DICOM PS3.16 2022b - Content Mapping Resource |
---|
This template describes the positions of objects and locations that are defined in the X-Ray source reference coordinate system (e.g., output measurement point, filter position, etc.).
Table TID 10051. Beam Position
The DateTime of the beginning of the time period over which the content items are applicable. |
|
The DateTime of the end of the time period over which the content items are applicable. |
|
Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template. |
|
The Output Measurement Point Position describes the position where the measure or machine estimate of radiation output occurs. It is described in the X-Ray source reference coordinate system. It may be the same position as the reference point position (the reference point may be defined in TID 10041). |
|
If a reference point is defined in TID 10041, its position in the X-Ray source reference coordinate system shall be described here. |
|
Provide the model and position of each X-Ray filter described in TID 10046, if known. |
|
Each X-Ray filter model and position description includes the identification of the X-Ray filter, which shall match the identification used in TID 10046. |
|
A 4-by-4 matrix of dimensionless numbers of the form defined in Section C.20.2.1.1 “Frame of Reference Transformation Matrix” in PS3.3 . The matrix describes the rigid transformation matrix (including translation and rotation) that transforms the reference coordinate system for the X-Ray filter to the X-Ray source reference coordinate system. The translation described by this matrix indicates the position and orientation of the X-Ray filter within the X-Ray source reference coordinate system. The RCS for the X-Ray filter shall be a right-handed Cartesian coordinate system. If the model data in Rows 11, 12, or 13 was not originally in a right-handed Cartesian coordinate system, it must transformed prior to inclusion in this template. |