DICOM PS3.16 2021e - Content Mapping Resource

TID 3560 Derived Hemodynamic Measurements

The Derived Hemodynamic Measurements Template consists of a CONTAINER containing measurements derived from one or more other measurements. These measurements are associated with a particular procedure phase, but not necessarily with a particular anatomic location.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3560. Derived Hemodynamic Measurements

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (122126, DCM, "Derived Hemodynamic Measurements")

1

U

2

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3614 “Valve Areas, Non-mitral”

$Units = EV (cm2, UCUM, "cm2")

$Equation = DT (122262, DCM, "Area = Flow / 44.5 * sqrt(Gradient[mmHg])")

3

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (251012002, SCT, "Mitral Valve Area")

$Units = EV (cm2, UCUM, "cm2")

$Equation = DT (122263, DCM, "MVA = Flow / 38.0 * sqrt(Gradient[mmHg])")

4

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3615 “Valve Areas”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (8277-6, LN, "Body Surface Area")

$Units = DT (cm2/m2, UCUM, "cm2/m2")

5

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3616 “Hemodynamic Period Measurements”

$Units = DT (s/min, UCUM, "s/min")

6

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3617 “Valve Flows”

$Units = DT (ml/s, UCUM, "ml/s")

7

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (54993008, SCT, "Cardiac Index")

$Units = DT (l/min/m2, UCUM, "l/min/m2")

8

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3529 “Arterial / Venous Content”

$Units = DT (ml/dl, UCUM, "ml/dl")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3618 “Hemodynamic Flows”

$Units = DT (l/min, UCUM, "l/min")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8736-1, LN, "FICK Cardiac Output")

$Units = DT (l/min, UCUM, "l/min")

11

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8750-2, LN, "FICK Cardiac Index")

$Units = DT (l/min/m2, UCUM, "l/min/m2")

12

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122229, DCM, "Arteriovenous difference")

$Units = DT (ml/dl, UCUM, "ml/dl")

13

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = BCID 3620 “Hemodynamic Ratios”

$Units = DT ({ratio}, UCUM, "ratio")

14

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122237, DCM, "Corrected Sinus Node Recovery Time")

$Units = DT (ms, UCUM, "ms")

15

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8861-7, LN, "Left Ventricular Stroke Work")

$Units = DT (gf.m, UCUM, "gf.m")

16

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8862-5, LN, "Right Ventricular Stroke Work")

$Units = DT (gf.m, UCUM, "gf.m")

17

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8863-3, LN, "Left Ventricular Stroke Work Index")

$Units = DT (gf.m/m2, UCUM, "gf.m/m2")

18

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (8864-1, LN, "Right Ventricular Stroke Work Index")

$Units = DT (gf.m/m2, UCUM, "gf.m/m2")

19

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122238, DCM, "Max volume normalized to 50mmHg pulse pressure")

$Units = DT (ml, UCUM, "ml")

20

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122239, DCM, "Oxygen Consumption")

$Units = DT (ml/min, UCUM, "ml/min")

$Equation = BCID 3664 “Oxygen Consumption Equations and Tables”

21

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (19217-9, LN, "Oxygen partial pressure at 50% saturation (P50)")

$Units = DCID 3500 “Pressure Units”

$Equation = BCID 3666 “P50 Equations”

22

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (11556-8, LN, "Blood Oxygen partial pressure")

$Units = DCID 3500 “Pressure Units”

23

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3618 “Hemodynamic Flows”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (8277-6, LN, "Body Surface Area")

$Units = DT (l/min/m2, UCUM, "l/min/m2")

24

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3619 “Hemodynamic Resistance Measurements”

$Units = DCID 3502 “Hemodynamic Resistance Units”

25

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3619 “Hemodynamic Resistance Measurements”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (8277-6, LN, "Body Surface Area")

$Units = DCID 3503 “Indexed Hemodynamic Resistance Units”

26

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122227, DCM, "Left to Right Flow")

$Units = DT (l/min, UCUM, "l/min")

27

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122228, DCM, "Right to Left Flow")

$Units = DT (l/min, UCUM, "l/min")

28

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (90096001, SCT, "Stroke Volume")

$Units = DT (ml, UCUM, "ml")

29

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (90096001, SCT, "Stroke Volume")

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (8277-6, LN, "Body Surface Area")

$Units = DT (ml/m2, UCUM, "ml/m2")

30

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (268384009, SCT, "Total blood volume")

$Units = DT (l, UCUM, "l")

31

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = DCID 3667 “Framingham Scores”

$Units = DT (%, UCUM, "%")

$Equation = DCID 3668 “Framingham Tables”


DICOM PS3.16 2021e - Content Mapping Resource