DICOM PS3.16 2021e - Content Mapping Resource

TID 4020 CAD Image Library Entry

Each instance of the Image Library Entry Template contains the Image SOP Class and Instance UIDs, and selected attributes for an image that facilitate spatial analysis without having to retrieve the entire set of referenced images. If values for the attributes are not present in the Image SOP Instance, then as many of the attributes as possible should be derived.

Table TID 4020. Parameters

Parameter Name

Parameter Usage

$ImageLaterality

Coded term or Context Group for Image Laterality

$ImageView

Coded term or Context Group for Image View

$ImageViewMod

Coded term or Context Group for Image View Modifier


Type:

Extensible

Order:

Significant

Root:

No

Table TID 4020. CAD Image Library Entry

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

IMAGE

1

M

2

>

HAS ACQ CONTEXT

CODE

EV (111027, DCM, "Image Laterality")

1

MC

Shall be present if (0020,0062) is in the Image Instance

$ImageLaterality

3

>

HAS ACQ CONTEXT

CODE

EV (111031, DCM, "Image View")

1

MC

Shall be present if (0054,0220) is in the Image Instance

$ImageView

4

>>

HAS CONCEPT MOD

CODE

EV (111032, DCM, "Image View Modifier")

1-n

MC

Shall be present if (0054,0222) is in the Image Instance

$ImageViewMod

5

>

HAS ACQ CONTEXT

TEXT

EV (111044, DCM, "Patient Orientation Row")

1

MC

Shall be present if (0020,0020) is in the Image Instance

6

>

HAS ACQ CONTEXT

TEXT

EV (111043, DCM, "Patient Orientation Column")

1

MC

Shall be present if (0020,0020) is in the Image Instance

7

>

HAS ACQ CONTEXT

DATE

EV (111060, DCM, "Study Date")

1

MC

Shall be present if (0008,0020) is in the Image Instance

8

>

HAS ACQ CONTEXT

TIME

EV (111061, DCM, "Study Time")

1

MC

Shall be present if (0008,0030) is in the Image Instance

9

>

HAS ACQ CONTEXT

DATE

EV (111018, DCM, "Content Date")

1

MC

Shall be present if (0008,0023) is in the Image Instance

10

>

HAS ACQ CONTEXT

TIME

EV (111019, DCM, "Content Time")

1

MC

Shall be present if (0008,0033) is in the Image Instance

11

>

HAS ACQ CONTEXT

NUM

EV (111026, DCM, "Horizontal Pixel Spacing")

1

MC

Shall be present if (0018,1164) or (0028,0030) is in the Image Instance

UNITS = EV (um, UCUM, "micrometer")

UNITS = EV (mm, UCUM, "millimeter")

12

>

HAS ACQ CONTEXT

NUM

EV (111066, DCM, "Vertical Pixel Spacing")

1

MC

Shall be present if (0018,1164) or (0028,0030) is in the Image Instance

UNITS = EV (um, UCUM, "micrometer")

UNITS = EV (mm, UCUM, "millimeter")

13

>

HAS ACQ CONTEXT

NUM

EV (112011, DCM, "Positioner Primary Angle")

1

UC

May be present if (0018,1510) is in the Image Instance

UNITS = EV (deg, UCUM, "deg")

14

>

HAS ACQ CONTEXT

NUM

EV (112012, DCM, "Positioner Secondary Angle")

1

UC

May be present if (0018,1511) is in the Image Instance

UNITS = EV (deg, UCUM, "deg")

15

>

HAS ACQ CONTEXT

NUM

EV (112226, DCM, "Spacing between slices")

1

UC

May be computed from the Image Position (Patient) (0020,0032) projected onto the normal to the Image Orientation (Patient) (0020,0037) if present; may or may not be the same as the Spacing Between Slices (0018,0088) if present.

UNITS = EV (mm, UCUM, "millimeter")

16

>

HAS ACQ CONTEXT

NUM

EV (112225, DCM, "Slice Thickness")

1

UC

May be present if Slice Thickness (0018,0050) is in the Image Instance.

