DICOM PS3.16 2021e - Content Mapping Resource

TID 11022 Imaging Agent Administration Injector Events

This template describes events occurring during the administration that are detected by an automated power injector.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 11022. Imaging Agent Administration Injector Events

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130233, DCM, "Imaging Agent Administration Injector Events")

1

M

2

>

CONTAINS

CODE

EV (130220, DCM, "Administration discontinued")

1

U

DCID 231 “Yes-No Only”

3

>

CONTAINS

CODE

EV (130234, DCM, "Imaging Agent Administration Injector Event Type")

1-n

M

DCID 71 “Imaging Agent Administration Injector Event Type”

4

>>

HAS PROPERTIES

DATETIME

EV (130235, DCM, "Injector Event Detection DateTime")

1

M

5

>>

HAS OBS CONTEXT

UIDREF

EV (130216, DCM, "Referenced Imaging Agent Administration Step UID")

1

U

6

>>

HAS OBS CONTEXT

UIDREF

EV (130262, DCM, "Referenced Imaging Agent Administration Phase UID")

1

U

7

>>

HAS OBS CONTEXT

TEXT

EV (130255, DCM, "Referenced Imaging Agent Identifier")

1

UC

IFF there is more than one syringe and the syringe/pump can be identified.

Shall be as defined in (130254, DCM, "Imaging Agent Identifier") Row 2 of TID 11002 “Imaging Agent Information”


Content Item Descriptions

Row 4

Date and time of occurrence of the injector event.

Row 5

UID of the performed step (as recorded in Row 3 of TID 11007) in which the injector event occurred.

Row 6

UID of the performed phase (as recorded in Row 3 of TID 11008) in which the injector event occurred.

Row 7

The Imaging Agent being administered when the event was detected.

DICOM PS3.16 2021e - Content Mapping Resource