DICOM PS3.16 2026d - Content Mapping Resource

TID 3313 Stress ECG Summary

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3313. Stress ECG Summary

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

EV (271921002, SCT, "ECG Finding")

1

U

Device Generated Test Summary

2

>

HAS OBS CONTEXT

CODE

EV (121005, DCM, "Observer Type")

1

M

EV (121007, DCM, "Device")

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1004 “Device Observer Identifying Attributes”

1

M

4

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (164931005, SCT, "ST Elevation")

$Units = DT (mV, UCUM, "mV")

$Derivation = EV (56851009, SCT, "Maximum")

$TargetSite = DCID 3001 “ECG Lead”

5

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (429622005, SCT, "ST Depression")

$Units = DT (mV, UCUM, "mV")

$Derivation = EV (56851009, SCT, "Maximum")

$TargetSite = DCID 3001 “ECG Lead”

6

INCLUDE

DTID 300 “Measurement”

1-n

U

$Measurement = EV (428550008, SCT, "T wave alternans")

$Units = DT (uV, UCUM, "uV")

$Derivation = EV (56851009, SCT, "Maximum")

$TargetSite = DCID 3001 “ECG Lead”

7

CODE

EV (365416000, SCT, "ST Segment Finding")

1

U

BCID 3231 “ST Segment Finding”

8

>

HAS PROPERTIES

CODE

EV (363698007, SCT, "Finding Site")

1-n

U

BCID 3232 “ST Segment Location”

9

>

HAS PROPERTIES

CODE

EV (116676008, SCT, "Associated Morphology")

1

U

BCID 3233 “ST Segment Morphology”

10

NUM

DT (122707, DCM, "Number of Ectopic Beats")

1

U

UNITS = EV ({beats}, UCUM, "beats")

11

>

HAS PROPERTIES

CODE

EV (116676008, SCT, "Associated Morphology")

1-n

U

BCID 3234 “Ectopic Beat Morphology”

12

CODE

DT (8884-9, LN, "Cardiac Rhythm")

1-2

U

BCID 3415 “Cardiac Rhythm”

13

>

HAS CONCEPT MOD

CODE

EV (109054, DCM, "Patient State")

1

M

DCID 3102 “Rest/Stress State”

14

CODE

EV (271921002, SCT, "ECG Finding")

1-n

U

BCID 3230 “ECG Finding”

15

>

HAS CONCEPT MOD

CODE

EV (109054, DCM, "Patient State")

1

U

BCID 3262 “ECG Patient State Value”


Content Item Descriptions

Rows 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14

Each observation (measurement or finding) may have a specific Observation DateTime (0040,A032) Content Item Attribute to indicate the time in the procedure at which the observation was made (e.g., time of maximum heart rate, or time of occurrence of an arrhythmia).

Rows 12, 13

This Concept and the associated Concept Modifier may be instantiated twice, once for resting state measurements, once for stress.