DICOM PS3.16 2026c - Content Mapping Resource

TID 10049 Radiation Field Area

This template contains a description of the radiation field area for a given X-Ray source, which accounts for collimation.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10049. Radiation Field Area

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130516, DCM, "Radiation Field Area")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

TEXT

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

5

>

CONTAINS

SCOORD3D

EV (130517, DCM, "Radiation Field Outline")

1-n

M

GRAPHIC TYPE = {POLYGON, ELLIPSE}

6

>>

HAS PROPERTIES

CODE

EV (130518, DCM, "Value Timing")

1

U

DCID 10073 “Value Timing”


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the Content Items are applicable.

Row 3

The DateTime of the end of the time period over which the Content Items are applicable.

Row 4

Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR.

Row 5

The Radiation Field Outline describes the area and shape of the radiation field. The points shall be coplanar and are defined in the X-Ray source reference coordinate system.