DICOM PS3.16 2022b - Content Mapping Resource

TID 5013 Follicles Section

This Template contains follicle metrics for left or right ovarian follicles.

Table TID 5013. Parameters

Parameter Name

Parameter Usage

$Laterality

Ovary laterality

$Number

The number of follicles


Type:

Extensible

Order:

Significant

Root:

No

Table TID 5013. Follicles Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

DT (24162005, SCT, "Ovarian Follicle")

3

>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

M

$Laterality

4

>

CONTAINS

NUM

$Number

1

U

5

>

CONTAINS

INCLUDE

DTID 5014 “Follicle Measurement Group”

1-n

U

6

>

CONTAINS

INCLUDE

DTID 5016 “LWH Volume Group”

1-n

U

$GroupName = EV (24162005, SCT, "Ovarian Follicle")

$Width = EV (103355008, SCT, "Width")

$Length = EV (410668003, SCT, "Length")

$Height = EV (121207, DCM, "Height")

$Volume = EV (121221, DCM, "Volume of ellipsoid")

$Method = DCID 7230 “Automation of Measurement”

$Method = DCID 12025 “OB-GYN Ultrasound Beam Path”


DICOM PS3.16 2022b - Content Mapping Resource