DICOM PS3.16 2021e - Content Mapping Resource

TID 5104 Vascular Ultrasound Measurement Group

This Template is an anatomy specific container of measurements.

Table TID 5104. Parameters

Parameter Name

Parameter Usage

$AnatomyGroup

The concept name of the vascular anatomy


Type:

Extensible

Order:

Significant

Root:

No

Table TID 5104. Vascular Ultrasound Measurement Group

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

$AnatomyGroup

1

M

2

>

HAS CONCEPT MOD

CODE

EV (106233006, SCT, "Topographical Modifier")

1

U

DCID 12116 “Vessel Segment Modifiers”

3

>

HAS CONCEPT MOD

CODE

EV (125101, DCM, "Vessel Branch")

1-n

U

DCID 12117 “Vessel Branch Modifiers”

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

M

$Measurement = DCID 12119 “Vascular Ultrasound Property”

$Derivation = DCID 3627 “Measurement Type”

5

>>

HAS CONCEPT MOD

CODE

EV (272518008, SCT, "Cardiac Cycle Point")

1

U

DCID 12233 “Cardiac Phase”

6

>>

HAS CONCEPT MOD

CODE

EV (309602000, SCT, "Temporal period related to eating")

1

U

DT (24863003, SCT, "Post-prandial")

7

>>

HAS CONCEPT MOD

CODE

EV (125105, DCM, "Measurement Orientation")

1

U

DCID 12118 “Measurement Orientation”

8

>>

HAS PROPERTIES

NUM

EV (125106, DCM, "Doppler Angle")

1

U

UNIT = EV (deg, UCUM, "degrees")

9

>>

HAS PROPERTIES

NUM

EV (125107, DCM, "Sample Volume Depth")

1

U

UNIT = EV (cm, UCUM, "cm")


Content Item Descriptions

Row 1

Specifies the anatomic context of the observations in the group.

Row 2

Details the anatomical location, e.g., proximal, middle, or distal

Row 3

The particular vessel branch, such as the inferior, medial or lateral

Row 5

Cardiac phase (systolic, diastolic), especially for aorta measurements

Row 6

Eating phase, especially for mesenteric and celiac measurements

DICOM PS3.16 2021e - Content Mapping Resource