DICOM PS3.16 2021e - Content Mapping Resource

TID 4305 Extra-prostatic Finding

This template can be used to report abnormal lymph nodes and skeletal metastases as part of prostate MRI report. At least bi-dimensional measurement of suspicious lymph nodes is recommended. Organization of the sub-parts of the template corresponding to lymph nodes and skeletal metastases is similar, with the exception that lymph nodes section is parameterized with CID 7600 “Lymph Node Anatomic Sites” for $TargetSite.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 4305. Extra-prostatic Finding

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130559, DCM, "Extra-prostatic Finding")

1

M

2

>

HAS OBS CONTEXT

TEXT

DT (112039, DCM, "Tracking Identifier")

1

U

3

>

HAS OBS CONTEXT

UIDREF

EV (112040, DCM, "Tracking Unique Identifier")

1

U

4

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1

U

BCID 6347 “Prostate MRI Extra-prostatic Findings”

5

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding site")

1-n

U

BCID 7600 “Lymph Node Anatomic Sites”

BCID 6348 “Prostate MRI Assessment of Extra-prostatic Anatomic Sites”

6

>>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

U

DCID 244 “Laterality”

7

>>

HAS CONCEPT MOD

CODE

EV (106233006, SCT, "Topographical modifier")

1

U

8

>

CONTAINS

INCLUDE

DTID 3909 “Best Illustration of Findings”

1-n

U

9

>

CONTAINS

INCLUDE

DTID 1410 “Planar ROI Measurements and Qualitative Evaluations”

1-n

MC

IF Rows 10, 11 are absent

$Measurement = BCID 218 “Quantitative Image Features”

$Units = BCID 7181 “Abstract Multi-dimensional Image Model Component Units”

$Derivation = BCID 7464 “General Region of Interest Measurement Modifiers”

$Method = BCID 6147 “Response Criteria”

$FindingType = Value of Row 4

$TargetSite = Value of Row 5

$TrackingID = Value of Row 2

$TrackingUID = Value of Row 3

$QualType = BCID 6333 “mpMRI Assessment Types”

$QualValue = BCID 6335 “mpMRI Assessment Values”

10

>

CONTAINS

INCLUDE

DTID 1411 “Volumetric ROI Measurements and Qualitative Evaluations”

1-n

MC

IF Rows 9, 11 are absent

$Measurement = BCID 218 “Quantitative Image Features”

$Units = BCID 7181 “Abstract Multi-dimensional Image Model Component Units”

$Derivation = BCID 7464 “General Region of Interest Measurement Modifiers”

$Method = BCID 6147 “Response Criteria”

$FindingType = Value of Row 4

$TargetSite = Value of Row 5

$TrackingID = Value of Row 2

$TrackingUID = Value of Row 3

$QualType = BCID 6333 “mpMRI Assessment Types”

$QualValue = BCID 6335 “mpMRI Assessment Values”

11

>

CONTAINS

INCLUDE

DTID 1501 “Measurement and Qualitative Evaluation Group”

1-n

MC

IF Rows 9, 10 are absent

$Measurement = BCID 218 “Quantitative Image Features”

$Units = BCID 7181 “Abstract Multi-dimensional Image Model Component Units”

$Derivation = BCID 7464 “General Region of Interest Measurement Modifiers”

$Method = BCID 6147 “Response Criteria”

$FindingType = Value of Row 4

$TargetSite = Value of Row 5

$TrackingID = Value of Row 2

$TrackingUID = Value of Row 3

$QualType = BCID 6333 “mpMRI Assessment Types”

$QualValue = BCID 6335 “mpMRI Assessment Values”

$ImagePurpose = EV (260753009, SCT, "Source")

12

>

CONTAINS

CONTAINER

EV (C0034375, UMLS, "Qualitative Evaluations")

1

MC

IF row 6 and 10 are absent

13

>>

CONTAINS

CODE

DCID 6333 “mpMRI Assessment Types”

1-n

U

BCID 6335 “mpMRI Assessment Values”

14

>>

CONTAINS

TEXT

DCID 6333 “mpMRI Assessment Types”

1-n

U


DICOM PS3.16 2021e - Content Mapping Resource