DICOM PS3.16 2021e - Content Mapping Resource

TID 4306 PI-RADS® Localized Abnormality Assessment

This template stores PI-RADS® assessment for all analyzed modalities, with the references to the annotations in the individual modalities. Assessment is not specific to lesions, since it can also be used to assess additional findings, including definitely benign findings.

When lesions are identified, only one lesion shall be designated as "Index lesion".

Type:

Extensible

Order:

Significant

Root:

No

Table TID 4306. PI-RADS® Localized Abnormality Assessment

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130559, DCM, "PI-RADS Localized Abnormality Assessment")

1

M

2

>

CONTAINS

CODE

EV (C110961, NCIt, "Index lesion")

1

U

BCID 231 “Yes-No Only”

3

>

CONTAINS

CONTAINER

EV (130561, DCM, "PI-RADS T2WI Lesion Assessment")

1

M

4

>>

CONTAINS

CODE

EV (RID50301, RADLEX, "PI-RADS T2WI PZ Lesion Assessment Category")

1

MC

XOR Row 5

DCID 6329 “PI-RADS® v2 T2WI PZ Lesion Assessment Category”

5

>>

CONTAINS

CODE

EV (RID50307, RADLEX, "PI-RADS T2WI TZ Lesion Assessment Category")

1

MC

XOR Row 4

DCID 6330 “PI-RADS® v2 T2WI TZ Lesion Assessment Category”

6

>>

CONTAINS

CODE

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

DCID 6335 “mpMRI Assessment Values”

7

>>

CONTAINS

TEXT

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

8

>

CONTAINS

CONTAINER

EV (130562, DCM, "PI-RADS DWI Lesion Assessment")

1

M

9

>>

CONTAINS

CODE

EV (RID50313, RADLEX, "PI-RADS DWI Lesion Assessment Category")

1

U

DCID 6331 “PI-RADS® v2 DWI Lesion Assessment Category”

10

>>

CONTAINS

CODE

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

DCID 6335 “mpMRI Assessment Values”

11

>>

CONTAINS

TEXT

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

12

>

CONTAINS

CONTAINER

EV (130563, DCM, "PI-RADS DCE Lesion Assessment")

1

M

13

>>

CONTAINS

CODE

EV (RID50319, RADLEX, "PI-RADS DCE Lesion Assessment Category")

1

M

DCID 6332 “PI-RADS® v2 DCE Lesion Assessment Category”

14

>>

CONTAINS

CODE

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

DCID 6335 “mpMRI Assessment Values”

15

>>

CONTAINS

TEXT

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

16

>

CONTAINS

CODE

EV (RID50295, RADLEX, "PI-RADS Lesion Assessment Category")

1-n

U

BCID 6328 “PI-RADS® v2 Lesion Assessment Category”

17

>

CONTAINS

CODE

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U

BCID 6335 “mpMRI Assessment Values”

18

>

CONTAINS

TEXT

DCID 6334 “mpMRI Assessment Types from PI-RADS®”

1-n

U


Content Item Descriptions

Rows 7, 11, 15, 18

The intent of those content items is to allow for communication of information that cannot be coded in the preceding CODE content items, and not to replicate the CODE items.

DICOM PS3.16 2021e - Content Mapping Resource