DICOM PS3.16 2021e - Content Mapping Resource

TID 10045 Radiation Technique

This template describes the technique parameters (e.g., X-Ray tube potential, X-Ray tube current, pulse rate, etc.) of the radiation exposure during a period of time.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10045. Radiation Technique

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130511, DCM, "Radiation Technique")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

TEXT

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

5

>

CONTAINS

NUM

EV (111634, DCM, "Half Value Layer")

1

UC

XOR Row 7

UNITS = EV (mm, UCUM, "mm")

6

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

7

>

CONTAINS

TABLE

EV (111634, DCM, "Half Value Layer")

1

UC

XOR Row 5

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (111634, DCM, "Half Value Layer")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

8

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

9

>

CONTAINS

NUM

EV (113733, DCM, "KVP")

1

MC

XOR Row 11

UNITS = EV (kV, UCUM, "kV")

10

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

11

>

CONTAINS

TABLE

EV (113733, DCM, "KVP")

1

MC

XOR Row 9

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113733, DCM, "KVP")

COLUMN 2 UNITS = EV (kV, UCUM, "kV")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

12

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

13

>

CONTAINS

NUM

EV (113734, DCM, "X-Ray Tube Current")

1

MC

XOR Row 15

UNITS = EV (mA, UCUM, "mA")

14

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

15

>

CONTAINS

TABLE

EV (113734, DCM, "X-Ray Tube Current")

1

MC

XOR Row 13

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113734, DCM, "X-Ray Tube Current")

COLUMN 2 UNITS = EV (mA, UCUM, "mA")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

16

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

17

>

CONTAINS

NUM

EV (113791, DCM, "Pulse Rate")

1

UC

XOR Row 19

UNITS = EV ({pulse}/s, UCUM, "pulse/s")

18

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

19

>

CONTAINS

TABLE

EV (113791, DCM, "Pulse Rate")

1

UC

XOR Row 17

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113791, DCM, "Pulse Rate")

COLUMN 2 UNITS = EV ({pulse}/s, UCUM, "pulse/s")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

20

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

21

>

CONTAINS

NUM

EV (113793, DCM, "Pulse Width")

1

UC

XOR Row 23

UNITS = EV (ms, UCUM, "ms")

22

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”

23

>

CONTAINS

TABLE

EV (113793, DCM, "Pulse Width")

1

UC

XOR Row 21

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113793, DCM, "Pulse Width")

COLUMN 2 UNITS = EV (ms, UCUM, "ms")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

24

>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Types”


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the content items are applicable.

Row 3

The DateTime of the end of the time period over which the content items are applicable.

Row 4

Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template.

DICOM PS3.16 2021e - Content Mapping Resource