DICOM PS3.17 2026d - Explanatory Information

UUUU Radiation Dose Structured Reporting (Informative)

UUUU.1 Purpose of This Annex

This Annex describes the use of the RDSR Objects by different system types, highlighting the encoding similarities and differences across modalities, as well as the differences between traditional RDSR (defined with Root TID 10001 and TID 10011) and the Enhanced RDSR (defined with Root TID 10040).

Multiple systems contributing to patient care during a visit may expose the patient to irradiation during diagnostic and/or interventional procedures. Each of those systems may record the dose in a Radiation Dose Structured Report (RDSR) information object. Radiation information reporting systems may take advantage of this information and create additional reports or summaries for a visit or parts of a procedure performed, since information is completely available as structured content. The Radiation Dose Structured Report (RDSR) Objects described in this Annex record the radiation output of the equipment.

This Annex includes Projection X-Ray RDSR, CT RDSR, and Enhanced RDSR, and it excludes Radiopharmaceutical RDSR (PET, SPECT), Patient RDSR, and radiation for treatment (which is encoded in the family of Radiotherapy objects). The use of the Patient RDSR Objects is described in Annex GGGG Patient Radiation Dose Structured Report Document (Informative) .

DICOM PS3.17 2026d - Explanatory Information