DICOM PS3.16 2024d - Content Mapping Resource

TID 5017 Follicle Count Group

This Template records a count of ovarian follicles of a given type and size range.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5017. Follicle Count Group

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130901, DCM, "Follicle Count Group")

1

M

2

>

HAS OBS CONTEXT

TEXT

EV (125010, DCM, "Identifier")

1

U

3

>

CONTAINS

CODE

EV (130908, DCM, "Follicle Type")

1

U

DCID 12010 “Follicle Type”

4

>

CONTAINS

NUM

EV (130902, DCM, "Diameter Range Minimum Value")

1

MC

IFF Row 6 is absent

UNITS = DT (mm, UCUM, "mm")

5

>

CONTAINS

NUM

EV (130903, DCM, "Diameter Range Maximum Value")

1

MC

IFF Row 4 is present

UNITS = DT (mm, UCUM, "mm")

6

>

CONTAINS

NUM

EV (130904, DCM, "Volume Range Minimum Value")

1

MC

IFF Row 4 is absent

UNITS = DT (mm3, UCUM, "mm3")

7

>

CONTAINS

NUM

EV (130905, DCM, "Volume Range Maximum Value")

1

MC

IFF Row 6 is present

UNITS = DT (mm3, UCUM, "mm3")

8

>

CONTAINS

NUM

EV (130906, DCM, "Number of Follicles in Range")

1

M

UNITS = EV (1, UCUM, "no units")


Content Item Descriptions

Row 2

Shall be unique among all groups of same laterality.

Rows 4-7

Specifies the size range (in terms of either follicle diameter or follicle volume) that defines this follicle count group.

Row 8

The number of follicles included in a given range count may have actually been measured or only estimated.

DICOM PS3.16 2024d - Content Mapping Resource