DICOM PS3.16 2022b - Content Mapping Resource

TID 4016 CAD Analyses Performed

This Template gathers two lists of analysis algorithms attempted, grouped by success/failure status. Algorithms not attempted are not mentioned in this sub-tree of the Document Root (e.g., TID 4000 “Mammography CAD Document Root”). This information forms the basis for understanding why a CAD report may produce no (or fewer than anticipated) analysis results.

The sub-tree formed by this Template is illustrated in Figure E.1-2 “Summary of Detections and Analyses Levels of Mammography CAD SR Content Tree” in PS3.17 .

Table TID 4016. Parameters

Parameter Name

Parameter Usage

$AnalysisCode

Coded term or Context Group for Analysis Performed


Type:

Non-Extensible

Order:

Significant

Root:

No

Table TID 4016. CAD Analyses Performed

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (111062, DCM, "Successful Analyses")

1

MC

Shall be present only if value of parent is (111222, DCM, "Succeeded") or (111223, DCM, "Partially Succeeded")

2

>

CONTAINS

INCLUDE

DTID 4018 “CAD Analysis Performed”

1-n

M

$AnalysisCode = $AnalysisCode

3

CONTAINER

EV (111024, DCM, "Failed Analyses")

1

MC

Shall be present only if value of parent is (111224, DCM, "Failed") or (111223, DCM, "Partially Succeeded")

4

>

CONTAINS

INCLUDE

DTID 4018 “CAD Analysis Performed”

1-n

M

$AnalysisCode = $AnalysisCode