.4.47110815.6
|
TID 11021
Same value as
1.21.4.6.2
|
1.24.2.6
|
Comment
|
Patient was afraid of procedure
|
TID 11021
|
1.24.3
|
Adverse Event
|
(95384003, SCT, "Injection Site Extravasation")
|
TID 11021
CID 60
|
1.24.3.1
|
Relative Time
|
(303110006, SCT, "After Procedure")
|
TID 11021
CID 61
|
1.24.3.2
|
Adverse Event Detection DateTime
|
20181012122100
|
TID 11021
|
1.24.3.3
|
Estimated Extravasation Volume
|
2
|
TID 11021
UNITS = EV (ml, UCUM, "ml")
Since 1.17.3 is "Injection Site Extravasation"
|
1.24.3.4
|
Referenced Imaging Agent Administration Step UID
|
1.2.3.4.47110815.10
|
TID 11021
Same value as
1.21.6.2
|
1.24.3.5
|
Referenced Imaging Agent Administration Phase UID
|
1.2.3.4.47110815.12
|
TID 11021
Same value as
1.21.6.9.2
|
1.24.3.6
|
Comment
|
Detected extravasation when removing needle
|
TID 11021
|
1.25
|
Imaging Agent Administration Injector Events
|
|
TID 11022
|
1.25.1
|
Administration discontinued
|
(373067005, SCT, "No")
|
TID 11022
CID 231
|
1.25.2
|
Imaging Agent Administration Injector Event Type
|
(130161, DCM, "Keep vein open started")
|
TID 11022
CID 71
|
1.25.2.1
|
Injector Event Detection DateTime
|
20181012121628
|
TID 11022
|
1.25.2.2
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11022
|
1.25.3
|
Imaging Agent Administration Injector Event Type
|
(130162, DCM, "Keep vein open ended")
|
TID 11022
CID 71
|
1.25.3.1
|
Injector Event Detection DateTime
|
20181012121958
|
TID 11022
|
1.25.3.2
|
Referenced Imaging Agent Identifier
|
INJECTOR_FLUSH_AGENT
|
TID 11022
|
1.26
|
Total Keep Vein Open Volume Administered
|
3
|
TID 11020
UNITS = EV (ml, UCUM, "ml")
|