1
|
|
|
CONTAINER
|
EV (122158, DCM, "ECG Global Measurements")
|
1
|
M
|
|
|
2
|
>
|
HAS OBS CONTEXT
|
INCLUDE
|
DTID 3715 “ECG Measurement Source”
|
1
|
U
|
|
|
3
|
>
|
CONTAINS
|
NUM
|
DT (2:16020, MDC, "Atrial Heart Rate")
|
1
|
U
|
|
UNITS = EV ({H.B.}/min, UCUM, "BPM")
|
4
|
>
|
CONTAINS
|
NUM
|
DT (2:16016, MDC, "Ventricular Heart Rate")
|
1
|
M
|
|
UNITS = EV ({H.B.}/min, UCUM, "BPM")
|
5
|
>
|
CONTAINS
|
NUM
|
DT (2:16160, MDC, "QT interval global")
|
1
|
M
|
|
UNITS = EV (ms, UCUM, "ms")
|
6
|
>
|
CONTAINS
|
NUM
|
DT (2:15876, MDC, "QTc interval global")
|
1
|
U
|
|
UNITS = EV (ms, UCUM, "ms")
|
7
|
>>
|
HAS PROPERTIES
|
CODE
|
DT (111001, DCM, "Algorithm Name")
|
1
|
U
|
|
DCID 3678 “QT Correction Algorithms”
|
8
|
>
|
CONTAINS
|
NUM
|
DT (2:15872, MDC, "PR interval global")
|
1
|
M
|
|
UNITS = EV (ms, UCUM, "ms")
|
9
|
>
|
CONTAINS
|
NUM
|
DT (2:16156, MDC, "QRS duration global")
|
1
|
M
|
|
UNITS = EV (ms, UCUM, "ms")
|
10
|
>
|
CONTAINS
|
NUM
|
DT (2:16168, MDC, "RR interval global")
|
1
|
M
|
|
UNITS = EV (ms, UCUM, "ms")
|
11
|
>
|
CONTAINS
|
NUM
|
DCID 3689 “ECG Global Waveform Durations”
|
1-n
|
U
|
|
UNITS = EV (ms, UCUM, "ms")
|
12
|
>
|
CONTAINS
|
NUM
|
DCID 3229 “ECG Axis Measurements”
|
1-n
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
13
|
>
|
CONTAINS
|
NUM
|
DT (2:16032, MDC, "Count of all beats")
|
1
|
U
|
|
UNITS = EV ({beats}, UCUM, "beats")
|
14
|
>
|
CONTAINS
|
NUM
|
DT (122707, DCM, "Number of Ectopic Beats")
|
1
|
U
|
|
UNITS = EV ({beats}, UCUM, "beats")
|
15
|
>>
|
HAS PROPERTIES
|
CODE
|
EV (116676008, SCT, "Associated Morphology")
|
1-n
|
U
|
|
BCID 3234 “Ectopic Beat Morphology”
|