UNITS = EV (mm, UCUM, "millimeter")

17

>

HAS ACQ CONTEXT

UIDREF

EV (112227, DCM, "Frame of Reference UID")

1

UC

May be present if Frame of Reference UID (0020,0052) is in the Image Instance.

18

>

HAS ACQ CONTEXT

NUM

EV (110901, DCM, "Image Position (Patient) X")

1

UC

May be present if Image Position (Patient) (0020,0032) is in the Image Instance, and is the first value of Image Position (Patient) (0020,0032) for the referenced image or frame.

UNITS = EV (mm, UCUM, "millimeter")

19

>

HAS ACQ CONTEXT

NUM

EV (110902, DCM, "Image Position (Patient) Y")

1

MC

Shall be present if Row 18 is present, and is the second value of Image Position (Patient) (0020,0032) in the Image Instance for the referenced image or frame.

UNITS = EV (mm, UCUM, "millimeter")

20

>

HAS ACQ CONTEXT

NUM

EV (110903, DCM, "Image Position (Patient) Z")

1

MC

Shall be present if Row 18 is present, and is the second value of Image Position (Patient) (0020,0032) in the Image Instance for the referenced image or frame.

UNITS = EV (mm, UCUM, "millimeter")

21

>

HAS ACQ CONTEXT

NUM

EV (110904, DCM, "Image Orientation (Patient) Row X")

1

UC

May be present if Image Position (Patient) (0020,0037) is in the Image Instance, and is the first value of Image Orientation (Patient) (0020,0037) for the referenced image or frame.

UNITS = EV ({-1:1}, UCUM, "{-1:1}")

22

>

HAS ACQ CONTEXT

NUM

EV (110905, DCM, "Image Orientation (Patient) Row Y")

1

MC

Shall be present if Row 21 is present, and is the second value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame.

UNITS = EV ({-1:1}, UCUM, "{-1:1}")

23

>

HAS ACQ CONTEXT

NUM

EV (110906, DCM, "Image Orientation (Patient) Row Z")

1

MC

Shall be present if Row 21 is present, and is the third value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame.

UNITS = EV ({-1:1}, UCUM, "{-1:1}")

24

>

HAS ACQ CONTEXT

NUM

EV (110907, DCM, "Image Orientation (Patient) Column X")

1

MC

Shall be present if Row 21 is present, and is the fourth value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame.

UNITS = EV ({-1:1}, UCUM, "{-1:1}")

25

>

HAS ACQ CONTEXT

NUM

EV (110908, DCM, "Image Orientation (Patient) Column Y")

1

MC

Shall be present if Row 21 is present, and is the fifth value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame.

UNITS = EV ({-1:1}, UCUM, "{-1:1}")

26

>

HAS ACQ CONTEXT

NUM

EV (110909, DCM, "Image Orientation (Patient) Column Z")

1

MC

Shall be present if Row 21 is present, and is the sixth value of Image Orientation (Patient) (0020,0037) in the Image Instance for the referenced image or frame.

UNITS = EV ({-1:1}, UCUM, "{-1:1}")

27

>

HAS ACQ CONTEXT

NUM

EV (110910, DCM, "Pixel Data Rows")

1

UC

May be present if Rows (0028,0010) is in the Image Instance.

UNITS = EV ({pixels}, UCUM, "pixels")

28

>

HAS ACQ CONTEXT

NUM

EV (110911, DCM, "Pixel Data Columns")

1

MC

Shall be present if Row 27 is present, and is the value of Columns (0028,0011) in the Image Instance.

UNITS = EV ({pixels}, UCUM, "pixels")


Content Item Descriptions

Patient Orientation Row

First (row) and second (column) components of Patient Orientation (0020,0020) in the Image IOD. See Section C.7.6.1.1.1 in PS3.3 .

Patient Orientation Column

Horizontal Imager Pixel Spacing

The second component of Imager Pixel Spacing (0018,1164) in the Image IOD. See Section C.8.11.4 in PS3.3 .

Vertical Imager Pixel Spacing

The first component of Imager Pixel Spacing (0018,1164) in the Image IOD. See Section C.8.11.4 in PS3.3 .

DICOM PS3.16 2021e - Content Mapping Resource