DICOM PS3.16 2021e - Content Mapping Resource

TID 5009 Fetal Biophysical Profile Section

This Template encodes scoring observations for fetal well-being evaluation as described by Manning, Antepartum Fetal Evaluation: Development of a Fetal Biophysical Profile Score, Am. J Obsetet Gynecol, 1980;136:787.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 5009. Fetal Biophysical Profile Section

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (125006, DCM, "Biophysical Profile")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1008 “Subject Context, Fetus”

1

MC

IF this Template is invoked more than once to describe more than one fetus

3

>

CONTAINS

NUM

EV (11631-9, LN, "Gross Body Movement")

1

MC

At least one of row 3-7 shall be present

UNITS = DT ({0:2}, UCUM, "range 0:2")

4

>

CONTAINS

NUM

EV (11632-7, LN, "Fetal Breathing")

1

MC

At least one of row 3-7 shall be present

UNITS = DT ({0:2}, UCUM, "range 0:2")

5

>

CONTAINS

NUM

EV (11635-0, LN, "Fetal Tone")

1

MC

At least one of row 3-7 shall be present

UNITS = DT ({0:2}, UCUM, "range 0:2")

6

>

CONTAINS

NUM

EV (11633-5, LN, "Fetal Heart Reactivity")

1

MC

At least one of row 3-7 shall be present

UNITS = DT ({0:2}, UCUM, "range 0:2")

7

>

CONTAINS

NUM

EV (11630-1, LN, "Amniotic Fluid Volume")

1

MC

At least one of row 3-7 shall be present

UNITS = DT ({0:2}, UCUM, "range 0:2")

8

>

CONTAINS

NUM

DT (11634-3, LN, "Biophysical Profile Sum Score")

1

U


Content Item Descriptions

Row 3-7

The numeric profile score of range 0-2

Row 8

The sum of rows 3-7. The range is from 0 to the maximum possible score according the items scored in rows 3-7.

DICOM PS3.16 2021e - Content Mapping Resource