DICOM PS3.16 2021e - Content Mapping Resource

TID 5240 Myocardial Strain Analysis

This Template is used to document strain measurements in the myocardium.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5240. Myocardial Strain Analysis

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (59776-5, LN, "Findings")

1

M

2

>

CONTAINS

CONTAINER

EV (125301, DCM, "Pre-coordinated Measurements")

1

M

3

>>

CONTAINS

INCLUDE

DTID 5301 “Pre-coordinated Echo Measurement”

1-n

U

$Measurement = DCID 12309 “Core Echo Strain Measurements”

$Preferred = DCID 12301 “Measurement Selection Reasons”

4

>

CONTAINS

CONTAINER

EV (125302, DCM, "Post-coordinated Measurements")

1

M

5

>>

CONTAINS

INCLUDE

DTID 5302 “Post-coordinated Echo Measurement”

1-n

U

$Preferred = DCID 12301 “Measurement Selection Reasons”

$Property = DCID 12311 “Echo Measured Strain Properties”


Content Item Descriptions

Row 4

Image Mode, Image View, Measurement Method, and Cardiac Cycle Point are all considered significant concepts for strain measurements.

DICOM PS3.16 2021e - Content Mapping Resource