DICOM PS3.16 2026a - Content Mapping Resource

TID 2125 Ganglion Cell Key Measurements

Type:

Extensible

Order:

Non-Significant

Root:

Yes

Table TID 2125. Ganglion Cell Key Measurements


Content Item Descriptions

Row 5

For example, TID 2120 may be invoked once for each eye, or multiple times for a trend series for a single eye.

Methods to measure the Ganglion Cell thickness vary widely. This template requires the SOP Instance creator application to specify which other cell layers, if any, are measured with the Ganglion Cell Layer proper (using a concept from CID 4286 “Ganglion Cell Measurement Extent”).

Applications also use various methods to identify different sectors of the retina when measuring the Ganglion Cell thickness. The sector definition used is specified by a concept from CID 4289 “Ganglion Cell Sector Methods”. Measurements matching the sector names defined by the method may be selected from CID 4288 “Ganglion Cell Sector Measurements”. If the sector method uses a rectangular or elliptical ROI, the sector measurements should include both the ROI width and height.

If optional measurements using (131298, DCM, "Average ganglion cell thickness in posterior pole grid") are present, the SOP Class of the instance must support the TABLE Content Item Value Type (e.g., the Extensible SR Storage SOP Class).

Receiving applications are cautioned about comparing or aggregating measurements from measurement groups with different measurement extents or sector methods.

DICOM PS3.16 2026a - Content Mapping Resource