DICOM PS3.16 2026a - Content Mapping Resource

TID 3714 ECG Lead Measurements

The ECG Lead Measurements Template provides a structure for measurements calculated on individual ECG leads.

As an Extensible Template, applications may include any ECG per lead measurements, such as integrals over time of various ECG wave voltages. The recommended vocabulary for such concepts is ISO/IEEE 11073-10102.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3714. ECG Lead Measurements

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122159, DCM, "ECG Lead Measurements")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (122148, DCM, "Lead ID")

1

M

BCID 3001 “ECG Lead”

3

>

HAS OBS CONTEXT

INCLUDE

DTID 3715 “ECG Measurement Source”

1

U

4

>

CONTAINS

NUM

DCID 3687 “Electrophysiology Waveform Duration”

1-n

U

UNITS = EV (ms, UCUM, "ms")

5

>

CONTAINS

NUM

DCID 3688 “Electrophysiology Waveform Voltage”

1-n

U

UNITS = EV (mV, UCUM, "mV")

6

>

CONTAINS

CODE

EV (365416000, SCT, "ST Segment Finding")

1

U

DCID 3233 “ST Segment Morphology”

7

>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

U

DCID 3680 “ECG Lead Noise Description”


DICOM PS3.16 2026a - Content Mapping Resource