DICOM PS3.16 2022b - Content Mapping Resource

TID 3217 Sub-segmental Data

The Sub-segmental Data Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for a sub-segment in a total analyzed segment.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3217. Sub-Segmental Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

DCID 3604 “Arterial Lesion Locations”

3

>>

HAS CONCEPT MOD

CODE

EV (106233006, SCT, "Topographical modifier")

1

U

DCID 3019 “Cardiovascular Anatomic Location Modifiers”

4

>

CONTAINS

CODE

EV (122554, DCM, "Segmentation Method")

1

M

DCID 3456 “Sub-segment Methods”

5

>

CONTAINS

INCLUDE

DTID 3219 “Segment Values”

1

U

6

>

CONTAINS

INCLUDE

DTID 3218 “Position in Arterial Segment”

1

M

7

>

CONTAINS

IMAGE

1

U


Content Item Descriptions

Row 7

Secondary Capture image with obstruction analysis contour. No purpose of reference is specified.

DICOM PS3.16 2022b - Content Mapping Resource