DICOM PS3.16 2021e - Content Mapping Resource

TID 5016 LWH Volume Group

This Template is a container for a group of measurements that assess the size of an anatomical structure using a volume derived from perpendicular diameters.

Table TID 5016. Parameters

Parameter Name

Parameter Usage

$GroupName

The name of the volume group that is an anatomical structure

$Volume

Concept name of volume measurement

$Length

Concept name of length measurement

$Width

Concept name of width measurement

$Height

Concept name of height measurement

$DistanceUnits

Units of linear (height, weight, length) distance measurement

$VolumeUnits

Units of volume measurement

$FunctionalCondition

Functional condition present during measurement

$Method

Value for Measurement Method

$FindingType

Type of the finding


Type:

Extensible

Order:

Significant

Root:

No

Table TID 5016. LWH Volume Group

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

$GroupName

1

M

1b

>

HAS OBS CONTEXT

TEXT

EV (125010, DCM, "Identifier")

1

U

1c

>

HAS CONCEPT MOD

CODE

EV (370129005, SCT, "Measurement Method")

1-n

U

$Method

2

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

MC

At least one of row 2, 3, 4, 5 shall be present

$Measurement = $Volume

$TargetSite = $GroupName

$Units = $VolumeUnits

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

MC

At least one of row 2, 3, 4, 5 shall be present

$Measurement = $Length

$TargetSite = $GroupName

$Derivation = DCID 3627 “Measurement Type”

$Units = $DistanceUnits

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

MC

At least one of row 2, 3, 4, 5 shall be present

$Measurement = $Width

$TargetSite = $GroupName

$Derivation = DCID 3627 “Measurement Type”

$Units = $DistanceUnits

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

MC

At least one of row 2, 3, 4, 5 shall be present

$Measurement = $Height

$TargetSite = $GroupName

$Derivation = DCID 3627 “Measurement Type”

$Units = $DistanceUnits

6

>

CONTAINS

CODE

EV (130324, DCM, "Functional condition present during acquisition")

1

U

$FunctionalCondition

7

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1

U

$FindingType


Content Item Descriptions

Row 1b

Shall be unique among all groups of same type of finding, site and laterality. E.g., "1", "2", etc. within all left ovarian follicle size measurements.

DICOM PS3.16 2021e - Content Mapping Resource