This is a generic section heading Template for any of the anatomical headings. Measurements within a section heading appear as groups (by image mode, acquisition protocol, and/or protocol stage).
Table TID 5202. Parameters
Parameter Name
|
Parameter Usage
|
$SectionSubject
|
The subject modifier of the section heading container
|
$MeasType
|
The concept name of the measurement
|
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 5202. Echo Section
|
NL
|
Rel with Parent
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
1
|
|
|
CONTAINER
|
EV (59776-5, LN, "Findings")
|
1
|
M
|
|
|
2
|
>
|
HAS CONCEPT MOD
|
CODE
|
EV (363698007, SCT, "Finding Site")
|
1
|
M
|
|
$SectionSubject
|
3
|
>
|
CONTAINS
|
CONTAINER
|
DT (125007, DCM, "Measurement Group")
|
1-n
|
M
|
|
|
4
|
>>
|
HAS CONCEPT MOD
|
CODE
|
EV (399264008, SCT, "Image Mode")
|
1
|
U
|
|
BCID 12224 “Ultrasound Image Modes”
|
5
|
>>
|
HAS CONCEPT MOD
|
CODE
|
DT (125203, DCM, "Acquisition Protocol")
|
1
|
U
|
|
|
6
|
>>
|
HAS CONCEPT MOD
|
TEXT
|
DT (125203, DCM, "Acquisition Protocol")
|
1
|
U
|
|
|
7
|
>>
|
HAS ACQ CONTEXT
|
CODE
|
EV (18139-6, LN, "Stage")
|
1
|
U
|
|
BCID 12002 “Ultrasound Protocol Stage Types”
|
8
|
>>
|
CONTAINS
|
INCLUDE
|
DTID 5203 “Echo Measurement”
|
1-n
|
M
|
|
$Measurement = $MeasType
$Method = CID 12227 “Echocardiography Measurement Method”
|
Content Item Descriptions